abbe 5 refractometer - bellingham and stanley · abbe 5 refractometer ... (35w, approx. 3000 hours...

8
Abbe 5 Refractometer FOR ACADEMIA, RESEARCH AND GENERAL LABORATORY APPLICATIONS The Abbe 5 is an affordable refractometer ideally suited for use where a wide refractive index measurement range is required such as in small contract laboratories or applications where sample throughput is relatively low. The instrument is also ideal for practical demonstrations and experiments in chemistry and physics laboratories of schools, colleges and universities. The instrument features both refractive index and Brix scales, with the prism temperature being monitored electronically and displayed on the integral digital display. Waterbath connections provide prism temperature control facilitating accurate readings to 4 decimal places RI (1 decimal Brix). Dispersion measurements of solids materials such as glass, contact lenses and fibre optics may be made in accordance with standard practice. The Abbe 5 refractometer is supplied with a glass calibration plate, contact liquid, and instruction manual and is packed in a hard flight case for easy transportation and storage. Limited warranty is 24-months. • Dual Scale • Refractive Index & 0-95 Brix • Dispersion Measurement • Optional Light Source

Upload: buidieu

Post on 27-Aug-2018

240 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Abbe 5 Refractometer - Bellingham and Stanley · Abbe 5 Refractometer ... (35w, approx. 3000 hours ) - 230V~, 50Hz ... Bellingham + Stanley has updated the original Model to meet

Abbe 5 RefractometerFOR ACADEMIA, RESEARCH AND GENERAL LABORATORY APPLICATIONS

The Abbe 5 is an affordable refractometer ideally suited for use where a wide refractive index measurement range is required such as in small contract laboratories or applications where sample throughput is relatively low. The instrument is also ideal for practical demonstrations and experiments in chemistry and physics laboratories of schools, colleges and universities.

The instrument features both refractive index and Brix scales, with the prism temperature being monitored electronically and displayed on the integral digital display. Waterbath connections provide prism temperature control facilitating accurate readings to 4 decimal places RI (1 decimal Brix). Dispersion measurements of solids materials such as glass, contact lenses and fibre optics may be made in accordance with standard practice.

The Abbe 5 refractometer is supplied with a glass calibration plate, contact liquid, and instruction manual and is packed in a hard flight case for easy transportation and storage. Limited warranty is 24-months.

• DualScale

• RefractiveIndex&0-95Brix

• DispersionMeasurement

• OptionalLightSource

Page 2: Abbe 5 Refractometer - Bellingham and Stanley · Abbe 5 Refractometer ... (35w, approx. 3000 hours ) - 230V~, 50Hz ... Bellingham + Stanley has updated the original Model to meet

Abbe 5 Refractometer

SpecificationsScale 0-95 °Brix / 1.30 - 1.70 RI

Resolution 0.1 / 0.0005

Operating temperature 5–70 °C

Temperatue resolution 0.1°C

Ambient temperature 5-40°C

Power supply (temperature) Battery (1 x 1.5V AA alkaline).Approx. life 12 months continual use

Order CodesCode Description

44-501 Abbe 5 refractometer supplied with glass calibration plate, contact liquid and instruction manual packed in a hard flight case for easy storage

44-520 Abbe 5 light source for mounting on instrument base. UK/EU plug. 230v~, 50-60Hz

44-525 Abbe 5 light source for mounting on instrument base. US plug. 110v~, 50-60Hz

44-522 Abbe 5 light source - spare bulb (pack of 5)

44-590 Spare prism box assembly

72-200 Traceable calibration plate (silica: 1.45839 ± 0.0001 RI at 20 °C)

10-43 Contact liquid (monobromonapthalene) for test plates and solid samples with an RI up to 1.65 RI. Quantity 6ml.

10-61 Contact liquid (methyl iodide) for test plates and solid samples with an RI up to 1.74 RI. Quantity 6ml.

Weights & Dimensions Weight (gross) 3.5 kg

Weight (net) 2.55kg

Dimensions (packed): 27 x 37 x 18cm (0.018m3)

InternationalLongfield Road Tunbridge Wells, Kent TN2 3EY United KingdomTel:+44(0)1892500400 Fax:+44(0)1892543115 [email protected]

www.bellinghamandstanley.com

United States90HorizonDrive Suwanee GA30024 UnitedStatesofAmericaTel:6788045730 Fax:6788045729 [email protected]

The ABBE RERACTOMETER derives its name from the 19th Century physicist and father of modern optical technology, Ernst Abbe (1840-1905). He was the co-founder of the Carl Zeiss Jena Optical Company and Schott Glasswerk.

