Page 1: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

R

R

Page 2: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

1003121RAG 6/21/00

Page 3: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

1

R

BiPAP systems are the subject of one or more of U.S. Patents #5148802, #5239995, #531937,#5433193, Canadian Patent #2, 024, 477, European Patent #EP0425092, German Patent#69021681.5-08, and other pending U.S. and foreign patents. BiPAP is a registered trademark ofRespironics, Inc.

© 2000 Respironics, Inc. All Rights Reserved

Page 4: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

2

R

Contents

Contents of the Package .............................................................................................. 5

1 Warnings, Cautions, and Notes ............................................................................. 6Warnings ................................................................................................................................................................................6Cautions .................................................................................................................................................................................7Notes ......................................................................................................................................................................................7Intended Use ..........................................................................................................................................................................8Contraindications ...................................................................................................................................................................8Precautions ............................................................................................................................................................................8Communication Statements for Optional Modem .................................................................................................................9Industry Canada Notice .......................................................................................................................................................10

2 Introduction to the Synchrony ............................................................................ 11Definitions ........................................................................................................................................................................... 11What is Bi-level Ventilation? ...............................................................................................................................................13What is the Synchrony? .......................................................................................................................................................14Symbols ...............................................................................................................................................................................15How to Contact Respironics ................................................................................................................................................15

3 A Tour of the Synchrony ...................................................................................... 16Overview .............................................................................................................................................................................16Control Panel .......................................................................................................................................................................17

Page 5: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

3

R

Panel Indicators ...................................................................................................................................................................18Display Indicators ................................................................................................................................................................18Breathing Circuit Connection ..............................................................................................................................................20Rear Panel ............................................................................................................................................................................20

4 Using the Synchrony............................................................................................. 21Installing the Air Filters .......................................................................................................................................................21Where to Place the Synchrony ............................................................................................................................................22Connecting the Breathing Circuit ........................................................................................................................................22Complete Synchrony Setup .................................................................................................................................................24Using DC Power ..................................................................................................................................................................25Plugging the Synchrony In (Using AC Power) ...................................................................................................................25Starting the Synchrony ........................................................................................................................................................26Using the Optional Ramp Feature .......................................................................................................................................26Operating the Synchrony .....................................................................................................................................................27

5 Alarms and What to Do ....................................................................................... 30High Priority Alarms ...........................................................................................................................................................32Medium Priority Alarms ......................................................................................................................................................35Low Priority Alarms ............................................................................................................................................................36

6 Troubleshooting .................................................................................................... 37Mask Discomfort and Corrective Actions ...........................................................................................................................38

7 Cleaning and Maintenance .................................................................................... 41Cleaning the Synchrony ......................................................................................................................................................41Cleaning or Replacing the Inlet Filters ................................................................................................................................42

Page 6: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

4

R

8 Using Oxygen with the Synchrony ...................................................................... 44

9 Using the Optional Modem.................................................................................. 46Sending Data at Startup .......................................................................................................................................................46Sending Data During Therapy .............................................................................................................................................49

10 Specifications......................................................................................................... 50

Page 7: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

5

R

Contents of the PackageThe BiPAP Synchrony package contains:

• BiPAP Synchrony• reusable circuit tubing• power cord• air filter cover• reusable foam filter• ultra-fine filter• User Guide

If any of the items are missing, please contact your homecare provider.

DisposableUltra-fine Filter

BiPAP SynchronyReusable Circuit

Tubing

Power Cord

Pac

kage

Con

tent

s

Your home care provider should have pre-assembled thetubing, exhalation port, and mask into a breathing circuit.

OPTIONS

• DC power cord• oxygen valve (required for oxygen use)• bacteria filter• humidifier• internal modem• modem cable• carrying case Reusable Foam

Filter

R

Air Filter Cover

User Guide

R

R

Optional ModemCable

Page 8: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

6

R

1 Warnings, Cautions, and Notes

Warnings

A Warning indicates the possibility of injury to you or theoperator of the device.

• This manual serves as a reference. The instructions inthis manual are not intended to supersede theinstructions of your care giver.

• You should read and understand this entire manualbefore using the Synchrony.

• The Synchrony is not intended to provide your totalventilatory requirement

• The prescription must only be adjusted by a trainedhome care provider.

• Use only the breathing circuit provided by your homecare provider.

• When using a breathing circuit that contains a mask withan integrated exhalation port or a circuit with a separateexhalation device, do not tape, seal, or otherwise blockthe vent openings. Doing so could result in suffocation.

• Oxygen supports combustion. Oxygen should not beused while smoking or in the presence of an open flame.

• Do not use the Synchrony in the presence of flammableliquids or gases. Do not clean the Synchrony withflammable fluids.

• If oxygen is used with the Synchrony, the oxygen flowmust be turned off when the Synchrony is not in use.

• Do not use the device in ambient temperatures above95 °F. If the device is used at ambient temperaturesabove 95 °F, the temperature of the breathing air mayexceed 106 °F, which could cause thermal irritation orinjury to your airway.

• Do not operate the Synchrony in direct sunlight or near aheating appliance because these conditions can increasethe temperature of the air coming out of the Synchrony.

• When the Synchrony is used with a humidifier, positionthe humidifier such that the water level in the humidifieris lower than you, and the humidifier is on the samelevel or lower than the Synchrony.

• Repairs and adjustments must be performed byRespironics-authorized service personnel ONLY.Unauthorized service could cause injury, invalidate thewarranty, or result in costly damage.

• Periodically inspect electrical cords for damage or signsof wear.

• To avoid electrical shock, unplug the Synchrony beforecleaning it.

Warnings

Page 9: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

7

R

Cau

tions

, Not

es

Cautions

A Caution indicates the possibility of damage to thedevice.

Caution: Federal law restricts this device to sale by or onthe order of a physician.

• The Synchrony may only be operated at temperaturesbetween 41 °F and 95 °F.

• A properly installed, undamaged reusable foam inletfilter is required for proper operation.

• Do not immerse the Synchrony or allow any liquid toenter the case or the inlet filter.

• Condensation may damage the Synchrony. Always allowthe Synchrony to reach room temperature before use.

Warnings continued• Immediately report any unusual chest discomfort,

shortness of breath, or severe headache toyour physician.

• If you detect any unexplained changes in the perfor-mance of the Synchrony, if the Synchrony is dropped ormishandled, if water is spilled into the enclosure, or ifthe enclosure is broken, seek the assistance of yourhome care provider.

Additional Warnings, Cautions, and Notes arelocated throughout this manual.

NOTE: Illustrations are for reference only. Occasionally,cosmetic changes may be made to the productthat do not affect the performance or specifica-tions of the product. These kinds of changes donot warrant a reprinting of this manual.

Notes

• Throughout this manual, you will see the symbol UPGRADE .When you see this symbol, it means that the featurebeing discussed is only available with the SynchronyS/T upgrade package. Your Synchrony may or may notbe provided with this upgrade. Contact your home careprovider for details.

Page 10: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

8

R

Intended Use

The Synchrony is intended to provide non-invasiveventilation in adult patients (>30 kg) for the treatment ofrespiratory insufficiency (a condition in which the patientcan maintain ventilation without mechanical support forsome period of time) or obstructive sleep apnea. Thisdevice may be used in the hospital or home.

