09sn70 sputtering system op_maual_eng

49
SYSTEM Magnetron Sputter MODEL(S/N) PSP 5004(09SN70) SUPPLIER SNTEK CUSTOMER UTHM MAGNETRON SPUTTERING SYSTEM SNTEK Co.,Ltd

Upload: muhd-syahid-murtuzah

Post on 21-Apr-2015

70 views

Category:

Documents


7 download

TRANSCRIPT

Page 1: 09sn70 Sputtering System Op_maual_eng

SYSTEM Magnetron Sputter

MODEL(S/N) PSP 5004(09SN70)

SUPPLIER SNTEK

CUSTOMER UTHM

MAGNETRON SPUTTERING SYSTEM

SNTEK Co.,Ltd

Page 2: 09sn70 Sputtering System Op_maual_eng

Contents

1. INSTALLATION

1.1 Preparations before installation ·································································································1

1.2 Basic check before operation ········································································································2

1.3 Checking period for each part ······································································································3

1.3 Explanation of warning and caution marks ····················································································4

2. SYSTEM CONFIGURATION

2.1 Block Diagram ································································································································5

2.2 Hardware configuration ··················································································································6

2.3 Program configuration ··················································································································11

3. SYSTEM OPERATION

3.1 Quick Start ····································································································································12

3.2 Operation.(Manual mode) ···············································································································17

1) PCW, CDA, GAS Supply ···········································································································17

2) System power on ····················································································································18

3) Chamber Pumping ··················································································································19

4) Gas injection ·······················································································································20

5) Working Pressure Control ·····································································································21

6) Sputter Gun Power Supply ···································································································24

4. MAINTENANCE

4.1 Replacement of sputter target ·····································································································25

4.2 Rotary pump oil check & Refill ···································································································27

5. Explanation of part set up & use

5.1 Baratron gauge ···························································································································29

5.2 ATM sensor ····································································································································34

5.3 CDA, Air regulator ······················································································································37

5.4 CDA, Air Pressure switch ············································································································39

5.5 PCW Pressure switch ····················································································································40

6. TROUBLESHOOTING ·······························································································································41

Page 3: 09sn70 Sputtering System Op_maual_eng

- 3 -

Appendix A.

Roles of various sensors and failure occurring situation ··························································43

Appendix B.

Interlock list ··································································································································44

Page 4: 09sn70 Sputtering System Op_maual_eng

1. INSTALLATION

1.1 Preparations before installation

Item Details

System S/N - 09SN70

Model - PSP 5004

Power source - AC 415V, 3phase, 30A

Earthing - 1 class earthing equivalent device

- Be sure to connect with Chamber Ass'y Body

PCW - DI water : recommended of 15~20 ℃.

- Optimum hydraulic pressure : 2~4 Kgf/Cm3

- Optimum flux : 15 L/min

- Connected with Φ12 urethane tube

CDA - Use compressed air without water

- Optimum air pressure : 4~6 Kgf/Cm3

- Connected with Φ6 urethane tube

Process Gas - 99.999% class high purity Gas recommended for Ar, N2, O2 Gas .

- Optimum pressure: 2~3 Kgf/Cm3 (25 ~ 40 psi)

- Connected with Lock, VCR Type

Purge Gas - Use of N2 gas recommended for venting.

- Optimum pressure: 2~3 Kgf/Cm3 (25 ~ 40 psi)

Page 5: 09sn70 Sputtering System Op_maual_eng

- 2 -

1.2 Basic check before operation

➀ Vacuum ChamberThe inside of process chamber should be kept clean consistently, and vacuum state should always be

maintained in no operation of the equipment. O-ring part should be no contamination to avoid leakage.

➁ Vacuum PumpThis plays an important role to generate and maintain vacuum. If Chamber inside is not clean, it

takes much time to maintain vacuum state, and trouble occurring frequency of vacuum pump would

increase.

- Dry pump type : a vacuum pump requiring not so frequent maintenance and repair, to be checked

occasionally if the supply state of purge N2 and PCW are normal.

-Oil pump type : Oil gauge at the back of pump should be checked. If oil color has changed or oil

volume is below optimum level, it should be replaced or supplemented for use.

