4/6/2018 1 Dual Energy X-Ray Absorptiometry (DXA) – Science, Technology, and Practice David M. Gauntt, Ph.D. UAB Medical Center Disclosures I am a (minor) GE stockholder I was involved in the design of two DXA machines (Expert and Achilles) when working for Lunar Corporation (now GE Lunar) My only clinical experience with DXA has been as a patient (no QA experience!) I grew up in Bedford MA, 2 miles from Hologic HQ

Upload: others

Post on 17-Jun-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

1

Dual Energy X-Ray Absorptiometry (DXA) –Science, Technology, and Practice

David M. Gauntt, Ph.D.UAB Medical Center

Disclosures

I am a (minor) GE stockholder

I was involved in the design of two DXA machines (Expert and Achilles) when working for Lunar Corporation (now GE Lunar)

My only clinical experience with DXA has been as a patient (no QA experience!)

I grew up in Bedford MA, 2 miles from Hologic HQ

Page 2: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

2

DXA Bone Densitometers

Image courtesy GE Healthcare

Image courtesy Hologic Corporation

Effective dose

Exam GE Prodigy Hologic QDR 4500

Lumbar spine 0.7 uSv 1.8 uSv (?)

Proximal femur 0.7 uSv 6 uSv

Whole body 0.6 uSv 4 uSv

Source: CRCPD Bone Densitometry White Paper (2006)

Page 3: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

3

Uses of DXA machines

Bone densitometry

Diagnosis (accuracy)

Therapy response (precision)

Fracture prediction (FRAX application)

Body composition

Total bone, total lean, total adipose

Vertebral morphology

Dual Energy X-ray Absorptiometry

ScienceTechnologyPractice

Page 4: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

4

Monoenergetic x-rays

ln

Monoenergetic x-rays

, ,

Log attenuation ln

Mass attenuation coefficient

, ≡

Areal density (g/cm2) ≡

Page 5: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

5

Dual energies

, ,

, ,

,

,

Dual energies

, , , , ,

Page 6: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

6

Material Decomposition

0.125

0.25

0.5

1

2

4

8

20 40 60 80 100 120 140

Mu/Rho (cm2/g)

Photon energy (keV)

Aluminum

PMMA

Compton

Photoelectric

Material Decomposition

0.125

0.25

0.5

1

2

4

8

20 40 60 80 100 120 140

Mu/Rho (cm2/g)

Photon energy (keV)

Aluminum PMMA Fi ed Aluminum Fi ed PMMA

Page 7: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

7

Material Decomposition

0

0.05

0.1

0.15

0.2

0.25

0.3

0.9 0.95 1 1.05 1.1 1.15

Coefficent o

f photoelectric

absorp

on

Coeffic

i

ent of Compton Sca ering

Bone

Aluminum

Lean

PMMA Adipose

Material Decomposition

‐0.2

0

0.2

0.4

0.6

0.8

1

1.2

1.4

1.6

‐0.5 0 0.5 1 1.5

Aluminum

equivalence

(g/cm2)

PMMA Equivalence (g/cm2)

Bone

Lean

Adipose PMMA

Aluminum

Page 8: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

8

Material Decomposition

‐0.2

0

0.2

0.4

0.6

0.8

1

1.2

‐0.2 0 0.2 0.4 0.6 0.8 1 1.2

Bone equivalence

(g/cm2)

Lean Equivalence (g/cm2)

Bone

Lean

Adipose

PMMA Aluminum

Example

Page 9: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

9

‐0.8 ‐0.6 ‐0.4 ‐0.2

0 0.2 0.4 0.6 0.8 1

1.2

‐5 0 5 10 15 20 25 30

Bone equivalence

(g/cm2)

Lean Equivalence (g/cm2)

Bone

Lean

Adipose

Total

Example

Example

‐0.8 ‐0.6 ‐0.4 ‐0.2

0 0.2 0.4 0.6 0.8 1

1.2

‐5 0 5 10 15 20 25 30

Bone equivalence

(g/cm2)

Lean Equivalence (g/cm2)

