Page 1: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

1

Atomic Absorption Atomic Absorption SpectroscopySpectroscopy

Page 2: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

2

Atomic Transitions: Excitation and Emission

Page 3: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

3

Types of Atomic Flame Spectroscopy

Page 4: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

4

Process for Metal Ion in Flame AA

Page 5: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

5

Double Beam AA Spectrometer

Page 6: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

6

Sample Intro Into Flame

Page 7: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

7

Burner Chamber and Nebulizer

Page 8: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

8

Monochromator

Page 9: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

9

Detection Limit

Page 10: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

10

Types of MachinesTypes of Machines

•Flame: absorption and emission•Graphite Furnace: absorption and emission•Plasma: emission–ICP: inductively coupled plasma

–ICP-MS: inductively coupled plasma mass spectrometry

Page 11: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

11

The FlameThe Flame

•Depending on the element, different temperatures are required.•So different fuel and oxidants are employed.•For typical elements, acetylene is the fuel and air is the oxidant.•But nitrous oxide, oxygen may be used as the ox.•Hydrogen and cyanogen may be the fuel.

Page 12: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

12

The FlameThe Flame

Page 13: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

13

Graphite FurnaceGraphite Furnace

•Graphite furnace AA machines are more expensive and more tempermental, but they have advantages.•The amount of sample required is very small, and autosamplers are used to ensure a very small, yet consistent amount is injected into the furnace.•Sample size is 1-100 L•The sensitivity is greater than flame.

Page 14: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

14

Graphite FurnaceGraphite Furnace

•In a graphite furnace, the sample is injected into a platform, where it can reside for several seconds before being atomized.•This long residence time results in the higher sensitivity and low sample requirement.•Again, it is wise to use an autosampler to take full advantage of the capabilities and sensitivities.

Page 15: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

15

Page 16: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

16

Graphite FurnaceGraphite Furnace

•In the chamber, a sample is first dried at a low temperature to remove the solvent.•If necessary to remove organic matter, it will then undergo pyrolysis (charring) at a higher temperature (1000 C)• Finally, the sample will be atomized.•Absorption and emission are both used in graphite furnaces.

Page 17: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

17

ICPICP

•In ICP, much higher temperatures are employed to create a plasma.•Of course, these are very expensive machines!•They utilize emission.

Page 18: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

18

ICP: Creating the PlasmaICP: Creating the Plasma

•Argon gas is ionized by a spark from a Tesla coil.• The ionized argon gas is heated to 6,000-10,000 C by the collision of free electrons with the argon atoms.

Page 19: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

19

Page 20: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

20

Page 21: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

21

ICP: The SampleICP: The Sample

•Meanwhile, the sample is injected into the system which aerosolizes it.•The aerosol is then dried by being carried through a heated tube by argon gas (not the plasma).•The analyte then reaches the plasma.•The analyte is then atomized, excited, and its emission is measured.

Page 22: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

22

Page 23: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

23

ICP-MSICP-MS

•In ICP-MS, a mass spec is coupled to the ICP so that extremely low detection limits are obtained.

Page 24: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

24

Types of Atomic Flame Spectroscopy

Page 25: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

25

Temperature EffectsTemperature Effects

•You know that as the temperature increases, the energy of atoms increases.•So as the temp goes up, the number of atoms in an excited state also increases.•This is very important for emission techniques.•The rule of thumb is that for every 10K increase in temp, there will be a 4% increase in excited state atoms, and hence a 4% increase in emission intensity.

Page 26: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

26

Temperature EffectsTemperature Effects

•Why is this?•If the baseline temp is increased, a higher population of atoms will be in the excited state.•For emission, the energy released when these excited state atoms relax is measured.•So the more atoms in the excited state, the more energy is released and measured.

Page 27: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

27

Temperature EffectsTemperature Effects

•Temperature has a much lower effect on absorption measurements.•For absorption, there are still mostly ground state atoms even at higher temperatures.•Since absorption measures the ground state atoms becoming excited state, there is very little difference as the temp is raised slightly.

Page 28: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

28

InterferencesInterferences

•There are several types of interferences that affect a measurement of abs or emission.–Chemical

–Spectral

–Ionization

–Matrix

Page 29: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

29

Chemical InterferencesChemical Interferences

•Chemical interferences occur when another substance is present in the sample that suppresses the atomization of the analyte.•Anions like sulfate and phosphate can bind to metal ions like calcium to lower the amount of calcium atoms that actually atomize.•So the calcium reading would be low (sound familiar?).

Page 30: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

30

Chemical InterferencesChemical Interferences

•To prevent this, another metal such as La3+ is added to preferentially remove the anion.•EDTA and other agents may also be employed.•These are called “releasing agents” as the remove the interfence, releasing the analyte so that it may be fully atomized.•Having a higher temp and higher fuel ratio may also help.

Page 31: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

31

Spectral InterferencesSpectral Interferences

•This occurs when another substance has a signal that overlaps with the analyte signal.•This would result in high measurements.•If you know this, or deduce this from control samples, the best thing is to choose another wavelength to monitor.

Page 32: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

32

Ionization InterferencesIonization Interferences

•This occurs usually with alkali metals, as they have low ionization energies.•As the atoms become excited, they don’t just get excited, a fraction will also ionize!•Of course, this would lead to low measurements as the ionized atoms do not give off the same signal.•Na is typically 5% ionized, while K is up to 33% ionized!

Page 33: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

33

Ionization InterferencesIonization Interferences

•If you want correct measurements of alkali metals, then an “ionization suppressor” is a must.•Strangely enough, an even “easier” to ionize additive like CsCl is added!•The ionization of the Cs atoms yields an electron rich environment in the flame.•This then suppresses the ionization of Na or K as they will pick up an electron as easily as they will lose it (Le Chatelier’s Principle).

Page 34: 1 Atomic Absorption Spectroscopy. 2 Atomic Transitions: Excitation and Emission

34

Matrix InterferencesMatrix Interferences

•If the matrix for the sample is very different than the solvent system for the calibration standards, then matrix effects may result.•This may be due to chemical, spectral, or ionization interferences present in the complicated sample matrix.•The method of standard additions is commonly employed to compensate for simple matrix effects.