19 chap 14 electron beam therapy
TRANSCRIPT
1
Chapter 14
Electron Beam Therapy
2
1. In early days, betatrons were used to produce electron beams, in modern times, linacs are used to produce electron beams.
2. Clinically useful energies are between 6 and 20-MeV.
3. Used for treating superficial tumors (skin, chestwall, boost to nodes, head/neck).
4. Relatively uniform dose in the target, fast dose drop off beyond the electron range.
3
14.1 Electron Interactions
Electrons interact with atoms by different processes through the Coulomb force. These processes are (1) inelastic collisions with atomic electrons (ionization/excitation); (2) inelastic collisions with nuclei (bremsstrahlung); (3) elastic collisions with atomic electrons; (4) elastic collisions with nuclei (no energy loss, large angle deflection).
In inelastic collisions, some of the kinetic energy is lost in producing ionization or converted to other forms of energy.
In elastic collisions, kinetic energy is not lost but it may be redistributed among the emerging particles.
In low-Z media (water, tissue), electrons lose energy through ionization and excitation.
In high-Z media (tungsten, lead), bremsstrahlung is important.
In ionization, if the ejected electron is energetic enough to cause further ionization, it is called secondary electron or -ray. (note: by definition, the energy of the -ray is < ½ of the incident electron energy)
Electrons continuously lose its energy traveling through the medium.
4
14.1 Electron Interactions – A. Rate of Energy Loss
colS
radS
2 MeV/cm
kge
kge
S
leade
watere
ecol
/1038.2
/1034.3
density)(electron
26
26
MeV-1 ~ minimumcolS
electronsenergy -high
material, Z-high
for efficient more
production lungbremsstrah
2EZS rad
radcoltot SSS
5
14.1 Electron Interactions – A. Rate of Energy Loss (polarization or density effect)
mediumdensegas SS
Because of polarization of the condensed medium. Atoms close to the incident electron track screen those remote from the track.
The ratio of (S/)water to (S/)air varies with energy, therefore, the conversion from dose-to-air(in chamber) to dose-to-water(phantom) varies with depth (because electron energy decreases with depth by ~ 2-MeV/cm in water).
6
14.1 Electron Interactions – A. Rate of Energy Loss (absorbed dose)
(E>energy carried away by -ray
Local energy deposition due to ionization and excitation
(unrestricted) Stopping power (S/) refers to the energy lost by a charged particle to the medium.
Restricted stopping power (L/)col,
(linear energy transfer LET) refers to energy absorbed by the medium. (collisions in which energy loss < )
dEL
EDcol
E
,
0
)(
SL
col ,
7
14.1 Electron Interactions – B. Electron Scattering
l
θ
When an electron pencil beam passes through a medium, it suffers multiple scattering, resulting in spread in both lateral position and direction. The spread in approximately Gaussian.
2222
2
/angle. scattering squaremean theis where
:power scatteringangular mass the
EZ
l
High-Z materials are used for electron scattering foil to spread out the electron beam. (recall that photon beam is spread out by the production of bremsstrahlung itself.)
8
14.2 Energy Specification and Measurement
Electron energy
Ele
ctro
n flu
ence
Accelerator tube
Scattering foil
phantom
At exit window, nearly monoenergetic
At patient surface, energy degraded and spread due to collision with scattering foil, air
Emax(0) Ea
Ep(0)
E(0)
z
At depth z, further energy degradation and spread
Ep(z)
Electron beam
9
14.2 Energy Specification and Measurement
depth
Per
cent
dep
th d
ose
100
50
R50 Rp
Rp : practical range
Most probable energy Ep:
Ep(0)=C1+C2Rp+C3Rp2
for water
C1 = 0.22 MeV
C2 = 1.98 MeV/cm
C3 = 0.0025 MeV/cm2
Mean energy at surface E(0):
E(0)=C4×R50
for water, C4 ~ 2.33 MeV/cm
Energy at depth:
p
ppp
R
zEzE
R
zEzE
1)0()(
1)0()(
10
14.3 Determination of Absorbed Dose
Absolute dose can be measured with:
ionization chamber
calorimetry
Fricke dosimetry
Relative dose can be measured with:
film: energy independence for electron beam
TLD
diode: often used for electron beam measurement.
