A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma David A Peake, 1 Reiko Kiyonami, 1 Yasuto Yokoi, 2 Yukihiro Fukamachi 2 and Yingying Huang 1 1 Thermo Fisher Scientific, San Jose, CA, USA; 2 Mitsui Knowledge Industry, Tokyo, Japan Poster Note 64485 Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma. Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC). Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections. Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development. Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey. 1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma. Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol. Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes. Table 1. LC gradient method utilized for sample analysis. Table 2. Mass Spectrometer settings utilized for sample analysis. Results Quantitative Results The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ). Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure. Time (min) Flow rate (uL/min) %A %B 0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95 6.0 500 5 95 6.1 500 95 5 8.0 500 95 5 HESI Source Settings Value MS Scan Settings Valu Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature ( O C) 450 Resolution 70,00 Capillary Temperature ( O C) 350 AGC Target 2.00E+ Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1 2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 Time (min) 0 200000 400000 600000 800000 0 1000 2000 3000 4000 5000 6000 7000 RT: 2.9428 RT: 2.9294 3.0178 3.1165 2.8591 3.2 3.4 3.6 3.8 4.0 Time (min) 0 200000 400000 600000 800000 1000000 0 1000 2000 3000 4000 5000 RT: 3.6080 RT: 3.5811 Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo Scientific TM Xcalibur data acquisition software. Data Analysis All data was processed utilizing Thermo Scientific TM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Upload: tranlien

Post on 04-Jun-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: A High Resolution Accurate Mass Approach for the ... · A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in ... linear response has remained

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma David A Peake,1 Reiko Kiyonami,1 Yasuto Yokoi,2 Yukihiro Fukamachi2 and Yingying Huang1

1Thermo Fisher Scienti� c, San Jose, CA, USA; 2Mitsui Knowledge Industry, Tokyo, Japan

Po

ster No

te 64

48

5

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

330.1494

329.9454329.9065 330.3360330.2271329.8745 330.2630330.0081 330.0590

Curve_1_Test_4a_112 #772 RT: 3.65 AV: 1 NL: 1.88E3T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

329.9455

329.9071

330.3363

329.8745

330.1040

330.1910

330.2039330.1495330.2629 330.2880330.2263

330.1180

330.1756

330.0599

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3103

468.2327468.1020 468.1486

Curve_1_Test_4a_134 #630 RT: 2.96 AV: 1 NL: 3.29E3T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3101

468.2329

468.1016

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma Keeley Murphy1, Jonathan Josephs1, Maciej Bromirski2, Olaf Scheibner2, 1Thermo Fisher Scientific, San Jose, CA USA; 2Thermo Fisher Scientific, Bremen, Germany

Conclusion In this experiment we described a method for the quantitation of Buprenorphine and Paroxetine. The LOQ was determined to be 10 pg/mL for each compound. The signal response across the concentration range of the assay was determined to be linear across 4 orders of magnitude, from 10 pg/mL to 100 ng/mL. Evaluation of the analytical scan speed determined that 11-15 scans were generated for chromatographic peak widths ranging between 3-4 seconds at the base. Additionally the spectrum of each analyte was examined near the target m/z to identify possible interferences. At the resolution setting of 70,000, all interferences present in the sample matrix were adequately resolved from the analyte of interest. Also the scan to scan mass accuracy at the LOQ was evaluated demonstrating a Δppm of less than +/- 5ppm for each scan across the analyte peak. These data interrogation options offer valuable method development and troubleshooting tools that provide greater confidence in the analytical method and quantitative results. Overall the Q Exactive Focus mass spectrometer was determined to be very well suited for robust and reproducible quantitation at very low concentration levels in biological matrixes.

References 1. A. Kaufmann, P. Butcher, K. Maden, S. Walker, M. Widmer, Analytica Chimica

Acta, 2010, 673, 60–72

2. Henry H. et al., Rapid Commun. Mass Spectrom. 2012, 26, 499–509

Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma.

Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC).

Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections.

Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development.

Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey.1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma.

Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol.

Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes.

Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo ScientificTM Xcalibur data acquisition software.

Data Analysis All data was processed utilizing Thermo ScientificTM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Figure 2. Calibration Curve for Buprenorphine and Paroxetine. Linear regression was performed using 1/x2 weighting.

Table 1. LC gradient method utilized for sample analysis.

Table 2. Mass Spectrometer settings utilized for sample analysis.

Results Quantitative Results

The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ).

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure.

Figure 5. Zoomed view of Buprenorphine at 10pg/mL (LEFT) and Paroxetine at 10pg/mL (Right). The figure illustrates the scans measured across the peak and the corresponding mass accuracy. Mass accuracy is displayed in parts per million. (ppm)

FIGURE 6. Replicate injections of Buprenorphine (Top) and Paroxetine (bottom) at the assay LOQ of 10pg/mL. Method integration settings where implement for all chromatogram integration and no manual integrations were performed. Additionally no smoothing was applied for any chromatographic peak generation.

A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 25,000 and 10pg/mL to 100,000pg/mL respectively (Figure 2 ).

The analytical scan speed of the instrument was evaluated at the LOQ for each sample. The number of scans across each analyte peak at the LOQ was assessed to ensure adequate scan speed in conjunction with the sample peak width produced by the chromatographic method. The chromatographic peak width for each analyte ranged between 3.6 and 4.2 seconds at the base, providing 11-15 scans for the target analytes at the LOQ (Figure 3).

.

Here we examine the scan to scan mass accuracy of the Q Exactive Focus with an external mass calibration for Buprenorphine and Paroxetine at assay LOQ of 10pg/mL (Figure 5). The mass accuracy for each scan across the analyte peaks was demonstrated to be less than 3ppm for both Buprenorphine and Paroxetine. The high level of mass accuracy from scan to scan allows the utilization of a narrow mass extraction window providing robust and reproducible results at low analyte concentration levels while in the presence of a biological matrix. Evaluation of the scan to scan mass accuracy provides a method development and troubleshooting tool that is unique to HRAM and the Orbitrap mass analyzer.

The spectra of target analytes at the LOQ can be used to verify method quality, determine the suitability of a given resolution setting, and help identify interferences that may inhibit sensitivity and reproducibility. Spectra from injections made at the LOQ for each compound were evaluated to identify unresolved interferences. At the resolution setting of 70,000 (FWHM at m/z 200) both compounds were adequately resolved from all co-eluting interferences (Figure 4). As seen in the figure, although interferences are present during the elution of each compound, the target analyte m/z was baseline resolved from nearby interferences at the 70,000 resolution setting. Additionaly the mass extraction window of +/- 5ppm used to generate the chromatographic peaks used for quantitation is illustrated in the figures. This narrow extraction window allows greater assay confidence and robustness by ensuring that signal contributing to the chromatographic peak area is resulting from the target analyte itself and not the nearby interferences. A narrow mass extraction windows and the resulting XIC do not provide analytical value without a high level of scan to scan mass accuracy across the entire analyte peak. Scan to scan mass accuracy can also offer an important indicator for indentifying potential analyte interferences as well for accessing overall assay performance. A loss in mass accuracy can provide an indication of an unresolved or co eluting interference and can be implemented as a quality control measure in validated analyses.

Figure 3. Evaluation of chromatographic peak width vs scan speed for Buprenorphine (left) and Paroxetine (right) the LOQ of 10pg/mL

Time (min)

Flow rate (uL/min) %A %B

0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95

6.0 500 5 95 6.1 500 95 5 8.0 500 95 5

HESI Source Settings Value MS Scan Settings Value

Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature (OC) 450 Resolution 70,000 Capillary Temperature (OC) 350 AGC Target 2.00E+04 Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1

RT: 2.5471 - 4.0071

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0Time (min)

0

200000

400000

600000

800000

0

1000

2000

3000

4000

5000

6000

7000

Rel

ativ

e Ab

unda

nce

RT: 2.9428

RT: 2.9294

3.0178 3.11652.8591

NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

NL:9.32E5m/z= 472.3335-472.3383 MS Curve_1_Test_4a_130

RT: 3.1978 - 4.8231

3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8Time (min)

0

200000

400000

600000

800000

1000000

0

1000

2000

3000

4000

5000

Rel

ativ

e Ab

unda

nce

RT: 3.6080

RT: 3.5811

NL:5.46E3m/z= 330.1483-330.1517 MS ICIS Curve_1_Test_4a_113

NL:1.13E6m/z= 336.1860-336.1894 MS ICIS Curve_1_Test_4a_113

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 5000 10000 15000 20000 25000pg/mL

0

5

10

15

20

Area

Rat

io

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 20 40 60 80 100pg/mL

0.00

0.02

0.04

0.06

0.08

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

0 20000 40000 60000 80000 100000pg/mL

0

10

20

30

40

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

20 40 60 80 100pg/mL

0.00

0.01

0.02

0.03

0.04

0.05

Area

Rat

io

RT: 2.7911 - 3.1104

2.80 2.82 2.84 2.86 2.88 2.90 2.92 2.94 2.96 2.98 3.00 3.02 3.04 3.06 3.08 3.10Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Relat

ive Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

RT: 3.4422 - 3.7632

3.46 3.48 3.50 3.52 3.54 3.56 3.58 3.60 3.62 3.64 3.66 3.68 3.70 3.72 3.74 3.76Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Relat

ive Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Buprenorphine Peak Width = 4.2 sec at base Scans Across Peak = 15

Paroxetine Peak Width = 4.8 sec at base Scans Across Peak = 18

The number of scans recorded across the width of each analyte peak correlates with the instrument specification of an analytical scan speed of 3Hz at the 70,000 resolution setting. Analysis of the prepared sample set at each concentration level demonstrated results that met or exceeded all of the previously described acceptance criteria indicating a high level of performance across the working range of the assay (Table 3)

Buprenorphine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 3.2% 4.4% 0 25 23.7 5.1% 1.9% 0 50 47.7 4.6% 1.5% 0

100 98.9 1.1% 1.5% 0 250 242 3.4% 1.0% 0 500 497 0.6% 0.6% 0

1000 1011 1.1% 0.7% 0 2500 2501 0.05% 0.8% 0 5000 5159 3.2% 1.0% 0

10,000 10413 4.1% 0.8% 0 25,000 25794 3.2% 0.9% 0

Paroxetine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 2.7% 8.6% 0 25 23.7 5.3% 1.8% 0 50 48.4 3.3% 1.2% 0

100 101 1.3% 1.5% 0 250 245 2.1% 1.0% 0 500 507 1.4% 0.6% 0

1000 1030 3.0% 0.5% 0 2500 2474 1.0% 1.2% 0 5000 5122 2.4% 0.7% 0

10,000 10286 2.9% 0.4% 0 25,000 24702 1.2% 0.6% 0 50,000 49916 0.2% 1.2% 0

100,000 99385 0.6% 1.2% 0

Table 3. Statistical Results for Buprenorphine and Paroxetine Replicates. Note: AVG %Diff represents the average difference from the linear regression fit of the calibration curve for each set of replicates at a given concentration point.

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

Rel

ativ

e Ab

unda

nce

468.2327

468.1020

468.1486

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

Rel

ativ

e Ab

unda

nce 329.9454329.9065

330.3360

330.2271329.8745

330.2630

330.0081330.2051

330.0590

Buprenorphine – Spectral Analysis Paroxetine – Spectral Analysis

RT: 2.8875 - 3.0177

2.89 2.90 2.91 2.92 2.93 2.94 2.95 2.96 2.97 2.98 2.99 3.00 3.01Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Rel

ativ

e Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

Δppm-2.1

Δppm-1.5

Δppm-1.8

Δppm-2.7

Δppm-1.8

Δppm-2.4 Δppm-1.5

Δppm-1.5

Δppm-1.8

Δppm-1.8

Δppm-1.5

Δppm-2.7

Δppm-0.9

Δppm-2.1

Δppm-1.8

Δppm-2.1 Δppm-0.6

RT: 3.5430 - 3.6824

3.55 3.56 3.57 3.58 3.59 3.60 3.61 3.62 3.63 3.64 3.65 3.66 3.67 3.68Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Rel

ativ

e Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Δppm-2.7

Δppm-2.7

Δppm-1.8

Δppm-2.1

Δppm-1.8

Δppm-1.5

Δppm-0.9

Δppm -1.5

Δppm-1.8 Δppm-1.5

Δppm-1.5

Δppm-2.4 Δppm-2.1

Δppm-1.8

Δppm-2.1

Δppm-0.6

Δppm-1.8

Δppm-0.8 Δppm-1.8

Figure 4. Spectra analysis of Buprenorphine and Paroxetine at 10pg/mL and 1ng/mL in rat plasma. The exact m/z for the target analyte has been underlined in each spectra and the 5ppm mass extraction window used to generate chromatographic peaks has been outlined. Common interferences across each sample concentration have been circled. All interferences present in the sample were demonstrated to be baseline resolved at the 70,000 resolution setting.

c:\xcalibur\...\curve_1_test_4a_117 03/06/15 15:47:02 Compound Mix 0.010ng/mL

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6083 NL:5.16E3m/z= 330.14835-330.15165 MS ICIS Curve_1_Test_4a_112

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6080 NL:5.46E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_113

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Rela

tive

Abun

danc

e

RT: 3.6000 NL:4.65E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_114

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6143 NL:5.33E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_115

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6224 NL:5.89E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_116

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

NL:6.73E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_117

c:\xcalibur\...\curve_1_test_4a_134 03/06/15 18:19:37 Compound Mix 0.100ng/mL

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9428 NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9450 NL:7.69E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_131

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9631 NL:6.88E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_132

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9446 NL:7.27E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_133

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9355 NL:6.57E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_134

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9356 NL:7.89E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_135

Chromatographic reproducibility is an import factor for determining assay quality and ruggedness. Consistent and reproducible signal response at the assay LOQ eliminates the need for manual chromatographic peak integration and improves the overall data confidence and the assay as a whole. Here we examine the chromatographic reproducibility of Buprenorphine and Paroxetine at the assay LOQ of 10pg/mL. Each analyte was analyzed in replicates of n=6. The chromatogram for each replicate can be seen in figure 6.

