7090 operation/service manualpaclp.tech/lab_instruments/7090 detector with gc/7090... ·...
TRANSCRIPT
7090 Series Nitrogen / Sulfur
GC Detectors
Installation/Operation/Service Manual
Rev E
P/N MAN 7090
8824 Fallbrook Dr. • HOUSTON, TEXAS 77064 USA
TELEPHONE: 800-444-TEST or 281-580-0339 • FAX: 281-580-0719
[email protected] • www.paclp.com
PAC 7090 Series IOS Manual
Rev E MAN 7090 Page i
Copyright © 2015 by Petroleum Analyzer Company, L.P.
All text, graphics, design, and other works contained herein are the copyrighted works of
Petroleum Analyzer Company, L.P (PAC).
All rights reserved. Any redistribution or reproduction of any materials contained
herein is strictly prohibited without the express written permission of the copyright
holder.
The PAC trademark, the PAC service mark, PAC’s logo trademark, the Antek trademark,
and Petroleum Analyzer Company, L.P.’s distinctive color trademark as used on its
instruments and in these materials, are trademarks, service marks, registered trademarks,
and/or registered service marks of Petroleum Analyzer Company, L.P. in the U.S. and
other countries and are its sole and exclusive property.
Microsoft® Windows®, Microsoft Windows XP®, and Windows Professional operating
systems are registered trademarks of Microsoft Corporation. Windows is a registered
trademark of Microsoft Corporation in the United States and other countries.
Other trademarks contained within these materials are the property of their respective
owners.
INFORMATION
The information contained within these materials is subject to change without notice.
PAC makes no warranties or guarantees, either expressed or implied in these materials,
including the warranty of merchantability or fitness for a particular purpose. At any time,
PAC may modify these materials, its instruments or its programs without notice and
subsequent versions of these materials may contain different information. These
materials could contain technical inaccuracies and/or typographical errors.
PAC does not assume responsibility for the accuracy of any translation of these materials.
About PAC
PAC is an international manufacturing and service organization with a portfolio that
spans petroleum, petrochemical, biofuels, environmental, food and beverage,
pharmaceutical and industrial analysis solutions. PAC provides advanced testing
equipment for laboratory, process on-line and field use for entities ranging from the
smallest to the largest enterprise businesses. PAC offers analytical solutions for a wide
variety of applications, including chromatographic systems and detectors, elemental,
laboratory, and on-line process analyzers, software applications, and spectroscopy.
More information about PAC is available at http://www.paclp.com.
PAC 7090 Series IOS Manual
Page ii MAN 7090 Rev E
Revision History
Revision Description Date
A First Edition Release October, 2002
B Update Chapter 7 October, 2002
C Update October, 2012
D Ozone updates, 9.3.1 and 9.3.2, new drawing January 2014
E Added MSDS’s and Sulfur or Nitrogen/Sulfur Simultaneous Setup, removed DWG
June 2015
THIS MANUAL CONTAINS CONFIDENTIAL AND PROPRIETARY INFORMATION…
…which is the property of PAC, L.P. The information contained in this document is subject to
change without notice.
This document is for information only. The manufacturer accepts no liability for errors contained
herein or for incidental or consequential damages arising from the furnishing, performance, or use
of this material.
Unless otherwise specified, each reference to names or individual trademarks and service marks
has the sole purpose of illustrating the product described here. Contents of this publication may
not be reproduced in any form or by any means (including electronic storage and retrieval or
translation into a foreign language) without prior agreement and written consent from the
copyright owner.
All rights reserved.
Copyright ©2015, by PAC, LP Houston, Texas.
Printed in the United States of America All rights reserved.
Contents of this publication may not be reproduced in any form without permission of the copyright owner.
01.18.2015
PAC 7090 Series IOS Manual
Rev E MAN 7090 Page iii
In This Manual
The 7090 SERIES Operation Manual contains the following sections:
Chapter 1 – General Information: Warranty and compliance information
concerning the 7090 Series detector.
Chapter 2 – Safety Notices and cautions specifically for the 7090 Series.
Chapter 3 – Introduction A general introduction to the 7090 Series detailing
the basic theory of operation.
Chapter 4 – Installation Steps to unpack, inspect and prepare the
7090 Series for use. This Chapter details the
optional equipment available.
Chapter 5 – System Overview An overview of the controls, connections and
general layout of the 7090 Series detector.
Chapter 6 – Primary tests General tests to ensure the integrity of the gas flows
and vacuums.
Chapter 7 – Start up basics The minimal conditions needed to start up the
detector.
Chapter 8 – Applications and Analytical Procedures Describes the general operation of the system.
Includes information on controls, connections,
functional tests and basic operating parameters.
Chapter 9 – Service Information (Programming and Troubleshooting) Details on servicing the 7090 Series of analyzers,
general and preventative maintenance
schedules, as well as removal and replacement
procedures.
Chapter 10 – Data Sheets Data Sheets
PAC 7090 Series IOS Manual
Page iv MAN 7090 Rev E
TABLE OF CONTENTS
1. GENERAL INFORMATION ................................................................................................1
1.1 FCC NOTICE ......................................................................................................................... 1
1.2 DISPOSAL ............................................................................................................................. 1
1.3 CE NOTICE ........................................................................................................................... 2
1.4 WARRANTY .......................................................................................................................... 3
1.5 RETURN POLICY .................................................................................................................... 5
1.6 CONTACT US ........................................................................................................................ 7
2. SAFETY ......................................................................................................................... 10
2.1 PROPER USE AND SAFETY ...................................................................................................... 10
2.2 CERTIFICATIONS .................................................................................................................. 11
2.3 SYMBOLS ........................................................................................................................... 12
2.4 ENVIRONMENTAL PROTECTION .............................................................................................. 13
PERSONAL PROTECTIVE EQUIPMENT (PPE) ......................................................................................... 11
3. INTRODUCTION ............................................................................................................ 14
3.1 INTENDED AUDIENCE ........................................................................................................... 14
3.2 PRINCIPLE OF OPERATION ..................................................................................................... 14
3.2.1 Principle of Operation: Nitrogen .............................................................................. 14
3.2.2 Principle of Operation: Sulfur ................................................................................... 15
4. INSTALLATION .............................................................................................................. 18
4.1 PRE-INSTALLATION REQUIREMENTS ........................................................................................ 18
4.1.1 Electrical Requirements: .......................................................................................... 18
4.1.2 Environmental and space requirements: ................................................................. 19
4.1.3 Materials Requirements ........................................................................................... 20
4.1.4 Materials Supplied by PAC: ...................................................................................... 20
4.1.5 Gas Requirements: ................................................................................................... 21
4.1.6 Plumbing Requirements: .......................................................................................... 22
4.2 INSTALLATION PROCEDURES .................................................................................................. 25
4.2.1 Unpacking and inspection ........................................................................................ 25
4.2.2 Shipping damage reports ......................................................................................... 25
4.3 PHYSICAL INSTALLATION ....................................................................................................... 27
4.3.1 Gases ........................................................................................................................ 27
4.3.2 Furnace Base ............................................................................................................ 28
4.3.3 Pyrotubes and furnace ............................................................................................. 29
4.3.4 Inlet adapter ............................................................................................................. 30
4.3.5 Gas transfer lines ...................................................................................................... 32
4.3.6 Signal output cables ................................................................................................. 32
4.3.7 Vents ........................................................................................................................ 33
4.3.8 Injection systems ...................................................................................................... 33
4.3.9 Vacuum Pump .......................................................................................................... 33
5. SYSTEM OVERVIEW ...................................................................................................... 34
PAC 7090 Series IOS Manual
Rev E MAN 7090 Page v
5.1 FRONT PANEL CONTROLS ...................................................................................................... 34
5.2 REAR PANEL ....................................................................................................................... 36
5.3 INTERNAL COMPONENTS ....................................................................................................... 39
5.3.1 Vacuum pump .......................................................................................................... 41
5.4 FURNACE ........................................................................................................................... 42
6. PRIMARY TESTS ............................................................................................................ 44
6.1 PRESSURE TESTS ................................................................................................................. 44
6.1.1 Gas supply (External) leak check .............................................................................. 44
6.1.2 Vacuum internal leak check ..................................................................................... 45
6.1.3 7090 pressure leak check procedure ....................................................................... 46
6.1.4 Resetting system to factory defaults ....................................................................... 48
6.2 GC POWER-UP ................................................................................................................... 48
6.3 MODEL 7090 POWER-UP ..................................................................................................... 48
6.3.1 Check list .................................................................................................................. 48
6.3.2 Power-up and functional tests ................................................................................. 49
6.3.3 Furnace power-up .................................................................................................... 50
7. START-UP BASICS ......................................................................................................... 54
7.1 BASELINE ADJUSTMENTS ....................................................................................................... 54
7.2 HIGH VOLTAGE .................................................................................................................... 54
7.3 OZONE GENERATOR ............................................................................................................. 54
8. APPLICATIONS AND ANALYTICAL PROCEDURES ............................................................. 56
8.1 IMPORTANT BASICS .............................................................................................................. 56
8.1.1 Gas and carrier flows ................................................................................................ 56
8.1.2 System flow considerations ..................................................................................... 58
8.1.3 System gas flow adjustments ................................................................................... 58
8.2 OPTIMIZING SULFUR DETECTION ............................................................................................ 59
8.3 OPTIMIZING NITROGEN DETECTION ........................................................................................ 64
9. SERVICE INFORMATION ................................................................................................ 66
9.1 FURNACE TEMPERATURE CONTROLLER PROGRAMMING PROCEDURE ............................................ 66
9.1.1 Initial power up ........................................................................................................ 66
9.1.2 Temperature Set Procedure ..................................................................................... 68
9.1.3 Auto Tune Procedure ............................................................................................... 68
9.2 THERMAL ELECTRIC COOLER (TEC) TEMPERATURE CONTROLLER PROGRAMMING PROCEDURE ........ 70
9.2.1 Initial Power up ........................................................................................................ 70
9.2.2 Temperature Set Procedure (see following text) ..................................................... 73
9.2.3 Auto Tune Procedure ............................................................................................... 73
9.3 BOARD ALIGNMENT PROCEDURES .......................................................................................... 74
9.3.1 Printed Circuit Boards Overview .............................................................................. 75
9.3.2 Ozone Generator Board – p/n 103603 ..................................................................... 76
9.3.3 Signal Processing (58141R3 Alignment Procedure) ................................................. 77
9.3.4 High Voltage Power supply voltage (58160) Adjustment ........................................ 79
9.3.5 Display Select Adjustment Procedure (58183) ........................................................ 80
9.4 FUSE REQUIREMENTS ........................................................................................................... 83
9.5 LEAK CHECK PROCEDURES .................................................................................................... 83
PAC 7090 Series IOS Manual
Page vi MAN 7090 Rev E
9.5.1 Mass Flow Controller Scaling Procedure and Vacuum Test ..................................... 83
9.5.2 Pressure Test (With Furnace) ................................................................................... 85
9.6 START-UP & SHUTDOWN PROCEDURES ................................................................................... 85
9.6.1 Start-up Procedure ................................................................................................... 85
9.6.2 Shutdown Procedure................................................................................................ 86
9.7 ROUTINE MAINTENANCE ...................................................................................................... 86
9.7.1 Daily .......................................................................................................................... 87
9.7.2 As Needed ................................................................................................................ 87
9.7.3 Reaction Chamber Cleaning Procedures .................................................................. 87
9.8 PRACTICAL TROUBLESHOOTING CHECKLIST .............................................................................. 88
9.8.1 Gases ........................................................................................................................ 88
9.8.2 Instrument Conditions ............................................................................................. 89
9.8.3 Vacuum Pump .......................................................................................................... 89
9.8.4 Furnace ..................................................................................................................... 89
9.8.5 Gas Chromatograph Conditions ............................................................................... 89
9.8.6 Proper Integrator Channel selected ......................................................................... 90
9.8.7 Detection Problems .................................................................................................. 90
9.8.8 Noise Problems ........................................................................................................ 90
9.8.9 Practical Troubleshooting Hints ............................................................................... 90
9.9 ANALYTICAL TROUBLESHOOTING ............................................................................................ 91
10. DATA SHEETS ............................................................................................................. 104
PAC 7090 Series IOS Manual
Rev E MAN 7090 Page vii
Left Blank
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 1
1. GENERAL INFORMATION
1.1 FCC Notice
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his expense.
1.2 Disposal
Correct disposal of Waste Electrical and Electronic Equipment in the end-of life Applicable in the European Union and other European countries with separate collection systems
This product is designed for exclusive professional use by commercial companies. This marking shown on the product or its literature, indicates that it should not be disposed with other household wastes at the end of its working life. To prevent possible harm to the environment or human health from uncontrolled waste disposal, please separate this from other types of wastes and recycle it responsibly to promote the sustainable reuse of material resources. Business users should contact the producer or the importer and check the terms and conditions of the purchase contract. If you have a separate agreement with your producer or importer such that you will take care of end-of-life disposal on your own, please ensure an environmentally sound disposal according to the legal regulations for electric and electronic waste equipment in your country. This product should not be mixed with other commercial wastes for disposal. The above WEEE-symbol is the official marking for equipment under the WEEE-scope. In some EC-Member states, "pure B2B equipment" is not necessarily marked with the waste bin-symbol. To provide a homogenous EC-wide procedure, PAC uses the marking in all EC-Member states.
NOTE:
User changes or modifications not expressly approved by PAC, L.P. could void the
user’s authority to operate the equipment. Use only accessories provided with the
7090 Series or PAC approved options for continued FCC compliance.
PAC 7090 Series IOS Manual
Page 2 MAN 7090 Rev E
1.3 CE Notice
CE Notice
The symbol indicates the 7090 Series’ compliance with the European Union (EU)
directives and standards listed below:
The symbol indicates PAC’s compliance with the European Union (EU) directives
and standards listed below:
EMC Directive 89/336/EEC
EN 55011: 1998 “Electrical equipment for measurement, control and
laboratory use. EMC requirements. Particular requirements. Test
configurations, operational conditions and performance criteria for
sensitive test and measurement equipment for EMC unprotected
applications”
This system contains devices classified for use in a typical Class A commercial
environment and is not designed or intended for use in other EMC environments. The
user of this system is obliged to properly use and install the system and to take all
steps necessary to remove sources of interference to telecommunications or other
devices.
Issued 2012
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 3
1.4 Warranty
STANDARD LIMITED WARRANTY OF
PETROLEUM ANALYZER COMPANY, L.P.
I. Limited Warranty
1. Limited Warranty. For each product sold, PAC offers to the original owner (“BUYER”) a limited warranty against failure to conform to the product specifications or any defects in material and workmanship for a period of twelve (12) months
from the date of installation of the product or eighteen (18) months from date of invoice, whichever is less (the “Initial Warranty”). If
a failure to conform to specifications or a defect in materials or workmanship is discovered within the Initial Warranty period, BUYER must promptly notify PAC in writing, which notification, in any event must be received no later than 14 months from the date
of installation of the product. Within a reasonable time after such notification, PAC will correct any failure to conform to
specifications or any defect in materials or workmanship, or in lieu of such repair, and at its sole option, shall replace the product, F.O.B. PAC’s city of shipment or refund the purchase price, less a reasonable reduction in such purchase price as determined by PAC.
In no event shall PAC be liable for consequential or special damages, or for transportation, installation, adjustment or other expenses,
which may arise in connection with such products or warranty claim. Products and parts sold by authorized PAC Distributors are covered by the Distributor’s terms and conditions.
During the Initial Warranty period, PAC offers a limited warranty on each part or product repaired or replaced by a PAC
service person for a period ending the later of (a) the remaining term of the product’s Initial Warranty or (b) ninety (90) days from the date of repair or replacement whichever is longer. After expiration of the Initial Warranty period, PAC offers a ninety (90) day
limited warranty on each part or product repaired or replaced by a PAC service person. PAC further warrants that the products and
parts it sells will conform to PAC’s written specifications therefor. The foregoing limited warranties cover parts and labor only and PAC does not warrant and will not reimburse the BUYER for any other costs relating repairing or replacing the product at issue. The
foregoing limited warranties apply only to the repair or replacement of defective parts and/or products and such determination will be in the sole discretion of PAC. The limited warranties of this Section I.1 are further subject to those warranty exclusions set forth
below.
2. Aftermarket. The use of third party parts in the operation or maintenance of the product or repairs or servicing by unauthorized service personnel immediately voids all further warranty obligations of PAC. PAC genuine consumables are sold on
an as-is basis and have no warranty beyond being shipped in good working order.
3. Limited Warranty Exclusions. Other than the warranties set forth in Section I.1., PAC disclaims any and all
express or implied warranties, including but not limited to the warranty of merchantability, fitness for a particular purpose and non-
infringement of the intellectual property of others. Provided, however, this Intellectual Property Warranty shall apply to the product
only so long as BUYER (i) does not modify the product, combine the product with other elements or use the product in a practice or a process, and such modification, combination or practice of which the product forms a part is the subject of such claim or allegation of
infringement and (ii) uses the product under ordinary conditions and for their intended purposes. PAC makes no warranty, express or
implied, as to the design, sale, installation or use of its products. PAC’s warranties will not be enlarged by, nor will any obligation or liability of PAC arise due to PAC providing technical advice, facilities or service in connection with any product. PAC disclaims any
and all other warranties and representations and PAC provides no warranty on the oral representations made by its personnel while
they are undertaking Services for BUYER. There is no warranty by PAC with respect to any product’s: (i) uninterrupted or error-free operation; (ii) actual
performance, other than the product’s capability to meet PAC’s specifications therefor; (iii) removal or installation from a worksite or
process; (iv) electronic components or associated accessories (including without limitation circuit boards and integrated circuits); (v) maintenance (including without limitation gasket and seal replacements, adjustments, minor repairs and other inspection
requirements, preventative or otherwise); (vi) use under inappropriate conditions or not in accordance with operating instructions; or
(vii) use in connection with the operation of a nuclear facility. There is no warranty for products determined to be, in PAC’s sole discretion, damaged or impaired as a result of (a) misuse, neglect or accident; (b) improper application, installation, storage or use;
(c) improper or inadequate maintenance or calibration; (d) operation outside of the published environmental specification;
(e) improper site preparation or maintenance; (f) unauthorized repairs or replacements; (g) modifications negligently or otherwise improperly made or performed by persons other than PAC; (h) BUYER-supplied software or supplies; (i) use in conjunction with or
interfacing with unapproved accessory equipment; (j) use of ABC-style or dry powder fire suppression agents; or (k) leaked sample
materials. To the extent a PAC product is used in connection with the operation of a nuclear power facility, BUYER agrees to indemnify and hold PAC harmless from any and all actions, claims, suits, damages and expenses arising from such use. PAC provides
no warranty on the oral representations made by its personnel while they are attempting to assist BUYER in the operation of a
product. This Standard Limited Warranty does not apply to items consumed by the products during their ordinary use, including but not limited to fuses, batteries, paper, septa, fittings, screws, fuses, pyrolysis, dryer or scrubber tubes, sample boats, furnaces or lamps.
