a simulation study of irregular respiratory motion and its dosimetric impact on lung tumors

Upload: xu-qin

Post on 07-Apr-2018

217 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    1/16

    A simulation study of irregular respiratory motion and its dosimetric impact on lung tumors

    This article has been downloaded from IOPscience. Please scroll down to see the full text article.

    2011 Phys. Med. Biol. 56 845

    (http://iopscience.iop.org/0031-9155/56/3/019)

    Download details:

    IP Address: 130.88.240.14

    The article was downloaded on 01/08/2011 at 16:00

    Please note that terms and conditions apply.

    View the table of contents for this issue, or go to thejournal homepage for more

    ome Search Collections Journals About Contact us My IOPscience

    http://iopscience.iop.org/page/termshttp://iopscience.iop.org/0031-9155/56/3http://iopscience.iop.org/0031-9155http://iopscience.iop.org/http://iopscience.iop.org/searchhttp://iopscience.iop.org/collectionshttp://iopscience.iop.org/journalshttp://iopscience.iop.org/page/aboutioppublishinghttp://iopscience.iop.org/contacthttp://iopscience.iop.org/myiopsciencehttp://iopscience.iop.org/myiopsciencehttp://iopscience.iop.org/contacthttp://iopscience.iop.org/page/aboutioppublishinghttp://iopscience.iop.org/journalshttp://iopscience.iop.org/collectionshttp://iopscience.iop.org/searchhttp://iopscience.iop.org/http://iopscience.iop.org/0031-9155http://iopscience.iop.org/0031-9155/56/3http://iopscience.iop.org/page/terms
  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    2/16

    IOP PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY

    Phys. Med. Biol. 56 (2011) 845859 doi:10.1088/0031-9155/56/3/019

    A simulation study of irregular respiratory motionand its dosimetric impact on lung tumors

    Y D Mutaf1, C J Scicutella, D Michalski, K Fallon, E D Brandner,

    G Bednarz and M S Huq

    Department of Radiation Oncology, University of Pittsburgh Medical Center, Pittsburgh,PA 15232, USA

    E-mail: [email protected]

    Received 5 August 2010, in final form 29 November 2010Published 17 January 2011

    Online at stacks.iop.org/PMB/56/845

    Abstract

    This study is aimed at providing a dosimetric evaluation of the irregular motion

    of lung tumors due to variations in patients respiration. Twenty-three lung

    cancer patients are retrospectively enrolled in this study. The motion of the

    patient clinical target volume is simulated and two types of irregularities

    are defined: characteristic and uncharacteristic motions. Characteristic

    irregularities are representative of random fluctuations in the observed target

    motion. Uncharacteristic irregular motion is classified as systematic errors in

    determination of the target motion during the planning session. Respiratory

    traces from measurement of patient abdominal motion are also used for the

    target motion simulations. Characteristic irregular motion was observed to

    cause minimal changes in target dosimetry with the largest effect of 2.5%

    0.9% (1) reduction in the minimum target dose (Dmin) observed for targets

    that move 2 cm on average and exhibiting 50% amplitude variations within a

    session. However, uncharacteristic irregular motion introduced more drastic

    changes in the clinical target volume (CTV) dose; 4.1% 1.7% reduction for

    1 cm motion and 9.6% 1.7% drop for 2 cm. In simulations with patients

    abdominal motion, corresponding changes in target dosimetry were observed

    to be negligible (

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    3/16

    846 Y D Mutaf et al

    technologies such as 4DCT (Rietzel et al 2005, Mutaf and Brinkmann 2008a), the changes in

    3D patient anatomy such as organ motion are measured over time. This additional knowledge

    is incorporated into the treatment planning process for the patient and used in conjunction

    with appropriate treatment delivery techniques. In modern radiotherapy, the 3D model of the

    patient anatomy is valid as long as it is sufficiently reproduced between the planning and theactual delivery stages of the radiotherapy. Similarly, the 4D model of the patient used for

    treatmentplanning purposes is only deemedvalidif the reproducibility of themotion is verified.

    Analogous to patient setup errors affecting the 3D localization of the target volume, a large

    component of error in 4D radiotherapy could be the target motion irregularities (Ozhasoglu

    and Murphy 2002, Chang et al 2007, Juhler Nottrup et al 2007) due to variations in patient

    respiration.

    Irregular breathing could be as large as the underlying respiratory motion itself.

    Respiratory variations could also have intra and inter-fractional components affecting the

    radiotherapy of mobile tumors in different ways. Using optical measurement methods, Juhler

    Nottrup etal (2007) evaluated the variations in the chest wall motion of 11 lung cancer patients

    and found out that inter-fractional variations can be up to ten times larger than the intra-

    fractional variations. Due to observed irregularities in patients respiration, Juhler Nottrupetal (2007) also concluded that margins to account for respiratory motion cannot be determined

    in one imaging session. Several techniques have been developed to mitigate the occurrence of

    irregularities in patient respiratory motion. Audio and visual feedback techniques have been

    demonstrated to be an effective way of achieving a consistent pattern of respiration. George

    et al (2006) subjected 24 lung cancer patients to audio-visual respiratory feedback techniques

    and observed significant improvement in the reproducibility of respiration. Gating technology

    (Ramsey et al 1999, Wagman et al 2003) is also used to prevent irradiation in the presence of

    drastic changes in breathing such as sneezing or coughing.

    Despite the existing studies demonstrating the amount of irregularities in respiratory

    motion, there is limited information on the dosimetric outcome of irregular breathing and

    whether or not they have a clinically significant impact on therapeutic goals of the irradiation.

    Due to radiation dispersion in tissue, penumbral effects and features of plan topology,

    geometric observations performed with radiographic imaging or optical methods are notdirectly translatable to dosimetric conclusions. In the context of patient setup errors, the

    amount of geometric miss and the associated target margin to ensure sufficient coverage are

    shown to be proportional but not the same (van Herk 2004). This is a result of differences in

    the shapes of target definition function perceived visually as a step function at the border of

    the target and the function of the dosimetric profile across the target with a gradual fall off.

