a simulation study of irregular respiratory motion and its dosimetric impact on lung tumors
TRANSCRIPT
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
1/16
A simulation study of irregular respiratory motion and its dosimetric impact on lung tumors
This article has been downloaded from IOPscience. Please scroll down to see the full text article.
2011 Phys. Med. Biol. 56 845
(http://iopscience.iop.org/0031-9155/56/3/019)
Download details:
IP Address: 130.88.240.14
The article was downloaded on 01/08/2011 at 16:00
Please note that terms and conditions apply.
View the table of contents for this issue, or go to thejournal homepage for more
ome Search Collections Journals About Contact us My IOPscience
http://iopscience.iop.org/page/termshttp://iopscience.iop.org/0031-9155/56/3http://iopscience.iop.org/0031-9155http://iopscience.iop.org/http://iopscience.iop.org/searchhttp://iopscience.iop.org/collectionshttp://iopscience.iop.org/journalshttp://iopscience.iop.org/page/aboutioppublishinghttp://iopscience.iop.org/contacthttp://iopscience.iop.org/myiopsciencehttp://iopscience.iop.org/myiopsciencehttp://iopscience.iop.org/contacthttp://iopscience.iop.org/page/aboutioppublishinghttp://iopscience.iop.org/journalshttp://iopscience.iop.org/collectionshttp://iopscience.iop.org/searchhttp://iopscience.iop.org/http://iopscience.iop.org/0031-9155http://iopscience.iop.org/0031-9155/56/3http://iopscience.iop.org/page/terms -
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
2/16
IOP PUBLISHING PHYSICS IN MEDICINE AND BIOLOGY
Phys. Med. Biol. 56 (2011) 845859 doi:10.1088/0031-9155/56/3/019
A simulation study of irregular respiratory motionand its dosimetric impact on lung tumors
Y D Mutaf1, C J Scicutella, D Michalski, K Fallon, E D Brandner,
G Bednarz and M S Huq
Department of Radiation Oncology, University of Pittsburgh Medical Center, Pittsburgh,PA 15232, USA
E-mail: [email protected]
Received 5 August 2010, in final form 29 November 2010Published 17 January 2011
Online at stacks.iop.org/PMB/56/845
Abstract
This study is aimed at providing a dosimetric evaluation of the irregular motion
of lung tumors due to variations in patients respiration. Twenty-three lung
cancer patients are retrospectively enrolled in this study. The motion of the
patient clinical target volume is simulated and two types of irregularities
are defined: characteristic and uncharacteristic motions. Characteristic
irregularities are representative of random fluctuations in the observed target
motion. Uncharacteristic irregular motion is classified as systematic errors in
determination of the target motion during the planning session. Respiratory
traces from measurement of patient abdominal motion are also used for the
target motion simulations. Characteristic irregular motion was observed to
cause minimal changes in target dosimetry with the largest effect of 2.5%
0.9% (1) reduction in the minimum target dose (Dmin) observed for targets
that move 2 cm on average and exhibiting 50% amplitude variations within a
session. However, uncharacteristic irregular motion introduced more drastic
changes in the clinical target volume (CTV) dose; 4.1% 1.7% reduction for
1 cm motion and 9.6% 1.7% drop for 2 cm. In simulations with patients
abdominal motion, corresponding changes in target dosimetry were observed
to be negligible (
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
3/16
846 Y D Mutaf et al
technologies such as 4DCT (Rietzel et al 2005, Mutaf and Brinkmann 2008a), the changes in
3D patient anatomy such as organ motion are measured over time. This additional knowledge
is incorporated into the treatment planning process for the patient and used in conjunction
with appropriate treatment delivery techniques. In modern radiotherapy, the 3D model of the
patient anatomy is valid as long as it is sufficiently reproduced between the planning and theactual delivery stages of the radiotherapy. Similarly, the 4D model of the patient used for
treatmentplanning purposes is only deemedvalidif the reproducibility of themotion is verified.
Analogous to patient setup errors affecting the 3D localization of the target volume, a large
component of error in 4D radiotherapy could be the target motion irregularities (Ozhasoglu
and Murphy 2002, Chang et al 2007, Juhler Nottrup et al 2007) due to variations in patient
respiration.
Irregular breathing could be as large as the underlying respiratory motion itself.
Respiratory variations could also have intra and inter-fractional components affecting the
radiotherapy of mobile tumors in different ways. Using optical measurement methods, Juhler
Nottrup etal (2007) evaluated the variations in the chest wall motion of 11 lung cancer patients
and found out that inter-fractional variations can be up to ten times larger than the intra-
fractional variations. Due to observed irregularities in patients respiration, Juhler Nottrupetal (2007) also concluded that margins to account for respiratory motion cannot be determined
in one imaging session. Several techniques have been developed to mitigate the occurrence of
irregularities in patient respiratory motion. Audio and visual feedback techniques have been
demonstrated to be an effective way of achieving a consistent pattern of respiration. George
et al (2006) subjected 24 lung cancer patients to audio-visual respiratory feedback techniques
and observed significant improvement in the reproducibility of respiration. Gating technology
(Ramsey et al 1999, Wagman et al 2003) is also used to prevent irradiation in the presence of
drastic changes in breathing such as sneezing or coughing.
Despite the existing studies demonstrating the amount of irregularities in respiratory
motion, there is limited information on the dosimetric outcome of irregular breathing and
whether or not they have a clinically significant impact on therapeutic goals of the irradiation.
Due to radiation dispersion in tissue, penumbral effects and features of plan topology,
geometric observations performed with radiographic imaging or optical methods are notdirectly translatable to dosimetric conclusions. In the context of patient setup errors, the
amount of geometric miss and the associated target margin to ensure sufficient coverage are
shown to be proportional but not the same (van Herk 2004). This is a result of differences in
the shapes of target definition function perceived visually as a step function at the border of
the target and the function of the dosimetric profile across the target with a gradual fall off.
