322 Vol.11 No.4 July 2014 Special Report CHINA FOUNDRY Celebrating the 10 th Anniversray 2004-2014 Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., Chief Engineer, Doctoral Supervisor. He is currently Director of New Technology Research & Development Center in the Shenyang Research Institute of Foundry. His research interests focus mainly on the research and application for advanced functional materials, structural materials and porous foam metals and their forming technologies. He has finished nearly 30 scientific research projects, which include the “863 Program”, the “973 Program”, the National Natural Science Foundation of China, National Science and Technology Major Project of the Ministry of Science and Technology of China, and key military projects. To date, he holds 7 national invention patents, and has won many honors, such as the National Award for Science and Technology Progress (second place), and the State Class Persons of National Talents Engineering. He is a recipient of the government special allowance. E-mail: [email protected] *Shen Gang: Ph.D., Professor and Vice Dean of School of Stomatology, Shanghai Jiao Tong University. Main research interests: biological and bio- molecular response of bone junctions in craniofacial regions to external forces. E-mail: [email protected] Received: 2014-06-10 Accepted: 2014-07-01 Zhou Linxi 1 , Yang Quanzhan 2 , Zhang Guirong 3 , Zhao Fangxin 2 , *Shen Gang 1 and *Yu Bo 2 1. Department of Orthodontics, Shanghai Ninth People`s Hospital, Shanghai Jiao Tong University, Shanghai 200011, China; 2. Shenyang Research Institute of Foundry, Shenyang 110022, China; 3. Shenyang Stomatology Hospital, Shenyang 110002, China Abstract: Biomedical metal materials with good corrosion resistance and mechanical properties are widely used in orthopedic surgery and dental implant materials, but they can easily cause stress shielding due to the significant difference in elastic modulus between the implant and human bones. The elastic modulus of porous metals is lower than that of dense metals. Therefore, it is possible to adjust the pore parameters to make the elastic modulus of porous metals match or be comparable with that of the bone tissue. At the same time, the open porous metals with pores connected to each other could provide the structural condition for bone ingrowth, which is helpful in strengthening the biological combination of bone tissue with the implants. Therefore, the preparation technologies of porous metal implants and related research have been drawing more and more attention due to the excellent features of porous metals. Selective laser melting (SLM) and electron beam melting technology (EBM) are important research fields of additive manufacturing. They have the advantages of directly forming arbitrarily complex shaped metal parts which are suitable for the preparation of porous metal implants with complex shape and fine structure. As new manufacturing technologies, the applications of SLM and EBM for porous metal implants have just begun. This paper aims to understand the technology status of SLM and EBM, the research progress of porous metal implants preparation by using SLM and EBM, and the biological compatibility of the materials, individual design and manufacturing requirements. The existing problems and future research directions for porous metal implants prepared by SLM and EBM methods are discussed in the last paragraph. Key words: additive manufacturing; SLM; EBM; porous metal implant; biocompatibility CLC numbers: TG249/TG132.3 + 3 Document code: A Article ID: 1672-6421(2014)04-322-10 B iomedical metal materials with high corrosion resistance, strength and toughness are widely used for bone surgery and dental implant materials. Conventional biomedical metal materials such as Ti6Al4V, 316L, CoCrMo and CoNiCrMo, etc. are mainly used for reparation of bony tissue under high load and complex stress conditions, such as bones, joints, and teeth [1] . Most metal implants used for clinical practice are fully dense, but their elastic moduli are obviously higher than that of bony tissues. This means that loadings can't be transferred efficiently from metal implants to the adjacent tissue, leading to stress

Upload: buidung

Post on 08-Feb-2018

223 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

322

Vol.11 No.4 July 2014Special ReportCHINA FOUNDRY Celebrating the 10th Anniversray 2004-2014

Additive manufacturing technologies of porous metal implants

* Yu BoPh.D., Chief Engineer, Doctoral Supervisor. He is currently Director of New Technology Research & Development Center in the Shenyang Research Institute of Foundry. His research interests focus mainly on the research and application for advanced functional materials, structural materials and porous foam metals and their forming technologies. He has finished nearly 30 scientific research projects, which include the “863 Program”, the “973 Program”, the National Natural Science Foundation of China, National Science and Technology Major Project of the Ministry of Science and Technology of China, and key military projects. To date, he holds 7 national invention patents, and has won many honors, such as the National Award for Science and Technology Progress (second place), and the State Class Persons of National Talents Engineering. He is a recipient of the government special allowance.E-mail: [email protected]

*Shen Gang: Ph.D., Professor and Vice Dean of School of Stomatology,

Shanghai Jiao Tong University. Main research interests: biological and bio-molecular response of bone junctions in craniofacial regions to external forces.E-mail: [email protected]

Received: 2014-06-10 Accepted: 2014-07-01

Zhou Linxi1, Yang Quanzhan2, Zhang Guirong3, Zhao Fangxin2, *Shen Gang1 and *Yu Bo2

1. Department of Orthodontics, Shanghai Ninth People`s Hospital, Shanghai Jiao Tong University, Shanghai 200011, China; 2. Shenyang Research Institute of Foundry, Shenyang 110022, China; 3. Shenyang Stomatology Hospital, Shenyang 110002, China