    Similarly, van Herk et al (2003) analytically demonstrated that the additional target margin

    required to compensate the motion of the target is substantially smaller than its full motion

    amplitude. This was also confirmed recently by Mutaf and Brinkmann (2008b) in a more

    extensive treatment planning study incorporating actual patient data. The aim of the current

    study is to extend the dosimetric investigation of mobile targets under respiration and evaluate

    changes in target coverage due to irregularities in its motion.

    2. Methods and materials

    The dosimetric effects of respiratory motion need to be evaluated in relation to the specific

    treatment delivery techniques used for compensation of the target motion. In this study, target

    motion was incorporated into treatment planning via an expansion of the target volume by

    an amount equal to the target excursion during the full respiratory cycle (i.e. no respiratory

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    4/16

    Irregular respiratory motion 847

    Table 1. Patient target volume statistics and their anatomical location are listed. The number offields and conformity of the plan dose to the target are also shown in simple and complex geometryplans.

    Simple plans Complex plans

    Patient no Anatomic location CTV (ml) Nfields Conformity Nfields Conformity

    1 Right lung (LL) 68.7 2 4.62 6 1.79

    2 Left lung (LL) 77.4 3 2.06 5 1.82

    3 Right lung (LL) 5.6 2 2.31 7 2.40

    4 Right lung (LL) 58.2 3 3.45 5 1.85

    5 Right lung (UL) 1.4 3 2.83 6 2.63

    6 Right lung (LL) 91.6 3 4.64 8 1.85

    7 Left lung (LL) 36.9 3 2.30 7 1.93

    8 Right lung (ML) 207.6 3 2.79 6 1.74

    9 Right lung (UL) 32.9 3 2.53 5 1.98

    10 Right lung (ML) 3.3 2 2.32 5 2.37

    11 Left lung (UL) 12.0 3 2.90 6 1.87

    12 Right lung (ML) 23.7 2 9.12 6 2.02

    13 Right lung (LL) 5.1 2 9.72 5 2.23

    14 Right lung (ML) 4.9 3 8.72 5 2.71

    15 Left lung (UL) 29.0 2 8.87 6 2.00

    16 Right lung (UL) 6.0 3 4.45 5 2.08

    17 Right lung (ML) 61.8 2 4.71 5 1.82

    18 Left lung (LL) 35.6 3 2.96 6 1.93

    19 Right lung (UL) 1.6 2 18.00 5 2.30

    20 Left lung (UL) 39.2 2 5.75 5 1.89

    21 Right lung (LL) 83.8 3 2.76 7 1.65

    22 Left lung (UL) 20.9 3 3.03 6 1.98

    23 Right lung (ML) 1.8 3 2.94 7 2.58

    Average 39.5 2.6 4.95 5.8 2.06Minimum 1.4 2 2.06 5 1.65

    Maximum 207.6 3 18.00 8 2.71

    LL: lower lobe; UL: upper lobe; ML: middle lobe.

    gating). Irregularities in respiratory motion are investigated using simulations of target motion

    and analysis of respective clinical treatment plans.

    2.1. Patient cohort and dosimetry

    Twenty-three lung cancer patients previously treated in our clinic are retrospectively analyzed

    through their informed consent according to an approved University of Pittsburgh institutionalreview board protocol. Clinical target volumes (CTV) ranged from 1.4 to 207.6 ml. Although,

    prescription doses varied between 50.0 and 66.0 Gy for actual patient treatments, the

    prescription doses were all normalized to 60.0 Gy to facilitate easier comparisons for the

    effects of respiratory motion. A summary of patient information related to treatment plans

    of the patients is provided in table 1. Due to the radio-sensitivity of lungs to low doses of

    radiation (Oetzel et al 1995, Graham et al 1999), the clinical treatment plans were designed to

    minimize irradiation volume and therefore consisted of two or three fields (usually in parallel

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    5/16

    848 Y D Mutaf et al

    (a) (b)

    Figure 1. Comparison of dose distributions in simple (a) and complex (b) geometry plans in axialand coronal views for patient 12.

    opposed and oblique orientations). In this paper, such 3D conformal plans are referred to as

    simple geometry plans where dose homogeneity within the target ranged between 95% and

    115% of the prescription dose. Adopting the RTOG formulation of conformity (Shaw et al

    2000) (defined as the ratio of 95% isodose volume to the target volume), the average target

    conformity in simple plans was 4.95. In addition to simple geometry plans, an alternative

    treatment plan was also generated incorporating more fields, ranging from five to eight, in

    order to produce more conformal target dosimetry. These plans, referred as complex geometry

    plans, achieved an average conformity of 2.06 in addition to improving dose homogeneity

    within the target (95% to 105% of prescription dose). The typical dose distributions in simple

    and complex geometry plans are demonstrated in figure 1 for one of the patients (patient 12).

    The corresponding dosevolume histograms for target and bi-lateral lungs are also presented

    in figure 2. In both type of plans, the selection of beams is restricted to the angles orthogonalto the longitudinal axis of the patients (transverse co-planar arrangement). All plans were

    manually optimized to conform to the shape of the target with static 0.5 cm MLCs and a block

    margin of 0.8 cm. Hard wedges were utilized for improvement of dose homogeneity. No

    intensity modulation techniques and optimization engines (i.e. IMRT) are utilized.

    Plans were further imposed with other constraints such as a maximum tolerable spinal

    cord dose of 45.0 Gy and a mean heart dose of 40.0 Gy, although the latter constraint was

    considered on a case-by-case basis depending on the location of the tumor.

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    6/16

    Irregular respiratory motion 849

    Figure 2. Comparison of target volume and bi-lateral lungs dosevolume histograms for simple(solid lines) and complex (dashed lines) geometry plans shown in figure 1.

