December 13, 2014 Sandra Velandia, AuD Ear Institute, Miller School of Medicine, University of Miami 1 A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND NRTs FOR THE NUCLEUS 422 AND CONTOUR ELECTRODE ARRAYS

Upload: others

Post on 23-May-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

December 13, 2014!Sandra Velandia, AuD!Ear Institute, Miller School of Medicine, University of Miami!

1

A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND NRTs FOR THE NUCLEUS 422 AND CONTOUR ELECTRODE ARRAYS

Page 2: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Sandra Velandia, AuD Diane Martinez, AuD Annelle Hodges, PhD!Rosemary Ojo, MD Simon Angeli, MD!

University of Miami Ear Institute

Miller School of Medicine!

2

Page 3: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Background!

•  A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace the straight Nucleus 22.!

!

•  Claudiu Treaba, Tom Roland and surgeons from Australia, Germany and the U.S. were major contributors in the design of the Contour Advance electrode.!

•  With this design the stimulation contacts are placed closer to the spiral ganglion cells by matching the natural shape of the cochlea.!

•  Focused spiral ganglion stimulation could be achieved with modiolar wall electrode placement.!

3

Page 4: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Background!

•  Research had shown that perimodiolar electrodes deliver unprecedented hearing performance, more focused stimulation and greater power efficiency.!

•  With introduction of the contour electrode, changes in programming as compared to the N22 straight electrode were noted.!

•  T and C levels seemed to be much more consistent.!!

4

Page 5: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

!

5

N22 map!

Contour map!

Page 6: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Background!

•  Skarzynski et al.2012: preliminary report described the new, thinner, lateral wall intracochlear electrode array that was incorporated into the Nucleus Straight Research Array (SRA) cochlear implant. !

•  Developed by the first author in collaboration with Cochlear Ltd., Sydney.!

•  The SRA, renamed Slim Straight TM array, is now available in the Nucleus CI422.!

6

Page 7: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Characteristics of the CA and the CI422!

7

Characteristics! Contour Advance Electrode Array!

CI422 Electrode Array!

Design Concept! •  Pre -Curved for perimodiolar wall placement (modiolus hugging)!

•  Focused spiral ganglion stimulation and greater power efficiency !

•  Slim Straight!•  Smooth surface to

minimize trauma!•  For lateral wall

placement !

Receiver Stimulator ! CIRE24! CIRE24!

Active Length ! 15 mm ! 20 mm !

Page 8: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Characteristics of the CA and the CI422!

8

Characteristics! Contour Advance Electrode Array!

CI422 Electrode Array!

Diameter at apex! 0.5 mm! 0.3 mm !Diameter at base! 0.8 mm! 0.6 mm !Diameter at tip! 0.2 mm!Electrode Contacts ! •  22 half- banded!

•  Medial facing!•  Larger surface

contact positioned in a non-uniform arrangement!

•  22 half- banded!•  Medial facing!

Page 9: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Case study-Map comparison!

9

CI 422!

Contour!

Page 10: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Examples of programming with the two electrodes!

10

Contour map!

CI422 map!

Page 11: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Hypothesis!

The modiolar wall placement of the Contour electrode produces more consistent proximity to the spiral ganglion cells and thus results in more consistent T and C levels, greater power efficiency, and less variation in impedances than the Slim Straight lateral wall design.!

11

Page 12: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Methods: Procedures!

•  Retrospective chart review!•  Data recorded at second programming and at 3 to 6 month

programming sessions!•  Data analyzed statistically to determine if actual differences exist

between the two electrode designs!

12

Page 13: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Methods: Procedures!

•  Data recorded included:!§  Threshold levels!§  Comfort levels!§  NRT thresholds!§  Dynamic range!§  Pulse width!§  Impedances!

13

Page 14: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Methods: Subjects!•  91 subjects for impedance comparison!

§  62 CI 422 patients implanted between 2012 and 2014!§  29 Contour Advance patients implanted between 2011

and 2014!§  No CI512 patients were included!

•  Gender!§  43 female!§  48 male!

•  Mean age at implant 46!•  Age range 7 months to 89 years!

