a tale of three (identical) scanners - ctug.org.uk tale of three (identical) scanners.pdf · a tale...
TRANSCRIPT
A tale of three
(identical) scanners
Gareth Iball, Lizzy Crawford, Samir Dawoud
Leeds Teaching Hospitals NHS Trust
This is an edited version of the presentation
that was given at the CT Users Group
meeting.
Background
October 2007: three Siemens Sensation 64
systems installed in new hospital wing
Serial numbers: 55220, 55224, 55226
Commissioned by MPE – all results very
comparable
Protocols set up by Siemens on one scanner,
copied across to other two systems
Into clinical use January 2008
Department layout
CT 1
CT 2
CT 3
Patient dose survey - 2008
Audit against current national DRLs
Head, abdo/pelvis, CAP, chest/liver, HRCT
No patient size information
Some comments on large/small patients
Abdomen/pelvis scans
Average DLP for abdo/pelvis scans
0
100
200
300
400
500
600
700
800
CT 1 CT 2 CT 3
DL
P (
mG
ycm
)
Average DLP
DRL DLP
120kV, 200 Q.Ref mAs, 24x1.2mm, 5mm images, pitch = 1.4, CARE Dose 4D on
Follow up
Why the difference?
CT 2 – used for all ICU/HDU patients
Many scanned with arms down higher doses
CT 3 – quite a few large patients in the sample
Suggested lowering mAs to 180 Q.Ref mAs
No subsequent image quality issues reported
3rd UK CT Dose Survey
Set trainees on abdomen/pelvis dose surveys
from PACS
CTDIvol, DLP, AP & lateral sizes
30 patients from across the 3 scanners
Chose 10 patients from each scanner in order to
compare
Found something strange…
His original data
CTDIvol noticeably higher for small patients on CT 2
All ‘arms down’ patients excluded from the data set
CTDIvol variation with patient size for abdomen/pelvis scans
0
2
4
6
8
10
12
14
16
0 200 400 600 800 1000 1200 1400 1600
Cross sectional area (cm2)
CT
DIv
ol
(mG
y)
CT 1
CT 2
CT 3
Log. (CT 1)
Log. (CT 3)
Log. (CT 2)
With extra data
CTDI variation with patient size for abdomen/pelvis scans
R2 = 0.8174
R2 = 0.5111
R2 = 0.5956
0
2
4
6
8
10
12
14
16
0 200 400 600 800 1000 1200 1400 1600
Cross sectional area (cm2)
CT
DIv
ol
(mG
y)
CT1
CT2
CT3
Log. (CT1)
Log. (CT2)
Log. (CT3)
Average CTDIvol figures
CTDIvol (mGy) “Small” “Average” “Large”
CT 1 7.4 10.6 12.0
CT 2 11.8 13.1 13.4
CT 3 9.0 10.0 11.7
Definitely something strange with CT 2
Suspect CARE Dose not functioning properly
Attempt at CT AEC testing
Other trainees looking
at method of routine
AEC testing
Modified CTDI
phantoms
mA values obtained
using DICOM Info
Extractor software
DICOM Info Extractor v1.0.0.0 www.infomed.gr
Initial CARE Dose 4D tests
mA variation for range of CARE Dose 4D settings
0.0
20.0
40.0
60.0
80.0
100.0
120.0
140.0
160.0
0 10 20 30 40 50 60 70 80 90 100 110 120 130 140 150
Image number
Ap
pli
ed
mA
s
Average/Weak
Average/Average
Average/Strong
Strong/Weak
Strong/Average
Strong/Strong
Weak/Average
Weak/Weak
Weak/Strong
CARE Dose 4D test
mA variation along length of modified CTDI phantoms
0
50
100
150
200
250
0 20 40 60 80 100 120 140 160 180
Image number
Ap
pli
ed
mA
CT1
CT2
CT3
We’d used a chest protocol, but all seemed ok now?
Check protocols
Routine abdomen/pelvis:
120kV, 180 Q.Ref mAs, 24x1.2mm, 1 & 5mm
images, pitch = 1.4
CARE Dose 4D settings: Weak/Strong
Same on all the scanners
Repeat dose survey
CTDI variation with patient size for abdomen/pelvis scans
R2 = 0.7584
R2 = 0.7685
R2 = 0.6594
0
2
4
6
8
10
12
14
16
0 200 400 600 800 1000 1200 1400
Cross sectional area (cm2)
CT
DIv
ol
(mG
y)
CT 1
CT 2
CT 3
Log. (CT 1)
Log. (CT 2)
Log. (CT 3)
Data from the same week as we performed the CARE Dose 4D tests
Retrospective doses – June ‘08
CTDI variation with patient size for abdomen/pelvis scans
R2 = 0.5916
R2 = 0.8268
R2 = 0.7533
0
5
10
15
20
25
0 200 400 600 800 1000 1200 1400 1600 1800
Cross-sectional area (cm2)
CT
DIv
ol
(mG
y)
CT1
CT2
CT3
Log. (CT2)
Log. (CT3)
Log. (CT1)
Further investigations…
All scans performed with clinical
abdomen/pelvis scan protocol
Modified CTDI phantoms
Offset modified CTDI phantoms
CIRS virtually human pelvis phantom
mA variation
Noise measurements
Modified CTDI phantoms
DLP (mGycm):
CT1: 263
CT2: 283
CT3: 276
Offset, modified CTDI phantoms
DLP (mGycm):
CT1: 261
CT2: 291
CT3: 266
Pelvis phantom
DLP (mGycm):
CT1: 482
CT2: 588
CT3: 504
Noise measurements
CT
number
Noise
(s.d.)
