a fast sequential injection analysis system for the simultaneous determination of ammonia and...
TRANSCRIPT
Microchim Acta 154, 31–38 (2006)
DOI 10.1007/s00604-006-0496-y
Original Paper
A Fast Sequential Injection Analysis System for the SimultaneousDetermination of Ammonia and Phosphate
Carsten Frank1;2;�, Friedhelm Schroeder1, Ralf Ebinghaus1;2, and Wolfgang Ruck2
1 GKSS, Institute for coastal research, Max-Planck-Strasse, D-21502 Geesthacht, Germany2 University of Lueneburg, D-21332 Lueneburg, Germany
Received August 6, 2005; accepted November 9, 2005; published online March 30, 2006
# Springer-Verlag 2006
Abstract. A flow system is described that is based
on sequential injection analysis (SIA) and is suitable
for the fast determination of ammonia and phosphate
in river and marine waters. It is applicable to nutri-
ent mapping in inhomogeneous coastal areas like the
Wadden Sea, and was optimised on several cruises on
the North Sea. The high sample throughput of 120
samples per analyte per hour and the low reagent con-
sumption (ammonia: 62.5 mL; phosphate: 60mL) were
achieved using a home made programme written in
the python programming language. The determination
of free reactive phosphate is based on the reaction of
phosphate with acidic molybdate to phosphomolybdate
which forms nonfluorescent ion pairs with rhodamine
6G. The remaining rhodamine fluorescence is detected
at 550 nm with an excitation at 470 nm. Ammonia
is determined with the help of o-phthaldialdehyde
and sodium sulfite. At 85 �C and a reaction time of
at least one minute a fluorescent species (exc. 365 nm,
em. 425 nm) is formed. The detection limits are
(3�) 0.3 mmol L�1 for phosphate and 1 mmol L�1 for
ammonia.
Key words: Phosphate; ammonia; sequential injection analysis;
marine waters.
Phosphorus as phosphate and nitrogen as nitrite,
nitrate and ammonia are the most important growth
limiting nutrients for phytoplankton in aquatic envi-
ronments [1]. In larger Seas like the North Sea phy-
toplankton accounts for more than 90% of the total
primary production [2]. Every change in the avail-
ability and composition of phosphorus and nitrogen
effects the phytoplankton population [3–5]. This again
has a direct or indirect effect on the phytoplankton
dependent food web and the whole ecosystem in the
affected area [6]. One example for a direct effect is the
increased amount and intensity of toxic and nontoxic
algal blooms [5].
To effectively emend past and prevent future errors
it is essential to determine and quantify the sources,
sinks and transport pathways of nutrients in the
problem areas (e.g. those defined by the OSPAR
Commission [7]). In areas with high hydrologic and
biologic dynamics like the Wadden Sea and the Elbe
estuary that quantification is difficult due to the in-
homogeneity of the water body. To succeed samples
have to be taken from a tight grid of sampling sta-
tions and out of different water depths. For larger
areas, like the Elbe estuary or parts of the North
Sea manual approaches for sampling and analysis
are too inefficient. An undulating towed vehicle like
the one developed and used by the BSH (Federal� Author for correspondence. E-mail: [email protected]
Maritime and Hydrographic Agency, Germany) [8]
could be used as a carrier for a fast in situ nutrient
analyser.
The towfish (developed and operated by the BSH)
is designed to dive and surface in a sawtooth like
pattern (see Fig. 1). It is already used quarterly to
monitor physical, chemical and biological parameters
(like temperature, pH, salinity, algal fluorescence) in
the whole North Sea. A nutrient analyser attached to
or integrated into that towfish has to meet high
demands. These demands refer to technical and method-
ical demands. Technical demands are weight, size,
buoyancy and power consumption, which all have to
be accounted for. Even more difficult to accomplish
are the methodical demands, like high reliability,
low reagent consumption and high sample frequency.