➂ MFC & Gas Line As the leakage of process gas used for generating Plasma may cause damage to devices or human body,

any damage or leakage of gas line should be checked at all times. In addition, by checking process

pressure occasionally, always be careful that problems in the process would not occur due to the gas

shortage during the process.

To supply gas in the required volume. MFC(Mass Flow Controller) is used. Gas pressure at the inlet

be adjusted to around 2~3Kg/Cm2(25~40 psi). the supply of high pressure gas may cause malfunction and

trouble of MFC.

➃ Air pressure supply part As vacuum valves of this equipment are operated by air pressure with constant pressure, you should

always check if the pressure of air pressure supply part is at 5~6 Kg/cm2.. If the pressure is

supplied below 5~6 Kg/cm2,, abnormal function of vacuum valves may occur. And, be careful that water

be not contained in the compressed air, since compressed air containing water may cause troubles of

air compressed valves. .

➄ Exhaust line You should check appearance and leakage of the exhaust line for treating exhaust gas discharged from

vacuum chamber, and, if any trouble found, use it after remedy.

➅ the state of Input power device and accessories Fixing state of input power cable, earthing state, operation state of sensor lamp and each switch

should be checked occasionally.

Page 6: 09sn70 Sputtering System Op_maual_eng

- 3 -

1.3 Checking period of each part

Item Checking period

Replace/Inspection·Correction/

Repair period

Daily Weekly Monthly Quarterly Weekly Monthly 6 Month Yearly

Cleanness of

Chamber inside- ◎

Input power 415V ±10V ◎

CDA Pressure

Supply Line

0.5 ~ 0.6 Mpa

(5 ~ 6kgf/cm2)◎

Process gas

line pressure

0.2 ~ 0.4 Mpa

(2 ~ 4kgf/cm2)◎

PCW Pressure

Supply Line

0.2 ~ 0.4 Mpa

(2 ~ 4kgf/cm2)◎

PCW flow 15L/min ◎

PCW Temp < 25 ℃ ◎

Vacuum gaugeCheck Base pressure

Working pressure◎ ◎

Low vacuum pump

(Oil type)

Maintain Oil gauge

over 80% ◎ ◎

High vacuum

pumpRPM = Normal ◎ ◎

RF Power

Supply

Max output power:

600W ◎ ◎

Gas control

device

(MFC)

Operation check of Gas

flow ◎ ◎

O-Ring Check surface state ◎ ◎

Viewport GlassCheck surface

contamination◎ ◎

RF Power

Cable

Check damages, short

circuit ◎ ◎

Fitting &

Accessaries◎ ◎

Page 7: 09sn70 Sputtering System Op_maual_eng

- 4 -

1.4 Explanation of warning and caution marks

Warning and caution marks Relevant contents

Operate according to the procedure

Notice

Refer to relevant manuals before use

Caution

Work requiring for clean tools, as the place is sensitive to dust

and other foreign materials.

Caution

Operation prohibited by non-authorized.

Caution

Use with care as it is risky

Warning: Life Threatening Voltage!

Danger of high voltage

Warning : High Temperature !

Be careful of burn due to high temperature

Warning : Electromagnetic field !

Danger of electromagnetic field

Danger : High frequency !

Users especially with cardiac pacemakers should work out of danger

range as the area requires cautions for high frequency emission.

Page 8: 09sn70 Sputtering System Op_maual_eng

- 5 -

2.SYSTEM CONFIGURATION

2.1 Block Diagram

Page 9: 09sn70 Sputtering System Op_maual_eng

- 6 -

2.2 Hardware Configuration

[Fig.1 System front]

[Fig.2 System bottom]

Page 10: 09sn70 Sputtering System Op_maual_eng

- 7 -

[Fig.3 System side]

[Fig.4 Sputter gun power connector]

Page 11: 09sn70 Sputtering System Op_maual_eng

- 8 -

[Fig.5 Sputter gun & process chamber]

Page 12: 09sn70 Sputtering System Op_maual_eng

- 9 -

[Fig.6 System control unit]

Page 13: 09sn70 Sputtering System Op_maual_eng

- 10 -

[Fig.7 System control back panel]

Page 14: 09sn70 Sputtering System Op_maual_eng

- 11 -

2.3 Program configuration

[Fig.8 Main control window]

[Fig.9 Data view window]

Page 15: 09sn70 Sputtering System Op_maual_eng

- 12 -

3. S YSTE M O PE R ATIO N

3.1 Q uick S tart

STEP 0 System Main Power ON

1) U nlock the key switch.

