[2] electrogravimetry

Upload: mheira-villahermosa

Post on 03-Apr-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/27/2019 [2] Electrogravimetry

    1/5

    7/12/20

    Electrogravimetry

    Potential is applied to cause a soluble species to

    be reduced or deposited on a solid electrode

    e.g., reduce Cu2+ onto Pt cathode

    Cu2+ (aq) + 2 e Cu (metal on Pt)

    Change in weight of dried cathode before & after

    deposition = amount of Cu in sample

    Assumptions:

    All Cu is plated out

    Nothing else plates out

    Cu2+(aq) + 2 e Cu(s) E

    o = 0.34 v

    O2 + 2 H+ + 2 e H2O E

    o = 1.23 v

    Cu2+ + H2O O2 + 2 H+ + Cu(s)

    For zero current,

    Ecell = ECuEO2,H2OUse Nernst Equation with Eos & concentrations

    Ecell = 0.34 (0.0592/2)log (1/[Cu2+])

    1.23 (0.0592/2)log{1/(PO2)[H+]2

    = 0.91 v

    Therefore, apply potential more negative than0.91 v

    to force system to reach an equilibrium where [Cu2+]

    is small (like 99.9% lower than the approximate

    starting concentration)

    Choose cathode potential to reduce equilibrium

    [Cu2+] to any desired value

    Must be cautious not to set potential too far negative

    to make sure nothing else is reduced

    Normally set conditions so that reduction is completein a reasonably short period of time

  • 7/27/2019 [2] Electrogravimetry

    2/5

    7/12/20

    Electrogravimetry Cell Pt electrodes

    Problem with simple electrogravimetry system Is that the potential of the cathode is not

    accurately known

    Eapplied = Ecathode- Eanode - EIR Eovervoltage

    Ej is insignificant and can be ignored here

    Magnitude of changes in EIR or Eovervoltage & EIRthroughout experiment are not known

    A similar problem exists at the anode due to

    H2O O2 + 2 H+

    Since and H+ & O2 change during electrolysis, anode

    potential also changes

    Much of this problem can be eliminated

    by using a three electrode system

    Use potentiometer to

    measure potential of cathoderelative to reference electrode

    then manually adjust slidewire

    contact to hold Ecathode at

    desired value. This operation

    must be repeated at intervalsduring electrodeposition

    because the current changes

    with conc. Bulk current stillflows from cathode to anode,

    but anode potential is nolonger important.

    Electrochemical cells resist the flow of current

    Effect of resistance on magnitude of current

    E = IR

    To generate a current of I amperes in the cell, a

    potential IRthat I more negative than the

    thermodynamic potential (Ecell) must be applied

    Eapplied = EcellIR

    IR drop in the cell can be minimized by having a

    very small cell resistance (high ionic strength) orusing a 3electrode cell

    Only a very small current passes between the

    working electrode and the reference electrode

  • 7/27/2019 [2] Electrogravimetry

    3/5

    7/12/20

    Three Electrode Configuration

    1) Reference electrode maintains fixed potential

    despite changes in solution composition

    2) Working electrode electrode of interest which is

    the cathode in this system

    3) Counter electrode (Auxiliary electrode) third

    electrode taking most of current flow (acts as current

    sink)

    Advantages of 3 electrode system

    1) Changes in concentration at counter

    electrode are not important, there is no

    effect on working electrode potential

    2) If no current flows through reference,

    there is no IRdrop (its potential is

    constant)

    Applications of Electrogravimetry

    1) Quantitative analysis (electrogravimetry)

    a. very accurate & precise,

    b. only measurement operation is weighing,

    c. Can get deposition reaction to go to any desired

    degree of completion by proper choice of

    potential,

    d. some degree of selectivity using potential

    2) Separations separate one species from another insolution by selectively plating it out and removing it

    from solution. This method is used to remove

    interferences particularly in electrochemical methods

    3) Preconcentration plating out metals from a

    large volume of solution (> 1 L) onto a small

    electrode effectively increases their

    concentration.

    Can also reoxidize metals back into a small

    volume of solution (e.g., 1 mL)

    the ratio of volumes gives preconcentration

    factor (i.e., 1000). Can also use other methods for analysis of

    metal on solid electrode surface (e.g., AA,

    ICP, XRF, XPS).

  • 7/27/2019 [2] Electrogravimetry

    4/5

    7/12/20

    4) Electrosynthesis (controlled potential

    electrolysis)

    used by organic chemists to perform

    oxidation or reduction reactions at bench

    scale.

    Takes advantage of ability to control

    potential & produce any oxidizing or

    reducing strength desired.

    5)Purification to remove trace

    metals from reagents by platingthem out of solution onto a

    large Hg pool electrode. This

    cell can also be used for

    electrosynthesis, etc.

    Coulometry

    Methods based on counting coulombs (C), thebasic unit of electrical charge (Q)

    Faradays Law

    Q M

    m =

    n F

    Where: M = molar mass (g/mole)m = mass (g)

    n = number of electrons (unitless)

    F = Faradays constant (96,500 C/mol)

    Fundamental assumption is that reaction is 100 %

    current efficient i.e, all coulombs go to oxidizing or

    reducing species of interest

    Kinds of coulometry

    1) Controlled Potential Coulometry

    t

    Q = i dt

    0

  • 7/27/2019 [2] Electrogravimetry

    5/5

    7/12/20

    2) Constant Current Coulometry

    Q = i t

    Involves nothing more than integrating

    area under the curve in

    chronoamperometry

    Can be referred to as chronocoulometry

    Care must be taken so that there is

    enough stuff to carry the current at

    electrode surface

    Rarely used anymore

    Major application is coulometric titrations wheretitrant is prepared electrochemically and standardized

    by counting/measuring coulombs

    e.g. bromine, Br2, as titrant

    2 Br Br2 + 2e

    1) Useful for titrants that cant be stored as stable

    solutions

    2) Small currents can be measured accurately so even

    very dilute titrants can be used

    3) Theoretically, the number of coulombs can be

    counted for any method where current is measured

    by integrating

    Examples of Coulometric Titrations

    Assayed

    Substance

    Reagent

    Generated

    Precursor Titration Type

    Br Ag+ Ag anode Precipitation

    Fe2+ Cl2 HCl Redox

    H2O I2, I3 - KI(pH