Slit Lamp Training

Tim BuckleyProduct Manager

Page 2: -1- Slit Lamp Training Tim Buckley Product Manager

-2-

BasicsDefinition and Applications

Definition

The purpose of a slit lamp is the biomicroscopy of the patient‘s eye under different lighting conditions. The slit lamp projects a bright and homogenously illuminated slit onto the eye which is variable in length, width, angle and light intensity.

Fields of Application

The primary field of application is the anterior eye segment (cornea, anterior chamber, lens, anterior vitreous).

Using additional optics enables the user to also examine the posterior eye segment as well as the anterior chamber angle.

Page 3: -1- Slit Lamp Training Tim Buckley Product Manager

-3-

BasicsDesign Principles

• components: microscope, slit projector, instrument base

• carrier arms for microscope and slit projector can be swiveled around a common axis

• swivelling axis is located in the focal plane of microscope and slit projector

Page 4: -1- Slit Lamp Training Tim Buckley Product Manager

-4-

BasicsDesign Principles - Biomicroscope

• Zeiss slit lamps:Galilei type microscope

– common front objective– parallel beam path– 3 or 5 magnification steps

• other manufacturers also offer Greenough type microscopes

– two separate, tilted beam paths

– only 2 magnification steps– very few accessories

Page 5: -1- Slit Lamp Training Tim Buckley Product Manager

-5-

BasicsDesign Principles - Slit Projector

• purpose: to project a slit image focused on the patient‘s eye

• the slit image is variable in length, width and angle

• light source: usually halogen (high color temperature)

• filters: blue, green (redfree), diffusor, heat absorbing filter

Page 6: -1- Slit Lamp Training Tim Buckley Product Manager

-6-

BasicsDesign Principles - Instrument Base

• functional coupling of carrier arms of microscope and slit projector

• slit projector and stereo microscope can independantly be swiveled around a common axis

• Axis is located below patient’s eye

• both slit image and observation are in focus in the axial plane

• three-dimensional positioning by joystick

Page 7: -1- Slit Lamp Training Tim Buckley Product Manager

-7-

Basic Functions of the Slit LampSlit Width

narrow slit circle shaped, if

fully opened

slit width is adjusted continuously

Page 8: -1- Slit Lamp Training Tim Buckley Product Manager

-8-

Basic Functions of the Slit LampSlit Length

short slit long slit

slit length is adjusted in steps and

continuously

Page 9: -1- Slit Lamp Training Tim Buckley Product Manager

-9-

Basic Functions of the Slit LampSlit Rotation

vertical slit horizontal slit vertical slit

slit rotation can be adjusted continuously by ±90°

Page 10: -1- Slit Lamp Training Tim Buckley Product Manager

-10-

Basic Functions of the Slit LampSlit Decentration

decentered slit

slit can be decentered continuously by ±4°

Page 11: -1- Slit Lamp Training Tim Buckley Product Manager

-11-

Basic Functions of the Slit LampTilting Prism

angle of incidence 0° angle of incidence 20°

tilting prism can be tilted by 0° to 20° continuously

positions 0°, 5°, 10°, 15°, 20° indexed

Page 12: -1- Slit Lamp Training Tim Buckley Product Manager

-12-

Types of Illumination

Forms of direct Illumination

Page 13: -1- Slit Lamp Training Tim Buckley Product Manager

-13-

Types of IlluminationDirect Diffuse Illumination

Principle• illumination of the eye with a

broad, unfocused light beam• usage of diffusor• microscope positioned at 0°• magnification 5x ... 12x