Abstract: Biomedical metal materials with good corrosion resistance and mechanical properties are widely used in orthopedic surgery and dental implant materials, but they can easily cause stress shielding due to the significant difference in elastic modulus between the implant and human bones. The elastic modulus of porous metals is lower than that of dense metals. Therefore, it is possible to adjust the pore parameters to make the elastic modulus of porous metals match or be comparable with that of the bone tissue. At the same time, the open porous metals with pores connected to each other could provide the structural condition for bone ingrowth, which is helpful in strengthening the biological combination of bone tissue with the implants. Therefore, the preparation technologies of porous metal implants and related research have been drawing more and more attention due to the excellent features of porous metals. Selective laser melting (SLM) and electron beam melting technology (EBM) are important research fields of additive manufacturing. They have the advantages of directly forming arbitrarily complex shaped metal parts which are suitable for the preparation of porous metal implants with complex shape and fine structure. As new manufacturing technologies, the applications of SLM and EBM for porous metal implants have just begun. This paper aims to understand the technology status of SLM and EBM, the research progress of porous metal implants preparation by using SLM and EBM, and the biological compatibility of the materials, individual design and manufacturing requirements. The existing problems and future research directions for porous metal implants prepared by SLM and EBM methods are discussed in the last paragraph.

Key words: additive manufacturing; SLM; EBM; porous metal implant;

biocompatibility

CLC numbers: TG249/TG132.3+3Document code: A Article ID: 1672-6421(2014)04-322-10

Biomedical metal materials with high corrosion resistance, strength and toughness are widely used for bone surgery and dental implant materials. Conventional biomedical

metal materials such as Ti6Al4V, 316L, CoCrMo and CoNiCrMo, etc. are mainly used for reparation of bony tissue under high load and complex stress conditions, such as bones, joints, and teeth [1].

Most metal implants used for clinical practice are fully dense, but their elastic moduli are obviously higher than that of bony tissues. This means that loadings can't be transferred efficiently from metal implants to the adjacent tissue, leading to stress

Page 2: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

323

Vol.11 No.4 July 2014Special Report CHINA FOUNDRYCelebrating the 10th Anniversray 2004-2014

Fig.1: The application fields of additive manufacturing

shielding between implants and bones. As a result, the stresses are always absorbed by the implants, causing amyotrophy and osteo-necrosis of bones around the implants, distortion of new bone and reduction of bearing capacity of connective parts [2]. Also the smooth surface of dense metal implants makes it difficult to combine well with the host bone. We noticed that surface treatments, such as sand blasting, spraying, or sintering, could increase the roughness and surface area of implants [3] and improve the adhesion strength of the implants with the host bone to a certain degree. However, the state of bone tissues covering around the implants is not changed and the low bonding strength problem has not been essentially solved. The elastic modulus of porous metals is lower than that of dense metals. By adjusting the pore parameters, the elastic modulus of the porous metals can be made to match that of the bone tissues [4]. The open and connected pores of the porous metals can provide space and channels for bone ingrowth, which helps bone tissue growth from the surface into the interior of the implant, strengthening the biological combination between bone tissues and the implants. The preparation technologies of porous metal implants and related research have drawn more and more attention due to the excellent features of porous metals [5].

At present, the preparation methods of porous metal implants mainly include space-holder, polymeric sponge dipping, self -propagation high-temperature synthesis and chemical vapor deposition [6-11]. However, for the bone repair implants, the size varies from one person to another, the shape is complicated and full of tiny details, making it difficult to flexibly control pore structure and liberally manufacture implants by considering different bone defect forms. Such additive manufacturing technologies as selective laser melting (SLM) and electron beam melting technology (EBM) are suitable for fabricating porous metal implants. The two technologies are critical in satisfying the individualized requirements. This is because they can not only produce adjustable elastic modulus and open porous metal implants, with pores connected to each other, but also give

the implants complicated interior and exterior shapes simply, rapidly, and accurately. The application of SLM and EBM in the field of additive manufacturing for fabricating porous metal implants is just getting started. Reviewing the research progress, understanding the problems and further research directions for porous implants will provide reference and enlightenment for further study of porous metal implant fabricated by additive manufacturing.

1 The technology status of SLM and EBM

Additive manufacturing technology (also known as rapid prototyping or 3D printing technology) is based on the idea of discrete accumulation forming. This applies the principle of material accumulating fabrication to manufacture complex three-dimensional parts under computer control by virtue of CAD data and model. It is a "bottom to top" material accumulation manufacturing method [12] compared to the traditional material cutting or machining technology. Such a method therefore can substantially shorten the processing because it omits the cutting tool, fixture and multi-step forming processes. In addition, the more complex the product shape, the more obvious the gain in manufacturing speed [13]. Additive manufacturing technology became available in the late 1980s. Since then, about 20 kinds of technology have been developed, such as Stereo lithography (SL), Laminated Object Manufacturing (LOM), Laser Cladding Forming (LCF), Selective Laser Sintering (SLS), Selective Laser Melting (SLM), and Electron Beam Melting (EBM). These are found in use in the fields of automation, consumer electronics, medical/dental, industrial machinery, aerospace, etc. The application of additive manufacturing in the medical/dental field shows a rising trend with increases of 13.6%, 15.9% and 16.4% in the years 2011, 2012, and 2013, respectively, as shown in Fig. 1.

