a scintillating fiber dosimeter for radiotherapy

28
1 G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007 A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY G. Bartesaghi INFN - Milano, Università degli Studi dell’Insubria - Como in collaboration with Azienda Ospedaliera Sant’Anna - Como

Upload: shad-pickett

Post on 01-Jan-2016

53 views

Category:

Documents


2 download

DESCRIPTION

A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY. G. Bartesaghi. INFN - Milano, Università degli Studi dell’Insubria - Como in collaboration with Azienda Ospedaliera Sant’Anna - Como. Radiotherapy in the cancer care Why a multichannel scintillator detector? - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

1G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

G. Bartesaghi

INFN - Milano, Università degli Studi dell’Insubria - Como

in collaboration with Azienda Ospedaliera Sant’Anna - Como

Page 2: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

2G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Outline

• Radiotherapy in the cancer care• Why a multichannel scintillator detector?• Single scintillating fiber prototype• Multifiber monodimensional detector• 2D prototype• Conclusions and outlooks

This work is part of the project PhoNeS (Photon Neutron Source) by INFN

supported by MIUR PRI 05

Page 3: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

3G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Radiotherapy in the cancer care

Data from EUROCARE 3 (European cancer registries study on cancer patients’ survival and care)

Three factors have a crucial role: - primary prevention

- earlier diagnosis

In the last decades, a constant progress has been made in the fight against cancer

- better treatment

Page 4: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

4G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

General treatment methods

- chemotherapy

- immunotherapy

- radiotherapy

- surgical removal

}

}

systemic

treatments

loco-regional

treatments

Radiotherapy exploits the effect of ionizing radiation on human tissues to eliminate tumour cells or to control their development.The radiation beam typically consists in electrons or photons (energy up to 20 MeV) emitted by a linear accelerator.

Radiotherapy is one of the most popular method in cancer treatment.Every year, more than 5 millions patients worldwide receive almost a radiotherapic treatment, exploiting more than 7500 linacs.

Page 5: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

5G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Quality assurance in Radiotherapy

For the future, the goal for radiotherapy is the same of the past

delivering a high dose in the target volume preserving the surrounding volume

All the procedures that ensure consistency of medical prescription and safe fulfilment of that prescription are part of the quality assurance.

The goal of this work is to develop of a dosimeter that, performing real time 2D measurement in the radiation beam, may simplify and accelerate the quality assurance controls.

Page 6: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

6G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Why a scintillator dosimeter?

Most popular dosimeters:

TLD small size time consumingrequiring great care

Ionization chambersand diodes

Radiographic films excellent 2D resolution long processing time not tissue equivalent

real time, accuratesingle channel not tissue equivalent

Page 7: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

7G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Why a scintillator dosimeter?

- no high voltage near the probe

A plastic multifiber scintillator dosimeter assures important advantages:

- real time measurement of 2D dose distribution

- high tissue equivalence

- a large dynamic range (100 keV – more than 20 MeV)

- excellent radiation hardness

- no pressure and temperature dependence

Page 8: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

8G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Single channel dosimeterAssembled in order to test the method feasibility

Scintillating fiber Bicron BCF-10 Material polystyrene

Diameter 1mm

Emission color blue

Emission peak 432 nm

Decay time 3.2ns

No. of photons per MeV ~ 8000

PM

TP

MT

PMT Hamamatsu R931BSpectral response 300 to 650 nm

Peak wavelength 400 nm

Num. of mult. stages 9

white fiber BCF-98 (diameter: 1mm, length: 300mm)

scintillating probe (diameter: 1mm, length: 10mm)

white fiber BCF-98 (diameter: 1mm, length: 300mm)

High energy electrons crossing a transparent medium produce Cherenkov light, which is not proportional to absorbed dose.

It is therefore necessary a twin detector, without the scintillating probe, in order to evaluate the Cherenkov contribution.

