abbott plum a+

85
5/01) ;g;h Chicago, IL 60064 430-95226-001 (Rev. Manuaf For use with List 11971-04 Abbott Laboratories System Operating

Upload: cillong101

Post on 02-Apr-2018

223 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 1/85

5/01)

;g;h Chicago, IL 60064

430-95226-001 (Rev.

Manuaf

For use with List 11971-04

Abbott Laboratories

System

Operating

Page 2: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 2/85

901111 430-95226-001 (Rev.

SE?TING GRAVITY FLOW ............................3-8

3.4 DISCONTINUING FLUID ADMINISTRATION ................................................................................. 3-8

3- 1

3.1 SETUP ......................................................................................................................................... 3-1

Unpacking ............................................................................................................................. 3-l

Tandem Carrier Instructions ................................................................................................. 3-2

3.2 A DMINISTRATION SETS .............................................................................................................. 3-2

Preparing the Administration Set .......................................................................................... 3-3

Priming the A dministration Set ............................................................................................ 3-4

Loading the Cassette ............................................................................................................. 3-5Preparing the Secondary Line ............................................................................................... 3-6

3.3 DISCONTINUING ELECTRONIC FLOW CONTROL

PlumSet Capabilities .............................................................................................................. 2-l

Air Management ................................................................................................................... 2-l

Battery ................................................................................................................................... 2-2

Biomedical ............................................................................................................................ 2-2

Options .................................................................................................................................. 2-2

Other Features ....................................................................................................................... 2-2

USER QUALIFICATION ................................................................................................................ 2-3

SYSTEM SELF-TESTS .................................................................................................................. 2-3

SECTION 3

INSTRUCTIONS FOR USE ........................................................................................................................

EATLJRES ................................................................................................................................... 2-l

Therapies ............................................................................................................................... 2-l

Line Programming Options .................................................................................................. 2-l

.....2-l

2.1

2.2

2.3

.......l-3

SECTION 2

DEVICE DESCRIPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

...l-2

Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .l-2

1.3 D EFINITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

..*............................................................... l-l

1.2 CONVENTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

- 1

1.1 P RODUCT OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

‘.

SECTION 1INTRODUCTION . . . .

Contents

Page 3: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 3/85

PumpOl) iv Plum A+ Infusion

6-3

Line B Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4

Malfunctions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4

430-95226-001 (Rev.

6-2General Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-2

Line A Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . .

ME.SSAGE~ . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-l

6.2 R ESPONSE TO ALARMS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

.................................................................................................... 5-15

SECTION 6ALARMS AND TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-l

6.1 W ARNING

PROG~WX~MINGULTISTEP 

B) .............................................................................. 5-5..........................

5.3 L OADING DOSE .......................................................................................................................... 5-9

5.4

(MG/HR OS

(MCG/KC~MIN ON A) ................................................................................................ 5-l

5.2 D OSE CALC

SETI-INGS ............................................................................................. 4-25

POSSIBLE NON-DELIVERY PROGRAMMED ............................................................................. 4-26

SECTION 5

THERAPIES (FOR ADVANCED USERS*)............................................................................................... 5-l

5.1 DOSE CALC

BACKPRIMING ......................................................................................................................... 4-25

CLEARING PROGRAM

1 L INE PUMPING ..................................................................... 4-23

STOP AND START WITH BOTH LINES PUMPING ...................................................................... 4-23

NURSE CALLBACK .................................................................................... 4-17

CONCURRENT DELIVERY ....................................................................................................... .4-18SELECT O PTION - VOLUMES INFUSED , PRESSURE /POST INFUSION R ATE , AND L IGHTING /C• ....... .4-20

STOP AND START WITH ONLY

RTFL4TION ................................................................................................................................ 4-10

SIMPLE DELIVERY USING DRUG LIST ..................................................................................... .4-12

4.24.3

4.4

4.5

4.6

4.7

4.8

4.94.10

4.11

4.12

4.134.14

4.15

4.16

4.17

4.18

4.19

SIMPLE DELIVERY USING DELAYED START ............................................................................ 4-14

PIGGYBACK DELIVERY ........................................................................................................... 4-15

PIGGYBACK WITH

CASECONTROLS .............................................................................................................. 4-5

POWER ON .................................................................................................................................. 4-7

SIMPLE DELIVERY ....................................................................................................................... 4-8

EXAMPLES OF AUTOMATIC CALCULATION ................................................................................ 4-9

REAR 

INDICATORS ................................................................................................................................ 4-5

..‘~ .................................. 4-3

OPERATING KEYS ............................................................................................ 4-3..........................

GETTING STARTED ...................................................................................................................... 4-l

System Self-Tests ................................................................................................................... 4-l

Data Retention ................................................................................

Contents

SECTION 4

BASIC OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-l

4.1

Page 4: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 4/85

5/01)

Enteral and H igh Viscosity Fluids ...................................................................................... 10-l

Containers ............................................................................................................................ 10-l

10.2 ACCESSORIES ........................................................................................................................... 10-l

System Operating Manual 430-95226-001 (Rev,

F~um ........................................................................................................ 10-l

Administration Fluids ......................................................................................................... 10-lADMINISTRATION  

CURVES ............................................................................................................................. 9-4

SECTION 10

SUPPLIES AND ACCESSORIES............................................................................................................. 10-l

10.1

~~LJM~ET  

BOLUS VOLUME RELEASED AFTER DOWNSTREAM OCCLUSIONS ARE CORRECTED ......................... 9-3

DELIVERY ACCURACY ....................................................................................................................... 9-4

DOWNSTREAM OCCLUSIONS ................................................................................. 9-3

RATE RANGE ................................................................................................................... 9-2

AIR-IN-LINE ALARM ......................................................................................................................... 9-2

OCCLUSION ALARM AND LIMITS ...................................................................................................... 9-3

TIME TO DETECT

BACKPRIMING ............................................................................................................................. 8-5

8.8 GENERAL .................................................................................................................................... 8-5

SECTION 9

SPECIFICATIONS ....................................................................................................................................... 9-l

PHYSICAL .......................................................................................................................................... 9-l

ELECTRICAL ...................................................................................................................................... 9-l

VTBI RANGE ..................................................................................................................................... 9-l

ENVIRONMENT ................................................................................................................................... 9-2DELIVERY

ACCESSORIES ............................................................................................................. 8-48.7

BATTERY O PERATION ................................................................................................................ 8-4

8.6 SETS A ND

PROFESSIONALS AND PATIENT R ELATED ............................................................ 8-1

8.3 CONCURRENT FLOW ................................................................................................................... 8-2

8.4 EPIDURAL A DMINISTRATION ...................................................................................................... 8-3

8.5

HEALTHCARE  

ARTIFACTS ................................................................................................................................. 8-1

8.2

................... 8-l

8.1

Contents

SECTION 7

MAINTENANCE AND STORAGE ............................................................................................................ 7-l

7.1 C LEANING , MAINTENANCE , AND STORAGE ................................................................................ 7-l

Cleaning and Sanitizing ......................................................................................................... 7-l

Battery M aintenance .............................................................................................................. 7-2

Storage .................................................................................................................................. 7-2

Service .................................................................................................................................. 7-2

SECTION 8PRECAUTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 5: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 5/85

5/01)- 1 430-95226-001 (Rev.

H km o d i a l y s i s

On co l o g y

M ob i l e n t en s i v e C u r e

N u t r i t i o n a l

OpDe c o v e r y

P e d i a t r i c s Cardiac C a t h L a b

N e o n a t o l g y Em e r g e n c y

B u r n U n i t

I CU /CCU Pos t

OR f i n e s t h e s i ae d i c a l l S u r g i c a l

L a b o r j 3 e t i v e r y l P o s t P a t t u m

& Barcode Directory.

A computer data port is also provided to interface with a host computer.

The Plum A+ is designed t o be used in most areas of patient care, including, but not limited to:

Gene r a l F l o or

identification by scanning a provided list of drug nameslocated on the Flash Tips

from multiple fluid container types.

The system includes a pumping m odule (hereafter called the pump) and an assortment of disposable IV sets(hereafter called a set), optional accessories, and this operator’s manual.

The optional Barcode Reader allows drug name

enteral or epidural infusions over a broad range ofinfusion rates

parenteral and

05 concurrent delivery. Therapymodes include:

. Standard Infusions

l Multistep Programming

l Loading Dose

l Dose Calculation

The Plum A+ is designed to deliver

A+TM Volumetric Infusion System is designed to meet the fluid delivery requirements of today’sevolving healthcare environments. The Plum A+ is a cassette based, two lines in and one line out, multi-function infusion pump. The pum p can be used for standard, piggyback.

8.

1 l Pr oduct Over view

The Plum

SECTION 1 Introduction

Page 6: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 6/85

l- 2 Plum A+ Infusion Pump/01)

.

Note: Figures are rendered as graphic representations to approximate the actual product. Therefore,figures may not exactly reflect the product.

430-95226-001 (Rev.

1This symbol directs the user to consult accompanying docum ents.a

M SERIOUS PATIENT OR USER INJURY.

Note: A Note highlights information that helps explain a concept or procedure.

RESULT  

FAXLXJRE

TO OBSERVE A CAUTION COULD

PRBVENT IRREVERSIBLE PRODUCT DAMAGE OR RARDWARE FAILURE.RAT COULDNF-ORMATION  

IT CONTAINSROCEDURE OR STATEMENT.

..sets are supplied Sterile and are fo r....

WARNINGS. CAUTI ONS, AND NOTE S

Alert messages used throughout this manual are described below. Pay particular attention to thesemessages.

WARNING

A WARNING MESSAGE CONTAINS SPECIAL SAFETY EMPHASIS AND MUST BE OBSERVED AT ALLTIMES. FAILURE TO OBSERVE A WARNING MESSAGE IS POTENTIALLY LIFE THREATENING.

CAUTION: A CAUTION USUALLY APPEARS IN FRONT OF A

I

Emphasis .

IBold

A Choose]

Sc reen d is pl ay s an d dev ic e l ab el s (as Th er ap yappropriate)

Dose Calculat ion

Softkey OptionsItalic]

Init ial Caps lowercase

3- 1, Priming Cassette)

Function or mode specif ic instruct ions Primary On/y: Attach an emptycontainer.

[ALL CAPS] Keys or touchsw itches on the deviceare described all caps in brackets

[START]

‘.

Reference to a section, figur e, or table (See Figur e

Introduction

1.2 Conventions

This section describes the conventions used throughout this manual, as follows:

Convention

Italic

Application Example

Page 7: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 7/85

5101)- 3 430-95226-001 (Rev.

Wa r n f n q .

System Operating Manual

K eu  

c ) T h a t a u n c t i on h a s b een n i t i a t e d in t h e p r o p er m a n n e r su c h a s a V a l i d

b) T h a t a n a t t e m p t h a s b een m a d e o u s e a u n c t i o n n h e w r o n g seq u en c e,

w i t h h e w r o n g v a l u es, o r a t h e w r o n g i m e su c h a s a n n v a l i d K ey W a r n -i n g .

- T h e a m o u n t o f f l u f d r em a i n i n g o b e d el i v er e d o h ev a t f e n t .

Wa r n i n g A n n d i c a t i o n o a d v i s e h e o p er a t o r :

a ) O f a p o ssi b l y h a z a r d o u s co n d i t i o n su c h a s a L ow -B a t t er y w a r n i n g .

VTB I V o l u m e T o B e n f u sed

Dfs -

I n f u s i o n Ra t e t a l P r essu r e L im i t . I t s a l so u sed t o vi ew D i s t a l P r essu r e in r ea l - t ime .

U n i t s A p r ed et er m i n ed d el i v er y q u a n t i t y a d o p t ed a s a st a n d a r d o f m ea s u r e .

mo d e s u s ed o set KVO R a t e a n dn f l t h i sOp t fons lVo l  r e ssu r e lPo s t S el e c t ed u s i n g

Open ) c a t h e t e r .

L i n e A Ca sset t e n l e t i n e o n h e o p c en t er o f h e c a s set t e w h en o o k i n g a t h e c a s -set t e r om h e r o n t o f h e D ev i c e .

L i n e B Ca sset t e n l et i n e o n h e o p r i g h t o f h e c a s set t e w h e n o o k i n g a t h e c a s -

set t e r om h e r o n t o f h e D ev i c e .M a l f u n c t i o n F a i l u r e o f h e d e vi c e o p a ss n t e r n a l n t eg r i t y o r p e r f o r m a n c e est s.

Pfgg ybac k T h e m o d e w h er e i n e B d e l i v er s i r st , o l l o w e d b y i n e A .

f l u i d a t a sl o w r a t e n or & r o p r ev en t oc cl u d i n g h e p a t i e n t

f l u i d f l o w .

KV O (K eep Ve i n D el i v er y o f

V T B I a t a d el i v er y r a t e set b y h e op e r a t o r .

E r r o r A c on d i t i o n r esu l t i n g w h e n a n o p er a t o r a t t em p t s a n o n a c cep t a b l e n p u t ,su c h a s a n n v a l i d k e y p r ess.

F l o w R a t e T h e r esu l t i n g r a t e o f

f n d e p e n d e n t j l o w r a t e s .D e l i v e r y

D e l f v e r y D e l i v er y op t i o n w h i c h a l l o w s i n es A a n d B o d el i v er c on c u r r e n t l y .

T o g e t h e rD e l i v er y R a t e T h e D el i v er y r a t e set by h e op er a t o r .

D u r a t i o n T h e i m e p e r i o d u sed o d e l i v er a

j l u f d d e l i v er ed n a d i sc r e t e

p e r i o d o f i m e.

A r ec o g n i z ed c a s set t e y p e n st a l l ed w i t h h e d o o r cl o s ed .

C o n c u r r e n t S i m u l t a n eo u s d el i v er y of w o f l u i d s a t

N o f l u i d

i s d e l i v er ed d i st a l o h e c a sset t e d u r i n g b a c k p r i m i n g .