This warranty is valid only if genuine PAC parts, consumables and standards are used in PAC products.
4. Resale products are goods (that are sold with PAC’s products) which are not manufactured by PAC and which are supplied as an accommodation to BUYER. PAC MAKES NO WARRANTY FOR RESALE PRODUCTS, EITHER EXPRESS
OR IMPLIED, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. THE
SOLE WARRANTY SHALL BE THAT OF THE RESALE PRODUCT MANUFACTURER. BUYER agrees that PAC has no liability for resale products beyond the services within PAC’s direct control necessary to
reasonably discharge the above stated responsibility and that PAC shall not be liable for delays caused by resale product manufacturer.
PAC 7090 Series IOS Manual
Page 4 MAN 7090 Rev E
PAC does not warrant products it does not manufacture except to the extent the warranty of the manufacturer of the product at issue
passes through or is otherwise assigned to PAC. BUYER further agrees that BUYER’s SOLE AND EXCLUSIVE REMEDY for PAC’s breach of the stated responsibility shall be limited to the difference between the resale product manufacturer’s price to PAC
and PAC’s price to BUYER for resale products.
5. Expenses for Non-Warranty Work. All repairs or replacements by PAC after the expiration of any applicable limited warranty period will be performed in accordance with PAC’s standard rate for parts and labor. Further, if upon PAC’s
inspection and review, PAC determines the condition of the product for which a warranty claim is submitted is not caused by a defect
in PAC’s material and workmanship, but is the result of some other condition, including but not limited to damage caused by any of the events or conditions set forth in Section I.3., BUYER shall be liable for all expenses incurred by PAC to conduct the inspection
and review of the product.
6. Exclusive Remedy. The limited warranty contained herein constitutes BUYER’s exclusive remedy with respect to products sold by PAC and PAC’s liability shall be exclusively limited to the written limited warranty specified herein. No
employee, representative or agent of PAC is authorized to either expressly or impliedly modify, extend, alter or change any of the
limited warranties expressed herein to BUYER. 7. Procedure and Costs. All limited warranty claims must be made in writing promptly following discovery of any
failure to conform to specifications or any defect in materials or workmanship. BUYER must hold products for inspection by PAC. If
requested by PAC, BUYER must send the product to PAC for inspection. Any such returns by BUYER will be at BUYER’s expense and BUYER will remain liable for any loss of or damage to the product during such product’s transportation to PAC. No products
will be sent to PAC for inspection unless PAC has authorized BUYER to do so.
8. Terms and Conditions. PAC’s General Terms and Conditions are incorporated herein by reference and BUYER accordingly agrees to be bound by the terms thereof.
II. Limitations on PAC Liability
1. In General. BUYER agrees that in no case shall PAC be liable for any special, incidental, consequential, or punitive
damages based upon any legal theory whether or not such damages are foreseeable. such damages include, but are not limited to, loss
of profits, loss of savings or revenue, loss of use of the product or any associated equipment, cost of capital, cost of any substitute equipment, facilities or services, downtime, the claims of third parties including customers, injury to property and, unless precluded
under applicable state law, bodily and personal injury. PAC’s total liability for any and all losses and damages arising out of any and
all causes whatsoever including, without limitation, defects in the product(s), services, software, or documentation supplied or breach of this agreement, shall in no event exceed the purchase price of the applicable product(s). BUYER agrees these limitations on PAC’s
liability are reasonable and reflected in the amounts charged by PAC for its products.
2. BUYER Data. If any data supplied by BUYER, whether in the form of BUYER specifications or pursuant to any purchase order or other documentation, proves to be inaccurate, any warranties or other related obligations of PAC relying thereon will be void.
3. Force Majeure. This Standard Limited Warranty does not cover and PAC shall not be liable for either direct or
consequential damage caused, either directly or indirectly, as a result of: (i) any act of God, including but not limited to natural disaster, such as floods, earthquakes, or tornadoes; (ii) damages resulting from or under the conditions of labor disputes, strikes or
riots, insurrection, civil commotion or war; (iii) damages or improper operation due to intermittent power line voltage, frequency,
electrical spikes or surges, unusual shock or electrical damage; (iv) accident, fire or water damage, neglect, corrosive atmosphere or causes other than ordinary use; (v) failure of supplies or transportation, or governmental action or; or (vi) any other causes beyond
Seller’s reasonable control.
4. Limitation on Warranty Claims. Prior to any obligation of PAC to perform any limited warranty service as set forth herein, BUYER must have: (i) paid all invoices to PAC in full, whether or not they are specifically related to the product at issue; and
(ii) notified PAC of the limited warranty claim within sixty (60) days from the date BUYER knew or had reason to know of the defect.
III. Compliance
1. In General. These Warranty Terms are subject to change without notice. PAC also retains the right to modify these
warranty terms in order to comply with policy or laws governing warranty issues in states or countries having specific remedies differing or additional to those described within this document.
2. Severability. If any one or more of the provisions or subjects contained in the Agreement shall for any reason be held
invalid, illegal, or unenforceable, it shall not affect the validity and enforceability of any other provisions or subjects.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 5
1.5 Return Policy
No product may be returned, whether in warranty or out of warranty, without first
obtaining approval from PAC, L.P. No replacements will be provided nor repairs be
made for products returned without such approval. A return authorization number must
accompany any returned product. The buyer will pay for the expense of returning the
unit to PAC, L.P. for service. The status of any product returned later than 30 days after
the issuance of a return authorization number will be subject to review.
Products may not be returned that are contaminated by radioactive materials, infectious
agents, or other materials constituting health hazards to PAC, L.P.'s employees.
RETURNED PRODUCT WARRANTY DETERMINATION
After PAC, L.P.’s examination, warranty or out-of-warranty status will be determined.
If a warranted defect exists, the product will be repaired at no charge and shipped pre-
paid back to the buyer. Warranty repairs do not extend the original warranty period.
If an out-of-warranty defect exists, the buyer shall be notified of the repair cost. At such
time, the buyer must issue a valid purchase order to cover the cost of the repair and
freight, or authorize the product to be shipped back as is, at the buyer’s expense. Failure
to obtain a purchase order number approval within fifteen days of notification will result
in the product being returned as is, at the buyer’s expense.
ON-SITE REPAIR
If a PAC L.P. product cannot be made functional by telephone assistance or by
installing replacement parts, and cannot be returned to PAC L.P. for repair, the
following policy applies:
PAC L.P. will provide an on-site field service representative in a reasonable amount of
time, provided that the customer issues a valid purchase order to PAC L.P. covering all
transportation, subsistence, and prevailing labor costs, including travel time, necessary
to complete the repair. For warranty field repairs, the customer will not be charged for
the cost of transportation, labor, or materials. Special service rates may apply when
service is rendered at times other than normal work periods.
Inspection
This instrument was thoroughly inspected and carefully packed before leaving our
factory. Responsibility for its safe delivery was assumed by the carrier upon acceptance
of the shipment. Claims for loss or damage sustained in transit must be made to the
carrier by the recipient as follows:
PAC 7090 Series IOS Manual
Page 6 MAN 7090 Rev E
Visible Loss or Damage
Note any external evidence of loss or damage on the freight bill or express receipt, and
have it signed by the carrier’s agent. Failure to adequately describe such external
evidence of loss or damage may result in the carrier’s refusing to honor your damage
claim. The form required to file such a claim will be supplied by the carrier.
Concealed Loss or Damage
Concealed loss or damage means loss or damage which becomes apparent when the
merchandise is unpacked and inspected. Should concealed loss or damage occur, make a
written request for inspection by the carrier’s agent within 15 days of the delivery date,
then file a claim with the carrier since the damage is the carrier’s responsibility.
By following these instructions carefully, we guarantee our full support of your claim to
be compensated for loss from shipping damage.
DO NOT - for any reason - return the instrument without first obtaining
authorization. In any correspondence to PAC, please supply the nameplate data,
including catalog number and serial number. The nameplate is located on the rear of the
instrument housing.
Record of Serial Number and Purchase Date
Catalog Number: ________________________
Serial Number: ________________________
Purchase Date: _________________________
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 7
1.6 Contact Us
If you have any questions about your “7090 Series Nitrogen / Sulfur GC Detector”
contact us using the information below:
PAC L.P.
8824 Fallbrook Drive
Houston, TX 77064
WEB: http://www.paclp.com
Main Number: 281 940-1803
Service: 800 444-TEST (8378) or 281 580-0339
Fax Number: 281 580-0719
Email Support: [email protected]
Additional contact information for our regional offices is on the next page.
PAC 7090 Series IOS Manual
Page 8 MAN 7090 Rev E
Web: http://www.paclp.com
Phone: USA
France
Germany
Singapore
Netherlands
Russia & CIS
China
South Korea
Thailand
Abu Dhabi
+1 281 940 1803
+33 231 264 300
+49 9343 6400
+65 6412 0890
+31 10 462 4811
+7 495 617 10 86
+86 10 65072236
+82 2 785 3900
+66 2 627 9410
+971 2 446 9671
Fax: USA
France
Germany
Singapore
Netherlands
Russia & CIS
China
South Korea
Thailand
Abu Dhabi
+1 281 580 0719
+33 231 266 293
+49 9343 640 101
+65 6412 0899
+31 10 462 6330
+7 495 913 97 65
+86 10 65072454
+82 2 785 3977
+66 2 627 9401
+971 2 446 9672
Sales: USA
France
Germany
Singapore
Netherlands
Russia & CIS
China
South Korea
Thailand
Abu Dhabi
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
Service: USA
France
Germany
Singapore
Netherlands
China
South Korea
Thailand
Abu Dhabi
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
mailto:[email protected]
PAC 7090 Series IOS Manual
Page 10 MAN 7090 Rev E
2. SAFETY
2.1 Proper use and safety
The 7090 Series of Nitrogen / Sulfur GC Detectors is designed to be installed, operated and
maintained as specified below and in other parts of this manual. Failure to operate a 7090
analyzer as specified may produce erroneous data and possible equipment failure. Some of the
conditions for proper and improper use of the detector are provided as follows:
PROPER USE
Normal Use – The Model 7090 Series detectors are advanced analytical systems for the
specific determination of nitrogen and/or sulfur containing compounds in simple or complex
matrices. Operation for any other use, and the results obtained from any other use, cannot be
guaranteed.
Corrosive Environments – The standard configuration of the 7090 Series detector is not
intended for use in corrosive environments.
Direct Sunlight – The detector should not be exposed to direct sunlight or inclement
weather conditions.
Ambient Temperature – The best accuracy and precision is obtained when the detector
is located in a temperature controlled environment.
Electrical Power – The electrical power source for the detector must be capable of
delivering greater than 1200 Watts of power from either a 115-VAC or 220-VAC source. Each
7090 detector is built to be powered by either 115 VAC or 220 VAC. Care must be taken to avoid
connecting the detector to the wrong AC power supply.
Sample Analysis – To ensure the sample being analyzed is representative of the current
process status, sample to the detector must be delivered by an approved, leak-free sample
handling system. The sample handling system must eliminate or minimize “dead volumes” and
areas in the flow path where new sample can be mixed with old sample.
System Leaks – Gas leaks in the system will result in poor performance and possible
damage to the detector.
Light Leaks – Light leaks in the detector or PMT housing will increase background
noise.
Gas Quality – Air and other gases supplied to the detector must be filtered and pure as
specified.
Preventative Maintenance – In order to ensure proper performance, preventative
maintenance (PM) procedures must be performed at regular intervals.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 11
Personal Protective Equipment (PPE)
Wear safety glasses.
Wear nitrile gloves or equivalent type.
Wear appropriate dusk mask (when handling and cutting the RCF).
All Caution and Warning notations in this manual, on tags and in other publications pertaining to
the installation, operation and maintenance of the 7090 Series detector must be read and
observed. Failure to consider the Caution and Warning messages can lead to severe personal
injury and/or equipment damage. Operators are urged to observe standard operation and
maintenance safety practices and the specific safety warnings and labels in this manual and on the
equipment itself.
2.2 Certifications
The system and its components have been designed with safety in mind. The design applies to the
use, the conditions and the instructions described in the documentation. The system:
Meets the requirements of the European Directives concerning health, hygiene and safety. The
standards applied are listed in the “CE Declaration of Conformity”.
For a copy of the corresponding declarations, please consult PAC.
If the user makes changes or modifications that are not expressly approved by PAC, the company
reserves the right to void the user’s authority to operate the equipment.
During performance or routine maintenance, the system operator may be exposed to potentially
dangerous chemicals, temperatures, electrical voltages and/or other hazards. In order to reduce
the personal risk involved, the user must follow established guidelines and accepted standards for
general analytical laboratory operation.
For established safety requirements, government safety and disposal standards, please refer to the
corresponding Material Safety Data Sheet (MSDS) for each PAC-supplied product and also
reference the corresponding vendor supplied MSDS for each sample and standard utilized with
the product.
WARNING:
Failure to follow the instructions and procedures contained in this manual could
result in death or serious injury. The instructions and procedures must be
performed by qualified and trained personnel only. Do not perform installation
or maintenance operations or procedures outside the scope of this manual.
PAC 7090 Series IOS Manual
Page 12 MAN 7090 Rev E
2.3 Symbols
This symbol (exclamation point in a triangle) indicates a general Caution or
Warning. A Caution alerts you to information that is important for
protecting the equipment and performance. A Warning alerts you to
information that is important for protecting yourself, others and equipment
from damage.
This symbol (a lightning bolt in a triangle) indicates an electrical shock
hazard. Inside the 7090 Series are exposed live voltage circuits. The 7090
Series should be opened only by qualified service personnel. To prevent
electrical shock hazard DO NOT operate the 7090 Series in rain or wet
conditions.
This symbol (three wavy lines over a straight line in a triangle) indicates a
hot surface. During operation, the 7090 Series and associated equipment
surfaces get hot. To prevent burns exercise caution when operating the
detector.
This symbol (a fire in a triangle) indicates a fire hazard. During operation,
the 7090 Series surfaces near the furnace get hot. To prevent burns exercise
caution when operating the detector.
NOTE:
A note pictogram denotes a recommendation that, if not followed, may
result in minor damage.
TIP:
A tip pictogram denotes a recommendation that can help in understanding
how to operate or service the apparatus.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 13
2.4 Environmental Protection
Observe all local regulations and recommendations for the disposal and recycling of parts and
materials that have been used or replaced during installation, operation, and maintenance tasks.
PAC 7090 Series IOS Manual
Page 14 MAN 7090 Rev E
3. INTRODUCTION
As a gas chromatographic detector, the 7090 Series is the most advanced analytical system
available for the specific determination of nitrogen and/or sulfur containing compounds in simple
or complex matrices. The 7090 provides unrivaled selectivity and sensitivity to nitrogen and/or
sulfur by rendering most organic compounds transparent.
3.1 Intended Audience
The contents of this document must be read, understood and observed in all points by each person
responsible for the preparation, transport, storage, installation, operation, maintenance, and repair
of the detector, before starting work with and on the instrument.
This particularly applies to the safety instructions in this manual.
3.2 Principle of Operation
3.2.1 Principle of Operation: Nitrogen
The 7090N is capable of two distinctly different modes of operation for nitrogen determination:
1) Total Nitrogen Mode and 2) Nitrosamine Mode. The Total Nitrogen Mode can be operated
simultaneously with the Total Sulfur Mode, but the Nitrosamine Mode cannot.
The principle of operation for the Total Nitrogen Mode begins with the complete, high
temperature oxidation of the entire sample matrix as illustrated in the equation (1) below.
(1) R-N + R-H + O2 → CO2 + H2O + NO + SO2 + MOx
where R-N is any chemically bound nitrogen compound and R-H is any non-nitrogenous organic
compound.
Oxidation products include CO2, H2O, NO, SO2, and various other oxides (designated MOx
above). The conversion of all chemically bound nitrogen to NO (nitric oxide) is quantitative. The
combustion gases are routed to the detector module for quantitation.
As seen in equation (2), the NO is reacted with O3 (ozone), produced by an on-board ozone
generator, to form NO2* (nitrogen dioxide in the excited state). As the excited species decays to
the ground state, light is emitted and detected at specific wavelengths by a photomultiplier tube.
This chemiluminescent emission is specific for nitrogen and proportional to the amount of
nitrogen in the original sample. Only chemically bound nitrogen is detected; atmospheric nitrogen
(N2) is not sensed.
(2) NO + O3 → NO2* + O2 → NO2 + hν
The principle of operation for the Nitrosamine Mode begins with the selective, high temperature
cleavage of NO from nitrosamine compounds in the sample as illustrated in equation (3). As each
component elutes from the GC column, the component is combined with oxygen at a temperature
of approximately 500°C. At this temperature, only N-nitroso compounds will thermally cleave to
yield NO (nitric oxide) and the nitric oxide formation is quantitative. The furnace gases are routed
through a membrane drying system to remove all water and then to the detector module for
quantitation. The nitric oxide is quantitated according to equation (2).
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 15
(3) R-N-NO → NO + (pyrolysis products)
Nitrogen calibration standards may be analyzed to produce calibration curves. External data
systems may be interfaced to the PAC 7090 for calculations and documentation.
