Bench TopInductively Coupled Plasma Optical Emission Spectroscopy
-
7/27/2019 01099401 Quantima Brochure
2/16
GBC has alwaysplaced a strongemphasis on qualityin all aspects of ouroperation, from design and
manufacture to the provisionof service and support to ourcustomers, and we are fullycommitted to continuousevaluation and improvementin all areas.
The GBC QualityManagement System hasbeen accredited to the ISO9001 quality standard byLloyds Register Quality
Assurance Limited. Thiscertification is your assurancethat the procedures andprocesses used to producethe goods and services whichGBC provides comply withthe relevant InternationalStandard, and demonstratescommitment to meeting theneeds and expectations ofour customers.
For over 30 years GBC has
been at the forefront ofscientific technologicaldevelopment, manufacturingand marketing a wide rangeof award winning, qualityscientific instruments.
GBCs product lines...
GBC Scientific Equipment
aawill advance peoples knowledge and
their capacity to enhance the quality of life
for all humankind.
HPLC ICP-OESAAS ICP-oTOFMS Rheometry UV-Vis XRD
Sensitive Technologyfor a Sensitive World
ISO 9001Quality
Accreditation
-
7/27/2019 01099401 Quantima Brochure
3/16
Excellence in Design
The 4th GenerationQuantima represents aquantum leap in technologyin ICP-OES packed with lots
of great features to give youthe technological edge!
3
The Quantima is at the forefrontof a long tradition of excellencein design and engineering.
It is the culmination of feedbackon a worldwide basis from bothdistributors and end users ofICP-OES. The Quantima is theresult of team development ofthe highest interactivity. Itsrobustness is only supersededby its reliablity andcompactness.
It is simple to operate, maintainand service. This instrument
raises the standards for trulyaffordable ICP.
The Quantima offers manynew innovative features.Plasma CamTM and PIPTM
ensures that the Quantimawill update the technologyavailable in your laboratory inone quantum leap.
Low Operating CostLower Argon Consumption
The Quantima uses less than11L/min for a typical aqueoussample. This is the lowest Argonconsumption of any ICP-OESavailable. With an 11% increasein monochromator speed the
Quantima is both faster andconsumes less Argon.
Improved Torch IgnitionSequence
The Quantimas torch IgnitionSequence has been greatlyimproved. The New IIS IntelligentIgnition StartupTM will vary theplasma startup sequence routinedepending on how long theplasma has been off. Thisenables quicker startup routinefor a plasma recently turned off.
Compact Benchtop Design
No matter what yourapplication, the Quantimawill analyse faster, moreaccurately and more cost
effectively than ever before!
Quantima is also available with an optional bench
-
7/27/2019 01099401 Quantima Brochure
4/16
Outstanding Features equal Outstanding Performance
The Quantima is suitable for adiverse range of applications.Method development andanalysis is easy. First timeusers can easily obtain
optimal performanceregardless of the application.The Quantimas outstandingfeatures providecomprehensive solutions foryour elemental analysis.
Routine or Research Analysis Applications
4
Computer control of over 230instrument parameters.
Precise mass flow control ofnebuliser, auxiliary* and
plasma* gases.
The Power and Durability ofa free-running 40.68 MHzRF generator which offersthe highest efficiency ofenergy transfer to your
sample.
Thermostatted Optics forunparalleled optical stability.
Choice of gratings* to offerbetter resolution/sensitivitydepending on yourapplication.
Capability to handle difficultmatrices such as 40%
dissolved salts using the
humidifier* - eliminating theneed for dilution and enhancingtrace level detectability.
Powerful Multi-tasking Windows
32 Bit software.
Sophisticated auto-optimisationsoftware.
Powerful semi quant facility.
Plasma Cam* enables userto view plasma from PC.
PIP* - Plasma IntegrityProtector significantlydecreases chances of a"doughnut" and associatedmelted torches.
Inbuilt chiller means no bulkyextra chillers required.