11
14.3 Determination of Absorbed Dose – output calibration
For photon beams, the output varies smoothly with field size.
For electron beams, the output does NOT vary smoothly with field size. This is because each applicator has its own collimator setting. For example, the output of a 10x10 applicator with a 10x10 insert may be different from that of a 15x15 applicator with a 10x10 insert.
Thus, for electron beams, it is important to measure the output of every applicator and every insert in clinical use. Do not assume that the output very smoothly with field size, especially when different applicators are involved.
For elongated or irregularly-shaped cutouts, the output should be individually measured.
12
14.3 Determination of Absorbed Dose – depth dose distribution
If ion chamber is used to measure electron beam depth doses, the conversion from depth-ionization to depth-dose involves the water-to-air stopping power ratio, which is depth-dependent. In addition, if the chamber is cylindrical, the measured depth-dose curve needs to be shifted to account for the effective point of measurement.
If diode is used, the diode response is taken as the depth-dose, no correction is needed.
Med Phys 14, 1060 (1987)
13
Med Phys 14, 1060 (1987)
14.3 Determination of Absorbed Dose – film dosimetry
Med Phys 16, 911 (1989)
Energy independence for electron relative dose measurement. The optical density can be taken as proportional to dose without correction.
14
Med Phys 16, 911 (1989)
14.3 Determination of Absorbed Dose – film dosimetry
15
14.3 Determination of Absorbed Dose – film dosimetry
Air gaps adjacent to film
Film sticking out the phantom
Film recess inside the phantom
Things to avoid with film dosimetry
16
14.3 Determination of Absorbed Dose – phantom
Water is the standard phantom.
Water-equivalent Plastic phantom (polystyrene, electron solid-water): same electron density (# electrons/cc), same effective-Z → same linear stopping power S, same linear angular stopping power.
Depth-dose measured in plastic phantom converted to depth-dose in water:
med
water
medeffmedw
watermed
watermedmedmedww
R
Rddd
SdDdD
50
50
)()(
waterPolystyrene
(clear)Polystyrene
(white)Acrylic
Electron solid water
1.000 1.045 1.055 1.18 1.04
eff 1.000 0.975 0.99 1.15 1.00
17
14.4 Characteristics of Clinical Electron BeamsCentral axis depth dose curves
6
18Modest skin sparing ↑energy ↑skin dose
Relatively uniform dose
Rapid dose drop-off for low energy electron beams, but disappears for high-energy electron beams
Bremsstrahlung x-ray contamination
The choice of beam energy is much more critical for electrons than for photons.
R80(cm) ~ E(MeV)/2.8R90(cm) ~ E(MeV)/3.2
increases with energy
dmax increases with energy for low-energy electrons, ~ 2.5cm for high-energy electrons (12-20 MeV)
18
14.4 Characteristics of Clinical Electron BeamsCentral axis depth dose curves – buildup region
Lower energy electrons scatter more and through larger angles, causing more rapid buildup, thus, the difference between the surface dose and maximum dose is larger.
Higher energy electrons scatter less and through smaller angles, causing less rapid buildup (in the extreme case, if there is no scatter, there will be no buildup).