Figure 3-1 Connections for gas intake on a nitrogen detector
3.2.2 Principle of Operation: Sulfur
The principle of operation for sulfur detection begins with the complete, high temperature
oxidation of the entire sample matrix as illustrated in equation (4) where R-S is any organic sulfur
compound and R-H is any non-sulfur containing organic compound. As each component elutes
from the GC column, the component is reacted with oxygen at temperatures from 700 to 1100°C.
Oxidation products include CO2, H2O, SO2, and various other oxides (designated MOx below).
The conversion of chemically bound sulfur to SO2 (sulfur dioxide) is quantitative. The oxidized
gases are then reacted with hydrogen in the reductive zone of the furnace, generating H2S and
other reduced sulfur species as illustrated in equation (5). These reduced sulfur species are then
passed through the reaction chamber where they react with ozone.
(4) R-S + R-H + O2 → CO2 + H2O + SO2 + MOx
(5) SO2 + H2 → H2S + Other Reduced Species
According to equation (6), the H2S and other reduced sulfur species are reacted with O3 (ozone),
produced by an onboard ozone generator to form SO2 * (sulfur dioxide in the excited state). As
the excited species decays to the ground state, light is emitted and detected, at specific
wavelengths, by a photomultiplier tube. This chemiluminescent emission is specific for sulfur and
is proportional to the amount of sulfur in the original sample.
(6) H2S + Other Reduced Species + O3 → SO2* → SO2 + hν
Sulfur calibration standards may be analyzed to produce calibration curves. External data systems
may be interfaced to the PAC 7090 Series for calculations and documentation.
NO Sample
Ozone inlet
Sample vent
Pressure
Sensor
PAC 7090 Series IOS Manual
Page 16 MAN 7090 Rev E
Figure 3-2 Connections for gas intake on a sulfur detector
H2S Sample
Ozone inlet
Sample vent
Pressure
Sensor
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 17
Left Blank
PAC 7090 Series IOS Manual
Page 18 MAN 7090 Rev E
4. INSTALLATION
4.1 Pre-installation Requirements
Figure 4-1 Model 7090 Front view
4.1.1 Electrical Requirements:
Stable 115 VAC, 50/60 Hz, 1200 Watts or Stable 230 VAC, 50/60 Hz, 1200 Watts.
The PAC 7090 system utilizes stable 115 VAC, 50/60 Hz or 230 VAC, 50/60 Hz. A 10 amp
circuit is required for 115 VAC and 5 amp for 230 VAC. A good earth ground is necessary for
proper operation.
WARNING:
A poor ground may present a severe shock hazard. If a replacement power plug is
used, confirm that the wiring is correct and that the plug is rated to carry the
maximum current.
AC power is supplied via a three conductor power cord. This cord is fitted with a three-prong
plug rated for 12 amps. This plug will mate to most 115 VAC receptacles. For 230 VAC and
some 115 VAC operations, a different plug may be supplied. In all cases, care should be taken to
ensure that the color code, as shown in the following figure, is followed exactly.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 19
Figure 4-2 Follow the color code when making power connections
NOTE:
The analyzer is set for either 115VAC or 230 VAC operation.
4.1.2 Environmental and space requirements:
PAC 7090 systems can operate over a wide range of environmental conditions. However, for
optimum performance, the following conditions are recommended:
Temperature: 10 to 27°C (50 to 80°F)
Humidity: 25 to 75 % (no condensation)
NOTE:
Operation of the instrumentation outside of these conditions may result in
unsatisfactory and erratic performance.
A corrosive atmosphere will, in time, cause problems with solder joints, gold-plated contacts, and
other reactive surfaces. Likewise, dust, smoke, or other airborne particulate matter that may be
swept in by the cooling fans may collect on internal surfaces. This particulate accumulation can
retain moisture and cause short circuits or prevent thermal radiation from heat transfer surfaces
resulting in component damage.
Model 7090 Size
Height: ..................................................... 61 cm (24 inches)
Width: ...................................................... 36 cm (14 inches)
Depth: ...................................................... 59 cm (23 inches)
Weight: ................................................... 28 kg (62 pounds)
Approximately 60 cm (24 inches) of bench space is required for the PAC 7090 system. The bench
should be capable of supporting at least 28 kg (62 pounds). This weight does not include the gas
chromatograph (GC). Adequate air space should be provided at the top and behind the instrument
for proper ventilation. Allow a minimum of 15 cm (6 inches) for ventilation. Additional space
PAC 7090 Series IOS Manual
Page 20 MAN 7090 Rev E
may be required if optional equipment or accessories are attached or used in conjunction with the
7090 system.
CAUTION:
Improper ventilation of the instrument may result in overheating and damage to
system components.
WARNING:
The furnace of the Model 7090 gets extremely hot! Care must be taken to ensure
that no temperature sensitive materials (plastics, solvents, chemicals, autosamplers,
compressed gas, etc.) are exposed to the furnace. Do not touch any part of the
furnace until it is completely cool – hot metal parts can cause severe burns!
4.1.3 Materials Requirements
Additional materials may be required if optional equipment or accessories are
attached or used in conjunction with the 7090 system.
4.1.4 Materials Supplied by PAC:
The PAC 7090 system includes the following modules and accessories:
Model 7090 Detector (with 7090 Accessory Kit*)
P/N 86060 Specified Furnace, 7090
* A 7090 accessory kit includes all the components that are not installed in the
7090 instrument. These include items such as the furnace, power and
thermocouple cables, the gas lines, power cord, signal cable, pyrotube(s), filters,
etc.
Part # Description
99506N–115V Model 7090 CLND, 115 V
99506N–230V Model 7090 CLND, 230 V
99506S–115V Model 7090 SCD, 115 V
99506S–230V Model 7090 SCD, 230 V
4.1.4.1 Optional 2nd Detector (Add-on):
The Optional 2nd
Detector includes the following modules and accessories:
2nd
7090 Detector (with its own Accessory Kit)
P/N 86060 Additional Specified Furnace, 7090
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 21
Part # Description
101757 Model 7090 SCD, 115V, Add-on
101758 Model 7090 SCD, 230V, Add-on
101759 Model 7090 CLND, 115 V , Add-on
101760 Model 7090 CLND, 230 V, Add-on
4.1.4.2 Installation Kits:
Each 7090 detector or 7090 add-on requires an installation kit* to mount the 7090
furnace on the gas chromatograph (GC).
PAC makes available kits for a wide variety of GCs. Contact PAC for supported models.
*Installation kits include all the components that are required to adapt the 7090
furnace to the specific GC. Items may include adapter plates, furnace base
assembly, heater, and GC cover with shield.
4.1.4.3 Optional Equipment:
P/N 80.24.401/402 Vacuum Pump Assembly
The vacuum pump must be capable of a vacuum of 25 torr at a flow of 100
ml/min.
P/N 100259 Oxygen Purifier, 7090
* Gas Chromatograph (capillary column injector required)
* Accessories for capillary column analysis
4.1.4.4 Required Materials Not Supplied By PAC:
* Oxygen supply, dry, regulated to 3.0 bar (40 psig)
* Helium supply, dry, regulated to 3.0 bar (40 psig)
* Hydrogen supply, dry, regulated to 3.0 bar (40 psig)
4.1.5 Gas Requirements:
Dry* oxygen: ............................. 3.0 bar (40 psig), 99.75 %, 6-25 mL/min
Dry* helium: .............................. 3.0 bar (40 psig), 99.995 %, 0-20 mL/min
Dry* hydrogen†: ....................... 3.0 bar (40 psig), 99.995 %, 0-200 mL/min
*Dry is defined as 5 ppm weight water, maximum. †Hydrogen is required only for sulfur instruments.
Each gas line must be a dedicated line or should have an independent valve to turn off gases when
the system is not in use. A pressure gauge installed after the valve will enhance the ability to
properly perform a leak test of the system.
“Chromatographic” or “Zero” grade hydrogen and helium, and 99.75% grade oxygen are
recommended to ensure trouble-free performance. Gas suppliers have different grades and
designations for gas. Care should be taken when choosing the gases to be used.
PAC 7090 Series IOS Manual
Page 22 MAN 7090 Rev E
All gas streams must be regulated at 3.0 bar (40 psig) and in-line gas purifiers are absolutely
recommended for all gas supplies. All factory adjustments and settings are performed at 3.0 bar
(40 psig). However, in certain applications, the oxygen or helium line pressure may need to be
increased to 4.5 bar (60 psig).
WARNING:
High pressure gases should be handled with extreme care. Ensure that all relevant
safety precautions are carefully followed and that all gas lines, regulators, gas
purifiers, etc. are specified for the intended use.
All supply gases should be delivered through clean, dry aluminum, copper, or stainless steel
tubing. Tubing should be carefully washed and dried to ensure contamination free gas. Use
methylene chloride or trichloroethane followed by methanol or acetone and air drying to clean
new tubing. Other solvents may be used provided they are capable of dissolving oils or greases.
All solvents should be reagent grade or better.
CAUTION:
When using a liquid leak detector, care should be taken not to wet any
electrical or electronic components, or contaminate the analytical flow system.
Severe component damage or errant results may occur.
4.1.6 Plumbing Requirements:
All threaded connections (with the exception of compression type connections) should be
assembled carefully with Teflon® tape to guard against contamination and leaks.
All gases must be dry and free of particulate matter. Filters, gas scrubbers, and dryers are
available from PAC and are recommended in all cases.
Each supply gas cylinder requires a two-stage regulator to reduce the cylinder pressure to the
desired working pressure. PAC can supply the recommended regulators. `Care should be taken to
ensure use of the proper regulator for each gas in use.
Figure 4-3 illustrates a general flow diagram for the supply gases to an PAC 7090 system. This
procedure should be followed for each gas supply used (i.e., oxygen, helium, hydrogen, etc).
Sulfur Setup
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 23
Figure 4-3 Instrument setup for nitrogen, sulfur, or nitrogen/sulfur simultaneous
Nitrogen Setup
Sulfur Setup
PAC 7090 Series IOS Manual
Page 24 MAN 7090 Rev E
Nitrogen / Sulfur Setup
Sulfur or Nitrogen/Sulfur Simultaneous Setup
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 25
4.2 Installation Procedures
Select a suitable location that meets the criteria as stated in Section 4.1.
CAUTION:
Only trained personnel with prior knowledge of all pertinent safety procedures and
operations should be allowed to install, operate and service the instrument.
After determining the location for the 7090 Series instrument, unpack and install the analyzer by
performing the following procedures.
4.2.1 Unpacking and inspection
1. Unpack each item listed on the packing list.
2. Inspect items for damage.
3. Inspect the exterior of the detector for damage.
4. Place the detector onto the predetermined lab bench or table.
4.2.2 Shipping damage reports
Shipments are inspected immediately prior to being packaged for transportation. Since our
packaging exceeds all freight specifications, your equipment should have arrived in perfect
condition.
Our terms of sale are through Ex Works at shipping point, meaning that title to these goods is
passed to the final receiver when the goods are transferred to the carrier at our factory. If any part
of the shipment is damaged, we suggest reporting the damage immediately to the transportation
company. The following procedures are suggested for reporting damaged equipment.
Common Carrier:
If the item is discovered at the time of delivery, have the driver note the nature and extent of the
damage on the carrier’s copy and on the delivery copy of the freight bill.
If the item is discovered after delivery, call the carrier immediately after discovering the damage
and request that an inspection be made as soon as possible. Perform the following steps when
damaged items are found after delivery.
1. Contact the carrier immediately for an inspection of the item(s).
NOTE:
It is essential that the damaged item(s) be left in the original shipping container(s).
PAC 7090 Series IOS Manual
Page 26 MAN 7090 Rev E
2. Obtain a copy of the inspection report to support your claim.
NOTE:
Claims on air shipments must be filed within 270 days. Other claims must be filed
within 9 months of the shipping date.
3. File the claim as quickly as possible, even if you have to estimate the dollar
amount of the damage.
Parcel Delivery Service (UPS, DHL, FEDEX, etc.):
NOTE:
All claims must be initiated by the receiver.
Follow the steps listed above to report damages to shipments by common carriers, and notify our
office in Houston, Texas by dialing 800-444-8378. Our Technical Service Department will help
you process your claim.
NOTE:
Do not return the system or any item to the factory without factory authorization.
United Parcel Service (UPS):
NOTE:
All claims with UPS must be initiated by the receiver.
Unless otherwise specified, UPS is the preferred primary carrier for PAC shipments. Follow the
steps listed above for reporting damages caused when items are shipped by common carriers, and
notify our Customer Service Department, which will take the necessary action for your claim.
NOTE:
Do not return the system or any item to the factory without factory authorization.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 27
4.3 Physical Installation
4.3.1 Gases
Refer to the figure below to locate the gas inlet and outlet fittings on the rear of the 7090 system.
Figure 4-4 Gas Connections
Compression Fittings
Using the copper gas lines supplied with the 7090 system, connect the supply gases to the proper
gas inlet fittings. Refer to the figure below for the recommended assembly procedure for the
swage-type compression fittings included in the accessory kit.
Figure 4-5 Compression Fittings
PAC 7090 Series IOS Manual
Page 28 MAN 7090 Rev E
1. Insert tubing through nut.
2. Slide ferrule over the tubing.
3. Insert tubing with ferrule into the fitting until tubing bottoms in fitting.
4. Secure nut finger tight.
5. Use a wrench to turn the nut ¾” turn. After all gas fittings are connected, the
system should be carefully checked for leaks. Do not attempt to operate the
system until all leak checks have been performed.
4.3.2 Furnace Base
Select a suitable mounting location for the Furnace Base. Ensure that the furnace will not come
into direct contact with any electrical wiring or any plastic or flammable components on the gas
chromatograph or accessory equipment. The PAC Model 7090 may be installed on many
different gas chromatographs with several configurations. PAC provides installation kits for a
variety of GCs. Contact PAC for a list of available kits.
PAC installation kits contain an exploded assembly diagram to illustrate the mounting of base
assembly, furnace, and accessories on top of the GC.
WARNING:
The furnace of the Model 7090 gets extremely hot! Make ensure no temperature
sensitive materials (plastics, solvents, chemicals, autosamplers, compressed gas,
etc.) are exposed to the furnace. Refer to Figure 4-6 for mounting locations.
Figure 4-6 Furnace mounting on a typical gas chromatograph (GC)
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 29
4.3.3 Pyrotubes and furnace
1. Locate the outer ceramic tube (P/N 12.94.302), the outer pyrotube nut or long
nut (P/N 10.25.753), and the 1/4" graphite ferrule (P/N 10.40.012), insert the
tube into the base and secure. Tighten by hand and then one quarter to one half
turn using a 1/2" wrench.
2. Slide the furnace assembly (P/N 86060) over the ceramic tube and tighten it
until it cannot be turned any further. Connect the furnace, power, and
thermocouple cables and route them to connections on the back of the 7090.
NOTE:
Mount the appropriate installation kit on top of the GC before installing the ceramic tube.
Figure 4-7 Furnace power and thermocouple connections
For Nitrogen detectors, perform the following steps.
1. Attach the 1/8” x 6 cm long tube (P/N 10.50.043) to the treated ¼” to 1/8”
reducing union with nut (P/N 10.10.042T).
2. Attach the filter (P/N 40.50.007) to the end of the 6 cm tube.
3. Attach the ¼” end of the reducing union with nut to top of the ¼” ceramic tube
with ¼” graphite ferrule (P/N 10.40.012).
4. Attach the nitrogen detector sample line from the 7090 to the top of the filter.
For Sulfur detectors, perform the following steps.
1. Disassemble the treated Tee fitting to make sure it has a ¼” graphite ferrule on
one end and a 1/8” graphite ferrule on the other end.
2. Attach the filter (P/N 40.50.007) to the shorter 6 cm tube.
PAC 7090 Series IOS Manual
Page 30 MAN 7090 Rev E
3. To the longer 10 cm tube, attach the 1/8” to 1/16” reducing union (P/N 30057).
4. Slide the 1/8” graphite ferrule over the 1/8” ceramic tube and slide the tube into
the Tee fitting.
5. Screw the 6 cm line and tighten to swage the ferrule holding the 1/8” ceramic
tube in place.
WARNING:
Over tightening the graphite ferrule will cause the 1/8" ceramic tube to break,
and the instrument will not detect sulfur compounds.
6. Loosely re-install the larger ¼” graphite ferrule and nut. Slide the whole
assembly and attach it to the ¼” ceramic tube. Tighten the nut to secure
assembly in place making sure that the 10 cm long line (the hydrogen feed tube)
points towards the back of the GC.
7. Attach the hydrogen gas line from the reducer fitting on the 10 cm line to the
hydrogen out fitting on the back of the 7090.
8. Attach the sulfur detector sample line from the 7090 to the top of the filter.
4.3.4 Inlet adapter
1. Remove the Column Adapter Assembly components from the Adapter Kit
P/N 19119 and assemble as follows:
Detector – Column Adapter Assembly P/N 19119
Insert the glass lined tubing, seal and straight union into the Detector Adapter (P/N 30380). Screw the union on finger tight.
Glass Lined Tubing (P/N 30429)
Sealing Ring (P/N 30140)
Straight Union (P/N 30282)
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 31
Insert the Detector Stem Adapter into
the bottom of the furnace base from the
GC oven side.
Push the Adapter up until it stops in the
furnace base.
Tighten the ¼” SS nut securely to
hold the Adapter in place.
PAC 7090 Series IOS Manual
Page 32 MAN 7090 Rev E
The capillary column installation is detailed in 6.1.3.3 after all system checks have been
performed.
4.3.5 Gas transfer lines
Route the Oxygen Transfer Line from the Base Assembly to the back of the 7090 and connect to
the Oxygen Out fitting.
4.3.6 Signal output cables
The Model 7090 is not equipped with any integration capabilities and requires some means of
interpreting data; therefore a recorder or integrator is required. Refer to the requirements for the
proper voltage input. Plug the recorder cable into the lmv, lv, or 10v connector located on the rear
panel of the 7090. See the following for info.