Bone

Lean

Adipose

Total

Page 10: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

10

Fat content correction

Image courtesy Hologic Corporation

Fat content correction

‐0.8 ‐0.6 ‐0.4 ‐0.2

0 0.2 0.4 0.6 0.8 1

1.2

0 5 10 15 20 25 30

Bone equivalence

(g/cm2)

Lean Equivalence (g/cm2)

Lean

Adipose Total so ssue

Page 11: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

11

Fat content estimation

‐0.6

‐0.4

‐0.2

0

0.2

0.4

0.6

0.8

1

1.2

‐5 0 5 10 15 20 25 30 Bone equivalence

So ssue (25% fat) Equivalence

Bone

Lean

Adipose

Total

Dual Energy X-ray Absorptiometry

ScienceTechnologyPractice

Page 12: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

12

Technologies - History

• Single Photon Absorptiometry• 1963 – Cameron and Sorenson

• Dual Photon Absorptiometry• ~1970, Cameron, Mazess, et al

• Dual X-Ray Absorptiometry• ~1980s (?); Stein, Mazess, et al

Source: Thorson and Wahner, JNucMed Vol 14, pp163-171US Patent Office

Single Photon Absorptiometry

LM Thorson, HW Wahner, “Single and Dual Photon Aborptiometry...”, J Med Vol 14, Sept 1986, pp163-171

Courtesy of Society of Nuclear Medicine and Molecular imaging.

Page 13: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

13

Dual Photon Absorptiometry

0

20

40

60

80

100

20 40 60 80 100 120 140

Yield

(%)

Photon energy (keV)

Gd153 spectrum

Detector

Page 14: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

14

Dual Photon Absorptiometry

0.125

0.25

0.5

1

2

4

8

20 40 60 80 100 120 140

Mu/Rho (cm2/g)

Photon energy (keV)

Bone Lean Adipose

Pulse height spectrum

0

2000

4000

6000

8000

10000

12000

Pulse count

PMT Pulse Height

Raw counts

Page 15: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

15

Pulse height spectrum

0

50

100

150

200

250 Pulse count

PMT Pulse Height

With 20cm water

Raster scan

Image courtesy GE Healthcare

Page 16: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

16

DXA Technologies

X-ray source• K-edge filter (GE)• kV switching (Hologic)

Beam geometry• Pencil beam (1st

generation)• Partial fan beam• Fan beam• Cone Beam

Photon counting Detectors (GE)• Na-I / PMT• CdTe (direct detection)• LYSO / SSPM

Energy integrating detectors• Scintillator/photodiode

array (Hologic)• Sandwich detector

Beam geometries

Pencil beam Wide fan beamNarrow fan beam

Page 17: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

17

Beam Geometries

Image courtesy GE Healthcare

X-ray source technologies

Characteristic GE/Lunar Hologic

Tube voltage 76 kV60kV, 120kV or100kV, 140kV

Filtration K-edgeSpinning filter

wheel*

* Six sector wheel with the following filter combinations• Air• Bone equivalent• Soft tissue equivalent• Copper• Copper + bone• Soft tissue + bone

Page 18: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

18

K-edge filtration

kV switching

Page 19: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

19

kV switching

Detector technologies

Photon counting (GE)• NaI with PMT• CdTe with direct conversion• LYSO with solid state PMT

Energy integrating• Ceramic detectors with photodiode

array (Hologic)• Sandwich detector (obsolete)

Page 20: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

20

K-edge filtration

0 20 40 60 80 100

Rela

ve photon flux

Photon energy (keV)