Applications• Overview• general assessment of anterior

eye, eye lids• assessment of contact lenses

Page 14: -1- Slit Lamp Training Tim Buckley Product Manager

-14-

Types of IlluminationDirect Diffuse Illumination

IOL

iris supported anterior chamber lens in

diffuse illumination

Bildquelle: Universitäts-Augenklinik Jena

Page 15: -1- Slit Lamp Training Tim Buckley Product Manager

-15-

Types of IlluminationDirect Focal Illumination - Optic Section

Principle• Illumination and observation are

focused in the same plane• slit width ca. 0,1 to 0,3mm

Applications• mainly findings in the cornea and

lens• opacities, scars, vessels• good perception of the depth of

findings

Page 16: -1- Slit Lamp Training Tim Buckley Product Manager

-16-

Types of IlluminationDirect Focal Illumination - Optic Section

Cataract

anterior cortical opacities, nucleosclerosis

and posterior opacities

Bildquelle: www.atlasophthalmology.com

Page 17: -1- Slit Lamp Training Tim Buckley Product Manager

-17-

Types of IlluminationDirect Focal Illumination - Optic Disc

Principle• Illumination and observation are

focused in the same plane• slit width ca. 2 to 4mm

Applications• mainly findings in the cornea and

lens• opacities, scars, vessels• good perception of shape and size

of findings

Page 18: -1- Slit Lamp Training Tim Buckley Product Manager

-18-

Types of IlluminationDirect Focal Illumination - Optic Disc

Cyst on Pupillary Edge

Cyst on pupillary edge stems from usage

of too strong miotica

Bildquelle: www.atlasophthalmology.com

Page 19: -1- Slit Lamp Training Tim Buckley Product Manager

-19-

Types of IlluminationDirect Focal Illumination - Conical Beam

Principle• assessment of particles floating in

the anterior chamber by illuminating with a light beam

• Tyndall‘s phenomenon• pinpoint illumination 0,3 - 0,5mm

Applications• assessment of particles in

aqueaous humor• inflammation cells, pigmented

cells, metabolic waste

Page 20: -1- Slit Lamp Training Tim Buckley Product Manager

-20-

Types of IlluminationDirect Focal Illumination - Conical Beam

cells in anterior chamber

cells in anterior chamber as a sign of

uveitis

Bildquelle: www.atlasophthalmology.com

Page 21: -1- Slit Lamp Training Tim Buckley Product Manager

-21-

Types of IlluminationTangential Illumination

Principle• a narrow light beam is projected

almost parallel along the structure to be observed

• elevated structures are visible by shadowing

Applications• elevated abnormities or changes in

the iris• tumors, cysts

Page 22: -1- Slit Lamp Training Tim Buckley Product Manager

-22-

Types of IlluminationTangential Illumination

Iris

Iris in tangential illumination

Page 23: -1- Slit Lamp Training Tim Buckley Product Manager

-23-

α α

Page 24: -1- Slit Lamp Training Tim Buckley Product Manager

-24-

Types of IlluminationSpecular Illumination

Principle• angle of incidence = angle of

reflection• observation and illumination have

same angle to perpendicular axis• slit width < 4mm

Applications• assessment of surfaces • assessment of tear film• endothelial cell layer

α α

Page 25: -1- Slit Lamp Training Tim Buckley Product Manager

-25-

Types of IlluminationSpecular Illumination

endothelial cells

endothelial cell layer magnified ca. 192x

α α

Bildquelle: Carl Zeiss Meditec

Page 26: -1- Slit Lamp Training Tim Buckley Product Manager

-26-

Types of Illumination

Forms of indirect Illumination

Page 27: -1- Slit Lamp Training Tim Buckley Product Manager

-27-

Types of IlluminationIndirect focal Illumination

Principle• illumination by stray light• slit is slightly decentered so that

stray light is created in direct neighbourhood of the finding

• slit width ca. 2 to 4mm

Applications• mainly corneal lesions and scars

Page 28: -1- Slit Lamp Training Tim Buckley Product Manager

-28-

Types of IlluminationIndirect focal Illumination

no example

Page 29: -1- Slit Lamp Training Tim Buckley Product Manager

-29-

Types of IlluminationDirect Retro-Illumination from the Iris

Principle• Illumination of the finding with

indirect light rflected from the iris• observation with light background• medium slit width, ca. 2 to 4mm