Nitrogen Purged optics -vacuum capability withoutmessy regular maintenance ofvacuum pumps.
Ability to directly analyse
organics without any oxygenaccessories.
With all of these powerfulfeatures, even a first time usercan immediately obtain themaximum performance from theQuantima regardless of theapplication.
The Quantima offers ease of usefor method development and
analysis which is unmatched byany other ICP.
* optional extra
-
7/27/2019 01099401 Quantima Brochure
5/16
High Performance Optics Technology
The Quantima optics havebeen designed to satisfy allthe fundamental figures-of-merit critical for spectroscopicinstrumentation
Unlimited Wavelengthselectivity
High Sensitivity
Highest resolution
Unparalleled Optical Stability
High Stray Light Rejection
Auto-optimisation of viewing
position
Optimised dual detector
response Widest dynamic range
High Productivity
Unmatched flexibility
The no compromise designphilosophy ensures that you canhave full confidence in theintegrity of data, regardless ofthe application.
Largest range of selectable
wavelengths
With its comprehensivewavelength database, theQuantima makes wavelengthselection simple. Relativesensitivities and all potentialspectral interferences aredisplayed. Choose the mostsensitive wavelength for tracedeterminations or less sensitivewavelengths for high
concentrations. You can choosethe best possible interference-free wavelengths when thecomplexity of the spectrumgenerated by your sample matrixdemands it.
The standard Quantimaconfiguration will cover thespectral range of all ICPdetectable elements from S, P, B,Hg or Al (Ultra Violet region) right
up to Na, Li, Cl and K (Visibleregion).
Maximum energy throughputfor high sensitivity
By using only efficient all-reflective optics and minimaloptical components light loss isminimised and highest sensitivityis assured.
Highest resolving power to0.004 nm
With a choice of specializedgratings you can choose whetheryou require wide wavelengthcoverage or high sensitivity and
An example of how the phosphorous line at 177.495 nm can be fullyresolved from a copper interference at 177.482 nm using 0.004 nmresolution
high resolution to satisfy your ICPapplication requirements. Theoptional 2,400 l/mm holographicgrating allows resolution as highas 0.004 nm allowing the
Quantima the resolution toperform measurements directly oninterference peaks even in themost demanding matrices. Thehigh resolution eliminates theneed for complex and timeconsuming mathematical peakdeconvolution software that canonly provide poor approximationsat separating these interferences.
Largest Range of Selectable Wavelengths
5
-
7/27/2019 01099401 Quantima Brochure
6/16
Outstanding Long-term Precision
6
Quantima - Mass flowcontrollers provideprecision, accuracy andstability
Innovative GBC engineeredmass flow controllers, fitted tothe nebuliser line andoptionally to each of theplasma and auxiliary gas lines,guarantees the highest level ofprecision and control. Thisunique feature ensuresprecision, accuracy andreproducibility of signalmeasurement, essential sinceanalyte intensities are highly
sensitive to fluctuations in gasflows.
With gas flows reproduciblyset to within 0.001 L/mintogether with the highprecision concentric nebuliserand efficient cyclonic spraychamber, the Quantimaroutinely achieves analyticalprecision better than 0.5%.
Total Argon flow is typically
less than 11 L/min foraqueous samples makingthe Quantima the lowestargon consumption ICP thatis available.
Excellent Long-termStability
Direct comparison withconventional gas controlsystems proves that mass flow
controllers achieve the highestdegree of gas flow controlwhich results in the bestpossible precision, long-termstability and outstandingreproducibility.
On screen monitoring of critical parameters
With typically less than 1% driftover a four hour period, theanalyst can spend more timeanalysing new samples andless time repeating calibrationstandards.
Unparalleled OpticalStability
The highest possible analyticalaccuracy and precision areachieved through the precisetemperature control of the0.75 m Czerny-Turnermonochromator. Theelimination of drift caused byvariations in ambienttemperature ensures theintegrity of analytical data overlong and unattended analyses.
