A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During
Hypoglycemia in the Diabetes Research in Children Network (DirecNet) Accuracy Study
Michael Tansey1, Roy Beck2, Bruce Buckingham3, Nelly Mauras4, Rosanna Fiallo-Scharer5, Dongyuan Xing2, William Tamborlane6, Katrina Ruedy2, and the
Diabetes Research in Children Network (DirecNet) Study Group.
1Department of Pediatrics, University of Iowa, Carver College of Medicine, Iowa City, IA; 2Jaeb Center for Health Research, Tampa, FL; 3Division of Pediatric
Endocrinology and Diabetes, Stanford University, Stanford, CA; 4Nemours Children’s Clinic, Jacksonville, FL; 5Barbara Davis Center for Childhood
Diabetes, University of Colorado, Denver, CO; 6Department of Pediatrics, Yale University School of Medicine, New Haven, CT.
![Page 2: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/2.jpg)
The CGMS has been used to follow glucose trends in patients with type1 diabetes, however the accuracy of this device to document hypoglycemia has not been completely evaluated. Recently, the manufacturer of CGMS(Medtronic MiniMed, Northridge, CA) developed a modified glucose sensor using a new fabrication process. To study the accuracy of these sensors, the DirecNet study group examined 91 patients, 3-17 years of age with type 1 diabetes. CGMS glucose values were compared with simultaneous reference venous glucose values, sampled every 30-60 minutes for 24 hours and every 5 minutes for up to 90 minutes during an insulin-induced hypoglycemia test. When all sensor values were analyzed, the 25 CGMSs using modified sensors were substantially more accurate than the 115 CGMSs using original sensors in median error (11% versus 19%, P<0.001) and percentage meeting ISO criteria (72% versus 53%, P<0.001). During hypoglycemia, both original and modified sensors were less accurate than during hyperglycemia, and the enhanced accuracy with the modified sensors was less apparent for hypoglycemia than hyperglycemia. As indicated in the Table below, the original sensors accurately identified only 36% of Reference Glucose (RG) values < 60 mg/dL vs 49% with the modified sensors. All of the modified sensor values were <100 mg/dL when the RG was <60 mg/dL whereas 15% of the original sensor values were >100 mg/dL. Only 8 of 26 overnight hypoglycemic episodes detected by original sensors were RG confirmed, whereas all 3 detected by modified sensors were RG confirmed. Conclusion: Recent modifications to the CGMS sensor have resulted in substantially better accuracy. However, even the modified sensors differ substantially from RG glucose values in the hypoglycemic range.
ABSTRACT
![Page 3: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/3.jpg)
Background• Continuous Glucose Monitoring System (CGMS™) was
the first glucose sensor to be approved by the FDA• CGMS is capable of measuring interstitial fluid glucose
every 5 minutes for up to 72 hours• In November, 2002 Medtronic MiniMed changed the
manufacturing process of the sensor that had been in place since 1999 by modifying the sensor fabrication process
• To evaluate the accuracy of the original and modified CGMS sensors in children and adolescents with type 1 diabetes mellitus, by comparing sensor values to reference glucose levels
Objective
![Page 4: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/4.jpg)
MethodsData pooled from three studies: • Inpatient Accuracy Study
– 91 patients admitted to a CRC for 24-26 hours to evaluate accuracy of the CGMS– One or two CGMS sensors inserted in each patient– Blood samples collected every 60 min during the day, every 30 min overnight, and every 5
minutes for up to 90 minutes during an IV insulin induced hypoglycemia test
• Outpatient Pilot Study– 15 patient pilot study designed to evaluate the feasibility of electronic data collection from a
computer used in the patients’ homes– CGMS used by 10 patients prior to 3-month visit for 48-72 hours – On 2 days of CGMS use, 8-point blood glucose testing performed with Ultra