While methods for the parallel determination of
ammonia and phosphate are already known (e.g.
[9, 10]), none of these methods fulfill all mentioned
requirements sufficiently.
In this presentation a modified SIA system is intro-
duced which complies with all methodical as well as
some technical demands for the integration into a
towfish. The system consists of one syringe pump,
one valve, one thermostate two detectors and one
personal computer. Next to a low reagent consump-
tion (ammonia 62.5 mL and phosphate 60 mL per
determination) it reaches a sample throughput of up
to 180 samples per hour and per analyte. The soft-
ware controlling the SIA instrument is tolerant con-
cerning common errors. The validity of the gained
results is improved by continuously monitoring sys-
tem performance.
Experimental
Reagents
All reagents were prepared with fresh drawn degassed deionised
water. Sigma (www.sigmaaldrich.com) analytical grade chemicals
were used, unless otherwise stated.
Ammonia
O-phthaldialdehyde (OPA) stock solution was prepared by dissolv-
ing 2 g of o-phthaldialdehyde (Sigma P-1378) in 25 mL ethanol.
This solution has to be shaken for several minutes to achieve com-
plete dissolution. 2 g of sodium sulfite were dissolved in 250 mL to
prepare the sulfite stock solution.
Ammonia reagent (AR) 7.5 g of disodium tetraborate deca-
hydrate were diluted to 250 mL. The solution was stirred until
complete dissolution and then transferred into a dark glass bottle.
5 mL of OPA stock solution were added. After stirring 500mL of
sulfite stock solution and 0.1 mL of a 30% Brij (Merck 101894,
www.merck.de) solution were added. After stirring the solution was
left to stand for several hours [11].
Ammonia standard stock solution was 1 g L�1 from Merck.
Phosphate
Rhodamine stock solution was prepared by dissolving 0.20 g of
rhodamine 6G in 100 mL water. Molybdate stock solution was made
by dissolving 12.8 g of ammonium heptamolybdate tetrahydrate
(Merck, analytical grade) in 100 mL water. To prepare reagent 1
(PR1) 200mL rhodamine 6G stock solution was added to 90 mL
water. 500mL 5% IGEPAL (Polyoxyethylene(�)octylphenyl ether,
branched) was added and the solution diluted to 100 mL. Reagent 2
(PR2) was prepared by adding 8.45 mL of 30% (v=v) hydrochloric
acid to about 75 mL of water. Add 4 mL of molybdate stock solution
and dilute to 100 mL. The reagents PR1 and PR2 were derived from
Wei et al. [12].
Phosphate standard stock solution was 1 g L�1 from Merck.
Instrumentation
A schematic diagram of the flow system of the automated water
measurement system (ferrybox [13, 14]) with its connections to the
Fig. 1. The towfish of the BSH (Federal Maritime and Hydro-
graphic Agency, Germany) pulled by the research vessel GAUSS.
The towfish is able to dive and surface in a sawtooth like pattern.
This is used to continuously gain depth profiles without stopping
the ship
Fig. 2. The schematic of the flowsystem of the ship and the auto-
mated water measurement system (ferrybox), which was used as
source for the on-line samples. The tests and validations were
performed on the research vessel Ludwig Prandtl using the sam-
ples provided by the ferrybox. SWI Sea water intake; SWD sea
water discharge; CP centrifugal pump; PP peristaltic pump; TP1 &
TP2 T-pieces; F band-pass filter
32 C. Frank et al.
SIA-system and the ship (research vessel Ludwig Prandtl) is shown
in Fig. 2. The sample is drawn by the centrifugal pump out of about
1.20 m water depth at the bow of the ship. It is pumped through the
ship until it reaches the laboratory container which contains the
ferrybox and the SIA instrument. The T-piece (TP1) is used to split
the sample stream between ferrybox and waste using the high pump
rates to ensure sample freshness. After debubbling 4.5 L min�1 of
the sample stream are delivered to a band-pass filter (Metrohm
series 13, www.metrohm.de; Filter paper: Schleicher & Schuell type
1573; pore size 12–25mm, www.schleicher-schuell.de). The peri-
staltic pump (PP) of the band-pass filter is used to pump the filtrate
to the SIA system (Fig. 3). There T-piece TP2 is used as a bypass.