    2.2. Target selection and contouring

    Target definitions and contouring adhered to ICRU guidelines (International Commission on

    Radiation Units and Measurements 1993, 1999). Simulation images were acquired from

    4DCT scans using a retrospective phase sorting technique producing ten phase images at

    equal temporal intervals (i.e. 10% of respiration period). Following institutional protocols,

    treatment simulation and planning was performed on the end-of-exhalation (EOE) phase

    dataset extracted from the 4DCT scan (image in which the tumor volume is located most

    superiorly, typically 50% phase image). The EOE image was preferred for planning since in a

    typical respiratory cycle, longest amount of time is spent at this phase and therefore produces

    3D images with least amount of motion artifacts. CTV was contoured on the EOE images

    by the patients primary radiation oncologist with consideration for the subclinical extent of

    the disease. The clinical ITV was finally generated using an asymmetric expansion, internal

    margin or IM, equal to the amplitude of the target motion. In order to investigate the effects of

    different amplitudes of motions, the CTV is enlarged by a specific IM amount which is chosen

    to be equal to the amount of simulated motion and its direction (equation ( 1)):

    ITVIM = CTV IM(x,y,z). (1)

    In this formalism, ITV volumes are created by adding 1 and 2 cm margins to CTV in theinferior direction as demonstrated in figure 3. These final structures are referred as ITV1 cmand ITV2 cm within the text, where ITV1 c m = CTV1 cm and ITV2 c m = CTV2 cm..

    2.3. Respiratory motion simulations

    The mechanics of respiration induces an expansioncontraction motion observed mostly in

    the cranio-caudal (CC) axis of the patient. Several radiological studies (Seppenwoolde et al

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    7/16

    850 Y D Mutaf et al

    Figure 3. An illustration of the target volumes (CTV and ITV) described in the text. Internaltarget volumes (ITV) are generated from clinical target volumes (CTV) via application of 1 and2 cm asymmetric internal margins in the inferior direction.

    2002, Murphy et al 2002) have confirmed that largest motion occurs along the CC axis of the

    patients. Therefore, motion simulations in this study are restricted to the CC axis of the patient

    where CTV center-of-mass is changed with respect to the amount of simulated motion.

    Respiratory motion simulations are performed via dose accumulation for voxels within

    the CTV boundaries through the length of time the respiratory motion assessed. At each

    dose sampling time, the target center-of-mass was moved with respect to the instantaneousamplitude of the respiratory motion but no deformations were allowed. Therefore, the total

    number of target voxels and their coordinates within the target reference frame remain the

    same. Due to the scan resolution of the CT images, target voxels have the dimensions of

    1.00 1.00 mm2 in the axial plane and are separated by 1.25 mm slice thickness. Similarly,

    the simulated motions were also digitized with respect to the scan resolution, i.e. 1.25 mm

    since the simulated motion is in the longitudinal axis of the patient. In order to improve

    processing time, the temporal resolution of the simulations are reduced such that the target

    dose sampling is only performed at 10 equally spaced times within a respiratory cycle (i.e.

    10% phase intervals).

    Two main forms of respiration motion models are used for the simulations: regularand

    irregularrespiration. In a regular respiration study, a single respiratory cycle with a functional

    form represented in equation (2) was used for target simulations (Lujan et al 2003). A plot

    of respiratory motion amplitude in this form is shown in figure 4. For simulation of a 1.0

    cm peak-to-peak motion (A0), a 1 cm internal margin was used for CTV expansion to ITV

    (ITV1 cm), and similarly, a 2 cm internal margin was used as the target volume for a 2.0 cm

    peak-to-peak motion:

    z() = A0 cos4( ). (2)

    Irregular motion profiles used in the study were created similarly by generating 100

    respiratory cycles which follow the functional form in equation (2). However, each respiratory

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    8/16

    Irregular respiratory motion 851

    Figure 4. Plot of the functional form of respiration used in the simulations. The peak-to-peakamplitude of the motion, A0, is used as the free parameter and adjusted accordingly as describedin the text.

    cycle within a session was varied by changing its peak-to-peak amplitude (A0) sampled from a

    Gaussian distribution with a preselected mean (A) and standard deviation (A). The examples

    of simulated irregular respiration cycles and the corresponding distribution of peak-to-peak

    amplitudes are shown in figure 5.

    Irregular respiratory motion simulations are further divided into two categories.

    Characteristic or random irregular motion where the irregularities in the respiratory

    motion during treatment are distributed randomly about the motion amplitude attributed

    to the CTV during the 4DCT imaging.

    Uncharacteristic or systematically irregular motion where there is a systematic shift

    between CTV motion amplitude determined at the time of 4DCT imaging and the actual

    respiratory behavior of the patient during treatment.

    In characteristic motion, the meanof the Gaussiandistribution for peak-to-peak amplitudes

    is chosen to be equal to the internal margin. For example, a set of respiratory cycles distributed

    about a mean of 1.0 cm (e.g. figure 5, upper panel) is characteristic for the simulations of CTV

    motion using the plan generated with a 1.0 cm internal margin (ITV1 cm). In uncharacteristic

    motion simulations, however, the internal margin between the CTV and ITV is smaller thanthe mean of the respiratory cycle amplitudes by 1 standard deviation. An example to such

    respiratory motion would be a set of respiratory cycles distributed according to a mean of 1.5

    cm and a standard deviation of 0.5 cm (e.g. figure 5, lower panel) used for the simulations

    with only a 1.0 cm internal margin (ITV1 cm).