Buprenorphine – 10pg/mL

Buprenorphine – 1ng/mL

Buprenorphine – 1ng/mL Zoomed

Paroxetine– 10pg/mL

Paroxetine– 1ng/mL

Paroxetine– 1ng/mL Zoomed

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

330.1494

329.9454329.9065 330.3360330.2271329.8745 330.2630330.0081 330.0590

Curve_1_Test_4a_112 #772 RT: 3.65 AV: 1 NL: 1.88E3T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

329.9455

329.9071

330.3363

329.8745

330.1040

330.1910

330.2039330.1495330.2629 330.2880330.2263

330.1180

330.1756

330.0599

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3103

468.2327468.1020 468.1486

Curve_1_Test_4a_134 #630 RT: 2.96 AV: 1 NL: 3.29E3T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3101

468.2329

468.1016

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma Keeley Murphy1, Jonathan Josephs1, Maciej Bromirski2, Olaf Scheibner2, 1Thermo Fisher Scientific, San Jose, CA USA; 2Thermo Fisher Scientific, Bremen, Germany

Conclusion In this experiment we described a method for the quantitation of Buprenorphine and Paroxetine. The LOQ was determined to be 10 pg/mL for each compound. The signal response across the concentration range of the assay was determined to be linear across 4 orders of magnitude, from 10 pg/mL to 100 ng/mL. Evaluation of the analytical scan speed determined that 11-15 scans were generated for chromatographic peak widths ranging between 3-4 seconds at the base. Additionally the spectrum of each analyte was examined near the target m/z to identify possible interferences. At the resolution setting of 70,000, all interferences present in the sample matrix were adequately resolved from the analyte of interest. Also the scan to scan mass accuracy at the LOQ was evaluated demonstrating a Δppm of less than +/- 5ppm for each scan across the analyte peak. These data interrogation options offer valuable method development and troubleshooting tools that provide greater confidence in the analytical method and quantitative results. Overall the Q Exactive Focus mass spectrometer was determined to be very well suited for robust and reproducible quantitation at very low concentration levels in biological matrixes.

References 1. A. Kaufmann, P. Butcher, K. Maden, S. Walker, M. Widmer, Analytica Chimica

Acta, 2010, 673, 60–72

2. Henry H. et al., Rapid Commun. Mass Spectrom. 2012, 26, 499–509

Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma.

Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC).

Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections.

Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development.

Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey.1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma.

Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol.

Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes.

Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo ScientificTM Xcalibur data acquisition software.

Data Analysis All data was processed utilizing Thermo ScientificTM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Figure 2. Calibration Curve for Buprenorphine and Paroxetine. Linear regression was performed using 1/x2 weighting.

Table 1. LC gradient method utilized for sample analysis.

Table 2. Mass Spectrometer settings utilized for sample analysis.

Results Quantitative Results

The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ).

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure.

Figure 5. Zoomed view of Buprenorphine at 10pg/mL (LEFT) and Paroxetine at 10pg/mL (Right). The figure illustrates the scans measured across the peak and the corresponding mass accuracy. Mass accuracy is displayed in parts per million. (ppm)

FIGURE 6. Replicate injections of Buprenorphine (Top) and Paroxetine (bottom) at the assay LOQ of 10pg/mL. Method integration settings where implement for all chromatogram integration and no manual integrations were performed. Additionally no smoothing was applied for any chromatographic peak generation.

A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 25,000 and 10pg/mL to 100,000pg/mL respectively (Figure 2 ).

The analytical scan speed of the instrument was evaluated at the LOQ for each sample. The number of scans across each analyte peak at the LOQ was assessed to ensure adequate scan speed in conjunction with the sample peak width produced by the chromatographic method. The chromatographic peak width for each analyte ranged between 3.6 and 4.2 seconds at the base, providing 11-15 scans for the target analytes at the LOQ (Figure 3).

.

Here we examine the scan to scan mass accuracy of the Q Exactive Focus with an external mass calibration for Buprenorphine and Paroxetine at assay LOQ of 10pg/mL (Figure 5). The mass accuracy for each scan across the analyte peaks was demonstrated to be less than 3ppm for both Buprenorphine and Paroxetine. The high level of mass accuracy from scan to scan allows the utilization of a narrow mass extraction window providing robust and reproducible results at low analyte concentration levels while in the presence of a biological matrix. Evaluation of the scan to scan mass accuracy provides a method development and troubleshooting tool that is unique to HRAM and the Orbitrap mass analyzer.

The spectra of target analytes at the LOQ can be used to verify method quality, determine the suitability of a given resolution setting, and help identify interferences that may inhibit sensitivity and reproducibility. Spectra from injections made at the LOQ for each compound were evaluated to identify unresolved interferences. At the resolution setting of 70,000 (FWHM at m/z 200) both compounds were adequately resolved from all co-eluting interferences (Figure 4). As seen in the figure, although interferences are present during the elution of each compound, the target analyte m/z was baseline resolved from nearby interferences at the 70,000 resolution setting. Additionaly the mass extraction window of +/- 5ppm used to generate the chromatographic peaks used for quantitation is illustrated in the figures. This narrow extraction window allows greater assay confidence and robustness by ensuring that signal contributing to the chromatographic peak area is resulting from the target analyte itself and not the nearby interferences. A narrow mass extraction windows and the resulting XIC do not provide analytical value without a high level of scan to scan mass accuracy across the entire analyte peak. Scan to scan mass accuracy can also offer an important indicator for indentifying potential analyte interferences as well for accessing overall assay performance. A loss in mass accuracy can provide an indication of an unresolved or co eluting interference and can be implemented as a quality control measure in validated analyses.

Figure 3. Evaluation of chromatographic peak width vs scan speed for Buprenorphine (left) and Paroxetine (right) the LOQ of 10pg/mL

Time (min)

Flow rate (uL/min) %A %B

0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95

6.0 500 5 95 6.1 500 95 5 8.0 500 95 5

HESI Source Settings Value MS Scan Settings Value

Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature (OC) 450 Resolution 70,000 Capillary Temperature (OC) 350 AGC Target 2.00E+04 Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1

RT: 2.5471 - 4.0071

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0Time (min)

0

200000

400000

600000

800000

0

1000

2000

3000

4000

5000

6000

7000

Rel

ativ

e Ab

unda

nce

RT: 2.9428

RT: 2.9294

3.0178 3.11652.8591

NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

NL:9.32E5m/z= 472.3335-472.3383 MS Curve_1_Test_4a_130

RT: 3.1978 - 4.8231

3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8Time (min)

0

200000

400000

600000

800000

1000000

0

1000

2000

3000

4000

5000

Rel

ativ

e Ab

unda

nce

RT: 3.6080

RT: 3.5811

NL:5.46E3m/z= 330.1483-330.1517 MS ICIS Curve_1_Test_4a_113

NL:1.13E6m/z= 336.1860-336.1894 MS ICIS Curve_1_Test_4a_113

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 5000 10000 15000 20000 25000pg/mL

0

5

10

15

20

Area

Rat

io

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 20 40 60 80 100pg/mL

0.00

0.02

0.04

0.06

0.08

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

0 20000 40000 60000 80000 100000pg/mL

0

10

20

30

40

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

20 40 60 80 100pg/mL

0.00

0.01

0.02

0.03

0.04

0.05

Area

Rat

io

RT: 2.7911 - 3.1104

2.80 2.82 2.84 2.86 2.88 2.90 2.92 2.94 2.96 2.98 3.00 3.02 3.04 3.06 3.08 3.10Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Relat

ive Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

RT: 3.4422 - 3.7632

3.46 3.48 3.50 3.52 3.54 3.56 3.58 3.60 3.62 3.64 3.66 3.68 3.70 3.72 3.74 3.76Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Relat

ive Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Buprenorphine Peak Width = 4.2 sec at base Scans Across Peak = 15

Paroxetine Peak Width = 4.8 sec at base Scans Across Peak = 18

The number of scans recorded across the width of each analyte peak correlates with the instrument specification of an analytical scan speed of 3Hz at the 70,000 resolution setting. Analysis of the prepared sample set at each concentration level demonstrated results that met or exceeded all of the previously described acceptance criteria indicating a high level of performance across the working range of the assay (Table 3)

Buprenorphine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 3.2% 4.4% 0 25 23.7 5.1% 1.9% 0 50 47.7 4.6% 1.5% 0

100 98.9 1.1% 1.5% 0 250 242 3.4% 1.0% 0 500 497 0.6% 0.6% 0

1000 1011 1.1% 0.7% 0 2500 2501 0.05% 0.8% 0 5000 5159 3.2% 1.0% 0

10,000 10413 4.1% 0.8% 0 25,000 25794 3.2% 0.9% 0

Paroxetine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 2.7% 8.6% 0 25 23.7 5.3% 1.8% 0 50 48.4 3.3% 1.2% 0

100 101 1.3% 1.5% 0 250 245 2.1% 1.0% 0 500 507 1.4% 0.6% 0

1000 1030 3.0% 0.5% 0 2500 2474 1.0% 1.2% 0 5000 5122 2.4% 0.7% 0

10,000 10286 2.9% 0.4% 0 25,000 24702 1.2% 0.6% 0 50,000 49916 0.2% 1.2% 0

100,000 99385 0.6% 1.2% 0

Table 3. Statistical Results for Buprenorphine and Paroxetine Replicates. Note: AVG %Diff represents the average difference from the linear regression fit of the calibration curve for each set of replicates at a given concentration point.

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

Rel

ativ

e Ab

unda

nce

468.2327

468.1020

468.1486

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

Rel

ativ

e Ab

unda

nce 329.9454329.9065

330.3360

330.2271329.8745

330.2630

330.0081330.2051

330.0590

Buprenorphine – Spectral Analysis Paroxetine – Spectral Analysis

RT: 2.8875 - 3.0177

2.89 2.90 2.91 2.92 2.93 2.94 2.95 2.96 2.97 2.98 2.99 3.00 3.01Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Rel

ativ

e Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

Δppm-2.1

Δppm-1.5

Δppm-1.8

Δppm-2.7

Δppm-1.8

Δppm-2.4 Δppm-1.5

Δppm-1.5

Δppm-1.8

Δppm-1.8

Δppm-1.5

Δppm-2.7

Δppm-0.9

Δppm-2.1

Δppm-1.8

Δppm-2.1 Δppm-0.6

RT: 3.5430 - 3.6824

3.55 3.56 3.57 3.58 3.59 3.60 3.61 3.62 3.63 3.64 3.65 3.66 3.67 3.68Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Rel

ativ

e Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Δppm-2.7

Δppm-2.7

Δppm-1.8

Δppm-2.1

Δppm-1.8

Δppm-1.5

Δppm-0.9

Δppm -1.5

Δppm-1.8 Δppm-1.5

Δppm-1.5

Δppm-2.4 Δppm-2.1

Δppm-1.8

Δppm-2.1

Δppm-0.6

Δppm-1.8

Δppm-0.8 Δppm-1.8

Figure 4. Spectra analysis of Buprenorphine and Paroxetine at 10pg/mL and 1ng/mL in rat plasma. The exact m/z for the target analyte has been underlined in each spectra and the 5ppm mass extraction window used to generate chromatographic peaks has been outlined. Common interferences across each sample concentration have been circled. All interferences present in the sample were demonstrated to be baseline resolved at the 70,000 resolution setting.

c:\xcalibur\...\curve_1_test_4a_117 03/06/15 15:47:02 Compound Mix 0.010ng/mL

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6083 NL:5.16E3m/z= 330.14835-330.15165 MS ICIS Curve_1_Test_4a_112

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6080 NL:5.46E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_113

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Rela

tive

Abun

danc

e

RT: 3.6000 NL:4.65E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_114

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6143 NL:5.33E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_115

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6224 NL:5.89E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_116

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

NL:6.73E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_117

c:\xcalibur\...\curve_1_test_4a_134 03/06/15 18:19:37 Compound Mix 0.100ng/mL

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9428 NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9450 NL:7.69E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_131

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9631 NL:6.88E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_132

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9446 NL:7.27E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_133

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9355 NL:6.57E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_134

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9356 NL:7.89E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_135

Chromatographic reproducibility is an import factor for determining assay quality and ruggedness. Consistent and reproducible signal response at the assay LOQ eliminates the need for manual chromatographic peak integration and improves the overall data confidence and the assay as a whole. Here we examine the chromatographic reproducibility of Buprenorphine and Paroxetine at the assay LOQ of 10pg/mL. Each analyte was analyzed in replicates of n=6. The chromatogram for each replicate can be seen in figure 6.

Buprenorphine – 10pg/mL

Buprenorphine – 1ng/mL

Buprenorphine – 1ng/mL Zoomed

Paroxetine– 10pg/mL

Paroxetine– 1ng/mL

Paroxetine– 1ng/mL Zoomed

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

330.1494

329.9454329.9065 330.3360330.2271329.8745 330.2630330.0081 330.0590

Curve_1_Test_4a_112 #772 RT: 3.65 AV: 1 NL: 1.88E3T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

329.9455

329.9071

330.3363

329.8745

330.1040

330.1910

330.2039330.1495330.2629 330.2880330.2263

330.1180

330.1756

330.0599

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3103

468.2327468.1020 468.1486

Curve_1_Test_4a_134 #630 RT: 2.96 AV: 1 NL: 3.29E3T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3101

468.2329

468.1016

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma Keeley Murphy1, Jonathan Josephs1, Maciej Bromirski2, Olaf Scheibner2, 1Thermo Fisher Scientific, San Jose, CA USA; 2Thermo Fisher Scientific, Bremen, Germany

Conclusion In this experiment we described a method for the quantitation of Buprenorphine and Paroxetine. The LOQ was determined to be 10 pg/mL for each compound. The signal response across the concentration range of the assay was determined to be linear across 4 orders of magnitude, from 10 pg/mL to 100 ng/mL. Evaluation of the analytical scan speed determined that 11-15 scans were generated for chromatographic peak widths ranging between 3-4 seconds at the base. Additionally the spectrum of each analyte was examined near the target m/z to identify possible interferences. At the resolution setting of 70,000, all interferences present in the sample matrix were adequately resolved from the analyte of interest. Also the scan to scan mass accuracy at the LOQ was evaluated demonstrating a Δppm of less than +/- 5ppm for each scan across the analyte peak. These data interrogation options offer valuable method development and troubleshooting tools that provide greater confidence in the analytical method and quantitative results. Overall the Q Exactive Focus mass spectrometer was determined to be very well suited for robust and reproducible quantitation at very low concentration levels in biological matrixes.