Mass flow controllers provide accuracy and stability
Exceptional stability and precisionachieved using mass flow controlcompared to conventional gas control
With sample gasflow optimised foreach element,
highest sensitivitiescan be achieved
with differentmatrices
-
7/27/2019 01099401 Quantima Brochure
7/16
Flexible and Robust Plasma
7
Flexibility allows best sensitivity regardless of element or matrix
Optimised dual detectorsystem
The Quantima's dual detectorsystem ensures maximum
sensitivity across the entirewavelength range. One detectoris sensitive below 300 nm andthe second detector for elementsabove 300 nm. The choice ofdetector is automatic.
Widest linear dynamic range -8 orders of magnitude
Eight orders of linear responseachieved by the dual detector
system permits analyticaldeterminations over the widestconcentration ranges.Concentrations from high percentto low parts per billion levels canbe confidently measured withoutdilution.
Flexible parameter settingsallows best conditions forevery element
For ICP determinations everyelement requires slightly differentnebuliser flow and viewing heightto obtain the optimum sensitivity.These critical parameters can beaccurately set and reproduciblyrecalled wherever needed.
Robust and reliable highefficiency plasma module
Crucial to the Quantima's powerand performance is a robust,reliable and efficient 40.68 MHzfree-running 1500 watt RFgenerator. Utilising the latest inhighly accurate fuzzy logiccontrollers,
the GBC designed RF systemprovides superb stability whichtranslates to unsurpassedanalytical performance. Thisfrequency also offers reduced
background levels for enhancedsignal to noise response.
With power settings controlled inprecise 1 W steps, the plasmaoperating power can be easilyconfigured even for the mostdifficult sample matrices witheach element individuallyoptimised.
The Highest Energy Transfer
from Plasma to SampleThe Mg II 280.270 nm/Mg I285.213 nm Intensity Ratioprovides a true measure of theefficiency of energy transfer fromplasma to sample.
Graphs showing effectthat varying viewingheight and nebuliserflow have on variouselements.
Professor J. M. Mermetdetermined that values greaterthan 10 are a measure ofsuperior performance. TheQuantima's efficient RF generator
and precise control of plasmaparameters all combine toprovide the highest possibleenergy transfer from plasma tosample. With values typicallyexceeding 15, the Quantima hasthe most robust plasma availableensuring maximum sensitivity andminimised matrix effect especiallywith difficult organic matrices. Aswell as better sensitivity inaqueous samples, the mostvolatile organic solvents can bedirectly aspirated without theneed to dilute or use of oxygenaccessories.
-
7/27/2019 01099401 Quantima Brochure
8/16
8
The traditional method of determiningthe nutritional composition of milk is toash it at high temperatures to removethe organic matter followed up by anacid digest. This process can take up to
a day to perform.
As well as time costs involved and needfor costly equipment and chemicalsthere is a possibility of either losingvolatile elements in the ashing stage orthe possibility of contamination in thedigestion stage.
As an alternative methodology, GBCchemists devised a 3% slurry techniquefor milk analysis.
This much simplified technique simplyrequires weighing a milk powder thenadding distilled/deionised water anddirectly analysing.
Direct, Accurate and Fast
Application - Direct analysis of Milk Powder using a Slurry Method
Element Wavelength Standard 1 Standard 2
Na 589.592 10.00 50.00
K 769.896 25.00 100.0
Ca 422.673 25.00 100.0
Mg 280.270 2.00 10.00
Fe 238.204 0.20 2.00
Zn 213.856 0.20 1.00
Cu 324.754 100.0 500.0
Mn 259.373 20.00 50.00
P 177.495 15.00 30.00
Element Nestle ReferencePowder ExpectedResult (mg/100g)
Ashed Samples(mg/100g)
3% SlurryResults
(mg/100g)
Na 116.9 - 126.4 115, 115.8 120, 120
K 588 - 710 680, 670 672,675
Ca 273.3 - 284.8 278.3, 276.7 279.3, 278.9
Fe 24.7 - 26.7 25.1, 25.2 25.1, 25.3
Mg 3.9 - 4.15 3.92, 3.95 3.93, 3.95
Zn 1.50 - 1.65 1.52, 1.52 1.62, 1.62
Cu 0.125 - 0.150 0.128, 0.127 0.142, 0.142
Mn 30.6 - 50.1 40.9, 41.8 40.8, 40.9
P 215.8 - 224.3 226.8, 227.3 223.0, 222.8
The Quantimas spacious sample introduction areaallows convenient access to all sample introductioncomponents. With quick release gas connectionsand hoses, changeover is rapid. The new Cool
Touch design ensures that the torch box never getshot, even when running at maximum power forhours, thus allowing physical removal quickly withouthaving to wait for components to cool.