2) Turn on the power (the green button).

- If the main power (220V) is applied normally, the red button (the 'O FF' button) is remained 'O N '.

3) Turn on the PC power.

4) Execute the control program on the PC screen.

※ As the actual Power O N /O FF buttons may differ from the image above, please use them according to

the actual system.

STEP 1 PCW, CDA Supply

1) PC W , C ompressed Air (C D A) Supply

2) System Alarm check

C aution

In case of alarm or warning signals, please check the corresponding parts and then proceed.

Page 16: 09sn70 Sputtering System Op_maual_eng

- 13 -

STEP 2 Sample Loading

1) Chamber vent

2) Sample Loading

1) Chamber Vent

- M/V, R/V close

- GSD/V open

- Vent/V open

- ATM signal ON check

2) Sample loading

- Main shutter open

- Sample Loading

- Main shutter close

C aution

Please use the clean tool when installing a sample.

Page 17: 09sn70 Sputtering System Op_maual_eng

- 14 -

STEP 3 System Pumping

1) TMP normal stand-by

- Rotary pump start

- Line gauge pressure check

- F/V open

- TMP start

- TMP normal signal check

2) Chamber low vacuum pumping

- F/V close

- R/V open

- Low vacuum T/V 100% open

- Chamber pressure check (Chamber pressure < 5.0E-2 Torr)

3) Chamber high vacuum pumping

- R/V close

- F/V open

- M/V open

- T/V 100% open

C aution

O peration of the vacuum valves or pumps should only be conducted by the qualified personnel.

Page 18: 09sn70 Sputtering System Op_maual_eng

- 15 -

STEP 4 Sample Deposition

1) Setting of C onditions for Sample Vapor D eposition

- Set the R PM .

- Set the heater temperature (Power O N ).

- C lose the G SD /V .

- Set the flow rate (M FC values) of process gas and open the gas valves.

- Adjust the process pressure (Throttle/V : set 'pressure mode')

- Set the D C or R F Power values.

- Set the vapor deposition time.

2) Sample Vapor D eposition

- O pen the gun shutter.

- Turn on the gun power (D C or R F).

- O pen the main shutter.

3) End of Sample Vapor D eposition

- Turn off the gun power.

- Turn off the heater power.

- C lose the gas valve.

- O pen the Throttle/V 100% .

- The sample stay in the chamber at least the heater temperature go down 200℃.

- C lose the M /V .

- O pen the G SD /V .

- O pen the Vent/V .

C aution

Please be careful of burns as temperature around the sample holder and the sputter gun is very high.

Page 19: 09sn70 Sputtering System Op_maual_eng

- 16 -

STEP 5 System OFF

1) Turn off the TMP.

- Stand-by during the time required for TMP to stop.

2) Close the F/V.

3) Pumping for Low Vacuum Chamber

- Open the R/V.

- Set the chamber pressure at < 5.0E-2 Torr.

- Close the R/V.

4) Stop the rotary pump.

5) Turn off the system power.

N otice

The process chamber must be switched to the 'vacuum' condition after use to protect from foreign materials and/or moisture.

Page 20: 09sn70 Sputtering System Op_maual_eng

- 17 -

3.2 O peration.(M anual mode)

N otice

P lease get fully familiar with the manuals and the sequence of system operation before use.

1) PC W , C D A, G AS Supply

PCW, CDA, Gas port size

Port Port spec Note Pressure spec

System PCW 1/2 inch Lock type 2~3kgf/cm2

Sputter Gun PCW 1/2 inch Lock type 2~3kgf/cm2, 3 L/min

CDA 1/4 inch Lock type 0.4~0.6Mpa

GAS 1/4 inch Lock type 0.2~0.3Mpa

Please install as shown in the picture below for PCW, CDA, Gas supplies before use.

N otice

P lease use high purity gas above 4N for the process.

C aution

Please set the pressure at 0.2 to 0 .3M pa by use of a regulator for the initia l gas supply to each gas line. U se of excessive pressure that this may cause malfunction or failure of M FC .