Similarly, van Herk et al (2003) analytically demonstrated that the additional target margin
required to compensate the motion of the target is substantially smaller than its full motion
amplitude. This was also confirmed recently by Mutaf and Brinkmann (2008b) in a more
extensive treatment planning study incorporating actual patient data. The aim of the current
study is to extend the dosimetric investigation of mobile targets under respiration and evaluate
changes in target coverage due to irregularities in its motion.
2. Methods and materials
The dosimetric effects of respiratory motion need to be evaluated in relation to the specific
treatment delivery techniques used for compensation of the target motion. In this study, target
motion was incorporated into treatment planning via an expansion of the target volume by
an amount equal to the target excursion during the full respiratory cycle (i.e. no respiratory
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
4/16
Irregular respiratory motion 847
Table 1. Patient target volume statistics and their anatomical location are listed. The number offields and conformity of the plan dose to the target are also shown in simple and complex geometryplans.
Simple plans Complex plans
Patient no Anatomic location CTV (ml) Nfields Conformity Nfields Conformity
1 Right lung (LL) 68.7 2 4.62 6 1.79
2 Left lung (LL) 77.4 3 2.06 5 1.82
3 Right lung (LL) 5.6 2 2.31 7 2.40
4 Right lung (LL) 58.2 3 3.45 5 1.85
5 Right lung (UL) 1.4 3 2.83 6 2.63
6 Right lung (LL) 91.6 3 4.64 8 1.85
7 Left lung (LL) 36.9 3 2.30 7 1.93
8 Right lung (ML) 207.6 3 2.79 6 1.74
9 Right lung (UL) 32.9 3 2.53 5 1.98
10 Right lung (ML) 3.3 2 2.32 5 2.37
11 Left lung (UL) 12.0 3 2.90 6 1.87
12 Right lung (ML) 23.7 2 9.12 6 2.02
13 Right lung (LL) 5.1 2 9.72 5 2.23
14 Right lung (ML) 4.9 3 8.72 5 2.71
15 Left lung (UL) 29.0 2 8.87 6 2.00
16 Right lung (UL) 6.0 3 4.45 5 2.08
17 Right lung (ML) 61.8 2 4.71 5 1.82
18 Left lung (LL) 35.6 3 2.96 6 1.93
19 Right lung (UL) 1.6 2 18.00 5 2.30
20 Left lung (UL) 39.2 2 5.75 5 1.89
21 Right lung (LL) 83.8 3 2.76 7 1.65
22 Left lung (UL) 20.9 3 3.03 6 1.98
23 Right lung (ML) 1.8 3 2.94 7 2.58
Average 39.5 2.6 4.95 5.8 2.06Minimum 1.4 2 2.06 5 1.65
Maximum 207.6 3 18.00 8 2.71
LL: lower lobe; UL: upper lobe; ML: middle lobe.
gating). Irregularities in respiratory motion are investigated using simulations of target motion
and analysis of respective clinical treatment plans.
2.1. Patient cohort and dosimetry
Twenty-three lung cancer patients previously treated in our clinic are retrospectively analyzed
through their informed consent according to an approved University of Pittsburgh institutionalreview board protocol. Clinical target volumes (CTV) ranged from 1.4 to 207.6 ml. Although,
prescription doses varied between 50.0 and 66.0 Gy for actual patient treatments, the
prescription doses were all normalized to 60.0 Gy to facilitate easier comparisons for the
effects of respiratory motion. A summary of patient information related to treatment plans
of the patients is provided in table 1. Due to the radio-sensitivity of lungs to low doses of
radiation (Oetzel et al 1995, Graham et al 1999), the clinical treatment plans were designed to
minimize irradiation volume and therefore consisted of two or three fields (usually in parallel
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
5/16
848 Y D Mutaf et al
(a) (b)
Figure 1. Comparison of dose distributions in simple (a) and complex (b) geometry plans in axialand coronal views for patient 12.
opposed and oblique orientations). In this paper, such 3D conformal plans are referred to as
simple geometry plans where dose homogeneity within the target ranged between 95% and
115% of the prescription dose. Adopting the RTOG formulation of conformity (Shaw et al
2000) (defined as the ratio of 95% isodose volume to the target volume), the average target
conformity in simple plans was 4.95. In addition to simple geometry plans, an alternative
treatment plan was also generated incorporating more fields, ranging from five to eight, in
order to produce more conformal target dosimetry. These plans, referred as complex geometry
plans, achieved an average conformity of 2.06 in addition to improving dose homogeneity
within the target (95% to 105% of prescription dose). The typical dose distributions in simple
and complex geometry plans are demonstrated in figure 1 for one of the patients (patient 12).
The corresponding dosevolume histograms for target and bi-lateral lungs are also presented
in figure 2. In both type of plans, the selection of beams is restricted to the angles orthogonalto the longitudinal axis of the patients (transverse co-planar arrangement). All plans were
manually optimized to conform to the shape of the target with static 0.5 cm MLCs and a block
margin of 0.8 cm. Hard wedges were utilized for improvement of dose homogeneity. No
intensity modulation techniques and optimization engines (i.e. IMRT) are utilized.
Plans were further imposed with other constraints such as a maximum tolerable spinal
cord dose of 45.0 Gy and a mean heart dose of 40.0 Gy, although the latter constraint was
considered on a case-by-case basis depending on the location of the tumor.
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
6/16
Irregular respiratory motion 849
Figure 2. Comparison of target volume and bi-lateral lungs dosevolume histograms for simple(solid lines) and complex (dashed lines) geometry plans shown in figure 1.
