a diluent study investigating a wide range of...a set of polar diluents were compared to the...

1
TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2013 Waters Corporation | LL-pdf A DILUENT STUDY INVESTIGATING A WIDE RANGE OF ORGANIC SOLVENTS COMPATIBILITY WITH UPC 2 FOR A MEDICINAL CHEMISTRY LABORATORY Janet Hammond 1 , Michael D. Jones 2 and Sean M. McCarthy 2 1 Waters Corporation, Manchester, UK; 2 Waters Corporation, Milford, MA, USA INTRODUCTION In the medicinal chemistry laboratory, synthetic reactions are carried out using a wide range of solvents. To analyze reaction mixtures, the chemist would favor injecting the reaction mixture directly onto a chromatographic instrument without any dilution. However; this is not practical when considering the high concentration of a reaction mixture, therefore simple dilutions are necessary. The impact of sample diluents on reversed phase separations in relation to peak shape, retention and other factors are well understood. It is generally understood to dissolve the analyte in a solvent with an elutropic strength inversely proportional to the strength of the chromatographic elution solvent. By this definition, non-polar solvents such as hexanes would be optimal diluents for Convergence Chromatography (CC) due to their equivalent elutropic strength to that of CO 2 . However when using UltraPerformance Convergence Chromatography (UPC 2 ), this can be challenging for the analysis of polar compounds due to insolubility with the ‘optimal’ choice of a non-polar solvent. Presently, knowledge of the appropriate diluent choice is under-developed In this study, we investigated the effect of different sample diluents on a probe analyte with a low retention factor (k’) when injected on the UltraPerformance Convergence Chromatography (UPC 2 ) system. Our study included a range of solvents which span a range of polarities and relative elutropic strengths. The goal of this work was to provide recommendations for diluent selection for the analysis of reaction mixtures. Figure 6: A comparison of peak distortion of polar solvents DMF diluted with non-polar solvent heptane/IPA 9:1 using CSH Fluoro-Phenyl 3.0 mm x 100mm; 1.7μm column Figure 5: A comparison of peak distortion with the polar solvent NMP diluted with non-polar solvent heptane/IPA 9:1 using CSH Fluoro-Phenyl 3.0 mm x 100mm; 1.7μm column Methanol Dimethyl sulphoxide (DMSO) Heptane/IPA 90:10 N-methyl pyrrolidone (NMP) Heptane/IPA 70:30 Dimethylformamide (DMF) 2-Propanol (IPA) Dimethyl Acetamide (DMA) Toluene Ethyl Acetate Acetonitrile Methanol/ Dichloromethane1:1 Dichloromethane THF/Heptane 1:1 Dioxane Tetrahydrofuran (THF) Methyl-t-butylether (MTBE) Table 1: Solvents used in the study RESULTS AND DISCUSSION Analyte Probe: 4-Hydroxy butyl-benzoate was the chosen analyte and prepared as 0.2mg/ml solution in a range of polar and non-polar solvents as shown in table 1. EXPERIMENTAL UPC 2 Instrument: ACQUITY UPC 2 w/ PDA detection Mobile Phase: 97% Methanol: 3% CO 2 (medical grade) Columns: UPC 2 BEH UPC 2 CSH Fluoro-Phenyl UPC 2 BEH 2-Ethyl Pyridine UPC 2 HSS C 18 SB Dimensions: 3.0 x 100mm; 1.7um Injection Vol.: see figure captions Run time: 3 minutes Column Temp.: 45 °C Flow Rate: 2.0 ml/min ABPR pressure: 2000 psi Wavelength: 254 nm CDS: MassLynx 4.1 A comparison between the non-polar solvent of heptane/IPA 9:1 and DMSO when varying injection volumes (Figure 1). In the case of DMSO, peak distortion is observed when the injection is increased to 2μl, however peak shape is Gaussian for the heptane/IPA 9:1 diluent. Once the