Page 3: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

324

Vol.11 No.4 July 2014Special ReportCHINA FOUNDRY Celebrating the 10th Anniversray 2004-2014

SLM and EBM are two kinds of precision forming technologies which are suitable for making metal parts with complex structures. The forming process of SLM is operated in an inert gas chamber, in which a laser scans the metal powder supplied by a powder spreader according to the designed paths. The bulk form is produced by metallurgical bonding of the metal powder layer by layer (Fig. 2) [14]. The EBM forming process is operated in a vacuum environment and the bulk form is produced by an electron beam in a similar way to the SLM method (Fig. 3). The main difference between SLM and EBM is the heat source, i.e., a laser for SLM and an electron beam for EBM. The advantages of the SLM method are the more developed technology, cheaper equipment and higher forming precision, due to a smaller particle size of the powder condensed by the small diameter of the laser spot. However, the forming speed of EBM (80 cm3·h-1) is 4 to 5 times faster than that of SLM resulting from its high energy (as high as 3000 W, 10 times larger than that of SLM) supply and larger particle size. Because of the larger particle size, the forming precision of EBM method is relatively low.

SLM was proposed by the Fraunhofer Institute in 1995 and the SLM system was first developed by the MCP Company

Fig. 2: Schematic illustration of the SLM process

Fig. 3: Schematic illustration of the EBM process [14]

in Germany in 2003. Since then, commercial equipment has been developed by many Germany companies, including EOS, Concept Laser, and ILT. A commercial SLM machine named EOSINT M270 equipped with a 200 W solid state Yb fiber laser was recently developed by the EOS Company. The spreading thickness of the powder layer is 20 to 100 μm. The spot diameter is 70 to 200 μm. The typical scanning speed is 750 mm·s-1, the dimensional accuracy of parts is 20 to 80 μm, and the surface roughness of parts is 10 to 15 μm. Parts made by SLM equipment have been successfully used in aerospace, automotive, medical, injection molding, and other fields. In China, the Rapid Manufacturing Center of Huazhong University of Science and Technology launched a diode pumped 150 W YAG laser and SLM devices using a 100 W fiber laser in 2003, as a prelude to China's research on SLM [15]. In 2007, South China University of Technology developed DiMetal-280 type of SLM prototype based on DiMetal-240 equipped with a Yb fiber laser. The spot diameter is 50 to 200 μm, the typical scanning speed is 200 to 600 mm·s-1, the dimensional accuracy is 20 to 100 μm, and the surface roughness is 20 to 30 μm. Beijing Institute of Aeronautical Manufacturing Engineering developed China’s largest SLM equipment (type LSF-M360) with forming dimensions of 350 mm × 350 mm × 400 mm. It seems that the research on SLM equipment and technology has made some progress in China. However, the key components of SLM equipment, such as the laser, focusing mirror, and high-speed scanning galvanometer are still dependent on imports.

EBM research began in the early 21st century. In 2003, the Swedish Arcam Company independently developed the earliest EBM for commercial use in the world. The largest forming dimensions are 200 mm × 200 mm × 160 mm. The spreading thickness of the powder layer is 50 to 200 μm. The scanning speed is over 1000 m·s-1. The beam spot diameter is 300 to 500 μm. The dimensional accuracy is 200 to 400 μm [16]. Many research institutes and universities in the USA, Japan, Britain, etc. have purchased EBM equipment from the Arcam Company to carry out research in multiple fields. At present, more than 100 EBM units are in use around the world. In 2004, Tsinghua University developed the first domestic EBM prototype (type EBSM150), with a maximum scanning speed of 2 m·s-1 [17]. In 2008, Tsinghua University again developed the second generation EBSM250 electron beam forming equipment, and the largest forming dimensions are 230 mm × 230 mm × 250 mm. This is based on the research on several key problems related to EBM process, such as the powder preheating process [18], scanning path planning, and mechanical properties of the part [19]. It was reported that the load-bearing components for Ti6Al4V aircraft manufactured by the EBM process have excellent mechanical properties which equal to that of the parts made by a forging process [20]. Furthermore, the porous titanium alloy made by the EBM method has also stimulated the research interest in Northwest Research Institute of Non-ferrous Metals and much work has been done [21].

The development of additive manufacturing technology of

Page 4: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

325

Vol.11 No.4 July 2014Special Report CHINA FOUNDRYCelebrating the 10th Anniversray 2004-2014

SLM and EBM is supported by the innovation of equipment, the advancement of the forming process and the development of metal powder. The preparation of metal powder materials is a very important part in the development of additive manufacturing technology. Additive manufacturing has special requirements for the metal powder, such as small particle size, narrow particle size distribution, high sphericity, good liquidity, and low oxygen content [22]. At present, the general preparation methods of metal powder for additive manufacture include Plasma Atomization, Plasma Spheroidization, Vacuum Induction-melting Inert Gas Atomization (VIGA) and Electrode Induction-melting Inert Gas Atomization (EIGA). The varieties of the metal powders are fewer. At present, only Ti6Al4V, CP-Ti, In625, CoCrMo, and 316L are available on the market. AP&C Metal Powder Division of Canada Raymor Industries Company

reserves the proprietary right of plasma atomization technology and is one of the suppliers of the main metal powders. Its products are characterized by a high degree of sphericity, high purity, low oxygen content, uniform particle size distribution, and high powder density. The metal powders being used by the Swedish Arcam Company come from the company. The USA’s Ametek employs hydrogenation and dehydrogenation plasma spheroidization and gas atomization to prepare spherically shaped metal powders with diameters of about 100 μm (Fig. 4), while ALD Vacuum Technologies of Germany uses VIGA and EIGA to produce high-quality metal powders marked with fine particle size and narrow particle size distribution (Table 1). However, in China, the high quality metal powders needed for additive manufacturing depend on imports.