References 1. A. Kaufmann, P. Butcher, K. Maden, S. Walker, M. Widmer, Analytica Chimica

Acta, 2010, 673, 60–72

2. Henry H. et al., Rapid Commun. Mass Spectrom. 2012, 26, 499–509

Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma.

Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC).

Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections.

Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development.

Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey.1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma.

Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol.

Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes.

Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo ScientificTM Xcalibur data acquisition software.

Data Analysis All data was processed utilizing Thermo ScientificTM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Figure 2. Calibration Curve for Buprenorphine and Paroxetine. Linear regression was performed using 1/x2 weighting.

Table 1. LC gradient method utilized for sample analysis.

Table 2. Mass Spectrometer settings utilized for sample analysis.

Results Quantitative Results

The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ).

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure.

Figure 5. Zoomed view of Buprenorphine at 10pg/mL (LEFT) and Paroxetine at 10pg/mL (Right). The figure illustrates the scans measured across the peak and the corresponding mass accuracy. Mass accuracy is displayed in parts per million. (ppm)

FIGURE 6. Replicate injections of Buprenorphine (Top) and Paroxetine (bottom) at the assay LOQ of 10pg/mL. Method integration settings where implement for all chromatogram integration and no manual integrations were performed. Additionally no smoothing was applied for any chromatographic peak generation.

A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 25,000 and 10pg/mL to 100,000pg/mL respectively (Figure 2 ).

The analytical scan speed of the instrument was evaluated at the LOQ for each sample. The number of scans across each analyte peak at the LOQ was assessed to ensure adequate scan speed in conjunction with the sample peak width produced by the chromatographic method. The chromatographic peak width for each analyte ranged between 3.6 and 4.2 seconds at the base, providing 11-15 scans for the target analytes at the LOQ (Figure 3).

.

Here we examine the scan to scan mass accuracy of the Q Exactive Focus with an external mass calibration for Buprenorphine and Paroxetine at assay LOQ of 10pg/mL (Figure 5). The mass accuracy for each scan across the analyte peaks was demonstrated to be less than 3ppm for both Buprenorphine and Paroxetine. The high level of mass accuracy from scan to scan allows the utilization of a narrow mass extraction window providing robust and reproducible results at low analyte concentration levels while in the presence of a biological matrix. Evaluation of the scan to scan mass accuracy provides a method development and troubleshooting tool that is unique to HRAM and the Orbitrap mass analyzer.

The spectra of target analytes at the LOQ can be used to verify method quality, determine the suitability of a given resolution setting, and help identify interferences that may inhibit sensitivity and reproducibility. Spectra from injections made at the LOQ for each compound were evaluated to identify unresolved interferences. At the resolution setting of 70,000 (FWHM at m/z 200) both compounds were adequately resolved from all co-eluting interferences (Figure 4). As seen in the figure, although interferences are present during the elution of each compound, the target analyte m/z was baseline resolved from nearby interferences at the 70,000 resolution setting. Additionaly the mass extraction window of +/- 5ppm used to generate the chromatographic peaks used for quantitation is illustrated in the figures. This narrow extraction window allows greater assay confidence and robustness by ensuring that signal contributing to the chromatographic peak area is resulting from the target analyte itself and not the nearby interferences. A narrow mass extraction windows and the resulting XIC do not provide analytical value without a high level of scan to scan mass accuracy across the entire analyte peak. Scan to scan mass accuracy can also offer an important indicator for indentifying potential analyte interferences as well for accessing overall assay performance. A loss in mass accuracy can provide an indication of an unresolved or co eluting interference and can be implemented as a quality control measure in validated analyses.

Figure 3. Evaluation of chromatographic peak width vs scan speed for Buprenorphine (left) and Paroxetine (right) the LOQ of 10pg/mL

Time (min)

Flow rate (uL/min) %A %B

0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95

6.0 500 5 95 6.1 500 95 5 8.0 500 95 5

HESI Source Settings Value MS Scan Settings Value

Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature (OC) 450 Resolution 70,000 Capillary Temperature (OC) 350 AGC Target 2.00E+04 Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1

RT: 2.5471 - 4.0071

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0Time (min)

0

200000

400000

600000

800000

0

1000

2000

3000

4000

5000

6000

7000

Rel

ativ

e Ab

unda

nce

RT: 2.9428

RT: 2.9294

3.0178 3.11652.8591

NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

NL:9.32E5m/z= 472.3335-472.3383 MS Curve_1_Test_4a_130

RT: 3.1978 - 4.8231

3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8Time (min)

0

200000

400000

600000

800000

1000000

0

1000

2000

3000

4000

5000

Rel

ativ

e Ab

unda

nce

RT: 3.6080

RT: 3.5811

NL:5.46E3m/z= 330.1483-330.1517 MS ICIS Curve_1_Test_4a_113

NL:1.13E6m/z= 336.1860-336.1894 MS ICIS Curve_1_Test_4a_113

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 5000 10000 15000 20000 25000pg/mL

0

5

10

15

20

Area

Rat

io

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 20 40 60 80 100pg/mL

0.00

0.02

0.04

0.06

0.08

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

0 20000 40000 60000 80000 100000pg/mL

0

10

20

30

40

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

20 40 60 80 100pg/mL

0.00

0.01

0.02

0.03

0.04

0.05

Area

Rat

io

RT: 2.7911 - 3.1104

2.80 2.82 2.84 2.86 2.88 2.90 2.92 2.94 2.96 2.98 3.00 3.02 3.04 3.06 3.08 3.10Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Relat

ive Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

RT: 3.4422 - 3.7632

3.46 3.48 3.50 3.52 3.54 3.56 3.58 3.60 3.62 3.64 3.66 3.68 3.70 3.72 3.74 3.76Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Relat

ive Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Buprenorphine Peak Width = 4.2 sec at base Scans Across Peak = 15

Paroxetine Peak Width = 4.8 sec at base Scans Across Peak = 18

The number of scans recorded across the width of each analyte peak correlates with the instrument specification of an analytical scan speed of 3Hz at the 70,000 resolution setting. Analysis of the prepared sample set at each concentration level demonstrated results that met or exceeded all of the previously described acceptance criteria indicating a high level of performance across the working range of the assay (Table 3)

Buprenorphine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 3.2% 4.4% 0 25 23.7 5.1% 1.9% 0 50 47.7 4.6% 1.5% 0

100 98.9 1.1% 1.5% 0 250 242 3.4% 1.0% 0 500 497 0.6% 0.6% 0

1000 1011 1.1% 0.7% 0 2500 2501 0.05% 0.8% 0 5000 5159 3.2% 1.0% 0

10,000 10413 4.1% 0.8% 0 25,000 25794 3.2% 0.9% 0

Paroxetine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 2.7% 8.6% 0 25 23.7 5.3% 1.8% 0 50 48.4 3.3% 1.2% 0

100 101 1.3% 1.5% 0 250 245 2.1% 1.0% 0 500 507 1.4% 0.6% 0

1000 1030 3.0% 0.5% 0 2500 2474 1.0% 1.2% 0 5000 5122 2.4% 0.7% 0

10,000 10286 2.9% 0.4% 0 25,000 24702 1.2% 0.6% 0 50,000 49916 0.2% 1.2% 0

100,000 99385 0.6% 1.2% 0

Table 3. Statistical Results for Buprenorphine and Paroxetine Replicates. Note: AVG %Diff represents the average difference from the linear regression fit of the calibration curve for each set of replicates at a given concentration point.

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

Rel

ativ

e Ab

unda

nce

468.2327

468.1020

468.1486

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

Rel

ativ

e Ab

unda

nce 329.9454329.9065

330.3360

330.2271329.8745

330.2630

330.0081330.2051

330.0590

Buprenorphine – Spectral Analysis Paroxetine – Spectral Analysis

RT: 2.8875 - 3.0177

2.89 2.90 2.91 2.92 2.93 2.94 2.95 2.96 2.97 2.98 2.99 3.00 3.01Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Rel

ativ

e Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

Δppm-2.1

Δppm-1.5

Δppm-1.8

Δppm-2.7

Δppm-1.8

Δppm-2.4 Δppm-1.5

Δppm-1.5

Δppm-1.8

Δppm-1.8

Δppm-1.5

Δppm-2.7

Δppm-0.9

Δppm-2.1

Δppm-1.8

Δppm-2.1 Δppm-0.6

RT: 3.5430 - 3.6824

3.55 3.56 3.57 3.58 3.59 3.60 3.61 3.62 3.63 3.64 3.65 3.66 3.67 3.68Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Rel

ativ

e Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Δppm-2.7

Δppm-2.7

Δppm-1.8

Δppm-2.1

Δppm-1.8

Δppm-1.5

Δppm-0.9

Δppm -1.5

Δppm-1.8 Δppm-1.5

Δppm-1.5

Δppm-2.4 Δppm-2.1

Δppm-1.8

Δppm-2.1

Δppm-0.6

Δppm-1.8

Δppm-0.8 Δppm-1.8

Figure 4. Spectra analysis of Buprenorphine and Paroxetine at 10pg/mL and 1ng/mL in rat plasma. The exact m/z for the target analyte has been underlined in each spectra and the 5ppm mass extraction window used to generate chromatographic peaks has been outlined. Common interferences across each sample concentration have been circled. All interferences present in the sample were demonstrated to be baseline resolved at the 70,000 resolution setting.

c:\xcalibur\...\curve_1_test_4a_117 03/06/15 15:47:02 Compound Mix 0.010ng/mL

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6083 NL:5.16E3m/z= 330.14835-330.15165 MS ICIS Curve_1_Test_4a_112

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6080 NL:5.46E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_113

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Rela

tive

Abun

danc

e

RT: 3.6000 NL:4.65E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_114

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6143 NL:5.33E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_115

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6224 NL:5.89E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_116

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

NL:6.73E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_117

c:\xcalibur\...\curve_1_test_4a_134 03/06/15 18:19:37 Compound Mix 0.100ng/mL

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9428 NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9450 NL:7.69E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_131

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9631 NL:6.88E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_132

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9446 NL:7.27E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_133

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9355 NL:6.57E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_134

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9356 NL:7.89E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_135

Chromatographic reproducibility is an import factor for determining assay quality and ruggedness. Consistent and reproducible signal response at the assay LOQ eliminates the need for manual chromatographic peak integration and improves the overall data confidence and the assay as a whole. Here we examine the chromatographic reproducibility of Buprenorphine and Paroxetine at the assay LOQ of 10pg/mL. Each analyte was analyzed in replicates of n=6. The chromatogram for each replicate can be seen in figure 6.

Buprenorphine – 10pg/mL

Buprenorphine – 1ng/mL

Buprenorphine – 1ng/mL Zoomed

Paroxetine– 10pg/mL

Paroxetine– 1ng/mL

Paroxetine– 1ng/mL Zoomed

Page 2: A High Resolution Accurate Mass Approach for the ... · A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in ... linear response has remained

2 A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

330.1494

329.9454329.9065 330.3360330.2271329.8745 330.2630330.0081 330.0590

Curve_1_Test_4a_112 #772 RT: 3.65 AV: 1 NL: 1.88E3T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

329.9455

329.9071

330.3363

329.8745

330.1040

330.1910

330.2039330.1495330.2629 330.2880330.2263

330.1180

330.1756

330.0599

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3103

468.2327468.1020 468.1486

Curve_1_Test_4a_134 #630 RT: 2.96 AV: 1 NL: 3.29E3T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3101

468.2329

468.1016

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma Keeley Murphy1, Jonathan Josephs1, Maciej Bromirski2, Olaf Scheibner2, 1Thermo Fisher Scientific, San Jose, CA USA; 2Thermo Fisher Scientific, Bremen, Germany

Conclusion In this experiment we described a method for the quantitation of Buprenorphine and Paroxetine. The LOQ was determined to be 10 pg/mL for each compound. The signal response across the concentration range of the assay was determined to be linear across 4 orders of magnitude, from 10 pg/mL to 100 ng/mL. Evaluation of the analytical scan speed determined that 11-15 scans were generated for chromatographic peak widths ranging between 3-4 seconds at the base. Additionally the spectrum of each analyte was examined near the target m/z to identify possible interferences. At the resolution setting of 70,000, all interferences present in the sample matrix were adequately resolved from the analyte of interest. Also the scan to scan mass accuracy at the LOQ was evaluated demonstrating a Δppm of less than +/- 5ppm for each scan across the analyte peak. These data interrogation options offer valuable method development and troubleshooting tools that provide greater confidence in the analytical method and quantitative results. Overall the Q Exactive Focus mass spectrometer was determined to be very well suited for robust and reproducible quantitation at very low concentration levels in biological matrixes.

References 1. A. Kaufmann, P. Butcher, K. Maden, S. Walker, M. Widmer, Analytica Chimica

Acta, 2010, 673, 60–72

2. Henry H. et al., Rapid Commun. Mass Spectrom. 2012, 26, 499–509

Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma.

Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC).

Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections.

Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development.

Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey.1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma.

Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol.

Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes.

Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo ScientificTM Xcalibur data acquisition software.

