1319824
TRANSCRIPT
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 1/42
Oxygen Guideline July 2012 Page 1 of 42
Approved on: July 2012
6.6
GUIDELINE /PROCEDURE
SUBJECT: OXYGEN THERAPY AND DELIVERY DEVICES FORPAEDIATRICS JHCH (Excluding NICU, ED, ICU)
DOCUMENT NUMBER: 6.6
DATE DEVELOPED: June 2006
DATES REVISED: November 2011, August 2012
DATE APPROVED: August 2012
REVIEW DATE: August 2016
DISTRIBUTION:
JHCH – H1, J1, J2A, J2Day Stay, Oncology Day Stay
PERSON RESPONSIBLE FOR MONITORING AND REVIEW:
Paediatric Respiratory CNC
COMMITTEE RESPONSIBLE FOR RATIFICATION AND REVIEW:
Kaleidoscope GNS Quality Committee
Keywords: children, device, oximetry, oxygen, therapy, SaO2
Disclaimer : It should be noted that this document reflects what is currently regarded as a safe and appropriate approachto care. However, as in any clinical situation there may be factors that cannot be covered by a single set ofguidelines, this document should be used as a guide, rather than as a complete authorative statement ofprocedures to be followed in respect of each individual presentation. It does not replace the need for theapplication of clinical judgment to each individual presentation.
SAFE WORK PRACTICE
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 2/42
Oxygen Guideline July 2012 Page 2 of 42
Approved on: July 2012
SAFE WORK PRACTICE
1. Oxygen is highly combustible; therefore anything that supports or initiates
combustion should be avoided e.g. smoking, candles.
2. All electrical equipment, wall oxygen and cylinder oxygen will be operated within the
guidelines of safe work practices for both patient and staff.
3. All oxygen delivery circuits are single patient use as per infection control guidelines.
TABLE OF CONTENTS:
DEFINITIONS / ABBREVIATIONS ................................................................................... 4
1. OUTCOMES............................................................................................................. 5
2. INDICATIONS .......................................................................................................... 5
3. PRECAUTIONS ....................................................................................................... 5
4. OXYGEN THERAPY PROCEDURE ......................................................................... 6
4.1 OBSERVATIONS ............................................................................................. 6
4.1.1 SIGNS AND SYMPTOMS OF RESPIRATORY DISTRESS ................... 7
4.1.2 OBSERVATION TO ASSESS THE SIGNS AND SYMPTOMS OF
RESPIRATORY DISTRESS .................................................................. 8
4.1.3 NORMAL RANGES IN BLOOD GAS ANALYSIS ................................... 9
5. NURSING CARE FOR ALL CHILDREN ON OXYGEN THERAPY ........................... 10
6. DELIVERY SYSTEMS .............................................................................................. 11
6.1 NASAL CANNULA ............................................................................................ 11
6.1.1 NASAL CANNULA PROCEDURE ......................................................... 12
6.1.2 ADDITIONAL NURSING CARE FOR NASAL CANNULARS ................. 12
6.2 NASAL CANNULA WITH WARMED HUMIDIFIED OXYGEN (WHO) VIA THEFISHER & ® PAYKEL HUMIDIFICATION SYSTEM ................................................ 13
TABLE 1: F & P HUMIDIFIER CURCUITS, CORRESPONDING PRONGS &
USE .................................................................................................. 14
6.2.1 WARMED HUMIDIFICATION ................................................................ 15
6.2.2 WHO EQUIPMENT ................................................................................ 16
TABLE 2: F & P BC RANGE NASAL CANNULA SIZE SELECTION TO USE
WITH CIRCUIT RT329 INFANT RESPIRATORY CARE SYSTEM .... 17
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 3/42
Oxygen Guideline July 2012 Page 3 of 42
Approved on: July 2012
TABLE 3: F & P OPTIFLOW JUNIOR NASAL CANNULA SIZE SELECTION TO
USE WITH CIRCUIT RT330 OPTIFLOW TUBING KIT ..................... 17
6.2.3 ASSEMBLE WHO EQUIPMENT ............................................................ 17
TABLE 4: WHO PROCEDURE ......................................................................... 19
6.2.4 ADDITIONAL NURSING CARE FOR CHILDREN ON WHO VIA NASALCANNULAE ........................................................................................... 28
6.2.5 TRANSPORTING CHILD WHILE ON FISHER & PAYKEL WHO
CURCUIT .............................................................................................. 29
TABLE 5: PROCEDURE FOR TRANSPORT ................................................... 29
6.2.6 CLEANING WHO ................................................................................... 30
6.2.7 TROUBLESHOOTING FOR WHO ......................................................... 30
6.3 FACE MASK ..................................................................................................... 316.3.1 PROCEDURE FOR FACE MASK APPLICATION .................................. 31
6.4 PARTIAL RE-BREATHER ................................................................................ 32
6.4.1 PARTIAL RE-BREATHER PROCEDURE .............................................. 33
6.5 NON RE-BREATHER BAG .............................................................................. 33
6.5.1 NON RE-BREATHER BAG PROCEDURE ............................................ 33
6.6 VARIABLE CONCENTRATION MASK ............................................................. 34
6.6.1 VARIABLE CONCENTRATION MASK PROCEDURE ........................... 35
6.7 BLOW OVER OXYGEN.................................................................................... 35
6.8.1 BLOW OVER OXYGEN PROCEDURE ................................................. 35
6.8 TRANSPORT OXYGEN VIA PORTABLE CYLINDER ...................................... 36
7. WEANING ................................................................................................................ 36
8. EQUIPMENT CHECK ............................................................................................... 37
9. DOCUMENTATION .................................................................................................. 37
DEPARTMENT OF HEALTH CIRCULARS ...................................................................... 38
AREA POLICIES .............................................................................................................. 38
AUTHOR & CONSULTATION .......................................................................................... 38
REFERENCES ................................................................................................................. 39
APPENDIX 1. DETERMINING FiO2 DELIVERED FROM THE OUTLET FLOW ............... 40
APPENDIX 2. AIR / OXYGEN BLEND FOR WHO ........................................................... 41
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 4/42
Oxygen Guideline July 2012 Page 4 of 42
Approved on: July 2012
DEFINITIONS / ABBREVIATIONS:
BP Blood pressure
BTF Between the Flags
CXR Chest x-ray
ED Emergency Department
FiO 2 Fract ion of insp ired oxyg en (%)
HR Heart rate
HFNP - 8 High-f low nasal pron g to 8 L/min l imit
ICU Inten sive Care Unit
IV or IVT Intrav enous Therapy
JHCH John Hunter Children’s Hospital
LOC Level of consc iou sness
L/kg L i t res of gas f low p er k i logram of b ody weight per
minute
L/min L i t res of gas f low per minute
NBM Ni l by mo uth
NICU Neonatal Inten sive Care Unit
NPA Nasoph aryngeal A spirate
MO Medical Off icer
O 2 Oxygen
PaO 2 Part ia l pressur e of oxyg en in arter ia l bloo d
Resps Respiratory rate
RDS Respiratory Distress Score
RR Respiratory rate
SaO 2
Oxygen saturat ion
SOP Stand ard Operat ing Procedure
SPOC Stand ard Paediatr ic Observat ion Chart
WHO Warmed Humid if ied Oxygen
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 5/42
Oxygen Guideline July 2012 Page 5 of 42
Approved on: July 2012
1. OUTCOMES:
To maintain adequate oxygenation to body tissues to ensure:
1. Hypoxia is minimised
2. Optimal respiratory status is achieved and maintained.
3. Safe and effective administration of oxygen is delivered.
NOTE:
In order to increase the fraction of inspired oxygen concentration (FiO2) available to
a patient, a variety of oxygen delivery devices are employed to administer medical
oxygen. The oxygen should be administered with humidity if possible.