19
14.4 Characteristics of Clinical Electron BeamsIsodose curves
Different machines → different collimation systems (scattering foil, monitor chamber, jaws, cones, air-gap to surface) → dose distribution
For low energy electron beams, isodose curves bulging out for all dose levels
20
14.4 Characteristics of Clinical Electron BeamsIsodose curves
But bulge out for low dose levels
For high energy electron beams, isodose curves constrict for high dose levels
21
14.4 Characteristics of Clinical Electron BeamsField flatness and symmetry
)maxd (e.g.depth reference aat plane reference aon
uniformindex
uniformindex
7.0A
A
A
A
50%
90%
edgegeometric
90%
or
axiscentralDD %103max
%22)()(
)()(
pDpD
pDpDsymmetry
+ ●●
+p-p
22
14.4 Characteristics of Clinical Electron BeamsBeam collimation
Dual scattering foil system
applicator
Collimator jaws open to a fixed predetermined size for a given electron energy and applicator size (do NOT change it !)
Variation of output with collimator jaws opening
23
14.4 Characteristics of Clinical Electron BeamsField size dependence
Output increases smoothly with field size (as defined by the insert with collimator jaws size fixed) due to increased phantom scatter and in-air scatter.
24
14.4 Characteristics of Clinical Electron BeamsField size dependence
The PDD increases with field size until it exceeds the lateral range of the electrons, then the PDD is almost constant with field size.
The depth of maximum dose, dmax, also increases field size until the lateral range is reached.
The output and PDD for small field electron beams need to be individually measured.
Small field size, PDD increases significantly with field size
large field size, PDD nearly constant
25
14.4 Characteristics of Clinical Electron BeamsField equivalence and square root method
For large fields (>10x10), the PDDs are nearly the same, thus, they are all equivalent.
For small circular fields, the equivalent field radius, Requiv, to a 2ax2a square field is: Requiv ~ 1.116a.
For small rectangular fields XxY, as a result of Gaussian pencil beam distribution, the PDD is related to that of square fields by:
YYXXYX DDD ,,,
26
g
f < 100 cm
14.4 Characteristics of Clinical Electron Beams
IQdmax
Electron source (virtual source)
g
IIsloped
slopef
df
g
I
Ior
df
gdf
I
I
gm
mgm
m
g
11
1
0
02
0
Virtual source
27
14.4 Characteristics of Clinical Electron BeamsX-ray contamination
6-12 MeV 0.5-1%
12-15 MeV 1-2%
15-20 MeV 2-5%
Dose due to x-ray contamination generated in the collimating system and in phantom
Dose due to x-ray contamination generally is not a concern, except for total skin electron therapy (TSET) in which the entire body is irradiated (six directions, thus, the x-ray contamination dose is increased 6-times).
28
14.5 Treatment Planning Choice of energy and field size
Choice of beam energy is dictated by the depth of the prescribed level, typically, 80-90%, (thus ~ E(MeV)/3 in cm). Similarly, the choice of field size depends on the constriction of the isodose curve of the dose level. (the margin between 90% and geometric field edge typically > 0.5 cm)
29
14.5 Treatment Planning Correction for air gaps and beam obliquity
Dose affected by air-gap (inverse square) and obliquity
dmax shifts toward the surface with increasing incident angle
Decreased penetration with increasing incident angle
30
Effect of oblique incidence
d
d’
Increased dose at shallow depth due to greater side scatter from neighboring pencil beams traversing through a larger depth (d’ > d)
Decreased dose at larger depth due to lack of side scatter since it is beyond the range of neighboring pencil beams
31
g
d
d
f
f = effective SSD (surface-to-virtual source distance)
θ
D(f+g,d)
D0(f,d)
),(),(),(2
0 dOFdgf
dfdfDdgfD
Obliquity factor
32
Obliquity factor
33
14.5 Treatment Planning Irregular surface
High dose due to extra scatterLow dose due to loss of scatter
High dose due to extra scatter
Low dose due to loss of scatter
34
14.5 Treatment Planning Tissue inhomogeneity
d
ze
) thicknessequivalent ofnt (coefficie
)1(
)(
CET
zd
zzdd
e
e
eeff
D2
),,(
df
dfAdfPDDD eff
eff
A
fAn approximation:
35
Homogeneous water phantom
Lung inhomogeneity
Without lung inhomogeneity correction
With lung inhomogeneity correction
36
Small inhomogeneity
M’
M
Material M’ has greater scattering power than material M
Cold spot
hot spot
37
14.5 Treatment Planning Use of bolus and absorbers
Bolus is used to (a) flatten out an irregular surface, (b) reduce penetration in parts of the field, and (c) increase surface dose
38
14.5 Treatment Planning
Problems of adjacent fields
Big gap, cold spot
Small gap, hot spot
39
14.5 Treatment Planning Problems of adjacent fields
9-MeV e- SSD = 100
6-MV x-ray SSD = 100
9-MeV e- SSD = 120
6-MV x-ray SSD = 100
Increased SSD leads to wider e-beam penumbra, resulting in larger areas of hot/cold spots
(due to setup clearance)
40
14.6 Field Shaping
Electron beam can be shaped by cutouts made of cerrobend or lead, placed at the applicator (cone) or directly on the skin.