Data Analysis All data was processed utilizing Thermo ScientificTM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Figure 2. Calibration Curve for Buprenorphine and Paroxetine. Linear regression was performed using 1/x2 weighting.

Table 1. LC gradient method utilized for sample analysis.

Table 2. Mass Spectrometer settings utilized for sample analysis.

Results Quantitative Results

The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ).

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure.

Figure 5. Zoomed view of Buprenorphine at 10pg/mL (LEFT) and Paroxetine at 10pg/mL (Right). The figure illustrates the scans measured across the peak and the corresponding mass accuracy. Mass accuracy is displayed in parts per million. (ppm)

FIGURE 6. Replicate injections of Buprenorphine (Top) and Paroxetine (bottom) at the assay LOQ of 10pg/mL. Method integration settings where implement for all chromatogram integration and no manual integrations were performed. Additionally no smoothing was applied for any chromatographic peak generation.

A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 25,000 and 10pg/mL to 100,000pg/mL respectively (Figure 2 ).

The analytical scan speed of the instrument was evaluated at the LOQ for each sample. The number of scans across each analyte peak at the LOQ was assessed to ensure adequate scan speed in conjunction with the sample peak width produced by the chromatographic method. The chromatographic peak width for each analyte ranged between 3.6 and 4.2 seconds at the base, providing 11-15 scans for the target analytes at the LOQ (Figure 3).

.

Here we examine the scan to scan mass accuracy of the Q Exactive Focus with an external mass calibration for Buprenorphine and Paroxetine at assay LOQ of 10pg/mL (Figure 5). The mass accuracy for each scan across the analyte peaks was demonstrated to be less than 3ppm for both Buprenorphine and Paroxetine. The high level of mass accuracy from scan to scan allows the utilization of a narrow mass extraction window providing robust and reproducible results at low analyte concentration levels while in the presence of a biological matrix. Evaluation of the scan to scan mass accuracy provides a method development and troubleshooting tool that is unique to HRAM and the Orbitrap mass analyzer.

The spectra of target analytes at the LOQ can be used to verify method quality, determine the suitability of a given resolution setting, and help identify interferences that may inhibit sensitivity and reproducibility. Spectra from injections made at the LOQ for each compound were evaluated to identify unresolved interferences. At the resolution setting of 70,000 (FWHM at m/z 200) both compounds were adequately resolved from all co-eluting interferences (Figure 4). As seen in the figure, although interferences are present during the elution of each compound, the target analyte m/z was baseline resolved from nearby interferences at the 70,000 resolution setting. Additionaly the mass extraction window of +/- 5ppm used to generate the chromatographic peaks used for quantitation is illustrated in the figures. This narrow extraction window allows greater assay confidence and robustness by ensuring that signal contributing to the chromatographic peak area is resulting from the target analyte itself and not the nearby interferences. A narrow mass extraction windows and the resulting XIC do not provide analytical value without a high level of scan to scan mass accuracy across the entire analyte peak. Scan to scan mass accuracy can also offer an important indicator for indentifying potential analyte interferences as well for accessing overall assay performance. A loss in mass accuracy can provide an indication of an unresolved or co eluting interference and can be implemented as a quality control measure in validated analyses.

Figure 3. Evaluation of chromatographic peak width vs scan speed for Buprenorphine (left) and Paroxetine (right) the LOQ of 10pg/mL

Time (min)

Flow rate (uL/min) %A %B

0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95

6.0 500 5 95 6.1 500 95 5 8.0 500 95 5

HESI Source Settings Value MS Scan Settings Value

Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature (OC) 450 Resolution 70,000 Capillary Temperature (OC) 350 AGC Target 2.00E+04 Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1

RT: 2.5471 - 4.0071

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0Time (min)

0

200000

400000

600000

800000

0

1000

2000

3000

4000

5000

6000

7000

Rel

ativ

e Ab

unda

nce

RT: 2.9428

RT: 2.9294

3.0178 3.11652.8591

NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

NL:9.32E5m/z= 472.3335-472.3383 MS Curve_1_Test_4a_130

RT: 3.1978 - 4.8231

3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8Time (min)

0

200000

400000

600000

800000

1000000

0

1000

2000

3000

4000

5000

Rel

ativ

e Ab

unda

nce

RT: 3.6080

RT: 3.5811

NL:5.46E3m/z= 330.1483-330.1517 MS ICIS Curve_1_Test_4a_113

NL:1.13E6m/z= 336.1860-336.1894 MS ICIS Curve_1_Test_4a_113

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 5000 10000 15000 20000 25000pg/mL

0

5

10

15

20

Area

Rat

io

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 20 40 60 80 100pg/mL

0.00

0.02

0.04

0.06

0.08

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

0 20000 40000 60000 80000 100000pg/mL

0

10

20

30

40

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

20 40 60 80 100pg/mL

0.00

0.01

0.02

0.03

0.04

0.05

Area

Rat

io

RT: 2.7911 - 3.1104

2.80 2.82 2.84 2.86 2.88 2.90 2.92 2.94 2.96 2.98 3.00 3.02 3.04 3.06 3.08 3.10Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Relat

ive Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

RT: 3.4422 - 3.7632

3.46 3.48 3.50 3.52 3.54 3.56 3.58 3.60 3.62 3.64 3.66 3.68 3.70 3.72 3.74 3.76Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Relat

ive Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Buprenorphine Peak Width = 4.2 sec at base Scans Across Peak = 15

Paroxetine Peak Width = 4.8 sec at base Scans Across Peak = 18

The number of scans recorded across the width of each analyte peak correlates with the instrument specification of an analytical scan speed of 3Hz at the 70,000 resolution setting. Analysis of the prepared sample set at each concentration level demonstrated results that met or exceeded all of the previously described acceptance criteria indicating a high level of performance across the working range of the assay (Table 3)

Buprenorphine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 3.2% 4.4% 0 25 23.7 5.1% 1.9% 0 50 47.7 4.6% 1.5% 0

100 98.9 1.1% 1.5% 0 250 242 3.4% 1.0% 0 500 497 0.6% 0.6% 0

1000 1011 1.1% 0.7% 0 2500 2501 0.05% 0.8% 0 5000 5159 3.2% 1.0% 0

10,000 10413 4.1% 0.8% 0 25,000 25794 3.2% 0.9% 0

Paroxetine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 2.7% 8.6% 0 25 23.7 5.3% 1.8% 0 50 48.4 3.3% 1.2% 0

100 101 1.3% 1.5% 0 250 245 2.1% 1.0% 0 500 507 1.4% 0.6% 0

1000 1030 3.0% 0.5% 0 2500 2474 1.0% 1.2% 0 5000 5122 2.4% 0.7% 0

10,000 10286 2.9% 0.4% 0 25,000 24702 1.2% 0.6% 0 50,000 49916 0.2% 1.2% 0

100,000 99385 0.6% 1.2% 0

Table 3. Statistical Results for Buprenorphine and Paroxetine Replicates. Note: AVG %Diff represents the average difference from the linear regression fit of the calibration curve for each set of replicates at a given concentration point.

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

Rel

ativ

e Ab

unda

nce

468.2327

468.1020

468.1486

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

Rel

ativ

e Ab

unda

nce 329.9454329.9065

330.3360

330.2271329.8745

330.2630

330.0081330.2051

330.0590

Buprenorphine – Spectral Analysis Paroxetine – Spectral Analysis

RT: 2.8875 - 3.0177

2.89 2.90 2.91 2.92 2.93 2.94 2.95 2.96 2.97 2.98 2.99 3.00 3.01Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Rel

ativ

e Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

Δppm-2.1

Δppm-1.5

Δppm-1.8

Δppm-2.7

Δppm-1.8

Δppm-2.4 Δppm-1.5

Δppm-1.5

Δppm-1.8

Δppm-1.8

Δppm-1.5

Δppm-2.7

Δppm-0.9

Δppm-2.1

Δppm-1.8

Δppm-2.1 Δppm-0.6

RT: 3.5430 - 3.6824

3.55 3.56 3.57 3.58 3.59 3.60 3.61 3.62 3.63 3.64 3.65 3.66 3.67 3.68Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Rel

ativ

e Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Δppm-2.7

Δppm-2.7

Δppm-1.8

Δppm-2.1

Δppm-1.8

Δppm-1.5

Δppm-0.9

Δppm -1.5

Δppm-1.8 Δppm-1.5

Δppm-1.5

Δppm-2.4 Δppm-2.1

Δppm-1.8

Δppm-2.1

Δppm-0.6

Δppm-1.8

Δppm-0.8 Δppm-1.8

Figure 4. Spectra analysis of Buprenorphine and Paroxetine at 10pg/mL and 1ng/mL in rat plasma. The exact m/z for the target analyte has been underlined in each spectra and the 5ppm mass extraction window used to generate chromatographic peaks has been outlined. Common interferences across each sample concentration have been circled. All interferences present in the sample were demonstrated to be baseline resolved at the 70,000 resolution setting.

c:\xcalibur\...\curve_1_test_4a_117 03/06/15 15:47:02 Compound Mix 0.010ng/mL

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6083 NL:5.16E3m/z= 330.14835-330.15165 MS ICIS Curve_1_Test_4a_112

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6080 NL:5.46E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_113

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Rela

tive

Abun

danc

e

RT: 3.6000 NL:4.65E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_114

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6143 NL:5.33E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_115

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6224 NL:5.89E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_116

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

NL:6.73E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_117

c:\xcalibur\...\curve_1_test_4a_134 03/06/15 18:19:37 Compound Mix 0.100ng/mL

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9428 NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9450 NL:7.69E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_131

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9631 NL:6.88E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_132

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9446 NL:7.27E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_133

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9355 NL:6.57E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_134

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9356 NL:7.89E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_135

Chromatographic reproducibility is an import factor for determining assay quality and ruggedness. Consistent and reproducible signal response at the assay LOQ eliminates the need for manual chromatographic peak integration and improves the overall data confidence and the assay as a whole. Here we examine the chromatographic reproducibility of Buprenorphine and Paroxetine at the assay LOQ of 10pg/mL. Each analyte was analyzed in replicates of n=6. The chromatogram for each replicate can be seen in figure 6.

Buprenorphine – 10pg/mL

Buprenorphine – 1ng/mL

Buprenorphine – 1ng/mL Zoomed

Paroxetine– 10pg/mL

Paroxetine– 1ng/mL

Paroxetine– 1ng/mL Zoomed

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

330.1494

329.9454329.9065 330.3360330.2271329.8745 330.2630330.0081 330.0590

Curve_1_Test_4a_112 #772 RT: 3.65 AV: 1 NL: 1.88E3T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

329.9455

329.9071

330.3363

329.8745

330.1040

330.1910

330.2039330.1495330.2629 330.2880330.2263

330.1180

330.1756

330.0599

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3103

468.2327468.1020 468.1486

Curve_1_Test_4a_134 #630 RT: 2.96 AV: 1 NL: 3.29E3T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3101

468.2329

468.1016

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma Keeley Murphy1, Jonathan Josephs1, Maciej Bromirski2, Olaf Scheibner2, 1Thermo Fisher Scientific, San Jose, CA USA; 2Thermo Fisher Scientific, Bremen, Germany

Conclusion In this experiment we described a method for the quantitation of Buprenorphine and Paroxetine. The LOQ was determined to be 10 pg/mL for each compound. The signal response across the concentration range of the assay was determined to be linear across 4 orders of magnitude, from 10 pg/mL to 100 ng/mL. Evaluation of the analytical scan speed determined that 11-15 scans were generated for chromatographic peak widths ranging between 3-4 seconds at the base. Additionally the spectrum of each analyte was examined near the target m/z to identify possible interferences. At the resolution setting of 70,000, all interferences present in the sample matrix were adequately resolved from the analyte of interest. Also the scan to scan mass accuracy at the LOQ was evaluated demonstrating a Δppm of less than +/- 5ppm for each scan across the analyte peak. These data interrogation options offer valuable method development and troubleshooting tools that provide greater confidence in the analytical method and quantitative results. Overall the Q Exactive Focus mass spectrometer was determined to be very well suited for robust and reproducible quantitation at very low concentration levels in biological matrixes.

References 1. A. Kaufmann, P. Butcher, K. Maden, S. Walker, M. Widmer, Analytica Chimica

Acta, 2010, 673, 60–72

2. Henry H. et al., Rapid Commun. Mass Spectrom. 2012, 26, 499–509

Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma.

Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC).

Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections.

Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development.

Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey.1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma.

Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol.

Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes.

Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo ScientificTM Xcalibur data acquisition software.

Data Analysis All data was processed utilizing Thermo ScientificTM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Figure 2. Calibration Curve for Buprenorphine and Paroxetine. Linear regression was performed using 1/x2 weighting.

Table 1. LC gradient method utilized for sample analysis.

Table 2. Mass Spectrometer settings utilized for sample analysis.

Results Quantitative Results

The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ).

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure.

Figure 5. Zoomed view of Buprenorphine at 10pg/mL (LEFT) and Paroxetine at 10pg/mL (Right). The figure illustrates the scans measured across the peak and the corresponding mass accuracy. Mass accuracy is displayed in parts per million. (ppm)

FIGURE 6. Replicate injections of Buprenorphine (Top) and Paroxetine (bottom) at the assay LOQ of 10pg/mL. Method integration settings where implement for all chromatogram integration and no manual integrations were performed. Additionally no smoothing was applied for any chromatographic peak generation.

A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 25,000 and 10pg/mL to 100,000pg/mL respectively (Figure 2 ).

The analytical scan speed of the instrument was evaluated at the LOQ for each sample. The number of scans across each analyte peak at the LOQ was assessed to ensure adequate scan speed in conjunction with the sample peak width produced by the chromatographic method. The chromatographic peak width for each analyte ranged between 3.6 and 4.2 seconds at the base, providing 11-15 scans for the target analytes at the LOQ (Figure 3).

.