    A characteristic irregular motion would, therefore, be the best-case scenario where the

    irregularities are essentially random in nature and CTV motion as observed in the simulation

    (e.g. with 4DCT) corresponds to the average target motion. In contrast, the uncharacteristic

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    9/16

    852 Y D Mutaf et al

    Figure 5. A time series plot of the simulated irregular respiration used in the study for the meanpeak-to-peak amplitude values (A) of 1.0cm (upperpanel) and1.5 cm (lowerpanel). A histogramof the cycle peak-to-peak amplitudes is plotted next to the corresponding plot of the respirationdata. Althoughthe continuous form of the full respiration datais shown (solid line), the simulationsare performed for respiration amplitudes sampled at phase intervals of 10% (filled circles). Thesampled motion is also digitized by 1.25 mm equal to the slice thickness of the CT images.

    irregular motion reflects a more extreme case where in addition to random variations in

    respiration, the determined CTV motion and internal margin are compromised by a systematic

    difference. This systematic error occurs when the amount of motion determined at the planning

    stage is consistently less than the average target motion observed during patients treatment.Characteristic and uncharacteristic irregular motion simulation parameters are listed

    below.

    Simulations with ITV1 cmA = 1.00 cm, A = 0.25 cm or 25% (characteristic)

    A = 1.00 cm, A = 0.50 cm or 50% (characteristic)

    A = 1.25 cm, A = 0.25 cm (uncharacteristic)

    A = 1.50 cm, A = 0.50 cm (uncharacteristic).

    Simulations with ITV2 cmA = 2.00 cm, A = 0.50 cm or 25% (characteristic)

    A = 2.00 cm, A = 1.00 cm or 50% (characteristic)

    A = 2.50 cm, A = 0.50 cm (uncharacteristic)

    A = 3.00 cm, A = 1.00 cm (uncharacteristic).These a priori choices of the mean and the standard value of the respiration amplitude

    distribution are intended to provide several threshold levels for their influence on target

    dosimetry rather than exact representations of what is observed with actual patients. In this

    respect, 25% variations as used in this study stand as a close approximation to what we

    observed within our patient population as well as what is reported by other investigators

    (Britton et al 2007). However, amplitude variations as large as 50% are not typically reported

    in the literature and therefore constitute a limiting case.

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    10/16

    Irregular respiratory motion 853

    Figure 6. Respiratory motion trace of a patient (abdomen motion). The average peak inhale andexhale amplitudes are emphasized with blue and red horizontal lines, respectively.

    In addition to the functional forms of irregular respiratory motion as described above,

    patients actual respiratory waveforms were also used for target motion simulations; however,

    these data were available only for a subset of patients, 13 out of 23 patients studied here.The respiratory signal represented the vertical motion of the abdominal surface recorded with

    the aid of reflective markers placed on patients abdomen and an infrared camera as available

    in a commercial respiratory management system (RPM, Varian Inc.). In clinical 4DCT

    procedures, these patients were subjected to audio-coaching feedback methods for improving

    the reproducibility of their respiratory pattern (Kini et al 2003, George et al 2006, Haasbeek

    et al 2008).

    The abdominal motion is used as surrogate for the actual tumor motion (Rietzel et al

    2005, Mutaf and Brinkmann 2008a) and has been shown to have a high correlation with the

    actual tumor motion in lung cancer patients (Ozhasoglu and Murphy 2002, Starkschall et al

    2004, Gierga et al 2005). However, this correlation is only relevant for the relative amounts

    of surrogate and target motions and therefore the surrogate motion is scaled in the simulations

    such that the average respiratory cycle from the entire surrogate waveform was normalized

    to selected peak-to-peak amplitudes. As described earlier, this normalization was chosen as

    1.0 cm for the average peak-to-peak motion of the surrogate data in simulations for plans

    with ITV1 cm and 2.0 cm for simulations with ITV2 cm. An example of patient respiration

    waveform is demonstrated in figure 6. The average respiration cycle data are used for

    regular motion simulation; however, the full respiration waveform is used for irregular motion

    simulations.

    In contrast to the synthesized respiratory motion data used for characteristic and

    uncharacteristic irregular motion simulations, the actual respiratory trace of the patients

    provides a more diverse spectrum of irregularities in breathing. While synthesized respiratory

    simulations only represent variations in inhale amplitude, actual motion data also include such

    irregular motion as variations in end-of-exhale (a.k.a. baseline-drift) as well as the presence

    of outlier respiratory cycles. However, due to the use of single session data, the simulations of

    irregular motion with patient motion traces only represent intra-fractional variations excludingpossible inter-fractional changes in breathing.

    2.4. Plan evaluation metrics

    Several metrics are utilized to assess the effect of irregular respiratory motion on dosimetric

    coverage for CTV. Respiratory motion mostly affects the cells at the margins of the target

    causing them to move outside the boundaries of the irradiation field and receive less than

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    11/16

    854 Y D Mutaf et al

    planned dose. Therefore, the target minimum dose (Dmin) is expected to be the most sensitive

    metric for dosimetric evaluation of structures under motion. In addition, global dosimetric

    evaluators such as the equivalent uniform dose (EUD) (Niemierko 1997, Choi and Deasy

    2002) and tumor control probability (TCP) (Webb and Nahum 1993, Okunieffet al 1995) are

    used for evaluations.The EUD is defined such that if a target is irradiated uniformly with this dose, it will

    result in the same clonogenic cell killing as the actual inhomogeneous dose distribution across

    the target volume. Assuming a homogeneous density of tumor clonogens within the target

    volume and ignoring cell repair and repopulation effects, Niemierkos (1997) EUD formalism

    was adopted for this study as shown in equation (3):

    EUD = Dref ln

    1N

    Ni=1 (SF2)Di/Dref

    ln(SF2). (3)

    In equation (3), the reference dose, Dref, is defined as the reference dose per fraction and

    usually taken as 2 Gy, whereas SF2, the surviving fraction of clonogens at 2 Gy, is set to 0.5.

    Using these parameters, the average EUD values were 61.1 Gy (range 59.7 to 63.1 Gy) and

    60.4 Gy (range 59.5 to 61.2 Gy) for simple and complex plans, respectively (note that the all

    prescription doses have been normalized to 60 Gy in all patient plans for the purpose of thisstudy).

