Features• Enhancedmonitorintegrationcapabilitieswithour
Cardiocap*/5andDash*seriesofmonitors
• Largecolorventilatordisplaywithcolorwaveformandalarmmessageindicator
• Lightweightandcompactforeasymaneuverability
• OptionalintegratedauxiliaryO2flowmeterandsuctioncontrol
Advanced Breathing System (ABS)• Onestepbag/ventswitchturnstheventilatoron/off
• Minimalnumberofpartsandtubeconnectionsmayhelptoreducethepotentialforleaksandmisconnects
• Easeofdisassembly(notools)
• Autoclavableandlatex-free
7100 Ventilator• VolumeandPressuremodeswithelectronicPEEP
• Exhaledvolume,airwaypressureandinspiredoxygenmonitoringcapabilities
• Directaccesstoventilatorparametersettings
• Smartalarmsdirectusertospecificproblemsandaffectedparameters
• Pressurebargraphforvisualreferenceonabreath-by-breathbasis(optionalpressurewaveformavailable)
Aespire*shownwithCardiocap*/5monitorandTec*7Vaporizers
Aespire7100ExceptionalperformanceCompactdesign
Improved low flow/reduced life cycle costs• Onlyonescheduledmaintenancecheckperyear
• Freshgasflowcompensation–automatically(availablewithtidalvolumecompensationoption)
• MinimumO2flowof50mL
• Dualairflowtubesstandardforhigherresolutionoflowflows
GEHealthcare
Physical specifications
Dimensions
Height: 134.5cm/52.9in
Width: 72cm/28.3in
Depth: 73cm/28.7in
Weight: Approximately108kg/238lbs
Top shelf
Weightlimit: 34kg/75lbs
Width: 66cm/26in
Depth: 40cm/15.75in
Work surface
Height: 81.7cm/32.2in
Size: 2160cm2/334in2
DIN rail
Sideofmachine: 34.5cm/13.6in
Drawers (internal dimensions)
Height: 17.5cm/6.9in
Width: 33cm/13in
Depth: 26.5cm/10.4in
Absorber bag arm (optional)
Armlength: 30.5cm/12in
Bagarmheight(adjustable): 87cm/34.3in 104cm/40.9in
Casters
Diameter: 12.5cm/5in
Brakes: Individuallocking frontcasters
Ventilator operating specifications
Positive End Expiratory Pressure (PEEP)
Type: Integrated,electronicallycontrolled
Range: OFF,4to30cmH2O (incrementsof1cmH2O)
Ventilator monitored values
Tidalvolume: 5to1500mL,1mLresolution
Minutevolume: 0to99.9L/min,0.1L/minresolution
Breathingrate: 0to65breathsperminute, 1breathperminuteresolution
Oxygenpercentage: 5%to110%,1%resolution
Airwaypressure:–9to99cmH2O,1cmH2Oresolution
Alarm settings
Tidalvolume(VTE): Low:OFF,5to1500mL High:20to1600mL,OFF
Minutevolume(VE): Low:OFF,0.1to10L/min High:0.5to30L/min,OFF
Inspiredoxygen (FiO2): Low:18to100% High:21to100%,OFF
Apneaalarm: MechanicalventilationON:<5mL breathmeasuredin30seconds
MechanicalventilationOFF:<25mL breathmeasuredin30seconds
Lowairwaypressure: Changeof<4cmH2OabovePEEP
Pressure(Plimit)range: 12to99cmH2O (incrementsof1cmH2O)
Sustainedairwaypressure:6to30cmH2O+PEEP (adjustedbasedonventilatorsettings)
Subatmosphericpressure: Paw<–10cmH2O
Alarmsilencecountdowntimer: 120to0seconds
Modes of ventilation
VolumeControlmode Withtidalvolumecompensation(optional)
Pressuremode(optional)
Ventilation parameters
Tidalvolumerange: 45to1500mL (VolumeControlmode)
Incrementalsettings: 45to100mL (incrementsof5mL)
100to300mL (incrementsof10mL)
300to1000mL (incrementsof25mL)
1000to1500mL (incrementsof50mL)
Pressure (PInspired)range: 5to50cmH2O (incrementsof1 cmH2O)
5to1500mL volumedelivery
Rate: 4to65breathsper minute(increments of1breathperminute)
Inspiratory/expiratoryratio: 2:1to1:6 (incrementsof0.5)
Inspiratorypauseadjust: 5%to60%of inspiratorytime (incrementsof5%)
Colorventilatorydisplay
Ventilator accuracy
Delivery/monitoring accuracy
Volumedelivery: >200mL=betterthan±10% SetTV
75to200mL=betterthan±20mL
<75mL=betterthan±15mL
Pressure(PInspired)deliveryrepeatability: ±2cmH2O
PEEPdeliveryrepeatability: ±2cmH2O
Volumemonitoring: >200mL=betterthan±10%
75to200mL=betterthan±20mL
<75mL=betterthan±15mL
Pressuremonitoring: Betterthan±2cmH2Oand±5% ofreading(whicheverisgreater)
Ventilator components
Flow transducer
Type: Variableorificeflowsensor
Dimensions: 22mmODand15mmID
Location: Inspiratoryoutletand expiratoryinlet
(Optionalautoclavablesensoravailable)
Oxygen sensor
Type: Galvanicfuelcell
Ventilator pneumatics
Pressurerangeatinlet: 240kPato700kPa/35psigto100psig
Peakgasflow: 70L/min+freshgasflow
Flowrange: 2to70L/min
Flowcompensationrange: 200mL/minto15L/min
Ventilator screen
Displaysize: 120mmx92mm
Displaydensity: 1/4colorVGA
Battery back-up
Backuppower: Demonstratedbatterytime undertypicaloperating conditionsis90+minutes whenfullycharged.Battery timeunderextremeconditions is30minutes.