Fig. 4: Additive manufacturing metal powders provided by Ametek: (a) Ti6Al4V powders using Plasma Spheroidization and (b) 316L powders using Gas Atomization

(a) (b)

Table 1: Technical data of additive manufacturing metal powders using VIGA and EIGA

Technical data of metal power VIGA EIGA

Typical powder alloy for metal additive manufacturing

CoCrMo, In625, In718 CP-Ti, Ti6Al4V, TiAl

Powder morphology Spherical Spherical

d50 [PSD mass median] 35-70 μm 60-100 μm

+10-45 μm +10-45 μm

Typical available size classes for +25-45 μm +25-45 μm

metal additive manufacturing +45-63 μm +45-63 μm

+45-105 μm +45-105 μm

2 The progress of porous metal implants made by SLM and EBM

Suitable pore structure, such as the dimension, shape, distribution, spatial direction and mutual connectivity, is conducive to the hard and soft human tissues ingrowth around the implants. The mutual chimeric and the biocompatibility could be realized between implants and human tissues [23].

The porous structure also could reduce the elastic modulus of metal implants. Pore structure, therefore, is the key factor for mechanical compatibility and biocompatibility of bone implants with the host bone. Reports on mechanical compatibility and biocompatibility of pore structure design and related research are seldom found. The recent study on the porous metals made by SLM or EBM is only based on the honeycomb [24] or crystal lattice structures [25], without a focus on the analysis of

Page 5: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

326

Vol.11 No.4 July 2014Special ReportCHINA FOUNDRY Celebrating the 10th Anniversray 2004-2014

design principles and pore structure criteria. In addition to the pore structure of implants, connection problems between the implants and host bone with inhomogeneous porous structure (Fig. 5) [26] still need to be considered. Even if the whole structure of the implant were porous and the elastic modulus of the implant matched that of the host bone, it is still possible that requirements for the strength of load-bearing parts and the stability of the connection between the implant and the host bone can not be satisfied. Therefore, an implant with a gradient pore structure was brought into application (Fig. 6) [27].

Fig. 5: The structure of bone tissue

Fig. 7: Porous acetabulum cup implants produced by Swedish Arcam Company and the illustration of application position

Fig. 6: Schematic of the connection interface between gradient pore structured implant and bone tissue [25]

A porous Ti6Al4V acetabulum cup implant with “a porous surface and a dense inner layer structure” fabricated by EBM has been successfully applied in clinical practice (Fig. 7) [28]. The porous surface can not only provide a good environment for bone ingrowth and decrease the elastic modulus difference between the implant and host bone, but also reduce the stress shielding effect; meanwhile, the dense inner layer can satisfy the strength requirement of the load-bearing parts and the connection stability requirement of the implants and host bone. A report on the Ti6Al4V dental implant with a porous surface and a dense inner layer structure by Traini et al. of Chieti-Pescara University, Italy, using the SLM method (Fig. 8) [29] showed that the rough surface of the implant is good for new bone ingrowth, and thus improves the connection strength between the implant and alveolar bone. It also showed that the elastic moduli of the dense layer and the porous surface layer are 104 ± 7.7 GPa and 77 ± 3.5 GPa, respectively, which reduce the elastic modulus difference between the implant and the human alveolar bone (the elastic modulus of a human alveolar bone is 20 ± 7.0 GPa), which is helpful in weakening the stress shielding effect. Lewis Mullen et al. [30] at the UK’s University of Liverpool developed a porous titanium implant by using an SLM facility (made by the MCP Company in Germany) equipped with a 200 W Yb fiber laser. The relationship between laser power and pore size was determined. When the pore dimensions were 600 μm, 800 μm, 1,000 μm and 1,200 μm, the appropriate requirement for laser power was 62 W, 96 W, 132 W and 167 W, respectively. The ideal powder particle size of the feeding material through optimization should be smaller than 45 μm. The fabricated porous titanium implant was characterized by a porosity of 10% to 95%, a pore wall thickness of 150 to 420 μm, a compressive strength of 0.5 to 350 MPa and precise dimensions. In addition, a method to adjust the porosity of porous metal by controlling the beam overlap procedure using SLM was suggested by Stamp et al. [31], and a porous metal implant with a porosity of 71%, average pore size of 440 μm, and compression strength of 70 MPa was obtained. It is worthy of note that Heinl et al. [32] at FAU in Germany, have successfully fabricated a porous Ti6Al4V

implant with an elastic modulus close to that of human bone by using EBM S12 made by the Arcam Company in Sweden. Recently an opening cup implant sample with different pore dimensions and porosity made with Ti6Al4V alloy was prepared using the EBM process [33, 34]. The pore dimensions and pore wall thickness are 765 to 1020 μm and 466 to 941 μm, respectively.