Into the T-piece a tubing (6.5 cm long, 0.8 mm i.d.) is inserted.
This tubing is used to draw fresh samples out of the filtrate stream.
The other end of the tubing is connected to the 17-port=1-channel
valve (V; Knauer Part-No. A1492, www.knauer.net). The SIA
system is composed of one syringe pump (CAVRO XL 3000,
www.tecansystems.com), one valve, two detectors (one fluorescence
detector exc. 365 nm, em. 425 nm build by ME Grisard GmbH,
www.me-grisard.de, and a Hitachi F1000 fluorescencespectrometer
exc. 470 nm, em. 550 nm, www.hitachi-hta.com) and four reaction
loops. The holding loop is 120 cm long (0.8 mm i.d.) and the reac-
tion loops are 60 cm long (0.8 mm i.d.). The Hitachi detector is
connected via an external multimeter with RS232 interface (HP
34401A, www.agilent.com) to the personal computer while the other
detector provides its own RS232 interface. Three reaction loops are
heated to 85 �C and are used to determine ammonia. The reaction
loop for the determination of phosphate is not temperature con-
trolled. A set of home made programmes written in python is used
to control the SIA instrument.
This programme set consists of three essential programmes and
four optional programmes. Two of the essential programmes are
used to control the detectors (one programme per detector). These
control programmes take care for the correct detector initialisation
and then switch into data server mode. The third essential pro-
gramme is the main programme which controls the SIA instrument
(syringe pump(s), valve(s), thermostate(s)). These three programmes
do only interact during peak determination. The main advantage of
this approach is the uninterrupted availability of the detector data.
Four more programmes can be used in conjunction with the three
essential programmes. Two are used to display the raw detector data
and two are used to display preliminary quantification results.
The structure of the main programme is shown in Fig. 4. The
programme starts with the initialisation of all used devices bringing
them into defined states. Due to the low reagent and sample con-
sumption the tubings (40 cm; 0.8 mm i.d.) connecting the reagents
and standards with the valve, have to be flushed with the respective
reagent or standard. The main loop starts with the ammonia readout
step. Directly after the startup none of the three reagent loops
is loaded with the sample-reagent-mix, so the actual reagent loop
is just flushed with carrier. Then the actual ammonia reagent loop is
loaded with the reagent-sample-mix (see Fig. 6). Phosphate is deter-
mined via phosphomolybdate and ion-pairing with rhodamine 6G in
the next step. Two factors limit the speed of the this step: The
dispersion of the reagents depend on the speed profile which is used
Fig. 3. Schematic of the sequential injection analysis (SIA) system
used for phosphate and ammonia analysis. A personal computer is
used to control syringe pump, valve and the detectors. SP Syringe
pump (2500 mL syringe); V 17 port=1 channel valve; C carrier
(degassed deionised water); hl holding loop; rl1–3 reaction loops
for the determination of ammonia; rl4 reaction loop for the deter-
mination of phosphate; D1 fluorescence detector (exc. 365, em.
425; for ammonia); D2 fluorescence detector (exc. 475, em. 550;
for phosphate); PR1 rhodamine reagent; PR2 acidic molybdate
reagent; AR OPA-reagent; Std standard; W waste
Fig. 4. Flow sheet of the home made software used to control the
SIA instrument. ‘x’ is the number of the ammonia reaction loop
used in the actual programme cycle. It is incremented prior to
every loop. The ‘wait’ times are calculated based on the time
needed for each step and the time left until the next ammonia
readout step has to occur
Sequential Injection Analysis of Ammonia and Phosphate 33
to push the reagent-sample-mix to the detector [15, 16]. This leads
to a compromise between speed and reaction efficiency. The limited
time resolution of the detector is the second speed limiting factor. At
this point the loop starts again with the ammonia readout step.