Here we examine the scan to scan mass accuracy of the Q Exactive Focus with an external mass calibration for Buprenorphine and Paroxetine at assay LOQ of 10pg/mL (Figure 5). The mass accuracy for each scan across the analyte peaks was demonstrated to be less than 3ppm for both Buprenorphine and Paroxetine. The high level of mass accuracy from scan to scan allows the utilization of a narrow mass extraction window providing robust and reproducible results at low analyte concentration levels while in the presence of a biological matrix. Evaluation of the scan to scan mass accuracy provides a method development and troubleshooting tool that is unique to HRAM and the Orbitrap mass analyzer.

The spectra of target analytes at the LOQ can be used to verify method quality, determine the suitability of a given resolution setting, and help identify interferences that may inhibit sensitivity and reproducibility. Spectra from injections made at the LOQ for each compound were evaluated to identify unresolved interferences. At the resolution setting of 70,000 (FWHM at m/z 200) both compounds were adequately resolved from all co-eluting interferences (Figure 4). As seen in the figure, although interferences are present during the elution of each compound, the target analyte m/z was baseline resolved from nearby interferences at the 70,000 resolution setting. Additionaly the mass extraction window of +/- 5ppm used to generate the chromatographic peaks used for quantitation is illustrated in the figures. This narrow extraction window allows greater assay confidence and robustness by ensuring that signal contributing to the chromatographic peak area is resulting from the target analyte itself and not the nearby interferences. A narrow mass extraction windows and the resulting XIC do not provide analytical value without a high level of scan to scan mass accuracy across the entire analyte peak. Scan to scan mass accuracy can also offer an important indicator for indentifying potential analyte interferences as well for accessing overall assay performance. A loss in mass accuracy can provide an indication of an unresolved or co eluting interference and can be implemented as a quality control measure in validated analyses.

Figure 3. Evaluation of chromatographic peak width vs scan speed for Buprenorphine (left) and Paroxetine (right) the LOQ of 10pg/mL

Time (min)

Flow rate (uL/min) %A %B

0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95

6.0 500 5 95 6.1 500 95 5 8.0 500 95 5

HESI Source Settings Value MS Scan Settings Value

Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature (OC) 450 Resolution 70,000 Capillary Temperature (OC) 350 AGC Target 2.00E+04 Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1

RT: 2.5471 - 4.0071

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0Time (min)

0

200000

400000

600000

800000

0

1000

2000

3000

4000

5000

6000

7000

Rel

ativ

e Ab

unda

nce

RT: 2.9428

RT: 2.9294

3.0178 3.11652.8591

NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

NL:9.32E5m/z= 472.3335-472.3383 MS Curve_1_Test_4a_130

RT: 3.1978 - 4.8231

3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8Time (min)

0

200000

400000

600000

800000

1000000

0

1000

2000

3000

4000

5000

Rel

ativ

e Ab

unda

nce

RT: 3.6080

RT: 3.5811

NL:5.46E3m/z= 330.1483-330.1517 MS ICIS Curve_1_Test_4a_113

NL:1.13E6m/z= 336.1860-336.1894 MS ICIS Curve_1_Test_4a_113

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 5000 10000 15000 20000 25000pg/mL

0

5

10

15

20

Area

Rat

io

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 20 40 60 80 100pg/mL

0.00

0.02

0.04

0.06

0.08

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

0 20000 40000 60000 80000 100000pg/mL

0

10

20

30

40

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

20 40 60 80 100pg/mL

0.00

0.01

0.02

0.03

0.04

0.05

Area

Rat

io

RT: 2.7911 - 3.1104

2.80 2.82 2.84 2.86 2.88 2.90 2.92 2.94 2.96 2.98 3.00 3.02 3.04 3.06 3.08 3.10Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Relat

ive Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

RT: 3.4422 - 3.7632

3.46 3.48 3.50 3.52 3.54 3.56 3.58 3.60 3.62 3.64 3.66 3.68 3.70 3.72 3.74 3.76Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Relat

ive Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Buprenorphine Peak Width = 4.2 sec at base Scans Across Peak = 15

Paroxetine Peak Width = 4.8 sec at base Scans Across Peak = 18

The number of scans recorded across the width of each analyte peak correlates with the instrument specification of an analytical scan speed of 3Hz at the 70,000 resolution setting. Analysis of the prepared sample set at each concentration level demonstrated results that met or exceeded all of the previously described acceptance criteria indicating a high level of performance across the working range of the assay (Table 3)

Buprenorphine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 3.2% 4.4% 0 25 23.7 5.1% 1.9% 0 50 47.7 4.6% 1.5% 0

100 98.9 1.1% 1.5% 0 250 242 3.4% 1.0% 0 500 497 0.6% 0.6% 0

1000 1011 1.1% 0.7% 0 2500 2501 0.05% 0.8% 0 5000 5159 3.2% 1.0% 0

10,000 10413 4.1% 0.8% 0 25,000 25794 3.2% 0.9% 0

Paroxetine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 2.7% 8.6% 0 25 23.7 5.3% 1.8% 0 50 48.4 3.3% 1.2% 0

100 101 1.3% 1.5% 0 250 245 2.1% 1.0% 0 500 507 1.4% 0.6% 0

1000 1030 3.0% 0.5% 0 2500 2474 1.0% 1.2% 0 5000 5122 2.4% 0.7% 0

10,000 10286 2.9% 0.4% 0 25,000 24702 1.2% 0.6% 0 50,000 49916 0.2% 1.2% 0

100,000 99385 0.6% 1.2% 0

Table 3. Statistical Results for Buprenorphine and Paroxetine Replicates. Note: AVG %Diff represents the average difference from the linear regression fit of the calibration curve for each set of replicates at a given concentration point.

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

Rel

ativ

e Ab

unda

nce

468.2327

468.1020

468.1486

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0R

elat

ive

Abun

danc

e 329.9454329.9065

330.3360

330.2271329.8745

330.2630

330.0081330.2051

330.0590

Buprenorphine – Spectral Analysis Paroxetine – Spectral Analysis

RT: 2.8875 - 3.0177

2.89 2.90 2.91 2.92 2.93 2.94 2.95 2.96 2.97 2.98 2.99 3.00 3.01Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Rel

ativ

e Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

Δppm-2.1

Δppm-1.5

Δppm-1.8

Δppm-2.7

Δppm-1.8

Δppm-2.4 Δppm-1.5

Δppm-1.5

Δppm-1.8

Δppm-1.8

Δppm-1.5

Δppm-2.7

Δppm-0.9

Δppm-2.1

Δppm-1.8

Δppm-2.1 Δppm-0.6

RT: 3.5430 - 3.6824

3.55 3.56 3.57 3.58 3.59 3.60 3.61 3.62 3.63 3.64 3.65 3.66 3.67 3.68Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Rel

ativ

e Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Δppm-2.7

Δppm-2.7

Δppm-1.8

Δppm-2.1

Δppm-1.8

Δppm-1.5

Δppm-0.9

Δppm -1.5

Δppm-1.8 Δppm-1.5

Δppm-1.5

Δppm-2.4 Δppm-2.1

Δppm-1.8

Δppm-2.1

Δppm-0.6

Δppm-1.8

Δppm-0.8 Δppm-1.8

Figure 4. Spectra analysis of Buprenorphine and Paroxetine at 10pg/mL and 1ng/mL in rat plasma. The exact m/z for the target analyte has been underlined in each spectra and the 5ppm mass extraction window used to generate chromatographic peaks has been outlined. Common interferences across each sample concentration have been circled. All interferences present in the sample were demonstrated to be baseline resolved at the 70,000 resolution setting.

c:\xcalibur\...\curve_1_test_4a_117 03/06/15 15:47:02 Compound Mix 0.010ng/mL

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6083 NL:5.16E3m/z= 330.14835-330.15165 MS ICIS Curve_1_Test_4a_112

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6080 NL:5.46E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_113

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Rela

tive

Abun

danc

e

RT: 3.6000 NL:4.65E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_114

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6143 NL:5.33E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_115

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6224 NL:5.89E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_116

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

NL:6.73E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_117

c:\xcalibur\...\curve_1_test_4a_134 03/06/15 18:19:37 Compound Mix 0.100ng/mL

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9428 NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9450 NL:7.69E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_131

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9631 NL:6.88E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_132

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9446 NL:7.27E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_133

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9355 NL:6.57E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_134

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9356 NL:7.89E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_135

Chromatographic reproducibility is an import factor for determining assay quality and ruggedness. Consistent and reproducible signal response at the assay LOQ eliminates the need for manual chromatographic peak integration and improves the overall data confidence and the assay as a whole. Here we examine the chromatographic reproducibility of Buprenorphine and Paroxetine at the assay LOQ of 10pg/mL. Each analyte was analyzed in replicates of n=6. The chromatogram for each replicate can be seen in figure 6.

Buprenorphine – 10pg/mL

Buprenorphine – 1ng/mL

Buprenorphine – 1ng/mL Zoomed

Paroxetine– 10pg/mL

Paroxetine– 1ng/mL

Paroxetine– 1ng/mL Zoomed

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

330.1494

329.9454329.9065 330.3360330.2271329.8745 330.2630330.0081 330.0590

Curve_1_Test_4a_112 #772 RT: 3.65 AV: 1 NL: 1.88E3T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

329.9455

329.9071

330.3363

329.8745

330.1040

330.1910

330.2039330.1495330.2629 330.2880330.2263

330.1180

330.1756

330.0599

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3103

468.2327468.1020 468.1486

Curve_1_Test_4a_134 #630 RT: 2.96 AV: 1 NL: 3.29E3T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3101

468.2329

468.1016

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma Keeley Murphy1, Jonathan Josephs1, Maciej Bromirski2, Olaf Scheibner2, 1Thermo Fisher Scientific, San Jose, CA USA; 2Thermo Fisher Scientific, Bremen, Germany

Conclusion In this experiment we described a method for the quantitation of Buprenorphine and Paroxetine. The LOQ was determined to be 10 pg/mL for each compound. The signal response across the concentration range of the assay was determined to be linear across 4 orders of magnitude, from 10 pg/mL to 100 ng/mL. Evaluation of the analytical scan speed determined that 11-15 scans were generated for chromatographic peak widths ranging between 3-4 seconds at the base. Additionally the spectrum of each analyte was examined near the target m/z to identify possible interferences. At the resolution setting of 70,000, all interferences present in the sample matrix were adequately resolved from the analyte of interest. Also the scan to scan mass accuracy at the LOQ was evaluated demonstrating a Δppm of less than +/- 5ppm for each scan across the analyte peak. These data interrogation options offer valuable method development and troubleshooting tools that provide greater confidence in the analytical method and quantitative results. Overall the Q Exactive Focus mass spectrometer was determined to be very well suited for robust and reproducible quantitation at very low concentration levels in biological matrixes.

References 1. A. Kaufmann, P. Butcher, K. Maden, S. Walker, M. Widmer, Analytica Chimica

Acta, 2010, 673, 60–72

2. Henry H. et al., Rapid Commun. Mass Spectrom. 2012, 26, 499–509

Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma.

Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC).

Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections.

Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development.

Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey.1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma.

Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol.

Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes.

Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo ScientificTM Xcalibur data acquisition software.

Data Analysis All data was processed utilizing Thermo ScientificTM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Figure 2. Calibration Curve for Buprenorphine and Paroxetine. Linear regression was performed using 1/x2 weighting.

Table 1. LC gradient method utilized for sample analysis.

Table 2. Mass Spectrometer settings utilized for sample analysis.

Results Quantitative Results

The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ).

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure.

Figure 5. Zoomed view of Buprenorphine at 10pg/mL (LEFT) and Paroxetine at 10pg/mL (Right). The figure illustrates the scans measured across the peak and the corresponding mass accuracy. Mass accuracy is displayed in parts per million. (ppm)

FIGURE 6. Replicate injections of Buprenorphine (Top) and Paroxetine (bottom) at the assay LOQ of 10pg/mL. Method integration settings where implement for all chromatogram integration and no manual integrations were performed. Additionally no smoothing was applied for any chromatographic peak generation.

A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 25,000 and 10pg/mL to 100,000pg/mL respectively (Figure 2 ).

The analytical scan speed of the instrument was evaluated at the LOQ for each sample. The number of scans across each analyte peak at the LOQ was assessed to ensure adequate scan speed in conjunction with the sample peak width produced by the chromatographic method. The chromatographic peak width for each analyte ranged between 3.6 and 4.2 seconds at the base, providing 11-15 scans for the target analytes at the LOQ (Figure 3).

.

Here we examine the scan to scan mass accuracy of the Q Exactive Focus with an external mass calibration for Buprenorphine and Paroxetine at assay LOQ of 10pg/mL (Figure 5). The mass accuracy for each scan across the analyte peaks was demonstrated to be less than 3ppm for both Buprenorphine and Paroxetine. The high level of mass accuracy from scan to scan allows the utilization of a narrow mass extraction window providing robust and reproducible results at low analyte concentration levels while in the presence of a biological matrix. Evaluation of the scan to scan mass accuracy provides a method development and troubleshooting tool that is unique to HRAM and the Orbitrap mass analyzer.

The spectra of target analytes at the LOQ can be used to verify method quality, determine the suitability of a given resolution setting, and help identify interferences that may inhibit sensitivity and reproducibility. Spectra from injections made at the LOQ for each compound were evaluated to identify unresolved interferences. At the resolution setting of 70,000 (FWHM at m/z 200) both compounds were adequately resolved from all co-eluting interferences (Figure 4). As seen in the figure, although interferences are present during the elution of each compound, the target analyte m/z was baseline resolved from nearby interferences at the 70,000 resolution setting. Additionaly the mass extraction window of +/- 5ppm used to generate the chromatographic peaks used for quantitation is illustrated in the figures. This narrow extraction window allows greater assay confidence and robustness by ensuring that signal contributing to the chromatographic peak area is resulting from the target analyte itself and not the nearby interferences. A narrow mass extraction windows and the resulting XIC do not provide analytical value without a high level of scan to scan mass accuracy across the entire analyte peak. Scan to scan mass accuracy can also offer an important indicator for indentifying potential analyte interferences as well for accessing overall assay performance. A loss in mass accuracy can provide an indication of an unresolved or co eluting interference and can be implemented as a quality control measure in validated analyses.