The color code is as follows:
RED = POSITIVE (+)
BLACK = NEGATIVE (-)
CLEAR or GREEN = SHIELD
Left Rear of 7090
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 33
4.3.7 Vents
The Model 7090 has an Ozone Scrubber to remove all ozone produced in the system. Venting of
the gases should produce no problem under normal situations, but if your sample contains any
potentially hazardous chemicals we recommend venting the vacuum exit into a hood to reduce
safety hazards. Check your local regulations regarding venting of gaseous materials.
4.3.8 Injection systems
Due to the nature of the samples, they can be easily absorbed onto metal surfaces. Care should be
taken to avoid the use of highly absorptive metals such as brass, copper, aluminum, etc. Use of
inert tubing is recommended in any cases where there is direct contact with the sample itself, such
as when utilizing a gas or a liquid sample valve. We also caution that most injector systems are
made of stainless steel components and sample contact with these metals is possible when using a
valve or syringe injection. When using a sample valve, we recommend completely bypassing the
injector.
4.3.9 Vacuum Pump
A vacuum pump is required in the operation of the 7090. The pump must be capable of a vacuum
of 25 torr at a flow of 100 ml/min. Do not connect the column to the detector base until a
preliminary leak check has been performed. Install a capillary plug fitting to the detector base as
shown in Figure 6.1. Do not turn the instrument on until a leak check of the entire system has
been performed (see Chapter 6).
WARNING:
Removal of the vacuum pump line from the scrubber assembly while furnace
and/or detection chamber are under vacuum could allow particulates to be
pulled into the reaction chamber.
PAC 7090 Series IOS Manual
Page 34 MAN 7090 Rev E
5. SYSTEM OVERVIEW
Before continuing with the setup of the PAC 7090 system, become thoroughly familiar with the
controls, connections, and general layout of the system.
NOTE:
For specific information concerning the function or operation of controls or
connections, consult the appropriate section of this manual or its addenda.
5.1 Front panel controls
This section explains the front panel controls and indicators to manipulate and monitor most of
the 7090 system’s analytical parameters, including detector sensitivity and baseline manipulation.
Figure 5-1 identifies the major front panel components and is followed by a brief explanation of
each.
NOTE:
All controls, except auto zero, are automatically initialized to the default settings upon
power-up. For more detailed information on the specific function(s) and
operation of the detector controls, refer to Chapter 8.
Figure 5-1 Front panel displays for the nitrogen analyzer (left) and sulfur (right)
1. Displays: for either high voltage set point or signal output.
2. Switch: for display of high voltage set point or signal output.
3. Sensitivity Selector: x1, x10, x25, x50 amplification factor for the Signal
Processing printed circuit board. The gain factor is used to increase or decrease
the sensitivity of the detector electronics. This is a multiplicative factor;
therefore, a factor of x50 is 50 times more sensitive than a factor of x1. This
control affects all analog outputs.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 35
4. Gain: The gain switch is the primary sensitivity control for the photomultiplier
tube preamplifier. HI gain is 100 times more sensitive than LOW gain. This
control affects all analog outputs.
5. Auto Zero: The auto zero button, when pushed, will automatically zero the
baseline of the detector. This function can be activated either manually or
remotely.
6. Ozone Generator On/Off Switch: Powers the ozone generator ON or OFF.
7. High Voltage On/Off: Powers the PMT voltage ON or OFF.
8. Nitrogen PMT Temperature Controller: See Section 9.2.1 for programming
procedures.
9. Furnace Temperature Controller: See Section 9.1 for programming
procedures.
10. Display LCD: Displays the value for the selection of furnace vacuum, detector
vacuum, ozone flow rate, oxygen flow rate, hydrogen flow rate, or furnace
vacuum set point.
11. Display Select: The Display Select Switch allows the user to scroll through
each of the six displays (Figure 5-2). Possible displays:
Figure 5-2 Status Lights
Furn Vac Displays the vacuum at the base of the furnace in
torr.
Det Vac Displays the vacuum in the Reaction Chamber in
torr.
O3 Displays the flow rate of ozone into the reaction cell.
O2 Displays the flow rate of oxygen into the furnace.
H2 Displays the flow rate of hydrogen into the furnace.
Furn Vac Set Displays the vacuum set point at which the Furn Vac
status light turns green.
12. Status Lights (six): The Status Lights provide a quick
inventory of the status of the instrument. Each light will change
from red to green as that portion of the system reaches a standby
mode and is ready for standard operations. All greens indicate that
the instrument is ready for normal operation.
O2 Oxygen Pressure LED: Indicates if the oxygen pressure is 30
psi or above. If oxygen is not present, the system cannot heat and
the power to the Ozone Generator will be turned off. If supply
pressure drops when a system is at temperature, the system will
cool down and the hydrogen solenoid will immediately turn off and
all lights will turn off except for the O2 light which will be red.
O3 Ozone On/Off Status LED: Indicates the current status of the
Ozone Generator.
The Generator will not operate if the oxygen pressure drops below
30 psi.
TEMP Temperature Status LED: The temperature status light is
designed to turn green when the instrument reaches a temperature
above 750°C and below 1100 °C.
PAC 7090 Series IOS Manual
Page 36 MAN 7090 Rev E
Furn Vac Furnace Vacuum LED: Furnace Pressure Status. Indicates the vacuum at the base
of the furnace is within limits. Typically this value in a system with no leaks is less
than 180 torr with an adequate vacuum pump. The light will not turn green until the
vacuum drops below 250 torr. The system will not heat if this light is not green.
H2 Hydrogen Status LED: The power to the hydrogen solenoid is set at the factory to
automatically turn on when the furnace reaches a temperature of 750 °C. If oxygen
pressure drops, the hydrogen solenoid will be automatically turned off, and the
LED will turn red, indicating that no power is applied.
Ready System Status LED: Indicates that the system is ready for use.
13. Mass Flow Controller Adjustments: The flow controllers are located behind the
front access door and can only be adjusted with a small screwdriver. To adjust
the flow controllers, insert the screwdriver into the slotted adjustment. Turning
clockwise will increase the flow and counterclockwise will decrease the flow.
14. High Voltage Power Supply (HVPS) Adjustments: The HVPS can only be
adjusted with a small screwdriver. To adjust the HVPS, insert the screwdriver
into the slotted adjustment for the correct detector, DET A is sulfur and DET B
is nitrogen. Turning clockwise will increase the voltage and counterclockwise
will decrease the voltage supplied to the PMT.
5.2 Rear panel
The following is a brief explanation of the function and operation of the rear panel controls and
connections of the PAC 7090 system. These controls and connections are primarily used for the
manipulation of gases, electrical power, external control cables, and signal distribution.
Most electrical and electronic connections are made from the rear panel of the 7090 system.
Connectors are present for Main Power, Remote Auto Zero, Detector Output Connections, and
Pyro Furnace and Thermocouple Control. Also included on the rear panel are the cooling fan,
vent, and fuses. Some electronic connections may be located on peripheral equipment. Refer to
Figure 5-3 for location of the major components on the rear panel.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 37
Figure 5-3 Rear panel
1. Furnace Thermocouple Connector: Interface connector between the 7090 system and the Furnace
Assembly Thermocouple.
2. Cooling Fan: This fan provides for removal of excess heat from the 7090 system.
3. Furnace Power Connector: Interface connector between the 7090 system and the Furnace
Assembly
PAC 7090 Series IOS Manual
Page 38 MAN 7090 Rev E
4. Scrubber Tube Assembly: The Scrubber Tube contains a catalyst for the complete elimination of
ozone prior to the vacuum pump. Do NOT attempt to operate the system without a Scrubber
Tube.
5. Detector Output Connections: Interface connectors for analog outputs of detector signals. Signal
levels of 1 millivolt, 1 volt, and 10 volts are available. Each connector is keyed for proper
polarity.
6. Sample Transfer Line: The Sample Transfer Line is a direct connection from the exit of the Furnace to
the Sample In fitting of the Reaction Chamber Assembly.
7. Plumbing Connections: (Refer to Figure 4-4.)
Oxygen Connect 3.0 Bar (40 PSI) oxygen source here.
Hydrogen* Connect 3.0 Bar (40 PSI) hydrogen source here.
Helium* Connect 3.0 Bar (40 PSI) Helium “He” source here.
Hydrogen Out The “H2 Out” line connects directly to the base of the furnace.
Oxygen Out The “O2 Out” is connected via the Oxygen Transfer Line to the
makeup fitting on the Detector Inlet.
*The hydrogen and the helium are only used on the sulfur detector.
Figure 5-4 Fuses
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 39
Fuse Description 115 VAC 230 VAC
1 & 2 Main Fuses (8 amp) (5 amp)
3 Fuse 70 VAC (transformer secondary) furnace voltage
(10 amp) (10 amp)
4 Fuse furnace transformer (8 amp) (5 amp)
5 Fuse 24 VAC (2 amp) (2 amp)
6 Fuse 24 V Transformer (2 amp) (1 amp)
7 Fuse Temperature Controller (1 amp) (1 amp)
8 Fuse 5 VDC Power Supply Ozone Generator (1 amp) (1 amp)
9 Fuse 15 V Power Supply Mass Flow Controllers (1 amp) (1 amp)
10 Fans (1 amp) (1 amp)
11 Detector Trays
(1 amp) (Slow-blo)
(1 amp) (Slow-blo)
NOTE:
The main fuses are located in the fuse drawer situated in the housing of the power
receptacle. Use a small screwdriver to release the latch and the drawer will slide out
easily.
5.3 Internal components
Below is a brief description of the major components and their functions:
PAC 7090 Series IOS Manual
Page 40 MAN 7090 Rev E
Figure 5-5 Internal Components
1. Signal Processing Board: The PMT signal is manipulated on this board by
selecting proper gains and ranges. The baseline controls are also performed by
this board.
2. Power Distribution Board: These boards contain all the fuses and provide the
main power distribution for the entire system.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 41
3. Display Board: The Display Board interprets all the conditions of the system
and gives the instrument user a current status at a glance. It contains all the
switches for operation and is located behind the front door access It is also
responsible for the measurement of the vacuum in the reaction chamber and at
the base of the oxidative/reductive furnace.
4. Signal Output Board: This board has the divider network to allow the signal to
be monitored as 10 volts, 1 volt, or 1 mv.
5. Ozone Generator Board: This board provides the circuitry to pulse the
transformer and produce ozone.
6. High Voltage Power Supply Board: This board provides the stable voltage to
operate the photomultiplier tube. This voltage can be monitored on the front
panel.
7. PMT Housings: The photomultiplier tubes are encapsulated here in a light free
environment. This housing should never be opened with voltage applied or
permanent damage to the PMT can occur.
8. Reaction Chamber: This is the mixing chamber for the reaction to occur by
uniformly mixing the ozone and the sample.
9. Hydrogen Solenoid: This solenoid is controlled so that it is energized only
when the furnace has reached the proper operating temperature and O2
pressure is above 30 psig.
10. Transformer: This transformer provides the 24 VAC to operate the HVPS
Board.
11. Transformer: This transformer provides the 70 VAC to operate the furnace.
12. 5V Power Supply: This provides the stable 5 VDC to operate the Ozone
Generator.
13. Ozone Generator: The Ozone Generator converts the oxygen into ozone
required for the detection of sulfur species.
14. Mass Flow Controllers: These control the flow of gases to the furnace and
ozone generator.
15. Temperature Control: This is the control for the furnace.
16. Testpoints, Sulfur (above) & Testpoints, Nitrogen (below): These are the
testpoints for ±15 VDC and 5 volt power supplies.
17. Line filter: reduces noise from random spikes from power source.
18. Nitrogen Thermoelectric Cooler and Fan: keeps PMT cool for nitrogen
detector.
19. 15V Power Supply: This provides the stable 15 VDC to operate the Mass Flow
Controllers.
5.3.1 Vacuum pump
The system requires a stable vacuum pump to operate. The pump must be capable of a vacuum of
25 torr at a flow of 100 ml/min. (See Section 4.3.9) The vacuum pump is connected directly to
the Scrubber Tube exit fitting. Refer to Figure 5-6.
PAC 7090 Series IOS Manual
Page 42 MAN 7090 Rev E
Figure 5-6 Vacuum pump connection
5.4 Furnace
The following is an explanation of the function and operation of the Furnace Module on the PAC
7090 system. Figure 5-7 illustrates the location of the main features and connections of the
furnace.
Figure 5-7 Furnace
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 43
1. Outer Pyrotube: Ceramic tube for sample oxidation
2. Inner Pyrotube: Ceramic tube for sample reduction
3. Thermocouple: Senses the furnace temperature
4. Furnace Power Cable: Supplies power to the heater element
5. Furnace Shield: Insulates the Furnace Assembly
6. Base Assembly: Mounts to the GC and connects the GC column to the furnace
7. Oxygen Transfer Tube (Sulfur Only): Supplies oxygen and helium to the
detector base of the furnace
8. Hydrogen Inlet Tee: Supplies hydrogen to the inlet
PAC 7090 Series IOS Manual
Page 44 MAN 7090 Rev E
6. PRIMARY TESTS
6.1 Pressure Tests
The gas flows for the PAC 7090 system will generally depend on the GC column(s), sensitivity
desired, and the specific application.
CAUTION:
To prevent system damage, all system components and accessories should
be properly installed and carefully checked for leaks before final gas flow
adjustments are made and the capillary column is installed. The gas supplies
should be connected as outlined in Section 4.3.1 of this manual.
Each gas line must be a dedicated line or have an independent valve to turn off gases when the
system is not in use. A pressure gauge installed after the valve will enhance the ability to properly
perform a leak test of the system.
6.1.1 Gas supply (External) leak check
1. Power off 7090 and turn off helium supply.
2. Turn on all other supply gases and set pressure to 3.0 bar (40 psi).
NOTE:
The H2 flow will display 00 until the instrument is fully installed. Record this setting
in Chapter 7.
Table 6.1
Factory Settings
Main Fuses
Oxygen Flow controller
Fuse furnace transformer
Hydrogen Flow controller
Fuse 24 VAC
Ozone Flow controller
Helium Makeup
CAUTION:
Do not over tighten shut-off valve. Turn counterclockwise only until resistance
is felt. Over tightening may damage the flow control devices.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 45
3. Turn off each gas at the valve located prior to the system and monitor the
pressure gauge. No noticeable pressure drop should be observed within five
minutes. If leak(s) are present, they are located prior to the system’s flow
controllers and are more than likely located in the manifolds providing gases
to the analyzer. Locate the leaks by testing each fitting with a liquid leak
detector such as SnoopTM
. Tighten the leaking fittings only enough to stop the
leak. Over tightening will damage the ferrule and fitting. Do not attempt to
operate the system or the gas lines until the leak(s) are located and
eliminated.
CAUTION:
When using a liquid leak detector care should be taken not to wet any electrical
or electronic components. Severe component damage may occur.
6.1.2 Vacuum internal leak check
1. Open supply gas valves and set regulators for 40 psi.
2. Perform a leak check of the external supply gases as described in the
previous section.
3. Ensure that there is a capillary plug fitting as shown in Figure 6-3.
4. Ensure that all plumbing connections are properly installed as described in
Section 4.3.
5. Ensure that the 7090 system is plugged into the proper power supply
(115 VAC or 230 VAC).
6. On the rear panel, turn on the Main Power.
7. Turn off ozone switch(es). Use the display select to scroll to O2 and O3 flows
and set flows to 0.
8. Turn on Helium supply.
9. On the front panel, use the display select to scroll to the FURN VAC display.
10. Ensure that the furnace temperature control is set to 25°C as described in
Section 6.3.3 so that the furnace will not heat until a proper leak check has
been performed.
11. Turn vacuum pump power on.
12. Watch the FURN VAC display to ensure that the displayed value is dropping.
CAUTION:
Do not use any liquid leak detection fluids while under vacuum.
PAC 7090 Series IOS Manual
Page 46 MAN 7090 Rev E
13. If there are no leaks, the FURN VAC and DET VAC values will read a Torr
value of less than 20. Record the value in Table 6.2 below. Record the DET
VAC value in Table 6.2. If there are no leaks, turn off vacuum and proceed to
Section 6.1.3.
Table 6.2
FURN VAC, torr _______________________
DET VAC, torr _______________________
14. If the values do not drop below 20, there is a leak that must be located prior to
operating the system or applying heat to the furnace.
15. If a leak is present, you may need to perform a more in depth leak check using
pressure to isolate the location of the leak.
16. The likely places for leaks are a break in the 1/ 4" ceramic pyrotube or the
Teflon detector sample inlet line. Carefully check these places first.
NOTE:
Internal vacuum sensors (U4 and U6) should be calibrated yearly to ensure optimal
performance.
6.1.3 7090 pressure leak check procedure
Pressure leak check preparation: Turn off ozone power, HVPS, and H2 and Helium flows. Set
furnace temperature to 0°C. Once furnace is below 200°C, turn off vacuum pump and allow
furnace vacuum reading to reach ambient pressure.
6.1.3.1 Total 7090 System leak check
1. Lower Oxygen pressure supply to 5-10 psi.
2. Adjust O2 and O3 flows to 25 ml/min.
3. Remove capillary column from base adapter and plug adapter input with the
blanking nut and ferrule provided in adapter kits.
4. Remove 1/4" Teflon line from exit of scrubber mounted at rear of 7090 and
plug the scrubber exit with 1/4" Swagelok plug.
5. Monitor both ozone and oxygen flows for several minutes.
a. If both oxygen and ozone flow readings drop to 0.0; the system has
passed the pressure leak check. Remove plugs, raise O2 pressure back to
normal (40 psi) and install column to base adapter.
b. If either oxygen or ozone flows do not drop to 0.0; a leak is present and
must be repaired. If the location of the leak is not known, use the
following section “Leak checking isolated components” to pinpoint the
leak.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 47
6.1.3.2 Pressure leak checking isolated components
This section will help locate a leak. Once the leak is repaired, go back to Section 6.1.3.1 and
verify that the total system is leak free.
A. Leak checking the furnace:
1. Remove 1/8" Teflon line from furnace exit and plug furnace exit with an 1/8"
Swagelok plug.
2. Monitor only the oxygen flow reading:
*If oxygen flow drops to 0.0; the furnace and connected lines are leak free. Go to step 3.