The Synchrony is intended for use with nasal masks andfull-face masks as recommended by Respironics.

Contraindications

The Synchrony should not be used if you have severerespiratory failure without a spontaneous respiratory drive.

If any of the following conditions apply to you, consultyour physician before using the Synchrony ventilator:

• inability to maintain a patent airway or adequately clearsecretions.

• at risk for aspiration of gastric contents.

• diagnosed with acute sinusitis or otitis media.

• allergic or hypersensitive to the mask materials wherethe risk from allergic reaction outweighs the benefit ofventilatory assistance.

• epistaxis, causing pulmonary aspiration of blood.

• hypotension.

Precautions

Immediately report any unusual chest discomfort, short-ness of breath, or severe headace to your physician.

If skin irritation or breakdown develops from the mask,refer to the mask instructions for appropriate action.

The following are potential side effects of noninvasivepositive pressure therapy:

• ear discomfort

• conjunctivitis

• skin abrasions due to noninvasive interfaces

• gastric distention (aerophagia)

Intended Use, C

ontraindications

Page 11: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

9

R

Communication Statements for OptionalModem

TYPES OF SERVICE

The Synchrony optional internal modem is designed to beused on standard device telephone lines. It connects to thetelephone line by means of a standard jack called theUSOC RJ-11C (or USOC FJ45S). Connection to tele-phone-company-provided coin service (central officeimplemented systems) is prohibited. Connection to partylines service is subject to state tariffs.

TELEPHONE COMPANY PROCEDURES

The goal of the telephone company is to provide you withthe best service it can. In order to do this, it may occasion-ally be necessary for them to make changes in theirequipment, operations or procedures. If these changesmight affect your service or the operation of your equip-ment, the telephone company will give you notice, inwriting, to allow you to make any changes necessary tomaintain uninterrupted service. In certain circumstances, itmay be necessary for the telephone company to requestinformation from you concerning the equipment that youhave connected to your telephone line. Upon request of thetelephone company, provide the FCC registration numberand the ringer equivalence number (REN); both of these

items are listed on the equipment label. The sum of all theRENs on your telephone lines should be less than five inorder to assure proper service from the telephone com-pany. In some cases, a sum of five may not be useable on agiven line.

I F PROBLEMS ARISE

If any of your telephone equipment is not operatingproperly, you should immediately remove it from yourtelephone line, as it may cause harm to the telephonenetwork. If the telephone company notes a problem, theymay temporarily discontinue service. When practical, theywill notify you in advance of the disconnection. If advancenotice is not feasible, you will be notified as soon aspossible. When you are notified, you will be given theopportunity to correct the problem and informed of yourright to file a complaint with the FCC. Contact yourtelephone company if you have questions about yourphone line. In the event repairs are ever needed on theSynchrony, they should be performed by Respironics, Inc.or an authorized representative of Respironics, Inc. Forinformation contact Respironics at the location shown onpage 13.

Com

mun

icat

ions

Sta

tem

ents

Page 12: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

10

R

Industry Canada Notice

NOTICE: The Industry Canada Label identifies certifiedequipment. This certification means that the equipmentmeets telecommunications network protective, operational,and safety requirements as prescribed in the appropirateTerminal Equipment Technical Requirements documents.The Department does not guarantee the equipment willoperate to the user’s satisfaction.

Before installing this equipment, users should ensure thatit is permissible to be connected to the facilities of thelocal telecommunications company. The equipment mustalso be installed using an acceptable method of connec-tion. The customer should be aware that compliance withthe above conditions may not prevent degradation ofservice in some situations.

Repairs to certified equipment should be coordinated by arepresentative designated by the supplier. Any repairs oralterations made by the user to this equipment, or equip-ment malfunctions, may give the telecommunicationscompany cause to request the user to disconnect theequipment.

Users should ensure for their own protection that theelectrical ground connections of the power utility, tele-phone lines and internal metallic water pipe system, ifpresent, are connected together. The precaution may beparticularly important in rural areas.

CAUTION: Users should not attempt to make suchconnections themselves, but should contact the appropriateelectric inspection authority, or electrician, as appropriate.

Ringer Equivalence Number (REN): The REN assignedto each terminal device provides an indication of themaximum number of terminals allowed to be connected toa telephone interface. The termination on an interface mayconsist of any combination of devices subject only to therequirement that the sum of the Ringer EquivalenceNumbers of all the devices does not exceed 5.

Com

munications S

tatements

Page 13: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

11

R

Intr

oduc

tion

Definitions

Apnea—A condition marked by the cessation ofspontaneous breathing.

CPAP—Continuous Positive Airway Pressure. TheSynchrony produces a single pressure level.

EPAP—Expiratory Positive Airway Pressure

High Priority Alarm —Alarm signal requiring animmediate response.

IPAP—Inspiratory Positive Airway Pressure

Low Priority Alarm —Alarm signal requiring operatorawareness.

Medium Priority Alarm— Alarm signal requiring aprompt response.

Low Minute Ventilation —A condition in which you arenot receiving a specified volume of air on a per minutebasis.

OSA—Obstructive Sleep Apnea.

Pressure Control (PC)—This is an optional bi-levelmode which responds to your inhalation by increasing thepressure. The length of the inhalation is controlled by theSynchrony. If you do not start inhaling within a set time,the Synchrony will automatically start inhalation.

Ramp—Ventilator Ramp, a feature that may increaseyour comfort when therapy is started. The Synchronyoutput starts at a low pressure (EPAP) and is increased ateach breath until IPAP is reached.

Rise Time—Rise time is the time it takes for theSynchrony to change from EPAP to IPAP. You can adjustthis time for your comfort. See Figure 2-1.

RR—Respiratory Rate

Spontaneous (S)—This is a bi-level mode whichresponds to both your inahalation and exhalation byincreasing pressure when you start to inhale anddescreasing pressure when you start to exhale. There is noautomatic delivery of a breath should you not inhale.

2 Introduction to the Synchrony

UPGRADE

Page 14: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

12

R

UPGRADESpontaneous/Timed (S/T)—This is an optional bi-levelmode which responds to both your inahalation andexhalation by increasing pressure when you start to inhaleand descreasing pressure when you start to exhale. If youdo not start inhaling within a set time, the Synchrony willautomatically start inhalation. When the Synchrony startsinhalation, it controls the time of inhalation andautomatically decreases the pressure for exhalation withina set time.

Timed (T)—This is an optional bi-level mode in whichthe Synchrony controls both inhalation and exhalationindependent of spontaneous breathing.

Introduction

UPGRADE

Page 15: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

13

R

What is Bi-level Ventilation?

Bi-level ventilation with the Synchrony VentilatorySupport System helps you to breathe by supplying twolevels of air pressure. The Synchrony provides a higherpressure—known as IPAP (Inspiratory Positive AirwayPressure)—when you inhale, and a lower pressure—known as EPAP (Expiratory Positive Airway Pressure)—when you exhale. The higher pressure makes it easier toinhale and the lower pressure makes it easier for you toexhale.

The Synchrony can also provide a single pressure level,known as CPAP (Continuous Positive Airway Pressure).