Figure 3. Evaluation of chromatographic peak width vs scan speed for Buprenorphine (left) and Paroxetine (right) the LOQ of 10pg/mL

Time (min)

Flow rate (uL/min) %A %B

0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95

6.0 500 5 95 6.1 500 95 5 8.0 500 95 5

HESI Source Settings Value MS Scan Settings Value

Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature (OC) 450 Resolution 70,000 Capillary Temperature (OC) 350 AGC Target 2.00E+04 Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1

RT: 2.5471 - 4.0071

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0Time (min)

0

200000

400000

600000

800000

0

1000

2000

3000

4000

5000

6000

7000

Rel

ativ

e Ab

unda

nce

RT: 2.9428

RT: 2.9294

3.0178 3.11652.8591

NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

NL:9.32E5m/z= 472.3335-472.3383 MS Curve_1_Test_4a_130

RT: 3.1978 - 4.8231

3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8Time (min)

0

200000

400000

600000

800000

1000000

0

1000

2000

3000

4000

5000

Rel

ativ

e Ab

unda

nce

RT: 3.6080

RT: 3.5811

NL:5.46E3m/z= 330.1483-330.1517 MS ICIS Curve_1_Test_4a_113

NL:1.13E6m/z= 336.1860-336.1894 MS ICIS Curve_1_Test_4a_113

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 5000 10000 15000 20000 25000pg/mL

0

5

10

15

20

Area

Rat

io

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 20 40 60 80 100pg/mL

0.00

0.02

0.04

0.06

0.08

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

0 20000 40000 60000 80000 100000pg/mL

0

10

20

30

40

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

20 40 60 80 100pg/mL

0.00

0.01

0.02

0.03

0.04

0.05

Area

Rat

io

RT: 2.7911 - 3.1104

2.80 2.82 2.84 2.86 2.88 2.90 2.92 2.94 2.96 2.98 3.00 3.02 3.04 3.06 3.08 3.10Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Relat

ive Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

RT: 3.4422 - 3.7632

3.46 3.48 3.50 3.52 3.54 3.56 3.58 3.60 3.62 3.64 3.66 3.68 3.70 3.72 3.74 3.76Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Relat

ive Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Buprenorphine Peak Width = 4.2 sec at base Scans Across Peak = 15

Paroxetine Peak Width = 4.8 sec at base Scans Across Peak = 18

The number of scans recorded across the width of each analyte peak correlates with the instrument specification of an analytical scan speed of 3Hz at the 70,000 resolution setting. Analysis of the prepared sample set at each concentration level demonstrated results that met or exceeded all of the previously described acceptance criteria indicating a high level of performance across the working range of the assay (Table 3)

Buprenorphine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 3.2% 4.4% 0 25 23.7 5.1% 1.9% 0 50 47.7 4.6% 1.5% 0

100 98.9 1.1% 1.5% 0 250 242 3.4% 1.0% 0 500 497 0.6% 0.6% 0

1000 1011 1.1% 0.7% 0 2500 2501 0.05% 0.8% 0 5000 5159 3.2% 1.0% 0

10,000 10413 4.1% 0.8% 0 25,000 25794 3.2% 0.9% 0

Paroxetine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 2.7% 8.6% 0 25 23.7 5.3% 1.8% 0 50 48.4 3.3% 1.2% 0

100 101 1.3% 1.5% 0 250 245 2.1% 1.0% 0 500 507 1.4% 0.6% 0

1000 1030 3.0% 0.5% 0 2500 2474 1.0% 1.2% 0 5000 5122 2.4% 0.7% 0

10,000 10286 2.9% 0.4% 0 25,000 24702 1.2% 0.6% 0 50,000 49916 0.2% 1.2% 0

100,000 99385 0.6% 1.2% 0

Table 3. Statistical Results for Buprenorphine and Paroxetine Replicates. Note: AVG %Diff represents the average difference from the linear regression fit of the calibration curve for each set of replicates at a given concentration point.

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

Rel

ativ

e Ab

unda

nce

468.2327

468.1020

468.1486

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

Rel

ativ

e Ab

unda

nce 329.9454329.9065

330.3360

330.2271329.8745

330.2630

330.0081330.2051

330.0590

Buprenorphine – Spectral Analysis Paroxetine – Spectral Analysis

RT: 2.8875 - 3.0177

2.89 2.90 2.91 2.92 2.93 2.94 2.95 2.96 2.97 2.98 2.99 3.00 3.01Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Rel

ativ

e Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

Δppm-2.1

Δppm-1.5

Δppm-1.8

Δppm-2.7

Δppm-1.8

Δppm-2.4 Δppm-1.5

Δppm-1.5

Δppm-1.8

Δppm-1.8

Δppm-1.5

Δppm-2.7

Δppm-0.9

Δppm-2.1

Δppm-1.8

Δppm-2.1 Δppm-0.6

RT: 3.5430 - 3.6824

3.55 3.56 3.57 3.58 3.59 3.60 3.61 3.62 3.63 3.64 3.65 3.66 3.67 3.68Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Rel

ativ

e Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Δppm-2.7

Δppm-2.7

Δppm-1.8

Δppm-2.1

Δppm-1.8

Δppm-1.5

Δppm-0.9

Δppm -1.5

Δppm-1.8 Δppm-1.5

Δppm-1.5

Δppm-2.4 Δppm-2.1

Δppm-1.8

Δppm-2.1

Δppm-0.6

Δppm-1.8

Δppm-0.8 Δppm-1.8

Figure 4. Spectra analysis of Buprenorphine and Paroxetine at 10pg/mL and 1ng/mL in rat plasma. The exact m/z for the target analyte has been underlined in each spectra and the 5ppm mass extraction window used to generate chromatographic peaks has been outlined. Common interferences across each sample concentration have been circled. All interferences present in the sample were demonstrated to be baseline resolved at the 70,000 resolution setting.

c:\xcalibur\...\curve_1_test_4a_117 03/06/15 15:47:02 Compound Mix 0.010ng/mL

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6083 NL:5.16E3m/z= 330.14835-330.15165 MS ICIS Curve_1_Test_4a_112

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6080 NL:5.46E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_113

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Rela

tive

Abun

danc

e

RT: 3.6000 NL:4.65E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_114

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6143 NL:5.33E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_115

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6224 NL:5.89E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_116

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

NL:6.73E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_117

c:\xcalibur\...\curve_1_test_4a_134 03/06/15 18:19:37 Compound Mix 0.100ng/mL

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9428 NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9450 NL:7.69E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_131

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9631 NL:6.88E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_132

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9446 NL:7.27E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_133

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9355 NL:6.57E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_134

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9356 NL:7.89E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_135

Chromatographic reproducibility is an import factor for determining assay quality and ruggedness. Consistent and reproducible signal response at the assay LOQ eliminates the need for manual chromatographic peak integration and improves the overall data confidence and the assay as a whole. Here we examine the chromatographic reproducibility of Buprenorphine and Paroxetine at the assay LOQ of 10pg/mL. Each analyte was analyzed in replicates of n=6. The chromatogram for each replicate can be seen in figure 6.

Buprenorphine – 10pg/mL

Buprenorphine – 1ng/mL

Buprenorphine – 1ng/mL Zoomed

Paroxetine– 10pg/mL

Paroxetine– 1ng/mL

Paroxetine– 1ng/mL Zoomed

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

330.1494

329.9454329.9065 330.3360330.2271329.8745 330.2630330.0081 330.0590

Curve_1_Test_4a_112 #772 RT: 3.65 AV: 1 NL: 1.88E3T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

329.9455

329.9071

330.3363

329.8745

330.1040

330.1910

330.2039330.1495330.2629 330.2880330.2263

330.1180

330.1756

330.0599

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3103

468.2327468.1020 468.1486

Curve_1_Test_4a_134 #630 RT: 2.96 AV: 1 NL: 3.29E3T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3101

468.2329

468.1016

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma Keeley Murphy1, Jonathan Josephs1, Maciej Bromirski2, Olaf Scheibner2, 1Thermo Fisher Scientific, San Jose, CA USA; 2Thermo Fisher Scientific, Bremen, Germany

Conclusion In this experiment we described a method for the quantitation of Buprenorphine and Paroxetine. The LOQ was determined to be 10 pg/mL for each compound. The signal response across the concentration range of the assay was determined to be linear across 4 orders of magnitude, from 10 pg/mL to 100 ng/mL. Evaluation of the analytical scan speed determined that 11-15 scans were generated for chromatographic peak widths ranging between 3-4 seconds at the base. Additionally the spectrum of each analyte was examined near the target m/z to identify possible interferences. At the resolution setting of 70,000, all interferences present in the sample matrix were adequately resolved from the analyte of interest. Also the scan to scan mass accuracy at the LOQ was evaluated demonstrating a Δppm of less than +/- 5ppm for each scan across the analyte peak. These data interrogation options offer valuable method development and troubleshooting tools that provide greater confidence in the analytical method and quantitative results. Overall the Q Exactive Focus mass spectrometer was determined to be very well suited for robust and reproducible quantitation at very low concentration levels in biological matrixes.

References 1. A. Kaufmann, P. Butcher, K. Maden, S. Walker, M. Widmer, Analytica Chimica

Acta, 2010, 673, 60–72

2. Henry H. et al., Rapid Commun. Mass Spectrom. 2012, 26, 499–509

Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma.

Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC).

Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections.

Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development.

Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey.1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma.

Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol.

Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes.

Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo ScientificTM Xcalibur data acquisition software.

Data Analysis All data was processed utilizing Thermo ScientificTM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Figure 2. Calibration Curve for Buprenorphine and Paroxetine. Linear regression was performed using 1/x2 weighting.

Table 1. LC gradient method utilized for sample analysis.

Table 2. Mass Spectrometer settings utilized for sample analysis.

Results Quantitative Results

The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ).

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure.

Figure 5. Zoomed view of Buprenorphine at 10pg/mL (LEFT) and Paroxetine at 10pg/mL (Right). The figure illustrates the scans measured across the peak and the corresponding mass accuracy. Mass accuracy is displayed in parts per million. (ppm)

FIGURE 6. Replicate injections of Buprenorphine (Top) and Paroxetine (bottom) at the assay LOQ of 10pg/mL. Method integration settings where implement for all chromatogram integration and no manual integrations were performed. Additionally no smoothing was applied for any chromatographic peak generation.

A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 25,000 and 10pg/mL to 100,000pg/mL respectively (Figure 2 ).

The analytical scan speed of the instrument was evaluated at the LOQ for each sample. The number of scans across each analyte peak at the LOQ was assessed to ensure adequate scan speed in conjunction with the sample peak width produced by the chromatographic method. The chromatographic peak width for each analyte ranged between 3.6 and 4.2 seconds at the base, providing 11-15 scans for the target analytes at the LOQ (Figure 3).

.

Here we examine the scan to scan mass accuracy of the Q Exactive Focus with an external mass calibration for Buprenorphine and Paroxetine at assay LOQ of 10pg/mL (Figure 5). The mass accuracy for each scan across the analyte peaks was demonstrated to be less than 3ppm for both Buprenorphine and Paroxetine. The high level of mass accuracy from scan to scan allows the utilization of a narrow mass extraction window providing robust and reproducible results at low analyte concentration levels while in the presence of a biological matrix. Evaluation of the scan to scan mass accuracy provides a method development and troubleshooting tool that is unique to HRAM and the Orbitrap mass analyzer.

The spectra of target analytes at the LOQ can be used to verify method quality, determine the suitability of a given resolution setting, and help identify interferences that may inhibit sensitivity and reproducibility. Spectra from injections made at the LOQ for each compound were evaluated to identify unresolved interferences. At the resolution setting of 70,000 (FWHM at m/z 200) both compounds were adequately resolved from all co-eluting interferences (Figure 4). As seen in the figure, although interferences are present during the elution of each compound, the target analyte m/z was baseline resolved from nearby interferences at the 70,000 resolution setting. Additionaly the mass extraction window of +/- 5ppm used to generate the chromatographic peaks used for quantitation is illustrated in the figures. This narrow extraction window allows greater assay confidence and robustness by ensuring that signal contributing to the chromatographic peak area is resulting from the target analyte itself and not the nearby interferences. A narrow mass extraction windows and the resulting XIC do not provide analytical value without a high level of scan to scan mass accuracy across the entire analyte peak. Scan to scan mass accuracy can also offer an important indicator for indentifying potential analyte interferences as well for accessing overall assay performance. A loss in mass accuracy can provide an indication of an unresolved or co eluting interference and can be implemented as a quality control measure in validated analyses.