2.2. Target selection and contouring
Target definitions and contouring adhered to ICRU guidelines (International Commission on
Radiation Units and Measurements 1993, 1999). Simulation images were acquired from
4DCT scans using a retrospective phase sorting technique producing ten phase images at
equal temporal intervals (i.e. 10% of respiration period). Following institutional protocols,
treatment simulation and planning was performed on the end-of-exhalation (EOE) phase
dataset extracted from the 4DCT scan (image in which the tumor volume is located most
superiorly, typically 50% phase image). The EOE image was preferred for planning since in a
typical respiratory cycle, longest amount of time is spent at this phase and therefore produces
3D images with least amount of motion artifacts. CTV was contoured on the EOE images
by the patients primary radiation oncologist with consideration for the subclinical extent of
the disease. The clinical ITV was finally generated using an asymmetric expansion, internal
margin or IM, equal to the amplitude of the target motion. In order to investigate the effects of
different amplitudes of motions, the CTV is enlarged by a specific IM amount which is chosen
to be equal to the amount of simulated motion and its direction (equation ( 1)):
ITVIM = CTV IM(x,y,z). (1)
In this formalism, ITV volumes are created by adding 1 and 2 cm margins to CTV in theinferior direction as demonstrated in figure 3. These final structures are referred as ITV1 cmand ITV2 cm within the text, where ITV1 c m = CTV1 cm and ITV2 c m = CTV2 cm..
2.3. Respiratory motion simulations
The mechanics of respiration induces an expansioncontraction motion observed mostly in
the cranio-caudal (CC) axis of the patient. Several radiological studies (Seppenwoolde et al
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
7/16
850 Y D Mutaf et al
Figure 3. An illustration of the target volumes (CTV and ITV) described in the text. Internaltarget volumes (ITV) are generated from clinical target volumes (CTV) via application of 1 and2 cm asymmetric internal margins in the inferior direction.
2002, Murphy et al 2002) have confirmed that largest motion occurs along the CC axis of the
patients. Therefore, motion simulations in this study are restricted to the CC axis of the patient
where CTV center-of-mass is changed with respect to the amount of simulated motion.
Respiratory motion simulations are performed via dose accumulation for voxels within
the CTV boundaries through the length of time the respiratory motion assessed. At each
dose sampling time, the target center-of-mass was moved with respect to the instantaneousamplitude of the respiratory motion but no deformations were allowed. Therefore, the total
number of target voxels and their coordinates within the target reference frame remain the
same. Due to the scan resolution of the CT images, target voxels have the dimensions of
1.00 1.00 mm2 in the axial plane and are separated by 1.25 mm slice thickness. Similarly,
the simulated motions were also digitized with respect to the scan resolution, i.e. 1.25 mm
since the simulated motion is in the longitudinal axis of the patient. In order to improve
processing time, the temporal resolution of the simulations are reduced such that the target
dose sampling is only performed at 10 equally spaced times within a respiratory cycle (i.e.
10% phase intervals).
Two main forms of respiration motion models are used for the simulations: regularand
irregularrespiration. In a regular respiration study, a single respiratory cycle with a functional
form represented in equation (2) was used for target simulations (Lujan et al 2003). A plot
of respiratory motion amplitude in this form is shown in figure 4. For simulation of a 1.0
cm peak-to-peak motion (A0), a 1 cm internal margin was used for CTV expansion to ITV
(ITV1 cm), and similarly, a 2 cm internal margin was used as the target volume for a 2.0 cm
peak-to-peak motion:
z() = A0 cos4( ). (2)
Irregular motion profiles used in the study were created similarly by generating 100
respiratory cycles which follow the functional form in equation (2). However, each respiratory
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
8/16
Irregular respiratory motion 851
Figure 4. Plot of the functional form of respiration used in the simulations. The peak-to-peakamplitude of the motion, A0, is used as the free parameter and adjusted accordingly as describedin the text.
cycle within a session was varied by changing its peak-to-peak amplitude (A0) sampled from a
Gaussian distribution with a preselected mean (A) and standard deviation (A). The examples
of simulated irregular respiration cycles and the corresponding distribution of peak-to-peak
amplitudes are shown in figure 5.
Irregular respiratory motion simulations are further divided into two categories.
Characteristic or random irregular motion where the irregularities in the respiratory
motion during treatment are distributed randomly about the motion amplitude attributed
to the CTV during the 4DCT imaging.
Uncharacteristic or systematically irregular motion where there is a systematic shift
between CTV motion amplitude determined at the time of 4DCT imaging and the actual
respiratory behavior of the patient during treatment.
In characteristic motion, the meanof the Gaussiandistribution for peak-to-peak amplitudes
is chosen to be equal to the internal margin. For example, a set of respiratory cycles distributed
about a mean of 1.0 cm (e.g. figure 5, upper panel) is characteristic for the simulations of CTV
motion using the plan generated with a 1.0 cm internal margin (ITV1 cm). In uncharacteristic
motion simulations, however, the internal margin between the CTV and ITV is smaller thanthe mean of the respiratory cycle amplitudes by 1 standard deviation. An example to such
respiratory motion would be a set of respiratory cycles distributed according to a mean of 1.5
cm and a standard deviation of 0.5 cm (e.g. figure 5, lower panel) used for the simulations
with only a 1.0 cm internal margin (ITV1 cm).