    TCP is also calculated to assist the dosimetric evaluations for the target and assess any

    dosimetric changedue to irregular respiratory motionin terms of itsrelative role in undermining

    local control, if any. TCP is defined as the Poisson probability for achieving complete tumor

    cell lethality (zero survival probability). Analysis of dose-response data from population-

    based studies results in several empirical formulations for TCP and we implemented the logit

    form as described in Okunieffet al (1995) and shown in equation (4):

    TCP =e

    450DD50D50

    1 + e

    450DD50D50

    . (4)

    Tumor control probability has a sigmoid shape which is modeled by the two free parameters

    D50 and 50, dose to achieve 50% control probability and the slope of the sigmoid curve atD50, respectively. Following the multi-institutional analysis of the dose response data for

    macroscopic lung tumors of Okunieff et al we adopted a D50 value of 51.2 Gy and 50 of

    0.83. For this set of parameters, the average TCP values were 0.66 (range 0.640.69) and 0.65

    (range 0.630.66) for simple and complex plans, respectively.

    It is not the intent of the current study to evaluate the validity of these treatment plan

    metrics or their selected parameters. These metrics are solely used to facilitate comparisons

    between target dosimetry under irregular respiratory motion as compared to its regular motion.

    Therefore, these metrics are only interpreted with respect to their relative values within

    corresponding comparisons. Furthermore, we define a figure of relativity for each metric

    discussed above that is equal to the ratio of the same metric for the target under irregular motion

    to that under regular motion. This is formulated in equation (5). Following this approach, we

    define rel-Dmin, rel-EUD and rel-TCP metrics for the targets and motions studied here:

    rel-M= Mirreg/Mreg. (5)

    3. Results

    The result of characteristic irregular motion simulations with 1.0 and 2.0 cm amplitudes are

    displayed in figure 7 for all 23 patients. The changes in the target minimum dose, represented

    by rel-Dmin, are also shown for simple and complex plans.

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    12/16

    Irregular respiratory motion 855

    Figure 7. Relative target minimum dose (rel-Dmin) results of characteristic irregular motionsimulations for all patients; mean motion amplitudes of 1 cm (upper plot) and 2 cm (lower panel)are shown.

    In characteristic motion simulations with a 1 cm average amplitude and 25% standard

    deviation, the maximum change in the minimum dose to CTV was 0.1% for both simple

    and complex plans. However no change in EUD or TCP was observed. A change in target

    dose was slightly larger when the variations in respiratory motion increased to 50% standard

    deviation with a maximum reduction of 0.6% in the minimum target dose for simple and 0.8%

    for complex plans. For this motion, maximum changes in both EUD and TCP were 0.2% in

    both simple and complex plans among all patients.

    When the motion amount was increased to 2 cm in amplitude, changes in all dosimetry

    metrics for CTV were larger for both 25% and 50% variations. The maximum change in the

    minimum dose to a target was 3.7% in the case of 50% variations. For 50% variations,

    the maximum EUD reduction was 2.1% and the TCP reduction was 3.0%. No substantial

    difference between simple and complex plan dosimetry was observed.Target dosimetry for uncharacteristic motion is also presented in comparison with the

    characteristic motion in figure 8 shown only for complex geometry plans. The effect

    of uncharacteristic motion on simple plans again was not substantially different than the

    complex plans. For plans with 1.0 and 2.0 cm internal margins in all patients, CTV dose

    was compromised more drastically in uncharacteristic irregular motion simulations compared

    to characteristic motion. The largest change in target dosimetry was observed for plan

    with a 2.0 cm margin and associated uncharacteristic irregular motion ( = 3.0 cm,

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    13/16

    856 Y D Mutaf et al

    Figure 8. Comparison of relative target minimum dose (rel-Dmin) in irregular motion simulationsfor characteristic and uncharacteristic motion traces; simulations with 1 cm (upper panel) and2 cm (lower panel) internal margins are shown.

    = 1.0 cm) such that the minimum target dose was reduced by 13.6% as compared to

    the regular motion.

    Simulations of respiratory motion using the actual abdominal (surrogate) motion of

    patients were also repeated for both 1.0 and 2.0 cm average motion amplitudes. The maximum

    change in the target minimum dose was 0.1% in both cases using complex plans. The

    maximum reduction in EUD was 0.3% and in TCP was 0.5% observed only with a 2 cm

    motion. Simple plans yielded similar results.

    The ranges of target dosimetry metrics among all the patients for all simulated motions

    are summarized in table 2.

    4. Discussion

    In this study, an evaluation of the dosimetric effects of irregular motion was presented

    for 23 lung cancer patients. Several types of irregular motion simulations are performed

    to assess target dosimetry under various irregular motion conditions. In the process, the

    interplay between the plan topology as demonstrated by simple and complex three-dimensional

    dosimetry and the presence of irregular respiratory motion was also investigated.

    Characteristic irregular motion type representative of random variations in the determined

    (i.e.planned) target motion caused non-substantive dosimetric effects. In mostsimulated cases,

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    14/16

    Irregular respiratory motion 857

    Table 2. Range of target dosimetry evaluation metrics for the patient population investigated inthis study. Relative metrics as tabulated here are defined in equation (5).