If you require a higher level of ventilatory support, theSynchrony can be upgraded to provide modes ofventilation that reinforce your breathing with machine-triggered breaths.

IPAPPressure

EPAPPressure

The Synchrony ProvidesTwo Levels of Ventilation.

Rise TimeTime

Figure 2-1 BiPAP Breathing Levels

You can adjust the Rise Time to make the pressure changemore comfortable. See page 27.

Intr

oduc

tion

NOTE: Throughout this manual, you will see the symbolUPGRADE . When you see this symbol, it means thatthe feature being discussed is only available withthe Synchrony S/T upgrade package. YourSynchrony may or may not be provided with thisupgrade. Contact your home care provider fordetails.

Page 16: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

14

R

Intr

oduc

tion

What is the Synchrony?

The Synchrony, shown in Figure 2-2, supplies air pressurethrough a breathing circuit. The circuit, shown in Figure2-3, consists of:

• circuit tubing to deliver the air from the Synchrony toyour interface (e.g., mask)

• a mask or other interface to deliver the prescribedpressure to your nose or nose and mouth, depending onwhich interface has been prescribed for you

• an exhalation port to vent exhaled air from the circuit

NOTE: The exhalation port may be part of the mask ormay be part of a separate exhalation device.

The system senses your breathing effort and changespressure levels when you inhale and exhale depending onthe mode of operation.

The Synchrony can operate on AC or DC power. The DCpower option is not intended as a battery backup. WhenDC power is obtained from a vehicle battery, the Syn-chrony should not be used while the engine of the vehicleis running.

Figure 2-2 The Synchrony

R

CircuitTubing

Exhalation Device(Typical)

Exhalation Ports(one port on eachside of elbow)

Patient Interface(Typical)

Circuit with SeparateExhalation Device

Circuit with Mask withIntegrated Exhalation Ports

Figure 2-3 Typical Breathing Circuits

Page 17: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

15

R

Symbols

The symbols shown at right are used on the Synchrony andthroughout this manual.

How to Contact Respironics

To have your unit serviced, contact your home careprovider. If you need to contact Respironics directly, call1-800-345-6443 or use the following address:

Sym

bols

AC Power Indicator

DC Power Indicator

Alarm Indicator

Type BF Applied Part

Class II for Protection Against Electric Shock

Attention, consult accompanying documents

Standby Key (Start/Stop)

Up Key

Down Key

Enter Key

Ramp Start/Alarm Silence Key

Symbol Meaning

Page 18: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

16

R

3 A Tour of the Synchrony

Overview

Figure 3-1 shows the location of:

• control panel• breathing circuit connection• carry handle• optional oxygen valve

Tour of the Synchrony

Figure 3-1 The Synchrony Front and Top

ControlPanel

BreatingCircuit

Connection

CarryHandle

OxygenValve

(Optional)

R

Page 19: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

17

R

Control Panel

The control panel contains the controls and indicators youwill need when you are using the Synchrony:

• AC power indicator• DC power indicator• Standby key• Control keys (Down, Up, Enter)• Ramp/Alarm Silence key• Yellow LED Alarm indicator• Red LED Alarm indicator• Display screen

The AC and DC power indicators light when theSynchrony is connected to AC or DC power.

The Standby key starts and stops the Synchrony.

The Control keys allow you to display information and setcontrols.

The Ramp/Alarm Silence key starts the ventilation ramp(see page 24) and silences an audible alarm for oneminute.

The LED Alarm Indicators alert you when a problemoccurs. See Chapter 5 for an explanation of what to dowhen an alarm is activated.

Figure 3-2 The Synchrony Control Panel

R

RedAlarm Indicator

YellowAlarm Indicator

AC PowerIndicator

DC PowerIndicator

StandbyKey

EnterKey

DownKey

UpKey

Ramp/Alarm Silence

Key

DisplayScreen

The Display shows you the current pressure controlsettings and displays alarm messages.

Con

trol

Pan

el

NOTE: If the display backlight is off, the first press of acontrol key ( , , or ) turns the backlight onand does nothing else. When you are instructed topress a key, this manual assumes that thebacklight is already on. When you are instructedto press a key, this manual assumes that thebacklight is already on.

Page 20: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

18

R

Indicators

AC Power indicator—a green LED that lights when ACpower is connected.

DC Power indicator—a green LED that lights when DCpower is connected.

Panel Indicators

Red Alarm Indicator:• When flashing indicates a high priority alarm.

• When on continuously indicates a loss of power or asilenced high priority alarm.

Yellow Alarm Indicator:• When flashing indicates a medium priority alarm.

• When on continuously indicates a low priority alarm ora silenced medium priority alarm.

Red

Yellow

Display Indicators

Activity Indicator ( )—This indicator shows whichpressure level you are receiving during therapy. It alsoshows when the Synchrony starts a timed breath (upgradeonly) and when an alarm is still active for the displayedmessage.

MODE: S 1/2

IPAP 15 cm H2O

EPAP 9 cm H2O

RR 20 BPM

Indicates the pressureis at the EPAP level

MODE: S/T 1/4

IPAP 15 cm H2O

EPAP 9 cm H2O

RR 20 BPM

Indicates the current breathwas started by the unit

ALARM: 1/2

Pat ient Disconnect

Indicates an active alarm message

The display uses indicators to show you the pressure levelsand guide you through the screens.

UPGRADE

Page 21: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

19

R

Selection Indicator ( )—This indicator appears next toan item that can be selected with the ENTER () key. Theindicator is moved by pressing the UP () or DOWN( ) key.

CONTROLS: 2/2

Rise Time: 3

Indicates that Rise Time will be selected for change

Ramp Indicator ( )—This indicator shows whenVentilation Ramp has been started by pressing the RAMP/SILENCE ( ) key.

MODE: S 1/2

IPAP 16 cm H2O

EPAP 9 cm H2O

RR 20 BPM

Indicates that theVentilation ramp has

been initiated

Navigation Indicator ( )—This indicator, next to a pagenumber, shows that the and keys can be used tochange pages. If it appears next to a value, it flashes andshows that the UP and DOWN keys can be used to changethe value.

MODE: S 1/2

IPAP 16 cm H2O

EPAP 9 cm H2O

RR 20 BPM

Indicates that the pageis scrolled with the

UP and DOWN keys

CONTROLS: 2/2

Rise Time: 3 sec

Dial Provider

Indicates that the Rise Timevalue can be changed

Progress Indicator ( )—This indicator empties from topto bottom to indicate the progress of certain tasks.

SYSTEM SELF TEST

VER:1.00 10.0 HOURS

Indi

cato

rs

Page Number—The current page and total number ofpages in a series of screens is indicated in the upper rightof the screen.

MODE: S 1/2

IPAP 15 cm H2O

EPAP 9 cm H2O

RR 20 BPM

Indicates that thisis page 1 of 2

Page 22: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

20

R

Breathing Circuit Connection

Figure 3-3 shows where the circuit tubing connects tothe Synchrony.

Rear Panel

Figure 3-4 shows the rear panel.

The DC inlet is where the DC plug is connected if you arerunning the Synchrony from a battery source.

The communications port is used by the home careprovider.

The DC cord retainer holds the DC cord. Used to keepthe plug from coming out of the Synchrony.