*If oxygen flow does not drop to 0.0; the leak is present at furnace or connecting lines. Reading
the oxygen flow will also indicate the size of the leak(s) you are looking for. After fixing a
furnace leak, proceed to step 3.
3. Remove plug from furnace sample vent and reconnect Teflon line back to
furnace sample vent.
B. Leak checking ozone assembly:
1. Disconnect union in Teflon line between ozone assembly and reaction
chamber.
2. Install 1/8" Swagelok plug on downstream side of union.
3. Watch for ozone flow reading.
* If ozone flow drops to 0.0, ozone assy. is leak free, go to step 4.
* If ozone flow does not drop to 0.0, leak is present in ozone generator assy.
or line to the assembly. After fixing leak, go to step 4.
4. Remove plug from union and reconnect ozone line to reaction chamber line.
C. Leak checking reaction chamber:
1. Ensure the PMT HVPS is switched off!
2. Remove 1/4" Teflon exit line from reaction chamber. Plug the exit fitting of
reaction chamber with 1/4" Swagelok plug.
3. Monitor the oxygen and ozone flow readings.
* If both flows drop to 0.0, reaction chamber, furnace and ozone assemblies are
leak free, go to step 4.
* If either flow does not drop to 0.0, leak is present in reaction chamber or
Tygon® line to detector vacuum sensor. Fix leak and go to step 4.
4. Remove 1/4 " plug from reaction chamber and reconnect 1/4" line to scrubber.
5. Leak check complete system again as in Section 6.1.3.
6. If a leak is still present, inspect the scrubber tube assembly. Tighten or reseal
the scrubber tube and check complete system again.
7. Lower oxygen pressure supply to 5-10 psi.
8. Remove capillary column from base adapter and plug adapter input with the
blanking nut and ferrule provided in adapter kit.
9. Remove 1/4" line from exit of scrubber mounted at the rear of the 7090 and
plug the scrubber exit with 1/4" Swagelok plug.
10. Monitor both ozone and oxygen flows.
PAC 7090 Series IOS Manual
Page 48 MAN 7090 Rev E
6.1.3.3 Capillary column installation
1. Remove the capillary plug.
2. Ensure that the proper flow is set on the capillary column.
NOTE:
The procedure is only necessary for fused silica columns.
3. Insert the capillary column 15 cm into the furnace. The distance is measured
from the tip of the column to the back of the nut.
6.1.4 Resetting system to factory defaults
1. Referring to Table 6.1, reset the Mass Flow Controllers on O2, H2, and O3 to
factory defaults.
NOTE:
H2 will not show any flow until the system is fully operational.
6.2 GC Power-up
Refer to your gas chromatograph manual and perform any diagnostic tests as required by the
manufacturer prior to the operation of the gas chromatographic system. Upon completion of these
tests, ensure that all gas chromatograph carrier gas flows are set, and that there is flow through
the capillary column without any leaks. Heat the capillary injector and the column oven to the
proper temperature. Refer to the column manufacturer's guidelines to establish the proper
temperatures. Heat the 7090 detector base to 280ºC.
6.3 Model 7090 Power-up
6.3.1 Check list
( ) Proper operating voltage determined? (Section 4.1.1)
( ) System located properly? (Section 4.1.2)
( ) Accessories (if any) installed?
( ) Gas line(s) installed? (Section 4.3.1)
( ) Pyrotubes and furnace installed properly? (Section 4.3)
( ) Leak check OK? (Section 6.1)
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 49
( ) All power switches OFF?
( ) Electrical/electronic interconnections correct? (Section 4.3.6)
( ) Power cord plugged into power source? (Section 4.1.1)
( ) GC flows are set and flowing? (Section 6.2)
( ) Vacuum pump connected and turned OFF?
( ) Hydrogen Valve OFF (Section 2)?
( ) Oxygen Valve OFF (Section 2)?
( ) Mass flow controllers returned to their original settings? (Section 6.1.4)
( ) All blanks filled in the prior sections of this manual? (Tables 6.1, 6.2)
CAUTION:
Become familiar with the location and use of all controls, indicators,
connections, and accessories, and carefully read all instructions before
attempting to operate any portion of the system.
6.3.2 Power-up and functional tests
Ensure that all physical setup procedures as outlined in Section 6.1 have been completed before
continuing with power-up and functional test sequence. Failure to do so may result in serious
damage to system components.
CAUTION:
Ensure that the system and all optional equipment and accessories are
connected to a power source of the correct voltage and wattage. Attaching
any portion of the system to an improper or inadequate power source may
cause severe damage.
Ensure that all power switches on the front panel are in the OFF position. Refer to Figure 5-1 for
locations of the switches on the basic system.
1. Turn the system Main Power Switch to the ON position. There should be an
audible sound of the circulation and cooling fans turning on. Upon power-up,
the Temperature Controller should start flashing the furnace temperature and
will alternately flash Alarm. This is due to the fact that the hydrogen solenoid
has not been energized and that the furnace is at a temperature below 920 °C.
2. Only the O2 Status LED should all be lit at this time. It should be red.
3. Turn on the Oxygen Supply Valve. The O2 LED should turn green. All others
should be red.
4. Turn on the Vacuum Pump. As the proper vacuum is obtained, the furnace vac
LED will fade from red to green if there are no leaks present and the flows are
properly adjusted.
PAC 7090 Series IOS Manual
Page 50 MAN 7090 Rev E
6.3.3 Furnace power-up
NOTE:
The furnace will not heat unless the FURN VAC and O2 LEDs are green.
Increase the furnace temperature to 950 °C by pressing the (*) key and simultaneously pressing
the up () arrow. The longer the up () arrow is pressed, the faster the number will increase. If
the value of 950 is missed, use the down () arrow and the (*) to reduce the set temperature. The
furnace is an isothermal control zone and is a standard feature with all 7090 systems. The primary
function of the furnace is to completely and quantitatively oxidize and reduce the sample. This
complete oxidation/reduction can only be successfully accomplished at very high temperatures.
The temperature range of the furnace is adjustable from ambient to 1100°C in 1°C increments.
NOTE:
As the furnace temperature increases past 925ºC, the hydrogen solenoid will
automatically be energized and the H2 LED will turn green. The Temp LED
will also turn green at this temperature.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 51
Figure 6-1 Nitrogen only system
PAC 7090 Series IOS Manual
Page 52 MAN 7090 Rev E
Figure 6-2 Sulfur only system
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 53
Figure 6-3 Simultaneous nitrogen/sulfur system
PAC 7090 Series IOS Manual
Page 54 MAN 7090 Rev E
7. START-UP BASICS
1. Ensure that all status lights except for O3 are green and that all temperatures are correct
and stable.
2. Ensure that the Signal Output Cable is installed as described in Section 4.3.6.
3. Press the Auto Zero button on the front panel.
4. Record the mass flow controller settings in Table 6.1.
7.1 Baseline adjustments
1. Select SIGNAL on the LED display.
2. Turn on the integrator or other recording device to plot the baseline.
3. The baseline should be relatively close to 0.0 mV.
NOTE:
A different baseline can be established by manually zeroing the baseline with the
Manual Zero Switch located on the front panel of the 7090NS.
4. Select high gain.
5. Select a sensitivity setting of x10.
7.2 High voltage
1. Select H.V. on LED display. The H.V. display should read relatively close to 0.
2. Turn the High Voltage Switch to the ON position. The H.V. display should read 700 or
more.
3. Select Signal on the LED display.
4. Monitor the baseline to ensure that there is a shift (positive) in the baseline.
5. Press the Auto Zero button on the front panel.
7.3 Ozone generator
6. Record Signal on 7090NS front panel display in Table 7.1.
7. Turn the Ozone Generator Switch to the ON position. (The O3 LED should turn green.)
8. Monitor the baseline to measure any shifts in the baseline. Allow at least 15 minutes for
the ozone generator and analytical system to stabilize.
9. Record Signal on 7090NS front panel display in Table 7.1.
TABLE 7.1
Signal with O3 off _______________________________________
Signal with O3 on _______________________________________
10. Press the Auto Zero button on the front panel.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 55
Left Blank
PAC 7090 Series IOS Manual
Page 56 MAN 7090 Rev E
8. APPLICATIONS AND ANALYTICAL PROCEDURES
8.1 Important basics
This section describes the general operation of the PAC 7090NS system. Included in this section
is information on controls, connections, functional tests, troubleshooting and other basic
operating parameters. For detailed information on the function and operation of specific controls
or modules, refer to the applicable section (s) of, or addenda to, this manual. Since the PAC
7090NS system is capable of analyzing a wide variety of materials for sulfur content, the
analytical parameters may vary. The operation conditions given in this section are general and
should be used only as guidelines.
CAUTION:
Do not attempt to operate the 7090NS system until all installation and setup
procedures outlined in this manual have been successfully completed.
Figure 8-1 PAC 7090NS system
8.1.1 Gas and carrier flows
Most gas flow adjustments are performed and monitored on the front panel of the 7090NS. Mass
flow controllers are provided for Pyrolysis Oxygen, Ozone Oxygen, and Hydrogen. Some flow
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 57
adjustments will be located on the GC to which the 7090NS system is attached. For detailed
information concerning these flow controls, please refer to the GC manual. Refer to Section 5.1
for location of the major components of the front panel of the 7090NS.
NOTE:
For more detailed information on the specific function(s) and operation of the
front panel controls, refer to Section 5.1.
NOTE:
Please refer to the 7090NS TEST Conditions Sheet for your instrument flows.
These are typical starting flows for a 7090NS system. The 7090NS system is factory calibrated
with thiophene in toluene and/or indole in toluene check standards. The actual mass flow
controller settings of your system are recorded in the attached 7090NS Test Conditions Sheet.
Please refer to this sheet as a starting point.
TABLE 8.1
Gas MFC Setting
Ozone Oxygen 25 mL/min.
Pyro Oxygen 8 mL/min.
Hydrogen 250 mL/min.
Oxygen Flow Measurement
Measure the oxygen without any power to the furnace. The flow is measured at the Oxygen Out
on the rear panel (refer to Figure 5.3).
Hydrogen Flow Measurement
The instrument must be turned on and ready for operation to be able to measure the hydrogen.
The oxygen must be at 30 psi minimum, the P1 torr value below 250, and the furnace above
750°C for the hydrogen solenoid to be energized. The flow is measured at the Hydrogen Out on
the rear panel.
Chromatographic Column Flow Measurements
Refer to the manufacturer’s Column Information Sheet for optimum conditions and optimum
column operating parameters.
PAC 7090 Series IOS Manual
Page 58 MAN 7090 Rev E
8.1.1 System flow considerations
Operating the 7090NS system with excess flows can reduce the ability of the pump to maintain
adequate pressure, and therefore, analytical results may vary. In a normally operating system, the
system flow considerations are as follows:
Furnace Temp = 950ºC (Top – Displayed)
= 1050ºC (Bottom – Not Displayed)
O2 Flow = 4 to 12 mL/min
H2 Flow = 110 to 350 mL/min
He Flow = 90 to 150 mL/min
O3 Flow = 20 to 90 mL/min
P1 Furnace Pressure = 80 to 350 torr
P2 Detector Pressure = 8 to 40 torr
Column Flow = 1 to 25 mL/min
Capillary Column = 0.100 to 0.530 mm
Figure 8-2 Normal operating system
8.1.2 System gas flow adjustments
The 7090NS system requires oxygen and hydrogen, (for sulfur mode only), to operate properly.
Oxygen is used in the oxidation step of the pyroreactor, and hydrogen is used in the reduction
step of the pyroreactor. See Section 3.2, equations (1) and (2) for reactions.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 59
8.2 Optimizing sulfur detection
1. Ensure that all flows are set according to factory preset. Refer to Section 8.1.1
for details.
2. Inject the check standards shipped with the instrument. (500 - 1000 ppb S as
thiophene in toluene)
3. When the instrument is optimized, the chromatogram will be comparable to the
one found in the 7090NS Test Conditions package.
The system flows may need adjusting when first received from the factory or when installing new
ceramic pyrotubes. The chromatograms in Figure 8-3 are general guidelines on what to expect
when adjusting the hydrogen and oxygen flows for sulfur analysis.
Figure 8-3 Chromatograms
Figure 8-3a Figure 8-3b Figure 8-3c Figure 8-3d
Flows set correctly H2 Flows – Too Low H2 Flows – Too High 02 Flows – Too Low
Use these chromatograms and the flowcharts on the following pages to optimize your system for
sulfur detection. See Section 8.3 to optimize your system for nitrogen.
To optimize for sulfur detection, use the following flow charts to determine the course of action.
PAC 7090 Series IOS Manual
Page 60 MAN 7090 Rev E
Figure 8.3a Figure 8.3b Figure 8.3c Figure 8.3d
Record all mass flow controller Settings in Table 6.1
See Page 58-59
See Page 57
See
Page 59
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 61
Figure 8.3a Figure 8.3b Figure 8.3c Figure 8.3d
Return to “Instrument Ready”
Page 60
See Page 56
See Page 58-59
See Page 59
Figure 8.3b
PAC 7090 Series IOS Manual
Page 62 MAN 7090 Rev E
Return to “Instrument Ready”
Page 60
Figure 8.3a Figure 8.3b Figure 8.3c Figure 8.3d
Return to “Start Here” Page 6059
Return to “Figure 8-3b”
Return to “Figure 8-3d”
See Page 57
See Page 56
See Page 59
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 63
Figure 8.3a Figure 8.3b Figure 8.3c Figure 8.3d
Return to “Start Here” Page 6059
Return to “Figure 8-3b”
Return to “Figure 8-3b”
Return to “Instrument Ready”
Page 60
See Page 56
See Page 57
See Page 58
Section 4.3
Section 9.5
PAC 7090 Series IOS Manual
Page 64 MAN 7090 Rev E
8.3 Optimizing nitrogen detection
Use the chromatograms in the Figures on the next page and the flowchart following to optimize
your system for nitrogen detection.
Figure 8-4 Optimizing flows
Figure 8-4a Figrue 8-4b
Flows set correctly Oxygen flow set too low
Record all mass flow controller Settings in Table 6.1
Figure 8.4a Figure 8.4b
See
Page 62
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 65
Return to “Instrument Ready”
Page 64
Figure 8.4b Figure 8.4a
Return to “Instrument Ready”
Page 64
See
Page 56
PAC 7090 Series IOS Manual
Page 66 MAN 7090 Rev E
9. SERVICE INFORMATION
9.1 Furnace temperature controller programming procedure
9.1.1 Initial power up
Action Controller Display
Turn instrument on
Initial message upon power up will flash alternatively between “inPt”
and “none”. The controller is requesting the type of thermocouple that will
be used.
To Set Thermocouple Type
To set the thermocouple type. Press and hold the * and then press the
button 4 times for the type K thermocouple. To accept the value, release
the *.
To advance to the next question press the button once.
To Select Celsius or Fahrenheit
The controller message will now flash alternatively between “unit” and
“none”. The controller is requesting the type of temperature display
preferred.
To set the temperature display: Press and hold the * and then press the
button once until the display shows oC. To accept the value, release the *.
To advance to the next question press the button 1 time.
Select Furnace Relay The controller message will now flash alternatively between “SP 1 .d” and
“none”. The controller is requesting the type of relay that is required.
To set the relay type. Press and hold the * and then press the - button 2 times
until the display for the relay type is SSd. To accept the value, release the *.
NOTE:
When the controller is initially turned on, it automatically enters programming
diagnostics at level 5 and requires the three questions listed above to be
answered before continuing with the other programming steps. To return to
different levels at any time, press the button repeatedly until the controller
message flashes alternatively between “level” and “5” or the existing level.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 67
Action Controller Display
The next questions are on level 2. To go to level 2, press the button 3 times
until the controller display flashes alternatively between “level” and “5”.
The controller will then be at the beginning of the level 5 programming.
To change levels, press and hold the * and then press the button 3 times
until the controller message flashes alternatively between “level” and “2”.
To accept the value, release the *.
Select Alarm Parameters
Note:
For Sulfur Detectors only, although it is always programmed
on both nitrogen and sulfur. This gives an “AL” display to
indicate a temperature lower than the setpoint.
To set the alarm type for the Hydrogen Solenoid, press the button 5 times
until the controller message flashes alternatively between “SP2.A” and
“none”.
To set the alarm output type, press and hold the * and then press the
button 5 times until the controller message display is “FS.Lo”. To accept the
value, release the *.
Set Furnace Power Limit %
To set the upper limit of the furnace, press the button 2 times and the
controller message display will flash alternatively between “PL. 1” and
“100”. Hold the * and press the button until the controller display
flashes alternatively between “PL. 1” and “85”.
The PL. 1 value should be set at a value low enough to heat slowly, yet reach
the desired temperature. A value too low will not reach temperature and a
value too high will heat very quickly which in some cases reduces the
overall life of the furnace. The 7090 will typically use a value of 85.
Set Furnace Upper Limit
To set the upper limit of the furnace, press the button 5 times and the
controller message display will flash alternatively between “hi.SC” and
“1200”.
To change the upper limit set point, press and hold the * and use the á button
or the button to adjust the setpoint value either up or down and adjust the
setpoint value to 950. To accept the value, release the or the and then
release the *.
PAC 7090 Series IOS Manual
Page 68 MAN 7090 Rev E
Action Controller Display
The next questions are on level 1.
To change to level 1, press and hold the * and then press the button 8 times until the
controller message flashes alternatively between “level” and “1 ”. To accept the
value, release the *.
Set the Cycle Time of the Furnace
Press the button 6 times until the controller message flashes alternatively between
“CyC.t” and “20”.
Press and hold the * and then press the button or the button to adjust for an
initial cycle time of 4.0. To accept the value, release the *.
Adjust the Hydrogen Solenoid Setpoint Temperature
Press the button 3 times until the controller message flashes alternatively between
“SEt.2” and “0”.
Press and hold the * and then press the button or the button to adjust for a
setting of 925 degrees. To accept the value, release the *.
To accept the parameters go to normal operation, simultaneously press the button
and the * button and hold them pressed until the display reads “PArk”.