A characteristic irregular motion would, therefore, be the best-case scenario where the
irregularities are essentially random in nature and CTV motion as observed in the simulation
(e.g. with 4DCT) corresponds to the average target motion. In contrast, the uncharacteristic
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
9/16
852 Y D Mutaf et al
Figure 5. A time series plot of the simulated irregular respiration used in the study for the meanpeak-to-peak amplitude values (A) of 1.0cm (upperpanel) and1.5 cm (lowerpanel). A histogramof the cycle peak-to-peak amplitudes is plotted next to the corresponding plot of the respirationdata. Althoughthe continuous form of the full respiration datais shown (solid line), the simulationsare performed for respiration amplitudes sampled at phase intervals of 10% (filled circles). Thesampled motion is also digitized by 1.25 mm equal to the slice thickness of the CT images.
irregular motion reflects a more extreme case where in addition to random variations in
respiration, the determined CTV motion and internal margin are compromised by a systematic
difference. This systematic error occurs when the amount of motion determined at the planning
stage is consistently less than the average target motion observed during patients treatment.Characteristic and uncharacteristic irregular motion simulation parameters are listed
below.
Simulations with ITV1 cmA = 1.00 cm, A = 0.25 cm or 25% (characteristic)
A = 1.00 cm, A = 0.50 cm or 50% (characteristic)
A = 1.25 cm, A = 0.25 cm (uncharacteristic)
A = 1.50 cm, A = 0.50 cm (uncharacteristic).
Simulations with ITV2 cmA = 2.00 cm, A = 0.50 cm or 25% (characteristic)
A = 2.00 cm, A = 1.00 cm or 50% (characteristic)
A = 2.50 cm, A = 0.50 cm (uncharacteristic)
A = 3.00 cm, A = 1.00 cm (uncharacteristic).These a priori choices of the mean and the standard value of the respiration amplitude
distribution are intended to provide several threshold levels for their influence on target
dosimetry rather than exact representations of what is observed with actual patients. In this
respect, 25% variations as used in this study stand as a close approximation to what we
observed within our patient population as well as what is reported by other investigators
(Britton et al 2007). However, amplitude variations as large as 50% are not typically reported
in the literature and therefore constitute a limiting case.
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
10/16
Irregular respiratory motion 853
Figure 6. Respiratory motion trace of a patient (abdomen motion). The average peak inhale andexhale amplitudes are emphasized with blue and red horizontal lines, respectively.
In addition to the functional forms of irregular respiratory motion as described above,
patients actual respiratory waveforms were also used for target motion simulations; however,
these data were available only for a subset of patients, 13 out of 23 patients studied here.The respiratory signal represented the vertical motion of the abdominal surface recorded with
the aid of reflective markers placed on patients abdomen and an infrared camera as available
in a commercial respiratory management system (RPM, Varian Inc.). In clinical 4DCT
procedures, these patients were subjected to audio-coaching feedback methods for improving
the reproducibility of their respiratory pattern (Kini et al 2003, George et al 2006, Haasbeek
et al 2008).
The abdominal motion is used as surrogate for the actual tumor motion (Rietzel et al
2005, Mutaf and Brinkmann 2008a) and has been shown to have a high correlation with the
actual tumor motion in lung cancer patients (Ozhasoglu and Murphy 2002, Starkschall et al
2004, Gierga et al 2005). However, this correlation is only relevant for the relative amounts
of surrogate and target motions and therefore the surrogate motion is scaled in the simulations
such that the average respiratory cycle from the entire surrogate waveform was normalized
to selected peak-to-peak amplitudes. As described earlier, this normalization was chosen as
1.0 cm for the average peak-to-peak motion of the surrogate data in simulations for plans
with ITV1 cm and 2.0 cm for simulations with ITV2 cm. An example of patient respiration
waveform is demonstrated in figure 6. The average respiration cycle data are used for
regular motion simulation; however, the full respiration waveform is used for irregular motion
simulations.
In contrast to the synthesized respiratory motion data used for characteristic and
uncharacteristic irregular motion simulations, the actual respiratory trace of the patients
provides a more diverse spectrum of irregularities in breathing. While synthesized respiratory
simulations only represent variations in inhale amplitude, actual motion data also include such
irregular motion as variations in end-of-exhale (a.k.a. baseline-drift) as well as the presence
of outlier respiratory cycles. However, due to the use of single session data, the simulations of
irregular motion with patient motion traces only represent intra-fractional variations excludingpossible inter-fractional changes in breathing.
2.4. Plan evaluation metrics
Several metrics are utilized to assess the effect of irregular respiratory motion on dosimetric
coverage for CTV. Respiratory motion mostly affects the cells at the margins of the target
causing them to move outside the boundaries of the irradiation field and receive less than
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
11/16
854 Y D Mutaf et al
planned dose. Therefore, the target minimum dose (Dmin) is expected to be the most sensitive
metric for dosimetric evaluation of structures under motion. In addition, global dosimetric
evaluators such as the equivalent uniform dose (EUD) (Niemierko 1997, Choi and Deasy
2002) and tumor control probability (TCP) (Webb and Nahum 1993, Okunieffet al 1995) are
used for evaluations.The EUD is defined such that if a target is irradiated uniformly with this dose, it will
result in the same clonogenic cell killing as the actual inhomogeneous dose distribution across
the target volume. Assuming a homogeneous density of tumor clonogens within the target
volume and ignoring cell repair and repopulation effects, Niemierkos (1997) EUD formalism
was adopted for this study as shown in equation (3):
EUD = Dref ln
1N
Ni=1 (SF2)Di/Dref
ln(SF2). (3)
In equation (3), the reference dose, Dref, is defined as the reference dose per fraction and
usually taken as 2 Gy, whereas SF2, the surviving fraction of clonogens at 2 Gy, is set to 0.5.
Using these parameters, the average EUD values were 61.1 Gy (range 59.7 to 63.1 Gy) and
60.4 Gy (range 59.5 to 61.2 Gy) for simple and complex plans, respectively (note that the all
prescription doses have been normalized to 60 Gy in all patient plans for the purpose of thisstudy).