Shielding thickness to achieve transmission < 5%
Shielding too thin (e.g. eye shield), causing dose buildup at depth immediately under shield
External shielding
41
14.6 Field Shaping
Lead thickness (in mm) required to stop primary electrons (transmitted dose due to bremsstrahlung photons generated in the shield) ~ MeV/2. For cerrobend, increase the thickness by 20%.
Shielding thickness vs electron energy
Measurement of transmission curves
42
14.6 Field Shaping Effect of blocking on dose rate
Output ratio = 1 / output-factor
Blocked field < electron lateral range (~ Rp/2)
When in doubt, individual dosimetry (output, depth-dose, isodose distribution) should be made for irregularly-shaped cutouts.
43
14.6 Field Shaping Internal shielding (e.g. protection of eye in the treatment of eyelid)
D
D’
Dose enhancement at the interface due to extra backscatter from the lead shield = D’/D
~30-60% enhancement
Lower energy, more scatter
Energy (MeV)
44
14.6 Field Shaping Internal shielding
No lead shield
Range of backscattered
electrons 1-2 cm in water
45
A thin layer of low-Z material (e.g.wax) can be placed in front of the lead shield to reduce backscatter
14.6 Field Shaping Internal shielding
Incident primary electrons
Depth in polystyrene upstream from the interface
46
14.6 Field Shaping Internal shielding - example
9-MeV electrons
Oral structure ch
eek
2 cm
(a) Energy immediately beyond cheek ~ 9-MeV – 2-MeV/cm x 2cm = 5 MeV
(b) Backscatter from 5-MeV electrons on lead ~ 56%
(c) Depth upstream from the interface to reach 10% dose is ~ 10 mm in polystyrene, or about 4mm of aluminum
aluminum
Pb shield
To protect oral structure, lead shield thickness ~ 5 MeV/2 = 2.5 mm
47
14.7 Electron Arc Therapy
Suitable for treating superficial tumors along curved surfaces.
Calibration of arc therapy beams
angle. i at the correction sqaure inverse Inv(i)
angle i at the chart) isodose (from P todose (P)D
min / rotations ofnumber n
(MU/min) rate dose
)()(2
)(
th
thi
0
1
0
D
where
iInvPDn
DPD
n
iiarc
48
14.7 Electron Arc Therapy Treatment planning
Beam energy
Dose increased at larger depth
Dose decreased at shallow depth
‘velocity effect’ (?)The effect depends on the field width and arc-size, i.e., the range of angles a given point is irradiated (exposed), it has nothing to do with the rotation speed.
49
14.7 Electron Arc Therapy Treatment planning
Scanning field width:
Smaller field width →
Lower dose rate (greater MU) →
Greater x-ray contamination dose (at the isocenter)
Smaller field width →
~Normal incidence at all angles (less surface curvature/obliquity effect)
Typically, field width 4 – 8 cm at isocenter.