The telephone jack is where the telephone line is pluggedin (if you have the optional modem) when you are asked tosend the home care provider data from the Synchrony.

The AC inlet is where the AC receptacle is connected ifyou are running the Synchrony from house voltage.

The air filter cover retains the air filters and helps toensure quiet operation.

Circuit C

onnections, Rear P

anel

Figure 3-3 Breathing Circuit Connection

Figure 3-4 Rear Panel

Breathing CircuitConnection

R

ACInlet

DCInlet

CommunicationsPort

DC CordRetainer

Air FilterCover

TelephoneJack

(optional internalmodem only)

Page 23: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

21

R

4 Using the SynchronyInstalling the Air Filters

Inst

allin

g th

e A

ir F

ilter

s

R

Reusable GrayFoam Filter

Disposable Ultra-fineFilter

Air FilterCover

Figure 4-1 Filters Installed

R

R

Figure 4-2 Filter Cover Installed

CAUTION: A properly installed, undamaged foamfilter is required for proper operation.

The Synchrony uses a gray foam filter that iswashable and reusable, and an optional white ultra-finefilter that is disposable. One filter of each kind is suppliedwith the Synchrony.

If your home care provider did not install the inlet airfilters, you must install at least the gray foam filter beforeusing the Synchrony.

STEP 1 Place the foam filter on top of theultra-fine filter (if using the ultra-fine filter).

STEP 2 Push the filters down into the air inlet recess atthe rear of the Synchrony as shown in Figure 4-1.

STEP 3 Place the bottom of the air filter cover into thebottom of the air inlet opening, making sure thatthe catches engage the lip of the opening.

STEP 4 As shown in Figure 4-2, swing the top of thecover into place and press down on it to engagethe cover catches in the air inlet opening. See Chapter 7 to clean or replace the filters.

Page 24: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

22

R

STEP 1 Connect one end of the circuit tubing to theoutlet of the bacteria filter (if using one) andconnect the inlet of the bacteria filter to thelarge connector on the Synchrony as shown inFigure 4-3.

If you are not using a bacteria filter, connect theend of the circuit tubing directly to the outletconnector on the Synchrony.

Connecting the Breathing Circuit

Figure 4-3 Connect the Tubing to the Outlet

Placing the S

ynchrony for Use

CircuitTubing

BacteriaFilter

(Optional)

R

NOTE: Follow the recommendations of your home careprovider for using the optional bacteria filter.

Where to Place the Synchrony

Place the Synchrony on its base somewhere within easyreach of where you will be using it. Make sure that the airinlet on the rear of the unit is not blocked. Place the uniton a hard, flat surface. If you block the air flow around theunit, the unit may not work properly.

If you are using a humidifier, be sure to place thehumidifier and unit on a placemat or other waterproofmaterial to protect your furniture from moisture. Positionthe humidifier such that the water level in the humidifier islower than you, and the humidifier is on the same level orlower than the ventilator. See the humidifier instructionsfor the complete setup.

Page 25: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

23

R

STEP 2 Connect the open end of the circuit tubing to theexhalation port as shown in Figure 4-4. If youare using a mask that has the exhalation portbuilt into the mask, connect the tubing directlyto the mask and skip to Step 4.

STEP 3 Connect the exhalation port to the maskconnector as shown in Figure 4-5.

STEP 4 Attach the headgear to the mask—see theinstructions that came with your headgear.

ExhalationPort

MaskConnector

Mask orOther

Interface

Figure 4-5 Connect the Mask

Con

nect

ing

the

Bre

athi

ng C

ircui

t

ExhalationPort

CircuitTubing

Figure 4-4 Connect the tubing to the Exhalation Port

Page 26: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

24

R

Complete Synchrony Setup

Figure 4-6 shows the completed breathing circuit setup forthe Synchrony.

Com

plete Setup

Figure 4-6 Synchrony Complete Breathing Circuit

CircuitTubing

Exhalation Port(Typical)

Patient Interface(Typical)

BacteriaFilter

(Optional)

R

Page 27: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

25

R

Plugging the Synchrony In (Using ACPower)

STEP 1 Plug the receptacle end of the AC cord into theAC inlet of the Synchrony as shown in Figure4-7.

STEP 2 Insert the plug end of the AC cord into thewall outlet as shown in Figure 4-7.

WARNING: Never plug the Synchrony AC cord intoan outlet that is controlled by awall switch.

Using DC Power

Plu

ggin

g th

e S

yn-

chro

ny In

STEP 1 Plug the receptacle end of the DC cord into theDC inlet of the Synchrony as shown in Figure4-8.

STEP 2 Leaving some slack in the DC cord, push theDC cord into the DC cord retainer.

STEP 3 Insert the adaptor end of the DC cord into theappropriate DC outlet. See the instructions thatcame with the DC cord for proper DCconnections.

R

R

R

DC CordRetainer

Figure 4-7 Plugging in the AC

Figure 4-8 Plugging in the DC

Page 28: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

26

R

Starting the Synchrony

STEP 1 Press the STANDBY ( ) key. The Synchronywill sound two beeps, briefly light the alarmindicators, and display the startup screenfollowed by the self-test screen. This is theinternal test performed by the Synchrony.

STEP 2 Put on your mask assembly when the air startsto flow.

NOTE: If the Synchrony does not sound the two beeps orilluminate the alarm indicators, do not try to usethe Synchrony. Contact your home care provider.

STEP 3 Make sure that no air is leaking from your maskinto your eyes. If it is, adjust the mask andSoftcap or headgear until air stops leaking intoyour eyes. See the instructions that came withyour mask.

STEP 4 If you are using the Synchrony while sleeping,try placing the tubing from the Synchrony overyour headboard. This may reduce tension on themask.

Starting the S

ynchrony

NOTE: A small amount of mask leak is normal andacceptable. Large mask leaks or eye irritationfrom an air leak should be corrected as soon aspossible.

STEP 5 Relax. Take normal, relaxed breaths throughyour nose.

NOTE: If you are having trouble with your mask, seeChapter 6 for some suggestions.

Using the Optional Ramp FeatureThe Synchrony is equipped with an optional ramp featurethat your home care provider can turn on or off. The Rampfeature starts the pressure at the EPAP level. The pressureis increased at each breath until IPAP is reached.

To start the Ramp feature, follow Steps 1 to 5 above, thenpress . The Ramp indicator will be shown at the topof the display. The diagram below illustrates how the rampchanges the pressure levels:

IPAPLEVEL

EPAPLEVEL

1 Breath

RampIncrement

(Set byHome CareProvider)

Page 29: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

27

R

Operating the Synchrony

MONITORING SCREENS

The screen that you see normally is called the Monitoringscreen. It shows you the current pressure measurementsand breathing rate.

The CPAP mode Monitoring screen is shown inFigure 4-9. If your Synchrony is not using the optionalmodem, the screen will show “1/1” in the upper right.

MODE: CPAP 1/2CPAP 10 cm H2ORR 20 BPM

Figure 4-9 CPAP Monitoring Screen

The S, S/T, PC, and T mode Monitoring screen is shown inFigure 4-10.