injection volume was increase to 5ul, peak distortion occurred irrespective diluent. The use of 100% methanol as the diluent was investigated for the probe analyte on three different columns. Each column resulted in a difference in retention (k’). The comparison of the peak shape from the three columns is shown in figure 2. This example suggests that analytes eluting with a lower k’ is more susceptible to peak distortion when using 100% MeOH. A set of polar diluents were compared to the non- polar diluent; used in Figure 1, by injection of the analyte probe on the UPC 2 BEH column (Figure 3). Each injection of the diluent study showed acceptable peak shape. It should be noted that retention of the DMA and DMF were observed and would interfere with analytes with lower k’. NMP retention was also observed, but exhibits low absorbance at 254nm which is a typical monitoring wavelength. The same experiment performed for Figure 3 was repeated by injection of the probe analyte on the UPC 2 CSH Fluoro-Phenyl column. The extent of the peak distortion was observed to be much more severe for this column when compare to the UPC 2 BEH results. Further investigations to explaining this phenomena would have to consider loadability studies such as those performed in Figure 1 for the different columns. *Diluent retention does not appear to be the major issue, but should be confirmed as part of the further investigation. Figure 3: A comparison of peak distortion of polar solvents DMSO, NMP, DMF, DMA and heptane/IPA 9:1 using a UPC 2 BEH 3mm x 100mm; 1.7μm column DMA DMF NMP DMSO Hep/IPA Figure 4: A comparison of peak distortion of polar solvents DMSO, NMP, DMF, DMA, and heptane/IPA 9:1 using a UPC 2 CSH Fluoro-Phenyl 3mm x100mm; 1.7μm column DMA DMF NMP DMSO Hep/IPA In many cases, a mixture of these solvents may be used for solubility or reaction rate purposes. The experiments from Figure 4 were taken further to explore varying percentages of the NMP and DMF solvent combined with heptane/IPA 9:1. Compositional mixtures of 2%, 5%, 10%, 20%, and 50% of each of the NMP and DMF were combined with the heptane/IPA mixture. The NMP results (Figure 5) show Gaussian peak shapes for the probe analyte up to a compositional mixture of [10% NMP : 90% heptane/IPA] on the Fluoro- Phenyl column. The varying compositional mixtures of DMF and heptane/IPA diluent (Figure 6) showed peak distortion with each compositional diluent mixture suggesting alternative solvents or an alternative column which provides a greater k’ to minimize the strong solvent effects should be considered when using DMF. Figure 2: Comparison of peak distortion when 3 different col- umns were used A) UPC 2 CSH Fluoro-Phenyl, B) UPC 2 BEH, and C) UPC 2 BEH 2-EP UPC 2 CSH Fluoro-Phenyl UPC 2 BEH UPC 2 BEH 2-EP Figure 1: Comparison of peak distortion between DMSO and Heptane/IPA 9:1 at various injection volumes 5μl DMSO 2μl DMSO 0.5μl DMSO 5μl Hep/IPA 2μl Hep/IPA 0.5μl Hep/IPA +2% DMF Hep/IPA +5% DMF +10% DMF +20% DMF +50% DMF Neat DMF +2% NMP Hep/IPA +5% NMP +10% NMP +20% NMP +50% NMP Neat NMP CONCLUSIONS Many of the solvents in Table 1 were highly compatible for use as diluents when using UPC 2 , with some precautionary exceptions... Strong diluent effects can be observed for some solvents on specific columns, such as the case with DMF and UPC2 Fluoro-Phenyl combinations. Lower k’ analytes are more susceptible to peak distortion. Dilution with a weaker elution solvent (such as 9:1 heptane:IPA) can minimize chromatographic band- spreading of the analyte caused by strong solvent effects.