    Characteristic motion (IM = ) rel-Dmin rel-EUD rel-TCP

    = 1.0 cm, = 0.25 cm (IM = 1.0 cm) 99.9100.0% 100.0% 100.0% = 1.0 cm, = 0.50 cm (IM = 1.0 cm) 99.2100.0% 99.8100.0% 99.8100.0%

    = 2.0 cm, = 0.5 cm (IM = 2.0 cm) 99.5100.0% 99.8100.0% 99.7100.0%

    = 2.0 cm, = 1.0 cm (IM = 2.0 cm) 96.299.6% 97.999.8% 97.099.7%

    Un-characteristic motion (IM = )

    = 1.25 cm, = 0.25 cm (IM = 1.0 cm) 99.0100.0% 99.8100.0% 99.8100.0%

    = 1.50 cm, = 0.50 cm (IM = 1.0 cm) 91.699.3% 98.299.9% 98.199.8%

    = 2.5 cm, = 0.5 cm (IM = 2.0 cm) 95.8100.0% 98.799.9% 98.499.9%

    = 3.0 cm, = 1.0 cm (IM = 2.0 cm) 86.493.6% 92.799.0% 90.899.0%

    Patient respiratory motion

    Normalized to 1 cm (IM = 1.0 cm) 99.9100.0% 99.9100.0% 99.9100.0%

    Normalized to 2 cm (IM = 2.0 cm) 99.0100.0% 99.7100.0% 99.5 100.0%

    the change in the target minimum dose was clinically insignificant (less than 2%) to require

    intervention or other compensation techniques. In the extreme scenario where the simulated

    irregular motion had a mean of 2 cm and showed 1 cm amplitude variations (at 1 standard

    deviation), the minimum target dose was dropped by 2.5% 0.9%. Although minimal, these

    observations confirm the expectation that the dosimetric effect of characteristic irregularities

    increases as the average amount of motion and its relative variations increases.

    However, there is no assurance that the target motion as determined during patient

    simulation (e.g. 4DCT) would be reflective of the average respiratory motion for the entire

    course of treatment. For normally distributed motion data, there is a 50% probability that the

    internal margin amount determined at simulation will be smaller than the average amplitude of

    the motion. Dosimetric consequences for such scenarios are investigated in this study via the

    simulations of uncharacteristic irregular motion where the chosen internal margin for the target

    is smaller than the mean of the entire peak amplitude distribution by 1 standard deviation.

    Uncharacteristic irregular motion resulted in clinically significant dosimetric

    consequences with minimum CTV doses reduced about 8.4% for a 1.0 cm internal margin

    (motion = 1.5 cm, = 0.5 cm) and up to 13.6% for a 2.0 cm internal margin (motion =

    3.0 cm, = 1.0 cm). This observation signifies that such systematic discrepancies between

    the determined target motion and the entire respiratory motion have more impact than the

    random variations in respiration. In fractionated radiotherapy, an example of this discrepancy

    could appear as inter-fractional variations of the respiratory motion between the simulation

    session where target motion is determined and subsequent treatment sessions.

    Results from the simulation of actual patient respiratory waveforms indicated that therespiration motion irregularities demonstrated by these patients did not cause detrimental

    dosimetric consequences. This parallels our observations with characteristic irregular motion.

    The similarity between the actual patient respiration motion and generated characteristic

    irregular motion simulations could be attributed to the fact that the patient respiratory data

    were collected with the help of audio-coaching techniques. Since coaching compels the patient

    to reproduce a respiratory pattern, the breathing irregularities are consequentially restricted

    to random fluctuations. Therefore, it can be argued that the coaching techniques help avoid

  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    15/16

    858 Y D Mutaf et al

    the occurrence of uncharacteristic respiratory motion which was observed to result in largest

    degradation for target dosimetry.

    5. Conclusion

    Determination of respiratory motion is a necessary first step for the management of respiratory

    motion in radiotherapy. The process also requires that determined respiratory motion is valid

    throughout the full course of patients treatment. However, this validity is compromised

    due to the varying degrees of breathing irregularities. In this study, we assessed the clinical

    implications and dosimetric consequences of these variations and categorized different types

    of breathing irregularities. Characteristic motion irregularities are not observed to have a

    considerable clinical effect on target dosimetry. Similarly, simulations of target motion using

    actual patient respiration data have not produced a clinically substantial dosimetric effect.

    However, uncharacteristic irregularities are shown to have critical dosimetric consequences

    for the target coverage. Coaching techniques are suggested as potential ways to avoid such

    inconsistencies and maintain a common baseline for patients respiration throughout the course

    of radiotherapy. Additionally, there was no substantial difference between the simple andcomplex treatment plan geometries incorporating different levels of target dose homogeneity

    and conformity.

    References

    Britton K R, Starkschall G, Tucker S L, Pan T, Nelson C, Chang J Y, Cox J D, Mohan R and Komaki R 2007

    Assessment of gross tumor volume regression and motion changes during radiotherapy for non-small-cell lung

    cancer as measured by four-dimensional computed tomography Int. J. Radiat. Oncol. Biol. Phys. 68 103646

    ChangJ etal 2007Observationof interfractional variations in lungtumor position using respiratory gated and ungated

    megavoltage cone-beam computed tomography Int. J. Radiat. Oncol. Biol. Phys. 67 154858

    Choi B and Deasy J O 2002 The generalized equivalent uniform dose function as a basis for intensity-modulated

    treatment planning Phys. Med. Biol. 47 357989

    George R, Chung T D, Vedam S S, Ramakrishnan V, Mohan R, Weiss E and Keall P J 2006 Audio-visual biofeedback

    for respiratory-gated radiotherapy: impact of audio instruction and audio-visual biofeedback on respiratory-

    gated radiotherapy Int. J. Radiat. Oncol. Biol. Phys. 65 92433

    Gierga D P, Brewer J, Sharp G C, Betke M, Willett C G and Chen G T 2005 The correlation between internal

    and external markers for abdominal tumors: implications for respiratory gating Int. J. Radiat. Oncol. Biol.

    Phys. 61 15518

    Graham M V, Purdy J A, Emami B, Harms W, Bosch W, Lockett M A and Perez C A 1999 Clinical dose-volume

    histogram analysis for pneumonitis after 3D treatment for non-small cell lung cancer (NSCLC) Int. J. Radiat.