Page 21: 09sn70 Sputtering System Op_maual_eng

- 18 -

2) System power on

Connection of main power cable

Voltage Current Pulse Cable spec

220 MAX 60A 3 10sq.

System power supply

1) Switch to the 'key switch unlock' status.

2) Switch to the 'System Power On' status. (the green button).

3) Turn on the control PC power and execute the sputter operation program.

4) Check the alarm signals as shown at the left bottom.

Window Alarm Normal Note

System Ready Lamp ONPC, PLC Communication Status

Water Pressure Lamp ON System PCW

Gun Water Flow Lamp ON Sputter Gun PCW

Gun Water Pressure Lamp ON Sputter Gun PCW

Air Pressure Lamp ON System compressed Air

Rotary Pump Trip Lamp OFF Rotary pump status

TMP Fault Lamp OFF TMP status

C aution

If an a larm is given, only use the system after correcting the cause of such alarm. Arbitrary changes in the alarm signals or adjustments in the sensitivity of sensors may cause a system fa ilure.

Page 22: 09sn70 Sputtering System Op_maual_eng

- 19 -

3) C hamber Pumping

Pumping for Low Vacuum Chamber

1) Start the rotary pump.

2) Check the current pressure at the line convection gauge.

(The pumping line is 'normal' if the value at the line gauge drops below 5.0E-2 Torr.)

3) Start pimping for the low vacuum chamber by opening the R/V.

4) Leave the manual T/V completely open during pumping for low vacuum chamber.

C aution

D O N O T disassemble the pumping line or the chamber blank port during pumping as this may affect the vacuum speed and the vacuum status of the chamber as well as breakdown of the rotary pump.

TMP start

1) For normalization of the TMP, close the R/V to stop pumping the chamber.

2) Stand-by the line gauge until the pressure at line gauge drops below 5.0E-3Torr.

(See the picture above)

3) Open the F/V.

4) Stand-by until the normal speed signal is turned on after starting the TMP.

Open the M/V to start the high vacuum pumping if the chamber pressure drops below 5.0E-2 Torr

upon the TMP normal signal.

When the chamber pressure is above 5.0E-2 Torr, open the M/V after repeating the 'low vacuum

pumping'.

Page 23: 09sn70 Sputtering System Op_maual_eng

- 20 -

4) G as injection

Either the low vacuum pumping or the high vacuum pumping status should be maintained to inject the

process gas into the chamber.

Set the MFC values for gas required for the process.

Supply the process gas.

1) G.M/V open

2) Gas valve open

Page 24: 09sn70 Sputtering System Op_maual_eng

- 21 -

5) W orking Pressure C ontrol

In the Throttle/V Pressure Mode

1) Set the Throttle/V into 'Remote' mode.

2) Enter the process pressure value to the window of Throttle/V by the unit of mTorr after supplying

the process gas.

3) Transmit the pressure value to the Throttle/V controller by using the set press button.

4) Start the pressure control by clicking the Throttle/V run button.

5) Wait until the input pressure value equals the current pressure value.

Window Control Normal

Set PressTransmit the input process pressure value to the T/V controller.

Set Pos Transmit the input T/V position value to the T/V controller.

Local Mode Switch the T/V controller into the 'Local Mode'.

Remote Mode Switch the T/V controller into the 'Remote Mode'.

Position Mode Adjust the T/V by the input T/V position value.

Pressure Mode Adjust the T/V by the input process pressure value.

Full Open Open the T/V 100%.

Full Close Close the T/V 100%.

Run Start the T/V Control.

Stop Stop the T/V Control.

Page 25: 09sn70 Sputtering System Op_maual_eng

- 22 -

In the Throttle/V Position Mode

1) Switch the Throttle/V Control Mode to the "Remote Mode'.

2) Enter the Throttle/V position value (0~100%) in the Throttle/V window after supplying the process

gas.

3) Transmit the Throttle/V value by using the Set Pos button.

4) Click on the Throttle/V Run button to start the position control.

5) Adjust the Throttle/V position value to search the required pressure value.

Page 26: 09sn70 Sputtering System Op_maual_eng

- 23 -

In the Throttle/V Controller Local Mode

1) Switch the Throttle/V Control mode to the 'Local' mode.

2) Use the keys on the Throttle/V Controller panel to enter the required pressure and position values in

the set point values (A~E).

3) Press the Set Point buttons (A~E) to start adjusting pressure for the chamber.