In China, there is less research on porous metal implants fabricated by SLM and EBM and there is only a small amount of literature about the investigation of pore structure design

Page 6: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

327

Vol.11 No.4 July 2014Special Report CHINA FOUNDRYCelebrating the 10th Anniversray 2004-2014

and forming techniques of porous metal implants. Based on the pore structure modeling method and the SLM manufacturing technology of porous implants, Dr. Xiao [35] of the South China University of Technology, proposed a method for an implicit surface which is represented by the implicit function of porous unit. According to the mechanical performance requirements of different defect locations of implant bone, different units can be constructed by changing the parameters of the implicit function. When the porous structure unit is mapped to the implant design domain, the design rules of porous structure based on SLM technology are then established. The porous structures of a hip joint and skull implants were designed and formed according to their individual requirements, which validate the feasibility of the design rules. Li [23] et al., of Shanghai Jiao Tong University, fabricated honeycomb porous Ti6Al4V implants using EBM equipment under the condition of vacuum < 0.5 Pa. The electron beam power was 4 kW, the processing layer thickness was 0.07 mm and the scan speed was 1 km·s-1. Scanning electron microscope observation indicated that porous titanium alloy prepared in this way has a three-dimensional interconnected pore structure consistent with the design shape. It also found that such porous Ti alloy presents about 15% shrinkage.

3 Studies on compatibility of porous metal implants

The compatibilities of the metal implants with the human body mainly include mechanical compatibility and biocompatibility, which relates to many aspects. Two very important aspects are the matching degree of the mechanical properties Fig. 9: A photograph of samples of (a) porous titanium

implant; and (b) coated with tantalum

Fig. 8: SEM photograph of porous titanium dental implant fabricated by SLM

between the implant and host bone, and the ingrowth condition for new bone provided by the implant. The stability to connect the implant and host bone, and the ability to induce new bone growth, are dependent on these two aspects.

A porous titanium implant [33] with diamond molecular structure was fabricated using the EBM process through the cooperation of Shanghai Jiao Tong University and the Institute of Metal Research of the Chinese Academy of Sciences. The implant was coated with a tantalum film using the Chemical Vapor Ceposition (CVD) method (Fig. 9). As a result, the porous titanium implant simultaneously has the combination of excellent mechanical performance and biological properties. The porosity, pore dimension and elastic modulus of the fabricated Ti6Al4V implant are 70.5±0.6 %, 700±50 μm and 11.3±0.4 GPa, respectively. The elastic modulus particularly satisfies the human cortical bone requirement. The compression strength, however, is lower than that of human cortical bone (Table 2) [36]. The USA Zimmer Company produced a porous tantalum dental implant

Page 7: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

328

Vol.11 No.4 July 2014Special ReportCHINA FOUNDRY Celebrating the 10th Anniversray 2004-2014

using the CVD method (Fig. 10). The porosity, pore dimensions and elastic modulus are in the range of 15% to 85%, 400 to 600 μm and 1.5 to 3 GPa, respectively. The elastic modulus of the implant is near to that of human cortical bone (Table 2) [37-39].

The biocompatibility of porous metal implants with the host bone depends on the pore structure and surface characteristics of the implant. It has been shown [40] that the ingrowth rate and the quantity of new bone into a porous metal implant are related to the pore structure of the implant. The pore structure characteristics can be described by the geometric parameters of the porosity and the pore diameter; the latter is the major factor influencing the growth of new bone. The results [41] given by Klawitter and Hulbert showed that the pore diameter should be larger than 100 μm in order to ensure blood circulation to the required bone tissue ingrowth. The growth speed and quantity of the new bone increase with the pore diameter. The results were further confirmed by Palmquist et al. [42] who embedded a titanium alloy implant with pore diameters from 500 μm to 700 μm into a sheep’s body for 26 weeks and found a lot of new bone tissues ingrown inside the implant. The porous structure of the implant is helpful to the new ingrown bone and improves the connection strength between the implant and the host bone. However, new ingrown bone cannot be induced if the surface of the porous metal implant is not modified by a coating treatment. Furthermore, the resultant corrosion products and pathological reaction after implantation in the human body will affect the normal biological function. Therefore, the coating modification treatment of the metal implants is a critical way of solving these problems. Nowadays, two kinds of coating materials, i.e. metallic and biological ceramic, are available for the coating of porous metals. One example is Ta coating fabricated by Li [36] who deposited tantalum metal on the surface of porous Ti6Al4V alloy using the CVD method. The tantalum coating combines well with the Ti6Al4V implant and does not possess biological toxicity caused by a metal ion dissolved into the blood. Therefore, it is bio-inert or lacks bioactivity in human

Table 2: Mechanical properties and porous parameters for bone tissues and porous metal implants [36-39]

Material Elastic Compression Yield Porosity Pore Fabricated

modulus (GPa) strength (MPa) Strength (MPa) % dimension (μm) method

Cortical bone 7 – 30 100 – 230 50 – 150 Cancellous bone 0.05 – 0.5 2 – 12 10 – 20 Porous Ti implant 11.3 ± 0.4 83.8 ± 4.2 74.6 ± 3.5 70.5 ± 0.6 700 ± 50 EBMPorous Ta implant 1.5 – 3 35 – 55 75 – 85 400 – 600 CVD

