www.sori.org.au Sydney Orthopaedic Research Institute Accuracy of PSI: control with navigation Sebastien Lustig MD, Prof , Sam Oussedik FRCS, Corey Scholes PhD, Myles RJ Coolican FRACS, David A Parker FRACS 1 CHU Lyon Nord, Centre Albert Trillat, France 2 Sydney Orthopaedic Research Institute, Australie

Upload: others

Post on 05-Feb-2022

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Accuracy of PSI: control with navigation

www.sori.org.au

Sydney Orthopaedic Research Institute

Accuracy of PSI: control with navigation

Sebastien Lustig MD, Prof, Sam Oussedik FRCS, Corey Scholes PhD,

Myles RJ Coolican FRACS, David A Parker FRACS

1 CHU Lyon Nord, Centre Albert Trillat, France2 Sydney Orthopaedic Research Institute, Australie

Page 2: Accuracy of PSI: control with navigation

www.sori.org.au

Page 3: Accuracy of PSI: control with navigation

www.sori.org.au

80,000 TKA withPSI in 2012

Page 4: Accuracy of PSI: control with navigation

www.sori.org.au

Literature …

0

5

10

15

20

25

2008 2009 2010 2011 2012 2013

80,000 PSI TKAs

Page 5: Accuracy of PSI: control with navigation

www.sori.org.au

Page 6: Accuracy of PSI: control with navigation

www.sori.org.au

Problem ?

Someone at computer making all the decision for you …

Page 7: Accuracy of PSI: control with navigation

www.sori.org.au

Problem ?

How can we control the accuracy of the process…

Page 8: Accuracy of PSI: control with navigation

www.sori.org.au

Aim Assess intraoperatively the accuracy

Of a Patient Specific Instrumentation

for TKA

Using a navigation system

PSI

vs

Navigation

Page 9: Accuracy of PSI: control with navigation

www.sori.org.au

Method Visionaire® Smith and Nephew

Page 10: Accuracy of PSI: control with navigation

www.sori.org.au

Method

60 consecutiv primary TKAs Genesis II PS cemented Navigation system Stryker

Precision®

Patient specific instrumentation Visionaire® (SN)

Positioning checked with CAS

(PSI used only if positioning was ideal – error less than 1° or 1mm)

Page 11: Accuracy of PSI: control with navigation

www.sori.org.au

Method

Page 12: Accuracy of PSI: control with navigation

www.sori.org.au

Method : Parameters assessed

Femur

Bone cuts :

- Frontal plane

- Sagittal plane

- Rotation

- Thickness

- Implant Size

Page 13: Accuracy of PSI: control with navigation

www.sori.org.au

Method : Parameters assessed

Tibia

Bone cuts :

- Frontal plane

- Tibial Slope

- Thickness

- Implant Size

Page 14: Accuracy of PSI: control with navigation

www.sori.org.au

Results – Alignement

Mean 0.6°

79.3% à +/- 3°

Page 15: Accuracy of PSI: control with navigation

www.sori.org.au

Tibial Slope

Mean -0.1°

80.7% à +/- 3°

Page 16: Accuracy of PSI: control with navigation

www.sori.org.au

Femoral rotation

Mean 0,6°

77.2% à +/- 3°

Page 17: Accuracy of PSI: control with navigation

www.sori.org.au

Femur : sagittal alignment

Mean 2.1°

65.4% à +/- 3°

Page 18: Accuracy of PSI: control with navigation

www.sori.org.au

Results Compared to preop plan :

Implants size : Femur 52% Tibia 50%

Bone cuts thickness:

Difference(mm)

Max (mm) % +/- 2mm

FemurMedial 0.0 + 1.2 -3.5 to 6.5 87.7Lateral 0.25 + 1.1 -6.5 to 6.5 87.7

TibiaMedial 0.1 + 1.2 -6.0 to 3.0 78.9Lateral 0.1 + 1.1 -7.0 to 3.0 78.9

Page 19: Accuracy of PSI: control with navigation

www.sori.org.au

Discussion20% mismatch > 3° for all parameters

35% mismatch > 3° for sagital positionning of the femur

« Mistakes » up to 8° for the frontal plane and 13° for the sagittal plane

Page 20: Accuracy of PSI: control with navigation

www.sori.org.au

“ (…) According to the above criteria, the custom cutting jigs were reliable in 34 out of 48 measurements (75 %)(…) ”

Page 21: Accuracy of PSI: control with navigation

www.sori.org.au

DiscussionConteduca KSSTA 2012

12 patients

Tibia frontal 1.2° (0°-5°)Tibia sagittal 3.8° (0°-7,5°)Femur frontal 1.2° (0°-6°)Femur sagittal 3.7° (0°-9°)

“In cases of the use of the custom made cutting jigs it is recommended to perform an accurate control of the alignment before making the cuts, for any step of the procedure”

Sagittal plane! !

Page 22: Accuracy of PSI: control with navigation

www.sori.org.au

DiscussionPreop MRI limited to the knee?

Sagittal plane! !

Page 23: Accuracy of PSI: control with navigation

www.sori.org.au

Limits

- Accuracy of the navigation system ? (1°-1mm)

- Parameters assessed are different between CAS and PSI ?

- Our results cannot be extrapolated to other systems

Page 24: Accuracy of PSI: control with navigation

www.sori.org.au

” (…) the error for total coronal alignment exceeded 3° for 27 % of the sample (…) “

n=30

“(…)Despite good coronal alignments of the single prosthetic components, the lower limb mechanical axis was not restored correctly in a number of patients. (…) ”

n= 50

Trumatch®(Depuy)

PSI® (Zimmer)

Page 25: Accuracy of PSI: control with navigation

www.sori.org.au

2014

Visonaire® (SN)Trumatch® (Depuy)Signature® (Biomet)

Page 26: Accuracy of PSI: control with navigation

www.sori.org.au

Conclusion

- Unsatisfactory accuracy of the visionairePSI system in that study,

- Origin of the error still to be determined

- Promising technology, need for improvement,

- Currently unacceptable risk for malalignment if used without intraoperative control.