9.1.2 Temperature Set Procedure
To set the furnace temperature, press and hold the * and use the button or the *
button to increase or decrease the desired temperature value. Release the * to accept the
setpoint value.
9.1.3 Auto Tune Procedure
To eliminate overshoots or undershoots at the setpoint value, it may be necessary to
autotune the controller. Ensure that the controller is at the approximate operating
temperature.
Press and hold the ⟨ button and the button for approximately 5 seconds or until the
display flashes between “tunE” and “oFF”.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 69
Action Controller Display
Press and hold the * and press the 'f' button 3 times until the display reads “At.SP”.
To accept the value, release the *.
Press and hold the 'f' button and the 'f' button for approximately 5 seconds or
until the display flashes between “tunE” and “At.SP” and “1000” or the
actual temperature of the furnace. This will take about 5 minutes to calculate
the “Cycle Time Value” and there will be a light in the upper left hand corner
of the display that will blink during the Autotune procedure.
At the end of the Autotune procedure the user must enter the calculated
value of the cycle time.
Press and hold the 'f' button and the â button for approximately 5 seconds or
until the display flashes between “tunE” and “oFF”.
Press the 'f' button 5 times until the display reads “CyC.t”.
Press and hold the * and press and hold the 1' button until the display does not
change any more and the display reads “A —” or the calculated value. If the
display reads “A 3.8”, then 3.8 is the optimum value. To accept the value, release
the 1' button and then release the *.
Press and hold the á button and the 1' button for approximately 5 seconds or
until the display flashes the actual furnace temperature.
The furnace should now stop at the desired temperature.
To Reset all Default Conditions
Simultaneously press the 'f' button and the 1' button and hold them pressed
until the display reads “tunE”.
Press the 1' button 1 time and the display will flash alternatively “LEVL” and
“1”.
The reset function is on level 3.
To change levels, press and hold the * and then press the 'f' button 2 times until
the controller message flashes alternatively between “level” and “3”. To
accept the value, release the *.
PAC 7090 Series IOS Manual
Page 70 MAN 7090 Rev E
Action Controller Display
Press the '1 button 12 times until the display flashes alternatively between
“reset” and “none”.
Press and hold the * and then press the '1 button 1 time until the controller
message display is “ALL” . To accept the value, release the *.
Simultaneously press the '1 button and the È button and hold them pressed until
the display flashes alternatively between “inPt” and “none”. The controller is now
ready to be reprogrammed.
9.2 Thermal Electric Cooler (TEC) Temperature Controller Programming Procedure
9.2.1 Initial Power up
Action Controller Display
Turn instrument on.
Initial message upon power up will flash alternatively between “inPt” and “none”. The
controller is requesting the type of thermocouple that will be used.
Set Thermocouple Type
To set the thermocouple type. Press and hold the * and then press the '1 button 10
times until the display is for the type RTD thermocouple. To accept the value,
release the *.
To advance to the next question press the '1 button 1 time.
To Select Celsius or Fahrenheit
The controller message will now flash alternatively between “unit” and “none”. The
controller is requesting the type of temperature display preferred.
To set the temperature display: Press and hold the * and then press the '1 button 1 time
until the display is oC. To accept the value, release the *.
To advance to the next question press the '1 button 1 time.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 71
Action Controller Display
Select Thermal Electric Cooler Relay The controller message will now flash alternatively between “SP 1 .d” and “none”.
The controller is requesting the type of relay that is required.
To set the relay type. Press and hold the * and then press the '1 button 2 times
until the display for the relay type is SSd. To accept the value, release the *.
NOTE:
When the controller is initially turned on, it automatically enters the
programming diagnostics at level 5 and requires the three questions listed
above to be answered before continuing with the other programming
steps to return to different levels at any time press the 1' button
repeatedly until the controller message flashes alternatively between “level”
and “5” or the existing level.
The next questions are on level 3. To go to level 3, press the button 3 times until the
controller display flashes alternatively between “level” and “5”. The controller will
then be at the beginning of the level 5 programming. To change levels, press and hold
the * and then press the 1' button 2 times until the controller message flashes
alternatively between “level” and “3”. To accept the value, release the *.
Select Reverse Mode
To set the relay to react opposite when firing the relay for the thermal electric
cooler, press the '1button 4 times until the display flashes alternatively between
“rEV.d” and “none”.
Press the * and the '1until the display is “1D.2R” Release the * and press the 1' 4 times
until the display flashes alternatively between “level” and “3”.
The next questions are on level 2. The controller is now at the beginning of the
level 3 programming. To change levels, press and hold the * and then press the â
button 1 time until the controller message flashes alternatively between “level” and
“2”. To accept the value, release the *.
Select Alarm Parameters
To set the alarm type for the Hydrogen Solenoid, press the '1 button 5 times until the
controller message flashes alternatively between “SP2.A” and “none”. To set the
alarm output type, press and hold the * and then press the '1 button 5 times until the
controller message display is “FS.Lo”. To accept the value, release the *.
PAC 7090 Series IOS Manual
Page 72 MAN 7090 Rev E
Action Controller Display
Set Thermal Electric Cooler Upper Limit
To set the upper limit of the thermal electric cooler, press the '1 button 3
times and the controller message display will flash alternatively between “hi.SC”
and “1200”.
To change the upper limit set point, press and hold the * and use the '1 button or the
1' button to adjust the setpoint value either up or down and adjust the setpoint value to
35. To accept the value, release the '1 or the 1' and then release the *.
Set Thermal Electric Cooler Lower Limit
To set the lower limit of the thermal electric cooler, press the '1 button 1 time
and the controller message will flash alternatively between “lo.SC” and “32”.
To change the lower limit set point, press and hold the * and then press and hold
the '1 button or the 1' button to scroll the setpoint value either up or down until the
setpoint value is -20. To accept the value, release the 1' button and then release the
*.
The next questions are on level 1.
To go to level 1, press the '1 button repeatedly until the controller display flashes
alternatively between “level” and “2”. The controller will then be at the beginning
of the level 2 programming. To change to level 1, press and hold the * and then press
the 1' button 1 time until the controller message flashes alternatively between “level”
and “1”. To accept the value, release the *.
Set the Cycle Time of the Thermal Electric Cooler
Press the '1 button 6 times until the controller message flashes alternatively between
“CyC.t” and “20”.
Press and hold the * and then press the '1 button or the 1' button to adjust for a initial
cycle time of 4.0.
To accept the parameters go to normal operation, simultaneously press the '1 button
and the 1' button and hold them pressed until the display reads “PArk”.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 73
9.2.2 Temperature Set Procedure (see following text)
To set the thermal electric cooler temperature, press and hold the * and use the '1 button or the 1' button to increase or decrease the desired temperature value. Release the * to accept the setpoint value.
Start with a setting of 4.
Action Controller Display
9.2.3 Auto Tune Procedure
To eliminate overshoots or undershoots at the setpoint value, the user may
need to autotune the controller. Ensure that the controller is at the
approximate operating temperature.
Press and hold the 'f' button and the 1' button for approximately 5 seconds or
until the display flashes between “tunE” and “oFF”.
Press and hold the * and press the 'f' button 3 times until the display reads “At.SP”.
Press and hold the 'f'button and the 1' button for approximately 5 seconds or until
the display flashes between “tunE” and “At.SP” and “5” or the actual
temperature of the thermal electric cooler. This will take about 5 minutes to
calculate the “Cycle Time Value” and there will be a light in the upper left hand
corner of the display that will blink during the Autotune procedure.
At the end of the Autotune procedure the user needs to enter the calculated
value of the cycle time.
Press and hold the 'f' button and the 1' button for approximately 5 seconds or
until the display flashes between “tunE” and “oFF”.
Press the 'f' button 5 times until the display reads “CyC.t”.
Press and hold the * and press and hold the âbutton until the display does not
change any more and the display reads “A —” or the calculated value. If the
display reads “A 3.8”, then 3.8 is the optimum value. To accept the value,
release the 1' button and then release the *.
Press and hold the 'f' button and the 1'button for approximately 5 seconds or until
the display flashes the actual thermal electric cooler temperature.
The thermal electric cooler should now rest at the desired temperature.
PAC 7090 Series IOS Manual
Page 74 MAN 7090 Rev E
Action Controller Display
To reset all default conditions
Simultaneously press the á button and the 1' button until the display reads
“tunE”.
Press the 1' button 1 time and the display will flash alternatively “LEVL” and
“1”.
The reset function is on level 3.
To change levels, press and hold the * and then press the Ç button 2 times until the
controller message flashes alternatively between “level” and “3”. To accept the
value, release the *.
Press the Ç button 12 times until the display flashes alternatively between
“reset” and “none”.
Press and hold the * and then press the Ç button 1 time until the controller
message display is “ALL” . To accept the value, release the *.
Simultaneously press the Ç button and the 1' button until the display flashes
alternatively between “inPt” and “none”. The controller is now ready to be
reprogrammed.
9.3 Board Alignment Procedures
See following page.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 75
9.3.1 Printed Circuit Boards Overview
The following is a list of printed circuit boards (PCB) in the Model 7090 with a brief
explanation of their functions.
1. 103603 —Ozone Generator: This board provides the firing for the ozone
generator transformer firing. (Refer to Section 9.3.2 for adjustment
procedures.)
2. 58137—Signal Pre-amp: This board is located within the photomultiplier
housing assembly. The photomultiplier tube (PMT) connects to this board and
the voltage to the photomultiplier tube (PMT) is distributed here. (No
adjustment required.)
3. 58141—Signal Processor: This board receives the signal from the Signal Pre-
amp board. The gain, sensitivity settings, and autozero are selected on this
board. The signal level is also driven from this board. (Refer to Section 9.3.3
for adjustment procedures.)
4. 58160—High Voltage Power Supply: This board provides the voltage to the
Signal Pre-amp PCB and ultimately to the PMT. The 0–5 vdc voltage that
controls the actual setpoint is received from the 7090 Display Select PCB
(58183). (Refer to Section 9.3.4 for adjustment procedures.)
5. 58167—Detector Interconnect: This board is the main distribution of the power
within the detector tray(s). The 58141 Signal Processor PCB (58141) and the
High Voltage Power Supply PCB (58160) both plug directly to this board. The
gain, autozero, and the sensitivity selection switches are located on this board
and are distributed accordingly. (No adjustment required.)
6. 58183—7090 Display Select: This board controls the selection of all available
displays other than those controlled at the detector tray(s) and provides status
lights for instrument parameters such as flows and vacuum levels. Control
parameters for all Mass Flow Controllers and high voltage to the High Voltage
Power Supply board (58160). (Refer to Section 9.3.5 for adjustment
procedures.)
7. 58185—7090 Signal Output: The output board is located on the rear panel of
the instrument and is the location for connecting the signal level outputs to
chart recorders or integrators. (No adjustment required.)
8. 58186—7090 Power Distribution: The power distribution controls the
distribution of all power to the analyzer. The power to the detector trays and all
power supplies and transformers are located on this board. The instrument
fuses are located on this PCB and can be easily accessed from the rear of the
instrument. (No adjustment required.)
9. 58193—Mass Flow Controller: The mass flow controllers are all connected
with this common board. It provides the + 15 VDC power requirement as well
as the control voltage and mass flow read to the 7090 Display Select PCB (58
183). (No adjustment required.)
10. 58210 —H2 Control LED Status: This board contains the LEDs that indicate
the switch position of the 7090 Display Select PCB (58183). (No adjustment
required.)
PAC 7090 Series IOS Manual
Page 76 MAN 7090 Rev E
CAUTION:
High voltages are present during operation of the ozone generator. Ensure that
safe work conditions are established and used.
NOTE:
The Ozone Generator board cannot be properly adjusted unless installed with a
working ozone cell.
NOTE:
The ozone circuit will not work if the pressure is not 30 psi. Also, the pressure sensor
will not disengage until pressure drops below 28.5 psi.
9.3.2 Ozone Generator Board – p/n 103603
1. Setup Ozone Generator / Ozone PCB test rig, as shown in picture below:
RotometerPN 23098
Pressure regulatorPN 21067
30 psi 25 cc/min
Ozone Generator
Ozone MeterTELEDYNE 454H
Ozone scrubberPN 71137-2
P-3Oxygen
Exhaust
2. Install jumpers to JP4, JP5, and JP6 on Ozone PCB.
3. Connect 5 VDC to the Ozone Generator PCB at location H2 with the Positive 2,
Negative Pin 1.
4. Leave H1 connector on Ozone PCB unconnected.
5. Set up a DMM to measure frequency.
6. Connect the DMM Black lead to TP3 on ozone PCB.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 77
7. Connect the DMM Red lead to TP6 on ozone PCB.
8. Start flow of Oxygen into Ozone Generator.
9. Apply power to the Ozone Generator.
10. Adjust potentiometer R1 so that DMM shows 120 ± 1Hz.
11. Allow the Ozone generator board to stabilize for 10 minutes.
12. The ozone meter should read between 1.5 and 3 % of ozone.
13. Turn off power and shut down oxygen flow to Ozone Generator.
14. Wait for 1 min for Ozone Generator to power down.
15. Configure the jumpers JP4 and JP6 on the Ozone PCB for the target instrument.
Leave H2 unconnected.
9.3.3 Signal Processing (58141R3 Alignment Procedure)
Figure 9-2 Signal Processing Board
58141 Connector Identification
J1 Remote Zero J10 Signal Out
J2 Auto Zero J11 Span
J3 Manual Zero J12 Scaled Output
J4 Sub-panel Board J13 Zero
J5 Display J14 High/Low
J6 Rear Output Board
(Signal & Auto Zero Combined) J15 PMT Power
J7 PMT Signal In J16 PM 390
J8 4-20 ma Output J17 External 4-20 ma
J9 Sensitivity
1. Place shorting jumper on JP2. (Display Select)
2. Place a shorting jumper on the left two pins of JP6. (Unscaled display)
PAC 7090 Series IOS Manual
Page 78 MAN 7090 Rev E
3. Place a shorting jumper on the right two pins of J11. (Scale option ON)
4. Place a shorting jumper on the left two pins of J7. (For field adjustment, this
jump must be made or a working PMT pre-amp must be installed )
5. Use a jumper to momentarily short the left side of R74 to the bottom pin of
JP3, JP4, or JP5 (+5 vdc). (Clears Zero register)
6. With a Digital Volt Meter, insert the black meter lead into TP1 (Ground). Insert
the red lead into TP2. Adjust R54 for 0.0 mvdc.
7. Use a jumper to short the left side of R7 to the left side of R21 (analog ground).
8. Remove the 16 pin ribbon cable from J4.
9. Switch S1#1 to the on position (S1#1 up = x50).
10. Move the Red lead to the right side of R3 1, adjust R1 for 0.0 mvdc.
11. Move the Red lead to the left side of R73, adjust R50 for 0.0 mvdc (± 5mvdc).
(Auto Zero Setpoint)
12. Move the jumper on JP6 from the left two pins to the right two pins. (Scaled
Display)
13. Place the Red lead to the bottom of R39, adjust R34 for -0.9 mvdc . (This is the
offset adjustment that allows a full scale output on the model 9000. The
adjustment should allow the raw data after the A/D module to be around 0.100
VDC. If more/less adjustment is required/desired the -0.9 mvdc is subject to
change to get a 0.100 VDC reading.)
14. Move the jumper on JP6 from right two pins to the left two pins. (Unscaled
display)
15. Completely remove the jumper installed in step 3 (left side of R7 to the left
side of R21–analog ground).
16. Move the red lead to the right side of R3 1.
17. Momentarily short the two pins of J1 to autozero the board.
18. If the board is properly adjusted and working, the voltage should stabilize
within 3.0 mvdc of 0.00.
19. Vary the millivolt output by applying a millivolt input or varying the voltage to
the PMT and auto zeroing several times. (This can also be accomplished by
shorting the bottom of JP5 to the left side of R74.)
20. To adjust the front panel display, measure R3 1 and adjust R30 so that the
display value is 100 times the result. e.g: 1.70 volts measured, display should
be 170.
21. To test the sensitivity settings measure R31 input with S1 #1 up and apply the
value to X (100%). e.g: 1.70 volts measured, display should be 170—Panel
display 170.
22. Move the jumper from JP2 to JP1
23. Return all switches to the down position and replace the 16 pin ribbon cable.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 79
Switch Position R31 Value Display Reading
#1 up only 1.70 VDC 170
#2 up only 0.85 VDC 085
#3 up only 0.34 VDC 034
#4 up only 0.17 VDC 017
9.3.4 High Voltage Power supply voltage (58160) Adjustment
Figure 9-3 High Voltage power supply board
9.3.4.1 Jumper Positions
1. Set JP1 to the Right (for VAC or VDC greater than 15V).
2. Set JP2 to the Right (for 15V or greater AC or DC).
3. Set JP3 to the Right (for 0-5VDC control voltage).
4. Set JP4 to the Left (for external control).
5. Short D3.
9.3.4.2 Board adjustment
1. Apply power to the board and adjust for an output of –600VDC. As the board is
set for external control, supply the control voltage of 3.0VDC that represents
600VDC and adjust P3 for 600VDC at TB1 positions 1 and 2.
2. Adjust P1 for the proper monitor voltage at TP1 and TP2 (ground). For 0-1VDC
monitor output adjust P1 for 0.6VDC.
PAC 7090 Series IOS Manual
Page 80 MAN 7090 Rev E
NOTE:
For a 0—5 VDC output, TP1 should be 1000=5v, 800=4v, 600=3v…etc. For a 0-1
VDC output, TP1 should be 1000=1v, 800=0.8v, 600=0.6v…etc.
9.3.5 Display Select Adjustment Procedure (58183)
9.3.5.1 Required Test Equipment
1. Digital Volt Ohm Meter
2. Vacuum Pump
3. Calibrated Vacuum Meter
4. 2 Pin Jumpers
5. 58183/58185 Test Fixture or completed instrument
9.3.5.2 Preliminary notes
The front panel lights will not be on unless there is 5 volts present from the pressure sensor.