Abbe designed an instrument for measuring dispersion which became the precursor for a refractometer – the so-called Abbe Refractometer..

Abbe5-EN/514

Page 3: Abbe 5 Refractometer - Bellingham and Stanley · Abbe 5 Refractometer ... (35w, approx. 3000 hours ) - 230V~, 50Hz ... Bellingham + Stanley has updated the original Model to meet

Abbe 60 DR & 95 RefractometersDIRECT READING MODELS FOR ACADEMIA, RESEARCH AND GENERAL USE

More stringent requirements of quality control and, in some cases, legislation, mean that greater accuracy is being demanded of refractometers. The Abbe 60 Direct Reading models, available in two measuring ranges, have been designed to meet these requirements.

The latest designs incorporate an externally mounted LED light source for sample illumination. A wavelength compensation device is used to ‘achromatise’ to the mean sodium wavelength (589.3nm). Refractive Index can be read directly from a scale divided to 0.001 allowing for estimation to ±0.0001. Temperature control can be achieved by circulation of water; prism temperature is displayed in digital form.

Direct reading models are particularly suitable for the measurement of sugar solutions and sugar content of food and drink products. They can also be used to determine the refractive index of pure liquids, solutions and formulated products including; flavourings, fragrances, oils and fats, solvents, plastics, resins, adhesives and many other industrial chemicals.

Instruments come complete with external power supply, standardisation plate and contact liquid, and instructions for use.Limited warranty is 24 months.

• Dualscale

• RefractiveIndexand°Brix

• IntegralLEDlightsource

Page 4: Abbe 5 Refractometer - Bellingham and Stanley · Abbe 5 Refractometer ... (35w, approx. 3000 hours ) - 230V~, 50Hz ... Bellingham + Stanley has updated the original Model to meet

Direct Reading Abbe 60 Refractometers

Specifications 60DR 6095

Range (Refractive Index / °Brix) 1.30 - 1.74 / 0-95 1.30-1.535 / 0-95

Scale divison (Refractive Index) 0.001 0.001

Scale division (°Brix) 0.5 0.2 (0-50)0.1 (50-95)

Precision (scale / by estimation) ±0.001 RI (±0.0001 RI) (±0.25 °Brix)

±0.2 (0-50)±0.1 (50-95)

Operating temperature 5–75 °C

Temperature resolution 0.1°C

Ambient temperature 5-40°C

Power supply 110-230V~, 50/60Hz

Order CodesCode Description

10-99 Abbe 60DR Refractometer supplied with calibration plate, contact liquid, power supply and instruction manual

10-03 Abbe 6095 Refractometer supplied with calibration plate, contact liquid, power supply and instruction manual

10-42 Volatile liquid cell

10-58 Divided cell for differential measurement

10-43 Contact liquid (monobromonaphalene) for test plates and solid samples with RI up to 1.65Quantity 6ml

10-61 Contact liquid (methylene iodide) for test plates and solid samples with RI up to 1.74 Quantity 3 ml

10-62 Contact liquid (SMI) for solid samples with RI to 1.79Quantity 10 ml

10-69 Abbe filter eye cap for use with spectral sources using appropriate interference filter. Manufactured to order - wavelngth defined at point of order

10-44 Glass test plate - approx. 1.5222 RI

10-46 Silica test plate - 1.45839 RI. Certificate of Verification showing traceabiltty available at extra cost

10-49 Polarising eyepiece for bi-refringent samples

10-59 Glass substrate for set resin samples (requires use of contact liquid code 10-43)

10-660 Abbe utility software (also available from www.bellinghamandstanley.com at no cost)

13-395 Spare desiccator

55-104 Spare power supply for Abbe refractometer (110-230V~, 50/60Hz)

The ABBE RERACTOMETER derives its name from the 19th Century physicist and father of modern optical technology, Ernst Abbe (1840-1905). He was the co-founder of the Carl Zeiss Jena Optical Company and Schott Glasswerk.