TCP is also calculated to assist the dosimetric evaluations for the target and assess any
dosimetric changedue to irregular respiratory motionin terms of itsrelative role in undermining
local control, if any. TCP is defined as the Poisson probability for achieving complete tumor
cell lethality (zero survival probability). Analysis of dose-response data from population-
based studies results in several empirical formulations for TCP and we implemented the logit
form as described in Okunieffet al (1995) and shown in equation (4):
TCP =e
450DD50D50
1 + e
450DD50D50
. (4)
Tumor control probability has a sigmoid shape which is modeled by the two free parameters
D50 and 50, dose to achieve 50% control probability and the slope of the sigmoid curve atD50, respectively. Following the multi-institutional analysis of the dose response data for
macroscopic lung tumors of Okunieff et al we adopted a D50 value of 51.2 Gy and 50 of
0.83. For this set of parameters, the average TCP values were 0.66 (range 0.640.69) and 0.65
(range 0.630.66) for simple and complex plans, respectively.
It is not the intent of the current study to evaluate the validity of these treatment plan
metrics or their selected parameters. These metrics are solely used to facilitate comparisons
between target dosimetry under irregular respiratory motion as compared to its regular motion.
Therefore, these metrics are only interpreted with respect to their relative values within
corresponding comparisons. Furthermore, we define a figure of relativity for each metric
discussed above that is equal to the ratio of the same metric for the target under irregular motion
to that under regular motion. This is formulated in equation (5). Following this approach, we
define rel-Dmin, rel-EUD and rel-TCP metrics for the targets and motions studied here:
rel-M= Mirreg/Mreg. (5)
3. Results
The result of characteristic irregular motion simulations with 1.0 and 2.0 cm amplitudes are
displayed in figure 7 for all 23 patients. The changes in the target minimum dose, represented
by rel-Dmin, are also shown for simple and complex plans.
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
12/16
Irregular respiratory motion 855
Figure 7. Relative target minimum dose (rel-Dmin) results of characteristic irregular motionsimulations for all patients; mean motion amplitudes of 1 cm (upper plot) and 2 cm (lower panel)are shown.
In characteristic motion simulations with a 1 cm average amplitude and 25% standard
deviation, the maximum change in the minimum dose to CTV was 0.1% for both simple
and complex plans. However no change in EUD or TCP was observed. A change in target
dose was slightly larger when the variations in respiratory motion increased to 50% standard
deviation with a maximum reduction of 0.6% in the minimum target dose for simple and 0.8%
for complex plans. For this motion, maximum changes in both EUD and TCP were 0.2% in
both simple and complex plans among all patients.
When the motion amount was increased to 2 cm in amplitude, changes in all dosimetry
metrics for CTV were larger for both 25% and 50% variations. The maximum change in the
minimum dose to a target was 3.7% in the case of 50% variations. For 50% variations,
the maximum EUD reduction was 2.1% and the TCP reduction was 3.0%. No substantial
difference between simple and complex plan dosimetry was observed.Target dosimetry for uncharacteristic motion is also presented in comparison with the
characteristic motion in figure 8 shown only for complex geometry plans. The effect
of uncharacteristic motion on simple plans again was not substantially different than the
complex plans. For plans with 1.0 and 2.0 cm internal margins in all patients, CTV dose
was compromised more drastically in uncharacteristic irregular motion simulations compared
to characteristic motion. The largest change in target dosimetry was observed for plan
with a 2.0 cm margin and associated uncharacteristic irregular motion ( = 3.0 cm,
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
13/16
856 Y D Mutaf et al
Figure 8. Comparison of relative target minimum dose (rel-Dmin) in irregular motion simulationsfor characteristic and uncharacteristic motion traces; simulations with 1 cm (upper panel) and2 cm (lower panel) internal margins are shown.
= 1.0 cm) such that the minimum target dose was reduced by 13.6% as compared to
the regular motion.
Simulations of respiratory motion using the actual abdominal (surrogate) motion of
patients were also repeated for both 1.0 and 2.0 cm average motion amplitudes. The maximum
change in the target minimum dose was 0.1% in both cases using complex plans. The
maximum reduction in EUD was 0.3% and in TCP was 0.5% observed only with a 2 cm
motion. Simple plans yielded similar results.
The ranges of target dosimetry metrics among all the patients for all simulated motions
are summarized in table 2.
4. Discussion
In this study, an evaluation of the dosimetric effects of irregular motion was presented
for 23 lung cancer patients. Several types of irregular motion simulations are performed
to assess target dosimetry under various irregular motion conditions. In the process, the
interplay between the plan topology as demonstrated by simple and complex three-dimensional
dosimetry and the presence of irregular respiratory motion was also investigated.
Characteristic irregular motion type representative of random variations in the determined
(i.e.planned) target motion caused non-substantive dosimetric effects. In mostsimulated cases,
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
14/16
Irregular respiratory motion 857
Table 2. Range of target dosimetry evaluation metrics for the patient population investigated inthis study. Relative metrics as tabulated here are defined in equation (5).
Characteristic motion (IM = ) rel-Dmin rel-EUD rel-TCP
= 1.0 cm, = 0.25 cm (IM = 1.0 cm) 99.9100.0% 100.0% 100.0% = 1.0 cm, = 0.50 cm (IM = 1.0 cm) 99.2100.0% 99.8100.0% 99.8100.0%
= 2.0 cm, = 0.5 cm (IM = 2.0 cm) 99.5100.0% 99.8100.0% 99.7100.0%
= 2.0 cm, = 1.0 cm (IM = 2.0 cm) 96.299.6% 97.999.8% 97.099.7%
Un-characteristic motion (IM = )
= 1.25 cm, = 0.25 cm (IM = 1.0 cm) 99.0100.0% 99.8100.0% 99.8100.0%
= 1.50 cm, = 0.50 cm (IM = 1.0 cm) 91.699.3% 98.299.9% 98.199.8%
= 2.5 cm, = 0.5 cm (IM = 2.0 cm) 95.8100.0% 98.799.9% 98.499.9%
= 3.0 cm, = 1.0 cm (IM = 2.0 cm) 86.493.6% 92.799.0% 90.899.0%
Patient respiratory motion
Normalized to 1 cm (IM = 1.0 cm) 99.9100.0% 99.9100.0% 99.9100.0%
Normalized to 2 cm (IM = 2.0 cm) 99.0100.0% 99.7100.0% 99.5 100.0%
the change in the target minimum dose was clinically insignificant (less than 2%) to require
intervention or other compensation techniques. In the extreme scenario where the simulated
irregular motion had a mean of 2 cm and showed 1 cm amplitude variations (at 1 standard
deviation), the minimum target dose was dropped by 2.5% 0.9%. Although minimal, these
observations confirm the expectation that the dosimetric effect of characteristic irregularities
increases as the average amount of motion and its relative variations increases.