Fig. 10: A porous tantalum dental implant using the CVD method produced by the Zimmer Company, USA

body; but cannot induce new bone ingrowth. The biological ceramic coating on the implant surface is not only beneficial to improve the bioactivity, induce the ingrowth of new bone, and promote osseo-integration; but is also helpful to cut off the toxic substances released by the metal substrate. Apatite coatings are representative of biological ceramic coatings. A novel method called bionic solution was developed to coat apatite on a substrate with a complex shape and porous structure. The basic process of this method is that the implant is immersed into the simulated body fluid (SBF), and an apatite coating similar to human bone is induced spontaneously on the implant surface. The biomimetic solution method can provide a reference in exploring surface modification technology of porous implants fabricated by SLM and EBM. But for SBF, technology is still at the stage of laboratory research. Some problems, such as the low bonding strength between the coating and the substrate and the long cycle of growth for the coating, still need to be solved [43].

4 Individual design and manufacture of porous metal implant

The bone shapes and defect morphologies vary between

different individuals. Porous metal implants should not only have the mechanical properties and pore structure suitable for the ingrowth of new bone, but also should have the characteristic of matching the individual’s shape.

Additive manufacturing is a simple, rapid method for the

Page 8: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

329

Vol.11 No.4 July 2014Special Report CHINA FOUNDRYCelebrating the 10th Anniversray 2004-2014

Fig. 11: (a) Porous titanium femoral bone implantdigital model and (b) physical photo

fabrication of a porous metal implant with complex pore structure and matching shape close to the morphology of the bone tissue to be replaced. This is because the design and the manufacture of the individual porous metal implant are based on the data of the patient's own bone. First, the defective bone part of a patient should be scanned layer by layer by using CT imaging. If the bone is totally defective, mirroring images can be generated according to the complete bone data of the symmetrical parts. From this CT image data of bone, a multi-layer section contour can be obtained. Further, the CT data could be translated into 3D CAD data and an individual model can be generated using reverse engineering and computer technology. On this basis, the individual design could be realized and the implant could be fabricated using additive manufacturing. Although the individualized implants can be comparable to the structure and the appearance of bone tissue with the aid of CT scanning and additive manufacturing technology, the implementation process is too complex to complete without

5 Existing problems and prospectsSLM and EBM are two kinds of simple, rapid additive manufacturing technology used to fabricate porous implant materials. These two technologies have unique advantages in precisely controlling pore size, spatial distribution, shape, etc. In these new developing additive manufacturing technologies of SLM and EBM, some problems still exist which urgently need to be solved. Therefore, further in-depth and meticulous research in terms of preparation of porous metal implants should be considered.

(1) The accuracy of the equipment, the quality of the metal powder and the level of the forming process are the key factors which influence the development of SLM and EBM technology. In China, research on SLM and EBM is still in its infancy; there is a long way to go considering the great gap between the developments of commercial equipment, metal

the help of professional image processing software. A famous and mature image processing software now being applied in the bone surgery field is named Mimics, which was developed by the Belgium Materialise Company. The Mimics is a set of reverse engineering software which is the bridge between the medical and the mechanical fields. Using this, the CT layer scanning image data can be quickly translated into CAD/CAM data format for the manufacturing of the defective bone parts and can be shown in highly accurate 3D model for design purposes. A porous artificial femoral model with individual features was built by Su et al. [44] through reconstructing CT data in three-dimensions with the aid of the Mimics software. The porous Ti6Al4V artificial femoral implant fabricated using the SLM method is shown in Fig. 11. However, additive manufacturing technology used for individualized bone implant fabrication is still at the exploratory stage, and it still needs further clinical validation and strict examination and approval of the registration process before it becomes reality in practice.

powder and forming technology, etc. that already exists abroad. It is suggested that China should focus on the research and development of commercial equipment with stable performance, high forming precision, low cost, and independent intellectual property rights; and extend the species of metal powder to the requirement of additive manufacturing. Only if the development of high quality metal powder and the supporting forming processes are simultaneously promoted can the substantive breakthrough for the preparation of porous metal implants using SLM and EBM be realized.

(2) Much work has been done abroad related to the laser power selection, scanning path scheme design, powder particle size optimization and clinical evaluation in preparation of porous metal implant using SLM and EBM. Part of the work has resulted in products on a large scale. China is still at the exploratory research stage in the field of porous metal fabrication using the additive manufacturing method.

Page 9: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

330

Vol.11 No.4 July 2014Special ReportCHINA FOUNDRY Celebrating the 10th Anniversray 2004-2014

Researchers in China should actively carry out the research on pore structure design by considering both the mechanical compatibility and biocompatibility of porous metal implant, and pay particular attention to the development of porous implants with both gradient pore and bionic structures.

(3) The mechanical properties and biocompatibility of the porous metal implants are closely related to the pore diameter and other pore parameters. In order to optimize the pore structure, it is necessary, therefore, to investigate the influence of pore parameters on the mechanical properties and biocompatibility of the porous metal systematically and thoroughly.