Power (w) 1200
Plasma Gas (L/min) 11.00
Auxiliary Gas (L/min) 0.50
Sample Gas (L/min) 0.50
Sample Gas (KPa) 280
Torch Height (mm) 9.0
Pump Speed (mL/min) 1.2Pump Speed (rpm) 15
PMT(V) 500
Scan Window 0.08nm 1st order 0.04nm 2nd order0.027nm 3rd order0.02nm 4th order
Data Points 40
Integration Time Per Point (s) 0.1
On Peak Integration (s) 2
Replicates 3
Background Correction Mode DynamicFast Pump between Sample (s) 15
Stabilization (s) 20
-
7/27/2019 01099401 Quantima Brochure
9/16
The periodical analysis of engineoil is an important application asfleet companies use this analysisto determine when major servicingis required. An increase in metal
concentration in sump oil canpinpoint potential engine trouble.The presence of tin may becaused by bearing wear. Thepresence of chromium, aluminiumor nickel may be because ofpiston or ring wear. Copper isindicative of coolant leaking intothe lubrication system via cracksor poor seals and silicon maysuggest dirt or sand intrusionindicating that the air filter needs
to be cleaned or replaced.This analysis is best performed byICP due to its speed and theability to analyse elements such asS and P.
The oil samples andorganometallic standards aredissolved in kerosene and directlyaspirated into the plasma withoutthe need for Oxygen Accessoriesrequired by conventional ICPs to
avoid carbon build up. TheQuantima can perform thisfunction without an accessory andin addition a pump speed of 60rpm can be used pumpingkerosene between samples tospeed up sample introductionwithout the plasma turning off.
Fast Sample Introduction
9
Viewing Height Individually Optimised 3-20 nmPower 1300 wNebuliser FlowOptimised for Pb 0.4 L/min (160 kPa)Plasma Gas Flow 12 L/min
Auxiliary Gas Flow 1.0 L/minSample Uptake Rate 1.0 mL/min
8 rpm
PMT Voltageindividually optimised from 340 - 600 Vdepending on concentration level
Measurement Conditions for Sample AnalysisMeasurement
Integration TimeOn Peak (s) 2Replicates (s) 3Background Correction DynamicScan WIndows 0.08 nm 1 st Order
0.04 nm 2nd Order0.027 nm 3rd Order0.02 nm 4th Order
Data Points 50Integration Time per Data Point (s) 0.1
Wear Metal Oil Analysis
ElementI - Atomic LineII - Ionic Line
Ag IAl
As lBiBA IICa IICd IICr IICu IFeHg IMg IMn IIMo IINa INi IIP IPb II
Wavelength(nm)
328.068167.081
188.979249.773455.403317.933214.438283.563324.754259.94184.95279.553257.610202.030588.95231.064177.495220.353
Detection Limit(ppb)
2.02.0
10.02.00.12.00.82.02.01.58.00.10.35.010.03.08.010.0
-
7/27/2019 01099401 Quantima Brochure
10/16
Accessories To Enhance Productivity
10
An argon humidifier is availableas an option to the Quantima.When fitted in series with thesample gas line and thestandard nebuliser, the argon
humidifier increases thetolerance of the sampleintroduction system to blockagesdue to high dissolved salts. Thisis achieved by bubbling argonthrough water to add a smallamount of humidity to the argongas supply. As the water vapourpasses through the nebuliserorifice it removes salt deposits asthey occur.