However, there is no assurance that the target motion as determined during patient
simulation (e.g. 4DCT) would be reflective of the average respiratory motion for the entire
course of treatment. For normally distributed motion data, there is a 50% probability that the
internal margin amount determined at simulation will be smaller than the average amplitude of
the motion. Dosimetric consequences for such scenarios are investigated in this study via the
simulations of uncharacteristic irregular motion where the chosen internal margin for the target
is smaller than the mean of the entire peak amplitude distribution by 1 standard deviation.
Uncharacteristic irregular motion resulted in clinically significant dosimetric
consequences with minimum CTV doses reduced about 8.4% for a 1.0 cm internal margin
(motion = 1.5 cm, = 0.5 cm) and up to 13.6% for a 2.0 cm internal margin (motion =
3.0 cm, = 1.0 cm). This observation signifies that such systematic discrepancies between
the determined target motion and the entire respiratory motion have more impact than the
random variations in respiration. In fractionated radiotherapy, an example of this discrepancy
could appear as inter-fractional variations of the respiratory motion between the simulation
session where target motion is determined and subsequent treatment sessions.
Results from the simulation of actual patient respiratory waveforms indicated that therespiration motion irregularities demonstrated by these patients did not cause detrimental
dosimetric consequences. This parallels our observations with characteristic irregular motion.
The similarity between the actual patient respiration motion and generated characteristic
irregular motion simulations could be attributed to the fact that the patient respiratory data
were collected with the help of audio-coaching techniques. Since coaching compels the patient
to reproduce a respiratory pattern, the breathing irregularities are consequentially restricted
to random fluctuations. Therefore, it can be argued that the coaching techniques help avoid
-
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
15/16
858 Y D Mutaf et al
the occurrence of uncharacteristic respiratory motion which was observed to result in largest
degradation for target dosimetry.
5. Conclusion
Determination of respiratory motion is a necessary first step for the management of respiratory
motion in radiotherapy. The process also requires that determined respiratory motion is valid
throughout the full course of patients treatment. However, this validity is compromised
due to the varying degrees of breathing irregularities. In this study, we assessed the clinical
implications and dosimetric consequences of these variations and categorized different types
of breathing irregularities. Characteristic motion irregularities are not observed to have a
considerable clinical effect on target dosimetry. Similarly, simulations of target motion using
actual patient respiration data have not produced a clinically substantial dosimetric effect.
However, uncharacteristic irregularities are shown to have critical dosimetric consequences
for the target coverage. Coaching techniques are suggested as potential ways to avoid such
inconsistencies and maintain a common baseline for patients respiration throughout the course
of radiotherapy. Additionally, there was no substantial difference between the simple andcomplex treatment plan geometries incorporating different levels of target dose homogeneity
and conformity.
References
Britton K R, Starkschall G, Tucker S L, Pan T, Nelson C, Chang J Y, Cox J D, Mohan R and Komaki R 2007
Assessment of gross tumor volume regression and motion changes during radiotherapy for non-small-cell lung
cancer as measured by four-dimensional computed tomography Int. J. Radiat. Oncol. Biol. Phys. 68 103646
ChangJ etal 2007Observationof interfractional variations in lungtumor position using respiratory gated and ungated
megavoltage cone-beam computed tomography Int. J. Radiat. Oncol. Biol. Phys. 67 154858
Choi B and Deasy J O 2002 The generalized equivalent uniform dose function as a basis for intensity-modulated
treatment planning Phys. Med. Biol. 47 357989
George R, Chung T D, Vedam S S, Ramakrishnan V, Mohan R, Weiss E and Keall P J 2006 Audio-visual biofeedback
for respiratory-gated radiotherapy: impact of audio instruction and audio-visual biofeedback on respiratory-
gated radiotherapy Int. J. Radiat. Oncol. Biol. Phys. 65 92433
Gierga D P, Brewer J, Sharp G C, Betke M, Willett C G and Chen G T 2005 The correlation between internal
and external markers for abdominal tumors: implications for respiratory gating Int. J. Radiat. Oncol. Biol.
Phys. 61 15518
Graham M V, Purdy J A, Emami B, Harms W, Bosch W, Lockett M A and Perez C A 1999 Clinical dose-volume
histogram analysis for pneumonitis after 3D treatment for non-small cell lung cancer (NSCLC) Int. J. Radiat.
