accurate determination of protein profiles in complex biological samples by maldi-tof ms
DESCRIPTION
Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS. Marvin L. Vestal SimulTOF Systems Lunchtime Workshop MSACL Tuesday, March 4, 2014. Commercial products introduced at ASMS 2012. V. SimulTOF Systems. Velocity and space focus. V g. 100 Linear. - PowerPoint PPT PresentationTRANSCRIPT
![Page 1: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/1.jpg)
Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS
Marvin L. VestalSimulTOF Systems
Lunchtime WorkshopMSACL
Tuesday, March 4, 2014
![Page 2: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/2.jpg)
Commercial products introduced at ASMS 2012
![Page 3: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/3.jpg)
Velocity and space focusSimulTOF Systems
Vg
V
Vg
V
Vg
V
100 Linear
200 Combo
300 Tandem
detector
detector
detector
Ion mirror
Ion mirror
Collision cell Ion accelerator
detector
timed ion selector
![Page 4: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/4.jpg)
1 shot
10 shots
Mixture of peptide standards, 100 femtomole/mL, 2.5 mm spotLaser spot ca. 50 mm, fluence 1.3x threshold, 0.4 attomole/spot=240,000 molecules
~24,000 ionsdetected
10%over allefficiency
![Page 5: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/5.jpg)
Serum sampleDiluted 1:10in sinipinic acid
Sensitivity, Dynamic Range, and Reproducibility are Key Metrics
Data from SimulTOF 100 Linear
![Page 6: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/6.jpg)
Serum sampleDiluted 1:10in sinipinic acid
Superposition of spectra from 5 different spots
Data from SimulTOF 100 Linear
Sensitivity, Dynamic Range, and Reproducibility are Key Metrics
![Page 7: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/7.jpg)
These are raw data. No normalization, background subtraction, smoothingor other data processing has been employed.
Expanded view
![Page 8: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/8.jpg)
Expanded view
Expanded view of highmass region serum
albumin
x20
x20
![Page 9: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/9.jpg)
Conventional Wisdom
•MALDI• is Not •Quantitive
![Page 10: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/10.jpg)
4 spectra 50 shots ea. On one sample spotIntensity varies by factor of 20@10 Hz requires 20 s
Your Old MALDI
![Page 11: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/11.jpg)
Data is acquired over mass range 1000-20,000 Da using 2 ns bins2 kHz laser rate50 shots averaged per spectrum1 mm/s snake raster for 5 passes at 500 mm over 3 mm dia spot Total travel 12 mm in 12 s generating 480 spectra with 50 mm resolutionSpectra with no significant peaks are not savedOnly 10% of sample used, 240,000 total laser shots possible
How do we make MALDI Quantitative????Data Acquisition, One Example
Data Processing (SimulTOF Wizard)
Spot AverageBaseline CorrectionSmoothCalibrate and Detect PeaksNormalizeBin by mass rather than timeQuantifyReport
Peak table includes:Intensity (number of ions/peak) Massstandard deviation (square root of number of ions/peak)Resolving power =m/dm=mass/FWHM
![Page 12: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/12.jpg)
Data is binned into mass bins using the following form:
mn+1=mn(1+r), mn=m0(1+r)n, mN=m0(1+r)N,m0 is the first mass, mN is the last mass, and N is the total number of bins.Log(mN/m0)=N[log(1+r)], r=10[x/N]-1, where x=log(mN/m0), for example, if mN/m0)=20, R=1/r=m/Dm=400Then N=1200
One approach is to detect peaks in the unknown spectrumAssign peaks to bins and sum intensity of all peaks in the bin.
An alternative approach is to sum intensity in each mass binand use this intensity as that for each bin.
The challenge is to determine which approach is best at minimizing instrumental artifacts due to such problems as incomplete separationof adjacent peaks.