(4) A biological ceramic coating on the surface of a porous metal implant can improve the biological activity. As a surface modification technique, the SBF method is suitable for forming a homogeneous ceramic coating on a complex and porous substrate. The problems for the SBF technology, however, are the low bonding strength between ceramics and the substrate and the longer growth circle of the ceramic coating. This should be regarded as one of the important research directions in developing the SBF technology.

(5) Fabrication of porous metal implants using SLM and EBM technologies is an interdisciplinary subject involving research on the thermo-physical properties, the melting and solidifying mechanism of the materials, processing techniques, compatibility with the human body and the clinical application of the implant, etc. This means that exchanges and cooperation between medical workers and materials workers are necessary in order to bridge the gaps and take advantage of the complementary disciplines. We expect that the technical difficulties will be surmounted as soon as possible in order to realize the application and popularization of porous metal implants.

References[1] Li Boqiong. The investigation of pore characteristic and

mechanical property of porous Ti. Dissertation, Dalian Jiao Tong University, 2005. (In Chinese)

[2] Hu Haibo, Liu Huiqun and Wang Jieen. Research Progress of Biomedical Porous Titanium and Its Alloys. Materials Review, 2012, 26(19): 262-270. (In Chinese)

[3] Li Qing, Wang Hua and Zhang Lixin. The Surface Treatment of Artificial Joint Metal Materials. Chinese Journal of Medical Instrumentation, 2002, 26 (1): 44-47. (In Chinese)

[4] Agrawal C M. Reconstructing the Human-Body Using Biomaterials. JOM, 1998, 50(1): 31-35.

[5] Ryan G and Pandit A. Fabrication methods of porous metals for use in orthopaedic applications. Biomaterials, 2006, 27: 2651-2670.

[6] Wang Xihan, Li Shujun, and Jia Mingtu. Porous Ti-24Nb-4Zr-8Sn alloy for biomedical applications fabricated by space-holder method. The Chinese Journal of Nonferrous Metals, 2010, 20(1): 967-971. (In Chinese)

[7] Zhao J, Lu X and Weng J. Macro-porous Ti-Based Composite Scaffold Prepared by Polymer Impregnating Method with Calcium Phosphate Coatings. Materials Letters, 2008, 62(17): 2921-2924.

[8] Stähli C, Bohner M, Bashoor-Zadeh M, et al. Aqueous Impregnation of Porous β-Tricalcium Phosphate Scaffolds. Acta Biomaterialia, 2010, 6 (7): 2760-2772.

[9] Kaya M, Orhan N, Kurt B, et al. The effect of solution treatment under loading on the microstructure and phase transformation behavior of porous NiTi shape memory alloy fabricated by SHS. J Alloys Compd., 2009, 475: 378-384.

[10] Badiche X, Forest S, and Guibert T. Mechanical properties and non-homogeneous deformation of open-cell nickel foams: application of the mechanics of cellular solids and of porous materials. Materials Science and Engineering, 2000, A289: 276-288.

[11] Ettel V A. Nickel foam. Taipeh Battery Forum, 1999: 5-10.[12] L i Dichen, Tian Xiaoyong, and Wang Yongxin, et a l .

Development of Additive Manufacturing. Electromachining & Mould, 2012, 1: 20-22. (In Chinese)

[13] Lu Bingheng and Li Dichen. Development of the Additive Manufacturing (3D printing) Technology. Machine Building & Automation, 2013, 42(4): 1-4. (In Chinese)

[14] Gong Shuili, Suo Jiangbo, and Li Huaixue. Development and Application of Metal Additive Manufacturing Technology. Aeronautical Manufacturing Technology, 2013, 13: 66-71. (In Chinese)

[15] Shi Yusheng, Lu Zhongliang, and Zhang Wenxian. The Technology and Equipment of Selective Laser Melting. China Surface Engineering, 2006, 19(5): 150-153. (In Chinese)

[16] Http://www.arcam.com.[17] Han Jiandong, Lin Feng, Qi Haibo. Effects of powder preheating

in electron beam selective melting process. Transactions of the China Welding Institution, 2008, 29(10): 77-81.

[18] Qi Haibo, Yang Minghui, and Qi Fangjuan. Numerical Simulation of Effects of Scanning Path on Electron Beam Selective Melting Process of TC4. Transactions of the China Welding Institution, 2009, 30(08): 5-8. (In Chinese)

[19] Yan Yongnian, Qi Haibo, and Lin Feng. Produced Three-Dimensional Metal Parts By Electron Beam Selective Melting. Chinese Journal of Mechanical Engineering, 2007, 43(06): 87- 92. (In Chinese)

[20] Bo Lin, Zhao Zhiguo, and Gong Haibo. Development of Laser Forming Technology for Titanium Alloy Aircraft Structure. Aeronautical Manufacturing Technology, 2013 (11): 40-44

[21] Jia Wenpeng, Tang HuiPing, and He Weiwei. The Development and Key Problems of Electr ic Beam Rapid Forming. Electromachining & Mould, 2010(2): 41-44.

[22] Yao Nina and Peng Xionghou. The Preparation Method of Metal Powder for 3D Printing. Sichuan Nonferrous, 2013, 12(4): 48-51.