Plasma Integrity Protector -PIPTM significantly reduces theoccurrence of molten torches.Occasionally, under certainconditions, the plasma can
collapse and form what iscalled a parasitic plasma ordoughnut. Here the plasmacollapses into the top of thetorch and if the power to theRF generator is shut nodamage occurs. However, if thedoughnut is left, such as wouldhappen in an unattendedanalysis, the torch would meltwithin a few seconds. GBC hasdevised and patented the PIPwhich is an option on theQuantima. The PIPTM
immediately shuts power to theRF Generator as soon as itdetects a change from anormal plasma to a parasiticplasma. The occurrence ofmolten torches are significantlyreduced.
ICP users often look at thePlasma to determine quickly
whether either the sample hasreached the plasma or whetherthe rinse solution has flushed itout. With an optional PlasmaCamTM, the end user can now
view this and more from theconvenience of the computerscreen.
Argon Humidifier allows up to 40% dissolved solids
The optional Argon Humidifier
30% NaCl solution excellent reproducibility over 4 hours usingstandard sample introduction components and Argon Humidifier.
Humidifier Software Icon
-
7/27/2019 01099401 Quantima Brochure
11/16
Best Performance - every application with automatic optimisation
Powerful, fast, accurate auto-optimisation software
For less experienced users, theQuantima software provides an
intelligent auto-optimisationsoftware package that simplifiesthe complex task of setting upinstrument operating parametersoptimised for individualapplications. Now there is noneed to understand the complexityof how all of the operatingparameters interact with eachother. There is no need to predictthe effect on sensitivity on aparticular wavelength as theoperating conditions are changed.Quantimas auto-optimisationsoftware will perform all themeasurements and calculationsaccording to your selectedoptimisation criteria. A Simplex isa mathematical procedure fordetermining optimum values for arange of inter-dependentparameters. GBC has modifiedthe traditional fixed stepmultivariate Simplex to usevariable step size as opposed to
the traditional fixed step size. Thisresults is a true optimisation valuebeing obtained more quickly andmore precisely and accurately.
User-selectable optimisationcriteria
Quantima software achievesoptimisation with any numberof variables according to user-selected criteria. (MaximumIntensity or SBR or minimumBEC or DL.) Criticalparameters such as Nebuliserflow, RF power and viewingheight, for example, each playan important role in achieving
the highest possible analyticalprecision, accuracy andsensitivity. The Quantimasoftware takes the guessworkout of setting parameters andturns inexperienced operatorsinto expert analysts.
Auto-Optimise Methods
A variable step size simple algorithm allows faster, more precise andaccurate optimisation compared to a fixed step
11
Actual auto-optimisation for 3different elements. Note that theoptimum torch viewing height andnebuliser flows are different.
-
7/27/2019 01099401 Quantima Brochure
12/16
12
True Multi-Tasking
User Friendly, Powerful Software
Ease of Use
The Quantima software, designedfor Windows, is logicallyorganized into clear modules for
simplified control of all instrumentand analysis functions. Designedfor ease of use and flexibility, thesoftware uses a combination ofeasily recognizable icons, functionkey controls, simple to edit menusand a notebook style layout tokeep you in control andorganized.
With User Name and passwordprotection your method and dataare now safe from unauthorizedusers.
To start an analysis can be as simple as a one click operation on the start ICON.While performing sample analysis, old data can be reviewed and printed, user definedcritical instruments status panels can be displayed and data can be sent to third partysoftware packages. With context sensitive Help only a click away and error-sensitiveprompts ready to respond, the truly multi-tasking software is easy to use.