MODE: S 1/2IPAP 10 cm H2OEPAP 5 cm H2ORR 20 BPM

Figure 4-10 S, S/T, PC, and T Mode Monitoring Screen

For the S/T, PC and T modes, “S” is replaced with “S/T,”“PC,” or “T.”

Note the indicator next to IPAP. It indicates the pressureis currently at the IPAP level. It will move from IPAP toEPAP as the pressure changes. Another indicator willappear next to the RR if the Synchrony starts a breath (Theunit will start a breath only in the optional S/T, PC, and Tmodes).

The upper right of the screen indicates that this is Page 1of 2, and the indicator shows that the or key willmove to Page 2 of 2.

Mon

itorin

g S

cree

nsWhen the Synchrony is operating, you can use twoscreens: the Monitoring screen and the Control screen.

The ramp feature can provide more comfort for youbecause of the gradual increase to full pressure.

Page 30: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

28

R

CONTROL SCREENS

If the modem is being used, you can press the key whilein the Monitoring screen. The Synchrony displays Page 2of 2, called the Control screen. In the CPAP mode, theControl screen contains one option as shown inFigure 4-11.

CONTROLS: 2/2 Dial Provider

Figure 4-11 CPAP Control Screen.

The selected control is the Dial Provider command. Withthis command selected, when you press , the Synchronyautomatically dials your home care provider to sendinformation about the Synchrony.

NOTE: The word “Provider” may be replaced with thename of your provider or the provider’s phonenumber.

The modem settings have been preset by your home careprovider. See Chapter 9 for instructions on installing thetelephone line into the rear panel of the Synchrony.

In the S, S/T, PC, and T modes, the Control screencontains two options (one if the modem is not beingused), as shown in Figure 4-12.

CONTROLS: 2/2 Rise Time: 3 Dial Provider

Figure 4-12 S, S/T, PC, and T Mode Control Screen

The selected control is the Rise Time parameter. With thiscommand selected, when you press , the parameter isavailable for change. See Page 10 for a description ofRise Time.

The next page shows you how to change the Rise Time.

Control S

creens

NOTE: If the display backlight is off, the first press of acontrol key ( , , or ) turns the backlight onand does nothing else. When you are instructed topress a key, this manual assumes that thebacklight is already on. When you are instructedto press a key, this manual assumes that thebacklight is already on.

Page 31: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

29

R

To change the Rise Time:

STEP 1 Use or to select the Rise Time parameter.

CONTROLS: 2/2 Rise Time: 3 Dial Provider

CONTROLS: 2/2 Rise Time: 3

CONTROLS: 2/2 Rise Time: 6

STEP 2 Press . The result is shown below.

STEP 3 Press or to change the value. The range is1 to 6; a setting of 1 is the fastest rise time, 6 isthe slowest.

STEP 4 Press when complete.

Allow the Synchrony to run several cycles to see if it feelsmore comfortable.

CONTROLS: 2/2 Rise Time: 6 Dial Provider

Ris

e T

ime

Page 32: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

30

R

Alarm

s

5 Alarms and What to Do

This chapter describes the Synchrony alarms and what youshould do if you get an alarm.

The Synchrony provides three alarm levels: high, medium,and low priority.

High priority alarms require immediate response. Thealarm signal consists of a flashing red alarm indicator anda sound that is either continuous, beeps, or repeats apattern of three beeps followed by two beeps. The displayhas the message “ALARM” at the top of the screen.

Medium priority alarms require prompt response. Thealarm signal consists of a flashing yellow alarm indicatorand a sound that repeats a pattern of three beeps. Thedisplay has the message “WARNING” at the top ofthe screen.

Low priority alarm require your awareness. The alarmsignal consists of a continuous yellow alarm indicator anda sound that repeats a pattern of two beeps. The displayhas the message “INFO” at the top of the screen.

The alarms indicators are LEDs located on the right side ofthe front panel, as shown in Figure 5-1.

Figure 5-1 Alarm Indicator Locations

R

High PriorityAlarm LED

Low/Medium PriorityAlarm LED

Page 33: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

31

R

The following example applies to most alarm conditions.You should follow these steps unless otherwise directed bythe alarm tables.

MODE: S 1/4IPAP 15 cm H2OEPAP 6 cm H2ORR 20 BPM

Alarm is resetDisplay returns toscreen that wasshowing at time ofalarm

STEP 1 Look at the alarm indicators and listen to thealarm sound.

STEP 4 Look up the alarm in the alarm tables andperform Your Action. For High and MediumPriority Alarms, the indicator will go offwhen the alarm condition is corrected.

STEP 2 Look at the display for text.

If more than one alarm is listed, the indicator showswhich alarm is still active.

STEP 3 Press the RAMP/SILENCE key.

ALARM 1/1 Pat ient Disconnect

(Note color and whether LED isflashing or solid)

ALARM 1/1 Pat ient Disconnect

CLEAR

indicator is offAlarm LED issolid

ALARM 1/1 Pat ient Disconnect

CLEAR

Alarm CLEARScreen is dis-playedAlarm LED issolid

indicator shows what alarmis activated.

STEP 5 Press the Enter key to clear the alarm.

Ala

rms

Page 34: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

32

R

Alarm

s Cannot be silenced;remove power fromunit and contacthome care provider.

Blank Screen Shuts down Synchrony FailureRed Flash

Blank Screen Power was lostwhile the unit wasproviding therapy.

Press to silence,restore power.

Shuts downRed Solid

Apnea An apnea eventoccurred duringtherapy.

Operates Report alarm tohome care provider.

Red Flash

Low PressureSupport

Malfunctioningunit.

Operates Check for thefollowing: dirty inletfilters, blocked airintake, excessiveleak in circuit. Ifalarm continues, callyour home careprovider.

Red Flash

Your ActionAlarmSound

AlarmLED

DisplayMessage

What theUnit Does

PossibleCause

High Priority Alarms

Page 35: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

33

R

Ala

rms

Your ActionAlarmSound

AlarmLED

DisplayMessage

What theUnit Does

PossibleCause

High Pressure Malfunctioningunit.

Operates Nothing. If alarmcontinues call yourhome care provider.

Red Flash

Low MinuteVentilation

OperatesRed Flash If you are havingrespiratory distressor breathingdifficulties, callyour home careprovider.

Check for thefollowing: dirtyinlet filters, blockedair intake, excessiveleak in circuit. Ifalarm continues,call your home careprovider.

Decreased breathingpattern or breathingrate.

Excessive leak orblockage.

Do not use the unit,call your home careprovider.

InvalidPrescriptionCallProvider

Shuts down The prescriptiondata is corrupted.

Red Flash

Page 36: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

34

R

Alarm

s

Your ActionAlarmSound

AlarmLED

DisplayMessage

What theUnit Does

PossibleCause

VentilatorInoperative(DisplaysCodes)

Internal devicefailure.

Shuts down Remove powerfrom the unit. Callyour provider.

Red Flash

VentilatorInoperativeBatteryFailure

Battery isdischarged.

Shuts down Remove DC powersource from theunit, replace thebattery, restorepower to the unit;or seek reliable ACpower source.

Red Flash

PatientDisconnect

Breathing circuit isdisconnected or hasa large leak.

Operates Reconnect thecircuit or fix theleak.

Red Flash

Low Pressure Excessive leak orblockage.