Oncol. Biol. Phys. 45 3239
Haasbeek C J, Spoelstra F O, Lagerwaard F J, van Sornsen de Koste J R, Cuijpers J P, Slotman B J and Senan S 2008
Impact of audio-coaching on the position of lung tumors Int. J. Radiat. Oncol. Biol. Phys. 71 111823
International Commission on Radiation Units and Measurements 1993 Prescribing, recording, and reporting photon
beam therapy ICRU Report 50 (Bethesda, MD: ICRU)
International Commission on Radiation Units and Measurements 1999 Prescribing, recording, and reporting photon
beam therapy ICRU Report 62 (Bethesda, MD: ICRU)
Juhler Nottrup T, Korreman S S, Pedersen A N, Aarup L R, Nystrom H, Olsen M and Specht L 2007 Intra- and
interfraction breathing variations during curative radiotherapy for lung cancer Radiother. Oncol. 84 408
Kini V R, Vedam S S, Keall P J, Patil S, Chen C and Mohan R 2003 Patient training in respiratory-gated radiotherapy
Med. Dosim. 28 711
Lujan A E, Balter J M and Ten Haken R K 2003 A method for incorporating organ motion due to breathing into 3D
dose calculations in the liver: sensitivity to variations in motion Med. Phys. 30 26439
Murphy M J, Martin D, Whyte R, HaiJ, OzhasogluC andLe Q T 2002 Theeffectiveness of breath-holding to stabilize
lung and pancreas tumors during radiosurgery Int. J. Radiat. Oncol. Biol. Phys. 53 47582
Mutaf Y D and Brinkmann D H 2008a An investigation of temporal resolution parameters in cine-mode four-
dimensional computed tomography acquisition J. Appl. Clin. Med. Phys. 9 2819
http://dx.doi.org/10.1016/j.ijrobp.2007.01.021http://dx.doi.org/10.1016/j.ijrobp.2007.01.021http://dx.doi.org/10.1016/j.ijrobp.2007.01.021http://dx.doi.org/10.1016/j.ijrobp.2006.11.055http://dx.doi.org/10.1016/j.ijrobp.2006.11.055http://dx.doi.org/10.1016/j.ijrobp.2006.11.055http://dx.doi.org/10.1088/0031-9155/47/20/302http://dx.doi.org/10.1088/0031-9155/47/20/302http://dx.doi.org/10.1088/0031-9155/47/20/302http://dx.doi.org/10.1016/j.ijrobp.2006.02.035http://dx.doi.org/10.1016/j.ijrobp.2006.02.035http://dx.doi.org/10.1016/j.ijrobp.2006.02.035http://dx.doi.org/10.1016/j.ijrobp.2004.12.013http://dx.doi.org/10.1016/j.ijrobp.2004.12.013http://dx.doi.org/10.1016/j.ijrobp.2004.12.013http://dx.doi.org/10.1016/S0360-3016(99)00183-2http://dx.doi.org/10.1016/S0360-3016(99)00183-2http://dx.doi.org/10.1016/S0360-3016(99)00183-2http://dx.doi.org/10.1016/j.ijrobp.2007.11.061http://dx.doi.org/10.1016/j.ijrobp.2007.11.061http://dx.doi.org/10.1016/j.ijrobp.2007.11.061http://dx.doi.org/10.1016/j.radonc.2007.05.026http://dx.doi.org/10.1016/j.radonc.2007.05.026http://dx.doi.org/10.1016/j.radonc.2007.05.026http://dx.doi.org/10.1016/S0958-3947(02)00136-Xhttp://dx.doi.org/10.1016/S0958-3947(02)00136-Xhttp://dx.doi.org/10.1016/S0958-3947(02)00136-Xhttp://dx.doi.org/10.1118/1.1609057http://dx.doi.org/10.1118/1.1609057http://dx.doi.org/10.1118/1.1609057http://dx.doi.org/10.1016/S0360-3016(01)02822-Xhttp://dx.doi.org/10.1016/S0360-3016(01)02822-Xhttp://dx.doi.org/10.1016/S0360-3016(01)02822-Xhttp://dx.doi.org/10.1120/jacmp.v9i4.2819http://dx.doi.org/10.1120/jacmp.v9i4.2819http://dx.doi.org/10.1120/jacmp.v9i4.2819http://dx.doi.org/10.1120/jacmp.v9i4.2819http://dx.doi.org/10.1016/S0360-3016(01)02822-Xhttp://dx.doi.org/10.1118/1.1609057http://dx.doi.org/10.1016/S0958-3947(02)00136-Xhttp://dx.doi.org/10.1016/j.radonc.2007.05.026http://dx.doi.org/10.1016/j.ijrobp.2007.11.061http://dx.doi.org/10.1016/S0360-3016(99)00183-2http://dx.doi.org/10.1016/j.ijrobp.2004.12.013http://dx.doi.org/10.1016/j.ijrobp.2006.02.035http://dx.doi.org/10.1088/0031-9155/47/20/302http://dx.doi.org/10.1016/j.ijrobp.2006.11.055http://dx.doi.org/10.1016/j.ijrobp.2007.01.021 -
8/4/2019 A Simulation Study of Irregular Respiratory Motion and Its Dosimetric Impact on Lung Tumors
16/16
Irregular respiratory motion 859
Mutaf Y D and Brinkmann D H 2008b Optimization of internal margin to account for dosimetric effects of respiratory
motion Int. J. Radiat. Oncol. Biol. Phys. 70 156170
Niemierko A 1997 Reporting and analyzing dose distributions: a concept of equivalent uniform dose Med.
Phys. 24 10310
Oetzel D, Schraube P, Hensley F, Sroka-Perez G, Menke M and Flentje M 1995 Estimation of pneumonitis risk
in three-dimensional treatment planning using dose-volume histogram analysis Int. J. Radiat. Oncol. Biol.
Phys. 33 45560
Okunieff P, Morgan D, Niemierko A and Suit H D 1995 Radiation dose-response of human tumors Int. J. Radiat.