20 cm Water 1.0 mm Lanthanum

Sandwich detectors

Page 21: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

21

Sandwich detectors

Dual Energy X-ray Absorptiometry

ScienceTechnologyPractice

Page 22: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

22

Clinical report

Image courtesy GE Healthcare

is the average for young healthy adults

is the average for adults of the same age and sex

Diagnosis

T-Score WHO Category

-1 or higher Normal

-2.5 to -1.5 Osteopenia

-2.5 or lower Osteoporosis

Page 23: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

23

Lumbar spine BMD

Lumbar Spine and Proximal Femur Bone Mineral Density, Bone Mineral Content, and Bone Area: United States, 2005-2008

https://www.cdc.gov/nchs/data/series/sr_11/sr11_251.pdf

Femoral neck BMD

Lumbar Spine and Proximal Femur Bone Mineral Density, Bone Mineral Content, and Bone Area: United States, 2005-2008

https://www.cdc.gov/nchs/data/series/sr_11/sr11_251.pdf

Page 24: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

24

Accuracy

Parameter Value

Absoluteaccuracy

+/- 10%

Inter-modelvariation

+/- 15%

Inter-unit variation +/- 2%

Source: International Society for Clinical Densitometry (ICSD)white paper “Precision Assessment and Radiation Safety for Dual-Energy X-ray Absorptiomentry

Therapy response

Precision• Scanner stability• Technologist skill

• Positioning• Tissue segmentation

Page 25: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

25

Scanner QA

Shewhart Chart

Avg + 2 Std. Dev.

Avg - 2 Std. Dev.

Page 26: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

26

Shewhart Rules

Machine failure if• Single BMD outside average ± 3 standard deviation• Two consecutive BMD on same side of average outside avg ± 2 S.D.• Four consecutive BMD on same side of average outside avg ± 1 S.D.• Ten consecutive BMD on same side of average• Two consecutive BMD differing by more than 4 S.D.

Sources

Sydney Lou Bonnick, Bone Densitometry in Clinical Practice: Application and Interpretation, Springer, p.115 (2009)

D.Pearson, S.A.Cawte, “Long-term quality control of DXA: A comparison of shewhart rules and cusum charts”, Osteoporois International, Vol 7 (4), pp 338-343 (1997)

Technologist Precision

In-vivo Precision Measurements• Measure BMD in 30 patients 2 times, or• Measure BMD in 15 patients 3 times

Least Significant Change (LSC)95% confidence interval

Written consent recommendedIRB approval not needed

≡ 2.77

Page 27: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

27

BMD Precision

Site Coefficient of variation

Leastsignificant

change

Lumbar spine 1.0-1.2% ~3%

Total hip 0.8-1.7% ~3.5%

Femoral neck 1.1-2.2% ~4.5%

Trochanter 1.2-1.5% ~3.8%

Therapy response

Parameter Lumbar spine Femoral neck

Baseline 0.850 g/cm2 0.865 g/cm2

2 years later 0.865 g/cm2 0.858 g/cm2

Difference +0.015 g/cm2 -0.007 g/cm2

LSC for this technologist

0.027 g/cm2 0.030 g/cm2

Page 28: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

28

Accreditation - ISCD

International Society of Clinical DosimetryAccreditation Program

• Follow manufacturer’s QA recommendations• Periodic (at least once/week) phantom scans• In-vivo precision test for every technologist

• Upon completion of training• When a new DXA system is installed• When the technologist’s skill level has changed.

Accreditation - CBMD

Canadian Bone Mineral Density Facility Accreditation Program(Ontario Association of Radiologists)

• “Approved” medical physicist• Weekend long workshop on DXA QA• At initial visit

• Establish “Shewhart” baselines• Determine LSC for facility/technologist• Measure entrance skin exposure• Measure scatter

• Annual review by physicist• Establish LSC for new staff

Source: Jeff Frimeth, MSc, MCCPM, CIIP

Page 29: 180404 DXA ScienceTechnologyPracticeamos3.aapm.org/abstracts/pdf/134-38481-435574-134875-1516237236.pdfSydney Lou Bonnick, Bone Densitometry in Clinic al Practice: Application and

4/6/2018

29

Accreditation - ANZBMS

Australian and New Zealand Bone and Mineral Society

Periodic tests• Daily phantom accuracy tests• Long term stability testsAcceptance tests and after major repair• Laser light positioning• Scan line and step spacing• Scan time indication• Entrance skin exposureOther• Staff scatter exposure measurements

Acknowledgements