A si n g l e u n i n t er r u p t e d d i s cr e t e v o l u m e o f

o r f l u i d r o m h e p r o x i m a l i n e a n d c a sset t e a i r r a p .

condf-

t i o n s w i l l c a u s e b ot h a u d i b l e a n d v i s i b l e a l a r m n d i c a t o r s. C on f i r m a t i o n o ft h e c on d i t i o n s u s u a l l u r e a u i r e d .

B a c k p r i m e

B o l u s

Cas s e t t e

I n s t a l l e d

R em o va l o f a i r

ope r a t o r . a t t e n t f o n . T h e s e

Definitions (General and C linical)

1.3 Defi ni t ions (General and Cl i ni cal)

A l a r m s A n y c on d i t i o n o f h e p u m p h a t r e q u i r es

Page 8: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 8/85

5/01)- 1 430-95226-001 (Rev.

Removal/Backpriming

l Air Detection-Proximall Air Detection-Distal

PLUMSET CAPABIL IT IES

l Anti F ree-F low Protection

l Direct Connection for syringe delivery

AIR MANAGEME NT

l Air Trap

l Air

flow rate range for both lines

l Automated Secondary drug delivery (Piggyback)

mL increments)L/hr (in 1oo-999

mL increments) flow rate range for both l ines

l Macro

mL/hr (in 0.1

LifeShield@ needleless protection systems makes the Plum A+ a convenient and cost-effectiveinfusion system.

THERAP IES

l Dose Calculation

l L oading Dose

l Multistep Programming

L INE PROGRAMMIN G OPT IONS

. Therapy Selection

l Nurse Call Back

l Delayed Start Setting

l Drug Label L ibrary

. Concurrent delivery

l Titration

l M icro 0.1-99.9

LifeCare Plum Series adm inistration sets and accessoriesand the

healthcare. standardization. With its primary line, secondary line,and piggyback fluid delivery capability, the Plum A+ is suited for a wide range of medical/surgical andcritical care applications. Full com patibility with

hospital-wide, as well as alternate site and home

SECTION 2

2.1 Features

The Plum A+ is a dual-line volumetric infusion system designed to meet the growing demand for

Page 9: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 9/85

2- 2 Plum A+ Infusion Pump/01)

ml/hr or less depending on delivery rate) or Continue Rate (CR) tocontinue at the current rate

l Variable Distal Pressure Settings

OTHER FEATU RES

Nonpulsatile volumetric accuracy

M icroprocessor control

Large liquid crystal display (LCD) screen

Panel back il lumination on mains power

L ockout switch

Standard full fill , partfi ll , syringe and vial useParenteral and nonparenteral (enteral) fluid delivery

Blood and blood products delivery

Wide range of Standard and Specialty administration sets

430-95226-001 (Rev.

1

l Plug-in Bar Code Reader for drug identification (Optional)

l Nurse Call Relay Connector

OPTIONS

l Volumes Infused (A, B, Total Volume)

l K VO at dose end (1.0

tempprary portable operation

BIOMEDICALl Serial Communication

l Upgradabil ity (F ield)

l Variable Rate Cap

. Alarm History

Six Hour Battery Recharge Time

l Long battery life (6 hours) for emergency backup and

Device Description

BATTERY

l Battery Gauge

l

Page 10: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 10/85

5/01)- 3 430-95226-001 (Rev.

con.fii readiness for use.

Failure during the Self-Tests is reported as a Malfunction Condition.

System Operating Manual

The pump performs a suite of System Self-Tests when power is applied to

APPROVEiD LABELING). EPIDURALADMINISTRATION OF DRUGS OTHER THAN THOSE INDICATED FOR EPIDURAL USE COULDRESULT IN SERIOUS INJURY TO THE PATIENT.

2.3 System Self-Tests

enteral, and epidural fluids and drugs and whole blood or red blood cell components. The training shouldemphasize preventing related IV complications, including appropriate precautions to prevent accidentalinfusion of air. The epidural route can be used to provide anesthesia or analgesia.

WARNING

ADMINISTER ONLY ANESTHETICS/ANALGESICS APPROVED FOR EPIDURAL ADMINISTRATION(AS INDICATED OR ALLOWED BY THE DRUGS’ FDA

User Qualification

2.2 User Qualif ication

The Plum A+ is intended for use at the direction or under the supervision of licensed physicians or certifiedhealthcare professionals who are trained in the use of the pump and the administration of parenteral,

Page 11: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 11/85

5/01)- 1 430-95226-001 (Rev.

SROULD

BE AVOIDED.

PUMP UND ER THESE CON D ITIONSHE  ICROCOMPUTER MIGHT BE D ISR UPT ED . O PE RA TION OF

2-WAY RADIOS, IT IS POSSIBLE THAT THE NORMAL OPERATION OF A SENSOR

OR

m TRE EVENT OF EXTREME

LEVELS OF INTERFERENCE, SUCH AS ENCOUNTERED NEXT To AN ELECTROSURGICAL GENERATOR.

CELLULAR TELEPHONES, OR

(EMI) CONDITIONS.

TRB PRESENCE OF MOST

ENCOUNTERED ELECTROMAGNETIC INTERFERENCE

JN  

ABBOTT -ORATORIES.

To set up the Plum A+, connect the mains cord to a properly earthed mains receptacle, unless temporarybattery operation is desired.

Note: Use mains power whenever possible. Store the Plum A+ connected to mains to confirm a fullycharged battery for em ergencies.

Note: If the quality of earth grounding source is in doubt, use battery power.

CAUTION: THE PLUM A+ SYSTEM IS DESIGNED TO OPERATE NORMALLY

PLUM A+ APPEARS TO BE DAMAGED, CONTACT=  F 

TRE BATTERY MAY

NOT BE CRARGED UPON RECEIPT.

Inspect the Plum A+ packaging for possible shipping damage. If damage is found, contact the deliverycompany immediately.

Use care when unpacking the Plum A+. Retain the packing slip and save all packing material in case thePlum A+ is damaged or fails the pump self-test and has to be returned to Abbott Laboratories.

Inspect the Plum A+ thoroughly for damage.

CAUTION:

IF IT APPEARS DAMAGED IN ANY WAY.LJJM A+RE

UNPACXING

AND INSTALLATION. Do NOT USE

DURING  XBRCISED  NLESS PROPER CARE ISAY OCCUR

‘.

This section explains the Plum A+ setup and administration set use.

3.1 Setup

WARNING

ARRANGE TUBING, CORDS, AND CABLES TO MINIMIZE THE RISK OF PATIENT STRANGULATIONOR ENTANGLEMENT.

UNPACKING

CAUTION: PR ODUCT D A MAGE

hstructions for UseECTION 3

Page 12: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 12/85

-~Plum A+ Infusion Pump- 2/01)

=EiFinger Gr ip

Pumping Chamber

430-95226-001 (Rev.

87 

IK!B

Flow Regulator

4

Line-nrnaty  

LifeCare Plum series administration sets. Become familiarwith the set components before preparing the administration set.

(List# 12276-O 1) to the pole before

attaching the pump(s) to the tandem carrier.

Make sure tandem carrier clamp is attached firmly to pole.

Make sure pumps are firmly attached to tandem carrier.

3.2 Administration Sets

The Plum A+ is compatible with the wide range of

TANDEM CARRIER INSTRUCTIONS

The Plum A+ may be safely and conveniently mounted on an IV stand or on a tandem carrier mounted on anIV stand.

Always mount the tandem carrier

Instructions for Use

Page 13: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 13/85

5101)- 3 430-95226-001 (Rev.System Operating Manual

l /2 full or to score mark. Do not

Note:To avoid getting Distal Air Alarms, confirm that the fluid bag is positioned higher than thepump. This should be set up prior to priming the cassette.

th$n proceed as follows:

Make sure flow regulator is closed.

Insert piercing pin into container outlet with a twisting motion.

Fill drip chamber

completely fill.

to about

Administration Sets

PREPARING THE ADMINISTRATION SET

Note: For detailed instructions, see administration set packaging.

Use the aseptic technique to prepare the administration set for priming,

Page 14: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 14/85

3- 4 Plum A+ Infusion Pump/01)

Instructions for Use

PRIMIN G THE ADMINISTRATION SET

To prime the administration set, proceed as follows:

1Invert the cassette.

Turn the flow regulator counter-clockwise until a drop of fluid is

seen in the pumping chamber.

Turn the cassette upright, then prime the remainder of the

administration set.

Push in the flow regulator to close. Confirm there is no flow.

430-95226-001 (Rev.

Page 15: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 15/85

5101)- 5 430-95226-001 (Rev.

.T,.

Open the cassette door by lifting up the handle.

Insert cassette into door guides. Close the cassette door. Confirm

there is no flow and no kinks appear in tubing.

Note: If flow is observed, close the tubing clamps and replace the administration set.

Note: The adm inistration set should be changed per CDC guidelines or healthcare provider policy.Discard after use.

System Operating Manual

.k

Administration Sets

LOADING THE CASSETTE

To load the primed cassette into the Plum A+, proceed as follows:

Page 16: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 16/85

3- 6 Plum A+ Infusion Pump/01)

5 o f

th is m anual.)

430-95226-001 (Rev.

4 an de&ions  

mL of fluid

into the syringe to clear air from the adapter filter.

To prepare the secondary line, use aseptic technique and proceed as follows:

With cassette door closed: loosen and remove white cap, then

discard. Add secondary tubing or syringe.

Open all proximal clamps before pressing [START].

Set up pum p per operating instructions. (See

vial cap when using vials on the secondaryline

l When using a syringe adapter, retract the plunger to draw approximately 1

1198648) to syringes smaller than 10 cc. (Syringes 10 cc or

larger can be attached directly to the secondary port of the cassette.)

l Syringes smaller than 3 cc should not be used.

l Attach a vial adapter (List 1188448) to the

19- gauge, 31 mm (or shorter) needle, if

the cassette secondary inlet port has a piercing reseal

l Attach syringe adapter (List

18- or

’efore preparing the secondary line, observe the following guidelines:

l Review the backpriming function

l Attach the secondary container using an

,

The Plum A+ features concurrent or piggyback delivery modes when therapy requires administering morethan one fluid.

In addition to standard containers, the Plum A+ can use syringes or vials on the secondary port. Thesecondary line can be prepared w ithout removing or repriming the cassette.

COEXPATIBILITP, CONCE NTRA TION ,

DELIVERY RATES, AND VOLUMES ARE ALL SUITABLE FOR CONCURRENT DELIVERY OR PIGGYBACK DELIVERY

(SECONDARY FOLLOWED AUTOMATICALLY BY PRIMARY) MODES.

THE DRUG LABELING TO CO N F IR M DRUG

Instructions for Use

PREPARING THE SECONDARY LINE

CAUTION: C ONS UL T

Page 17: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 17/85

5/01)- 7 430-95226-001 (Rev.

Via l : Attach the vial adapter to the secondary

adapter into the vial, if necessary.

port. Backprime the air from the vial

System Operating Manual

Administration Sets

V i a l o r S y r i n g e : Secure the container to the cassette door using the

op t i o n a l container support arm (List 12095-04-01).

Page 18: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 18/85

3- 8 Plum A+ Infusion Pump/01)

@ CLOSE DOOR

CASSETTE

DISCARD SET AND FLUIDCONTAINER PER HOSPITALPROCEDURE

430-95226-001 (Rev.

0 OPEN DOOR, REMOVE

@ PRESS STOP

TO PREVENT ACCIDENTAL FREE FLOW

NOTE: IF DOOR OPEN ALARM OCCURS, TURN PUMP OFF

3.4 D I SCONT I NU I NG FL U I D ADM IN ISTRAT ION

CASSElTE SHOULD BEIN UPRIGHT POSITION

REGUIATORCOUNTER-CLOCKWISE

NOTE:

GRAVITY RATE BYTURNING FLOW

@ SET

& REMOVE CASSETTE

NOTE: FLOW REGULATOR CLOSESAUTOMATICALLY WHEN DOOR IS OPENED

@ OPEN DOOR

Op_PRIBlARY AND SECONDARY

FLUIDS.

If electronic flow needs to be stopped, the precision flow regulator on the cassette can be used to manuallyset and maintain fluid flow in a gravity mode (see the f o l l o w i n g g r a p h i c s) .

S~ONDARY SET OR R EMOVE

SECONDARY CONTAINER FROM SECONDARY PORT TO PREVENT MIXING

PRIMARY ORLAMP ON

Instructions for Use

3.3 D I SCONT I NU I NG EL ECTRON IC F L OW CONTROL &

SETT I NG GRAV I TY F LOW

CAUTION: B EFORE OPE N IN G DOOR , CLOSE

Page 19: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 19/85

5101)- 1 430-95226-001 (Rev.

+ Silence Key

e Selection Key

DOWN/RIGHT

UP/LEFT

- Numeric Keys

Softkeys

+ LCD

THE SYSTEM SELF -TES TS .

Connect the m ains cord to mains power, then confirm that the Charging/Line indicator illuminates (abovethe [ON/OFF] key on the front panel).

tine B Flow Indicator

FAILS LUM A+ IN SER VI CE IF ITRE

-Decimal Point Key

Basic Layout of Front Panel Display and Keypad

4.1GETTING STARTED

This section details the Plum A+ instrument setup procedures.

SYSTEM SELF -TESTS

CAUTION: Do NOT P LACE

Y

___)

Numeric Field Clear Key

k

ON/OFF Key

moom

Charging/LineIndicator

- IrlrStart and Stop Keys

I

ge Region

xkA

Messa!

j

Status Region

Working Region

Softkeys Label Region

e-b

_I_)

Message Region

____)

Working Region

SECTION 4 Basic Operation

Line A Flow Indicator

Status Region

Page 20: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 20/85

4- 2 Plum A+ Infusion Pump/01)

and T r o u b l e s h o o t i n g ) .