Upload: others

Post on 27-Jan-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

  • TO DOWNLOAD A COPY OF THIS POSTER, VISIT WWW.WATERS.COM/POSTERS ©2013 Waters Corporation | LL-pdf

    A DILUENT STUDY INVESTIGATING A WIDE RANGE OF ORGANIC SOLVENTS COMPATIBILITY WITH UPC2 FOR A

    MEDICINAL CHEMISTRY LABORATORY

    Janet Hammond1, Michael D. Jones2 and Sean M. McCarthy2 1Waters Corporation, Manchester, UK; 2Waters Corporation, Milford, MA, USA

    INTRODUCTION In the medicinal chemistry laboratory,

    synthetic reactions are carried out using a

    wide range of solvents. To analyze reaction

    mixtures, the chemist would favor injecting

    the reaction mixture directly onto a

    chromatographic instrument without any

    dilution. However; this is not practical when

    considering the high concentration of a

    reaction mixture, therefore simple dilutions

    are necessary.

    The impact of sample diluents on reversed

    phase separations in relation to peak shape,

    retention and other factors are well

    understood. It is generally understood to

    dissolve the analyte in a solvent with an

    elutropic strength inversely proportional to

    the strength of the chromatographic elution

    solvent. By this definition, non-polar solvents

    such as hexanes would be optimal diluents

    for Convergence Chromatography (CC) due to

    their equivalent elutropic strength to that of

    CO2. However when using UltraPerformance

    Convergence Chromatography (UPC2), this

    can be challenging for the analysis of polar

    compounds due to insolubility with the

    ‘optimal’ choice of a non-polar solvent.

    Presently, knowledge of the appropriate

    diluent choice is under-developed

    In this study, we investigated the effect

    of different sample diluents on a probe

    analyte with a low retention factor (k’) when

    injected on the UltraPerformance

    Convergence Chromatography (UPC2) system.

    Our study included a range of solvents which

    span a range of polarities and relative

    elutropic strengths. The goal of this work was

    to provide recommendations for diluent

    selection for the analysis of reaction mixtures.

    Figure 6: A comparison of peak distortion of polar solvents

    DMF diluted with non-polar solvent heptane/IPA 9:1 using CSH Fluoro-Phenyl 3.0 mm x 100mm; 1.7µm column

    Figure 5: A comparison of peak distortion with the polar

    solvent NMP diluted with non-polar solvent heptane/IPA 9:1 using CSH Fluoro-Phenyl 3.0 mm x 100mm; 1.7µm column

    Methanol Dimethyl sulphoxide

    (DMSO)

    Heptane/IPA 90:10 N-methyl pyrrolidone

    (NMP)

    Heptane/IPA 70:30 Dimethylformamide (DMF)

    2-Propanol (IPA) Dimethyl Acetamide (DMA)

    Toluene Ethyl Acetate

    Acetonitrile Methanol/

    Dichloromethane1:1

    Dichloromethane THF/Heptane 1:1

    Dioxane

    Tetrahydrofuran

    (THF)

    Methyl-t-butylether

    (MTBE)

    Table 1: Solvents used in the study

    RESULTS AND DISCUSSION

    Analyte Probe:

    4-Hydroxy butyl-benzoate was the chosen analyte

    and prepared as 0.2mg/ml solution in a range of

    polar and non-polar solvents as shown in table 1.

    EXPERIMENTAL UPC2

    Instrument: ACQUITY UPC2 w/ PDA detection

    Mobile Phase: 97% Methanol: 3% CO2 (medical grade) Columns: UPC2 BEH

    UPC2 CSH Fluoro-Phenyl UPC2 BEH 2-Ethyl Pyridine

    UPC2 HSS C18 SB

    Dimensions: 3.0 x 100mm; 1.7um

    Injection Vol.: see figure captions

    Run time: 3 minutes

    Column Temp.: 45 °C

    Flow Rate: 2.0 ml/min

    ABPR pressure: 2000 psi

    Wavelength: 254 nm

    CDS: MassLynx 4.1

    A comparison between the non-polar solvent of

    heptane/IPA 9:1 and DMSO when varying injection

    volumes (Figure 1). In the case of DMSO, peak

    distortion is observed when the injection is increased

    to 2µl, however peak shape is Gaussian for the

    heptane/IPA 9:1 diluent. Once the injection volume

    was increase to 5ul, peak distortion occurred

    irrespective diluent.

