20131025 - dynamic response
DESCRIPTION
Dynamic Transducer CalibrationTRANSCRIPT
![Page 1: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/1.jpg)
Today
Dynamic Transducer Calibration
Theory 1
![Page 2: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/2.jpg)
Terminology pinpointTerminology pinpoint
“response characteristic”:
relation between stimulus and response (when the stimulus varies as function of time, the response characteristic is the transfer function)
“sensitivity”: prime derivative of the change of the response of a measuring instrument on the change of the stimulus
“static sensitivity”: sensitivity of a measurement instrument whereas stimulus and response are not function of time.
“offset” – “bias”: constant value of reading whereas stimulus is null (if not corrected leads to systematic error)
“response time” – “settling time”: time interval between the instant when a stimulus changes and the instant when the response reaches and remains within specified boundaries
![Page 3: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/3.jpg)
Dynamic Calibration
The black box model:
Suitable for every transducer regardless of the physical principles used for measuring.
Allows for a quick categorization of transducers.
1
0
n
i
ii xxy
TRANSDUCER input output
STATIC MODEL DYNAMIC MODEL
j
j
in
j
j
ii
out
i
idt
tqdb
dt
tqda
x, qin y, qout
![Page 4: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/4.jpg)
Dynamic Calibration
Common transducer black box model:
Neglects the offset part as it can usally be corrected.
TRANSDUCER input output
DYNAMIC MODEL (limited to second order)
tqbtqa
dt
tdqa
dt
tqda inout
outout0012
2
2
x, qin y, qout
![Page 5: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/5.jpg)
Dynamic Calibration
0th order model:
More theoretical than real, always an approximation
Usally means “the measurement chain (transducer AND conditioning) is faster than the signal”
Usually suitable for:
Estensimetric measurements
Optical measurements
Displacement measurements
DYNAMIC MODEL
tqbtqa inout 00
STATIC SENSITIVITY
0
0
ab
ks
![Page 6: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/6.jpg)
Dynamic Calibration
1st order model:
Has a dynamic parameter which express the time delay between input and output: the time constant
Usually suitable for:
Thermal measurements
Speed measurements
DYNAMIC MODEL
tqbtqatqa inoutout 001 STATIC SENSITIVITY
0
0
ab
ks
TIME CONSTANT
0
1
aa
![Page 7: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/7.jpg)
2nd order model:
Has a dynamic parameter which express the maximum bandwidth: the natural pulse
Usually suitable for:
Force measurements
Acceleration measurements
Dynamic Calibration
DYNAMIC MODEL
tqbtqatqatqa inoutoutout 0012
STATIC SENSITIVITY
0
0
ab
ks
NATURAL PULSE
2
0
aa
n
DAMPING 20
1
2 aa
a
![Page 8: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/8.jpg)
Why is the frequency response function so important?
Every periodic function can be described by the composition of sine waves of given amplitude, phase and frequency
Frequency Response
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 -4
-3
-2
-1
0
1
2
3
4
0 0.1 0.2 0.3 0.4 0.5 0.6 0.7 0.8 0.9 1 -4
-3
-2
-1
0
1
2
3
4
![Page 9: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/9.jpg)
Frequency Response:
Tells us the relationship between input and output in the frequency domain
Useful for handling period and quasi-period signals
Needed to estimate transducer readiness
Dynamic Calibration
0
0.2
0.4
0.6
0.8
1
1.2
1.4
0 5 10 15 20 25 30
qo
ut/
(ks*q
in)
ω [rad/s]
0th order 1st order 2nd order
![Page 10: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/10.jpg)