Applications• Infiltrations, small scars, corneal

vessels, micro cysts, vacuoles• with this illumination findings are

made visible with high contrast

Page 30: -1- Slit Lamp Training Tim Buckley Product Manager

-30-

Types of IlluminationDirect Retro-Illumination from the Iris

Keratitis Superficialis Punctata

finding after moderate cauterization by

acid, defects of epithelium and conjunctiva

have been stained with bengal rose

Bildquelle: www.atlasophthalmology.com

Page 31: -1- Slit Lamp Training Tim Buckley Product Manager

-31-

Types of IlluminationIndirect Retro-Illumination from the Iris

Principle• Illumination of the finding with

indirect light reflected from the iris• observation with dark background• medium slit width, ca. 2 to 4mm

Applications• Infiltrations, small scars, corneal

vessels, micro cysts, vacuoles

Page 32: -1- Slit Lamp Training Tim Buckley Product Manager

-32-

Types of IlluminationIndirect Retro-Illumination from the Iris

Keratitis Punctata, contact lens wearer

multiple erosions of the central cornea due

to inappropriate contact lens fittingBildquelle: www.atlasophthalmology.com

Page 33: -1- Slit Lamp Training Tim Buckley Product Manager

-33-

Types of IlluminationRetro-Illumination from the Lens

Principle• Illumination of the finding with

indirect light reflected from the lens• observation with light background• medium slit width, ca. 2 to 4mm

Applications• corneal defects, foreign bodies,

scars• (type of illumination not frequently

used)

Page 34: -1- Slit Lamp Training Tim Buckley Product Manager

-34-

Types of IlluminationRetro-Illumination from the Lens

no example

Page 35: -1- Slit Lamp Training Tim Buckley Product Manager

-35-

Types of IlluminationRetro-Illumination from the Fundus

Principle• Illumination of the finding with

indirect light reflected from the fundus

• observation with red/yellowish background

• dilated pupil

Applications• abnormities in the anterior

vitreous, lens, anterior chamber, cornea

• findings are visible like silhouettes

Page 36: -1- Slit Lamp Training Tim Buckley Product Manager

-36-

Types of IlluminationRetro-Illumination from the Fundus

Aniridia

missing iris and zonular cataract made

visible by retro-iluminationBildquelle: www.atlasophthalmology.com

Page 37: -1- Slit Lamp Training Tim Buckley Product Manager

-37-

Types of IlluminationIris-Transillumination

Principle• transillumination of the iris by

indirect light reflected from the fundus

• half dilated pupil (3 to 4mm)• Illumination and observation at ca.

Applications• Visualization of defects of the

pigment layer of the iris

Page 38: -1- Slit Lamp Training Tim Buckley Product Manager

-38-

Types of IlluminationIris-Transillumination

Albinism

Iris-Transillumination shows the light

transmission of the irisBildquelle: www.atlasophthalmology.com

Page 39: -1- Slit Lamp Training Tim Buckley Product Manager

-39-

Types of IlluminationSclerotic Scatters

Principle• Illumination of the limbus region

with a broad light beam at an angle of 45° - 60°, decentered slit

• total reflection of the incoming light at inner corneal boundaries (endothelium and epithelium)

Applications• scars, foreign bodies, corneal

defects• irregularities in the cornea cause

straylight

Page 40: -1- Slit Lamp Training Tim Buckley Product Manager

-40-

Types of IlluminationSclerotic Scatters

corneal scar

corneal scarring after infection

Bildquelle: www.atlasophthalmology.com

Page 41: -1- Slit Lamp Training Tim Buckley Product Manager

-41-

Fundus Observation and Gonioscopy

Page 42: -1- Slit Lamp Training Tim Buckley Product Manager

-42-

Fundus Observation and GonioscopyContact Glasses

Contact Glasses

• additional tool for fundus observation with the slit lamp

• mostly direct: erect and non mirrored image of the fundus

• required: dilated pupil, use of gliding liquid

Bildquelle: www.ocular-instruments.com

Microscope

Fundus Image

Page 43: -1- Slit Lamp Training Tim Buckley Product Manager

-43-

Fundus Observation and GonioscopyLenses

Lenses

• additional tool for fundus observation with the slit lamp

• mostly indirect: upside-down and mirrored image of the fundus (convex optics)