    Oncol. Biol. Phys. 45 3239

    Haasbeek C J, Spoelstra F O, Lagerwaard F J, van Sornsen de Koste J R, Cuijpers J P, Slotman B J and Senan S 2008

    Impact of audio-coaching on the position of lung tumors Int. J. Radiat. Oncol. Biol. Phys. 71 111823

    International Commission on Radiation Units and Measurements 1993 Prescribing, recording, and reporting photon

    beam therapy ICRU Report 50 (Bethesda, MD: ICRU)

    International Commission on Radiation Units and Measurements 1999 Prescribing, recording, and reporting photon

    beam therapy ICRU Report 62 (Bethesda, MD: ICRU)

    Juhler Nottrup T, Korreman S S, Pedersen A N, Aarup L R, Nystrom H, Olsen M and Specht L 2007 Intra- and

    interfraction breathing variations during curative radiotherapy for lung cancer Radiother. Oncol. 84 408

    Kini V R, Vedam S S, Keall P J, Patil S, Chen C and Mohan R 2003 Patient training in respiratory-gated radiotherapy

    Med. Dosim. 28 711

    Lujan A E, Balter J M and Ten Haken R K 2003 A method for incorporating organ motion due to breathing into 3D

    dose calculations in the liver: sensitivity to variations in motion Med. Phys. 30 26439

    Murphy M J, Martin D, Whyte R, HaiJ, OzhasogluC andLe Q T 2002 Theeffectiveness of breath-holding to stabilize

    lung and pancreas tumors during radiosurgery Int. J. Radiat. Oncol. Biol. Phys. 53 47582

    Mutaf Y D and Brinkmann D H 2008a An investigation of temporal resolution parameters in cine-mode four-

    dimensional computed tomography acquisition J. Appl. Clin. Med. Phys. 9 2819

    http://dx.doi.org/10.1016/j.ijrobp.2007.01.021http://dx.doi.org/10.1016/j.ijrobp.2007.01.021http://dx.doi.org/10.1016/j.ijrobp.2007.01.021http://dx.doi.org/10.1016/j.ijrobp.2006.11.055http://dx.doi.org/10.1016/j.ijrobp.2006.11.055http://dx.doi.org/10.1016/j.ijrobp.2006.11.055http://dx.doi.org/10.1088/0031-9155/47/20/302http://dx.doi.org/10.1088/0031-9155/47/20/302http://dx.doi.org/10.1088/0031-9155/47/20/302http://dx.doi.org/10.1016/j.ijrobp.2006.02.035http://dx.doi.org/10.1016/j.ijrobp.2006.02.035http://dx.doi.org/10.1016/j.ijrobp.2006.02.035http://dx.doi.org/10.1016/j.ijrobp.2004.12.013http://dx.doi.org/10.1016/j.ijrobp.2004.12.013http://dx.doi.org/10.1016/j.ijrobp.2004.12.013http://dx.doi.org/10.1016/S0360-3016(99)00183-2http://dx.doi.org/10.1016/S0360-3016(99)00183-2http://dx.doi.org/10.1016/S0360-3016(99)00183-2http://dx.doi.org/10.1016/j.ijrobp.2007.11.061http://dx.doi.org/10.1016/j.ijrobp.2007.11.061http://dx.doi.org/10.1016/j.ijrobp.2007.11.061http://dx.doi.org/10.1016/j.radonc.2007.05.026http://dx.doi.org/10.1016/j.radonc.2007.05.026http://dx.doi.org/10.1016/j.radonc.2007.05.026http://dx.doi.org/10.1016/S0958-3947(02)00136-Xhttp://dx.doi.org/10.1016/S0958-3947(02)00136-Xhttp://dx.doi.org/10.1016/S0958-3947(02)00136-Xhttp://dx.doi.org/10.1118/1.1609057http://dx.doi.org/10.1118/1.1609057http://dx.doi.org/10.1118/1.1609057http://dx.doi.org/10.1016/S0360-3016(01)02822-Xhttp://dx.doi.org/10.1016/S0360-3016(01)02822-Xhttp://dx.doi.org/10.1016/S0360-3016(01)02822-Xhttp://dx.doi.org/10.1120/jacmp.v9i4.2819http://dx.doi.org/10.1120/jacmp.v9i4.2819http://dx.doi.org/10.1120/jacmp.v9i4.2819http://dx.doi.org/10.1120/jacmp.v9i4.2819http://dx.doi.org/10.1016/S0360-3016(01)02822-Xhttp://dx.doi.org/10.1118/1.1609057http://dx.doi.org/10.1016/S0958-3947(02)00136-Xhttp://dx.doi.org/10.1016/j.radonc.2007.05.026http://dx.doi.org/10.1016/j.ijrobp.2007.11.061http://dx.doi.org/10.1016/S0360-3016(99)00183-2http://dx.doi.org/10.1016/j.ijrobp.2004.12.013http://dx.doi.org/10.1016/j.ijrobp.2006.02.035http://dx.doi.org/10.1088/0031-9155/47/20/302http://dx.doi.org/10.1016/j.ijrobp.2006.11.055http://dx.doi.org/10.1016/j.ijrobp.2007.01.021
  • 8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors

    16/16

    Irregular respiratory motion 859

    Mutaf Y D and Brinkmann D H 2008b Optimization of internal margin to account for dosimetric effects of respiratory

    motion Int. J. Radiat. Oncol. Biol. Phys. 70 156170

    Niemierko A 1997 Reporting and analyzing dose distributions: a concept of equivalent uniform dose Med.

    Phys. 24 10310

    Oetzel D, Schraube P, Hensley F, Sroka-Perez G, Menke M and Flentje M 1995 Estimation of pneumonitis risk

    in three-dimensional treatment planning using dose-volume histogram analysis Int. J. Radiat. Oncol. Biol.

    Phys. 33 45560

    Okunieff P, Morgan D, Niemierko A and Suit H D 1995 Radiation dose-response of human tumors Int. J. Radiat.