Page 9: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

9G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Linear Accelerator

Clinac Varian 2100Electron beam:

Energy: 6, 9, 12, 16, 20 MeV

Dose rate: from 100 to 600 MU/min

Field: up to 25x25 cm2

1 Monitor Unit = 1 Gy at the build up

(maximum dose region)

Photon beam:

Energy: 6, 18 MV

Dose rate: from 100 to 600 MU/min

Field: up to 40x40 cm2

Particles are emitted in bunches (100-300 Hz, depending on the dose rate)

Radiotherapy Unit

Ospedale S. Anna

Como (Italy)

Page 10: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

10G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Read out electronics

The read out electronics is based on a charge integrating ADC (12 channels), with the trigger taken from the linac itself

3s

7s

Page 11: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

11G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Treatment couch

Linac

Detector in PMMA phantom

Read out electronics1 m

Measurement set up

In order to test the scintillating dosimeter, a Depth Dose Scan is performed, the detector is inserted in a PMMA phantom

Page 12: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

12G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Our measurements are compared with data taken by a diode Scanditronix DEB101 (1.25 mm3)

Depth Dose Curves in Electron Beams

Field size: 10x10cm2

Page 13: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

13G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Depth Dose Curves in Electron Beams

Field size: 10x10cm2

Page 14: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

14G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Field size: 10x10cm2

Depth Dose Curve in Photon Beams

Page 15: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

15G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Crossline scan

In a 6 MV photon beam, the detector is moved

across the 10x10cm2 field.

Page 16: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

16G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Multifiber 1D prototype

The single channel detector demonstrates the feasibility of a scintillating fiber dosimeter easy to handle and accurate.

Next step is to develop a …

PMT

8 scintillating fibers (Bicron BCF-10, 1 cm long) inserted in a PMMA phantom are readout by a multichannel PMT.

Each fiber is coupled with a twin white one for Cherenkov subtraction.

Page 17: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

17G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

PMT Hamamatsu H8711

Spectral response 300 to 650 nm

Peak wavelength 420 nm

Electron transit time

10.9 ns

Scintillating fiber Bicron BCF-10

Material polystyrene

Diameter 1mm

Emission color blue

Emission peak 432 nm

Decay time 3.2ns

No. of photons per MeV

~ 8000

The PMT is a 16 channel, fast and low cross talk device.

Page 18: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

18G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Intensity response

The detector was positioned at different distances from the source in order to test the intensity response

Data fitted with

y = A/x2 + B

Page 19: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

19G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Reproducibility test

5 Gy are given 20 times

Distribution of the charge read by the 5 scintillating fibers

(each channel normalized to 100)

The distribution RMS is about 0.5%

Page 20: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

20G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Depth Dose Curve in Photon Beams

Field size: 10x10cm2

Page 21: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

21G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Depth Dose Curve in Electron Beams

Field size: 10x10cm2

Page 22: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

22G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Crossline scan

In a 6 MV photon beam, at 1.5 cm depth, the detector is moved across the 10x10cm2 field.

Page 23: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

23G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Gain equalization

In order to evaluate the gain factor of the different channels the detector is positioned vertically and irradiated in a 6MV photon

beam twice: from the top and from the bottom side

The photon exponential attenuation is exploited to get the gain factors

FiberRelative

gain factors1 1.000

2 0.576

3 0.368

4 0.581

5 0.976

6 0.339

Page 24: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

24G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Gain equalization

Before and … … after the equalization

Page 25: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

25G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Multifiber 2D prototype

1 cm scintillating fiber + 1 cm white fiber in every pit

8 scintillating fibers are inserted vertically in a PMMA phantom (20x20x4 cm3)

16 white fibers (35 cm long) exiting from the bottom side of the phantom carry the light to the PMT

Page 26: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

26G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Multifiber 2D prototype

Page 27: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

27G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Depth Dose Scan

Field size: 10x10cm2

Page 28: A SCINTILLATING FIBER DOSIMETER FOR RADIOTHERAPY

28G. Bartesaghi, INFN Milano, 11th Vienna Conference on Instrumentation, Feb 19-24, 2007

Conclusions …

- a scintillating fiber dosimeter can perform accurate real time measurement

- good spatial resolution

- good response at different radiation intensities

- excellent reproducibility

… and Outlooks- development of the 2D dosimeter increasing the number of channels (up to 256 or more)

- some equalization methods are under evaluation

- development of an integrated electronics (connection to PC with a USB/parallel port)