    The use of 100% methanol as the diluent was

    investigated for the probe analyte on three different

    columns. Each column resulted in a difference in

    retention (k’). The comparison of the peak shape

    from the three columns is shown in figure 2. This

    example suggests that analytes eluting with a lower

    k’ is more susceptible to peak distortion when using

    100% MeOH.

    A set of polar diluents were compared to the non-

    polar diluent; used in Figure 1, by injection of the

    analyte probe on the UPC2 BEH column (Figure 3).

    Each injection of the diluent study showed acceptable

    peak shape. It should be noted that retention of the

    DMA and DMF were observed and would interfere with

    analytes with lower k’. NMP retention was also

    observed, but exhibits low absorbance at 254nm

    which is a typical monitoring wavelength.

    The same experiment performed for Figure 3 was

    repeated by injection of the probe analyte on the

    UPC2 CSH Fluoro-Phenyl column. The extent of the

    peak distortion was observed to be much more

    severe for this column when compare to the UPC2

    BEH results. Further investigations to explaining this

    phenomena would have to consider loadability studies

    such as those performed in Figure 1 for the different

    columns. *Diluent retention does not appear to be the major issue, but should be confirmed as part of the further investigation.

    Figure 3: A comparison of peak distortion of polar solvents

    DMSO, NMP, DMF, DMA and heptane/IPA 9:1 using a UPC2 BEH 3mm x 100mm; 1.7µm column

    DMA

    DMF

    NMP

    DMSO

    Hep/IPA

    Figure 4: A comparison of peak distortion of polar solvents

    DMSO, NMP, DMF, DMA, and heptane/IPA 9:1 using a UPC2 CSH Fluoro-Phenyl 3mm x100mm; 1.7µm column

    DMA

    DMF

    NMP

    DMSO

    Hep/IPA

    In many cases, a mixture of these solvents may be used for solubility or reaction rate purposes. The

    experiments from Figure 4 were taken further to explore varying percentages of the NMP and DMF solvent

    combined with heptane/IPA 9:1. Compositional mixtures of 2%, 5%, 10%, 20%, and 50% of each of the

    NMP and DMF were combined with the heptane/IPA mixture. The NMP results (Figure 5) show Gaussian peak

    shapes for the probe analyte up to a compositional mixture of [10% NMP : 90% heptane/IPA] on the Fluoro-

    Phenyl column. The varying compositional mixtures of DMF and heptane/IPA diluent (Figure 6) showed

    peak distortion with each compositional diluent mixture suggesting alternative solvents or an alternative

    column which provides a greater k’ to minimize the strong solvent effects should be considered when using

    DMF.

    Figure 2: Comparison of peak distortion when 3 different col-

    umns were used A) UPC2 CSH Fluoro-Phenyl, B) UPC2 BEH,

    and C) UPC2 BEH 2-EP

    UPC2 CSH Fluoro-Phenyl

    UPC2 BEH

    UPC2 BEH 2-EP

    Figure 1: Comparison of peak distortion between DMSO and

    Heptane/IPA 9:1 at various injection volumes

    5µl DMSO

    2µl DMSO

    0.5µl DMSO

    5µl Hep/IPA

    2µl Hep/IPA

    0.5µl Hep/IPA

    +2% DMF

    Hep/IPA

    +5% DMF

    +10% DMF

    +20% DMF

    +50% DMF

    Neat DMF

    +2% NMP

    Hep/IPA

    +5% NMP

    +10% NMP

    +20% NMP

    +50% NMP

    Neat NMP

    CONCLUSIONS Many of the solvents in Table 1 were highly

    compatible for use as diluents when using UPC2, with

    some precautionary exceptions...

    Strong diluent effects can be observed for some

    solvents on specific columns, such as the case with

    DMF and UPC2 Fluoro-Phenyl combinations.

    Lower k’ analytes are more susceptible to peak

    distortion.

    Dilution with a weaker elution solvent (such as 9:1

    heptane:IPA) can minimize chromatographic band-

    spreading of the analyte caused by strong solvent effects.