Start pump. If the self-test alarm recurs, remove the Plum A+ from service and contact the local AbbottLaboratories representative.

430-95226-001 (Rev.

Alarms  e c t i o n 6,

MAINTAINBD AND CHARGED BATTERY H ELPS CONFIRM PROPER OPERATION.

To confirm battery is fully charged, connect the Plum A+ to AC power for a minimum of six hours while inthe OFF mode.

If an alarm occurs during the power on self-test, identify the alarm message, then take corrective action(see

APROPERLY

L.oaciing Cassette n Section 3 for details of correct procedure.

When the Charging/Line indicator is off and the battery symbol on the display is flashing, this indicatesthat the Plum A+ is operating on low battery power and should be recharged.

CAUTION: Do NOT OPERATE THE PLUM A+ ON PATIENTS WITH THE -BATTERY REMOVED. USE OF A

safeiy systems is performed whenever the Plum A+is turned on, to verify readiness for operation.

Note: If the quality of earth grounding source is in doubt, use battery power.

Press the [ON/OFF] key to turn the power on. Check screen display and listen for a beep to indicate theaudio is working. Wait for the self-tests to complete. If successful, put a fully primed macro cassette into the

cassette door and close the door. See

Basic Operation

A systematic self-testing of the processing, delivery, and

Page 21: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 21/85

5/01)- 3 430-95226-001 (Rev.

lf“double beep” is not heard, the pump will turn itself back onthinking the AC power was lost and revert to battery back-uppower.

The [START] key is used to begin infusion and as finalconfirmation of programming.

The [STOP] key is used to stop the current delivery onone line; used in conjunction with soft keys to stopdelivery on both lines.

The [SILENCE] key is used to silence an audible alarmduring actions to correct its cause.

System Operating Manual

confirm~proper shutdown of pump, do not unplug

until “double beep” is heard (approximately 5 seconds).

Operating Keys

DATA RETENTION

Delivery program settings and programming option selections are retained in non-volatile memory. If thePlum A+ has been turned off for longer than four hours, all delivery settings are cleared and programmingoption selections are restored to their default selections for next use.

4.2Operating Keys

The [ON/OFF] key is used to control the power of thePlum A+.

Note: All deliveries must be stopped in order to turn thepower off.

Note: To

Page 22: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 22/85

4- 4 Plum A+ Infusion Pump/01)

softkey.

430-95226-001 (Rev.

Softkeys are located at the bottom of the main displayand serve a variety of functions. What each key doesis indicated by the text in the display above the actual

mL).

Basic Operation

The [SELECT] key is used to move the highlighting

cursor between the programming fields.

The numeric keys are used to enter values for anyhighlighted field requiring numeric data.

The [CLEAR] key will cause a highlighted numericfield to be cleared in p reparation for data entry.

The [DECIMAL POINT] key is used to manually enternumbers other than who le numbers (i.e., 1.2

Page 23: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 23/85

5101)Rev. - 5 430-95226-001

car&m each lockout activation.

Note: If the lockout switch is put in the “Locked” positionwhile the pump is off, the [ON/OFF] key allows the unit toturn on.

System Operating Manual

a key while lockout is active results in thedisplay of the lockout enabled message. This action isrecommended to

“up” position, disabling all front panel

keys except [STOP].

Note: Pressing of

lqdicators

4.3 ndicators

CHARGE

The Charge/Line Indicator is illuminated to indicate

the battery is currently charging.

Note: If the device is plugged into AC power, with a batteryinstalled, and the Charge Indicator is not illuminated, contacttechnical support.

The fluid drop symbol, when flashing, indicates anactive infusion on Line A and/or Line B.

Battery Capacity Symbol (located on right side of

message region on display) indicates relative chargewith num ber of white boxes from left to right.

This example indicates the battery is about two-thirdsfull.

4.4 Rear Case Controls

The audio level rotary knob, located on the back,adjusts the alarm audio level.

The toggle switch, located below the audio levelcontrol, activates the lockout function when the leveris placed in the

Page 24: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 24/85

4- 6 Plum A+ Infusion Pump/01)

while’infusing.

Barcode Wand (optional) is used to scan desired drugnames from provided list. It plugs into the Barcode

Wand Port, located in the back of the pump.

430-95226-001 (Rev.

DataPort is used to interface with a host

computer.

Note: Do not operate

<

The RS-232

0The connector located below the lockout switch is theNurse Call Relay Connector, for providing remotenotification of a nurse call event.

0

Page 25: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 25/85

5/01)- 7 430-95226-001 (Rev.

VTBI onboth lines, as well as setting items, such as the KVO and NurseCallback options, to their default values.

Note: If no settings currently exist or all settings are zero, thisscreen will be bypassed.

System Operating M anual

Program]when changing ONLY Line Aor B.

Note: This clears all delivery parameters, such as Rate and

[c&f 

[Yes]to clear all settings.

Clear ALL settings when cha nging patients or multipleprogramming parameters.

Use

PLIRI A+ ON PATIENTS WITH THE

BATTERY REMOVED. USE OF A PROPERLY MAINTAINED AND CHARGED

BATTERY HELPS CONFIRM PROPER OPERATION.

2. To set up the Plum A+, insert selected Plum Set and closecassette holder door lever.

Note: If cassette is already inserted, this screen is skipped.

Example of Clear Settings Screen

3. The on-screen message “Clear Settings?” may appear.Press A

THB

DATAPORT MAY NOT OPERATE PROPERLY.

After self-testing successfully completes, the display willproceed to the next screen.

Note: If battery is missing or defective, a “Battery not installed”message with a ‘Continue without battery” prompt will appear (ACpower only).

CAUTION: Do NOT OPERATE

c

CAUTION: THE PUMP MUST BE INITIALLY STARTED ON AC POWER,

OTHERWISE THE

Power On

4.5 Power On

SETUP

Clear Settings?

1. Press [ON/OFF] to turn on the Plum A+ and you w ill seethe System Self-Tests screen.

Page 26: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 26/85

Iufusien Pump- 8 Plum A+/01)

mmum

minutes is 59.

8. Use keypad to change minutes.

430-95226-001 (Rev.

When all parameters are displayed, changing any oneparameter will result in other parameters automatically beingcomputed. To understand the Automatic Calculation feature, seeSec t ion 4.7 Examp l es o f Au t oma t i c Ca l c u l a t i o n ,

7 . Use [SELECT] to change from hours to minutes.

Note: The maximum hours allowed is 99 and the

99:59, dashes (- -)appear in place of hours.

5 . Use [SELECT] key to move dow n to the DUR ATION field.

6. Use keypad to change the hours part of duration ifdesired.

Note:

VTBI amount, use keypad.

Note: While the VTBI is being entered, the duration is automaticallycomputed. If the computed value is greater than

VTBI field.

Note: Upon entry to this screen, and to all lower screens from hereon, the Barcode Reader (BCR) is enabled as indicated by “WandActive” when attached to the Plum A+. To use the BCR, swipe tip ofBCR w and across entire barcode, from dot to dot, on the Abbottsupplied drug list card. The drug name also can be selected using A[Program Options].

4 . To enter

hEmin

Examp le of Main Delivery Screen (with cleared settings).

Note: For clarity, whole number font sizes will appear larger than

partial number font sizes (numbers on the right-side of the decimal).

1. To Program a delivery rate, press A [ A ] .

2. To enter delivery rate, use numeric keypad.

3. Use [SELECT] key to move down to the

DO5  

hrzmin

Duration

0O:OO 

=mL

Duration

VTBI

mLhr

hnmin

Rate 5.5

0O:OO 

lnlsroppEolsroppEDls1

Duration

Basic Operation

4.4 Simple Delivery

Page 27: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 27/85

5101)- 9 430-95226-001 (Rev.

VTBIEnter

DURATION[RATE]

System Operating Manual

VTBI [DURATION]

Enter

VTBI [DURATION]

Enter

DURATION

Enter

DURAT/ONl

Enter

DURATIONKeep RATE

Change RATE Enter

Kee0 RATE ITBI

[AUTOCALC]

Enter

Action

DUFUUION=OO:OO

1 st Action 2nd

\rrBl=O.O

PATE]

AT KVO

I~~~!fONTB l  

VTBI [RATE]

Change

VTBI Keep RATE [DURATION]

Change

DURATION

Keep

VTBI DURATION]

Change

) DURATION]

Change RATE Change

VTBIKeep

W-W

W HILE RUNNING

(T ITRATION )

1 st Action 2nd Action [AUTOCALC]

Change RATE

I

PATE]

Enter RATEEnter

DURATION

VTBIEnter

DURATION

1 DURATION]

Enter

VTBIEnter

[AUTO&UC]

Enter RATE

1st Action 2nd Action

4.7 Examples of Automatic Calculation

A T S TARTUP

%z

A A A A ,

B  

jusf,after beingprogrammed and started.

Note: Th e Line A Fluid Flow Indicator also will flash to indicatepumping.

A

AN D DURATION BEFORE

PRESSING [START].

9. Press [START] to begin infusion.

Example of Main Delivery Screen

mLlhr

CAUTION: VERIFY RATE, VTBI,

lbte 0.0

I

5.5

I BSTOPPEDPUMPING

Examnles of Automatic Calculation

A

Page 28: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 28/85

4- 10 Plum A+ Infusion Pump/01)

f one minute. As shownin this example, the delivery shows 3 hours and 20 minutes (not 3minutes and 20 seconds).

Note: [START] accepts the change. Delivery is changed to the newRATE and DURATION and display returns to Main D elivery Screen.

430-95226-001 (Rev.

I n n .

THE FOLLOWING 3 SCREENS ARE EXAMPLES OFCHANGED VALUES.

EXAMPLE OF CHANGED RATE.

4 . From the M ain Delivery Screen, press A [ A ] .

5 . Use keypad to change the RATE value.

Note: The display time is only accurate to

[Op t i ons /V01 

PRESSING [START].

3. After any change, press [START] to initiate new Rate andreturn to the Main Delivery Screen.

Note: Volume Infused is an incrementing value which is cleared in

AN D DURATION BEFORE

VTBI. To understand the AutomaticCalculation feature, see Sec t i o n 4.7 E x am p l e s o f A u t om a t i c

C a l c u l a t i o n ,

CAUTION: VERIFY RATE. VTBI.

[BJ at the Main Delivery Screen.

2 . Enter the desired Rate using the keypad.

Note: When Rate is changed, the duration is automaticallyrecomputed using the current

hrmin

1. Select Line A [A ] (or A

03:20  

\ A A A A,

Duration

%z 

03:20 h

A

,AlAlAlA]

Duration

A

Basic Operation

4.8 Titration

r

I

Page 29: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 29/85

5101)- 11 430-95226-001 (Rev.

AN D DURATION BEFORE

PRESSING [START].

System Operating Manual

[SI’ART] accepts the change. Delivery is changed to the newDURATION and RATE and display returns to Main Delivery Screen.

CAUTION: VERIFY RATE. VTBI,

O f A u t o m a t i cC a l c u l a t i o n ,

Note:

4.7JZxampIes  VTBI value. To understand the

Automatic Calculation feature, see Section

[STAKE?] accepts the change. Delivery is changed to the newVTBI and DU RATION and display returns to Main Delivery Screen.

EXAMPLE OF CHA NGED DURATION FIELD.

8. Use [SELECT] to highlight the Duration field, then use thekeypad to change it’s value.

Note: When D URATION is changed, RATE is automaticallyrecomputed using the current

Examples o f A u t o m a t i c Calculation,

Note:

withrespect to the current Rate. To understand the Automatic Calculationfeature, see Section 4.7

VTBI is changed, the Duration is recomputed

VTBI:value.

Note: When the

VTBI field.

the keypad to change the

VTBI VALUE.

[SELECT] key to highlight the

DO:mhnmin

6. Use

7. Use

EXAMPLE OF CHANGED

hrzmin

Duration

06:40  

Titration

Duration

Page 30: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 30/85

4- 12 Plum A+ Infusion Pump/01)

VTBl field and enter avalue. The Duration will be autom atically calculated.

4. Before or after entering the simple delivery program,press A [Program Options].

The Default screen for Program Options appears.

5. Press A [Drug List].

430-95226-001 (Rev.

[Therapy],and then A [Program O p t i o n s ] . If you select a Drug Name first, it willbe cleared upon entry in Therapy mode.

The Line A Programm ing Screen appears.

2 . Enter a RATE using the keypad.

3. Use [SELECT] to highlight the

Calc) you

must first select which line you want (A or B), next select A

[f3j) at the Main Delivery Screen,

Note: To select a Drug Nam e using a Therapy (i.e., Dose

hcmin

Callback No

1. Select Line A [A] (or A

lO0  tart for

wud&aivl

Delay

1 Program Options

hrmin

A

WmL

Duration 50:00

VTBI

hrzminO:OO 

mVhr

Duration

m  

““,:“’ 0.0

Rate

Jp,& 0.0

0.0

I

0.0

I BfOPPED STOPPEDA

Basic Operation

4.9 Simple Deli very u sin g D rug L ist

Page 31: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 31/85

5/01)- 13 430-95226-001 (Rev.

AND DURATION BEFORE

PRESSING [START].

10. Press [START] to confirm the programming, begin thedelivery, and return to the Main delivery screen.

Notice the drug name remains in the upper-left corner.

System Operating Manual

[Enter]to indicate selection and return to theProgram Options screen. Or, press A [Cancel/Back] if nochoice is acceptable.

Drug name is now visible in the upper-left corner.

9. Press A [Enter] or A [Cancel/Back] to return to the LineProgramming screen.

CAUTION: VERIFY RATE, VTBI,

U p ] ) until itshows, then use [SELECT] to highlight choice.

8. Press A

desired’drug name.