Figure 3. Evaluation of chromatographic peak width vs scan speed for Buprenorphine (left) and Paroxetine (right) the LOQ of 10pg/mL

Time (min)

Flow rate (uL/min) %A %B

0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95

6.0 500 5 95 6.1 500 95 5 8.0 500 95 5

HESI Source Settings Value MS Scan Settings Value

Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature (OC) 450 Resolution 70,000 Capillary Temperature (OC) 350 AGC Target 2.00E+04 Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1

RT: 2.5471 - 4.0071

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0Time (min)

0

200000

400000

600000

800000

0

1000

2000

3000

4000

5000

6000

7000

Rel

ativ

e Ab

unda

nce

RT: 2.9428

RT: 2.9294

3.0178 3.11652.8591

NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

NL:9.32E5m/z= 472.3335-472.3383 MS Curve_1_Test_4a_130

RT: 3.1978 - 4.8231

3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8Time (min)

0

200000

400000

600000

800000

1000000

0

1000

2000

3000

4000

5000

Rel

ativ

e Ab

unda

nce

RT: 3.6080

RT: 3.5811

NL:5.46E3m/z= 330.1483-330.1517 MS ICIS Curve_1_Test_4a_113

NL:1.13E6m/z= 336.1860-336.1894 MS ICIS Curve_1_Test_4a_113

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 5000 10000 15000 20000 25000pg/mL

0

5

10

15

20

Area

Rat

io

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 20 40 60 80 100pg/mL

0.00

0.02

0.04

0.06

0.08

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

0 20000 40000 60000 80000 100000pg/mL

0

10

20

30

40

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

20 40 60 80 100pg/mL

0.00

0.01

0.02

0.03

0.04

0.05

Area

Rat

io

RT: 2.7911 - 3.1104

2.80 2.82 2.84 2.86 2.88 2.90 2.92 2.94 2.96 2.98 3.00 3.02 3.04 3.06 3.08 3.10Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Relat

ive Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

RT: 3.4422 - 3.7632

3.46 3.48 3.50 3.52 3.54 3.56 3.58 3.60 3.62 3.64 3.66 3.68 3.70 3.72 3.74 3.76Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Relat

ive Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Buprenorphine Peak Width = 4.2 sec at base Scans Across Peak = 15

Paroxetine Peak Width = 4.8 sec at base Scans Across Peak = 18

The number of scans recorded across the width of each analyte peak correlates with the instrument specification of an analytical scan speed of 3Hz at the 70,000 resolution setting. Analysis of the prepared sample set at each concentration level demonstrated results that met or exceeded all of the previously described acceptance criteria indicating a high level of performance across the working range of the assay (Table 3)

Buprenorphine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 3.2% 4.4% 0 25 23.7 5.1% 1.9% 0 50 47.7 4.6% 1.5% 0

100 98.9 1.1% 1.5% 0 250 242 3.4% 1.0% 0 500 497 0.6% 0.6% 0

1000 1011 1.1% 0.7% 0 2500 2501 0.05% 0.8% 0 5000 5159 3.2% 1.0% 0

10,000 10413 4.1% 0.8% 0 25,000 25794 3.2% 0.9% 0

Paroxetine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 2.7% 8.6% 0 25 23.7 5.3% 1.8% 0 50 48.4 3.3% 1.2% 0

100 101 1.3% 1.5% 0 250 245 2.1% 1.0% 0 500 507 1.4% 0.6% 0

1000 1030 3.0% 0.5% 0 2500 2474 1.0% 1.2% 0 5000 5122 2.4% 0.7% 0

10,000 10286 2.9% 0.4% 0 25,000 24702 1.2% 0.6% 0 50,000 49916 0.2% 1.2% 0

100,000 99385 0.6% 1.2% 0

Table 3. Statistical Results for Buprenorphine and Paroxetine Replicates. Note: AVG %Diff represents the average difference from the linear regression fit of the calibration curve for each set of replicates at a given concentration point.

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

Rel

ativ

e Ab

unda

nce

468.2327

468.1020

468.1486

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

Rel

ativ

e Ab

unda

nce 329.9454329.9065

330.3360

330.2271329.8745

330.2630

330.0081330.2051

330.0590

Buprenorphine – Spectral Analysis Paroxetine – Spectral Analysis

RT: 2.8875 - 3.0177

2.89 2.90 2.91 2.92 2.93 2.94 2.95 2.96 2.97 2.98 2.99 3.00 3.01Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Rel

ativ

e Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

Δppm-2.1

Δppm-1.5

Δppm-1.8

Δppm-2.7

Δppm-1.8

Δppm-2.4 Δppm-1.5

Δppm-1.5

Δppm-1.8

Δppm-1.8

Δppm-1.5

Δppm-2.7

Δppm-0.9

Δppm-2.1

Δppm-1.8

Δppm-2.1 Δppm-0.6

RT: 3.5430 - 3.6824

3.55 3.56 3.57 3.58 3.59 3.60 3.61 3.62 3.63 3.64 3.65 3.66 3.67 3.68Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Rel

ativ

e Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Δppm-2.7

Δppm-2.7

Δppm-1.8

Δppm-2.1

Δppm-1.8

Δppm-1.5

Δppm-0.9

Δppm -1.5

Δppm-1.8 Δppm-1.5

Δppm-1.5

Δppm-2.4 Δppm-2.1

Δppm-1.8

Δppm-2.1

Δppm-0.6

Δppm-1.8

Δppm-0.8 Δppm-1.8

Figure 4. Spectra analysis of Buprenorphine and Paroxetine at 10pg/mL and 1ng/mL in rat plasma. The exact m/z for the target analyte has been underlined in each spectra and the 5ppm mass extraction window used to generate chromatographic peaks has been outlined. Common interferences across each sample concentration have been circled. All interferences present in the sample were demonstrated to be baseline resolved at the 70,000 resolution setting.

c:\xcalibur\...\curve_1_test_4a_117 03/06/15 15:47:02 Compound Mix 0.010ng/mL

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6083 NL:5.16E3m/z= 330.14835-330.15165 MS ICIS Curve_1_Test_4a_112

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6080 NL:5.46E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_113

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Rela

tive

Abun

danc

e

RT: 3.6000 NL:4.65E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_114

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6143 NL:5.33E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_115

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6224 NL:5.89E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_116

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

NL:6.73E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_117

c:\xcalibur\...\curve_1_test_4a_134 03/06/15 18:19:37 Compound Mix 0.100ng/mL

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9428 NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9450 NL:7.69E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_131

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9631 NL:6.88E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_132

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9446 NL:7.27E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_133

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9355 NL:6.57E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_134

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9356 NL:7.89E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_135

Chromatographic reproducibility is an import factor for determining assay quality and ruggedness. Consistent and reproducible signal response at the assay LOQ eliminates the need for manual chromatographic peak integration and improves the overall data confidence and the assay as a whole. Here we examine the chromatographic reproducibility of Buprenorphine and Paroxetine at the assay LOQ of 10pg/mL. Each analyte was analyzed in replicates of n=6. The chromatogram for each replicate can be seen in figure 6.

Buprenorphine – 10pg/mL

Buprenorphine – 1ng/mL

Buprenorphine – 1ng/mL Zoomed

Paroxetine– 10pg/mL

Paroxetine– 1ng/mL

Paroxetine– 1ng/mL Zoomed

Page 3: A High Resolution Accurate Mass Approach for the ... · A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in ... linear response has remained

PN64485-EN 0615S

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

330.1494

329.9454329.9065 330.3360330.2271329.8745 330.2630330.0081 330.0590

Curve_1_Test_4a_112 #772 RT: 3.65 AV: 1 NL: 1.88E3T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

329.9455

329.9071

330.3363

329.8745

330.1040

330.1910

330.2039330.1495330.2629 330.2880330.2263

330.1180

330.1756

330.0599

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3103

468.2327468.1020 468.1486

Curve_1_Test_4a_134 #630 RT: 2.96 AV: 1 NL: 3.29E3T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e Ab

unda

nce

468.3101

468.2329

468.1016

A High Resolution Accurate Mass Approach for the Quantitation of Buprenorphine and Paroxetine in Rat Plasma Keeley Murphy1, Jonathan Josephs1, Maciej Bromirski2, Olaf Scheibner2, 1Thermo Fisher Scientific, San Jose, CA USA; 2Thermo Fisher Scientific, Bremen, Germany

Conclusion In this experiment we described a method for the quantitation of Buprenorphine and Paroxetine. The LOQ was determined to be 10 pg/mL for each compound. The signal response across the concentration range of the assay was determined to be linear across 4 orders of magnitude, from 10 pg/mL to 100 ng/mL. Evaluation of the analytical scan speed determined that 11-15 scans were generated for chromatographic peak widths ranging between 3-4 seconds at the base. Additionally the spectrum of each analyte was examined near the target m/z to identify possible interferences. At the resolution setting of 70,000, all interferences present in the sample matrix were adequately resolved from the analyte of interest. Also the scan to scan mass accuracy at the LOQ was evaluated demonstrating a Δppm of less than +/- 5ppm for each scan across the analyte peak. These data interrogation options offer valuable method development and troubleshooting tools that provide greater confidence in the analytical method and quantitative results. Overall the Q Exactive Focus mass spectrometer was determined to be very well suited for robust and reproducible quantitation at very low concentration levels in biological matrixes.

References 1. A. Kaufmann, P. Butcher, K. Maden, S. Walker, M. Widmer, Analytica Chimica

Acta, 2010, 673, 60–72

2. Henry H. et al., Rapid Commun. Mass Spectrom. 2012, 26, 499–509

Overview Purpose: To demonstrate a High Resolution Accurate Mass (HRAM) approach for the quantitation of Buprenorphine and Paroxetine in the presence of protein precipitated rat plasma.

Methods: Sample analysis was performed using Single Ion Monitoring (SIM) with generic mass spectrometer source conditions and chromatographic separation utilizing ultra high performance chromatography (UHPLC).

Results: Sample analysis demonstrated a lower limit of quantitation (LOQ) of 10pg/mL for both analytes, a linear signal response of more than four orders of magnitude, a coefficient of variance of less than ten percent for all replicate injections.

Introduction As potential leads for successful drug candidates move through the drug discovery process there is a need for quantitative in vitro and in vivo analysis at each step of the journey. The requirement for quantitative assays that provide sensitivity, ruggedness, and linear response has remained a constant but as mass spectrometer technology has progressed over time other considerations also play an important factor in meeting the needs of these assays. Ease of use, simplified method development, and troubleshooting tools, now play an important factor when choosing the most appropriate technology for a particular assay and the corresponding method development.

Quantitation by high resolution accurate mass (HRAM) delivers reliable results for analyses at extremely low concentration levels and in complex matrices . HRAM enables direct analysis and monitoring of the ionized target analyte through SIM scanning, without the need to improve selectivity through fragmentation or multiple reaction monitoring (MRM). HRAM analysis at a resolution setting of 50,000 (FWHM at m/z 200) or greater has been routinely demonstrated to effectively resolve interferences present in complex matrixes such as plasma and honey.1,2 Data collection in SIM mode also provides additional information regarding possible interferences near the target analyte m/z. This additional information can be highly beneficial in method development, method optimization, and troubleshooting while eliminating the need for additional injections or complex scan experiments. Here we examine the quantitative performance of the Thermo Scientific™ Q Exactive™ Focus mass spectrometer for the analysis of Buprenorphine and Paroxetine in rat plasma.

Methods Sample Preparation Crashed plasma stock solutions were prepared using an Acetonitrile (ACN) crash at a ratio of 3:1, ACN to plasma. The resulting solution was centrifuged at 10,000rpm for 10 minutes. The supernatant was removed and added to an equivalent volume of water to make the final crashed plasma stock solution. Stock solutions of Buprenorphine and Paroxetine at 1mg/mL were diluted in the crashed plasma stock to concentration ranges of 10pg/mL to 25,000pg/mL and 10pg/mL to 100,000pg/mL respectively. Isotopically labeled Buprenorphine-D4 and Paroxetine-D6 were added at each concentration level as an internal standard to produce a final internal standard concentration of 1ng/mL. All reagents were obtained from Cerilliant Corporation, Round Rock, Texas, at 1mg/mL in methanol.

Liquid Chromatography Chromatographic separation was achieved using Thermo Scientific™ Vanquish™ UHPLC System. Samples were injected (5uL) onto a 2.1 x 100mm, 1.9um Thermo Scientific™ Hypersil Gold aQ column. Gradient elution was accomplished using water + 0.1% formic acid (FA) (A) and acetonitrile + 0.1% formic acid (FA) (B), with a 6 minute gradient at a flow rate of 500uL/min (Table 1). Total run time including column equilibration was approximately 8 minutes.

Mass Spectrometry Compounds were analyzed utilizing a Thermo Scientific™ Q Exactive Focus™ MS with heated electrospray ionization (HESI-II). Generic source conditions suitable for a 500uL/min LC flow rate were applied for all data collection (Table 2). Data was acquired in selected ion monitoring (SIM) mode with an external mass calibration and collected and using Thermo ScientificTM Xcalibur data acquisition software.

Data Analysis All data was processed utilizing Thermo ScientificTM Xcalibur processing data . All chromatographic integration was accomplished using automated processing settings. No manual integration or smoothing was applied to any chromatographic data.

Figure 2. Calibration Curve for Buprenorphine and Paroxetine. Linear regression was performed using 1/x2 weighting.

Table 1. LC gradient method utilized for sample analysis.

Table 2. Mass Spectrometer settings utilized for sample analysis.

Results Quantitative Results

The main goals for any rugged quantitative assay are reaching the lowest possible limit of quantitation, providing a linear response, and maintaining good reproducibility. To evaluate the performance and reproducibility of the LC/MS method, calibration curves were analyzed with replicates of n=6. Linearity and reproducibility were calculated across the working range of the curve. The limit of quantitation (LOQ) was defined as the lowest concentration level that is both within <20% difference of the linear fit and <20% RSD for each group of replicate concentration points. Using the above criteria, the LOQ was determined to be 10pg/mL for both Buprenorphine and Paroxetine. (Figure 1) A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 100ng/mL (Figure 2 ).

© 2015 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is not intended to encourage use of these products in any manner that might infringe the intellectual property rights of others.

Figure 1. Chromatograms for Buprenorphine (left) and Paroxetine (right) at the LOQ of 10pg/mL. Target analyte is displayed at 10pg/mL in the top chromatogram. The internal standard chromatogram at 1ng/mL is displayed below. No smoothing was applied during the chromatographic peak integration and data processing procedure.

Figure 5. Zoomed view of Buprenorphine at 10pg/mL (LEFT) and Paroxetine at 10pg/mL (Right). The figure illustrates the scans measured across the peak and the corresponding mass accuracy. Mass accuracy is displayed in parts per million. (ppm)

FIGURE 6. Replicate injections of Buprenorphine (Top) and Paroxetine (bottom) at the assay LOQ of 10pg/mL. Method integration settings where implement for all chromatogram integration and no manual integrations were performed. Additionally no smoothing was applied for any chromatographic peak generation.

