1) basic respiratory mechanics relevant for mechanical ventilation

Upload: irwandi-irwandi

Post on 07-Aug-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    1/14

    1

    ( ( ( ( (

    Peter C. Rimensberger, MDProfessor of Pediatrics, Critical Care and Neonatology

    Service of Neonatology and Pediatric Intensive CareDepartment of Pediatrics

    University Hospital of GenevaGeneva, Switzerland

    t

    t

    t

    t

    t

    t

    t

    t

    Volume Change

    Time

    Gas Flow

    Pressure Difference

    Basic respiratory mechanicsrelevant for mechanical ventilation

    Volume

    Pressure

    V

    P

    C = V P

    Compliance

    CRS influences V T

    PEEP PIP

    V o

    l u m e

    Surfactant

    for example with anintervention likesurfactant therapy:

    The Feature of the Tube

    Airway Resistance

    Pressure Difference = Flow Rate x Resistance of the Tube

    Resistance is the amount of pressure required to deliver a givenflow of gas and is expressed in terms of a change in pressuredivided by flow.

    R = PFlow

    P1 P2

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    2/14

    2

    Mechanical response topositive pressure application

    Active inspiration

    Passive exhalation

    R = ! P / V (cmH2O/L/s)

    Equation of motion: P = (1 / C rs ) V + R rs V.

    .

    Mechanical response topositive pressure application

    Compliance: pressure - volume

    Resistance: pressure - flow

    C = V / P (L/cmH2O)

    Pressure – Flow – Time - Volume

    Time constant: T = Crs x Rrs

    Volume change requires time to take place. When a step change in pressure is applied, the instantaneous change involume follows an exponential curve, which means that, formerly faster, itslows down progressively while it approaches the new equilibrium.

    P = ( 1 / Crs ) x V + Rrs x V.

    Time constant: T = Crs x Rrs

    Pressure – Flow – Time - Volume

    Will pressure equilibrium

    be reached in the lungs?

    A: Crs = 1, R = 1B: Crs = 2

    (T = R x 2C)

    C: Crs = 0.5(T = R x 1/2C)(reduced inspiratorycapacity to 0.5)

    D: R = 0.5(T = 1/2R x C)

    E: R = 2(T = 2R x C)

    Effect of varying time constants on the change

    in lung volume over time(with constant inflating pressure at the airway opening)

    A: Crs = 1, R = 1

    C: Crs = 0.5(T = R x 1/2C)(reduced inspiratorycapacity to 0.5)

    E: R = 2(T = 2R x C)

    Ventilators deliver gas tothe lungs using positivepressure at a certain rate .

    The amount of gasdelivered can be limited by time, pressure orvolume.

    The duration can be cycled by time, pressure or flow.

    Volume Change

    Time

    Gas Flow

    Pressure Difference

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    3/14

    3

    Nunn JF Applied respiratory physiology 1987:397

    ( )

    Understanding Time Constants at the bedside

    ! All about the FLOW wave form

    What would you do here?

    TiTe

    Optimizing tidal volume delivery at set pressure

    Lucangelo U, Bernabe F, and Blanch L Respir Care 2005;50(1):55– 65

    Flow termination and

    auto-PEEP detection

    PEEP i

    Trappedvolume

    Dhand, Respir Care 2005; 50:246

    Correction for Auto-PEEPgive bronchodilators and/or increase expiratory time

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    4/14

    4

    Appropriate inspiratory time

    Vt 120 ml

    Vt 160 ml Vt 137 ml

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    5/14

    5

    Zero flow Zero flow Zero flow

    Total insptime

    Total insptime

    Total insptime

    Aim for zero flow condition being less than ! of total insp. time

    How long should you set the Ti ?

    Alveolar

    Plateau

    Brown, DiBlasi Respir Care 2011;56(9):1298 –1311.

    How long should you set the Ti ?The patient will guide you !

    Proximal Airway Pressure

    Alveolar Pressure

    Look at the flow – time curve !

    P

    V.

    Cave! Ti settings in presence of anendotracheal tube leak

    Flow

    LeakTime

    The only solution in presenceof an important leak:Look how the thorax moves

    Time

    Volume

    Leak Vex < Vinsp

    Who

    s Watching the Patient?