7. When list does not contain desired generic drug name orgroup name press A [Page Down] (or A [Page

VZ: “ ’ 0.0

The Drug L ist Screen appears, in mainly alphabeticalorder.

Note: Default entry is “No Drug Selected”. This selection may beused later to remove an incorrect or unwanted drug nam e from thedisplay.

6. Use [SELECT] to highlight

6.0

0.1

r&k,

Dc+mlwd

5.0

1 BSTOPPEDPUMPING

hcmir:

Callback No

A

form:00 

wmld*ntw

Delay Start

aopamm1 Program Options

Dotutamine

A

Diltiiem

Diuretic

Dexmetatomidinecytarabine

Cydogporin

Wnd_

Amikacin

Program Drug List

(d-PA)lteplase Alpnstadil

Alfentanil

Simple Delivery using Drug List

Page 32: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 32/85

4- 14 Plum A+ Infusion Pump/01)

F'rogrcun Options]. If you program the D elayedStart first, it will be cleared upon entry into Therapy mode.

The Default screen for Program Options appears.

2. Enter values for desired hours and/or minutes.

3. Press A [Enter-to confirm delayed start.

CAUTION: VERIFY RATE, VTBI, AND DURATION BEFOREPRESSING [START].

4 . Press [START] to confirm programming.

Line A is delayed for the time en tered.

Note: Delayed line drip ind icator is solid, not flashing. It will beginflashing when actually pum ping.

430-95226-001 (Rev.

[Therapy], and then AB), next select A

Calc)you must first select which line you want (A or

,

Note: To program a Delayed Start using a Therapy (i.e., Dose

Vd’nf 0.0

1. Before or after entering the simple delivery program,press A [Program Options].

mL-O  

O-Ot%r 

1

5.0

I BSTOPPEDDELAYEDA

hrzminoa 

BOO hr:min

Delay Start for

Wud_

Delay Start for

1 Program Options

BmL

Duration 50:00 hr:min

A

VTBI

IL/hr  Rate 5.0

Basic Operation

4.10 Simple Deli ver y Usin g Delayed Start

Page 33: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 33/85

5101)- 15 430-95226-001 (Rev.System Operating Manual

VTBI field.

I P r w r c u n

O p t i o n s ] .

4 . To enter delivery rate, use numeric keypad.

5. Use [SELECT] key to move down to the

BCRwandacross entire barcode, from dot to dot, on the Abbott supplied druglist card. The drug name also can be selected using A

Plum A+. To use the BCR, swipe tip of

Mode]toswitch to “Piggyback”.

3. Press [SELECT] to highlight the RATE field.

Note: Upon entry to this screen, and to all lower screens from here,the Barcode Reader (BCR) is enabled as indicated by “Wand Active”when attached to the

‘1

The Delivery Mode field is highlighted.

Note: If an entry was previously programmed, RATE is highlighted.If no previous entries, MODE is highlighted.

2. If field reads “Concurrent”, press A [Change

[Bl.t&rogram a Piggyback

delivery, press A

will stop L ine A and infuse L ine B

until VTBI completes. Line A will automatically restart.

1. With pumping started on Line A,

PIGGYBACK DEL IVERY - I nfusion

0O:OO hr:min

0090 hr:min

Duration

vy 0.0

Duration

z;Er 0.0

20.0

I

5.5

1 BUMPING1 STOPPEDA

I Piggyback Deli ver y

Piggyback Delivery

4.1

Page 34: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 34/85

4- 16 Plum A+ Infusion Pump/01)

/Yes] to continue.

Examp le of Main Delivery Screen just after beingprogrammed and started.

430-95226-001 (Rev.

4-18).

10. Press A

-)appear in place of hours. To understand the Automatic Calculationfeature, see Section 4.7 Examples o f A u t o m a t i c C a l c u l a t i o n ,

7 . Use [SELECT] key to move dow n to the DUR ATION field.

8. (Optional) Use keypad to change duration.

When all parameters are displayed, changing any oneparameter w ill result in other pa rameters automaticallybeing computed.

CAUTION: V ERI FY RATE. VTBI. AND DURATION BEFO RE

PRESSING [START].

9. Press [START] to begin infusion.

Note: If Line A has a Delay Start remaining, a message will appearwarning of a possible non-delivery (see s e c t i o n

99:59, dashes (-

VTBI amount, use keypad.

Note: While the VTBI is being entered, the duration is automaticallycomputed. If the computed value is greater than

“ “ , : ” 0.1

6. To enter

133

20.0

m%“, , 

1

5.5

1 BPUMPINGDELAYED

hnmin

A

OOa  

mL

Duration

ml 100

mL/hr

hcmin

Mode Piggyback

Rate 133

00:30  

Basic Operation

Mode Piggyback

Duration

Page 35: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 35/85

5/01)- 17 430-95226-001 (Rev.

PRESSING [START].

3. Press [START] to confirm program setting and return toMain Delivery screen.

At end of P iggyback delivery, with Callback enabled forthe line, Callback alarm is issued.

4. Press [SILENCE] to clear alarm.

System Operating Manual

AND DURATION BEFORE

ofYes or No is a biomedical setting and mustbe configured as described in the Technical Service Manual.

CAUTION: VERIFY RATE. VTBI,

“Yes”, then press A [Enter].

Note: Callback default

,

Note: To set a nurse callback, the pump m ust be stopped. This

function can be selected either before or after entering the Piggybackdelivery program.

The default screen for Program Op tions appears.

2. Use [SELECT] to highlight “Callback” field and press A[Yes/No], if necessary to obtain

hcmin

1. Press A [Program Options].

00:30  

mL

Duration

m T31

mL/hr

wmdA0.r.m

Mode Piggyback

Rate 200

1 B

I

PROGRAM

CallbaCk

htzminO:oO  

MIU

1elay Start for

wuld 

DC:30 hcmin

Program Options B

mL

Duration

m  TBI

mUhr

&%++a

Mode Piggyback

Rate 200

v&d  (B

Piggyback with Nurse Callback

4.12 Piggyback with Nur se Call back

PROGRAM

Page 36: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 36/85

4- 18 Plum A+ Infusion Pump/01)

O p t i o n s J

Verify change to “Concurrent” mode.

3. Use [SELECT] to highlight Rate field.

430-95226-001 (Rev.

f P r o g r a m

alI lower screens from here,the Barcode Reader (BCR) is enabled as indicated by “Wand Active”when attached to the Plum A+. To use the BCR. swipe tip of BCR wandacross entire barcode, from dot to dot, on the Abbott supplied druglist card. The drug name also can be selected using A

Mode]toswitch to “Concurrent”.

Note: Upon entry to this screen, and to

[f3J

The Delivery Mode field is highlighted.

Note: If an entry was previously programmed, RATE is highlighted.If no previous entries, MODE is highlighted.

2 . If field reads “Piggyback”, press A [Change

ceitain alarm conditions, one linemay stop while the other line may continue to deliver.

1. With pumping started on line A, to program a Concurrentdelivery rate, press A

D E L J V E R Y RATE RANGES ).

Note: In concurrent mode with

FOR  

(SEE

SECT ION

FLOW, FOR INF O R M A TION ON C ONC URRENT

DELIVERY MODES.

S EE S ECTION

PIGGYRACK  

LABEL&G TO CONFIRM DRUG

COMPATIBILITY, CONCENTRATION, DELIVERY RATES, AND VOLUME AREALL SUITABLE FOR CONCURRENT AND

II

CAUTION: C ONSULT DRUG

~. 

hrrmin

CONCURRENT DELIVERY - Simultaneous delivery of two fluids atindependent flow rates.

mL

Duration 00: 00

Vi-B1 0.0

mUhr

BRate 0.0

-ode

hcminO:OO  

mL

Duration

VrBl 0.0

mUhr

bxq

Rate 0.0

w*

Mode

(B

““nl:“’ 0.0

PROGRAM

I0.1

O- 0, 

\rIAlAlAj

5.5

m

-PUMPING/

I

Basic Operation

4.13 Concur r ent Deli ver y

l-

Page 37: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 37/85

Page 38: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 38/85

4- 20 Plum A+ Infusion Pump/01)

Back-

430-95226-001 (Rev.

the,cleared quantity, press A [Cancel

In@

2. Use [SELECT] to highlight an option.

3. Press A [Choose] to select the Volumes Infused screen.

This will bring up the Volumes Infused screen whichdefaults to Total Volume.

4. Use [SELECT] to scroll to Line A or B.

5. To reset Line A (or B) Volume Infused, use [SELECT] to

highlight and press [CLEAR].

6. To restore

A,[OptionsNol 

mL

1. To select Options Screen, press

m  mL

Total Vol

mL

tine B 0.0LineA 393

IBl

Volumes Infused

IAl OPTIONS

Lighting/Con~rast

Basic Operation

4.14 Select Opti on - Volumes n fused, Pr essure/Post I n f us i on

Rate, and L igh ting/Contr ast

Pressure/Past Infusion Rate

Page 39: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 39/85

5101)30-95226-001 (Rev.- 21System Operating Manual

in the Technical Service Manual.

KVO or Rate is a biomedical setting

and must be configured as described

@ll continue at the Rate programmed.

Note: Option will be for both line A and line B selections.

15. To change Distal Pressure Limit when highlighted, entervalue between 1.0 and 15.0 psi.

16. Press A [Enter] to keep changes and return to the MainDelivery screen, or A [Cancel/Back] to restore originalvalues Options screen.

Note: Continue Rate default of

whichever is less.

Rate- The pump

mL/hr or the last pro-grammed Rate,

[KVO/Rate]KVO- The pum p will revert to a fixed KVO of 1

[Choose]to continue.

Note: Prior to changing Pressure/Post Infusion Rate, pump mustbe stopped.

13. Use [SELECT] to highlight Continue Rate or DistalPressure Limit field.

14. Continue Rate allows a choice when VTBI is completed.Press A

t;he Main Delivery

screen to return to the Options screen.

11. Use [SELECT] to highlight the Pressure/Post InfusionRate option.

12. Press A

lng fromOptions/l/o/  

softkey.

9. When clearing of quantities is completed, press A [Enter]to return to the Main Delivery screen.

10. Select A

A‘ [Cancel/Back]

valumes, use [SELECT] to highlight Total Vol,

then press [CLEAR].

8. To restore all quantities, press

psi

7 . To reset all

0.0 ressus

psi

Dii

Presrwe Limit 6.0&l  mateanlinw  

IBl

Pressure/Post Infusion Rate

Al OPTIONS

I

Select Option- Volumes Infused, Pressure/Post Infusion Rate, and Lighting/Contrast

Volumes Infused

Lighting/Contrast

Page 40: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 40/85

4- 22 Plum A+ Infusion Pump/01)

[Cancel/Back]to restoreprevious settings and return to the Options screen.

430-95226-001 (Rev.

[Enter]to keep change(s) and return to theMain Delivery screen, or A

highlight’Backlight Intensity orDisplay Contrast.

21. While viewing display for desired effect, press either A[Increase Setting] or A [Decrease Setting] to change level.

22. Press A

Cighting/Contrast Option.

19. Press A [Choose] to continue.

20 . Press [SELECT] key to

from the Main Deliveryscreen to return to the Options screen.

18 . Use [SELEC T] key to highlight

lnfl  Options/Vo/  

m

17. Select A

[A l OPTIONS

Lighting/Contrast

Display Contrast

IB

Volumes Infused

Pressure/Post Infusion Rate

\ I OPTIONS

Basic Operation

Page 41: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 41/85

5/01)

[CanceYBack]to return to the Main Delivery screen.

Example of STOPPED delivery.

2. To discontinue infusion, press [ON/OFF].

3. Clamp secondary tubing if connected.

4. Open door, remove set and close door. Or, press [START]to resume infusion.

4.16 Stop and Start wi th B oth L in es Pumping

1. Press [STOP] at the Main Delivery Screen.

System Operating M anual 430-95226-001 (Rev.

amming screen. Press A

lSTOP], the line is stopped,

the display will show the progr

[BJ is pressed prior toA] or Af 

<A A A A,

1. Press [STOP] at the Main Delivery Screen.

Note:

Iniol  Options 

v;:“’ 0.0

A

,“,a:;, 0.0

8.5

1 B

100

PUMPING1 STOPPED

Stop and Start with only 1 Line Pumping

4.15 Stop and Start wi th on ly 1 L in e Pumping

A)

Page 42: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 42/85

4- 24 Plum A+ Infusion Pump/01)

All]. Or, todiscontinue infusion, press [ON/OFF].

5. Clamp secondary tubing if connected.

6. Open door, remove set and close door.

430-95226-001 (Rev.

[Stati Start A], A [Start B], or A

PRESSING [START].

3 . To resume infusion, press [START].

4 . Select A

Note: This is an example using A [Stop AU].

CAUTION: VERIFY RATE, VTBI, AND DURATION BEFORE

[BJJ is pressed prior to [STOP], the display willshow the programming screen. Press A [Cancel/Back] to return to theMain Delivery screen.

Example of stopped infusion.

20.4

2. Press A [Stop A], A [Stop B], or A [Stop All]

Note: If A [A ] or A

“y 

my;;, 50.0

3.0

rAISX)PPED

25.0

Basic Operation

Page 43: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 43/85

5/01)- 25 430-95226-001 (Rev.

8.

System Operating Manual

Program]when changing ONLY Line Aor

[c/ear  

61 1. Press [STOP] at the Main Delivery Screen.

Clear ALL settings when changing patients or multipleprogramming parameters.

Use A

I STOPPEDPUMPING

[Back Prime] until fluid pum ped fromline A to line B has expelled the air from cassette air trapinto Secondary container.

Note: Fluid flow is at maximum rate possible.

3 . Upon release of A [Back Prime], the cassette test isperformed to confirm readiness for pumping. Pressing[START] will resume programmed delivery.

4.18 Clear in g Program Settings

A

[Back Prime].