Frequency Response:
Given a tolerance boundary can give us the (passband) bandwidth
Dynamic Calibration
0
0.2
0.4
0.6
0.8
1
1.2
1.4
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5
qo
ut/
(ks*q
in)
ω [rad/s]
0th order 1st order 2nd order
Tolerance boundaries
(e.g. 5%)
Bandwidth (5%)
Bandwidth (5%)
![Page 11: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/11.jpg)
Frequency Response
0th order model:
1ins
out
qk
q
0
0.2
0.4
0.6
0.8
1
1.2
0 2 4 6 8 10 12 14 16 18 20
qo
ut/
(ks*q
in)
τω [rad]
magnitude
0
0.2
0.4
0.6
0.8
1
0 2 4 6 8 10 12 14 16 18 20
qo
ut/
(ks*q
in)
τω [rad]
phase
![Page 12: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/12.jpg)
Frequency Response
1st order model:
1
22tan
1
1
ins
out
qk
q
0
0.2
0.4
0.6
0.8
1
1.2
0 2 4 6 8 10 12 14 16 18 20
qo
ut/
(ks*q
in)
τω [rad]
magnitude
-100
-50
0
0 2 4 6 8 10 12 14 16 18 20
qo
ut/
(ks*q
in)
τω [rad]
phase
![Page 13: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/13.jpg)
Frequency Response
2nd order model:
EACH FREQUENCY RESPONSE CURVE DEPENDS ON THE DAMPING PARAMETER
n
n
nn
ins
out
qk
q 2tan
21
1 1
22
2
![Page 14: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/14.jpg)
Frequency Response
2nd order model:
0
0.5
1
1.5
2
2.5
3
0 0.5 1 1.5 2 2.5 3
qo
ut/
(ks*q
in)
ω/ωn
magnitude
0.2 0.3 0.6 0.8
-200
-150
-100
-50
0
0 0.5 1 1.5 2 2.5 3
qo
ut/
(ks*q
in)
ω/ωn
phase
![Page 15: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/15.jpg)
Frequency Response
0
0.2
0.4
0.6
0.8
1
1.2
0 0.5 1 1.5 2 2.5 3
qo
ut/
(ks*q
in)
ω/ωn
magnitude
BANDWIDTH
Frequency response chart can be used to estimate
the response components of a given stimulus
![Page 16: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/16.jpg)
Dynamic Calibration
How can I assess my transducer order
and quantify its dynamical properties? 1. Assess the frequency range of interest
2. Perform an excitation task in that range:
Studying IMPULSE RESPONSE
Studying STEP RESPONSE
Recording a DISCRETE SINE SWEEP
Recording a CONTINOUS SINE SWEEP
3. Compare with known models responses
4. Use a fitting method (such as LS)
![Page 17: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/17.jpg)
Impulse Response:
Tells us what to expect from impulses and sudden variations of stimulus
Useful for identifying measurement systems dynamical characteristics
Useful for estimating settling time
Impulse Response
0 5 10 15-0.2
0
0.2
0.4
0.6
0.8
1
1.2
time [s]
qo/k
sqi []
0th order
1st order
2nd order
![Page 18: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/18.jpg)
Step Response:
Tells us what to expect from impulses and sudden variations of stimulus
Useful for identifying measurement systems dynamical characteristics
Useful for estimating settling time
Step Response
0 1 2 3 4 5 6 7 8 9 100
0.2
0.4
0.6
0.8
1
1.2
1.4
time [s]
qo/k
isqi []
0th order
1st order
2nd order
![Page 19: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/19.jpg)
Exercise 9: Dynamic Calibration
95.03
1ln1
00
000
01
0
0
0
1
00001
0
iS
iS
tt
iS
iSoo
ioo
Sioo
qk
qtt
tt
qk
qeqkq
qkqq
aaqa
bqq
a
a
abkqbqaqa
Step Response (first order instrument) Instrument
dynamical
properties
![Page 20: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/20.jpg)
eqq
te
qkq
qkqqq
aa
a
aa
qa
bqq
a
aq
a
a
abkqbqaqaqa
oo
oo
n
t
iS
iSoo
nn
o
n
iooo
Siooo
n
min
max
22
20
2
20
1
20
0
0
0
1
0
2
000012
1arcsin1sin1
1
2
2
Addendum: Dynamic Calibration
Step Response (second order inst.) Instrument
dynamical
properties
![Page 21: 20131025 - Dynamic Response](https://reader031.vdocuments.net/reader031/viewer/2022020106/563db943550346aa9a9ba032/html5/thumbnails/21.jpg)
Addendum: Dynamic Calibration
Step Response (second order inst.)