    Pierson, IN: Tobin, Principles and Practice of Critical Care Monitoring

    Flow also influences Ti

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    6/14

    6

    Flow alters rate of change in lung volume:Rheotrauma

    • Sheer Stress is a stress parallel to thesurface of interest (strain)

    • An important cause of VILI, especiallyin circumstances of atelectasis and/orhigh flows

    Adapted from Jane Pillow

    Progressive increase in inspiratory time

    Lucangelo U, Bernabe F, and Blanch L Respir Care 2005;50(1):55– 65

    CAVE: you have to set Frequency and Ti (Control) or Ti and Te (TCPL)

    time

    f l o w

    insufflation

    exhalation

    0

    auto-PEEP

    Premature flow termination duringexpiration = gas trapping =deadspace (Vd/Vt) will increase

    Expiratory flow waveform:normal vs pathologic

    Expiration is important too

    Premature flow termination during expiration = gas trapping

    Auto-PEEP Analysing time settings of the respiratory cycle

    1) Too short Te " intrinsic PEEP 1) Correct Te

    2) Unnecessary long Ti 2) Too short Ti

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    7/14

    7

    V o

    l u m e

    PressurePressure

    V o

    l u m e

    Pressure Ventilation Volume Ventilation

    Decreased Tidal Volume Increased Pressure

    ComplianceCompliance

    Decreased PressureIncreased Tidal Volume

    and do not forget, the time constant (T = Crs x Rrs) will change too!

    Compliance decrease

    Lucangelo U, Bernabe F, and Blanch L Respir Care 2005;50(1):55– 65

    0.4 (s) 0.35 (s) 0.65 (s) 0.5 (s)

    52 cmH2O 33 cmH2O

    Change in compliance after surfactant =change in time constant after surfactant

    PEEP PIP PEEP PIP

    pre post

    V o

    l u m e

    Adapted from Kelly E Pediatr Pulmonol 1993;15:225-30

    C dyn did not change, C stat improved with surfactant

    C stat

    C stat

    Cdyn

    Cdyn

    PEEP PIP

    post

    C stat

    Cdyn

    Effect of high airway resistance on flow

    Time (ms)

    0 100 200 300 400 500 T i d a

    l V o

    l u m e

    ( m L / k g

    )

    0

    2

    4

    6

    Time (ms)

    0 100 200 300 400 500 T i d a

    l V o

    l u m e

    ( m L / k g

    )

    0

    2

    4

    6

    Time (ms)

    0 100 200 300 400 500 T i d a

    l V o

    l u m e

    ( m L / k g

    )

    0

    2

    4

    6!"#$%& &()*! !"#$%& ()*+(, - ." !"# &%$ +(, - .*()*"/+

    +,-,#, ./0+ 123%&.),$,$4#%), 5.6,%6,7! !"#&%- ()*+(, - ." !"# &%$ +(, - .*()*"/+

    82#").9 &()* 5.6,%6,! !"#&%0 ()*+(, - ." !"# &%-1 +(, - .*2*()*"/+

    Model: Ventilation with a pressure of 12 cmH2O overPEEP in various conditions (respiratory pathologies anddisease stages) in a neonate

    courtesy (modified) from Jane Pillow, Perth

    3 – 5 time constants

    2

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    8/14

    8

    Volume Guarantee: New Approaches in Volume Controlled Ventilation for Neonates. Ahluwalia J, Morley C, Wahle G. Dräger Medizintechnik GmbH. ISBN 3-926762-42-X

    Too short Te will not allowto deliver max. possibleVt at given P

    " will induce PEEPi" increases the risk for

    hemodynamic instability

    Too short Ti willreduce delivered Vt

    " unneccessary highPIP will be applied" unnecessary highintrathoracic pressures

    The patient respiratory mechanics dictate the maximal respiratory frequency

    Time constant: = Crs x Rrs

    To measure compliance andresistance - is it in the clinicalsetting important ?