Note: Backpriming is available only when all delivery is stopped.Secondary container is necessary to receive the air.

2. Press and hold A

airtrap.

1. To initiate a backprime, press A

BACKPRIMING Removal of air or fluid from the proximal line andcassette

R%zAA A A,primesck A

“Z:” 0.0.9

:;tr 0.0

1 B

5.5

1 STOPPED

Backpriming

4.17 Backprimi ng

A) STOPPED

Page 44: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 44/85

4- 26 Plum A+ Infusion Pump/01)

[SIXRT] key is pressed, the device checksthe delivery program(s) for non-delivery conditions (such as Line BPiggyback without Line A or delayed start) that could permit a periodof non-delivery. If so, the display presents this warning screen.

1. Press A [Yes] to continue w ith delivery. Or, press A [N O]to cancel.

430-95226-001 (Rev.

anyway?

Note: Whenever a valid

continw 

[CanceVBack]to return to the Main Deliveryscreen or program new settings.

4.19 Possibl e Non-Deli ver y Pr ogrammed

Programlwhen changing ONLY Line Aor B.

Example screen of “Cleared Program”.

5 . Press A

hnmin

2. Select Line A (or B if active) from the stopped MainDelivery Screen.

3. Select A [C/ear Program].

Note: If Line A or B is programmed with a Therapy (i.e., DoseCalculation), Rate, VTBI and Duration will not display values. Thevalues to be cleared may be viewed in applicable programming screenunder A [ T h e r a p y ] .

4 . Select A [Yes].

Clear ALL settings when changing patients or multipleprogramming parameters.

Use A [C/ear

0O:OO 

mL

Duration

VTBI 0.0

ml/hr 

&++a

Rate

wind  

03:20 hr:min

Clear Line A Settings?

PROGRAM

mL

Duration

VTBI 22.0

mUhr 

Al PROGRAM

Rate

Basic Operation

Page 45: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 45/85

5/01)- 1 430-95226-001 (Rev.

[A] at the Main Delivery Screen.

The Line A Programm ing screen appears.

2. Press A [Therapy].

Programlwhen changing ON LY Line Aor B.

1. Press A

Allows programming Dose Rates inalternative units of measure. Dose Calculation can be used in SimpleDelivery, Loading Dose, and Multistep.

Clear ALL settings when changing patients or multiple

programming parameters.

Use A [C/ear

0O:OO hr:min

DOSE CALCULATION -

mL/hr

Duration

m  

(mcg/kg/min on A)

Rate

Calc 

Calc isused, the delivery steps are in the same unit of measure, over the same

period of time, from the same container.

5.1 Dose

(mcg/min), from the same container.

Allows a sequential program to deliver up to 10 steps; fluid volumes anddelivery rates may be programmed for each step. The program may beentered based on Rate and Volume or Volume and Time. If Dose

Calc is used, the Loading Dose andMaintenance Dose are in the same unit of measure, over the same periodof time

Multistep-

Allows programming Dose Rates in alternative units of measure. DoseCalculation can be used in Simple Delivery, Loading Dose, and Multistep.

Allows programming of an initial infusion rate for a specific volume,followed automatically by a maintenance rate from the same container

(e.g., a fluid challenge). If Dose

Dose-

Calculation-

Loading

* Advanced user s ar e hose wh o have r eceived additi onal

train in g on programm in g and use of un ctions.

Dose

(For Advanced Users*)herapieS  ECTION 5

Page 46: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 46/85

5- 2 Plum A+ Infusion Pump/01)

‘ield.

Note: The Concentration elements and weight value must beentered before the (SELECT] key allows access to remaining fields.

430-95226-001 (Rev.

entered, use [SELECT] to highlight the next programming

Calc screen appears

to provide for entry of delivery program parameters.

3. Use keypad to enter desired values. Once each value is

[Choose]to continue.

The next page of the Program Dose

Options1 after selecting desired Therapy and Units.

The Container Concentration Selection screen appears

whenever the selected Dose Units are gram-based (e.g.,

grams, mg, mcg, or ng) otherwise this screen is skipped.

Note: Default choice for possible gram-based ContainerConcentration Unit is mg or the previously selected option.

6. Use [SELECT] key to highlight the ContainerConcentration units to use in the delivery.

7. Press A

mcg/kg/min option or the previouslyselected option.

4. Use [SELECT] key to highlight the units to use in the

delivery.

5. Press A [Choose] to continue.

Note: Previously selected Drug Nam es and Delayed Start entrieswill be cleared if entered prior to Therapy selection. To reselect,press A [Program

ap’pears.

Note: Screen defaults to the

Calc screen

‘:

The Program Dose

’ Choosejto continue.

mr)*pmrkl

The Select Therapy screen for Line A appears.

Note: Dose Calculation is the only available option if Line A isinfusing.

3. Press A

o.cra 

-kg

Cone in Container

Weight

Calc

Drug

Al Program Dose

Calc IProaram Dose

Therapies (For Advanced Users*)

Loading Dose

Multistep

I A

Page 47: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 47/85

5/01)

Options], ,press [START] to begin infusion.

System Operating Manual 430-95226-001 (Rev.

VTBI value and any desired A

[ P r o g r am

AND DURATION BEFORE

PRESSING [START].

12 . After entering the desired

2a drug name from the internal list functions exactly asin Basic Operation for the Simple Delivery in Section 4.

CAUTION: VERIFY DOSE, RATE, VTBI,

assignindescribe

VTBI value is entered or changes, Duration isautomatically updated.

Note: The A [ P r o g r a m O p t i o n s] path to setting Delayed Start or

VTBI using the keypad.

Note: As the

11. Enter a value for

VTBI field.

Autom’atic Calculation feature, seeSec t ion 4.7 Ex amp l e s o f A u t oma t i c Ca l c u l a t i o n ,

10. Use [SELECT] to scroll down to the

bs

9. Continue to enter values using the keypad and advancing

to the next field using [SELECT].

Note: As the Dose value is entered, the Rate is automaticallycalculated. To understand the

0watkelKY

rnL

Weight 70.0 kg

250 mgcnc

WMd-CalcProgram Dose

-MN0

A

-

CalcProgram Dose

ml.&,

rA

kc*

7.9

ml_mm

lr#g!nlin  0.0

mL

70.0 kg

mg 25000

wvKIAdi4#CalcProgram Dose

(mca/ka/min on A)

A

Calc  ose

Page 48: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 48/85

5- 4 Plum A+ Infusion Pump/01)

I

430-95226-001 (Rev.

%, 

mql*phnin Dose

The M ain Delivery Screen appears, to display the

information just entered.

Note: The Therapy Type is displayed as well as the Dose Rate andDose Units.

1M

Vi:“’ 0.0

z;;, 0.0

0.1

~ZZL4&C4

7.9

1 STOPPEDPUMPING

Therapies (For Advanced Users*)

4

Page 49: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 49/85

5101)- 5 430-95226-001 (Rev.

programmed prior to Therapy selection. To reselect,press A [Program Options] after selecting desired Therapy and Units.

System Operating Manual

mcg/kg/min option or thepreviously selected option.

Note: Previously selected Drug Nam es and Delayed Start entrieswill be cleared if

Calc screen appears.

Note: The screen defaults to the

[Choosejto continue.

The Program Dose

[B] at the Main Delivery screen.

The Line B Programming screen appears.

2. Press A [Therapy].

The Select Therapy screen for Line B appears.

The screen defaults to Dose Calculation or the

previously selected option.

Note: Dose Calculation is the only available therapy while Line B

is running.

3. Press A

‘1

1. Press A

Programlwhen changing ONLY Line A

or B.

[C/ear 

:,

Clear ALL settings when changing patients or multiple

programming parameters.

Use A

Allows programming Dose Rates in

alternative units of measure. Dose Calculation can be used in Simple

Delivery, Loading Dose, and Multistep.

0O:OO hr:min

DOSE CALCULATION -

0.0

Duration

Inf ol  mL”  

9-o?ehr 

1 B

5.5

1 STOPPEDPUMPING

(mg/hr on B)

A

Calc 

(mg/hr on B)

5.2 Dose

Calc ose

Page 50: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 50/85

5- 6 Plum A+ Infusion Pump/01)

[SELECT] to highlight the next field.

9. To override the Piggyback Delivery Mode default, scroll

down to Mode using [SELECT], then press A [Change

Mode].

Note: Delivery Mode Option is available on Line B, its default isPiggyback, and the Weight field is gone.

430-95226-001 (Rev.

(Cone) and a value

must be entered to continue.

8. Use keypad to enter desired values. Once entered, use

Calc screen appears.

The screen defaults to Concentration

[Choose]to continue.

The next page of the Program Dose

‘_

The Container Concentration Selection screen appears

whenever the selected Dose Units are gram-based (e.g.,

grams, mg, mcg, or ng) otherwise this screen is skipped.

Note: Default choice for possible gram-based ContainerConcentration Unit is mg or the previously selected option.

6. Use [SELECT] key to highlight the ContainerConcentration units to use in the delivery

7. Press A

(mg/hr in this sample) is highlighted,

press A [Choose].

x A A,

4. Continue to scroll using [SELECT] key until desired field is

highlighted.

5. When desired field

Beck

,A

Therapies (For Advanced Users*)

Drug Cow in Container

Page 51: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 51/85

5/01)- 7 430-95226-001 (Rev.

Calc screen entered before it canbe changed.

System Operating Manual

cannot be changed while line is infusing. Theline must be stopped and the Dose

confiied bypressing [START], it

List] with an entry for a selection.

The drug n ame will appear in the upper-right corner.

Note: I f a drug name was previously entered and

P e l i v e r y Ra t e Ra n g e n Sec t i o n 9 . The ratio of concurrentrates has recommendations specified in Sec t i o n 8.3.

14. To update and identify the infusing drug, enter the drug

name by using either the BC R or via A [Program Options]

then A [Drug

[BJ

Upon [STA RT], the Main Delivery Screen shows delivery

status for both A and B lines.

Note: See

OptionsJ

13. Select A

AND DURATION BEFORE

PRESSING [START].

12. To start the delivery, press [START].

Note: To select Delayed Start, press A [Program

/?isplayed.

11. Continue entering the RATE value, if desired.

Note: As Hate is entered, the Duration and Dose are automaticallyadjusted as individual numbers are entered.

Note: If the Duration value is changed, the Rate and Dose areautomatically adjusted.

CAUTION: VERIFY DOSE, RATE, VTBI.

than es to dashes when thecomputed value is larger than what can be

VTBI value entered, theDOSE and DUR ATION fields are automatically updated. To

understand the Automatic Calculation feature, see Sec t i o n 4.7Examp l es o f Au t oma t i c Ca l c u l a t i o n ,

Note: The Duration field temporarily

with-a  

-

10. Continue entering desired values using the keypad. Use

[SELECT] to advance to the next field.

Note: When entering the RATE,

Wmd_I Balc

mL

Program Dose

Infol 

mLI hr

5.0

n&e

1 PUMPING

1.0

mL

PUMPING

loa &J  

wAc6u

I Balc

Mode Concurrent

Program Dose

mLo0  00  one

pdive&nd  I Balc

B)

Program Dose

(m&r onalc  ose

Page 52: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 52/85

5- 8 Plum A+ Infusion Pump/01)

I.

430-95226-001 (Rev.

AND DURATION BEFORE

PRESSING [START].

15. Press [START] to confirm entries and return to the Main

Delivery screen.

s,

CAUTION: VERIFY DOSE, BATE, VTBI,

Dose 

mLInl 0.2.0 Vd

mL1 hr

c%c%z

1.0 Rate 0.5

D% o  

1 EPUMPINGPUMPING

Therapies (For Advanced Users*)

Page 53: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 53/85

5101)

[BJ at the Main Delivery screen.

The selected Line (A or B) Programm ing screen appears.

2. Press A [Therapy].

The Select Therapy screen for Line A (or B) appears.

Note: This example shows line A being programmed.

3. Use [SELECT] to scroll down to

4. Press A [Choose] to continue.

Loading Dose.

System Operating M anual 430-95226-001 (Rev.

Program]when changing ONLY Line A

or B.

1. Select A [A] (or A

(mcg/min). from the same container.

Clear ALL settings when changing patients or mu ltiple

programming parameters.

Use A [C/ear

Calc is used, the

Loading Dose and Maintenance Dose are in the same unit of measure,over the same period of time

,by a maintenance rate fromthe same container (e.g.. a fluid challenge). If Dose

AIIOWS programming of an initial infusion rate for

a specific volume, followed automatically

Wsnd_

Doss Calculation

Multistep

L OADING DOSE -

Al PROGRAM

Dose

Multistep

AJxka

Loading

VJand Al PROGRAM

0O:OO hnmin

y: 0.0

Duration

:;tr 0.0

0.0

Loading Dose

5.3 Loading Dose

/A(

0.0

Page 54: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 54/85

5- 10 Plum A+ Infusion Pump/01)

mL/hr

depending on delivery mode.

8. After the desired value has been entered, u s e [SELECT] to

highlight the next field.

430-95226-001 (Rev.

use [SELECT] to

highlight another option.

6. When desired field is highlighted, press A [Choose].

The next page of the Program Loading Dose screen

appears.

7. Enter a rate using the keypad from 0.1 to 999

dr Choose]to continue,

mL/hr option or the previouslyselected option.

5. Press A

WE00

The Program Loading Dose screen appears.

Note: The screen defaults to the

0o.m2 0.0 0.0

0.0mmL

“pgp 2.k

<AA AA

Chmse Back

1Proaram Loadina Dose

Therapies (For Advanced Users*)

I A

Page 55: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 55/85

5101)- 11 430-95226-001 (Rev.System Operating Manual

Exaniples o f A u t o m a t i cC a l c u l a t i o n ,

10. After the values for the loading dose Step-l are entered,

use [SELECT] to highlight the maintenance dose Step-2.