Oxygen is a drug and should be given only when ordered by a medical officer, or by
a registered nurse as a nurse initiated order in an emergency situation. A nurse
initiated order for oxygen can be given as an intervention whilst patient is reviewed.
Oxygen is to be ordered on the intervention section of an age appropriate Standard
Paediatric Observation Chart (SPOC).
Oxygen therapy is given to maintain oxygen saturations above 95% unless altered
calling criteria are documented on the SPOC.
2. INDICATIONS:
Hypoxia.
Increased work of breathing, increased heart rate with progressive decrease inSaO2.
3. PRECAUTIONS:
In patients with chronic carbon dioxide retention whose stimulus to breathe is adecreased partial pressure of oxygen in arterial blood (PaO2), oxygen
administration may depress respiratory drive. Careful monitoring of these patients
for hypoventilation is required during oxygen therapy1.
Persons with severe lung disease need to be maintained at the prescribed oxygen
saturation range.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 6/42
Oxygen Guideline July 2012 Page 6 of 42
Approved on: July 2012
4. OXYGEN THERAPY PROCEDURE:
4.1 OBSERVATIONS:
Children’s oxygenation status should be assessed in a well-lit room if possible by
assessing clinical signs and symptoms and recording baseline observations.
Observations must be recorded on an age appropriate NSW Health Standard
Paediatric Observation Chart (SPOC) with the times entered as the observations
are completed.
The response of the nurse assessing the child will depend on the clinical
assessment, level of deterioration and which zone the observations are recorded in.
See Kaleidoscope Recognition of the Deteriorating Paediatric Patient Guideline,
Number 3.19.
In the acute presentation, frequent observations should be attended and recorded.
The first set of observations must be initiated within 10 minutes of the child’s
deterioration then half hourly to hourly depending on the condition of the child2.
Patients who are stable with ongoing treatment of oxygen need hourly heart rate,
respiratory rate, SaO2, and gas flow. Temperature and BP can be attended fourth
hourly if child stable.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 7/42
Oxygen Guideline July 2012 Page 7 of 42
Approved on: July 2012
4.1.1 SIGNS AND SYMPTOMS OF RESPIRATORY DISTRESS:
Neurolog ical Status - Anxiety
- Restlessness initially progressing to
agitation, confusion, disorientation
and loss of consciousness.
Increased Work o f Breathing - Apnoea
- Change in pattern of breathing including;
respiratory rate, rhythm and effort.
- Nasal flaring
- Grunting
- Mouth breathing
- Use of accessory muscles and recession
Vital Sign s - Tachycardia
- Hypertension
- Tachypnoea
- Bradypnoea may indicate further
deterioration
- Decreasing O2 saturation
Skin Colour - Mild hypoxia initially as pale colour due to
peripheral vasoconstriction.
- Clammy
- Later signs progressing to bluish grey
[cyanosis] most easily seen as lip
discolouration3.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 8/42
Oxygen Guideline July 2012 Page 8 of 42
Approved on: July 2012
4.1.2 OBSERVATIONS TO ASSESS THE SIGNS AND SYMPTOMS OF
RESPIRATORY DISTRESS INCLUDE:
a) Respiratory rate.
b) Respiratory effort and work of breathing as above.
c) Heart rate.d) SaO2 (in any amount of O2).
An oximetry reading:
95% in room air is desirable unless an altered calling criterion is
documented on the SPOC.
90-95% indicates a need for increased observation and report to the
nurse in charge as per SPOC zone protocol.
< 90% requires a rapid response as per SPOC zone protocol.
The response to the SaO2 reading is dependent:
On the child’s normal baseline observation. Children with chronic
illnesses such as congenital heart disease, chronic obstructive pulmonary
disease, pulmonary hypertension, cystic fibrosis and bronchopulmonary
dysplasia should be assessed individually as their oxygen requirements
may be different. An altered calling criterion will need to be documented
on the SPOC if a child’s stable baseline observations are outside the
accepted values.
On work of breathing and heart rate and not just a decrease in SaO2
alone.
e) Oxygen – Flow rate of gases and device.
f) Capillary refill.
g) Blood pressure.
h) Level of consciousness.
i) Pain score.
j) Temperature.
k) Blood glucose level if necessary.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 9/42
Oxygen Guideline July 2012 Page 9 of 42
Approved on: July 2012
4.1.3 TABLE 1: NORMAL RANGES for ARTERIAL BLOOD GAS4
Measurement ArterialBlood
Clinical Notes
Acid-base status(pH)
7.35 – 7.45 Most important acid-base value; detectsacidosis or alkalosis
Partial pressureof carbon dioxide(PaCO2)
35 – 45mm Hg
Measures adequacy of ventilation andrespiratory contribution of acid-baseabnormality (respiratory acidosis)
Partial pressureof oxygen (PaO2)
(sea level)
80 – 100mm Hg
Indicates driving pressure that causes oxygen-hemoglobin binding; varies with age and
barometric pressure.
Bicarbonate(HCO3)
21 – 28mEq/L
Measures metabolic contribution to acid-baseabnormality (metabolic acidosis); calculatedfrom pH and Pco2
Base excess (BE) -2 to +2 Reflects deviation of bicarbonate concentrationfrom normal
Saturation ofhemoglobin withoxygen (SaO2)
96 – 98% Indicates abnormality of oxyhemaglobinassociation and dissociation; may bemeasured directly or calculated from Po2, pHand body temperature
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 10/42
Oxygen Guideline July 2012 Page 10 of 42
Approved on: July 2012
5. NURSING CARE FOR ALL CHILDREN ON OXYGEN
THERAPY:
Oxygen is a drug and should be given only when ordered by a medical officer, or by
a registered nurse as a nurse initiated order in an emergency situation. A nurse
initiated order for oxygen can be given as an intervention whilst patient is reviewed.
Observations as above per guideline.
All children on oxygen therapy MUST be connected to the Massimo System
Patients who are stable with ongoing treatment of oxygen need hourly heart rate,
respiratory rate, SaO2, and gas flow. Temperature and BP can be attended fourth
hourly if child stable.