50
14.7 Electron Arc Therapy
Location of isocenter:
Approximately equi-distance from the contour surface from all angles, and
The depth of isocenter > electron range, so that dose from primary electrons is not accumulated (but dose from contaminated bremsstrahlung x-rays cannot be avoided).
Treatment planning
51
14.7 Electron Arc Therapy Treatment planning
Field shaping
Gradual dose falloff at both ends of the arc
Use surface shield to better define the dose distribution
Isodose distribution calculated by computer treatment planning system
52
14.8 Total Skin Irradiation
Stanford technique
2-9 MeV electron beams are useful for treating superficial lesions covering a large areas of the body (e.g. mycosis fungoides)
A. Translational technique
B. Large field technique
53
14.8 Total Skin Irradiation Field flatness
3 weighted fields
Arc field vs stationary field
54
14.8 Total Skin Irradiation X-ray contamination
X-ray contamination along the beam central-axis
Reduce x-ray contamination by angling the central axis away from the patient
55
14.8 Total Skin Irradiation Field arrangement
15°15°
56
14.8 Total Skin Irradiation Dose distribution
The depth-dose curve and dmax shift toward the surface, due to oblique incident angles.
With the 6-field technique, dose uniformity of ±10% can be achieved in general, except in areas with large surface irregularities (e.g. inner thigh) where supplementary irradiation may be needed.
Bremsstrahlung dose in patient midline is approximately doubled due to the opposed beam arrangement.
57
14.8 Total Skin Irradiation Modified Stanford technique (dual field angle)
~ 400 cm
10-15°
10-15°
films
Plastic screen
58
14.8 Total Skin Irradiation Modified Stanford technique (calibration)
Single Dual-angle field
polystyrene
Parallel plate chamber
replion
poly
air
gasPTpolyP PPL
NCMD
,
=1 for parallel-plate chamber
waterpoly
water
poly
polyPwaterPS
DD
=1 P at surface
P
59
14.8 Total Skin Irradiation Modified Stanford technique (treatment skin dose)
Dual-angle field
All 6 fields
Total skin dose from all 6 dual-angle fields:
BDD polyPpolyS
BDD waterPwaterS 2.5 ~ 3.0
60
14.8 Total Skin Irradiation Modified Stanford technique (in-vivo dosimetry)
Although an overall surface dose uniformity of ±10% can be achieved, there are localized regions of extreme non-uniformity of dose on the patient’s skin. (e.g. sharp body projections, curved surfaces,…)
TLDs are most often used for in-vivo dosimetry.
61
r
14.9 Treatment Planning Algorithms
Pencil beam based on multiple scattering theory
)()(),(
)(/)(),0(
)(),0()(
2),0(2),(
22
),0(),(
2
)(
2
2
0
)(
0
222
)(
22
22
22
z
ezDzrd
zzDzd
zzdzD
rdrezdrdrzrd
ezdzrd
r
zr
p
rp
rp
r
zrp
rp
yxr
zrpp
r
r
r
x
y
Dose due to an infinite field size beam
62
14.9 Treatment Planning AlgorithmsPencil beam based on multiple scattering theory
xt
rrrr
p
yx
z
yx
p
dtexerf
z
yberf
z
yberf
z
xaerf
z
xaerf
zDzyxD
dydxzyyxxdzyxD
z
ezDzyxd
yx
0
2,
)(2
2
2,
22
2)(
where
)()()()(4
)(),,(
:2b2a size fieldr rectangulafor
''),','(),,(
)(2)(),,(
63
14.9 Treatment Planning AlgorithmsPencil beam based on multiple scattering theory (lateral spread parameter, )
')')('()'(2
1)(
:equation Eyges
2
0'
22 dzzzzz
lz
z
zx
Modified Eyges equation
Mass angular scattering power
64
14.9 Treatment Planning AlgorithmsPencil beam based on multiple scattering theory (implementation)
For more accurate electron-beam dose calculation, Monte Carlo methods are available on modern-day commercial treatment planning systems.