Note: The loading dose therapy has two steps. The first is theloading dose followed by the maintenance dose. These words willappear on the Main Screen as each step is delivered.

II. Enter the Rate using the keypad .

12. Use [SELECT] to highlight the next programming field.

mL

using the keypad.

Note: After a Rate and VTBI value are entered, the Duration isautomatically computed. Manually changing the Duration field willautomatically recompute the Rate. To understand the AutomaticCalculation feature, see Sec t i o n 4.7

w:w

9. Enter Volume To B e Infused (VTBI) from 0.1 to 9999

C&X2

1

“h%LiAw v-mlate

1oadzdse  l Program

w:oo0.0 M):ffio 55.0

tuxlinlmDuratimTBI ate 

WUdAC4iV8I Program Loading Dose

lYJ:W

A

w:c62 0.0 0.0

&O “;FmT

mUhrvrslate

A&wsnd  ( Program Loading Dose

Loading Dose

A

Page 56: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 56/85

5- 12 Plum A+ Infusion Pump/01)

Callback.option.

Note: Nurse Callback enables an alarm when a change in thedelivery on a line occurs such as completing Step- 1 and startingStep-2. Or on line B, a piggyback delivery is completed and line A isstarting its delivery.

15. Enter a delay time (less than or equal to 24 hours) using

the keypad.

Note: Not available if a delivery is taking place on the line.

16. After selecting a desired value, use [SELECT] to highlight

the callback feature.

430-95226-001 (Rev.

VTBI values are selected, theDuration is automatically updated.

14. Press A [Program Options] to enable setting a Delayed

Start or to enable the Nurse

hcmin

13. Enter the desired VTBI using the keypad.

Note: After the desired Rate and

OOa

lO 0 hnmin

Delay Start for

Optioiow

Delay Start for

1rogramA

WandMiveI Program Loading Dose

Therapies (For Advanced Users*)

A

Page 57: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 57/85

5/01)

AND DURATION BEFOREPRESSING [START].

19. Press [START] to begin delivery.

The M ain Delivery Screen appears.

Note: The words “DELAYED” and “Loading Dose” appear in theupper-left comer. These indicate the options wh ich have beenselected.

After the programmed delay period has expired, the

Plum A+ starts “Pumping”. This is indicated where

“Delayed” previously was displayed.

System Operating Manual 430-95226-001 (Rev.

VThI,  ATE, 

“$ ” 0.0

17. Select using A [Yes/No].

Callback changes from NO to YES.

Note: Nurse Callback cannot be changed while a delivery is takingplace on the line.

18. Press A [Enter] to accept the changes.

Note: A [ C a n c e l / B a c k ] returns the fields to their original valueand displays the previous screen.

CAUTION: VERIFY DOSE,

mL/hr

0.5

Iwe 0.0

-Do-

500

IBPUMPING STOPPEDA

“$” 0.0

“’

0.0

m!% 

Mhprrr

500

1 BSTOPPEDDELAYED

WM(I*

A

1 Program Loading Dose

m

A

OD:O5 hr:min

Callback

Start for

wllnd-

Delay

) Program Options

Loadina Dose

A

Page 58: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 58/85

5- 14 Plum A+ Infusion Pump101)

VTBI of last line is changed.

430-95226-001 (Rev.

“KVO”.If Continue Rate option w as selected, pumping c o n t i n u e s at originalrate.

21. Press [SILENCE] to stop alarm sound.

Note: Warning message continues to flash and audio alarm willreturn after a two minute period of silence, unless delivery isstopped or

mL/hr or lessdepending on delivery rate and “Pumping” mode changes to

Therapies (For Advanced U sers*)

After the first step has completed, the Main Delivery

Screen changes to the second step.

Note: A flashing “Callback Line A” message and the audio alarmsound indicates activation of that option.

20. The user should acknowledge the Callback message and

stop the audio alarm by pressing [SILENCE].

Upon completion of delivery, the screen shows a

flashing “Line A VTBI comp lete” message and the

audible alarm sounds.

Note: The KVO rate will change to the rate of 1.0

Page 59: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 59/85

5/01)

[Chooselto continue.

System Operating Manual 430-95226-001 (Rev.

cianging ONLY Line A

or B.

1. Select A [A] at the Main Delivery screen.

The Line A Programm ing screen appears.

2. Press A [Therapy].

The Select Therapy screen appears.

Use [SELECT] to scroll down to Multistep.

Press A

Program]when C/ear  

Calc is used, the delivery steps’ are in the same unit ofmeasure, over the same period of time, from the same container.

Clear ALL settings when changing patients or multiple

programming parameters.

Use A

+ndVolume or Volume and

Time. If Dose

Rrogrammed for each step.The program may be entered based on Rate

MULTISTEP- Allows a sequential program to deliver up to 10 steps:fluid volumes and delivery rates may be

I Dose Calculation

Loading Dose

W M d A C % V aAl PROGRAM

whnd&WO

Loading DoseMultistep

Al PROGRAM

0O:OO hr:min

mL

Duration

VTBI 0.0

mL/hr 

W&_

Rate

Al PROGRAM

Multistep Programming

5.4 Mul tistep Programming

Page 60: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 60/85

5/01) Plum A+ Infusion Pump

diluent volume before any other entries.

430-95226-001 (Rev.

(Cone). Enter

value using keypad.

Note: The user must enter the concentration drug amount and

[Choose]to continue, or use [SELECT] to

highlight another option.

8. When desired field is highlighted, press A [Choose].

9. The default field is the concentration field

[Choose]to go to the second page of the

Program Multistep screen.

The Container Concentration Selection screen appears

whenever the selected Dose Units are gram-based (e.g.,

grams, mg, mcg, or ng) otherwise this screen is skipped.

Note: Default choice for possible gram-based ContainerConcentration Unit is mg or the previously selected option.

7. Press A

mcg/hr.

6. Press A

mL/hr option or the previouslyselected option.

5. Use [SELECT] to highlight the desired setting. This

example is using

::: 0 .0 0 .0 0. 00.0 0.0

The Program M ultistep screen appears.

Note: Screen defaults to the

ztIEoo:W.oW

OOZW 0.0z!ir%? z

WMd_I Program Multistep

Ax.AA

Cone in Container

Back

Drug  

I Program Multistep

I Program M ultistep

A

Therapies (For Advanced Users*)

Program M ultistep

A

Page 61: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 61/85

5/01)- 17 430-95226-001 (Rev.

VTBI) areautomatically re-computed, depending on the prog ramming order.

System Operating Manual

VTBI is entered, the Duration value isautomatically computed.

15. Use [SELECT] to highlight the Rate field.

Note: If Rate is changed, Dose and Duration (or

VTBI value is automatically computed w hen theDuration value is entered. If

VTBI field.

Note: The

14. Use [SELECT] to highlight the Duration or

C a l c u l a t f o n ,

f A u t o m a t i cxa~npZes  

(Diluent-Amount).

11. Enter value using keypad.

12.

13.

Use [SELECT] to highlight the next field.

Enter Dose using keypad.

Note: The Rate value is automatically computed. To understand theAutomatic Calculation feature, see Section 4.7

mL

10 . Use [SELECT] to highlight the next programming field

500gaw 50.0

WMd*ruvsI Program Multistep

rnL

A

so0  geix 50.0

v&a MMI Program Multistep

oo:w 0.0 0.0

A

0.030I?303 0.0 0.0.lxa

00 0.0 0.0 0.0 0.0

3i

00z 00: 00:

mLRnLcmin  gmr

rnL g  ow 50.0

WMd*aMI

) Program Multistep

Multistep Programming

A

Page 62: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 62/85

Page 63: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 63/85

5/01)- 19 430-95226-001 (Rev.

usi.ng the keypad.

Note: The Concentration drug amount or Diluent volume cannotbe changed while a delivery is in progress.

If Dose value is changed, the Duration and Rate will

automatically change.

Note: Only the current delivering step and future steps can bechanged.

System Operating Manual

**” in a Step number field indicates the step is deliveringor “d” indicates each completed step.

26. Enter a value

i-A%

Note: The

Pumeing on Step-3.

25. To change a program parameter while pumping, press A

mL/hr, the Dose rate and units also willbe displayed.

Example of M ultistep Screen

‘_

Note: In units other than

DURATION BEFORE

PRESSING [START].

24. Press [START] to begin the delivery. The Main Delivery

Screen displays Step-l in the upper-left corner.

AND  ERIFY DOSE, RATE, VTBI,

mL

CAUTION:

5X  gone 50.0

WMd*mveI Program Multistep

1

A

ME2  I Program

-1 Dose50

v”,:“’ 0.0

0.0

37.5

mzehr 

8-3

2.5

1 BSTOPPEDPUMPING

mr Dose

A

m,

“2:” 0.0

,“,a:;, 0.0

0.5

%P l

2.0

1 BSTOPPEDPUMPING

Multistep Programming

A

Page 64: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 64/85

5- 20 Plum A+ Infusion Pump/01)

VTBIvalue.

430-95226-001 (Rev.

‘.

Note: The Values are now updated on the Main Delivery Screen.

The pump reverts to KVO when last Step completes.

The last step can be restarted by entering anew

Deljvery Screen.

DOSE, RATE, VTBI, AND DURATION BEFORE

PRESSING [START].

27. Press [START] to begin infusion with the changed

program and return to the Main

Therapies (For Advanced Users*)

CAUTION: VERIFY

Page 65: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 65/85

5/01)- 1 430-95226-001 (Rev.

%y another message upon the nextvalid input key press.

messa e will be shown to explain theproblem. All such messages will be cleared or replaced

warning  

AC p ow e r , o n l y a b o u t 1 5m i n u t es o f b a t t e r y i f e r em a i n s .

Note: Other error and warning messages will occur when appropriate for the currentselection context. For instance, when a change to a required parameter would result in

clearing some delivery settings, the warning message will appear when that field isselected. When an input error has just been made, the three quick beeps of an Invalid KeyWarning will sound and an appropriate

m A j w .

T h e b a t t e r y v o l t a g e s ess h a nt h e d e p l et ed h r esh o l d a n dc h a r g e ev el s h i g h er h a n h el o w c h a r g e h r esh o l d .

Wa r n i n g : C h a r g er

Se r v i c e

W a r n i n g : L o w

B a t t e r y

T h e c h a r g i n g ci r c u i t r y s n o t

b eh a v i n g a s ex p ec t ed ( ch a r g ev ol t a g e s n o t ch a n g i n g ) a n d

c a n r e d u c e b a t t e r y i f e .

P um p n o t p l u g g ed n o AC

powe r .

S i l e n c e k e y s p r e ss ed .

P l u g n t o

s e c t f o n 4 . 4R e a rC a seCo n t r o l s ) , t h e n r e t u r n t o

i s en a b l e d . en a b l e d p o si t i o n .

B a t t er y S er v i c eWa r n f n g

T h e b a t t e r y c a p a c i t y s g r ea t e r R e p l a c e b a t t e r y .

t h a n 2 00

( e x c e p t STOP )w h e n L o c k o u t sw f t c h  

P l a c e L o c k o u t sw i t c h f n d f s a b l e d p o s f t f o n [ s e e

I t u r n o ff

L o c k o u t E n a b l ed

w h i l e a d e l i v er y s n p r o g r e ss.

P r e s s i n g a n y k e y w h f l e d e l f v e r f n g

M essages

SECTION 6 Alarms and Troubleshooting

6.1 Warning

Page 66: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 66/85

6- 2 Plum A+ Infusion Pump/01)

p u m p i n g o p e r a t i n g .

430-95226-001 (Rev.

l i n e A

( o r B ) c o n t a i n e r d i s co n n e c t ed o r

o bst r u c t ed d u r i n g n o n d el i v er y( s t a r t u p ) .

E2 3 3 ,E 2 3 4

D i s t a l A i r Wh e n a i r s d e t ec t ed a t h ed i st a l sen s o r a n d ex c eed s h el i m i t .

E250 D o o r op en w h i l e P um p d o o r o p en ed w h i l e

Prox O c c l A tE l 85 s t a r t u p

P r o x i m a l i n e o bst r u c t ed ,

N sf t e

c l o t t ed o r p o si t i o n a l , p r e ssu r el i m i t set o o o w .

D i st a l O c c l u si o n

E l87 N eg D i st a l P u m p oo h i g h a bove p a t i en t or

O c c l u s i o n d e f ec t i v e se t .

El 83,

E1 8 1 ,E l86

D i s t a l l i n e k i n k ed .

r e p r i m e c a s s e t t eu s i n g st a n d a r d ec h n i q u e .

C l o se d o o r w i t h c a sset t ei n s e r t e d .

A t t e m p t i n g o u s e u n r e c og n i z ed

c a s s e t t eUn r e c o g n i z e d

casse t t e

N o A c t i on A l a r m P u m p p r og r a mm i n g

i n c omp l e t e .

E l 02 I n f u s er d l e 2

M f n u t es

P um p s ON w i t h o u t o p er a t o r

i n p u t .

El 03 M em o r y F a i l u r e T h e p u m p d i d n o t a c cep t h ep r ev i o u s p r o gr a mm ed d a t a .

P r o x Ai r ,

B a c k p r i m e

W h en h e o t a l p r o x i m a l a i r

d e t ec t e d e x c eed s h e i m i t .E 2 3 0

E l 8 0,

doo r , h e n m m ed i a t e l y o p enc l a m p . Co n t i n u e p r og r a m m i n g .Re p l a c e set f p r o b l em p er si s t s.

Note: If clamps are not opened prior to

completion of cassette test, peak proximalocclusion may recur.

R em o v e a n d

bolus

following a distal occlusion.

L owe r p u m p o r r ep l a c e s et .