Oxygen therapy is given to maintain oxygen saturations above 95% unless an
altered calling criterion is documented on the SPOC.
Children who receive oxygen therapy via the Fisher and Paykel Humidifier at
a flow rate of 1L/kg/min with a ratio of oxygen to air of 1:1, which equates to
an approximate FiO2 of 60% oxygen, can be nursed in the paediatric wards of
JHCH.
Team leader should be notified BEFORE an increase in oxygen concentration
or flow rate is given.
A medical review should be attended after an increase in oxygen concentration or
flow rate is given.
Patient delivery device position is important as dislodgement will result in a loss of
respiratory support
If feeding is required, respiratory status should be observed closely
If child is NBM, IV fluid replacement or NG therapy should be commenced
Oral and nasal hygiene and normal saline drops as necessary.
Suctioning as required with caution. If suctioning is indicated there should be two
nurses with child for this procedure with emergency equipment ready to use
therefore checked prior to commencement of the procedure.
Document all changes in oxygen administration on the SPOC.
Oxygen is to be ordered on the intervention section of the SPOC.
Document all interventions on patient’s notes.
All other observations and interventions are to be documented on the paediatric
respiratory observation chart.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 11/42
Oxygen Guideline July 2012 Page 11 of 42
Approved on: July 2012
6. DELIVERY SYSTEMS:
The choice of delivery system is to be determined by the rate of oxygen
required as calibrated against the clinical assessment of work of breathing,
heart rate, oximetry reading and the age of the child.
It is to be determined by flow rate and FiO2 required (See Appendix 1)
The concentration of oxygen delivered is variable and depends on gas flow
rate, fit of the device, ventilatory pattern and anatomic dead space5.
The devices used to deliver oxygen are:
6.1 Nasal Cannulae
6.2 Nasal Cannulae with warmed humidified oxygen (WHO) via the
Fisher & Paykel ® Humidification System6.3 Simple Mask
6.4 Partial Rebreather Mask
6.5 Non – Rebreather Mask
6.6 Variable Concentration Mask (Venturi Or Air-Entrainment Mask)
6.7 “Blow over ” oxygen
6.8 Transport oxygen via portable cylinder
6.1 NASAL CANNULA:
L/Min FiO2
0.125 – 2 1L/min = 24% and 2L/min = 28%
Nasal cannulae from wall oxygen are the delivery method of choice for infants,
toddlers, older children and adolescents, who require a flow rate up to 2 L/min.
Nasal cannulae deliver a maximum of 28% FiO2. One litre of oxygen will deliverapproximately 24% FiO2 (See Appendix 1).
Do not deliver a flow rate greater than 2 L/min from wall oxygen as it is
uncomfortable for the child. If a higher flow rate is required an alternate delivery
system should be considered.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 12/42
Oxygen Guideline July 2012 Page 12 of 42
Approved on: July 2012
6.1.1 NASAL CANNULAE PROCEDURE:
1. Obtain the appropriate size nasal cannulae (ie infant, intermediate infant, child
or adult).
2. AQUAPAK humidifier is to be used when a child is having oxygen directly
from the wall. Attach AquaPakTM
to the wall outlet.3. Attach an appropriate (high, low or ultra-low flow) oxygen meter to wall oxygen
and ensure it is working.
4. Explain the procedure to the child and parents.
5. Assess adequacy of oxygen delivery via the child’s nasal passages.
6. Clean nares as needed.
7. To minimise skin irritation place thick Duoderm on the cheeks before securing
the prongs to the face with Hyperfix. Measure and cut Hyperfix ready for use.8. Attach the nasal prongs to the flow meter; adjust the oxygen flow to the required
amount of oxygen ordered.
9. Gently fit the prongs into the nostrils making sure the curves of the prongs follow
the contour of the nasal passage.
10.Ensure the tubing is under or over the child’s ears and behind their head.
Anchor the tube by sliding the plastic ring up until the tubing is sitting
comfortably around the child’s head.
11. Secure prongs to Duoderm with the premeasured Hyperfix. Nasal prongs should
be changed when necessary or at least weekly.
12. Connect child to Massimo System for continual monitoring
13. Check every four hours:
Integrity of nasal passages and skin area under prongs
The prongs are inserted into the nostrils correctly
Kinking or twisting does not obstruct the flow of oxygen.
6.1.2 ADDITIONAL NURSING CARE FOR CHILDREN ON WALL OXYGEN VIA
NASAL CANNULAE: (Also See Section 5 – Nursing Care for All Children On
Oxygen Therapy)
A maximum of 2 L/ minute via nasal prongs. A higher flow may cause discomfort
for the child unless a Fisher & Paykel ® Humidification System is used.
If child is requiring more that 28%, wall oxygen delivery system needs to be
changed.
Nasal prong position is important as dislodgement will result in a loss of
respiratory support.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 13/42
Oxygen Guideline July 2012 Page 13 of 42
Approved on: July 2012
Observe nares for drying, cracking or bleeding of nasal mucosa when attending
cares. Nasal toilets should be attended and normal saline drops as necessary
6.2 NASAL CANNULA WITH WARMED HUMIDIFIED OXYGEN (WHO) VIA
THE FISHER & PAYKEL ® HUMIDIFICATION SYSTEM
Each Fisher & Paykel Humidifier Circuit has its own corresponding set of nasal prongs.
The different circuits and nasal prongs are not interchangeable. There are three (3) circuits
that may be used at John Hunter Children’s Hospital. These circuits are to be used with
the MR850 heated humidifier. See Table 1 below:
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 14/42
Oxygen Guideline July 2012 Page 14 of 42
Approved on: July 2012
Table 1: Fisher & Paykel Humidifier Circuits, Corresponding Prongs and Use
Circuit Prongs Use Gas Flow FiO2 Weight
RT329 Circuit -
nfant Respiratory
Care System
BC Range of
Cannula
(See Table 2)
Prem infant – Paediatric
Original circuit used at JHCH
Circuit of choice for routine WHO
2 – 8 L/min
Max 1L/kg/min
in ward
Variable.
Dependent on gas
flow which is made
of oxygen / air
blend. (See
Appendix 2)
Maximum FiO2 is
60%, which equates
to a ratio of oxygen
and air of 1:1 can be
nursed in paediatric
wards
Prem
infant to
>5kgs
RT330 Circuit -
Optiflow Tubing
it
Optiflow
Nasal
Cannula.
(See table 3)
Premature infant – Paediatric
New Circuit Released in June 2012
Used for patients on the Bronchiolitis HFNP WHO RCTrial
Can be used for sicker older children (up to 22kgs) needing flows
>8L/min in the ward
1 – 20 L/min
Max 20L/min
for children
enrolled in
the trial
1kgs –
20kgs
RT203 Adult
Optiflow Circuit
MaxVenturi
Oxygen System
Small,
medium and
large adult
prongs.
OPT542
This circuit is used in ICU and not generally in the ward.
If a bigger child is in need of flows >25L/min this system may be
considered.