Operates Check for thefollowing: dirty inletfilters, blocked airintake, excessiveleak in circuit. Ifalarm continues, callyour home careprovider.

Red Flash

Page 37: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

35

R

Medium Priority Alarms

BatteryVoltage TooHigh

Wrong ormalfunctioningbattery or charger.

Operates Remove DC powerfrom the unit, useon AC or call homecare provider.

YellowFlash

YellowFlash

Low Battery Battery nearlydischarged.

If on battery, findanother; if on AC,exchange the batteryand charge.

Operates

Your ActionAlarmSound

AlarmLED

DisplayMessage

What theUnit Does

PossibleCause

Ala

rms

Page 38: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

36

R

Low Priority Alarms

On start-up,informationmessage that batteryis being used toprovide power.

Battery inUse

Operates Press to confirmthat you wish tooperate on batterypower.

YellowSolid

Call forService(Codes aredisplayed)

Minor internal error. Continue to use,write down errorcode, call yourhome care provider.

OperatesYellowSolid

MomentaryLoss ofPower

Power was lost forless than 30 secondswhile the unit wasproviding therapy.

Press to clear theinformationmessage.

OperatesYellowSolid

PowerFailure,Battery inUse

Line (AC) powerwas lost, the unit isnow operating onbattery (DC) power.

Check AC power,seek reliable powersource.

OperatesYellowSolid

Your ActionAlarmSound

AlarmLED

DisplayMessage

What theUnit Does

PossibleCauseA

larms

Page 39: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

37

R

6 Troubleshooting

This chapter lists problems you could experience withyour Synchrony or mask, and presents possible solutions.

Problem

Does not operate when you press .

The air out of the mask is muchwarmer than usual.

Why it Happened

If the power LED is off—no power atthe outlet or the Synchrony isunplugged.

If the power LED is on—problems inthe Synchrony.

Inlet filters may be dirty.

The Synchrony may be operatingdirect sunlight or near a heater.

The room temperature may betoo high.

What You Can Do

Check the outlet power and verifythat the Synchrony is plugged in.

Contact your home care provider forservice.

Clean or replace the inlet air filters asdescribed in Chapter 7. If the Syn-chrony is in direct sunlight or near aheater, move it. If the problempersists, contact your home careprovider.

Lower the room temperature. If theproblem persists, contact your homecare provider.

Trou

bles

hoot

ing

Page 40: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

38

R

Mask P

roblems

Mask Discomfort and Corrective Actions

Problem

Mask feels uncomfortable to wear.

Significant air leakage around themask.

Redness occurs when the maskcushion comes in contact withthe skin.

Why it Happened

1. Improper headgear adjustment.

2. Improper mask fitting.

1. Improper headgear adjustment.

2. Improper mask fitting.

1. Improper mask fitting.

2. Improper mask cleaning.

What You Can Do

Check the headgear adjustment asdescribed in the headgearinstructions.

Contact your home care provider fora refitting or a different size of mask.

Check the headgear adjustment asdescribed in the headgearinstructions.

Contact your home care provider fora refitting or a different size of mask.

Contact your home care provider fora refitting or a different size of mask.

Be sure to thoroughly rinse aftercleaning to remove residue. See themask cleaning instructions.

Page 41: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

39

R

Problem

Redness occurs when the maskcushion mask accessory comes incontact with the skin.

Sore or dry eyes.

Runny nose.

Why it Happened

3. Irritation or allergic reaction tomask material.

1. Mask not positioned properly.

2. Improper mask fitting.

Nasal reaction to air flow.

What You Can Do

Use a barrier between your skin andthe mask, such as 3M’s Microfoam®

or Squibb’s Duoderm®.

Check the headgear adjustment asdescribed in the headgearinstructions.

Contact your home care provider fora refitting or a different size of mask.

Call your doctor.

Mas

k P

robl

ems

Page 42: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

40

R

Problem

Throat or nose dryness.

Nasal, sinus or ear pain.

Why it Happened

1. Air is too dry.

2. Improper mask fitting.

Sinus or middle ear infection

What You Can Do

1. Increase the room humidity.

2. Consult with your doctor or homecare provider about using aRespironics humidifier with theSynchrony.

Stop using the Synchrony andcontact your doctor.

Mask P

roblems

Page 43: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

41

R

7 Cleaning and Maintenance

Cleaning the Synchrony

Before cleaning or performing any routine maintenance,always be sure the Synchrony is not operating anddisconnect the Synchrony from the power source.

NOTE: The following cleaning instructions are for theSynchrony only. To clean the accessories, refer toeach accessory’s instruction sheet.

CAUTION: Do not immerse the Synchrony or allowany liquid to enter the enclosure, inletfilter, or any openings.

Clean the front panel and exterior of the enclosure asneeded using a cloth dampened with water and a milddetergent. Allow the Synchrony to dry completely beforeplugging in the power cord.

Gently wash the reusable circuit tubing in a solution ofwarm water and a mild detergent. Rinse thoroughly andallow to air dry.

Cle

anin

g an

d M

aint

enan

ce

Page 44: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

42

R

Inlet Filters

Cleaning or Replacing the Inlet Filters

The Synchrony has two removable filters at the air inlet.The gray foam filter is washable and reusable. Theoptional white ultra-fine filter is disposable.

NOTE: Dirty inlet filters may cause high operatingtemperatures and may affect Synchrony perfor-mance. Regularly examine the inlet filters asneeded for integrity and cleanliness.

Step 1 Verify that the Synchrony is not operating, anddisconnect the power cord from the wall outlet.

Step 2 As shown in Figure 7-1, remove the filter panelby pressing down on the top of the panel torelease the tabs, then swinging the panel outfrom the Synchrony’s body.

Step 3 Remove the filters from the enclosure as shownin Figure 7-2. The top filter is the reusable grayfoam filter. The bottom filter is the optionaldisposable white ultra-fine filter.

Step 4 Check the filters to see if they are dirty or torn.

Figure 7-1 Removing the Air Filter Cover

Figure 7-2 Removing the Air Filters

R

R

Reusable FoamFilter

Disposable Ultra-fineFilter

R

R

Page 45: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

43

R

Step 5 If needed, wash the foam filter in warm waterand a mild detergent. Rinse the filter thoroughlyto remove all detergent residue. Allow the filterto completely dry before reinstalling it. If thefoam filter is torn, replace it.

Step 6 If the ultra-fine filter is dirty or torn, replace it.

Step 7 Reinstall the filters, with the ultra-fine filter onthe bottom. Slide the filters into the air inlet atthe rear of the Synchrony and push them downinto the recess.

Step 8 Replace the filter panel.

Contact your home care provider to order filters.

NOTE: To clean the breathing circuit accessories, refer toeach accessory’s instruction sheet.

Inle

t Filt

ers

Page 46: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

44

R

Using O

xygen

8 Using Oxygen with theSynchrony

If your doctor has prescribed oxygen to be used with yourSynchrony, follow these instructions to connect oxygen tothe breathing circuit.

WARNING: If you are using oxygen, yourSynchrony must be equipped with theoptional oxygen valve, shown in Figure8-1. Failure to use the oxygen valve couldresult in a fire hazard. Contact your homecare provider if you require an oxygenvalve.