E x am i n e i n e A ( or B ) o r k i n k s o r

c l o sed c l a m p s . O p e n a n d c l o s e

l i n e A o r B . C or r ec t

a s n e cessa r y . B a c k p r i m e n t ol i n e B u b i n g o r sy r i n g e

C h e c k d i st a l u b i n g , c h ec k V

si t e, r eset p r e ssu r e im i t .Note: See Section 8.8 for detailed

instructions for avoiding a

[ S TAw r l o r c on t i n u ep r o g r a mm i n g .

B eg i n p r o g r a m m i n g o r turn p u m pO F F .

Ree n t e r a l l p r o gr a m m ed d a t a .

C h e c k o r cl a m p s , o r em p t yc on t a i n er s o n

r e cogn f zed

casse t t e .

P r e s sl 01

T u r n o f f o r n s er t

[START] to resume infusion.

Note: Alarm messages begin with an alphanumeric code for tracking purposes only. Iftroubleshooting does not correct the problem, record the code number and contacttechnical support.

1.

2.

3.

4.

5.

GENERAL ALARMS

E l 0 0

confinning back switch is in the down position.

Press [SILENCE] key.

Identify/Observe Alarm condition.

Correct Alarm condition (see chart below).

Press

Alarms and Troubleshooting

6.2 Response o Alarms

Disable Lockout by

Page 67: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 67/85

5/01)

s y r i n g e

System Operating Manual 430-95226-001 (Rev.

V T B I .

E x a m i n e i n e A o r k i n k s o r

c l o sed c l a m p s .

C h e ck o r cl a m p s , o r em p t y

c o n t a i n er . C o r r ec t a s n e c essa r y .B a c k p r i m e n t o line B t u b i n g or

r de tec ted

B a c k p r i m e e x c e e d s h e im i t o r h e i n e .

P r o g r a m n ew

r A , Wh en p r o xi m a l a frox Af

l i n e Ar o x i m a l i n e o bst r u c t ed ,

SteD.

o r f o r

a n y st ep n M u l t i s t ep ” h e r a p yp r o v i d ed h e st e p s n o t h e a st

E l 9 1

E 2 3 2

Ca l l b a c k L i n e A A l a r m p r o gr a m m ed b y use r .

C h a n g e n d el i v er y o r L o a d i n g

D o se o M a i n t e n a n c e S t ep

E l 6 1

E l 8 4,

E l 9 0,

P r e s s f n g o f a n y k e y e x c e p t l S TOP ] U n l o ck o ck o u t sw i t c h .d u r f n g d e l f v e r y w h f l e a n y a l a r m f s

a c t i v e a n d o c k o u t s en a b l e d .

L INE A ALARMS

.

l o c k o u t s en a b l e d .

E 2 5 4 L o c k o u t E n a b l ed

& l f v er y ,o r o p en f n g t h e d oo r ,w h f l e  

Resnonse to Alarms

F a i l u r e d i s t a l oc cl u si o n o r a i r n

c a s set t e d e t ec t ed a t st a r t u p .

set , c l o se d o o r . B a c k p r i m e n t o

l i n e B u b i n g o r sy r i n g e fa p p r o p f fa t e. f a l a r m r ec u r s o n

r et e st , r ep l a c e set . f a l a r mr e p ea t s , r e p l a c e p u m p .

E 2 5 2 D e p l e t ed B a t t e r y T o o m u c h i m e o n b a t t er y P l u g n t o AC p ow er .

p ow e r .

E 2 5 3 L o ck o u t V i o l a t i o n Th e u s e o f he STOP ) k e y du r ng U n l o c k o ck o u t sw i t c h .

Page 68: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 68/85

5/01) Plum A+ Infusion Pump

1 s y r i n g e .

ma l f u n c t i o n r ep ea t s , r e p l a c e

430-95226-001 (Rev.

1 fneB

D o s e o M a i n t e n a n c e S t ep , orfor a n y st ep n M u l t i st ep ”

c on t a i n er d i sc on n ec t ed o r

t h e r a p y p r o v i d ed h e st ep isn o t h e a st st e p .

P re s s S I LENCE ) .

o b s t r u c t e d . I

P r o gr a m n ew V T B I .

Note: Occlusion caused by empty

container, backprime into line B tubing

E x a m i n e line B f o r kinks o r

or syringe.

c l o sed c l a m p s .

W h en p r o x i m a l a i r d e tec t ede x c e e d s h e im i t o r h e i n e .

C h e c k o r cl a m p s , o r em p t yc o n t a i n er . C o r r ec t a s n e c essa r y .B a c k p r i m e n t o i n e B u b i n g o r

V T B I c om p l et e o n i n e B .

Cha n g e n d e l i v e r y o r

p i g g y b a c k d e l i v er y , L o a d i n g

P r o x i m a l i n e o bst r u c t ed ,

aMALFUNCTIONS

A l a r m p r o g r a m m ed b y u ser .

3L

h -ox A i r B ,B a c k p r i m e

E l 89

E23 1

& c l B

E l 8 8,

V T B I

c om p l e t e

El 82 , P r ox

Alarms and Troubleshooting

L INE B ALARMS

E l 60 L i n e B

Page 69: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 69/85

5/01)- 1 430-95226-001 (Rev.

the cassette door using cotton-tipped swabs saturatedwith cleaning solution. The cassette door may be unlatched from the door handle to facilitate cleaning.

To unlatch the cassette door from its handle, tilt the device back, open the cassette door, then push thedoor release tab to open the door fully.

all of the elements behind

S p o r i c i d f n@ S p or i c i d i n n t er n a t i o n a l P er m a n u f a c t u r e r ’sr e c o mm e n d a t i o n

H o u s eh o l d b l e a c h V a r i o u s P er h o sp i t a l p r o ced u r es: d on o t ex c eed o n e p a r t b l e a c hi n en p a r t s w a t er

O n a routine basis, clean

Co v e r a g e TMHBV S t er f s C o r p o r a t i o n , A d i v i s i o n P er m a n u f a c t u r e r ’so f C a l g o n V est a l L a b o r a t o r i e s r e c o mm e n d a t i o n

TM Dfversey Corporation P er m a n u f a c t u r e r ’s

r e c o mm e n d a t i o n

Ma n u - K l e n z@ C a l g o n V est a l L a b or a t o r i es Per manufacturer’srecommendation

Formula C

@ se C a l g on V est a l L a b o r a t o r i es P er m a n u f a c t u r e r ’sr e c o mm e n d a t i o n

Ed f s o n i t e@ S . M . E d i s o n C o . P er m a n u f a c t u r e r ’ sr e c o mm e n d a t i o n

Ve s phene I

in the following list or mild, nonabrasive soapy water.

S u p e r

Turn the Plum A+ off with the [ON/OFF] switch, then disconnect the Plum A+ from mains power.

The exposed surfaces of the Plum A+ may be cleaned with a lint-free cloth dampened by one of therecommended cleaning solutions

‘.

The cleaning, maintenance, and storage of the Plum A+ are described in this section.

CLEANIN G AND SANITIZING

For proper maintenance of the Plum A+, observe the following cleaning and sanitizing guidelines.

Establish a routine weekly schedule for cleaning the Plum A+. To clean the Plum A+, proceed as follows:

SECTION 7 Maintenance and Storage

7.1 Cleani ng, M ain tenan ce, and Stor age

Page 70: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 70/85

7- 2 Plum A+ Infusion Pump/01)

Plum A+ connected to mains power

l Switch the Plum A+ off using the [ON/OFF] key

SERVICE

All servicing or adjustments to the Plum A+ should be referred to qualified technical personnel. A technicalservice manual may be ordered from the local Abbott Laboratories sales office.

430-95226-001 (Rev.

maxImum battery charge and to prolong battery life, connect the Plum A+ to mains powerwhenever possible.

STORAGE

To prolong the life of the Plum A+, observe the following storage precautions:

l Store the Plum A+ away from excessive heat, cold, and humidity

l Store the

ml/hr or 500 ml total volume delivered, whicheveroccurs first.

Note: For optimum battery life, the Plum A+ should be operated on battery power for six continuoushours at least once every six months and then charged for a minimum of six hours.

The battery charges whenever the Plum A+ is connected to mains power. If the Plum A+ is switched off,recharge takes approximately six hours. Recharge takes longer if the Plum A+ is turned on.

As a general rule, the more often the battery is partially discharged and recharged, the sooner it will need tobe replaced. Consult a qualified biomedical technician for battery replacement if necessary.

To maintain

fully charged,new battery provides six hours of operation at 125

b%AINS POWER

IMMEDIATELY.

The Plum A+ is battery powered for emergency backup and temporary portable operation. A

PLUM A+ TORE  OUNDS. CONNECT

USE OF A

PROPERLY MAINTAINED AND CHARGED BATTERY HELPS CONFIRM PROPER OPERATION.

CAUTION: IF THE LOW-BATTERY ALARM

THE BATTERY REMOVED.IT a  LUM A+ ON PATIENTS

.$4ANUFAcTuRBR FOR SPECIFIC

INFORMATION.

BATTERY MAINTENANCE

CAUTION: Do NOT OPERATE THE

CLEANING SOLUTIONS VARY: CONSULT THETI3B

DISINFECTING PROPERTIES OF

CLBANING SOLUTIONS SHOULD ONLY BE USED As DIRECTED.

(M‘O), OR RADIATION.

CAUTION: To AVOID PUMP DAMAGE,

CLEANANY PART OF THE PLUM A+.

CAUTION: Do NOT STERILIZE BY HEAT, STEAM, ETHYLENE OXIDE

CL&S, OR NE EDL ES TOS FINGER N A IL S, PAPERBJECTS SUCH

BEti? AMMONIUM CHLORIDE.

CAUTION: N EVER USE SHARP

DIMETWL 

ABBOTT LABORATORIES MAY RESULT IN PRODUCT DAMAGE. Do NOT USE COMPOUNDS

CONTAINING COMBINATIONS OF ISOPROPYL ALCOHOL AND

SLOWLY DEGRADE COM PONENTS MADE

FROM SOME PLASTIC MATERIALS. USING ABRASIVE CLEANERS OR CLEANING SOLUTIONS NOT

RECOMMENDED BY

MTIZING SOLUTIONS MAY

PLUM A+ IN ANYFLUIDS OR CLEANING SOLUTIONS.

CAUTION: CERTAIN CLEANING AND

THE  ECRANICXL OR ELECTRONIC DAMAGE, DO NOT IMMERSE

Maintenance and Storage

CAUTION: To AVOID

Page 71: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 71/85

5101)- 1 430-95226-001 (Rev.

allsuitable for secondary, concurrent and piggyback delivery modes.

co&ii drug compatibility, concentration, delivery rates, and volumes are

a-way radios in close proximity ofthis device may affect its operation.

8.2 H ealth care Prof ession als and Patient Related

In vitro studies have suggested that packed red blood cells with unusually high hematocrit be diluted withblood-compatible fluids, such as 0.9% sodium chloride injection, to decrease hemolysis and increase flowrate.

Setting the primary rate greater than the secondary rate will result in a more rapid infusion of any residualsecondary drug remaining in the line and the cassette.

Consult drug labeling to

(EMI) conditions. In the event of extreme levels of Interference, such as encountered next to anelectrosurgical generator, it is possible that the normal operation of a sensor or microcomputer might bedisrupted. Even in this event, the outcome would likely be a false alarm or detected system malfunction andwould not result in a hazard to patient or operator.

Use of radio frequency emitting devices such as cellular telephones and

Plum A +, observe the following precautions and hazards.

8.1 Ar tif acts

Nonhazardous, low-level electrical potentials are commonly observed when fluids are administered usinginfusion devices. These potentials are well within accepted safety standards, but may create artifacts onvoltage-sensing equipment such as ECG , EMG. and EEG m achines. These artifacts vary at a rate that is

associated with the infusion rate. If the monitoring machine is not operating correctly or has loose ordefective connections to its sensing electrodes, these artifacts may be accentuated so as to simulate actualphysiological signals. To determine if the abnormality in the monitoring equipment is caused by theinfusion device instead of some other source in the environment, set the Infusion device so that it istemporarily not delivering fluid. Disappearance of the abnormality indicates that it was probably caused bythe electronic noise generated by the infusion device. Proper setup and m aintenance of the m onitoringequipment should eliminate the artifact. Refer to the appropriate monitoring equipment systemdocumentation for setup and maintenance instructions.

The Plum A+ system is designed to operate normally in the presence of m ost encountered electromagneticinterference

SECTION 8 Precautions

The Plum A+ has been designed and manufactured to be safe, reliable, and easy to use. This section details

precautions and possible hazards.

For safe operation of the

Page 72: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 72/85

8- 2 Plum A+ Infusion Pump/01)

mL/hr or greater, the

other infusion can be programmed at any desired rate.

Note:The total of the primary rate plus the secondary rate cannot exceed 500 mL/hr.

430-95226-001 (Rev.

mL/hr.

l If the critical drug (with half-life less than 6 minutes) is to be infused at 5.1

ml/hr should not be accompanied by an infusion programmed any faster than 35

mL/hr the

other infusion should be no faster than ten times the critical drug’s rate. Dopamine, for example,delivered at 3.5

mL/hr.

l If the critical drug (with half-life less than 6 minutes) is to be infused at 2.0 to 5.0

mL/br should not be accompanied by an infusion programmed any faster than7.5

mL/hr, the

other infusion should be no faster than 5 times the critical drug’s rate. Dopamine, for example,delivered at 1.5

infused at less than 2.0

Critical D r u g s, h i s sec t i o n ) using the Plum A+ in theConcurrent mode, the following delivery rate guidelines should be observed:

l If the critical drug (with half-life less than 6 minutes) is to be

RATE

When delivering short half-life critical drugs (See

DELIVERY 

outgas. In these cases, an air eliminating filter may be used.

Repeated opening and closing of the door may defeat the proximal air-in-line alarm and’may cause a distalair-in-line alarm, requiring repriming.