The setup of this system is outside the scope of this guideline. See
HNE Health Guideline and Procedure: Humidified Oxygen Delivery via
the MaxVenturi Oxygen Diluter with MR850 heated Humidifier in
Intensive Care Adults)
10 - 60L/min
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 15/42
Oxygen Guideline July 2012 Page 15 of 42
Approved on: July 2012
6.2.1. WARMED HUMIDIFICATION:
One of the complications of oxygen therapy, even for relatively short periods of
time, is the drying of the upper airway, which causes secretions to be thicker and
harder to expectorate6, 7
.
Humidification minimizes this inflammation, mucosal drying and retention ofsecretions
6, 8.
When there is not enough humidity in the inspired gas, nasal passages become dry
and painful. There is also a higher risk of the development of atelectasis or
infection6, 8
.
Oxygen is to be humidified for comfort of the infant or child using the Fisher &
Paykel ® Humidification system if possible.
Warmed humidification is advantageous during oxygen administration as cold, dry
inspired gas causes detrimental physiological changes within the airway such as
moisture and heat loss from the mucosa, airway secretions with decreased volume
and abnormal rheology, decreased mucociliary clearance, and inflammatory
reactions due to thermal injury or fluid imbalance9. Additionally extra energy is
required to warm the gases in the airways.
Normal mucosal function is needed to maintain airway patency and lung
compliance9.
The inspired gas flows are heated to near body temperature (37.1°) and are fully
saturated with water vapour 10
.
Warmed humidification oxygen / air flow is only used in conjunction with the Fisher
& Paykel ® MR850 Humidification System at present.
The Fisher & Paykel ® Humidification System is used at the discretion of the
nursing team leader and medical team.
The minimum and maximum gas flow rate through a Fisher & Paykel ®
Humidification system changes depending on the circuit and prongs used. See
Table 1 for Fisher & Paykel Humidifier Circuits, Corresponding Prongs and Use
Children who receive oxygen therapy via the Fisher and Paykel Humidifier at a flow
rate of 1L/kg/min with a ratio of oxygen to air of 1:1, which equates to an
approximate FiO2 of 60% oxygen, can be nursed in the paediatric wards of JHCH.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 16/42
Oxygen Guideline July 2012 Page 16 of 42
Approved on: July 2012
A child can be transferred from Emergency Department to the ward if the child is
stabilised on WHO of 1L/kg/min with a ratio of oxygen to air of 1:1, which equates to
a FiO2 of 60% oxygen.
To calculate the flow rate of 1L/kg/min on the WHO, the child’s weight will need to
be rounded up or down to the closest whole number. . For example a child weighing
3.3 kgs will be managed as a 3 kg child; an 11.8 kg child will be managed as a 12
kg child.
Children are NOT to be nursed in the wards if the gas flow rate exceeds 1 L/kg/min
with a ratio of oxygen to air of 1:1, which equates to a FiO2 of 60% oxygen unless
special precautions are initiated.
For babies <4kgs an oxygen analyzer is to be used with WHO.
The oxygen analyzer becomes less accurate when the gas flow decreases as air is
entrained diluting the oxygen concentration.
The oxygen analyzer is to be removed when gas flows are weaned as the reading
will be incorrect.
6.2.2 WHO EQUIPMENT
Fisher and Paykel provide three (3) humidifier circuits that are used with the Fisher
& Paykel Humidifier base (MR850).
Each humidifier circuit has its own corresponding prongs. Humidifier circuits and
prongs are not interchangeable.
Any of these circuits may be used in the JHCH wards.
The choice of circuit will depend on child’s size, weight and flow requirements.
Ensure you have corresponding nasal prongs and circuit suitable to the child’s
weight and flow requirements. See Table 1 for details.
If using Fisher and Paykel Optiflow nasal cannulae, Duoderm and Hyperfix are not
necessary as this product comes with adhesive (WigglepadTM
) attached to the
prongs.
If the WigglepadTM
adhesive is loose or does not adhere replace WigglepadTM
.
DO NOT STICK OPTIFLOW NASAL CANNULAE TUBING TO FACE WITH
HYPERFIX.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 17/42
Oxygen Guideline July 2012 Page 17 of 42
Approved on: July 2012
Table 2: FISHER AND PAYKEL BC RANGE NASAL CANNULA SIZE SELECTION TO
USE WITH CIRCUIT RT329 INFANT RESPIRATORY CARE SYSTEM10
Cannula Description Minimum flow Maximum patient flow
BC2425 Premature 2L/min 6L/min
BC2435 Neonate 2L/min 6L/min
BC2745 Infant (term newborn) 2L/min 7L/min
BC2755 Intermediate infant
(<5 kg)
2L/min 7L/min
BC3780 Paediatric
(> 5 kg)
2L/min 8 L/min
Table 3: FISHER AND PAYKEL OPTIFLOW JUNIOR NASAL CANNULA SIZE
SELECTION TO USE WITH CIRCUIT RT330 OPTIFLOW TUBING KIT11
The following codes should be used as per the current manufacturer’s instructions andshould be utilized as a guide when selecting nasal cannula.
Colour
Code
Product Item
Code
Approx
Weight Range
Max Flow Rate
(L/Min)
Wigglepad
Item CodeRed Premature Size OPT312 <2kg 8 OPT010
Yellow Neonatal Size OPT314 1 – 8kg 8 OPT012
Purple Infant Size OPT316 3 – 15kg 20 OPT012
Green Pediatric Size OPT318 12 – 22kg 25 OPT012
6.2.3 ASSEMBLE WHO EQUIPMENT
Fisher & Paykel Healthcare ® Humidifier base (MR850) on IV stand.
Choose circuit and prongs appropriate to the child’s condition, weight and flow
requirements.
o Fisher & Paykel Healthcare ® RT329 Infant Respiratory Care System.
Discard short piece of blue tubing as it is not used in the set-up.
o Appropriate size BC range nasal cannula. See Table 2: Fisher and Paykel
BC Range Nasal Cannula Size Selection.
OR
o Fisher & Paykel Healthcare ® RT330 Circuit - Optiflow Tubing Kit.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 18/42
Oxygen Guideline July 2012 Page 18 of 42
Approved on: July 2012
o Appropriate size Optiflow nasal cannula. See Table 3: Fisher And Paykel
Optiflow Junior Nasal Cannula Size Selection To Use With Circuit RT330
Optiflow Tubing Kit.
Fisher & Paykel Temperature / flow probe (900MR868)
Fisher & Paykel Heater wire adaptor (900MR801)
One (1) piece of green tubing approximately 1m – 1.3m long
Two (2) pieces of green tubing approximately 70cm long
Y connector
Analyzer if child weighs 3 – 4 kgs
Pathology bag to hold blue cap if analyzer is being used
1 litre bag of water for irrigation.
15L/min Oxygen flow meter x 2
15L/min air flow meter
Oximeter and appropriate size disposable probe. (See Kaleidoscope Guideline 6.4,
Paediatric Oximetry).