During the fourth run of the main loop the first reaction loop is
read out. This readout procedure starts with a ‘wait’ step which
ensures that the reaction time is reached before the actual determi-
nation starts. The ‘wait’ times are calculated based on the time
needed for each step and the time left until the next ammonia
readout step is due. The reaction time of the ammonia determination
was changed from 60 to 90 s (also decreasing the sample throughput
to 120 samples analyte and hour) while using the ME Grisard
detector to compensate the poor detector performance.
The contents of the main loop can be changed any time with the
changes taking effect in the next cycle. This is especially advanta-
geous during method development, but can also be of great help in
the field. This flexibility can, for example, be used to optimise a
speed profile during method development or to add an additional
standard to the sample sequence in the field. The sample sequences
(see Fig. 5) were introduced to improve the quality of the achieved
data by continuously monitoring system performance.
The sequences of reagents and sample are schematically pictured
in Fig. 6. The sequence for the phosphate determination contains
three reagent segments (PR1 and PR2) and the sample segment. For
the determination of ammonia only one reagent is used. However,
due to problems with the sensitivity of the detector, the amount and
volume of the reagent and sample segments were increased to reach
a satisfying sensitivity (a strategy proposed by Wang et al. [17]). So
three reagent and two sample segments were used (Fig. 6). Earlier
experiments using the Hitachi F1000 detector instead of the one
made by ME Grisard have shown significant better results with less
segments and lower volumes. However, only one of these detectors
was available and it had to be used for phosphate.
The methods for the determination of phosphate and ammonia are
both based on fluorescence determination. Phosphate is determined
via the formation of phosphomolybdate which quantitatively extin-
guishes the fluorescence of the rhodamine 6 g. The reaction of
ammonia with o-phthaldialdehyde (OPA) and sulfite is used to
measure the ammonia concentration.
Results
System Speed
Although the SIA is considered to be slower than the
FIA [18] it was accelerated to an at least similar
speed. This was achieved mainly using the self made
software which intensively uses the features integrated
into the syringe pump (CAVRO XL 3000). One of the
Fig. 5. An example for a sample sequence. At a rate of 120
samples per hour 6 minutes are needed to complete this se-
quence. The standard and blank values are used to monitor system
performance
Fig. 6. Sequence of reagents in the holding loop prior to injection
into the reagent loop. Volumes in mL. AR Ammonia reagent; PR1
phosphate reagent 1; PR2 phosphate reagent 2
Fig. 7. Schematic diagram of the readout step for the phosphate
determination. The speed ramps are used to prevent the formation
of air bubbles in the tubings due to vacuum conditions during
deceleration. (1) The reagents are pumped through the reaction
loop in direction to the detector. (2) The remaining fluorescence is
determined in the detector. (3) Reaction loop and detector are
flushed. (4) Syringe command
34 C. Frank et al.
most advantageous features are the free configurable
acceleration and deceleration slopes. These speed
ramps can be used to achieve high pump rates without
pressure peaks during acceleration and deceleration
phases. The speed ramp used for the phosphate deter-
mination is displayed schematically in Fig. 7.
A multichannel stopped flow approach was chosen
to successfully integrate a high throughput ammonia
determination. The system was tested for up to four
heated reaction loops reaching more than 200 samples
per hour. Since the reaction is interrupted before com-
pletion (70% of the final fluorescence intensity are
reached after 60 s reaction time) the timing has to
be very precise. A precision of about �0.2 seconds
was achieved using the timing described above.