[23] Li Xiang, Wang Chengtao, and Zhang Wenguang. Fabrication of Porous Ti6Al4V Implant Using Electron Beam Melting and Mechanical Properties. Journal of Shanghai Jiao Tong University, 2009, 43(12): 1946 - 1949. (In Chinese)

[24] Wang Zhiyang, Shen Yifu, and Gu Dongdong. Preparation of Honeycomb-like Porous 316L Stainless Steel by Selective Laser Melting. Rare Metal Materials and Engineering, 2011, 40(1): 161 - 164. (In Chinese)

[25] He Weiwei, Jia Wenpeng, and Liu Haiyan. The Application of Rapid Prototyping Technology in Porous Metals Materials. Rare Metals Letters, 2008, 27(8): 1-6. (In Chinese)

[26] Http://www.quizlet.com.[27] Mont M A and Hungerford D S. Proximally Coated Ingrowth

Prostheses: A Review. Clinical Orthopaedics and Related Research, 1997, 344: 139-149.

[28] Murr L E, Gaytan S M, and Ramirez D A. Metal fabrication by

Page 10: Additive manufacturing technologies of porous metal Additive manufacturing technologies of porous metal implants * Yu Bo Ph.D., ... or sintering, could increase the roughness and surface

331

Vol.11 No.4 July 2014Special Report CHINA FOUNDRYCelebrating the 10th Anniversray 2004-2014

additive manufacturing using laser and electron beam melting technologies. J Mater Sci Techn, 2012, 28(1): 1-44.

[29] Traini T, Mangano C, and Sammons R I. Direct Laser metal sintering as a new approach to fabrication of isoelastic functionally graded material for manufacture of porous titanium dental implants. Dental Materials, 2008, 24: 1525-1533.

[30] Lewis M, Robin C S, and Wesley K B. Selective Laser Melting: A Regular Unit Cell Approach for the Manufacture of Porous, Titanium, Bone In-Growth Constructs, Suitable for Orthopedic Applications. Journal of Biomedical Materials Research Part B, Applied Biomaterials, 2008, 11: 325-334.

[31] Stamp R, Fox P, and O'Neill W. The development of a scanning strategy for the manufacture of porous biomaterials by selective laser melting. J Mater Sci Mater Med, 2009, 20(9): 1839-1848.

[32] Heinl P, Muller L, and Korner C. Cellular Ti-6Al-4V structures with interconnected macro porosity for bone implants fabricated by selective electron beam melting. Acta Biomaterialia, 2008, 4: 1536 -1544.

[33] Parthasarathy J. and Starly B. Mechanical Evaluation of Porous Titanium (Ti-6Al-4V) Structures with Electron Beam Melting (EBM). Journal of the Mechanical Behavior of Biomedical Materials, 2010: 249-259.

[34] Parthasarathy J, Starly B and Raman S. The Design for Additive Manufacture of Functionally Graded Porous Structures with Tailored Mechanical Properties for Biomedical Applications. Journal of Manufacturing Processes, 2011: 1-11.

[35] Xiao Dongming. Modeling of Porous Structure of Implants and Direct Manufacturing by Selective Laser Melting. Dissertation, South China University of Technology, 2013.

[36] Li Xiang, Yu Xiaoming, and Wang Chengtao. Preparation and Characterization of Porous Titanium Alloy Scaffold Coated with Tantalun. Rare Metal Materials and Engineering, 2012, 41(11):

2049 - 2052. (In Chinese)[37] Zardiackas L D, Parsell D E, and Dillon L D. Wear evaluation of

cobalt-chromium alloy for use in a metal-on-metal hip prosthesis. J Biomed Mater Res B: Appl Biomater, 2001, 58: 180-188.

[38] Shimko D A, Shimko V F, and Sander E A. Effect of porosity on the fluid flow characteristics and mechanical properties of tantalum scaffolds. J Biomed Mater Res B: Appl Biomater, 2005, 73: 315-324.

[39] Sevilla P, Aparicio C, and Planell J A. Comparison of the mechanical properties between tantalum and nickel-titanium foams implant materials for bone ingrowth applications. Journal of Alloys and Compounds, 2007, 439: 67-73.

[40] Garrett R, Abhay P, and Dimitrios P A. Fabrication methods of porous metals for use in orthopaedic applications. Biomaterials, 2006, 27: 2651-2670.

[41] Klawitter J J and Hulbert S F. Application of porous ceramics for the attachment of load bearing orthopaedic application. J Biomed Mater Res, 1971, (5): 161-166.

[42] Palmquist A, Snis A, Emanuelsson L, et al. Long-term biocompatibility and osseo-integration of electron beam melted, free-form-fabricated solid and porous titanium alloy: Experimental studies in sheep. Journal of Biomaterials Applications, 2011, 27(8): 1003-1016.

[43] Song Yunjing, Li Musen, and Wen Shulin. New Advances in Research on Fabrication of Bioactive Ceramic Coating by Biomimetic Method. Materials Review, 2002, 16(5): 33-35. (In Chinese)

[44] Su Xubin, Yang Yongqiang, and Yu Peng. Development of porous medical implant scaffolds via laser addit ive manufacturing. Transactions of Nonferrous Metals Society of China, 2012, 22: 181-187.