Throttle/V controller panel

No Key Description

1 MODE To enter to the 'System Setup' mode.

2 UP To move to the System Setup mode and to the Parameter Set mode in turn, also to increase the setup values in the Parameter Set mode

3 DOWN To move to the System Setup mode and to the Parameter Set mode in turn, also to decrease the setup values in the Parameter Set mode.

4 SHIFT To select setup values for the System Setup mode, also to select the setup digit in the Parameter Setup mode.

5 ENT. To be used at the beginning and at the end of entering parameters.

6 OPEN To open the valve completely.

7 CLOSE To close the valve completely.

8 STOP To stop the Open/Close/Run valve operations.

9 A To select and start to control the Set point A.

10 B To select and start to control the Set point B.

11 C To select and start to control the Set point C.

12 D To select and start to control the Set point D.

13 E To select and start to control the Set point E.

14 LEARN -

15 ZERO To reset the current pressure value to '0'.

Page 27: 09sn70 Sputtering System Op_maual_eng

- 24 -

6) Sputter G un Power Supply

Enter the DC or RF Power values after the process pressure is stabilized.

The set vapor deposition time will be counted after the main shutter is opened.

Switch to the RF or DC Power Local mode.

1) Press the remote/local switch located in front of the generator to switch OFF the LED in order to

enter the DC Power Local mode.

2) To switch to the RF Power Local Mode, replace the interface connector (D-sub 25-pin) terminal

located at the rear of the generator with the one for local use.

Page 28: 09sn70 Sputtering System Op_maual_eng

- 25 -

4. MAINTENANCE

4.1 Replacement of sputter target

Switching to the RF / DC Power Local Mode.

1) Open the Target Shutter and remove the target shield cover.

2) Remove particles at the front and side of the target (by using N2 Blower).

3) Remove the clamp bolts located in front of the target. Take care that the target should not fall.

4) Remove the target after removing the target clamp bolts. Take care that the magnet should not be

detached from the gun (for the Gun type where magnet is exposed at the rear of the target).

5) Use the correct clamp and bolt to fit the target thickness during replacement.

6) Install the target tightly close to the gun.

7) Remove particles at the front and side of target after fixing all clamp bolts (by using N2 Blower)..

8) Fit the target shield cover.

9) When fitting the target shield cover, keep a distance of 2~3mm from the target.

C aution

Poor contact between the target and the gun may cause excessive heat to generate on the target surface during processes resulting in deformation or creak of the target.

Page 29: 09sn70 Sputtering System Op_maual_eng

- 26 -

Page 30: 09sn70 Sputtering System Op_maual_eng

- 27 -

4.2 R otary Pump O il C heck & R efill

For oil check and replacement, please refer to the attached sub-manual.

Before check and/or replacement, please check if the TMP is turned off and the rotary pump

is stopped.

1) Keep the rotary pump stopped.

2) Refill the oil through the oil inlet when the oil gauge at the rear of the rotary pump is

below the appropriate level (80~90% of the full gauge).

3) When the oil is discolored (clear in normal cases), refill with new oil after emptying oil in use

completely through the oil outlet.

C aution

Pump checkup and oil refill/replacement should be conducted only by the authorized personnel.

C aution

Before checking and replacing pumps, please turn off the power.

C aution

If an excessive amount of moisture flows into the pump, it may impair performance and/or cause breakdown of the pump.

C aution

Always use the correct type of oil in service.

Page 31: 09sn70 Sputtering System Op_maual_eng

- 28 -

Page 32: 09sn70 Sputtering System Op_maual_eng

- 29 -

5. Explanation of part set up & use

Notice

Instruction for the part used is a material to describe the parts commonly

applicable to the System delivered by our company. The relevant parts should

be dissembled, assembled, and calibrated accordingly.

Meanwhile, as this is a brief instruction for system users, you need to refer

to other relevant manuals for more details.

5.1 Baratron gauge

1) Use and related data

Maker Use Reange of the use

MKS InstrumentProcess pressure control

(Working pressure)MAX 10 Torr

Baratron gauge performance specification data

Page 33: 09sn70 Sputtering System Op_maual_eng

- 30 -

Notice

As the model of Baratron 627B type used in this system is a self heating type

of gauge, warming up time is required before use.