Ensure correct, functional emergency equipment (including suction catheters) are
available.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 19/42
Oxygen Guideline July 2012 Page 19 of 42
Approved on: July 2012
TABLE 4: WHO PROCEDURE
Photos by HNE Health Medical Photography Department.
PROCEDURE ADDITIONAL INFORMATION
EQUIPMENT REQUIRED. Ensure you have corresponding nasal prongs
and circuit suitable to the child’s weight and flow
requirements. (See Table 1, F & P Humidifier
Circuits, Corresponding Prongs And Use).
WHO Set ups can be assembled in advance
except for puncturing the water bag.
Set up in treatment room and not next to patient
area
Not all children need an oxygen analyser
however smaller infants weighing 3 to 4 kgs
should have an analyser connected to the
humidifier.
The set up for both RT329 and RT330 are the same
RT329 Infant Respiratory Care
System
RT330 Circuit - Optiflow
Tubing Kit.
CONNECT HUMIDIFIER BASE TO IV POLE:
Place Humidifier Base on metal bracket and
attach to the IV pole.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 20/42
Oxygen Guideline July 2012 Page 20 of 42
Approved on: July 2012
FIT THE CHAMBER:
Depress the light blue finger guard above the
temperature readout.
Slide the water chamber onto the humidifier base.
HANG THE WATER BAG:
Hang the 1 litre water bag from the IV pole.
Remove the blue protector caps from top of water
chamber.
Unwind the waterfeed set from white storage
frame.
Spike the waterbag and water will automatically
feed into the chamber.
The bag should be at least 50cm above the
chamber.
CONNECT THE CIRCUIT:
DISCARD SHORT PIECE OF BLUE TUBING
from RT329 circuit set-up as it is not used.
Connect the elbow of the blue inspiratory circuit
to the chamber.
50cmbetweenbag andchamber
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 21/42
Oxygen Guideline July 2012 Page 21 of 42
Approved on: July 2012
Connect the yellow end of the heater wire
adaptor plug into the yellow socket on side of
machine.
Ensure the arrows on the plug line up with the
arrow on the machine.
Connect the 3 way plug of the heater wire
adaptor to the blue inspiratory circuit which has
been attached to the humidifier.
Connect the blue end of the temperature probe
into blue socket on the side of the humidifier.
Ensure the arrows on the plug line up with the
arrow on the machine.
Connect the middle plug of the temperature
probe into circuit on humidifier.
Be careful of the prongs as the temperature
probe is fragile and breaks easily.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 22/42
Oxygen Guideline July 2012 Page 22 of 42
Approved on: July 2012
Connect the end of the temperature probe to
the patient end of the circuit.
Hang circuit from IV stand to keep clean while
you continue with set up.
RT329 Circuit
RT330 Circuit
Connect the white pressure manifold with 3
ports to the chamber. The ports are:
1. Top blue port is the pressure pop off
valve
2. The small middle nipple is a pressure
port for pressure lines, which we do not
utilize.
3. Lower port with blue cap is the oxygen
analyzer port.
DO NOT DISCARD THIS BLUE CAP IF REMOVED
(see below for instructions)
SET UP OXYGEN TUBING:
Take the two shorter pieces of green tubing
(approx. 70cm) and connect to the two top
sections of the Y connector.
Take the longest piece of green tubing (approx.
1m – 1.3m) and connect to the lower end of the
Y connector.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 23/42
Oxygen Guideline July 2012 Page 23 of 42
Approved on: July 2012
Take the longest piece of oxygen tubing which
is attached to the end of the Y connector and
connect it to the top of the white pressure
manifold.
Hang green tubing from IV stand to keep clean
while you continue with set up.
CONNECT NASAL CANNULA TO CIRCUIT:
Push distal end of nasal cannula onto blue
inspiratory circuit.
If using RT329 Circuit connect
a nasal cannula from the BC
size selection
If using RT330 Circuit connect
a nasal cannula from the
Optiflow Junior size selection.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 24/42
Oxygen Guideline July 2012 Page 24 of 42
Approved on: July 2012
COMPLETE CIRCUIT:
The humidifier may be turned on before it is
attached to wall gas outlets however the
humidifier will not reach its desired temperature
without gas flow.
RT329 Circuit
RT330 Circuit
CONNECT HUMIDIFIER TUBING TO THE WALL
OXYGEN AND AIR SUPPLY:
Take humidifier to bedside.
Add a two way oxygen connector to the wall
oxygen meter before connecting green tubing.This allows O2 high flow meter to be available
for emergency use.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 25/42
Oxygen Guideline July 2012 Page 25 of 42
Approved on: July 2012
GAS FLOW:
Connect the two pieces of green tubing to the
oxygen and air flow meters.
Record the air and oxygen flows on the
paediatric respiratory observation chart.
POSITION OF HUMIDIFIER:
Do not set up the WHO under vents or in adraught.
o Cold air on circuit will cause increased
condensation11
.
Ensure humidifier and chamber are positioned
lower than the patient.
o This allows any condensation to collect in
the tubing away from the patient.
TURN HUMIDIFIER ON:
The humidifier will take 10-15 minutes to reach
37.0ºC.
The humidifier needs flow from the wall to reach
desired temperature.
Humidifier must be warmed to at least 30º C before
connecting child to high flow. If the high flow gas is
cool and dry it is uncomfortable for the child.
The button on the right hand side of the humidifier
changes the mode. Keep the humidifier at the
default setting which is Invasive Mode. This is the
recommended setting to provide optimal
humidification for WHO.
Aim is to deliver optimal humidity 37°C (44mg/L
H2O). Temperature readout on humidifier will
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 26/42
Oxygen Guideline July 2012 Page 26 of 42
Approved on: July 2012
fluctuate between 35-40°C to achieve this.
SAFETY ALERT:
Do not touch the metal heater disk of the base unit
or the bottom of the humidifier chamber as both
become very hot and will burn if the unit is turned
on.
Ensure there is not excessive condensation (water
pooling) in the tubing or cannula. Check that the
tubes are not kinked or tangled.
ATTACHING AN OXYGEN ANALYSER:
An oxygen analyser should be used on infants
3 –4 kgs.
Older children do not require an oxygen
analyser.
To attach an analyzer remove blue plug from
the lower port of the white pressure manifold.
DO NOT DISCARD BLUE PLUG.
Place blue plug in a pathology plastic bag and
attach to the IV stand for safe keeping so it can
be used again once analyser is removed.
Calibrate analyser to room air before placing
oxygen analyser sensor into port
Stand oxygen analyser on patient self
CONNECT CHILD TO WHO:
NOTE:
Connect child to Masimo System for continual
monitoring.
The nasal cannula size should be approximately
half the diameter of the nares (no seal should
be created).
Ensure at least a 2mm gap to the septum2.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 27/42
Oxygen Guideline July 2012 Page 27 of 42
Approved on: July 2012
If an occlusive seal is formed between the
prongs and nares excessive pressure may be
generated2.