We have evaluated performance with R=400 and R=2000
Mass Binning
![Page 13: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/13.jpg)
Comparing Spectra
Use standard vector algebra on normalized and binned spectra
Difference in spectra computed by
(Na -Nb )/Square Root (Na +Nb )
For each bin
![Page 14: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/14.jpg)
347 spectra 50 shots ea.12s @2000 Hz
![Page 15: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/15.jpg)
0.5 mm
3 mm
17,400 shots recorded out of 24,000 shots applied10% of sample scanned in 12 sSpot 176
![Page 16: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/16.jpg)
4 spectra 50 shots ea. On one sample spotIntensity varies by factor of 20@10 Hz requires 20 s
Your Old MALDI
![Page 17: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/17.jpg)
Processed spectrum from same sample spot as above Our New MALDI
![Page 18: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/18.jpg)
Spectra from 4 different sample spots super imposed Our New MALDI
![Page 19: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/19.jpg)
![Page 20: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/20.jpg)
Binned 2000 rp
![Page 21: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/21.jpg)
Saliva Sample Prep
Individuals spit into tubes, ~ 11 AM. Participants DK, DP, KP, MV.
Spin 5 min.
Dilute 10 ul of supernatant into 90 ul of 5 mg/ml HCCA matrix in 75% acetonitrile / 0.1% TFA Spot 2ul per well.
First set of Sample tubes saved in refrigerator at 4 C for 6 days. Not stirred
Fresh samples collected ~ 11AM.
Spin again, repeat as above.
2 ul sample to 20 ul of matrix for rerun
Samples for DK, DP, KP, MV. 6 each.
![Page 22: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/22.jpg)
dk
dpkp
mv
dp
![Page 23: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/23.jpg)
DN
DN/s
dp-dp
mv-dp
Normalized to TIC, 400 RP
![Page 24: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/24.jpg)
kp-dp
dk-dp
Normalized to TIC, 400 RP
DN/s
![Page 25: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/25.jpg)
kp-dp
Normalized to TIC, 400 RP
DN/s
![Page 26: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/26.jpg)
dp-dp
mv-dp
Normalized to Base Peak, 400 RP
DNDN/s
dp-dp
![Page 27: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/27.jpg)
kp-dp
dk-dp
dp-dp
mv-dp
Normalized to Base Peak, 2000 RP
DNDN/s
![Page 28: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/28.jpg)
mv
dkmv
![Page 29: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/29.jpg)
mv
Fresh samples
After 6 days at 4C
No change
![Page 30: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/30.jpg)
dp
mv
Fresh samples
After 6 days at 4CNo change
![Page 31: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/31.jpg)
mv
dp
Fresh samples
After 6 days at 4C
![Page 32: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/32.jpg)
dp
Fresh samples
After 6 days at 4C
![Page 33: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/33.jpg)
The Future of Diagnostic MSMALDI-TOF
Very fast (full spectrum/laser shot @5 kHz)Tradeoff between speed and sensitivity
Can be interfaced with variety of separationsFully automated, no operator required
SimulTOF Model 100 Linear MALDI -sensitivity, dynamic range, speed, and simplicity -wide mass range for proteins, peptides, and small molecules SimulTOF Model 200 Combo MALDI -adds high resolution reflector -accurate mass of peptides and small molecules; SimulTOF Model 300 MALDI MS-MS -multiplexed TOF-TOF -high resolution precursor selection -identification and quantitation of peptides and small molecules.
Modular BenchTop Systems
Highest Performance at Lowest Cost
![Page 34: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/34.jpg)
Sample AcquisitionAnd ProcessingSpotting on MALDIPlate and addingMatrixSample ID
MALDI-TOFMassSpectrometer
Data processingInterpretationStorage and reportingMay include resultDependent modsTo MS method
user
applicationreport
massspectra
MSmethod
sampleinformation
sampleinformation
sampleinformation
MALDIsampleplate
SampleData
Pathogen IdentificationCancer typing directly from serum,
Tissue imaging of Proteins for
cancer typing
Tissue imaging of
Small molecules
for drug disposition
Biomarker Validation
Clinical assays of Biomarkers
Protein array reader
![Page 35: Accurate determination of protein profiles in complex biological samples by MALDI-TOF MS](https://reader035.vdocuments.net/reader035/viewer/2022062521/5681668c550346895dda55b3/html5/thumbnails/35.jpg)
Conclusions• Resolving power 500-1000 over wide range is routine• Normalization to TIC or base peak removes most of amplitude
variation• Each spot will yield up to 200,000 shots without degrading
resolving power or accuracy and giving dynamic range limited only by chemical noise
• Results might be improved by multiple levels of dilution and use of alternative matrices
• Mass error <30 ppm across the plate over the full mass range with single peak automatic calibration
• Dynamic range up to100,000