9.3.5.3 Power Supply Test Points
1. TP3 Common
2. TP2 +15 VDC
3. TP4-15 VDC
4. TP9 +5 VDC
9.3.5.4 Meter adjustment
1. Turn selector knob to the 3rd position (Ozone).
2. Insert black lead into TP3.
3. Measure 5 vdc on pin 2 of JP1.
4. Install a jumper on JP1.
5. Insert the red lead into TP13.
6. Adjust P7 until the front panel display reads 500 (Ignore all decimals if
present).
7 Remove jumper.
9.3.5.5 Oxygen Pressure Sensor
1. Ensure that the Oxygen Pressure sensor light comes on when:
a. O2 is greater than 30 psi.
b. Test fixture O2 Sensor Simulation switch is energized.
c. When JP2 Oxygen Bypass is jumped.
9.3.5.6 Furnace Vacuum Adjustment
1. Turn the selector switch to the Furnace vacuum read. (Position 1).
2. Adjust P2 until display reads 760.
3. Using an approved vacuum sensor, determine the torr reading of a vacuum
pump. For testing purposes, assume 30 torr is the low end vacuum.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 81
4. Apply known vacuum to the P1 port on the vacuum sensor.
5. Adjust P1 for the torr value determined in step 4.
6. Remove the vacuum and adjust P2 until the display reads 760.
7. Repeat steps 5–7 until no further adjustments are required.
9.3.5.7 Detector Vacuum Adjustment
1. Turn the selector switch to the Furnace vacuum read. (Position 1).
2. Adjust P5 until display reads 760.
3. Using an approved vacuum sensor, determine the torr reading of a vacuum
pump. For testing purposes, assume 30 torr is the low end vacuum.
4. Apply known vacuum to the P1 port on the vacuum sensor.
5. Adjust P4 for the torr value determined in step 4.
6. Remove the vacuum and adjust P5 until the display reads 760. 14. Repeat steps
5-7 until no further adjustments are required. 7.3.5.8
9.3.5.8 Temperature Span Adjustment
1. Apply mv source to thermocouple input on 58186 printed circuit board or heat
the furnace to 950°C.
2. Ensure ribbon cable is properly connected.
3. Insert the black lead into the common (TP3).
4. Insert the red lead into the High setpoint Test Point (TP6).
5. Adjust P15 until the DVM displays 42.0 mvdc.
6. Insert the red lead into the Low Setpoint test point (TP5).
7. Adjust P14 until the DVM displays 36.5 mvdc.
8. Monitor both the Temperature LED and the Cal 3200 display as the mv source
is increased and decreased. Ensure that the LED is green only between the
reported temperature of 900° and 1050°.
9. Monitor the Hydrogen LED and ensure that the LED turns green at the reported
temperature of 925°.
9.3.5.9 Mass Flow Controller Adjustment
1. Measure the voltage at TP17 with the red lead to ensure a stable 10 volt
reference.
2. If there are no mass flow controllers present, use jumpers to temporarily
simulate MFCs by placing jumpers on RC2 pins (9,10 Ozone), (11,12 O2),
(13,14 Hydrogen).
3. On the front panel, turn the display selection switch to the “Ozone” (Position
3).
4. Measure the voltage at TP12 with the red lead.
5. Adjust P9 until the DVM is 2.5 vdc.
6. Ensure that the front panel display meter reads 25.0 (cc/min). Flow must be
present to read correctly.
7. Measure the actual flow to ensure the proper flow from the Mass Flow
Controller.
PAC 7090 Series IOS Manual
Page 82 MAN 7090 Rev E
8. On the front panel, turn the display selection switch to the “Oxygen” (Position
4).
9. Measure the voltage at TP11 with the red lead.
10. Adjust P10 until the DVM is 1.0 vdc.
12. Ensure that the front panel display meter reads 10.0 (cc/min). Flow must be
present to read correctly.
13. Measure the actual flow to ensure the proper flow from the Mass Flow
Controller.
14. On the front panel, turn the display selection switch to the “Hydrogen”
(Position 5).
15. Measure the voltage at TP18 with the red lead.
16. Adjust P8 until the DVM is 2.5 vdc. (This is the maximum voltage that can be
applied to the Hydrogen controller.)
17. Measure the voltage at TP10 with the red lead.
18. Adjust P11 until the DVM is 1.0 vdc.
19. Ensure that the front panel display meter reads 100 (cc/min). Flow must be
present to read correctly.
20. Measure the actual flow to ensure the proper flow from the Mass Flow
Controller
9.3.5.10 Furnace Vacuum Setpoint Adjustment
1. Turn the selector switch to the Furnace Vacuum Setpoint read. (Position 6).
2. Adjust P3 fully until display reads 250. 7.3.5.1
9.3.5.11 High Voltage Adjustment Channel A
1. Measure the voltage at TP16.
2. Adjust P12 until the voltage reads 5.0 VDC.
3. Measure across J2.
9.3.5.12 High Voltage Adjustment Channel B
1. Measure the voltage at TP 15.
2. Adjust P13 until the voltage reads 5.0 VDC.
3. Measure across J1.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 83
9.4 Fuse Requirements
Fuse Description 115 VAC 230 VAC
b & c Main Fuse (8 amp) (5 amp)
d Fuse 70 VAC (transformer secondary)
Furnace voltage
(10 amp) (10 amp)
e Fuse Furnace Transformer (8 amp) (5 amp)
f Fuse 24 VAC (2 amp) (2 amp)
g Fuse 24 V Transformer (2 amp) (1 amp)
h Fuse temperature controller (1 amp) (1 amp)
i Fuse 5 VDC Power Supply Ozone Generator (1 amp) (1 amp)
j Fuse 15 V Power Supply Mass Flow controllers (1 amp) (1 amp)
k (1 amp) (1 amp)
l Detector trays (1 amp)
(Slo-blo)
(1 amp)
(Slo-blo)
NOTE:
The main fuses are located in the fuse drawer located in the housing of the power
receptacle. Use a small screwdriver to release the latch and the drawer will easily
slide out.
9.5 Leak Check Procedures
9.5.1 Mass Flow Controller Scaling Procedure and Vacuum Test
1. Connect the “Sample IN” line to the exit of the furnace.
2. Connect the “Hydrogen Out” to the Furnace Base “Hydrogen Inlet”.
3. Connect the “Oxygen Out” to the side arm of the Furnace Inlet Adapter
Assembly.
4. Install a column plug on the bottom arm of the Furnace Inlet Adapter
Assembly.
5. Turn the O3 (ozone), O2 (oxygen to the furnace base), H2 (Hydrogen to the
Furnace base) trimpots 100 % counterclockwise. (OFF)
6. Attach a 50 psi gas oxygen line with a pressure gauge and a toggle valve to the
bulkhead labeled “Oxygen In” on the rear panel.
7. Open the oxygen valve to apply pressure to the analyzer.
8. Turn off the O2 valve, and monitor the pressure to ensure that the O2 pressure
does not drop which would indicate a leak in the supply line or a leaking Mass
Flow Controller (MFC).
9. If there are no leaks, turn the “Display Select” knob to the O3 (Ozone) position.
PAC 7090 Series IOS Manual
Page 84 MAN 7090 Rev E
10. Attach a digital flow meter to the exit of the scrubber tube vent.
11. The initial reading of the digital flow meter should be 0.
12. Turn the O3 (ozone) adjustment trimpot until the front panel display reads 25.0
the display.
13. The digital flow meter should read 25.0 ±3 %.
14. If the flow is good, turn the “Display Select” knob to the O2 (oxygen to the
furnace base) position.
15. Turn the O2 (oxygen to the furnace base) adjustment trimpot until the front
panel display reads 10.0.
16. The digital flow meter should read 35.0 ±3 %.
17. Turn the O3 (ozone), O2 (oxygen to the furnace base) trimpots 100 %
counterclockwise. (OFF)
18. Disconnect the digital flowmeter from the scrubber vent.
19. Disconnect the furnace heater cable from the rear panel of the Model 7090 to
prevent heating.
20. Attach a known good vacuum to the scrubber vent.
21. Turn the “Display Select” knob to the P1 position.
22. Turn the vacuum pump on and ensure that the display on the front panel
indicates a vacuum less than 035 torr.
23. Attach a 50 psi gas hydrogen line to the bulkhead labeled “Hydrogen In” on the
rear panel.
24. Open the cylinder valve to pressurize the line.
25. Immediately close the cylinder valve and monitor the pressure gauges on the
cylinder. If the pressure drops, this is an indication of a leak in the supply line.
26. If the pressure does not drop, open the cylinder valve and let it remain open.
27. Disconnect the hydrogen out line from the rear of the analyzer and plug the line
with a brass plug.
28. Attach a digital flowmeter to the “Hydrogen Out” bulkhead.
29. On the 58183 printed circuit board, turn the P8 trimpot 100 % clockwise.
(OFF)
30. Trick the Hydrogen Solenoid to “Open” by:
a. Changing the Cal3200 SP1 alarm setting to 0°C or,
b. Using a millivolt source as a thermocouple to indicate 950°C
31. Turn the H2 (hydrogen to the furnace base) trimpots 100 % clockwise. (ON).
32. Increase P8 until the digital flow meter is at 625 cc/min (maximum H2 flow).
33. Turn the H2 (hydrogen to the furnace base) trimpots 100% counterclockwise.
(OFF).
34. Return the Hydrogen Solenoid to normal operation.
35. Disconnect the digital flow meter from the “Hydrogen Out” bulkhead.
36. Reattach the “Hydrogen Out” line to the “Hydrogen Out” bulkhead on the rear
of the analyzer.
37. Turn off the vacuum pump and allow the vacuum to bleed.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 85
38. Disconnect the column plug from the bottom arm of the Furnace Inlet Adapter
Assembly.
39. Connect the capillary column to the base of the Furnace Inlet Adapter
Assembly.
40. Turn the vacuum pump on.
41. Turn the ozone flow to 25 cc/min.
42. Increase the O2 flow to 10.0 cc/min.
43. The P1 value should be approximately 110 torr.
If a leak is suspected and cannot be found, a pressure test might help localize the leak.
9.5.2 Pressure Test (With Furnace)
Do not pressurize above 10 psi.
1. Adjust the oxygen supply pressure to 10 psi.
2. Plug the bulkhead on the rear panel labeled “Hydrogen Out” and “Hydrogen
In”.
3. Plug the exit of the ozone scrubber tube.
4. Connect the 1/8" “Sample Inlet” to the furnace vent.
5. Connect the furnace base O2 Inlet to the bulkhead on the rear panel labeled
“Oxygen Out”.
6. Turn on the Model 7090.
7. Turn the hydrogen, oxygen and ozone flow adjustments located on the front
panel 100 % fully clockwise (On).
8. Open the toggle valve for the oxygen and pressurize the system.
9. Turn off the toggle valve and ensure that the pressure holds for 3 minutes.
10. If leak cannot be found, bypass the furnace by connect the 1/8" “Sample Inlet”
to the bulkhead on the rear panel labeled “Oxygen Out”.
a. If pressure holds, there is a leak in the furnace.
b. If pressure drops, the leak is in the instrument plumbing and
interconnections.
11. Once there are no leaks, remove plugs.
12. Turn the hydrogen, oxygen and ozone flow adjustments located on the front
panel 100% fully counterclockwise (OFF).
13. Return the oxygen cylinder pressure to 50 psi.
9.6 Start-up & Shutdown Procedures
For optimum performance, the instrument should remain on at all times to maintain stability.
Shutdown the instrument only during maintenance or long periods of inactivity. Recalibration of
instrument is typically required after any shutdown or when normal gas flows are not maintained.
9.6.1 Start-up Procedure
1. Oxygen and helium should be off at the cylinders.
2. Turn on main power switch. O2 status light should be red.
3. Open oxygen cylinder. O2 status light should turn green, all other status lights
should be red. If O2 light is not green, check cylinder pressure.
PAC 7090 Series IOS Manual
Page 86 MAN 7090 Rev E
4. Turn on vacuum pump. Furnace vacuum should be < 120 torr, and status light
should turn green. Detector vacuum should be < 20 torr. Furnace pressure may
increase with temperature.
5. While furnace heats, check oxygen and ozone flows. They should be at preset
conditions.
6. Hydrogen solenoid should turn on when temperature reaches ~925°C (sulfur
instruments only).
7. Temperature should be set at 950°C. When this is reached temperature status
light should be green. All status lights except for O3 should now be green.
8. Set the detector display for high voltage and turn on the high voltage power
supply. Value should be set at ~750 volts.
9. Set the detector display to sensitivity and press the autozero switch. The
display should approach and remain near 000.
10. Turn on the ozone generator. The display (signal) should increase. At this point
all lights should be green.
11. Allow oxidation period of preferably overnight.
12. Open toggle valve for helium make up flow. P1 pressure will increase 100 to
150 torr. (sulfur instruments only).
13. Slowly bring the hydrogen flow (if sulfur instrument) up until the display reads
the recommended setting. Turn clockwise until display value is reached.
Furnace vacuum should be approximately the predetermined value. Detector
vacuum should be < 30 torr.
9.6.2 Shutdown Procedure
A long-term shutdown is defined as a period of no analytical operation for more than four days. If
the system is to be shut down for a long period of time, the following steps should be taken:
1. Turn off hydrogen at mass flow controller (sulfur instruments only).
2. Turn off helium make up flow using toggle valve.
3. Turn off ozone generator.
4. Turn off high voltage power supply.
5. Allow oxidation period.
6. Program furnace temperature to zero.
7. Allow furnace to cool to ambient temperature.
8. When furnace is cool, turn off vacuum pump.
9. Turn off main power, then turn off oxygen and hydrogen at their cylinders.
9.7 Routine Maintenance
The PAC 7090 System is a relatively low maintenance analyzer. Regular care of a few items should keep
the system operating at peak level. The following is a recommended routine maintenance
schedule:
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 87
9.7.1 Daily
1. Check cylinder supply pressure for all supply gases.
2. Check all gas flows via mass flow controller display.
3. Check all analytical parameters including temperatures, detector settings, etc.
4. Check the validity of any calibration curve being used by analyzing at least one
standard of known concentration and comparing the results to the existing
curve.
9.7.2 As Needed
1. Lenses may need to be cleaned from time to time. Residue may accumulate on
the lenses of the reaction chamber. Generally this is only performed on a
yearly basis, but if response decreases steadily or the system exhibits poor
repeatability, it may be an indication that the lenses need cleaning.
2. Replace any items in the analytical flow path which have degraded with use
(i.e. Pyrotubes, interconnect tubing, O-rings, etc.).
3. Empty and recharge the scrubber with fresh material. Refill kits (P/N 72001)
are available from PAC. (Normally done once a year)
9.7.3 Reaction Chamber Cleaning Procedures
In the event the lens in the detector reaction chamber needs cleaning or replacement, the
following steps should be taken:
1. Turn the main power switch of the 7090 system to the OFF position and
unplug the power cord from the power source.
WARNING:
Ensure that the power cord has been unplugged from the power source before
continuing. If power is applied while servicing internal components, a severe
electric shock hazard may exist.
2. Remove the side panel of the 7090 system.
3. Locate the sulfur detector module on the left of the 7090 system.
4. Carefully remove all interconnecting tubing associated with the detector
reaction chamber assembly, noting their location and orientation.
5. Remove the reaction chamber assembly from the PMT housing by rotating
the reaction chamber clockwise while holding the PMT housing stationary.
PAC 7090 Series IOS Manual
Page 88 MAN 7090 Rev E
CAUTION:
The PMT is extremely light sensitive. All operations involving the exposure of the
PMT should be conducted in darkness or a very low light situation.
6. Carefully extract the reaction chamber assembly from the housing noting its
orientation.
7. Carefully cover the open end of the PMT housing with a soft lint-free cloth.
NOTE:
At this point, the reaction chamber assembly may be repaired or replaced with a new
assembly.
8. Unscrew the lens ring and carefully remove the lens ring and cutoff lens,
noting their location.
9. Carefully clean the lenses and the reaction chamber with a soft, lint-free cloth
and a solvent such as acetone or methanol.
10. Dry the lenses and the reaction chamber with a soft, lint-free cloth and
ensure that no particles or lint are on the lenses.
11. Install a new silicone O-ring and install the cutoff lens and the lens ring. Secure
the lens ring.
12. Install the reaction chamber assembly into the PMT housing.
13. Replace all interconnecting tubing.
CAUTION:
Ensure that all interconnections are properly installed. Improper connections may
cause damage to system components.
14. Replace side panels.
15. Perform the start-up sequence outlined in Chapter 5 and confirm proper
operation of the repaired reaction chamber assembly.
16. Resume normal operation.
9.8 Practical Troubleshooting Checklist
9.8.1 Gases
1. Gases on
2. Oxygen pressure at 40 psi
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 89
3. Hydrogen (if present) is at 40–65 psi
4. Leak check performed
9.8.2 Instrument Conditions
1. Main Power Switch is on
2. All lights are on
3. Fuses are good
4. Proper gain selection (Hi/Low)
5. Proper sensitivity selection (x50, x25, x10, x5, x1)
6. Auto zero light off
7. Displayed signal level ~ zero
8. Autozero light comes on when selected
9. High Voltage switch on and display shows voltage
10. O3 switch on
11. Display shows voltage change
12. Furnace should heat or be at temperature
13. All lights green
9.8.3 Vacuum Pump
1. Vacuum pump is on
2. Fuses are good
3. Display vacuum changes from 760 when the vacuum pump is turned on
9.8.4 Furnace
1. Furnace is at temperature
2. Fuses are good
3. Temperature controller is programmed
4. Cables are connected
5. O2 pressure must be 40 psi to heat
6. Vacuum must be below setpoint
9.8.5 Gas Chromatograph Conditions
1. Injector temperature
2. Injection valve
3. Septum
4. Column type and selection
5. Column temperature
PAC 7090 Series IOS Manual
Page 90 MAN 7090 Rev E
6. Detector temperature
7. Column flows
9.8.6 Proper Integrator Channel selected
1. Ensure appropriate integrator channel is selected.
9.8.7 Detection Problems
1. PMT voltage on
2. High Voltage on
3. Temperature correct
4. No gas leaks
5. Flows correct
6. Reaction chamber lens is clean
7. Pyrotubes are oxidized and in good condition
8. Instrument cooling fans turning and unobstructed
a. Main fan (rear panel)
b. Thermal Electric Cooler (Nitrogen only)
9.8.8 Noise Problems
1. PMT voltage not too high
2. Flows not too high
3. No light leaks
4. No gas leaks
5. Gains and settings not too high
6. Check main power in.
9.8.9 Practical Troubleshooting Hints
Answers to the following questions will help the Service Department better troubleshoot and
determine the problem with your instrument.