1. Gel hands.
2. Humidifier must be warmed to at least 30o Celsius
before connecting child to the gas flow as cool dry
gas is uncomfortable for child.
3. Explain the procedure to the child and parents.
4. Clean nares as needed.
5. If using the RT329 Circuit, the BC range of nasal
prongs will need to be taped to cheek.
6. Place a piece of thick Duoderm along each cheek.
(aids protection of skin from pressure) Measure
and cut Hyperfix ready for use.
7. If using the RT330 Circuit, the Optiflow range of
nasal prongs have tape already attached.
8. Remove the adhesive cover of the WigglepadTM
9. Ensure the gas flow is on
10. Ensure the prongs are the correct size for the age
of the child.
11. Gently fit the prongs into the nostrils making sure
the curves of the prongs follow the contour of the
nasal passage.
12. Secure prongs to Duoderm with the premeasured
Hyperfix or remove Wigglepad
TM
backing andsecure to cheeks.
13. DO NOT STICK OPTIFLOW NASAL CANNULAE
TUBING TO FACE WITH HYPERFIX.
14. If the WigglepadTM
adhesive is loose or does not
adhere replace WigglepadTM
.
15. Ensure the tubing is under or over the child’s ears
and behind their head.
16. Nasal prongs should be changed when necessary
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 28/42
Oxygen Guideline July 2012 Page 28 of 42
Approved on: July 2012
or at least weekly
17. Check every four hours:
Integrity of nasal passages and skin area under
prongs
Integrity of skin on face, ears and head from
metal ringed tubing if using Optiflow Junior
nasal cannula.
The prongs are inserted into the nostrils
Correctly.
Kinking or twisting does not obstruct the flow of
Oxygen10
.
6.2.4 ADDITIONAL NURSING CARE FOR CHILDREN ON WHO VIA NASAL
CANNULAE: (Also See Section 6 – Nursing Care for All Children On
Oxygen Therapy)
The minimum gas flow rate through a Fisher & Paykel RT329 Circuit is 2L/min and
maximum 8L/min.
The minimum gas flow rate through a Fisher & Paykel RT330 Circuit is 1L/min to a
maximum 20L/min.
The maximum gas flow rate for a child on the ward is 1L/kg/min.
The maximum oxygen flow rate for a child on the ward is 60%.
Hourly observation of circuit, water level and prong position
If the child’s oxygen / air flow requirements increase beyond 1 L/kg/min of
flow or 60% oxygen the child needs urgent review +/- transfer to ICU
An oxygen analyzer is to be used on infants 3-4kgs
Be aware that if an oxygen analyzer is used on flows < 1L/kg/min, the read out may
be less accurate due to the child entraining air.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 29/42
Oxygen Guideline July 2012 Page 29 of 42
Approved on: July 2012
6.2.5 TRANSPORTING CHILD WHILE ON WHO CURCUIT
To transfer paediatric patients within JHCH who are on WHO take humidifier with
patient.
Transfer the patient with the Fisher & Paykel MR850 turned off. This should be
done when transfer is imminent. The circuit will be warm enough to warm the
oxygen during the transfer. Transfer the patient with an oximeter with appropriate
alarm limits set, an ambu bag/mask, and emergency equipment as appropriate
TABLE 5: PROCEDURE FOR TRANSPORTING PATIENT ON WHO
Obtain an oxygen cylinder for transport and
check that it is full enough to support the
duration of transfer.
Check cylinder for appropriate flow meter
(capable of delivering 2L/min).
Disconnect the long piece of green tubing
from the bottom of the Y connector and
attach it to the flow meter on the oxygen
cylinder.
The end of this piece of tubing will remain
attached to the humidifier manifold for
transfer.
Set flow while flow meter is vertical.
Oxygen Cylinders may be laid on side with
flow meter always kept in upright position
to read accurately.
Adjust flow to 2L/min
Transfer child with Fisher & Paykel MR850
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 30/42
Oxygen Guideline July 2012 Page 30 of 42
Approved on: July 2012
6.2.6 CLEANING OF WHO
Each circuit, nasal prong and water supply line is disposable.
Y piece is re-usable alcohol wipe reuse oxygen tubing after cleaning with alcohol
wipes
For long term patients, change each circuit after 7 days.
Clean the humidifier base, small grey heater lead with yellow connectors, and stand
with large alcohol wipes after use.
Leads are thoroughly wiped over with alcohol wipes and re set up to decrease the
risk of cross infection and damage. The grey leads are not disposable.
Set up the unit for the next patient.
6.2.7 TROUBLESHOOTING FOR WHO
With the RT329 circuit, gas flows < 0.5 L/m may cause the humidification alarm to
be triggered – increase the gas flow.
Temperature alarms triggered - The blue probes must be pushed into the circuit
carefully and completely in order to monitor the humidity and temperature in the
circuit. Ensure that the connections are firmly matched to each other. The humidifier
base will light up the area of concern on its diagram.
Also refer to Fisher & Paykel Healthcare ® Reference cards attached to the
humidifier unit.
Watch for “rain-out” in the clear prong tubing as this can cause a lavage into the
infant if tubing is raised above the level of the child’s face. This may result in
aspiration.
Watch for condensation in the blue tubing. To remove the condensation from the
tubing, hold up the blue tubing above where you can see the condensation or the
pool of water forming and drain the fluid back into the water chamber.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 31/42
Oxygen Guideline July 2012 Page 31 of 42
Approved on: July 2012
6.3 FACE MASK
L/Min FiO2 5 – 10 45 – 60%
The face mask is applied over the mouth and nose. This increases the size of the
oxygen reservoir so that a higher flow rate can be administered.
Improper mask fit may allow room air to dilute oxygen concentration excessively
The vent holes in the mask allow room air to be inspired in addition to the oxygen
being delivered, and the exhaled carbon dioxide to be released.
A minimum gas flow of 5 L/min must be maintained in order to prevent
accumulation of carbon dioxide with subsequent re-breathing
For patient comfort, humidification should be used6,12
.
6.3.1 PROCEDURE FOR FACE MASK APPLICATION:
1. Obtain appropriate size mask and oxygen tubing.
2. Attach 15 litre flow meter to the oxygen outlet on the wall and ensure it is working
3. Explain procedure to the child and family.
4. Securely attach the oxygen tubing to the gas source and set the oxygen to the flowrate required. (To determine the FiO2 that is being delivered from a flow rate see
Appendix 1).
5. Check for gas flow through the device before placing mask on patient.
6. Place the mask over the child’s mouth and nose. Mold the metal strip on the bridge
of the nose portion of the mask to fit the patient’s face ensuring a snug fit to create
a seal.
7. Slip elastic strap over head and above or below both ears and adjust to ensure a
secure but comfortable fit. Gently pull the ends of the strap until the mask is secure
and comfortable
8. Observe child closely:
Connect child to Massimo System for continual monitoring
Continuous monitoring and hourly recording of child’s work of breathing and
observations
Monitor child’s compliance with mask.