WARNING: Oxygen supports combustion. Oxygenshould not be used while smoking or inthe presence of an open flame.

WARNING: Turn off the oxygen supply when theSynchrony is not in use.

Figure 8-1 Optional Oxygen Valve

OxygenValve

(Optional)

R

To add oxygen to the circuit, the oxygen source mustcomply with the local regulations for medical oxygen. Theoxygen flow into the oxygen valve cannot exceed 15 lpmand the pressure cannot exceed 50 psi.

Page 47: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

45

R

Step 1 As shown in Figure 8-2, attach oxygen tubingto the rear oxygen valve port. The oxygenvalve prevents oxygen from entering theSynchrony enclosure when the Synchronyis stopped.

Step 2 Connect another section of oxygen tubing tothe front valve port, then to the oxygenenrichment adaptor in the breathing circuit orto one of the sample ports on the mask.

Step 3 Turn the Synchrony on before turning on theoxygen supply.

Step 4 Turn the oxygen flow on.

Figure 8-2 Attaching oxygen tubing to theSynchrony

Figure 8-3 Attaching oxygen tubing to mask port

CircuitTubing

OxygenEnrichmentAttachment

OxygenValve

From OxygenSource

R

R

ExhalationPort

Oxygen Tubingfrom OxygenValve

Oxygen tubing from the oxygen valve can be connectedeither to the mask or to an oxygen enrichment adaptor. Ifusing the adaptor, you must place it in the breathing circuitbetween the air outlet of the Synchrony and the breathingcircuit tubing.

Usi

ng O

xyge

n

Page 48: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

46

R

Using O

ptional Modem

9 Using the Optional Modem

If the optional modem has been installed in yourSynchrony, you may be asked to dial your home careprovider and have the Synchrony send information usingyour telephone line. You can send the data when themachine is starting up or during therapy.

Sending Data at Startup

STEP 1 Plug a telephone line from the rear of theSynchrony to a convenient telephone outlet. SeeFigure 9-1.

STEP 2 Connect the Synchrony to a power source andpress .

STEP 3 When the Self-Test screen appears (See Figure9-2), press .

STEP 4 When the Dial Provider screen appears (Figure9-3), press to select the Dial Providercommand, then press to contact the providerand download information.

R

R

2

1

Figure 9-1 Connecting the Telephone Line

SYSTEM SELF TEST

VER:1.00 10.0 HOURS

Figure 9-2 Self-Test Screen

Dial Provider

EXIT

Figure 9-3 Dial Provider Screen(Continued on next page)

Page 49: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

47

R

STEP 5 The SENDING REPORT screen appears asshown in Figure 9-4.

The STATUS changes from “Dialing” to“Connected” to “Complete” under normalcircumstances.

If you press when ABORT is displayed in theSENDING REPORT screen, the call will stopand the STATUS line will go toSTATUS: Complete.

STEP 6 When the data has been sent, the STATUS linewill change to Complete, and the ABORT linewill change to EXIT as shown in Figure 9-5.

STEP 7 Press to EXIT the SENDING REPORTscreen and return to the Dial Provider screen(Figure 9-3). Press again to return to the self-test screen.

If a problem occurs while attempting to send the data, theSTATUS line will tell you what the problem is. See thefollowing table for a list of the possible STATUS messagesand what you can do.

SENDING REPORT

STATUS:Dial ing ABORT

Figure 9-4 Sending Report Screen

SENDING REPORT

STATUS:Complete EXIT

Figure 9-5 Call Complete

NOTE: If it is convenient, you may leave the Synchronyconnected to the telephone outlet. It will notinterfere with your telephone.

Usi

ng O

ptio

nal M

odem

Page 50: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

48

R

Using O

ptional Modem

STATUS PossibleMessage Cause What to Do

No Modem Modem is not installed Call your home care provider.properly.

No Dial Tone A phone is off the hook Press to exit, check the telephone line connection. Lift the receiver ofor telephone line is your telephone to make sure that you have a dial tone.not working.

Line Busy The caregiver’s line is The unit will try to make the connection again after a slight delay. Press in use. if you wish to stop the redialing. Your home care provider has set the

modem to try several times before stopping.

Send Failed 1. If occurs shortly after 1. Press to exit, contact your home care provider.first try, the Synchronyunit does not have a phonenumber entered into it.

2. If occurs after many 2. Press to exit; try again later.“Line Busy” and “Retry”messages, the Synchronyunit tried too many times,with the line busy.

Table 9-1 Problems with Calling the Home Care Provider

Page 51: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

49

R

STEP 1 Connect a telephone line to the Synchrony asshown in Figure 9-1.

STEP 2 Press . The Synchrony displays theCONTROLS screen, as shown in Figure 9-6.

STEP 3 If the Dial Provider command is alreadyselected (as shown by the indicator, press .Otherwise, press to select the command, then

press .

STEP 4 Follow Steps 5 to 7 on the page 45.

Sending Data During TherapyCONTROLS: 2/2 Dial Provider

CONTROLS: 2/2 Rise Time: 3 Dial Provider

CPAP Controls

S, S/T, T, PC Controls

Figure 9-6 Controls Screens

Usi

ng O

ptio

nal M

odem

Page 52: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

50

R

Specifications

ENVIRONMENTAL

PHYSICAL

Dimensions:

Weight:

ELECTRICAL

AC Voltage:

DC Voltage:

AC Current:

100 to 240 V , 50/60 Hz

12 V

1.25 A maximum

12" L x 7" W x 6" H

Approximately 6 lb

Operating Storage

Temperature

Humidity

Atmospheric Pressure

41 °F to 95 °F15 to 95% Relative

(non-condensing)

83 kPa to 102 kPa(5600 feet to sea levelat standard pressure)

-4 °F to 140 °F15 to 95% Relative

(non-condensing)

50 kPa to 106 kPa(20000 feet to sea levelat standard pressure)

10 Specifications

Page 53: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

51

R

PRESSURE

4 to 30 cm H2O

DC Current:

Protection against electric shock:

Degree of protection againstelectric shock:

Degree of protection againstharmful ingress of water:

Modes of Operation:

ElectromagneticCompatibility:

Fuses:

5.5 A maximum

Class II

Type BF Applied Part

Ordinary Equipment, IPX0

Continuous

The BiPAP Synchrony Ventilatory Support System meets the requirements ofEN 60601-1-2.

There are no user-replaceable fuses.

Spe

cific

atio

ns

Outlet:

Page 54: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

52

R

Specifications

CONTROL

ACCURACY

Respironics circuit connectors comply with ISO 5356-1.

CONNECTORS

Parameter Range Accuracy

IPAP 4 to 30 cm H2O ± 2 cm H

2O*

EPAP 4 to 15 cm H2O ± 2 cm H

2O*

CPAP 4 to 20 cm H2O ± 2 cm H

2O*

Rate 0 to 30 BPM (PC and S/T); Greater of 1 BPM or ± 10% of setting when

4 to 30 BPM (T) measured over a four minute period.

Timed Inspiration 0.5 to 3.0 sec ± 10% of setting + 0.1 sec

Rise Time 1 to 6, corresponding to 100 to 600 msec ± 25%*

Ventilation Ramp Disabled, 0.5, 1.0, 2.0, 3.0 cm H2O per breath ± (0.1 + 0.25 of setting) cm H

2O/breath.