2 hours ≥ 1Torr, 4 hours ≤ 1Torr

Notice

As a controllable range is different from Barantron gauge Full scale Range,

you need to refer to the table and relevant manual.

ex> When using 10Torr Baratron gauge, minimum range of effective pressure is

5.0E-3 Torr, controllable minimum range 5.0E-2 Torr.

Page 34: 09sn70 Sputtering System Op_maual_eng

- 31 -

2) Baratron gauge calibration

Execute if sample results are different under the same conditions in process or execute in

accordance with inspection and correction period,

Notice

Warm up time should be kept before the progress of Baratron gauge CAL.

Caution

As problems may occur when operated arbitrarily, persons except authorized or

responsible ones are prohibited from the operation of Baratron gauge CAL.

1) After stan-by of the process chamber for more than 2 hours at the high vacuum state, check if

the degree of chamber vacuum is below 1.0E-4 Torr. (Checking possible by Pirani gauge)

Caution

Arbitrary CAL progress prohibited until reaching at the degree of vacuum

practicable for CAL.

2) After reaching at the degree of vacuum in the process chamber suggested in the above table,

measure the voltage of Baratron gauge.

3) Measure the DC voltage at the required position using voltage testor, (For the position, refer

to the picture below)

Page 35: 09sn70 Sputtering System Op_maual_eng

- 32 -

4) Progress after adjusting the voltage measuring value of PLC A/D 10 pin to 0V(zero voltage). At

this time, be sure to progress with more than 2 persons.

5) Zero voltage adjusting position

Warning

As this is Baratron gauge of heating type, be careful of the burn by the

surface while working

6) At the time of voltage adjustment, if the measuring value of PLC A/D 10 pin indicates 0V(zero

voltage) after turning to the left & right using a precision driver of (-) type, then stop the

progress of CAL.

Page 36: 09sn70 Sputtering System Op_maual_eng

- 33 -

3) Removal of Baratron gauge

At the time of inspection/correction of Baratron gauge, and separating work for A/S treatment,

Baratron gauge, progress according to the following procedure.

1) Progress a chamber Vent, and convert into atmospheric state.

2) Remove a Gauge power cable.(For cable removal, remove a fixing bolt first using a driver.

3) Separate from chamber after removing the fixing clamp.

4) To prevent the sensor part from the contamination after removing a vacuum gauge, put on the

sensor part the cover in protection caps, foils, or wraps, Separating work should be done by clean tools.

Notice

Sensitive area to dust and other foreign materials. Work required with

Clean tools.

Page 37: 09sn70 Sputtering System Op_maual_eng

- 34 -

5.2 ATM sensor

1) Use and related data

Maker Use Range of the use

Autonics

Check the atmospheric (760Torr)

signal

(ATM signal)

0 ~ 100kpa

(0 ~ 750 Torr)

ATM sensor performance specification data

Page 38: 09sn70 Sputtering System Op_maual_eng

- 35 -

2) ATM sensor calibration

When the gas pressure of Vent Gas(GN2) is at the normal state, and if vent state at more than

average vent time is sustained ,

Checking method of the atmospheric state at troubles of ATM sensor

1) When venting is done with Gas(GN2), if the chamber inside is at the atmospheric pressure or

over-pressure state(760Torr이상), gas comes out to the outside at the adhering part of Door and other

vacuum-applied items.

If the above occasion happens, progress the CAL(zero point) according to the following procedure.

1) Check if the chamber is at atmospheric state or over.

2) Convert the program into Manual mode(If in progress in Auto state, stop the Auto state)

3) Close the Vent valve.

4) Check the contact point of the current situation of ATM sensor.

Normal Signal in ATM Status Abnormal Signal in ATM Status

5) If No. 2 contact signal coms out at ATM state, progress Zero point CAL.

6) If pushing the up/down buttons( ) for over 1 second at the same time, On cones on No. 1

Contact signal.

6) Check On of ATM lamp on the program screen.

Page 39: 09sn70 Sputtering System Op_maual_eng

- 36 -

3) CN range change of No. 1 Contact point of ATM sensor

Execute if there are frequent progresses of ATM zero point CAL.

The output operation mode of ATM sensor used in this system is F-6 window mode.