• Outpatient Randomized Trial (RCT)– Ongoing study of 200 patients– All patients provided with home computer and UltraSmart meter– CGMS used by all patients for 48-72 hours at Baseline, 3 and 6 months– On 2 days of CGMS use, 8-point blood glucose testing performed with UltraSmart – Data available from Baseline on 103 patients
![Page 5: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/5.jpg)
Demographics
Inpatient Study
(N=91) Pilot Study
(N=10) RCT
(N=103)
Mean Age 9.9 ± 4.1 yrs 12.7 ± 3.2 yrs 12.0 ± 2.7
Gender 45 M, 46 F 5 M, 5 F 60 M, 43 F
Race 85% White 80% White 88% White
Duration T1DM 4.7 ± 3.2 yr 6.6 ± 2.8 yr 5.2 ± 3.1 yr
Mean HbA1c 7.8 ± 1.2% 7.4 ± 0.7% 8.3 ± 0.9%
Diabetes Therapy 57% MDI, 43% Pump 30% MDI, 70% Pump 50% MDI, 50% Pump
![Page 6: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/6.jpg)
Data Analysis• CGMS glucose measurements were matched to reference
measurements within 2.5 min of each other
– Inpatient Study reference values: IV blood samples measured in a central laboratory
– Pilot Study and RCT reference values: home glucose meter
• CGMS time was adjusted by 2.5 min to account for averaging of glucose levels made over the previous 5 min
• When comparing CGMS monitors simultaneously, sensor values were matched within 2.5 min of each other
![Page 7: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/7.jpg)
Sensory Accuracy for Reference Glucose Values <60 mg/dL
Inpatient Study Pilot and RCT Studies Original Sensors Modified Sensors Modified Sensors
N=356 N=45 N=61 Sensor Glucose Value N(%)
<60 129 (36%) 22 (49%) 34 (56%) 61-80 108 (30%) 16 (36%) 18 (30%)
81-100 65 (18%) 7 (16%) 5 (8%) 101-120 21 (6%) 0 2 (3%)
>120 33 (9%) 0 2 (3%) Absolute Difference *
Median (25th, 75th %tiles) 20 (8, 37) mg/dL 15 (9, 24) mg/dL
10 (5,18)mg/dL Values within 15 mg/dL of reference N(%) 147 (41%) 23 (51%) 44 (72%)
* Absolute value of sensor glucose minus reference glucose.
![Page 8: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/8.jpg)
Sensor AccuracyInpatient Study
Pilot and RCT Studies
Original Sensors
N=4643
Paired Values
Modified Sensors
N=999
Paired Values
Modified Sensors
N=1262
Paired Values
Relative Absolute Difference Median
(25th, 75th percentiles)
19%
(9%, 34%)
11%
(5%, 23%)
12%
(5%, 23%)
ISO Criteria Met According to Reference Glucose Level
≤70 mg/dL 37% 44% 67%
71-120 mg/dL 44% 64% 65%
121-180 mg/dL 52% 71% 71%
181-240 mg/dL 60% 77% 75%
>240 mg/dL 59% 80% 77%
All Values 54% 72% 72%
![Page 9: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/9.jpg)
Effect of Glucose Concentration on Accuracy
Assessed by Median RAD%
0%
5%
10%
15%
20%
25%
30%
35%
40%
0 50 100 150 200 250 300 350 400
Reference Glucose (mg/dL)
Median RAD
Original CGMS
(Inpatient)
Modified CGMS
(Inpatient)
ModifiedCGMS
(Pilot / RCT)
![Page 10: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/10.jpg)
Conclusions
• The modified CGMS sensor offers improved accuracy compared to the original sensor at all glucose ranges that were measured
• There is decreased accuracy of both sensors in the hypoglycemic ranges when compared to the reference glucose
![Page 11: A Comparison of the Original vs. Modified Continuous Glucose Monitoring System (CGMS™) Sensor During Hypoglycemia in the Diabetes Research in Children](https://reader036.vdocuments.net/reader036/viewer/2022082818/56649efd5503460f94c11a39/html5/thumbnails/11.jpg)
• Barbara Davis Center
– H. Peter Chase– Rosanna Fiallo-Scharer– Jennifer Fisher– Barbara Tallant
• University of Iowa– Eva Tsalikian– Michael Tansey– Linda Larson– Julie Coffey– Amy Sheehan
• Nemours Children’s Clinic– Tim Wysocki– Nelly Mauras– Keisha Bird– Kelly Lofton
• Stanford University– Bruce Buckingham– Darrell Wilson– Jennifer Block– Paula Clinton
• Yale University– William Tamborlane– Stuart Weinzimer– Elizabeth Boland– Kristen Sikes– Amy Steffen
• Jaeb Center for Health Research– Roy Beck– Katrina Ruedy– Craig Kollman– Dongyuan Xing– Cynthia Silvester