The screen displays the VTBI (volume to be infused) in integers when value is above 99.9. Any fraction of amilliliter delivered is not displayed, but is retained in memory.

8.3 Concurrent F low

mL of unexpectedfluid delivery.

A small amount of fluid is expelled from the set (less than 0.05 ml) each time the door is opened or closedwith a set installed. If potent drugs are being used, take appropriate action to guard againstovermedication of the patient.

Before disconnecting a syringe from the cassette, pull up the plunger slightly to avoid spilling the fluid. Forrigid containers (for example, vials), close the upper slide clamp, open the cassette door, then remove andinvert the cassette (ports down).

Air bubbles may form distal to the cassette as result of normal out-gassing of dissolved air in the fluid.This may occur if chilled solution is in use, if the pump is mounted significantly above the patient, or whenusing certain fluids known to routinely

sebondary container from thesecondary port to prevent mixing of primary and secondary fluids.

Although unlikely, failure of certain robust mechanical components such as the anti-free flow mechanismor valve control springs could cause fluid delivery limited to the contents of the fluid container. Single faultfailure of certain electronic/motor control components would result in no more than 5

Precautions

Connecting more than one pump Into one patient line may significantly affect the infusion rate of at leastone of the pumps.

Arrange tubing, cords, and cables to minimize the risk of patient strangulation or entanglement.

Before opening the door, close clamp on the primary line or remove the

Page 73: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 73/85

5/01)Rev.  - 3 430-95226-001

“epidural” stickers indicating ongoing epidural administration are recommended.

Administration of drugs via the epidural route should be limited to personnel familiar with associatedtechniques and patient managem ent problems. Proper epidural placement of the catheter is essential since

System Operating Manual

mL/hr or less.

Note: This list of critical drugs is not intended to be all-inclusive of critical drugs or drugs with a shorthalf-life.

The clinician should become familiar with the pharmacodynamics of any critical drug beforeadministration.

This information is presented to inform clinicians of a rare situation that could be misinterpreted if they areunfamiliar with this phenom enon.

8.4 Epidur al A dmin istrat ion

Recommended use of the epidural route is to provide anesthesia or analgesia for periods up to 96 hours.

This device can be used to administer only those anesthetics/analgesics approved for epiduraladministration (as indicated or allowed by the drugs’ FDA approved labeling). Epidural administration of

drugs other than those Indicated for epidural use could result in serious Injury to the patient.

For epidural administration, the use of Abbott Laboratories catheter, List No. 1193 or List No. 6947, pumpsets without Y-sites, and

IV) Include:

l Dobutamine l Nitroglycerin

l Dopamine l Nitroprusside

l Epinephrine l Norepinephrine

l Esmolol l oxytocin

l Isoproterenol l Procainamide

l Lidocaine l Trimethaphan

For these drugs, the concurrent flow guidelines should be followed when the infusion rate of the drug will be5

CRITICAL D RUGS

Examples of drugs with a short half-life (approximately 6 minutes or ess when given

‘.

5 Times the Critical Drug Rate10 Times the Critical Drug Rate

Any Desired Ratio

I , 

Maximum Rate of Accompanying Infusion

mL/hr

5.1 or Greater

- 5.0mL/hr

2.0- 1.9

Half-life (less than 6 minutes)Critical Drug Infusion Rate

0.5

ha l f - 1 i J e c r i t i c a l d r u g s n C on c u r r en t m o d e. Individualpatient responses may vary requiring adjustment of delivery rates.

Delivery Rate Guidelines

Short

NoteThese guidelines apply only when Infusing short

Epidural Administration

Page 74: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 74/85

8- 4 Plum A+ Infusion Pump/01)

ml/hr.

See S ec t i o n 1 0 for information on sets and accessories.

430-95226-001 (Rev.

19-gauge or larger needle or catheter at the venipuncture site for viscous fluids if operating at ratesgreater than 500

mL/hr or less) the use of thick-walled microbore Plum sets isrecommended. This will reduce the amount of the fluid bolus that may be delivered when a distal lineocclusion is released.

Use List 11986-48, syringe adapter when using syringes 10 cc or smaller on the secondary line. (Syringesmust be larger than 3 cc.)

Use a

nonblood filters are not for use in the administration of blood, bloodproducts, emulsions, suspensions, or any medications not totally soluble in the solution beingadministered. These medications may be administered through the lower Y-injection site, below the filter.

When infusing at low delivery rates (5.0

LifeCare IV Infusion sets with integral

LifeCare Plum Sets can be used with the Plum A+. See individual set instructions foradditional information.

Administration sets should be changed per CDC guidelines or healthcare provider policy. Discard after use.

HELPS CONFIRM PROPER OPERATION.

The battery may not be fully charged upon receipt. Connect the Plum A+ to AC power for at least six hours.

Use AC power whenever possible. Connect the Plum A+ to AC power during storage to ensure a fullycharged battery for emergencies. If quality earth grounding source is in doubt, use battery power.

If the low-battery alarm sounds, connect the Plum A+ to AC power immediately.

8.6 Sets and Accessor ies

Only compatible

WITEI THE BATTERY REMOVED. USE OF A

PROPERLY MAINTAINED AND CHARGED BATTERY

PLUM A+ ON PATIENTS

fully-equipped and staffed environment for at least 24 hours following completion of drug administration by theepidural route. DELAYED RESPIRATORY DEPRESSION FOLLOWING CONTINUOUS EPIDURALADMINISTRATION OF PRESERVATIVE-FREE MORPHINE SULFATE HAS BEEN REPORTED.

The epidural space has 58 openings through which fluid can exit. Pressure buildup during administrationis transient. However, if a large volume of fluid is administered over a short time period, the pressure willtake longer to return to normal. If overdelivery occurs during administration, observe the patient closely forsigns of spinal cord compression (disorientation, headache, transient neuralgias) and drug overdose.

8.5 Battery Operation

CAUTION: Do NOT OPERATE THE

Orrimetry) is recommended for continuous monitoring of thepatient during epidural administration. Patients must be observed frequently for side effects in a

Precautions

catheter migration could result in intravascular or intrathecal administration. Facilities practicing epiduraladministration must be equipped with resuscitative equipment, oxygen, naloxone, and other resuscitativedrugs. Adequate monitoring equipment (e.g.,

Page 75: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 75/85

5/01)- 5 430-95226-001 (Rev,

Backpriming

8.7 Backpriming

Backpriming is not recomm ended for reconstituting secondary containers containing dry powders.

To avoid pressurization when backpriming Into a syringe or a vial, the user m ust confirm that these

containers have sufficient empty space to accept the backprimed fluid.

8.8 General

Possible explosion hazard exists if used in the presence of flammable anesthetics.

Do not place Plum A+ in service if it fails the self-test.

Do not operate the Plum A+ with the case opened.

Keep the cassette door securely closed while the Plum A+ is not in use, to avoid cassette door damage.

Use the following procedure to avoid the administration of a bolus following a distal occlusion:

If a secondary container is in use. clamp proximal tubing before opening cassette door

Open cassette door and remove the cassette

Open the flow regulator briefly to dissipate the pressure and then close it

Eliminate the source of occlusion

Reinsert the cassette and close the cassette door

Open all clamps and resume infusion

System Operating Manual

Page 76: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 76/85

5101)- 1 430-95226-001 (Rev.

mL increments)L (in 1

mL. increments)

100 to 9999

mL (in 0.1

VD C Max

Current- 0.25 Amps Max

Contact Rating- 3 Watts Max

VTBI RANGE

VTBI Range: 0.1 to 99.9

Norrnally-

Open (NO) to Normally Closed (NC) system.

Voltage-30

(NO)Note: Contact the Technical Services Center to make an

internal adjustment to change the device from

12/78 standard.

NURSE-CALL alarm is factory set for Normally- Open

AAMI/ANSI SCL (ungrounded)

mIJhr on one line.

Risk current limits meet

mL/hr or greater on one line.

The battery charges whenever the Plum A+ is connected to

AC power. The recharge time is approximately six hours

with the device operating at 125

mL at 126

mIJhr or less, or deliver

500

F2,25OV-, 0.5 A. (internal)

Hospital-grade AC cord. 10 ft long, with transparent plug

and retainer plate.One sealed, lead-acid, rechargeable 6 V battery, internal to

device.

With a new fully charged battery, the infuser shall operate

for a minimum of six hours at 125

Fl,

6”, excluding pole clampprotrusion and power cord storage.

Approximately 9.5 Ibs. with battery.

High-impact plastic

120 V-, 50-60 Hz, 35 VA.

8” X 8” X

SECTION 9 Specifications

PHYSICALDimensions:

Weight:

Casing:

ELECTRICAL

Power R equirements:

Fuses:

Power Cord:

Battery:

Battery Life:

Recharge:

Electrical Leakage:

NURSE-CALL System:

Circuitry Ratings:

Approximately

Page 77: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 77/85

9- 2 Plum A+ Infusion pump/01)

mL

430-95226-001 (Rev.

mL, Total 1.0lumSet (Proximal): Bolus at 0.5

mLnL out of 2.5mL or larger

Cumulative 0.26PlumSet (Distal): Bolus 0.1

Muir or the last primary delivery rate, whichever is less

AIR-IN-LINE ALARM

mLfhr cumulative (A+B) maximum

KVO: 1.0

PlumSet: 500

Muir minimum for each line

increraents)

Concurrent Delivery: 0.5

mL  L/hr (in 1999

mL increments)

100

mIJhr (in 0.199.9

- 90% (40” C Max)

DELIVE RY RATE RANGE

Lines A and B: 0.1

equivalent pressure

Relative Humidity: 10

- 3,000m) or10,000 feet (0

60” C

Atmospheric Pressure: 0

-20” to

40” C

Storage Tem perature:

5” to

Specifications

ENVIRONMENT

Operating Temperature:

Page 78: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 78/85

5101)- 3 430-95226-001 (Rev.System Operating Manual

mL

* Baseline backpressure is 0 psig *

mL 0.70slg 0.595 

mLL 0.30pslg 0.24m.Uhr 6

mLL 0.63PQE 0.555 

mLL 0.28~51~ 0.23J.Jhr 6

ImLJhr 6 psig 17 minutes

Maximum Time to Alarm

(Macro)

47 seconds

106 seconds

22 minutes

15 psig 45 minutes 49 minutes

* Baseline backpressure is 0 psig *

BOLUS VOLUM E REL EASED AFTER DOWNSTREAM OCCLUSIONS ARE CORRECTED

Rate Occlusion Alarm Typical Bolus Volume Maximum Bolus Volume

Pressure Setting (Macro) (Macro)

25

mIfhr 6 psig 38 seconds

15 psig 98 seconds

k user-selectable.(without alarm): Factory default setting is 6 psig.

Maximum Infusion Pressure: 20 psig

TIM E TO DETECT DOWNSTREAM OCCLUSIONS

Rate Occlusion Alarm Typical Time to Alarm

Pressure Setting (Macro)

25

occluded.

Distal Pressure Limit 1 to 15 psig. The maximum pressure

alamr sounds after the distal set

tubing or set outlet fitting becomes occluded.

Proximal Occlusion: The PROXIMAL OCCLUSION alarm sounds if the tubing

proximal to the cassette becomes

Specifications

OCCLUSION ALARM AND LI MITS

Distal Occlusion: The DISTAL OCCLUSION

Page 79: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 79/85

9- 4 Plum A+ Infusion Pump/01)

as a straight

line. The graph also presents maximum and minimum average delivery rate error for this interval plotted by

averaging delivery errors over intervals of 2, 5, 11, 19 and 3 1 minutes (“Trumpet Curve”).

430-95226-001 (Rev.

p& T wi : The graphs following the

Example plot flow rates at 30 second intervals for the first 2 hours and for the 96th hour of delivery. The

graph plots mean delivery rate error (Average of 3 pum ps) for the 2nd hour and the 96th hour

HogRUMPET CURVES

The Trumpet Cu rve Graphs following the Example show representative maximum and minimum percent

flow rate deviation from the programmed rate over time. This information was developed in accordance with

IEC 60601-2-24: 1998, Sub-Clause 50.102. Refer to this standard for detailed information.

1, For higher ratios, the absolute per-

centage of delivery for the lowest rate may be affected by n o more than 0 .5%. W hen air is present in the bub-

ble trap, the absolute percentage of delivery for the lowest rate may be affected by up to 2.0%. Whenvariations in container height are present, the absolute percentage of delivery for the lowest rate may be

affected by up to 2.5% for up to 24 inches of container height differences.

& B) are delivering, the ratio of delivery for the fluid with

the lowest rate may be affected by as m uch as 5% for ratios up to 9 to

+28 inches. A typical deviation under these conditions is

0.5%.

Concurrent Delivery Effect- When both lines (A

majr affect delivery accuracy by no

more than 2% for variations in the range -15 to

mL/hr) for backpressures up to 15 psig. A typical

deviation under these conditions is 0.3%. Negative backpressure may affect delivery accuracy no more than

2% for pressures up to -2 psig (56 inches of water). A typical deviation under these conditions is 0.3%.

Filling Head Effect- Variations in filling head (such as container height)

f 10%.

Backpressure Effect- Positive backpressure on the distal line may affect delivery accuracy by no more than0.5% per psig (2% per psig for rates between 0.1 and 0.9

rnL/‘hr, the delivery rate

error is less than or equ al to

mL/hr during normal operating conditions. For use of the device at rates below 1

2 5% over the course of 48 hours at a programmed rate of 1 to

999

DELIVE RY ACCURACY

The Plum A+ Infusion pump was designed and m anufactured to maintain a volumetric delivery rate error of

the total fluid delivered of less than or equal to

Page 80: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 80/85

Page 81: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 81/85

Page 82: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 82/85

Page 83: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 83/85

Page 84: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 84/85

Page 85: Abbott Plum a+

7/27/2019 Abbott Plum a+

http://slidepdf.com/reader/full/abbott-plum-a 85/85