    Use the flow curve for decision makingabout the settings for respiratory rate andduty cycle in the mechanical ventilator

    The saying we ventilate at 40/min or with a Ti of 0.3 is a testimony of no understanding !

    AlveolarPressure

    AirwayPressure

    Residual Flows

    Avoid ventilation out of control !

    t

    t

    t

    t

    t

    t

    t

    t

    Modes of Ventilation

    • Controlled – not synchronized – IMV

    • Controlled - synchronized – AC / PC – SIPPV / SIMV

    • Assisted - synchronized – Pressure Support / NAVA

    • Pressure versus Volume Ventilation – PC / VC / VG / TTV

    • High Frequency Ventilation – HFOV / HFJV / HFPPV

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    9/14

    9

    Pressure Control- A/C or PTV (VAC)

    Assist/Control

    Inspiration: Flow or pressuretriggered

    Breath Initiation (Inspiratory Triggering)

    IMVControlled

    – not synchronized

    IMV

    IMV versus SIMV

    SIMV

    SIMV versus IMV in neonatal ventilation

    Conclusions: We found that SIMV wasat least as efficacious as conventionalIMV, and may have improved certainoutcomes in BW-specific groups.

    Bernstein G et al. J PEDIATR 1996;128:453 -63 Greenough A et al. Cochrane Database of Systematic Reviews 2004, Issue 3. CD000456.

    " Better entrainement of respiratory muscles ?

    Observational study conducted in 349 ICUs from 23 countries

    Propensity score stratified analysis to compare 350 patientsventilated with SIMV-PS with 1,228 patients ventilated with A/Cventilation.

    Primary outcome was in-hospital mortality.

    No significant differences in in-hospital mortality After adjustment for propensity score, overall effect of SIMV-PS versus A/C on in-hospital mortality was not significant(OR, 1.04; 95% CI, 0.77-1.42; P=0.78).

    Ortiz et al. CHEST 2010; 137(6):1265–1277

    Outcomes of Patients Ventilated With SIMV + PSA Comparative Propensity Score Study

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    10/14

    10

    Effects of NIMV, synchronized NIMV (S-NIMV), and NCPAPon ventilation, gas exchange, and infant-ventilator interaction

    Chang H-Y, Claure N et al. Pediatr Res 69: 84–89, 2011

    Non-synchronized NIMV

    Synchr onizedNIMV

    Synchronized intermittent NIMV

    Effects of NIMV and synchronized NIMV (S-NIMV) onventilation, gas exchange, and infant-ventilator interaction

    Chang H-Y, Claure N et al. Pediatr Res 69: 84–89, 2011

    Patient-Ventilator Asynchrony with SIMV(in a Traumatically Injured Population)

    SIMV with set breathing frequencies of > 10 breaths/minwas associated with increased asynchrony rates (85.7% vs14.3%, p= 0.02).

    " No difference in ventilator days, ICU or hospital stay,

    percent discharged home, or mortality between theasynchronous and non-asynchronous subjects.

    Robinson BR et al. Respir Care 2013;58(11):1847–1855

    Pressure Control- A/C or PTV

    Assist/Control

    FSV

    PSV

    Inspiration: Flow or pressuretriggered

    Inspiration: Flow or pressuretriggered

    Pressure Support

    Breath Initiation (Inspiratory Triggering)

    Pressure Control- A/C or PTV

    Assist/Control

    FSVPSV

    Inspiration: Flow triggeredExpiration: Flow cycled

    Inspiration: Flow triggeredExpiration: Time cycled

    Pressure Support

    Breath termination (Cycling off)

    Pressure Ventilation - Waveforms

    Flow

    Pressure

    Tinsp.PIP

    PeakFlow

    25%

    Pressure Control Pressure Support

    Inspiratory time or cycle off criteria

    Ti given by thecycle off criteria

    Pressure Control versus Pressure Support

    Ti set

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    11/14

    11

    Flow Termination Sensitivity

    Nilsestuen J Respir Care 2005; 50:202-232

    ; : .

    . . ; :

    :

    I

    I .

    I

    . ; ;

    Early flow termination will reduce delivered Vt

    The non synchronized patient during Pressure-Support(inappropriate end-inspiratory ow termination criteria)

    Nilsestuen J Respir Care 2005;50:202–232.