9. Do not use for children who are vomiting or have excess secretions due to risk of
aspiration.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 32/42
Oxygen Guideline July 2012 Page 32 of 42
Approved on: July 2012
10. If the patient’s oxygen requirements increase consider a partial re-breather or non
re-breather bag and mask.
6.4. PARTIAL RE-BREATHER
L/Min FiO2 6 - 15 35 – 70%
The partial re-breathing bag is similar to the face mask but with the addition of an
oxygen reservoir bag and side ports.
By increasing the oxygen reservoir, the oxygen concentration can be increased to
70%.
The side ports are covered with one-way discs to prevent room air entering the
mask on inspiration and subsequently reducing the FiO213
.
The reservoir bag must remain inflated on inspiration and the oxygen flow rate
regulated so that it is sufficient enough to only deflate the bag by ⅓ on
inspiration6,12
.
When the child inspires, the air is drawn from the bag and through the holes in themask.
On expiration the first ⅓ of air is blown back into the reservoir bag. This air comes
from the anatomic dead space and is still rich in oxygen, humidification and
contains very little carbon dioxide.
On the next inspiration the child inhales part of the previously exhaled air together
with 100% oxygen. This system results in less dilution by room air . 6
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 33/42
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 34/42
Oxygen Guideline July 2012 Page 34 of 42
Approved on: July 2012
Recommended flow rate is minimum 8 to 15 litres (90-99% FiO2).
The valves between the mask and the reservoir should rise on inspiration and
lower on exhalation. Valves located on external mask surface should open
during exhalation.
6.6. VARIABLE CONCENTRATION MASK
(VENTURI OR AIR-ENTRAINMENT MASK)
L/Min FiO2
12 15 L/min Colour coded jets are attached to mask to deliver desired amount of
oxygen
Blue = 24%; Yellow = 28%; White = 31%; Green = 35%;
Pink = 40%; Orange = 50%
Colour and percentage of oxygen can change between different
manufacturers. Check product information on pack when usung.
Delivers a more accurate concentration of Oxygen than Hudson masks, and is able
to be calibrated with the device and easily monitored.
Requires a minimum flow rate of at least 5 litres / min and is dependent upon the
Oxygen concentration required.
Venturi mask (also called an air-entrainment mask) uses a nozzle to accelerate the
oxygen flow and mix it with air in a precise ratio. The venturi mask can deliver from
24% to 50% oxygen by using different adapters with different sized nozzle openings
varying the size of the opening where the room air enters the system (called
entrainment ports)5.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 35/42
Oxygen Guideline July 2012 Page 35 of 42
Approved on: July 2012
6.6.1 VARIABLE CONCENTRATION MASK PROCEDURE:
Follow steps 1 through 10 as per face mask procedure.
Additional Steps
Adjust oxygen flow as per chart using correct colour attachment to deliver required
oxygen %.
6.7 BLOW OVER OXYGEN
This practice is designed for short periods only.
Can be used if child is requiring small amounts of oxygen to correct transient
respiratory compromise.
Can be used when the child is receiving nasal cannula oxygen and has increased
oxygen requirements with feeding or physical activity.
Owing to dilutional effects of ambient air, the effective FiO2, which is the
hypopharyngeal, may be low and is unpredictable14
.
Blow over oxygen is a transient measure and oxygen supplementation is to be
administered by other means if child remains unstable.
The Fio2 cannot be measured.
6.7.1 BLOW OVER OXYGEN PROCEDURE:
Connect oxygen tubing to 100% oxygen.
Adjust flow meter to deliver oxygen at 5 to 10 L/min.
Hold tube with or without mask attached above child’s nose and mouth.
Deliver oxygen by:
- Oxygen mask. Deliver oxygen close to the infant's face. Allow some oxygen to
escape around the mask.
- If using oxygen tubing on its own cup hands around tubing and hold over the
infant's face.
- Once the infant is pink, withdraw the oxygen gradually until the infant remains
pink on room air.
- If the child remains compromised trial child on nasal prongs, mask, or bag and
mask depending clinical assessment15
.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 36/42
Oxygen Guideline July 2012 Page 36 of 42
Approved on: July 2012
6.8 TRANSPORT OXYGEN VIA PORTABLE CYLINDER
No need for humidification as only used for short periods
Cylinders may be laid on side
Flow meters must always be kept in upright position to read accurately
Check cylinder for fullness
Check cylinder for appropriate flow meter.
Attach oxygen tubing and delivery device
Adjust flow and place on child
7. WEANING
If the child is under the care of the specialist paediatric respiratory team the
weaning of oxygen is per order of the specialist.
To be commenced when child’s clinical condition is improving as evidenced by
decreased work of breathing and normal respiratory rate, normal cardiovascular
parameters and O2 saturations >95% or as stated in the altered calling criteria.
Infants enrolled in the HFNP WHO Randomised Control Trial (July 16th 2012 –
August 31st 2014 or trial closeout) are to be weaned according to the trial standard
operating procedure (SOP) No.6 – HFNP WHO RCT Procedure Manual.
8. EQUIPMENT CHECK
Care must be taken that all connections are secure and are tight, otherwise a
reduced volume or concentration will be administered.
Check the oxygen is flowing through the flow meter by turning the oxygen supply on
and placing finger at the outlet to feel flow.
Flow meters must be vertical to be read accurately.
Is the tubing twisted, kinked or blocked in some way? Check that all tubing is
connected securely and not kinked at any place.
Ensure the child’s mouth or nose is not obstructed by food, secretions or other
articles.
Ensure the regulator/flow meters are on the correct setting as ordered by the
medical officer6.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 37/42
Oxygen Guideline July 2012 Page 37 of 42
Approved on: July 2012
9. DOCUMENTATION
Document the initiation of oxygen therapy, changes in therapy, and the effect
and tolerance of therapy.
All “continuous” and “prn” oxygen therapy must be documented at least once per
shift in the patient’s chart. The following is required documentation
Usage of therapy (continuous or prn)
Mode of delivery
Device
FIO2 and/or liter flow
SpO2
Indication12
Oxygen therapy must be documented on the SPOC hourly.
All other observations and interventions are to be documented on the oxygen chart.
REFERENCES:
1. Warren G. Maguson Clinical Centre: National Institute of Health. Oxygen therapy
procedure. Bethesda, USA
2. Clinical Excellence Commission (2010). Between the flags, keeping patients safe:
Paediatric clinical emergency response system (CERS).
3. Aylott, M. (2006). Observing the sick child: Part 2a respiratory assessment.
Paediatric Nursing, 18, 9 ,38-44.
4. McCance, K., & Huether, S.E. Pathophysiology: The biological basis for disease in
adults and children. 4th Ed. Mosby, Missouri, USA.
5. Pruitt, W. and Jacobs, M. (2003). Breathing lessons: Basics of oxygen therapy.
Nursing 33, 10. 43-45.