• non contact

• required: dilated pupil

Bildquelle: www.ocular-instruments.com

microscope

Inverted fundus image

Page 44: -1- Slit Lamp Training Tim Buckley Product Manager

-44-

Fundus Observation and GonioscopyGonioscopy

Three mirror contact glass

• Goldmann contact glass

• central lens: posterior pole

• 73° mirror: equator

• 67° mirror: ora serrata

• 59° mirror: anterior chamber angle

Bildquelle: www.ocular-instruments.com

central lens

Page 45: -1- Slit Lamp Training Tim Buckley Product Manager

-45-

Fundus Observation and GonioscopyGonioscopy

Three mirror contact glass

• Goldmann contact glass

• central lens: posterior pole

• 73° mirror: equator

• 67° mirror: ora serrata

• 59° mirror: anterior chamber angle

Bildquelle: www.ocular-instruments.com

73° mirror

Page 46: -1- Slit Lamp Training Tim Buckley Product Manager

-46-

Fundus Observation and GonioscopyGonioscopy

Three mirror contact glass

• Goldmann contact glass

• central lens: posterior pole

• 73° mirror: equator

• 67° mirror: ora serrata

• 59° mirror: anterior chamber angle

Bildquelle: www.ocular-instruments.com

67° mirror

Page 47: -1- Slit Lamp Training Tim Buckley Product Manager

-47-

Fundus Observation and GonioscopyGonioscopy

Three mirror contact glass

• Goldmann contact glass

• central lens: posterior pole

• 73° mirror: equator

• 67° mirror: ora serrata

• 59° mirror: anterior chamber angle

Bildquelle: www.ocular-instruments.com

59° mirror

Page 48: -1- Slit Lamp Training Tim Buckley Product Manager

-48-

Fundus Observation and GonioscopyExample: Fundus

retinal scar

Bildquelle: UAK Jena / Carl Zeiss

microscope

Page 49: -1- Slit Lamp Training Tim Buckley Product Manager

-49-

Fundus Observation and GonioscopyExample: Anterior Chamber Angle

blood in chamber angle

vessels in chamber angle

Page 50: -1- Slit Lamp Training Tim Buckley Product Manager

-50-

Examination using Fluorescein

Page 51: -1- Slit Lamp Training Tim Buckley Product Manager

-51-

Examination using Fluorescein

Principle• Fluorescein is inserted into the

conjunctival sac and fills, for example, intracellular spaces

• dye is excited with blue light(λ 450 ... 500 nm)

• contrast reducing straylight is blocked with barrier filter (yellow filter λ > 530 nm)

Applications• corneal lesions / defects• contact lens fitting

Page 52: -1- Slit Lamp Training Tim Buckley Product Manager

-52-

Applanation Tonometry

Page 53: -1- Slit Lamp Training Tim Buckley Product Manager

-53-

Applanation Tonometry

Principle (according to Prof. Goldmann)

• diameter of the applanated area: 3,06 mm

• counteracting forces of corneal rigidity and capillary attraction cancel each other out

• applanated force (AF) equals the intraocular pressure (IOP)

Applications• measurement of intraocular

pressure

IOP AF

Page 54: -1- Slit Lamp Training Tim Buckley Product Manager

-54-

Applanation Tonometry

measuring figures

a: applanation force

too low

b: correct setting

c: applanation force

too high

Page 55: -1- Slit Lamp Training Tim Buckley Product Manager

-55-

Length and Angle Measurement

Page 56: -1- Slit Lamp Training Tim Buckley Product Manager

-56-

Längen- und Winkelmessung

Length and Angle Measurement

• Diameter of cornea and pupil

• height of conjunctival folds

• anterior chamber depth

• toric contact lenses

Page 57: -1- Slit Lamp Training Tim Buckley Product Manager

-57-

Redfree Examination

Page 58: -1- Slit Lamp Training Tim Buckley Product Manager

-58-

Redfree Examination

• Objects with high portion of red (e.g. fundus, vessels)

• contrast enhancement through green filter