Interferences Between Ammonia and Phosphate
Determination
The contemporaneous use of ammoniamolybdate as
reagent and ammonia as analyte in the same system
was not accompanied by any problems. This was
tested in the lab and during campaigns by deactivating
the phosphate or the ammonia determination for some
minutes and monitoring the effects on the preliminary
data. No difference in the signals of the standards and
the sample was observed.
System and Data Reliability
An improved system reliability was achieved by split-
ting the original large programme into a set of smaller
programmes. Each of these programmes has to ac-
complish only a small amount of tasks like gaining
the data from the detector, saving the data and deliver-
ing parts of these data on request. All non critical
tasks, like the visualisation of the data in a graph,
were moved to separate programmes which can be
used at will. This lead to a significant increase in
system stability and simplified debugging in case of
an error. Another improvement was the integration of
basic error correction features for comparatively com-
mon errors like erroneous data transmissions.
Especially during the first field tests the quality of
the achieved data is always questionable. Variations in
the overall system performance can be caused by tem-
perature differences, wind, vibrations and other fac-
tors. To improve the quality of the data gained with
this SIA system the system of sample sequences was
adopted. This was done using two or more additional
valve ports to attach two or more standards. During
the cruises performed to optimise and validate this
device the determination sequence usually comprises
of 50% standards and 50% samples as shown in Fig. 5.
Effect of Sample Salinity
Sample dilution and standard addition were used to
investigate the effect of the salinity of the sample on
the determined nutrient concentration. Both experi-
ments showed, that the salinity of the sample has
no distinguishable influence on the result of the
determination.
Detection Limits and Range Switching
In this SIA system two different fluorimetric methods
are used. While ammonia is determined via a fluores-
cent molecule which is formed by ammonia, sulfite
and o-phthaldialdehyde, the formation of phosphomo-
lybdate extinguishes quantitatively the fluorescence of
rhodamine 6 g, and therefore phosphate is determined
by the absence of fluorescence. Due to the strong
fluorescence of the rhodamine which is very effec-
tively quenched by phosphomolybdate, the linear
range of this system is quite small. This is compen-
sated using up to three different sample volumes for
different expected phosphate concentrations (50, 17
and 8 mL) [20].
Using 50mL as sample volume a detection limit
(3�) of 0.04 mmol L�1 can be reached for phosphate
in the laboratory. On the ship the situation is comple-
tely different. There the detection limit varies between
0.04 up to 0.3mmol L�1. This is most probably caused
by variations of environmental parameters like tem-
perature, wind, air humidity and incidence of light
which have an effect on the spectrometer, the reaction
loop and the reagents. These effects can most prob-
ably be reduced by exchanging the detector with one
more suitable for shipboard use. Up to 4 mmol L�1
phosphate can be measured using a sample volume
of 8mL.
Ammonia can be quantified directly using the fluo-
rescence intensity (peak area) and shows a calibration
graph with a good linearity up to 20mmol L�1. The
detection limit can be as low as 0.02mmol L�1 using
the Hitachi fluorescence detector in the laboratory.
Because that detector was already in use for the deter-
mination of phosphate a detector produced by ME
Grisard was used. That one has an inferior perfor-
Sequential Injection Analysis of Ammonia and Phosphate 35
mance compared to the Hitachi detector. By increas-
ing the reaction time, the sample volume and the
amount of reagent and sample segments a detection
limit of about 0.05mmol L�1 was achieved in the
laboratory. However, on the ship the average detection
limit reached values of about 1 mmol L�1 (range from
0.07 to 1.50mmol L�1) which was caused mainly by
failed or incorrect data transmissions between de-
tector and computer. Again a detector more suitable
for this environment seems to be the solution to this
problem.