14

Page 15: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Methods: Subjects!•  54 subjects with complete data:!

§  40 CI 422 users!§  14 Contour users!

•  Gender!§  28 females!§  25 males!

•  Mean age at implantation 46 years!•  Age range 5-89 years!

15

Page 16: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Methods: Subjects!

!•  Inclusion criteria included:!

§  Programmed by counting T levels!§  Consistent and reliable judgment of C levels or presence of stapedial

reflexes!

16

Page 17: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Results!

Are T and C levels different for the two electrode designs?!!This could not be analyzed with current data. Variations in PW could not be taken into account due to sample sizes!

17

Page 18: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Results!

Are pulse widths different between the two electrode designs?!!!

18

25! 37! 50!

422! 57%! 38%! 5%!

Contour! 79%! 21%! 0%!

Page 19: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Results!Are impedances different for the two electrode designs? No significant differences at 95% confidence intervals!!!

19

5.00!6.00!7.00!8.00!9.00!

10.00!11.00!12.00!13.00!

1! 2! 3! 4! 5! 6! 7! 8! 9!10!11!12!13!14!15!16!17!18!19!20!21!22!

Impe

danc

e!Electrode!

Contour!

Slim!

Contour(95%CI)!

Slim(95%CI)!

5.00!6.00!7.00!8.00!9.00!

10.00!11.00!12.00!13.00!

1! 2! 3! 4! 5! 6! 7! 8! 9! 10!

11!

12!

13!

14!

15!

16!

17!

18!

19!

20!

21!

22!

Impe

danc

e!

Electrode!

Contour!

Slim!

Contour(95%CI)!

Slim(95%CI)!

2nd visit

3 to 6 mo visit

Page 20: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Results!Are dynamic ranges different between the two electrode designs?!No significant differences at 95% confidence intervals for any electrodes!

!!

20

45.00!

50.00!

55.00!

60.00!

65.00!

70.00!

1! 2! 3! 4! 5! 6! 7! 8! 9!10!11!12!13!14!15!16!17!18!19!20!21!22!

Dyn

amic

Ran

ge!

Electrode!

Contour!

Slim!

Contour(95%CI)!

Slim(95%CI)!

45.00!

50.00!

55.00!

60.00!

65.00!

70.00!

1! 2! 3! 4! 5!6! 7! 8! 9!10!11!12!13!14!15!16!17!18!19!20!21!22!

Dyn

amic

Ran

ge!

Electrode!

Contour!

Slim!

Contour(95%CI)!

Slim(95%CI)!

2nd visit!

3 to 6 mo visit!

Page 21: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Results!Are NRTs different between the two electrode designs?!No significant differences at 95% confidence intervals for any electrodes!

!!

21

120.00!130.00!140.00!150.00!160.00!170.00!180.00!190.00!200.00!

1! 2! 3! 4! 5! 6! 7! 8! 9!10!11!12!13!14!15!16!17!18!19!20!21!22!

NR

T!Electrode!

Contour!

Slim!

Contour(95%CI)!

Slim(95%CI)!

120.00!130.00!140.00!150.00!160.00!170.00!180.00!190.00!200.00!

1! 2! 3! 4! 5! 6! 7! 8! 9!10!11!12!13!14!15!16!17!18!19!20!21!22!

NR

T!

Electrode!

Contour!

Slim!

Contour(95%CI)!

Slim(95%CI)!

2nd visit!

3 to 6 mo visit !

Page 22: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Conclusions!!•  Even though there were no significant differences at

95% confidence intervals, there were trends that reflected our clinical impressions and may bear further study.!

•  There is too much variability in programming across patients to generalize (levels needs).!

•  Clinical application: Every patient has to be programmed on an individualized basis. We cannot assume all 422 DRs are narrower across the array, or that all Contour impedances are lower.!

22

Page 23: A COMPARISON OF IMPEDANCES, DYNAMIC RANGES AND … · 2018-04-03 · Background! • A pre-curved, perimodiolar electrode array was introduced by Cochlear Corporation in 2000 to replace

Future directions!

•  Speech perception results !•  Hearing preservation!