    Oncol. Biol. Phys. 32 122737

    Ozhasoglu C and Murphy M J 2002 Issues in respiratory motion compensation during external-beam radiotherapy

    Int. J. Radiat. Oncol. Biol. Phys. 52 138999

    Ramsey C R, Scaperoth D, Arwood D and Oliver A L 1999 Clinical efficacy of respiratory gated conformal radiation

    therapy Med. Dosim. 24 1159

    Rietzel E, Pan T and Chen G T 2005 Four-dimensional computed tomography: image formation and clinical protocol

    Med. Phys. 32 87489

    Seppenwoolde Y, Shirato H, Kitamura K, Shimizu S, van Herk M, Lebesque J V and Miyasaka K 2002 Precise and

    real-timemeasurement of 3D tumor motion in lungdue to breathingand heartbeat, measuredduring radiotherapy

    Int. J. Radiat. Oncol. Biol. Phys. 53 82234

    Shaw E, Scott C, Souhami L, Dinapoli R, Kline R, Loeffler J and Farnan N 2000 Single dose radiosurgical treatment

    of recurrent previously irradiated primary brain tumors and brain metastases: final report of RTOG protocol

    9005 Int. J. Radiat. Oncol. Biol. Phys. 47 2918

    Starkschall G, Forster K M, Kitamura K, Cardenas A, Tucker S L and Stevens C W 2004 Correlation of gross tumor

    volume excursion with potential benefits of respiratory gating Int. J. Radiat. Oncol. Biol. Phys. 60 12917

    van Herk M 2004 Errors and margins in radiotherapy Semin. Radiat. Oncol. 14 5264

    van Herk M, Witte M, van der Geer J, Schneider C and Lebesque J V 2003 Biologic and physical fractionation effects

    of random geometric errors Int. J. Radiat. Oncol. Biol. Phys. 57 146071

    Wagman R, Yorke E, Ford E, Giraud P, Mageras G, Minsky B and Rosenzweig K 2003 Respiratory gating for liver

    tumors: use in dose escalation Int. J. Radiat. Oncol. Biol. Phys. 55 65968

    WebbS andNahum A E 1993 A modelfor calculating tumourcontrol probability in radiotherapyincluding theeffects

    of inhomogeneous distributions of dose and clonogenic cell density Phys. Med. Biol. 38 65366

    http://dx.doi.org/10.1016/j.ijrobp.2007.12.025http://dx.doi.org/10.1016/j.ijrobp.2007.12.025http://dx.doi.org/10.1016/j.ijrobp.2007.12.025http://dx.doi.org/10.1118/1.598063http://dx.doi.org/10.1118/1.598063http://dx.doi.org/10.1118/1.598063http://dx.doi.org/10.1016/0360-3016(95)00009-Nhttp://dx.doi.org/10.1016/0360-3016(95)00009-Nhttp://dx.doi.org/10.1016/0360-3016(95)00009-Nhttp://dx.doi.org/10.1016/0360-3016(94)00475-Zhttp://dx.doi.org/10.1016/0360-3016(94)00475-Zhttp://dx.doi.org/10.1016/0360-3016(94)00475-Zhttp://dx.doi.org/10.1016/S0360-3016(01)02789-4http://dx.doi.org/10.1016/S0360-3016(01)02789-4http://dx.doi.org/10.1016/S0360-3016(01)02789-4http://dx.doi.org/10.1016/S0958-3947(99)00006-0http://dx.doi.org/10.1016/S0958-3947(99)00006-0http://dx.doi.org/10.1016/S0958-3947(99)00006-0http://dx.doi.org/10.1118/1.1869852http://dx.doi.org/10.1118/1.1869852http://dx.doi.org/10.1118/1.1869852http://dx.doi.org/10.1016/S0360-3016(02)02803-1http://dx.doi.org/10.1016/S0360-3016(02)02803-1http://dx.doi.org/10.1016/S0360-3016(02)02803-1http://dx.doi.org/10.1016/S0360-3016(99)00507-6http://dx.doi.org/10.1016/S0360-3016(99)00507-6http://dx.doi.org/10.1016/S0360-3016(99)00507-6http://dx.doi.org/10.1016/j.ijrobp.2004.07.707http://dx.doi.org/10.1016/j.ijrobp.2004.07.707http://dx.doi.org/10.1016/j.ijrobp.2004.07.707http://dx.doi.org/10.1053/j.semradonc.2003.10.003http://dx.doi.org/10.1053/j.semradonc.2003.10.003http://dx.doi.org/10.1053/j.semradonc.2003.10.003http://dx.doi.org/10.1016/j.ijrobp.2003.08.026http://dx.doi.org/10.1016/j.ijrobp.2003.08.026http://dx.doi.org/10.1016/j.ijrobp.2003.08.026http://dx.doi.org/10.1016/S0360-3016(02)03941-Xhttp://dx.doi.org/10.1016/S0360-3016(02)03941-Xhttp://dx.doi.org/10.1016/S0360-3016(02)03941-Xhttp://dx.doi.org/10.1088/0031-9155/38/6/001http://dx.doi.org/10.1088/0031-9155/38/6/001http://dx.doi.org/10.1088/0031-9155/38/6/001http://dx.doi.org/10.1088/0031-9155/38/6/001http://dx.doi.org/10.1016/S0360-3016(02)03941-Xhttp://dx.doi.org/10.1016/j.ijrobp.2003.08.026http://dx.doi.org/10.1053/j.semradonc.2003.10.003http://dx.doi.org/10.1016/j.ijrobp.2004.07.707http://dx.doi.org/10.1016/S0360-3016(99)00507-6http://dx.doi.org/10.1016/S0360-3016(02)02803-1http://dx.doi.org/10.1118/1.1869852http://dx.doi.org/10.1016/S0958-3947(99)00006-0http://dx.doi.org/10.1016/S0360-3016(01)02789-4http://dx.doi.org/10.1016/0360-3016(94)00475-Zhttp://dx.doi.org/10.1016/0360-3016(95)00009-Nhttp://dx.doi.org/10.1118/1.598063http://dx.doi.org/10.1016/j.ijrobp.2007.12.025