-
7/27/2019 01099401 Quantima Brochure
13/16
Specifications
13
Data for Future Reference
Method Development MadeEasy
Method parameters are logicallygrouped on individual pages
and accessed via appropriatelynamed tabs on the side of eachpage. All of the developmenttools required are located in theMethod module.
Powerful Graphics Module
The powerful graphics moduleis used to display and overlaywavelength scans forinvestigating potentialinterferences, definingbackground correction positionsor for determining the effect onintensity caused by changinggas flows, viewing height,power, wavelength order ormany other operatingparameters. You can also usethe chart recorder facility toinvestigate performance trendswhen optimising, to monitor thestability of the instrument orspecific parameters, or to
determine spray chamberwashout times so that you canaccurately set and store theappropriate rinse andstabilization times.
Comprehensive WavelengthLibrary
Direct access to the built inwavelength library providesinformation on relative line
sensitivity, backgroundequivalent concentrations andpotential interferences about32,000 wavelengths.
With graphical display ofpotential spectral interferencesavailable at all times,wavelength selection becomesvery easy, even for complex
matrices.
Linking multiple applications maximises the potential of unattendedoperation
User customisable status panels. User can choosefrom over 200 parameters and a range of displayor output models.
In the unlikely situation wherean analyte has an interferenceand another wavelengthcannot be used the Quantimasoftware provides interference
corrections to be used.
-
7/27/2019 01099401 Quantima Brochure
14/16
Extend the Quantimas Capabilities
14
Accessories
SDS720 Autosampler
The SDS720 is a precisionengineered X-Y-Z autosampler.The SDS720 provides accurate
and fast analysis due to itsdurable, simple, adaptable,reliable and sturdy design.Supplied with sample racks tohold 240 sample vials each ofapproximately 14 mL and astandards rack to hold 10standard vials each with avolume of approximately 50mL. Plastic sample andstandard vials are supplied.PTFE and PEEK is used to
provide a metal free liquid flowpath. Variable continuous flowsample probe rinse station withperistaltic pump minimisessample contamination andcarryover. Software controlsinclude rinse time, delay time,number of replicates, rescalerate, recalibration rate,measurement time and analysisorder. Full random accesscapability is standard. Up to360 samples can be loaded
using 7 mL tubes and a further360 samples using theoptional rack extension givinga total of 720 samples.Optional clean enclosure alsoavailable.
HG3000PII Hydride Generator
Use the HG3000PII hydridegenerator for the determination ofsub parts per billion of hydride
forming elements such as As, Hg,Sb, Se, Pb, Bi, Sn, Ge and Te. TheHG3000PII hydride generatoraccessory incorporates precisionglassware for highly efficient mixingof reactants and gas liquidseparation to ensure reproducibilityand high sensitivity. This easy touse accessory takes less than aminute to interchange from theQuantima standard set of sampleintroduction components and istotally made from inertpolypropylene material for longlasting performance, essential inhighly corrosive environments.
Programmable TemperatureSpray Chamber
The programmable temperaturespray chamber (PTSC) features
electronically controlledtemperature using an inbuiltpeltier device.
The PTSC provides all the benefitsof a temperature controlled ICPsample introduction system in acompact, convenient, costeffective package. Anytemperature between -10 and+60C can be selected to achieveoptimum conditions for any
application.
For maximum convenience thePTSC can be controlled from aPC either through suppliedBluetooth interface or suppliedUSB interface. The cyclonic spraychamber is encapsulated in aspecial heat conducting polymerwhich provides excellent sensitivityand precision with fast washout.
HF-resistant samplingaccessories
For samples digested inhydrofluoric acid, apolypropylene spray chamber,
used in conjunction with aceramic nebuliser and torchsample tube, provides an inertenvironment for sampleintroduction. This system can alsobe used for aqueous sampleswithout compromise.
-
7/27/2019 01099401 Quantima Brochure
15/16
Rapid Service Response
GBC has many factory trainedservice representatives world-wide so you can be assured of
a rapid response to yourservice requirements.