A linear response was observed across four orders of magnitude for Buprenorphine and Paroxetine ranging from 10pg/mL to 25,000 and 10pg/mL to 100,000pg/mL respectively (Figure 2 ).

The analytical scan speed of the instrument was evaluated at the LOQ for each sample. The number of scans across each analyte peak at the LOQ was assessed to ensure adequate scan speed in conjunction with the sample peak width produced by the chromatographic method. The chromatographic peak width for each analyte ranged between 3.6 and 4.2 seconds at the base, providing 11-15 scans for the target analytes at the LOQ (Figure 3).

.

Here we examine the scan to scan mass accuracy of the Q Exactive Focus with an external mass calibration for Buprenorphine and Paroxetine at assay LOQ of 10pg/mL (Figure 5). The mass accuracy for each scan across the analyte peaks was demonstrated to be less than 3ppm for both Buprenorphine and Paroxetine. The high level of mass accuracy from scan to scan allows the utilization of a narrow mass extraction window providing robust and reproducible results at low analyte concentration levels while in the presence of a biological matrix. Evaluation of the scan to scan mass accuracy provides a method development and troubleshooting tool that is unique to HRAM and the Orbitrap mass analyzer.

The spectra of target analytes at the LOQ can be used to verify method quality, determine the suitability of a given resolution setting, and help identify interferences that may inhibit sensitivity and reproducibility. Spectra from injections made at the LOQ for each compound were evaluated to identify unresolved interferences. At the resolution setting of 70,000 (FWHM at m/z 200) both compounds were adequately resolved from all co-eluting interferences (Figure 4). As seen in the figure, although interferences are present during the elution of each compound, the target analyte m/z was baseline resolved from nearby interferences at the 70,000 resolution setting. Additionaly the mass extraction window of +/- 5ppm used to generate the chromatographic peaks used for quantitation is illustrated in the figures. This narrow extraction window allows greater assay confidence and robustness by ensuring that signal contributing to the chromatographic peak area is resulting from the target analyte itself and not the nearby interferences. A narrow mass extraction windows and the resulting XIC do not provide analytical value without a high level of scan to scan mass accuracy across the entire analyte peak. Scan to scan mass accuracy can also offer an important indicator for indentifying potential analyte interferences as well for accessing overall assay performance. A loss in mass accuracy can provide an indication of an unresolved or co eluting interference and can be implemented as a quality control measure in validated analyses.

Figure 3. Evaluation of chromatographic peak width vs scan speed for Buprenorphine (left) and Paroxetine (right) the LOQ of 10pg/mL

Time (min)

Flow rate (uL/min) %A %B

0 500 95 5 0.1 500 95 5 0.5 500 70 30 4.5 500 67 33 5 500 5 95

6.0 500 5 95 6.1 500 95 5 8.0 500 95 5

HESI Source Settings Value MS Scan Settings Value

Spray Voltage (V) 4000 Scan Type SIM Vaporizer temperature (OC) 450 Resolution 70,000 Capillary Temperature (OC) 350 AGC Target 2.00E+04 Sheath Gas Pressure (Arb) 45 IT Fill Time (ms) 260 Aux Gas Pressure (Arb) 15 Ion Sweep Gas Pressure (Arb) 1

RT: 2.5471 - 4.0071

2.6 2.7 2.8 2.9 3.0 3.1 3.2 3.3 3.4 3.5 3.6 3.7 3.8 3.9 4.0Time (min)

0

200000

400000

600000

800000

0

1000

2000

3000

4000

5000

6000

7000

Rel

ativ

e Ab

unda

nce

RT: 2.9428

RT: 2.9294

3.0178 3.11652.8591

NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

NL:9.32E5m/z= 472.3335-472.3383 MS Curve_1_Test_4a_130

RT: 3.1978 - 4.8231

3.2 3.4 3.6 3.8 4.0 4.2 4.4 4.6 4.8Time (min)

0

200000

400000

600000

800000

1000000

0

1000

2000

3000

4000

5000

Rel

ativ

e Ab

unda

nce

RT: 3.6080

RT: 3.5811

NL:5.46E3m/z= 330.1483-330.1517 MS ICIS Curve_1_Test_4a_113

NL:1.13E6m/z= 336.1860-336.1894 MS ICIS Curve_1_Test_4a_113

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 5000 10000 15000 20000 25000pg/mL

0

5

10

15

20

Area

Rat

io

BuprenorphineY = -0.000895104+0.000754696*X R^2 = 0.9984 W: 1/X^2

0 20 40 60 80 100pg/mL

0.00

0.02

0.04

0.06

0.08

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

0 20000 40000 60000 80000 100000pg/mL

0

10

20

30

40

Area

Rat

io

ParoxetineY = -0.000226672+0.000438837*X R^2 = 0.9986 W: 1/X^2

20 40 60 80 100pg/mL

0.00

0.01

0.02

0.03

0.04

0.05

Area

Rat

io

RT: 2.7911 - 3.1104

2.80 2.82 2.84 2.86 2.88 2.90 2.92 2.94 2.96 2.98 3.00 3.02 3.04 3.06 3.08 3.10Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Relat

ive Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

RT: 3.4422 - 3.7632

3.46 3.48 3.50 3.52 3.54 3.56 3.58 3.60 3.62 3.64 3.66 3.68 3.70 3.72 3.74 3.76Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Relat

ive Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Buprenorphine Peak Width = 4.2 sec at base Scans Across Peak = 15

Paroxetine Peak Width = 4.8 sec at base Scans Across Peak = 18

The number of scans recorded across the width of each analyte peak correlates with the instrument specification of an analytical scan speed of 3Hz at the 70,000 resolution setting. Analysis of the prepared sample set at each concentration level demonstrated results that met or exceeded all of the previously described acceptance criteria indicating a high level of performance across the working range of the assay (Table 3)

Buprenorphine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 3.2% 4.4% 0 25 23.7 5.1% 1.9% 0 50 47.7 4.6% 1.5% 0

100 98.9 1.1% 1.5% 0 250 242 3.4% 1.0% 0 500 497 0.6% 0.6% 0

1000 1011 1.1% 0.7% 0 2500 2501 0.05% 0.8% 0 5000 5159 3.2% 1.0% 0

10,000 10413 4.1% 0.8% 0 25,000 25794 3.2% 0.9% 0

Paroxetine Standard

Conc.Level (pg/mL)

Mean Calc Amount

AVG %Diff

Area Ratio % CV

Excluded Replicates

10 10.3 2.7% 8.6% 0 25 23.7 5.3% 1.8% 0 50 48.4 3.3% 1.2% 0

100 101 1.3% 1.5% 0 250 245 2.1% 1.0% 0 500 507 1.4% 0.6% 0

1000 1030 3.0% 0.5% 0 2500 2474 1.0% 1.2% 0 5000 5122 2.4% 0.7% 0

10,000 10286 2.9% 0.4% 0 25,000 24702 1.2% 0.6% 0 50,000 49916 0.2% 1.2% 0

100,000 99385 0.6% 1.2% 0

Table 3. Statistical Results for Buprenorphine and Paroxetine Replicates. Note: AVG %Diff represents the average difference from the linear regression fit of the calibration curve for each set of replicates at a given concentration point.

Curve_1_Test_4a_149 #629 RT: 2.96 AV: 1 NL: 3.76E4T: FTMS + p ESI SIM ms [465.00-475.00]

468.00 468.05 468.10 468.15 468.20 468.25 468.30 468.35 468.40 468.45m/z

0.0

0.2

0.4

0.6

0.8

1.0

1.2

1.4

1.6

1.8

2.0

2.2

2.4

2.6

2.8

3.0

Rel

ativ

e Ab

unda

nce

468.2327

468.1020

468.1486

Curve_1_Test_4a_150 #774 RT: 3.65 AV: 1 NL: 4.23E4T: FTMS + p ESI SIM ms [328.00-338.00]

329.85 329.90 329.95 330.00 330.05 330.10 330.15 330.20 330.25 330.30 330.35 330.40m/z

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

Rel

ativ

e Ab

unda

nce 329.9454329.9065

330.3360

330.2271329.8745

330.2630

330.0081330.2051

330.0590

Buprenorphine – Spectral Analysis Paroxetine – Spectral Analysis

RT: 2.8875 - 3.0177

2.89 2.90 2.91 2.92 2.93 2.94 2.95 2.96 2.97 2.98 2.99 3.00 3.01Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

5500

6000

6500

7000

Rel

ativ

e Ab

unda

nce

2.9428

2.94722.9383

2.9338

2.9294 2.9519

2.9249

2.9566

2.9202

2.9613

2.9660

2.9155

2.9707

2.9108

2.9061 2.9754

2.98012.9895

NL:7.02E3m/z= 468.3085-468.3131 MS Curve_1_Test_4a_130

Δppm-2.1

Δppm-1.5

Δppm-1.8

Δppm-2.7

Δppm-1.8

Δppm-2.4 Δppm-1.5

Δppm-1.5

Δppm-1.8

Δppm-1.8

Δppm-1.5

Δppm-2.7

Δppm-0.9

Δppm-2.1

Δppm-1.8

Δppm-2.1 Δppm-0.6

RT: 3.5430 - 3.6824

3.55 3.56 3.57 3.58 3.59 3.60 3.61 3.62 3.63 3.64 3.65 3.66 3.67 3.68Time (min)

0

200

400

600

800

1000

1200

1400

1600

1800

2000

2200

2400

2600

2800

3000

3200

3400

3600

3800

4000

4200

4400

4600

4800

5000

5200

5400

Rel

ativ

e Ab

unda

nce

3.6080

3.5989

3.61273.5945

3.59003.6174

3.5856 3.6221

3.62683.5811

3.6315

3.5766

3.57223.6362

3.5675

3.6409

3.64563.56283.6503

NL:5.46E3m/z= 330.1483-330.1517 MS Curve_1_Test_4a_113

Δppm-2.7

Δppm-2.7

Δppm-1.8

Δppm-2.1

Δppm-1.8

Δppm-1.5

Δppm-0.9

Δppm -1.5

Δppm-1.8 Δppm-1.5

Δppm-1.5

Δppm-2.4 Δppm-2.1

Δppm-1.8

Δppm-2.1

Δppm-0.6

Δppm-1.8

Δppm-0.8 Δppm-1.8

Figure 4. Spectra analysis of Buprenorphine and Paroxetine at 10pg/mL and 1ng/mL in rat plasma. The exact m/z for the target analyte has been underlined in each spectra and the 5ppm mass extraction window used to generate chromatographic peaks has been outlined. Common interferences across each sample concentration have been circled. All interferences present in the sample were demonstrated to be baseline resolved at the 70,000 resolution setting.

c:\xcalibur\...\curve_1_test_4a_117 03/06/15 15:47:02 Compound Mix 0.010ng/mL

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6083 NL:5.16E3m/z= 330.14835-330.15165 MS ICIS Curve_1_Test_4a_112

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6080 NL:5.46E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_113

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

Rela

tive

Abun

danc

e

RT: 3.6000 NL:4.65E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_114

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

500

1000

1500

2000

2500

3000

3500

4000

4500

5000

Rela

tive

Abun

danc

e

RT: 3.6143 NL:5.33E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_115

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

RT: 3.6224 NL:5.89E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_116

RT: 3.0925 - 4.1000

3.2 3.4 3.6 3.8 4.0Time (min)

0

1000

2000

3000

4000

5000

Rela

tive

Abun

danc

e

NL:6.73E3m/z= 330.14835-330.15165 MS ICIS curve_1_test_4a_117

c:\xcalibur\...\curve_1_test_4a_134 03/06/15 18:19:37 Compound Mix 0.100ng/mL

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9428 NL:7.02E3m/z= 468.3085-468.3131 MS ICIS Curve_1_Test_4a_130

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9450 NL:7.69E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_131

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9631 NL:6.88E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_132

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9446 NL:7.27E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_133

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

Rela

tive

Abun

danc

e

RT: 2.9355 NL:6.57E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_134

RT: 2.4000 - 3.5000

2.4 2.6 2.8 3.0 3.2 3.4Time (min)

0

1000

2000

3000

4000

5000

6000

7000

Rela

tive

Abun

danc

e

RT: 2.9356 NL:7.89E3m/z= 468.3085-468.3131 MS ICIS curve_1_test_4a_135

Chromatographic reproducibility is an import factor for determining assay quality and ruggedness. Consistent and reproducible signal response at the assay LOQ eliminates the need for manual chromatographic peak integration and improves the overall data confidence and the assay as a whole. Here we examine the chromatographic reproducibility of Buprenorphine and Paroxetine at the assay LOQ of 10pg/mL. Each analyte was analyzed in replicates of n=6. The chromatogram for each replicate can be seen in figure 6.

Buprenorphine – 10pg/mL

Buprenorphine – 1ng/mL

Buprenorphine – 1ng/mL Zoomed

Paroxetine– 10pg/mL

Paroxetine– 1ng/mL

Paroxetine– 1ng/mL Zoomed

Africa +43 1 333 50 34 0Australia +61 3 9757 4300Austria +43 810 282 206Belgium +32 53 73 42 41Canada +1 800 530 8447China 800 810 5118 (free call domestic)

400 650 5118

Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0Finland +358 10 3292 200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591

Japan +81 45 453 9100Korea +82 2 3420 8600Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55New Zealand +64 9 980 6700Norway +46 8 556 468 00

Russia/CIS +43 1 333 50 34 0Singapore +65 6289 1190Spain +34 914 845 965Sweden +46 8 556 468 00Switzerland +41 61 716 77 00UK +44 1442 233555USA +1 800 532 4752

www.thermoscientific.com©2015 Thermo Fisher Scienti� c Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scienti� c and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scienti� c products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Speci� cations, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.