Abbe designed an instrument for measuring dispersion which became the precursor for a refractometer – the so-called Abbe Refractometer

InternationalLongfield Road Tunbridge Wells, Kent TN2 3EY United KingdomTel: +44 (0) 1892 500400 Fax: +44 (0) 1892 543115 [email protected]

www.bellinghamandstanley.com

United States90 Horizon Drive Suwanee GA 30024 United States of AmericaTel: 678 804 5730 Fax: 678 804 5729 [email protected]

AbbeDR813

Page 5: Abbe 5 Refractometer - Bellingham and Stanley · Abbe 5 Refractometer ... (35w, approx. 3000 hours ) - 230V~, 50Hz ... Bellingham + Stanley has updated the original Model to meet

Abbe 60 ED & LR Refractometers ULTIMATE PRECISION FOR RESEARCH AND ACADEMIA ACROSS THE VISIBLE SPECTRUM

The high accuracy models are designed for use with monochromatic light sources such as the sodium and spectral range available from Bellingham+Stanley. A PC program for conversion of scale readings to refractive index (RI) or Brix (Sugar%) is available as FREEWARE from the B+S website. With good temperature control and use of the micrometer Vernier Scale, an average accuracy of ±0.00004 RI can be achieved across the specified measuring range. Temperature can be controlled by use of a water bath and circulator; prism temperature can be read from the digital display on the instrument.

High Accuracy Abbe 60 Refractometers are ideal for use in academic, research and development, quality control and other industrial laboratories where the highest accuracy and precision is often sought. The Abbe 60/ED Refractometer conforms to the requirements of ASTM D1218 and is used widely throughout the petroleum industry.

Instruments come complete with external power supply, digital thermometer, traceable standardisation plate and contact liquid, sodium D line calibration tables, and instructions for use. Limited warranty of 24 months.

• AbbeNo.anddispersionmeasurement

• ConformstoASTMD1218andD1747

Page 6: Abbe 5 Refractometer - Bellingham and Stanley · Abbe 5 Refractometer ... (35w, approx. 3000 hours ) - 230V~, 50Hz ... Bellingham + Stanley has updated the original Model to meet

High Precision Abbe 60 Refractometer

Specifications 60ED 60LR

Range (equivalent to) 1.30 - 1.74 RI 1.20-1.70 RI

Precision (mean @ 589nm) ±0.00004

Operating temperature 5–75 °C

Temperature resolution 0.1°C

Ambient temperature 5-40°C

Power supply 110-230V~, 50/60Hz

Order Codes Code Description

10-04 Abbe 60ED Refractometer supplied with calibration plate, contact liquid, power supply and instruction manual

10-06 Abbe 60LR Refractometer supplied with calibration plate, contact liquid, power supply and instruction manual

59-01 Free standing sodium lamp outfit with bulb (35w, approx. 3000 hours ) - 230V~, 50Hz

59-06 Free standing sodium lamp outfit with bulb (35w, approx. 3000 hours)- 110V~, 60Hz

59-20 Free standing spectral lamp outfit (without spectral bulb) - 230V~, 50Hz

59-25 Free standing spectral lamp outfit (without spectral bulb) - 110V~, 60Hz

59-50 Spare sodium bulb (35w, approx. 3000 hours)

59-33 Mercury bulb (approx. 200 hours)

59-37 Cadmium bulb (approx. 200 hours)

59-39 Helium bulb (approx. 200 hours)

10-42 Volatile liquid cell

10-58 Divided cell for differential measurement

10-43 Contact liquid (monobromonaphalene) for test plates and solid samples with RI up to 1.65Quantity 6ml

10-61 Contact liquid (methylene iodide) for test plates and solid samples with RI up to 1.74 Quantity 3 ml

10-62 Contact liquid (SMI) for solid samples with RI to 1.79Quantity 10 ml

10-69 Abbe filter eye cap for use with spectral sources using appropriate interference filter. Manufactured to order - wavelngth defined at point of order

10-44 Glass test plate - approx. 1.5222 RI

10-46 Silica test plate - 1.45839 RI. Certificate of Verification showing traceabiltty available at extra cost

10-49 Polarising eyepiece for bi-refringent samples

10-59 Glass substrate for set resin samples (requires use of contact liquid code 10-43)

10-660 Abbe utility software (also available from www.bellinghamandstanley.com at no cost)

13-395 Spare desiccator

55-104 Spare power supply for Abbe refractometer (110-230V~, 50/60Hz)

Note: Abbe 60ED & 60LR must be used in conjunction with a monochromatic light soure such as those listed above. Adequate care must be taken when laser or other high intensity light is used. Aways read the health & safety instructions before use.