Signal output of F-6 window mode

ON(stand-by state) range change method of No. 1 Contact point

Notice

Progress after reading sufficiently the relevant manuals when changing inside

parameters of ATM sensor.

Page 40: 09sn70 Sputtering System Op_maual_eng

- 37 -

5.3 CDA, Air regulator

1) Use and related data

Maker Use Range of the use

SMCOperate Diaphragm valve and

Cylinder

Detting pressure 0.4 ~ 0.6 Mpa

(4 ~ 6kgf/cm2)

2) Adjustment of used pressure

If Main CDA or Air pressure is changed, adjust the gauge to the using pressure with Pressure

control knob, and thereafter convert the knob into the Lock state.

Notice

After rotating the knob toward (-) direction at maximum at the initial

pressure setting, progress at no pressure supplied to the regulator out.

Use after moving the indicating label to the desired place for the using range,

Page 41: 09sn70 Sputtering System Op_maual_eng

- 38 -

3) Water removal

If contained water is supplied through filter while using CDA or Air, moisture gathers near the

filter. If moisture stays while checking, remove it.

1) Rotate the moisture removing nozzle at the lower part of regulator toward the Unlock direction.

2) If a little air venting sound is heard while rotating, sustain at that state.

3) While the inside compressed air comes out, the moisture stayed inside flows outward.

4) When moisture removal is completed, rotate the nozzle again to the Lock direction to prevent the

inside compressed air from flowing outward.

Caution

If the moisture removal nozzle is open too much, that may bring about

temporary low pressure, resulting in system alarm, or stops or malfunctions

of valves and other parts in operation.

Progress prohibited for danger in process.

Page 42: 09sn70 Sputtering System Op_maual_eng

- 39 -

5.4 CDA, Air Pressure switch

1) Use and related data

Maker Use Renage of the use

SMC CDA, Air Pressure check sensorSetting pressure 0.4 ~ 0.6 Mpa

(4 ~ 6kgf/cm2)

2) Adjustment of using pressure

Main CDA or Air pressure is being supplied within the usable range, but adjust if Air pressure

alarm occurs.

1) Adjust by rotating the upper lever to (-)(+) directions until the Air pressure lamp is ON

2) If the regulator pressure is used at 4Mpa, and if Air pressure sensor is set at 4Mpa, Air

pressure alarm may occur frequently, Thus, if the regulator pressure is used at 0.4Mpa, the range of Air

pressure sensor needs to be set at below 0.4Mpa (about 0.3~0.35Mpa) for use.

Notice

As the operating pressure of Diaphragm valve used in this system is 0.4~0.7Mpa

with Main Regulator pressure of 0.6Mpa, Air sensor needs to be set at 0.4Mpa

for use.

Alert in

Lamp Off

Page 43: 09sn70 Sputtering System Op_maual_eng

- 40 -

5.5 PCW Pressure switch

Maker Use Range of the use

- PCW pressure check sensorSetting pressure 0.2 ~ 0.3 Mpa

(2 ~ 3kgf/cm2)

2) Adjustment of the using pressure

PCW pressure is being supplied within the usable range, but adjust if a pressure alarm happens,

1) Adjust by rotating the upper lever to Water pressure (-)(+) directions, until the Water

pressure is ON.

Caution

Too low setting of a Water pressure sensor may cause normal signal output of a

Water pressure sensor under no actual PCW supply. If system operation is

progressing without the supply of PCW, system part deformation or troubles may

follow.

Notice

The figures of PCW Flow meter are not in proportion to those of PCW pressure.

Even if the Pressure is measured high, actual figure of PCW Flow meter may be

measured low.

Adjuster

Adjuster

Page 44: 09sn70 Sputtering System Op_maual_eng

- 41 -

6. TROUBLESHOOTING

(1) If there is any trouble in vacuum, that is, if vacuum vent is not done properly or the Leakage

rate is relatively high

<Cause>

- breakage of vacuum O-Ring

- if the degree of vacuum is not indicated correctly due to the breakage of vacuum sensor

- if there is no vacuum due to the failure of piping and gas line

- failures of vacuum valves.

<Measure>

- If any breakage of O-Ring occurs due to the long use of vacuum O-Ring or

mistake of workers

- Replace an O-ring, check occasionally the state of Vacuum O-ring,

and apply Vacuum grease.

- Be careful of the impact as vacuum sensors are sensitive to the impact.