Oncol. Biol. Phys. 32 122737
Ozhasoglu C and Murphy M J 2002 Issues in respiratory motion compensation during external-beam radiotherapy
Int. J. Radiat. Oncol. Biol. Phys. 52 138999
Ramsey C R, Scaperoth D, Arwood D and Oliver A L 1999 Clinical efficacy of respiratory gated conformal radiation
therapy Med. Dosim. 24 1159
Rietzel E, Pan T and Chen G T 2005 Four-dimensional computed tomography: image formation and clinical protocol
Med. Phys. 32 87489
Seppenwoolde Y, Shirato H, Kitamura K, Shimizu S, van Herk M, Lebesque J V and Miyasaka K 2002 Precise and
real-timemeasurement of 3D tumor motion in lungdue to breathingand heartbeat, measuredduring radiotherapy
Int. J. Radiat. Oncol. Biol. Phys. 53 82234
Shaw E, Scott C, Souhami L, Dinapoli R, Kline R, Loeffler J and Farnan N 2000 Single dose radiosurgical treatment
of recurrent previously irradiated primary brain tumors and brain metastases: final report of RTOG protocol
9005 Int. J. Radiat. Oncol. Biol. Phys. 47 2918
Starkschall G, Forster K M, Kitamura K, Cardenas A, Tucker S L and Stevens C W 2004 Correlation of gross tumor
volume excursion with potential benefits of respiratory gating Int. J. Radiat. Oncol. Biol. Phys. 60 12917
van Herk M 2004 Errors and margins in radiotherapy Semin. Radiat. Oncol. 14 5264
van Herk M, Witte M, van der Geer J, Schneider C and Lebesque J V 2003 Biologic and physical fractionation effects
of random geometric errors Int. J. Radiat. Oncol. Biol. Phys. 57 146071
Wagman R, Yorke E, Ford E, Giraud P, Mageras G, Minsky B and Rosenzweig K 2003 Respiratory gating for liver
tumors: use in dose escalation Int. J. Radiat. Oncol. Biol. Phys. 55 65968
WebbS andNahum A E 1993 A modelfor calculating tumourcontrol probability in radiotherapyincluding theeffects
of inhomogeneous distributions of dose and clonogenic cell density Phys. Med. Biol. 38 65366
http://dx.doi.org/10.1016/j.ijrobp.2007.12.025http://dx.doi.org/10.1016/j.ijrobp.2007.12.025http://dx.doi.org/10.1016/j.ijrobp.2007.12.025http://dx.doi.org/10.1118/1.598063http://dx.doi.org/10.1118/1.598063http://dx.doi.org/10.1118/1.598063http://dx.doi.org/10.1016/0360-3016(95)00009-Nhttp://dx.doi.org/10.1016/0360-3016(95)00009-Nhttp://dx.doi.org/10.1016/0360-3016(95)00009-Nhttp://dx.doi.org/10.1016/0360-3016(94)00475-Zhttp://dx.doi.org/10.1016/0360-3016(94)00475-Zhttp://dx.doi.org/10.1016/0360-3016(94)00475-Zhttp://dx.doi.org/10.1016/S0360-3016(01)02789-4http://dx.doi.org/10.1016/S0360-3016(01)02789-4http://dx.doi.org/10.1016/S0360-3016(01)02789-4http://dx.doi.org/10.1016/S0958-3947(99)00006-0http://dx.doi.org/10.1016/S0958-3947(99)00006-0http://dx.doi.org/10.1016/S0958-3947(99)00006-0http://dx.doi.org/10.1118/1.1869852http://dx.doi.org/10.1118/1.1869852http://dx.doi.org/10.1118/1.1869852http://dx.doi.org/10.1016/S0360-3016(02)02803-1http://dx.doi.org/10.1016/S0360-3016(02)02803-1http://dx.doi.org/10.1016/S0360-3016(02)02803-1http://dx.doi.org/10.1016/S0360-3016(99)00507-6http://dx.doi.org/10.1016/S0360-3016(99)00507-6http://dx.doi.org/10.1016/S0360-3016(99)00507-6http://dx.doi.org/10.1016/j.ijrobp.2004.07.707http://dx.doi.org/10.1016/j.ijrobp.2004.07.707http://dx.doi.org/10.1016/j.ijrobp.2004.07.707http://dx.doi.org/10.1053/j.semradonc.2003.10.003http://dx.doi.org/10.1053/j.semradonc.2003.10.003http://dx.doi.org/10.1053/j.semradonc.2003.10.003http://dx.doi.org/10.1016/j.ijrobp.2003.08.026http://dx.doi.org/10.1016/j.ijrobp.2003.08.026http://dx.doi.org/10.1016/j.ijrobp.2003.08.026http://dx.doi.org/10.1016/S0360-3016(02)03941-Xhttp://dx.doi.org/10.1016/S0360-3016(02)03941-Xhttp://dx.doi.org/10.1016/S0360-3016(02)03941-Xhttp://dx.doi.org/10.1088/0031-9155/38/6/001http://dx.doi.org/10.1088/0031-9155/38/6/001http://dx.doi.org/10.1088/0031-9155/38/6/001http://dx.doi.org/10.1088/0031-9155/38/6/001http://dx.doi.org/10.1016/S0360-3016(02)03941-Xhttp://dx.doi.org/10.1016/j.ijrobp.2003.08.026http://dx.doi.org/10.1053/j.semradonc.2003.10.003http://dx.doi.org/10.1016/j.ijrobp.2004.07.707http://dx.doi.org/10.1016/S0360-3016(99)00507-6http://dx.doi.org/10.1016/S0360-3016(02)02803-1http://dx.doi.org/10.1118/1.1869852http://dx.doi.org/10.1016/S0958-3947(99)00006-0http://dx.doi.org/10.1016/S0360-3016(01)02789-4http://dx.doi.org/10.1016/0360-3016(94)00475-Zhttp://dx.doi.org/10.1016/0360-3016(95)00009-Nhttp://dx.doi.org/10.1118/1.598063http://dx.doi.org/10.1016/j.ijrobp.2007.12.025