InternationalLongfield Road Tunbridge Wells, Kent TN23EY United KingdomTel: +44 (0) 1892 500400 Fax:+44(0)1892543115 [email protected]

www.bellinghamandstanley.com

United States90 Horizon Drive Suwanee GA30024 United States of AmericaTel:6788045730 Fax:6788045729 [email protected]

Abbe60EN/813

Page 7: Abbe 5 Refractometer - Bellingham and Stanley · Abbe 5 Refractometer ... (35w, approx. 3000 hours ) - 230V~, 50Hz ... Bellingham + Stanley has updated the original Model to meet

Model D7 PolarimeterFOR ACADEMIA, RESEARCH AND GENERAL USE

For over 40-years, the Model D Polarimeter and sodium lampoutfit has been the primary choice of leading industrial andacademic institutions throughout the world, offering practical assessment of optically active materials combined with high accuracy measurement.

Changes in legislation, particularly within the sugar and pharmaceutical sectors, has demanded a shift towards digital instrumentation, however, the need for a high quality opticalpolarimeter is still paramount where circumstances dictate that an optical instrument should be used. Such applications include use in contract laboratories of low sample throughput and of course the continued use in both physics and chemistry departments within the education field.

Bellingham + Stanley has updated the original Model to meet the requirements of 21st Century applications. instrument body has been replaced with a new ‘low weight’ component that reduces shipping costs, significantly important where a number of instruments are supplied to one destination and the sodium lamp has been replaced with a lightweight, low cost LED light source with an interference filter and integral intensity control. Retaining the measurement specification of the original Model D, the new D7 provides the user with one of the highest specification optical measurements of rotation at a fraction of the cost of its predecessor. Limited warranty is 24 months.

• LEDLightSource

• IntensityControl

• DualRange

• 0-360°Angularand-30to130°Z(Sugar)

Page 8: Abbe 5 Refractometer - Bellingham and Stanley · Abbe 5 Refractometer ... (35w, approx. 3000 hours ) - 230V~, 50Hz ... Bellingham + Stanley has updated the original Model to meet

Model D7 Polarimeter

SpecificationsScale 0-360 °A / -30 to 130 °Z

Sub Division (scale resolution) 1 °A / 1 °Z

Resolution (micrometer drum) 0.05 / 0.1

Precision (estimation) ±0.02 / ±0.05

Operating temperature 5-45 °C

Power supply 90-230V~, 50/60Hz

Code Description

31-02 Model D7 Polarimeter with integral LED light source, intensity control, 200mm “bubble type” sample tube, power supply and instructions for use

Poalrimeter Tubes

Code Description Length Diameter Volumen/100

35-2935-3035-28

Bubble - to clear bubble from field of viewMost suited to Model D7

10020050-200

8 5

35-4635-4735-45

Centre fill – for easy filling and placement of ADPtemperature sensor

10020050-200

8 5

35-5735-5835-56

Cup – funnel shaped centre fill for viscous samples10020050-200

8 5

35-1035-11 Metal end – centre fill for aggressive chemicals and solvents 100

200 8 5

36-7736-78

Centre fill - water jacket for temperature controlled measurements

100200 8 5

35-71 Low volume - stainless steel - Leur 50 5 2

35-76 Low volume - stainless steel - Leur - water jacket 50 5 2

All lengths in millimetres. Volumes in millilitres. All collar sizes 30mm diameter

Quartz PlatesCode Description Value

34-2034-2134-22

Standard Quartz Control Plate supplied with a Certificate of Conformity at 589nm

+100 °Z (+34.6 °A) +15 °Z (+5.2 °A)- 30 °Z (-10.4 °A)

90-803 UKAS Certificate of Calibration for Quartz Control Plate at 589nm and 20°C

Weights and DimensionsWeight (gross) 8.3 kg

Weight (net) 5.5 kg

Dimensions (packed): 55 x 39 x 28 cm

InternationalLongfield Road Tunbridge Wells, Kent TN23EY United KingdomTel:+44(0)1892500400 Fax:+44(0)1892543115 [email protected]

www.bellinghamandstanley.com

United States90HorizonDrive Suwanee GA30024 UnitedStatesofAmericaTel:6788045730 Fax:6788045729 [email protected]

Mo_D7-EN/813