Replace the sensor if broken.

- Check occasionally the state of gas line and replace if any failure.

- Replace immediately if a vacuum valve is in failure.

(2) If no discharge even at the vacuum state.

<Cause>

- short circuit of electric wire connected between the electrode and the Vacuum Chamber

- Connection between the electrode and the Ground(Short state)

- when we want to generate the discharge at too high vacuum or too low vacuum

- Power supply failure

<Measure>

- Cut off Main Power, check the state of connected electric wire, and if any failure, replace.

- Install the electrode again and check the discharge state.

- Check if any failure of vacuum gauge, adjust to the proper degree of vacuum,

and start the discharge.

- Repair and replace the Power supply(request to the maker)

Page 45: 09sn70 Sputtering System Op_maual_eng

- 42 -

(3) If Plasma discharge is not constant

<Cause>

- If the gas supply is not done or not normal

- If it is turned into the vacuum state improper to discharge

<Measure>

- Check any locking failure of gas line including gas pressure and air pressure valves,

and, if any failure, replace.

- Check the operation of MFC, and, if any failure, replace

- Check the motion of vacuum gauge, and, if any failure, repair or replace.

(4) In case of severe arc,

<Cause>

- Local discharge phenomena occur due to particles by heavy contamination

at the Vacuum Chamber inside

<Measure>

- Arc generates from the excessive electric current inside Vacuum Chamber.

First of all, clean the Vacuum Chamber.

- After loading the sample properly, and start the process again.

(5) In case of sustaining the high degree of vacuum

<Cause>

- Failures of gas supply valves and MFC

<Measure>

- Check the gas line, operating state of gas pressure or air compressure valves,

and the degree of air compressure. Take actions for gas supply.

(6) Others

- When an excessive current cut-off switch is down, check the short circuit of supply power

and electric wire, secure the power and replace the disconnected wire.

If any electric problem happens, down the line breaker and check any failures.

Page 46: 09sn70 Sputtering System Op_maual_eng

- 43 -

Appendix A.

- Roles of various sensors and failure occurring situation.

(1) Water Pressure Switch

Action Detects a water pressure, Interlock happens in case of

insufficient pressure

Interconnection below setting pressure : Interlock, above setting pressure :

signal ON

Failure

1) below setting pressure & signal ON state

- All devices can operate the signal ON, but device damages

happen due to the heat by cooling failure of Vacuum Flange

2) above setting pressure & signal OFF state

- Processing operation does not work in Interlock state

(2) Air Pressure Switch

Action Detects an air pressure. Interlock happens in case of insufficient

pressure

Interconnection below setting pressure : Interlock, above setting pressure :

signal ON

Failure

1) below setting pressure & signal ON state

- All devices can operate with signal ON, but smooth

operation is impossible.

2) above setting pressure & signal OFF state

- Processing operation does not work in Interlock state.

(3) ATM Sensor (Atmospheric sensor)

Action Detects an atmospheric pressure. Signal On when reaching at

the atmospheric pressure,

Interconnection below atmospheric pressure : Interlock, above atmospheric

pressure : signal ON

Failure

1) above atmospheric pressure & signal OFF state

- Consistent Vent Valve Open state and constant Vent Gas

inflow cause damages to devices due to high pressure.

Page 47: 09sn70 Sputtering System Op_maual_eng

- 44 -

Appendix B.

- Interlock list

1. Main Valve(M/V) is not set in the TMP and currently not used.

- Basic Interlock setting is done.

- Openable below 5*10-2 Torr Convection Gauge Sensor Display value

- M/V Open possible only for F/V Open

2. TMP S/W is currently not used.

- On possible only for BR.P operation.

3. Vent Valve(V/V) can operate only in R/V and M/V Off.

- V/V Automatic Off if the value of ATM Sensor Display is 0.00 Torr (S.P 1 On).

4. Gas Valve can operate in On of Gas Main Valve(G.M/V).

Page 48: 09sn70 Sputtering System Op_maual_eng

MEMO

Page 49: 09sn70 Sputtering System Op_maual_eng

- 46 -

987-1, Gosaek-dong, kwonsun-gu

Suwon-Shi, Gyonggi-Do, Korea

TEL : 82-31-299-3823

FAX : 82-31-299-3829

http: www.sntek.com