Remote Diagnostics
Easy to install, user friendlysoftware enables GBC toprovide you with completeon-line remote instrumentdiagnostics and troubleshooting. With over 200instrument status items, the
user can create customisedstatus panels to monitor over200 parameters. By using thelog to file facility, anyparameters can be logged toa file and emailed to GBC.
Enhance Productivity and Sensitivity
15
The temperature can be setas low as -10 and +60 canbe selected to achieveoptimum conditions for anyapplication.
By holding the spraychamber at a stabletemperature, the long termstability is improved.
The sensitivity of manyanalyses is enhanced byrunning the spray chamberat elevated temperatures,particularly important for
samples with limitedvolume.
Ultrasonic Nebuliser
The combination of highnebulisation efficiency andsubsequent desolvation by theultrasonic nebuliser results insensitivities typically 5 to 20fold greater than thoseachieved by conventional
pneumatic type nebulisers.
Setup is rapid, taking lessthan one minute tointerchange from theQuantimas standard set ofsample introduction
components and operation issimplified with automatictemperature control of theheating and coolingfunctions. The compactdesign allows you toconveniently place theultrasonic nebuliser onQuantimas standard built-inbench, ensuring optimum useof your laboratory space.
An ultrasonic nebuliser is alsoavailable for use with organicsamples.
GBC global agents
-
7/27/2019 01099401 Quantima Brochure
16/16
Designed and manufactured byGBC Scientific Equipment Pty Ltd
A.C.N. 005 472 686
GBC reserves the right tochange specifications withoutprior notice.
GBC publication number01-0994-01
August 2011 Australia
GBC SCIENTIFIC EQUIPMENT
Manufacturer of world-classscientific instruments andaccessories: AAS, HPLC, ICP-OES, ICP-oTOFMS, Rheometry,UV-Vis and XRD.
2-4 Lakewood BoulevardBraeside, Victoria 3195
Australia
Telephone 61 3 9588 6666
Facsimile 61 3 9588 6677
email [email protected] www.gbcsci.com
All trade-marks and trade-namesare the property of theirrespective owners.
Quantima SpecificationsPart Number: 99-0552-00
RF GeneratorFree running, air cooled generator. Auto start from switch or keyboard.Operating Frequency 40.68 MHz.
Grating1,800 l/mm holographic grating allows wide analysis range from 160 - 800 nm and up to6 pm resolution.
Optional 2400 l/mm grating allows analysis range from 160 - 640 nm with up to 4 pm resolution.
Mass Flow ControlNebuliser Gas flow regulated using mass flow controller.Optional mass flow control available on Plasma and Auxiliary flows.
RF PowerComputer controlled from 500 W to 1,500 W with auto tuning.
Safety InterlocksInterlocks on plasma compartment door, gas pressure, low/high mains voltage, plasma failure.
EMC Standards
Meets the following EMC standards:EN61326-1:2006.
Operating Voltage200-240 V AC 10%, 3.5 kVA, 20A.
Frequency50/60 Hz
Dimensions1520 mm (W) x 850 mm (H) x 730 mm (D)
Weight
245 kg
Accessories
99-0582-00 SDS720 Autosampler99-0126-03 Hydride Generator40-0254-00 Programmable Temperature Spray Chamber99-0322-00 Ultrasonic Nebuliser, 220 V99-0323-00 Ultrasonic Nebuliser, 110 V99-0566-00 Ultrasonic Nebuliser with Desolvator, 220 V99-0567-00 Ultrasonic Nebuliser with Desolvator, 110 V97-3764-00 Humidifier97-3718-00 Plasma Cam
97-3719-00 PIP40-0182-00 2,400 l/mm High Resolution/Sensitivity Grating97-3750-00 3 Channel Mass Flow Controller97-1929-00 4 Channel Peristaltic Pump97-3753-00 HF Sample Introduction Kit95-0035-00 Organic Samples Kit99-0562-00 Quantima Bench includes Storage Shelves and Wheels