    Pressure-Support and ow termination criteria

    Pressure-Support and ow termination criteria

    Increase in RR, reduction in VT, increase in WOB Nilsestuen JRespir Care 2005

    Termination Sensitivity = Cycle-off CriteriaHow to set in presence of a leak ?

    Flow

    Peak Flow (100%)

    TS 5%

    Tinsp. (eff.)

    Leak

    Set (max)Tinsp .

    Time

    TS 30%

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    12/14

    12

    F l o w

    % of MAX Flow

    PROS

    • Adapts to Patientsphysiology as itchanges

    • Adapts to Ti• Set max Ti• Delivers better V T

    CONS

    Leak• Asynchrony- Termination Sensitivity too

    high = breath starving- Termination Sensitivity too

    low = breath stack• Caution using PSV in devices with

    a FIXED Termination Sensitivity• COMPLICATED

    PSV: Termination Sensitivity

    Leak Cycle off with changes in mechanics

    10%

    Flow

    Time

    T1

    T2

    T3

    Compliance• Describes the ability of the

    respiratory system to moveduring an applied pressure

    • CRS = ! V/ ! P• Slope of the P-V curve

    CRS

    P-V loops and C RS

    Improved C RS

    WorseCRS

    GoodCRS =Zoneof VT

    Overdistension =Bad C RS

    Atelectasis =Bad C RS

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    13/14

    13

    Why isnt the P-V relationship linear?

    Ventilator pressures

    Alveolar volumeOpening and closing

    pressures

    Superimposed pressure

    0123456789

    c m H

    2 O

    InflationLimb

    DeflationLimb

    Crotti AJRCCM 2001

    Salazar & Knowles JAP 1964

    VOLUME HISTORY Oxyg

    Lung Volume+ Diffusion andperfusion

    Key points: Basic respiratory mechanicsduring mechanical ventilation

    • An understanding of time constants, complianceand resistance will help you individualise treatmentirrespective of modality

    • General principles of relationships between

    inspiratory/expiratory times and V T are: – Absolute max V T is determined by C RS and DP – Ti relative to t determines what max V T is delivered

    • Flow and pressure waveforms assist at thebedside unlocking the mechanics of the lung

    What do airway pressures mean?

    P alv = P pl + P tp

    Where Ptp is transpulmonary pressure, Palv is alveolar pressure, and Ppl is pleural pressure

    P tp = P alv - P pl

    Pplat 25 cm H 2O,PCV

    Pplat = Palv;Pplat " Transpulmonary Pressure (Ptp)

    0 cm H 2O

    P tp = 25

    Pplat 25 cm H 2O,PCV

    Pplat 25 cm H 2O,PCV

    Pplat = Palv;Pplat " Transpulmonary Pressure (Ptp)

    +10 cm H 2O0 cm H 2O

    P tp = 25 P tp = 15

  • 8/19/2019 1) Basic Respiratory Mechanics Relevant for Mechanical Ventilation

    14/14

    P tp = P aw x [E cw / (E tot )]

    when airway resistance is nil (static condition)P aw = P Pl + P tp

    E tot = E L + E cw

    Gattinoni LCritical Care 2004,8: 350-355

    soft stiff

    Pplat 25 cm H 2O,PCV

    Pplat 25 cm H 2O,PCV (PSV)

    Active inspiratory effort

    Pplat 25 cm H 2O,PCV

    Pplat = Palv;Pplat " Transpulmonary Pressure (Ptp)

    +10 cm H 2O0 cm H 2O -10 cm H 2O

    P tp = 25 P tp = 15 P tp = 35

    Agostini E J Appl Physiol 14:909-913, 1959

    Neonates require in general lower pressurethan children and adults because of respiratory

    system behaviour and not because they are small

    Pressure – Flow – Time - Volume

    . V o r V

    P or V

    Loops

    V /

    P / V

    . tCurves

    but we still have to decide about respiratory rate, tidalvolumes, pressure settings …

    … that have to be adapted to the patients respiratorymechanics

    The ventilator display can help us….