Measured over the full duration of the

programmed ramp.

*Measured at the patient end of circuit with a Whisper Swivel II exhalation port and no flow. Dynamic pressureaccuracy is ± 5 cm H

2O measured at the patient end of the circuit with a Whisper Swivel II and varying flow conditions.

UPGRADE

Page 55: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

53

R

TRIGGERING

ACCURACY OF MEASURED AND CALCULATED PARAMETERS

PARAMETER

Respiratory Rate

Exhaled Tidal Volume

Exhaled Minute Ventilation

Leak Rate

ACCURACY

Greater of ±1 BPM or ±10% of reading when measured over a fourminute period

± (25 + 0.15 of reading) ml

15% of reading + 1 L/min

15% of reading + 5 L/min

Respironics BiPAP® AutoTrak Sensitivity™

Spe

cific

atio

ns

Page 56: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

54

R

Index

IndexAAC Indicator 17, 18AC Power

Inlet 20Plugging 25

Accuracy Specifications 52Activity Indicator 18Air Filter

Cover 20Air Filters

Installing 21Alarm Indicators

High Priority Alarm 18Medium/Low Priority Alarm 18

AlarmsGeneral Instructions 31High Priority Defined 11Low Priority Defined 11

ApneaDefined 11

BBacteria Filter 22Bi-Level Ventilation 13Breathing Circuit

Complete 24Connection 16, 22

CCarry Handle 16Cautions 7Circuit Tubing 14Cleaning and Maintenance 41Cleaning The Synchrony 41Cleaning/Replacing the Inlet Filters 42–43Comfort control 17Communication Statement 9Communications Port 20Connections

AC Inlet 20Breathing Circuit 20, 22Communications Port 20DC Inlet 20

Page 57: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

55

R

Rear Panel 20Telephone Jack 20

Contacting Respironics 15Contents 5Contraindications 8Control Keys 17Control Panel 16

Controls and Indicators 17Details 17

Control ScreensS, S/T, PC, T Mode 28

ControlsControl Keys 17Down Key 17Enter Key 17Ramp/Silence Key 17Standby Key 17Up Key 17

Corrective ActionsMask 38Synchrony 37

CPAPDefined 11Monitoring Screen 27

DData

Sending to Care Giver 46DC cord retainer 20DC Indicator 17, 18DC Inlet 20DC Power 25Definitions

Apnea 11CPAP 11EPAP 11High Priority Alarm 11IPAP 11Low Minute Ventilation 11Low Priority Alarm 11OSA 11Pressure Control Mode 11Ramp 11Rise Time 11Spontaneous Mode 11Spontaneous/Timed Mode 12Timed Mode 12

Dialing with Modem 28Display 17Display Indicators 18

Activity 18

Inde

x

Page 58: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

56

R

Index

Navigation 19Progress 19Ramp 19Selection 19

Down Key 17

EElectrical Specifications 50Enter Key 17Environmental Specifications 50EPAP 13

Defined 11Exhalation Port 23Expiratory Positive Airway Pressure 13

FFilters

Air Filter Cover 20Bacteria (Optional) 22Cleaning/Replacing 42–43Gray Pollen 21Installing 21Ultra-Fine 21

GGray Pollen Filter 21

HHandle 16High Priority Alarm Defined 11Humidifier

Placement 22

IIndicators 18

AC 17, 18Activity 18DC 17, 18Display 18Medium/Low Priority Alarm 18Navigation 19Progress 19Ramp 19Selection 19

Inlet FiltersCleaning/Replacing 42–43

Inspiratory Positive Airway Pressure 13Installing the Air Filters 21Intended Use 8Interface 14Introduction 11IPAP 13

Defined 11

Page 59: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

57

R

LLimited Warranty 60Location 22Low Minute Ventilation

Defined 11Low Priority Alarm

Defined 11

MMaintenance 41Mask 14Mask Discomfort

Corrective Action 38Medium/Low Priority Alarm Indicator 18Modem 46

Communication Statement 9Sending Data 46

Monitoring ScreenCPAP 27S, S/T, PC, T Mode 27

NNavigation Indicator 19

OOperating 27Optional Oxygen Valve 16OSA, Defined 11Oxygen

Connecting 44Optional Valve 16, 44Valve 45

PPackage Contents 5Physical Specifications 50Plugging the Unit In 25Pressure Assist Mode Defined 11Pressure Specifications 51Problems 37Progress Indicator 19Provider

Dialing with Modem 28

RRamp

Using 26Ramp Defined 11Ramp Indicator 19Ramp/Silence Key 17

Inde

x

Page 60: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

58

R

Index

Rear Panel 20Respironics, How to Contact Us 15Rise Time

Controlling 28Defined 11

SS, S/T, PC, T Mode

Control Screens 28Monitoring Screen 27

ScreensCPAP Monitoring 27Monitoring 27S, S/T, PC, T Mode Monitoring 27

Selection Indicator 19Sending Data 46Specifications

Accuracy 52Electrical 50Environmental 50Physical 50Pressure 51

Spontaneous ModeDefined 11

Spontaneous/Timed ModeDefined 12

Standby Key 17Starting the Synchrony 26Symbols 15Synchrony

Breathing Circuit 14Cleaning 41Introduction 11Operating 27Starting 26Touring 16Using 21What it is 14Where to Locate 22

TTelephone

Connecting the Modem 46Telephone Jack 20Timed Defined 12Touring 16Trigger Sensitivity 53Troubleshooting 37Tubing 14

Breathing Circuit 20, 22

Page 61: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

59

R

UUltra-Fine Filter 21Up Key 17Using the Synchrony 21

VVentilation, Bi-Level 13

WWarnings 6–7Warnings, Cautions, and Notes 6–7Warranty 60

Inde

x

Page 62: 1003121 RAG 6/21/00 - CPAP |Portables Oxygen Concentrators Files...The following are potential side effects of noninvasive positive pressure therapy: • ear discomfort • conjunctivitis

60

R

Limited Warranty

Respironics, Inc. warrants that the BiPAP Synchrony System shall be free from defects of workmanship and materialsand will perform in accordance with the product specifications for a period of one (1) year from the date of sale byRespironics, Inc. to the dealer. If the product fails to perform in accordance with the product specifications, Respironics,Inc. will repair or replace – at its option – the defective material or part. Respironics, Inc. will pay customary freightcharges from Respironics, Inc. to the dealer location only. This warranty does not cover damage caused by accident,misuse, abuse, alteration, and other defects not related to material or workmanship.

Respironics, Inc. disclaims all liability for economic loss, loss of profits, overhead or consequential damages which maybe claimed to arise from any sale or use of this product. Some states do not allow the exclusion or limitation ofincidental or consequential damages, so the above limitation or exclusion may not apply to you.

This warranty is given in lieu of all other express warranties. In addition, any implied warranties—including anywarranty of merchantability or fitness for the particular purpose—are limited to one year. Some states do not allowlimitations on how long an implied warranty lasts, so the above limitation may not apply to you. This warranty givesyou specific legal rights, and you may also have other rights which vary from state to state.