Application and Validation
In order to optimise and validate the newly developed
system, several cruises to the German part of the
North Sea and Wadden Sea were performed. These
areas are characterised by strong nutrient concentra-
tion gradients (Wadden Sea) and low nutrient concen-
trations (North Sea). Therefore the chosen transects
perform an ideal validation scheme for the detection
of strong concentration gradients as well as for the
quantification of low nutrient concentrations directly
on the sea. One transect of a cruise performed in
August 2004 is described here in detail (map see
Fig. 8). The results of this experiment are shown in
the Figs. 9 and 10. The trend of the determined ammo-
nia and phosphate concentration was as expected from
earlier cruises and literature [21]. Starting with com-
paratively high values in Norddeich (the ship’s cen-
trifugal pump (CP see Fig. 2) was switched on about
2–4 km off the harbour after reaching 2–3 m water
depth) the concentrations decreased along the tideway
until the open sea was reached. The concentration of
about 0.5mmol L�1 for both, ammonia and phosphate,
remains stable until the area influenced by the Elbe
river is reached. Both values increase again during the
last part of that trip. About 7 km off Cuxhaven har-
bour the ammonia detector stopped to deliver data and
so the measurement was terminated.
In addition to the on-line measurements off-line
samples were taken. These samples were sent to a
commercial lab which uses an SFA system to quantify
nutrients [22, 23]. Five of these off-line samples were
splitted and determined also off-line with the method
presented here. The results of these measurements are
displayed in the Figs. 9, 10 and in Table 1. The first
Table 1. Comparison of off-line data with data from the external
lab
Slope Intercept r2
Phosphate 0.95 0.010 0.996
Ammonia 1.35 �0.737 0.902
Fig. 8. Map of the transect from Norddeich to Cuxhaven. 5th of
August 2004
Fig. 9. Transect from Norddeich to Cuxhaven. Phosphate data.
Rectangles indicate lab samples
Fig. 10. Transect from Norddeich to Cuxhaven. Ammonia data.
Rectangles indicate lab samples
36 C. Frank et al.
two figures present the results of the on-line mea-
surements as described above and the results of the
samples sent to the lab. These graphs indicate that
the tendency of the on-line and lab results is the same
for phosphate (r2¼ 0.93) and similar for ammonia
(r2¼ 0.63). However the absolute values differ signif-
icantly. The difference between on-line ammonia
data and the lab results are caused by contamination,
inappropriate sample storage and by detector noise.
Furthermore the flow paths of the seawater differ for
the on-line and the lab samples. All lab samples were
drawn at TP1 (Fig. 2), filtered using Schleicher &
Schuell spartan 30=0.2 (pore size 0.2mm) filters
and directly stored in the freezer. Whatman GF=F
(pore size 0.7mm) filters were used for the five splitted
samples. The on-line samples however had to pass
the ferrybox with its integrated additional centrif-
ugal pump. This pump increases the number of
destroyed plankton cells whose contents are released
into the water. In the case of phosphate additional
problems were caused by the increased turbulence
in the pump and the ferrybox which leads to
an increased desorption of phosphate ions from
suspended particles. That water is then filtered
using Schleicher & Schuell type 1573 (pore size
12–25mm) filter paper in the band pass filter. The
increased amount of phosphate ions due to the turbu-
lence caused by centrifugal pump, the increased
amount of particles with adsorbent phosphate [24] in
the sample due to the coarse filter and the increased
number of destroyed plankton cells also lead to
higher phosphate concentrations in the on-line sample
stream.
Due to the significant differences between the on-
line data and the data provided by the external lab the
off-line results were also compared to the lab data.
The results of this comparison are given in Table 1.
The correlation between the off-line and lab results is
very good for phosphate and acceptable for ammonia.
While sample pretreatment and storage is difficult for
both analytes [25, 26], the procedure used in this case
is generally less optimal for ammonia due to the mate-
rial of the viles (PE) and the freezing and melting
processes.
A better approach to validate on-line methods like
the one presented here is to use another automated
method which also can be used on-line at a slower
pace. Another possibility would be a manual method
which can be carried on the ship directly after a sam-
ple is taken.