1. Does the problem stop when the ozone generator is turned off?
2. Does the problem stop when the high voltage is turned off?
3. Does the baseline move when shadows are on the instrument or when light is
shined on the instrument?
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 91
4. Does the detector respond when a light leak is created?
5. Is furnace vacuum below 760, but not below setpoint? Could mean a broken
inner pyrotube.
6. Is furnace vacuum below setpoint when hydrogen is flowing? Is detector
vacuum below 35 torr?
7. Is furnace vac increasing? Could mean an obstruction in oxygen or supply
lines.
9.9 Analytical Troubleshooting
This section is a guide for isolating and correcting problems which may arise while operating the
PAC 7090 system.
NOTE:
The following guide is primarily concerned with troubleshooting the 7090
system. This guide should be used in conjunction with normal gas
chromatograph and column troubleshooting guides. To use this guide, simply
locate the symptom in the left column and follow suggestions for probable cause
and solution.
Before attempting to perform any troubleshooting and before taking any action to remedy an
apparent problem, ensure that all components of the analytical system and any accessories are
properly assembled.
PAC 7090 Series IOS Manual
Page 92 MAN 7090 Rev E
Symptom Probable Cause Solution
Baseline Problems: Baseline off scale - positive(+) Zero control not adjusted
PMT voltage too high
Contaminated gas supply
Contaminated pyrotube
Contaminated transfer
tubing
Contaminated reaction
chamber
Light leak
Press autozero.
Adjust PMT voltage.
Change gas supply.
Re-oxidize or replace
pyrotube.
Clean or replace tubing.
Clean reaction chamber.
Repair light leak.
Baseline off scale - negative(-) Zero control not adjusted Press autozero.
Baseline drift - positive(+) Contaminated gas supply
Contaminated septum
Contaminated pyrotube
Contaminated transfer
tubing
Contaminated reaction
chamber
Gas leak
Vacuum leak
Column bleed
Change gas supply.
Replace septum.
Re-oxidize pyrotube or replace
pyrotube. Clean tubing or
replace.
Clean reaction chamber.
Repair gas leak.
Repair vacuum leak.
Recondition column, reduce
temperature, replace column.
Baseline drift - negative(-) Contaminated septum
Contaminated pyrotube
Moisture in gas supply or
gas supply lines quenching
ozone
Replace septum.
Re-oxidize pyrotube or replace
and oxidize new pyrotube.
Change to new gas supply,
replace inline gas purifiers.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 93
Baseline shifts during analysis Change in system gas flow
during sample introduction
Sample size too large
Septum leak
Column bleed
Incomplete combustion of
the sample
Sample reacting with the
sample
Faulty fan motor can cause
internal temperature
instability
Check for obstruction or leak
in analytical system.
Reduce sample size.
Replace septum.
Condition column, reduce
temperature, or replace
column.
Ensure that the furnace
temperatures, split flow, and
gas flows are correct.
Convert to a non-reactive
injection system.
Check instrument fan to be on
and free from obstruction.
Baseline shift – anytime Change in gas system flows
Gas leak
Septum leak
Column bleed
Light leak
Vacuum leak
Liquid trapped in transfer
line
Moisture present in reaction
chamber housing
Faulty fan motor can cause
internal temperature
instability
Check for obstruction or leak
in analytical system or quality
of gas supply and lines.
Repair gas leak.
Replace Septum.
Condition column, reduce
temperature or replace column.
Repair light leak.
Repair vacuum leak.
Ensure that all lines are clean
and dry.
Ensure reaction chamber is
clean and dry.
Check instrument fan to be on
and free from obstruction.
PAC 7090 Series IOS Manual
Page 94 MAN 7090 Rev E
Baseline noise Unstable power source
Unstable power supply
Faulty signal cable
Faulty PMT
Faulty pre-amp printed
circuit board
Faulty amplifier printed
circuit board
Noisy ozone generator
Faulty gas regulator
Light leak
Unstable temperatures
Contaminated furnace
interface
Photomultiplier (PMT) not
cooled (Nitrogen only)
Moisture present in reaction
chamber housing
Faulty fan motor can cause
internal temperature
instability
Stabilize power.
Replace power supply.
Replace signal cable.
Replace PMT.
Replace pre-amp, printed
circuit board.
Replace amplifier printed
circuit board.
Replace ozone generator cell
or assembly.
Replace regulator.
Repair light leak.
Ensure that the ambient
temperature and the instrument
temperature zones are stable.
Clean furnace interface tubing.
Check setting or replace
cooler.
Ensure reaction chamber is
clean and dry.
Check instrument fan to be on
and free of obstructions.
Baseline Problems:
See following page.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 95
No Peaks Gas leak
No ozone
No sample introduction
Contaminated or spent
pyrotube
No PMT voltage
Loose cable
Faulty PMT
Faulty pre-amp printed
circuit board
Faulty amplifier printed
circuit board
Capillary column plugged
Broken inner pyrotube
(sulfur only)
Sample reacting with the
sample injection system
Septum leak
Repair gas leak.
Check ozone generator.
Check sample injector or inject
a known standard.
Re-oxidize pyrotube or replace
and oxidize new pyrotube.
Check PMT voltage.
Secure all cables.
Replace PMT.
Replace pre-amp printed
circuit board.
Replace amplifier printed
circuit board.
Break cm (in) off the inlet of
the capillary column.
Replace inner pyrotube and
oxidize pyrotubes.
Convert to a non-reactive
injection system.
Replace septum.
Negative Peaks Signal cable attached
backwards
Improper gas flows
Reverse + and – leads
Adjust gas flows.
Ghost peaks Contaminated gas supply
Contaminated septum
Contaminated pyrotube
Contaminated transfer
tubing
Contaminated column
Contaminated or reactive
injector
Change gas supply.
Replace septum.
Re-oxidize or replace
pyrotube.
Clean or replace transfer
tubing.
Condition or replace column.
Clean or replace injector
assembly.
Tailing Peaks Gas leak
Septum leak
Change in system gas flow
during sample introduction
Incorrect flows
Poor sample introduction
Replace gas leak.
Replace septum.
Check for obstruction in
analytical system.
Ensure that flows are correct
Check injection.
Split Peaks Poor capillary column
Wrong column material for
sample components
Incorrect temperatures
Replace capillary column.
Use different type of column.
Check all temperature zones.
PAC 7090 Series IOS Manual
Page 96 MAN 7090 Rev E
Low Response Gas leak
Septum leak
Change in system gas flow
during sample introduction
Incorrect flows
Poor sample introduction
Incorrect Gain, sensitivity
or PMT Voltage
Sample size too small
Dirty reaction chamber or
lens
Insufficient ozone
production
Replace gas leak.
Replace septum.
Check for obstruction in
analytical system.
Ensure that flows are correct
Check injection.
Check instrument settings.
Increase sample size or
decrease split ratio.
Clean reaction chamber and
lens.
Ensure ozone generator is on
and is producing sufficient
ozone
Changes in response Gas leak
Septum leak
Change in system gas flow
during sample introduction
Contaminated septum
Contaminated gas supply
Contaminated gas pyrotube
Unstable power source
Faulty gas regulator
Poor sample introduction
No sample introduction
Incorrect flows
Change in ozone
production
Change in PMT voltage
Abnormal ambient
temperature change Light
leak
Faulty PMT
Faulty pre-amp printed
circuit board
Faulty amplifier printed
circuit board
Dirty reaction chamber or
lens
Faulty fan motor can cause
internal temperature
instability
Replace gas leak.
Replace septum.
Replace septum.
Replace septum.
Change gas supply.
Oxidize or replace pyrotube.
Stabilize power.
Replace regulator.
Check injection.
Check sample injector.
Ensure that flows are correct.
Check ozone generator
performance. Check PMT
voltage.
Stabilize ambient temperature.
Repair light leak.
Replace PMT.
Replace pre-amp printed
circuit board.
Replace amplifier printed
circuit board
Clean reaction chamber and
lens.
Check instrument fan to be on
and free from obstruction.
Flat top peaks Saturation of electronics Reduce sample size,
concentration, or gain,
sensitivity or PMT voltage; or
increase split introduction.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 97
Nonlinear curve Saturation of electronics
Contaminated gas supply
Contaminated pyrotube
Poor sample introduction
No sample introduction
Incorrect flows
Change in ozone
production
Reduce sample size,
concentration, or gain,
sensitivity or PMT voltage; or
increase split introduction.
Change gas supply.
Oxidize or replace pyrotube
Check injection.
Check sample injector.
Ensure flows are correct.
Check Ozone Generator
performance.
Small linear range Saturation of electronics
Contaminated gas supply
Contaminated pyrotube
Poor sample introduction
No sample introduction
Incorrect flows
Change in ozone
production
Reduce sample size,
concentration, or gain,
sensitivity or PMT voltage; or
increase split introduction.
Change gas supply.
Oxidize or replace pyrotube.
Check injection.
Check sample injector.
Ensure flows are correct.
Check Ozone Generator
performance.
PAC 7090 Series IOS Manual
Page 98 MAN 7090 Rev E
Non-repeatable results Saturation of electronics
Contaminated septum
Contaminated gas supply
Contaminated pyrotube
Poor sample introduction
Incorrect flows
Flow path restrictions
Sample frequency too fast
Change in ozone
production
Change in PMT voltage
Ambient temperature
change
Light leak
Faulty PMT
Faulty pre-amp printed
circuit board
Faulty amplifier printed
circuit board
Gas leak
Improper sample size
Incomplete combustion of
sample
Unstable temperatures
Noise baseline
Faulty fan motor can cause
internal temperature
instability
Reduce sample size,
concentration, or gain,
sensitivity or PMT voltage; or
increase split introduction.
Replace septum.
Change gas supply.
Oxidize or replace pyrotube.
Check injection.
Ensure flows are correct.
Check fittings, ferrules,
pyrotube, and vents for
restrictions.
Reduce sample frequency
Check ozone generator
performance.
Check PMT voltage
Stabilize ambient temperature.
Repair light leak.
Replace PMT.
Replace pre-amp printed
circuit board.
Replace amplifier printed
circuit board.
Repair gas leak.
Adjust sample size.
Ensure that furnace
temperature and oxygen flow
are correct.
Ensure that the ambient
temperature and the instrument
temperature zones are stable.
Stabilize baseline.
Check instrument fan to be on
and free from obstruction.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 99
Inaccurate results Inaccurate calibration curve
Inaccurate calibration
standards
Saturation of electronics
Contaminated septum
Contaminated gas supply
Contaminated pyrotube
Poor sample introduction
Incorrect flows
Change in ozone
production
Change in PMT voltage
Ambient temperature
change
Light leak
Faulty PMT
Faulty pre-amp printed
circuit board
Faulty amplifier printed
circuit board
Gas leak
Incomplete combustion of
sample
Unstable temperatures
Faulty fan motor can cause
internal temperature
instability
Re-calibrate the system.
Construct new calibration
standards.
Reduce sample size,
concentration, or gain,
sensitivity or PMT voltage; or
increase split introduction.
Replace septum.
Change gas supply.
Oxidize or replace pyrotube
Check sample injector.
Ensure flows are correct.
Check ozone generator
performance.
Check PMT voltage.
Stabilize ambient temperature.
Repair light leak.
Replace PMT.
Replace pre-amp printed
circuit board.
Replace amplifier printed
circuit board.
Repair gas leak.
Ensure that furnace
temperature and oxygen flow
are correct.
Ensure that the ambient
temperature and the instrument
temperature zones are stable.
Check instrument fan to be on
and free from obstruction.
Coking (sooting) Incomplete combustion of
sample
Incorrect flows
Gas leak
Too much sample
Ensure that furnace
temperature and oxygen flow
are correct.
Ensure that flows are correct.
Repair gas leak.
Reduce sample size or increase
split introduction.
Furnace problems Oxygen pressure below 40
psi
Increase pressure.
PAC 7090 Series IOS Manual
Page 100 MAN 7090 Rev E
Furnace not heating Vacuum not below setpoint
Temperature controller is
not programmed
Furnace cable is
disconnected
Furnace fuse is blown
Transformer fuse is blown
Temperature Controller
fuse is blown
Furnace element is burnt
out
Faulty solid state relay
(SSR)
Correct vacuums.
Program Controller.
Connect furnace cable.
Replace Fuse.
Replace Fuse.
Replace Fuse.
Replace furnace.
Replace solid state relay
(SSR).
No temp readout-existing
temperature
Loose thermocouple
connector
Defective thermocouple
Faulty solid state relay
(SSR)
Secure thermocouple zone
connector.
Replace thermocouple.
Replace solid state relay
(SSR).
Overheating of a temperature
zone
Improper setpoint
Shorted thermocouple
Faulty solid state relay
(SSR)
Program controller.
Replace thermocouple.
Replace solid state relay
(SSR).
Gas system / Flow problems
Mass flow controllers indicate no
flow
No supply pressure
Closed control valve
Leaks
Restricted gas line
No power to Mass Flow
controllers (MFC)
Supply pressure greater
than 60 psi
Supply pressure less than
40 psi
Defective Mass Flow
controller (MFC)
Mass Flow Controllers
(MFC) board not secure
Loose 16-pin ribbon cable
to Mass Flow Controller
Board
Flow adjustments not made
Check supply gas.
Adjust control valve.
Correct leaks.
Correct restrictions.
Check +15VDC power supply
and fuses.
Correct gas supply pressure.
Correct gas supply pressure.
Replace defective Mass Flow
controller (MFC).
Secure board.
Secure ribbon cable.
Adjust flow settings.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 101
No ozone Ozone switch off
Oxygen pressure below 40
psi
Leaks in ozone system
Faulty ozone generator cell
Faulty ozone generator
printed circuit board
Faulty ozone generator
transformer
Faulty +5VDC power
supply or fuse
Turn ozone on
Increase oxygen pressure.
Correct leaks.
Replace ozone generator cell.
Replace ozone generator
printed circuit board.
Replace ozone generator
transformer.
Replace +5VDC power supply
or fuse.
No hydrogen flow Hydrogen solenoid not
energized
Flow not set
Defective Mass Flow
controller
Check hydrogen solenoid
parameters (See hydrogen
solenoid not energized below).
Correct setting.
Replace Mass Flow Controller.
Hydrogen solenoid not energized
24 VAC not present
Temperature controller not
properly set for alarm
Temperature below alarm
setpoint
Vacuum not below setpoint
Oxygen pressure below 40
psi
Check 24 VAC input/output
fuses.
Program controller.
Increase temperature.
Correct vacuums.
Increase pressure.
Furnace vacuum (P1) not reading
760 when system is open to
atmosphere
Board not calibrated Replace or recalibrate.
Furnace vacuum (P1) not below
setpoint with vacuum applied
Too much flow
Broken pyrotube
Plugged oxygen supply line
to furnace
Leaks
Bad vacuum pump
Board not calibrated
Column broken
Check gas flow settings.
Replace pyrotube.
Ensure oxygen supply line is
clean and unobstructed.
Repair leaks.
Repair or replace.
Replace or recalibrate.
Replace column.
Control Problems
No PMT voltage displayed PMT switched OFF
24VAC transformer input
fuse blown (rear panel)
24 VAC transformer ouput
fuse blown (rear panel)
24 VAC fuse blown (in
tray)
Faulty high voltage power
supply PC board
Shorted high voltage cable
Switch PMT to ON.
Replace fuse.
Replace fuse.
Replace fuse.
Replace high voltage power
supply board.
Replace cable.
PAC 7090 Series IOS Manual
Page 102 MAN 7090 Rev E
No baseline deflection when High
Voltage is turned on
High voltage not present at
photomultiplier tube
Active signal no present or
detected
Loose or defective high
voltage cable or connector
Check voltage at PMT.
Monitor signal level at
instrument display when High
Voltage is turned on or off
Secure high voltage cable
connector.
No baseline deflection when
ozone generator is turned on
Ozone not being generated
Active signal not present or
detected
+5VDC not present at
ozone generator board
Check for presence of ozone
Monitor signal level at
instrument display when ozone
monitor is turned on or off
Correct -5VDC problem.
No Auto-Zero control Loose ribbon cable
Improperly adjusted
amplifier board
PMT voltage too high
Open signal from PMT
Faulty pre-amplifier PC
board
Faulty amplifier PC board
Secure ribbon cable.
Adjust board.
Lower PMT voltage.
Check signal cable from PMT.
Replace pre-amplifier PC
board.
Replace amplifier PC board.
System Status LEDs not
energized
Oxygen pressure is below
40 psi
Increase oxygen pressure to 40
psi.
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 103
Left Blank
PAC 7090 Series IOS Manual
Page 104 MAN 7090 Rev E
10. DATA SHEETS
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 105
PAC 7090 Series IOS Manual
Page 106 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 107
PAC 7090 Series IOS Manual
Page 108 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 109
PAC 7090 Series IOS Manual
Page 110 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 111
NOTE: There is no MSDS-10 page.
PAC 7090 Series IOS Manual
Page 112 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 113
PAC 7090 Series IOS Manual
Page 114 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 115
PAC 7090 Series IOS Manual
Page 116 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 117
PAC 7090 Series IOS Manual
Page 118 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 119
PAC 7090 Series IOS Manual
Page 120 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 121
PAC 7090 Series IOS Manual
Page 122 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 123
PAC 7090 Series IOS Manual
Page 124 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 125
PAC 7090 Series IOS Manual
Page 126 MAN 7090 Rev E
PAC 7090 Series IOS Manual
7090 Series User Manual • Rev D MAN 7090 Page 127
Petroleum Analyzer Company, LP.
Web site: www.paclp.com
For a complete list of contact information, refer to the Contacts section of this publication.