6. The Children’s Hospital at Westmead. (2008) Oxygen therapy and delivery devices
procedure.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 38/42
Oxygen Guideline July 2012 Page 38 of 42
Approved on: July 2012
7. Sim, M., Dean, P., Kinsella, J., Black, R., Carter, R., and Hughes, M. (2008)
“Performance of Oxygen Delivery Devices when the breathing pattern of respiratory
failure is simulated”. Anaesthesia, 63(9), 938 -940
8. Fell, H. and Boehm, M. (1998) “Easing the discomfort of oxygen therapy”, Nursing
Times, 94(38), 57-58.9. Williams, R., Rankin, N., Smith, T., Galler D. and Seakins, P. (1996). Relationship
between the humidification and temperature of inspired gas and the function of the
airway mucosa. Critical Care Medicine 24(11)
10. Fisher & Paykel Healthcare Product Literature. RT329 Infant Oxygen Therapy
usage System. 2008
11. Fisher & Paykel Healthcare Product Literature. (2012). Optiflow Junior: Easy care,
effective flow.12.Jevon, P. (2007). “Respiratory Procedures” Nursing Times, 103(32), 26 – 27
13. Mayo Health Care, Australia. Product information
http://www.mayohealthcare.com.au/products/Resp_oxygen_variableconcent_
mask.htm
14. Frey, B. and Shann, F. (2002). Oxygen administration in infants. Arch Dis Child
Fetal Neonatal Ed, 88. F84-F88
15. Princess Margaret Hospital for Children. (2010). Paediatric Nursing Practicemanual: 7.2.3 Humidified High Flow Nasal Cannula Therapy for Infants. Perth
DEPARTMENT OF HEALTH CIRCULARS:
NSW Health, (2006) Policy Directive: newborn Infants – Safe oxygen administration
AREA POLICIES:
3.19 Recognition of the deteriorating paediatric patient
6.8 Fisher & Paykel ® Humidification System
6.4 Paediatric Oximetry
The Upper Hunter Emergency Guidelines: July 2004
John Hunter Neonatal Unit Oxygen Therapy Policy: May2003
The New Children’s Hospital, Westmead Oxygen Policy: 2000.
AUTHOR:
Bernadette Goddard – Paediatric General Respiratory Clinical Nurse Consultant
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 39/42
Oxygen Guideline July 2012 Page 39 of 42
Approved on: July 2012
CONSULTATION:
Dr Bruce Whitehead – Director of Tertiary services / Respiratory Staff specialist JHCH
Linda Cheese – Paediatric Respiratory Clinical Nurse Consultant
Dr Larry Roddick – Paediatric Staff SpecialistDr Mark Lee – Paediatric Emergency Staff Specialist
Elizabeth Kepreotes – Clinical Improvement Coordinator
Leanne Lehle – H1 NUM
Elizabeth Newham – Paediatric Nurse Educator
APPROVED: CPGAG 16th July 2012
KGN Quality and Safety Committee – 24
th
July 2012
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 40/42
Oxygen Guideline July 2012 Page 40 of 42
Approved on: July 2012
Appendix 1: Determining FiO2 Delivered from the Outlet Flow
OXYGEN DELIVERY DEVICE OUTLET FLOW RATE
[LITRES / MIN ]
FiO2
Nasal Cannulae 1
2
24%
28%
Simple Mask 5
6
7
89
10
45%
50%
55%60%
60-65%
Partial Rebreather 6
8
10
12
15
35%
50%
60%
60-65%
65-70%
Non – Rebreather
10
12
15
70-80%
80-85%
85-90%
Venturi Mask Flow rate as indicated on
entrainment device
24%, 28%
30%, 35%40%, 50%
Disposable infant head box
‘Foldadome’
10
15
up to 50%
over 50%
If available confirm with oxygen
analyser
FiO2 is influenced by patient inspiratory demand. FiO2 amounts are derived from the
manufacturer’s product information and are approximations for the adult population.
THESE FIGURES ARE A GUIDE AND WILL DIFFER FOR THE PAEDIATRIC POPULATION. The exception is Venturi mask FiO2 figures which should remain constant regardless of
inspiratory demand.
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 41/42
Oxygen Guideline July 2012 Page 41 of 42
Approved on: July 2012
Appendix 2: Air and Oxygen Blends for WHO Fisher & Paykel Healthcare ® (2012)
FiO2O2 Flow
L/min
Air Flow
L/min
TOTAL
Flow L/minFiO2
O2 Flow
L/min
Air Flow
L/min
TOTAL
Flow L/min
21% 0 2
2
21% 0 7
7
30% 0.2 1.8 30% 1 6
40% 0.5 1.5 40% 2 5
50% 0.7 1.3 50% 3 4
60% 1 1 60% 3.5 3.5
21% 0 3
3
21% 0 8
8
30% 0.5 2.5 30% 1 7
40% 0.8 2.2 40% 2 6
50% 1 2 50% 3 5
60% 1.5 1.5 60% 4 4
21% 0 4
4
21% 0 9
9
30% 0.5 3.5 30% 1 8
40% 1 3 40% 2 7
50% 1.5 2.5 50% 3.5 5.5
60% 2 2 60% 4.5 4.5
21% 0 5
5
21% 0 10
10
30% 0.5 4.5 30% 1 9
40% 1 4 40% 2.5 7.5
50% 2 3 50% 3.5 6.5
60% 2.5 2.5 60% 5 5
21% 0 6
6
30% 0.5 5.5
40% 1.5 4.5
50% 2.5 3.5
60% 3 3
8/9/2019 1319824
http://slidepdf.com/reader/full/1319824 42/42
Air and Oxygen Blends for WHO Fisher & Paykel Healthcare ® (2012)
FiO2O2 Flow
L/min
Air Flow
L/min
TOTAL
Flow L/minFiO2
O2 Flow
L/min
Air Flow
L/min
TOTAL
Flow L/min
21% 0 11
11
21% 0 16
16
30% 1 9 30% 2 14
40% 3 8 40% 4 12
50% 4 7 50% 6 10
60% 5.5 5.5 60% 8 8
21% 0 12
12
21% 0 17
17
30% 1.5 10.5 30% 2 15
40% 3 9 40% 4 13
50% 4.5 7.5 50% 6 11
60% 6 6 60% 8.5 8.5
21% 0 13
13
21% 0 18
18
30% 1.5 11.5 30% 2 16
40% 3 10 40% 4.5 13.5
50% 5 8 50% 6 12
60% 6.5 6.5 60% 9 9
21% 0 14
14
21% 0 19
19
30% 1.5 12.5 30% 2 17
40% 3.5 10.5 40% 4.5 14.5
50% 5 9 50% 7 12
60% 7 7 60% 9.5 9.5
21% 0 15
15
21% 0 20
20
30% 1.5 13.5 30% 2 18
40% 3.5 11.5 40% 5 15
50% 5.5 9.5 50% 7.5 12.5
60% 7.5 7.5 60% 10 10