Discussion
A sequential injection analysis system is presented
which is suitable for the fast determination of ammo-
nia and phosphate in coastal waters. The system was
successfully tested on several cruises on the north sea.
One transact of these cruises was used to validate this
system. The methods used in this device were opti-
mised for the use in a towfish system.
The presented methods are especially useful for
three application fields. First it can be used for high
resolution surface mapping of ammonia and phos-
phate especially in coastal areas. Furthermore it is
useful for long term monitoring due to the low amount
of reagents used in this system, and the reliability
which comes with the integration of standards which
can be determined contemporaneous with the sam-
ples. A third field would be the integration of more
parameters like nitrate or silicate and use this system
as general purpose nutrient analyser.
In the near future the presented method will be
transferred to significant smaller components (e.g. a
Lee LPV pump instead of a CAVRO XL 3000, smaller
detectors with integrated analog to digital converters)
and then integrated into the towfish.
References
[1] Redfield A C, Ketchum B H, Richards F A (1963) The
sea vol. 2. The composition of seawater. Wiley, New York
p 26
[2] Bolin B, Degens E T, Kempe S, Ketner P (1979) The global
carbon cycle. Wiley, UK
[3] Skogen M D, Søiland H, Svendsen E (2003) J Marine Systems
46: 23
[4] Hegarty S G, Villareal T A (1998) J Exp Marine Biol Ecology
226: 241
[5] Anderson D M, Glibert P M, Burkholder J M (2002) Estuaries
25: 704
[6] Breitburg D L, Sanders J G (1999) Limnol Oceanograph
44: 837
[7] www.ospar.org
[8] Schmuhl A, Foge D, Becker G, Rasmus R, Schulz A (2001)
Sea Technol 42: 27
[9] Tovar A, Moreno C (2001) Quim Anal 20: 37
[10] Fernandes R N, Reis B F (2002) Talanta 58: 729
[11] Kerouel R, Aminot A (1997) Marine Chem 57: 265
[12] Wei F S, Wu Z X, Ten E J (1989) Anal Lett 22: 3081
[13] Petersen W, Petschatnikov M, Schroeder F, Colijn F (2003)
Ferrybox systems for monitoring coastal waters. Proceedings
of the Third International Conference on EuroGOOS, 325
[14] Wehde H, Petersen W, Petschatnikov M, Schroeder F, Colijn F
(2003) Development and distribution of plankton observed
with a FerryBox system for monitoring coastal waters. ICES
Annual Science Conference, September 24–27
[15] Ruzicka J, Hansen E H (1988) Flow injection analysis,
2nd edn. J. Wiley and Sons
Sequential Injection Analysis of Ammonia and Phosphate 37
[16] Karlberg B, Pacey G E (1989) Flow injection analysis.
A practical guide. Elsevier
[17] Wang X D, Cardwell T J, Cattrall R W, Dyson R P, Jenkins G E
(1998) Anal Chim Acta 368: 105
[18] www.flowinjection.com
[19] CAVRO Scientific Instruments, inc (1999) CAVRO XL 3000
series modular digital pump: operators manual. P=N 724043
[20] Frank C, Schroeder F, Ebinghaus R, Ruck W (2006) Using
sequential injection analysis for fast determination of phos-
phates in coastal waters. Talanta (accepted)
[21] Ebenh€ooh W, Kohlmeier C (2004) Ecological Modelling 174:
241
[22] Grasshoff K, Kleming K, Ehrhardt M (1999) Methods of
seawater analysis, edn. 3. Wiley, Chichester
[23] BranþLuebbe (2002) Auto Analyzer Applications, Method
No. G-171-96 Rev. 6
[24] Liu M, Hou L (2002) Environ Geol 42: 657
[25] Aminot A, Kerouel R (1997) Anal Chim Acta 351: 299
[26] Gardolinski P C F C, Hanrahan G (2001) Water Res 35:
3670
38 Sequential Injection Analysis of Ammonia and Phosphate