CT 1 40.77 23.93
CT 2 41.98 19.73
CT 3 44.12 24.84
Investigations with Siemens
Day 1:
MPE AEC tests performed again – same results
Siemens performed their CARE Dose 4D – all ok
Still no clearer
Suggestion from Siemens
“One possibility for the observed behaviour is, that the scan protocols used at the three scanners, although having the same scan parameter settings, may originate from different Siemens protocols.
The mAs adaptation of CARE Dose 4D is based on the ref. mAs and the related reference attenuation. The latter is stored in the system and depends on the body part for which the original Siemens protocol was built.
If for instance the protocol at scanner 2 is based on a thorax protocol, while the protocols at scanners 1 and 3 are based on a pelvis protocol, the dose at scanner 2 will be higher, because the ref. mAs at scanner 2 is related to the typical attenuation of a thorax, while the ref. mAs of scanner 1 and 3 is related to the (higher) attenuation of a typical pelvis.”
Recommended test
“Use for a test the same original Siemens
protocol (e.g. Pelvis Routine) for all three
scanners and scan the pelvis phantom
(accurate phantom positioning is essential) -
is the dose now identical?”
Investigations with Siemens
Day 2:
Scan pelvis phantom on clinical and Siemens
standard protocols
Protocol comparison
mA variation along the length of the CIRS pelvis phantom
0
100
200
300
400
500
600
700
0 50 100 150 200 250 300
Image number
Ap
pli
ed
mA
CT 2 clinical protocol
CT 2 Siemens protocol
CT 3 Siemens protocol
DLP (mGycm):
CT2 clinical: 588
CT2 Siemens: 551
CT3 Siemens: 551
Shows no problem with CARE Dose but a protocol problem. Siemens were correct
Investigations with Siemens
Day 3:
Delete all scan protocols from CT 2 and replace
with scan protocols from CT 3 (hoping that there
were no other protocol errors on CT 3…)
After protocol replacement
mA variation along the length of the CIRS pelvis phantom
0
100
200
300
400
500
600
700
0 50 100 150 200 250 300
Image number
Ap
pli
ed
mA
CT 1
CT 2
CT 3
DLP (mGycm):
CT1: 482
CT2: 493
CT3: 504
Follow up dose survey
CTDIvol variation with patient size for abdomen/pelvis scans
0
2
4
6
8
10
12
14
16
18
20
0 200 400 600 800 1000 1200 1400
Cross sectiona area (cm2)
CT
DIv
ol
(mG
y)
CT 1
CT 2
CT 3
Log. (CT 1)
Log. (CT 2)
Log. (CT 3)
Summary
Even though the protocols appeared identical
they were very different!
Reference attenuation is different for each
body part and affects dose significantly
No display of the body part on which a
protocol is based
Problem only identified because we have 3
identical scanners
Implications for Leeds
Protocols to only be created/amended by CT
Team Leaders
More regular & robust patient dose audit
programme
Need a system for regularly auditing scan
protocols – IR(ME)R requirement
How?? We have 13 scanners in Leeds Trust
alone, maybe 50-100 protocols per scanner…
Could it happen elsewhere?
DLP vs patient size for VHRCT
0
100
200
300
400
500
0 200 400 600 800 1000 1200 1400
Cross sectional area (cm²)
DL
P (
mG
ycm
)
Leeds Sensation 64
Leeds Definition 64
Leeds Sensation 16
Log. (Leeds Sensation 64)
Log. (Leeds Definition 64)
Log. (Leeds Sensation 16)
64 slice: 120kV, 100 Q.Ref mAs, 64x0.6mm, pitch=1.2, CD4D: Weak/Strong
16 slice: 120kV, 100 Q.Ref mAs, 16x0.75mm, pitch=0.85, CD4D: Average/Average
Could it happen elsewhere?
64 slice: 120kV, 100 Q.Ref mAs, 64x0.6mm, pitch=1.2, CD4D: Weak/Strong
16 slice: 120kV, 100 Q.Ref mAs, 16x0.75mm, pitch=0.85, CD4D: Average/Average
DLP vs patient size for VHRCT
0
100
200
300
400
500
0 200 400 600 800 1000 1200 1400
Cross sectional area (cm²)
DL
P (
mG
yc
m)
Leeds Sensation 64
Leeds Definition 64
Leeds Sensation 16
York Sensation 16
Log. (Leeds Sensation 64)
Log. (Leeds Definition 64)
Log. (Leeds Sensation 16)
Log. (York Sensation 16)
mA variation along length of Lungman phantom
0
50
100
150
200
250
0 50 100 150 200 250 300
Image number
Ap
pli
ed
mA
Clinical VHRCT 100mAs
Siemens VHRCT 100mAs
Siemens abdomen modified to 100mAs
Siemens VHRCT modified to 80mAs
No complaints from Radiologists about the image quality on this scanner.
All Siemens scanners in Trust using 100 Q.Ref mAs for VHRCT. Can we now reduce
them all to 80mAs to match the dose on the 16 slice machine?
Scan Lungman phantom
DLP Clinical = 203
DLP Siemens = 250
DLP Abdomen = 203
DLP Siemens 80mAs = 197
Conclusions
Siemens protocols built for specific body
parts & shouldn’t be used for anything else
Raised a lot of issues for ourselves and
Siemens
Working through how to address them all
It could be happening to you
Thanks to
Gillian Ward, Becky Artschan, Dan Shaw
(Medical Physics, LTH)
Altan Senvar, Paul Playford, Anna Sedgwick
(Siemens Healthcare UK)