Batterytype: Internalrechargeablesealed leadacid
Communication port
Serialinterface: IsolatedRS-232C compatibleport
Anesthetic agent delivery
Delivery
Vaporizers: Tec*5,Tec6Plus,Tec7
Numberofpositions: 2
Mounting: Tool-freeinstallation Selectatec*manifold interlocksandisolates vaporizers
Tec6Plusvaporizer Tec7vaporizer
Electrical specifications
Current leakage
100/120V: <300µA
Power
Powerinput: 100-120Vac,50/60Hz
Powercord: Length:5m/16.4ft Rating:15A@120Vac
Inlet/outlet modules
Systemcircuit breakers: 15A
Outlets 4outletsonback,3-2A,1-3Aindividual(optional): breakers,optionalisolationtransformer
Pneumatic specifications
Auxiliary common gas outlet
Connector: ISO22mmODand15mmID
Gas supply
Pipelineinputrange: 240kPato600kPa/35psigto88psig
Cylinderinput: Pinindexedinaccordancewith CGA-V-1orDIN(nutandgland); containsinputfilterandcheckvalve
Note:Maximum3cylinders; twoinboardmounted,one outboardmounted.
Primaryregulatordiaphragmminimumburstpressure: 2758kPa/400psig
Primaryregulatornominaloutput: <338kPa/49psig Pinindexedcylinder connections
O2 controls
Method: Proportionatedecreaseof N2OwithreductioninO2pressure
Supplyfailurealarm: Range:193kPato221kPa/ 28psigto32psig
Soundsatmaximumvolume every10seconds
O2flush: Range:25to75L/min
Flowmeters
O2ranges: 0.05to0.95L/min and1.0to15.0L/min; MinimumO2flow: 50mL/min±25mL
N2Oranges: 0to0.95L/minand 1.0to10.0L/min
Airrange: 0to0.95and1to15L/min
Pneumatic specifications, continuedCalibration: Percentof Accuracy fullscaleflow (%offlowrate)
100 ±2.5% 90 ±2.5% 80 ±2.6% 70 ±2.7% 60 ±2.9% 50 ±3.1% 40 ±3.4% 30 ±4.0% 20 ±5.0% 10 ±8.1%
Calibrationconditions:* 20°C/68°F,101.3kPa/760mmHg
* Differentbreathingcircuitpressures,barometricpressures ortemperatureschangeflowtubeaccuracy.
Hypoxic guard system
Type: MechanicalLink-25*
Range: Providesanominalminimum25% concentrationofoxygenin O2/N2Omixture
Materials
Allmaterialsincontactwithpatientbreathinggasesarefreeofnaturalrubberlatex.
Environmental specifications
System operation
Temperature: 10°to40°C/50°to104°F
Humidity: 15to95%relativehumidity (non-condensing)perIEC68-2-3
Altitude: –440to3565m/500to800mmHg
System storage
Temperature: –15°to50°C/–5°to122°F
Humidity: 10to95%relativehumidity (includingcondensing)perIEC68-2-3
Altitude: –440to5860m/375to800mmHg
Oxygencellstorage: –15°to50°C/5°to122°F 1 0to95%relativehumidity 500to800mmHg
Electromagnetic compatibility
Immunity: Complieswithallrequirements ofEN60601-1-2
Emissions: CISPR11group1classB
Approvals: UL2601-1, CSAC22.2#601.1 EN/IEC60601-1 CE0197
Breathing circuit specifications
Operational modes
Breathingcircuitiscirclemodeonly
Carbon dioxide absorbent canister
Absorbentcapacity: 800g
Integratedexpiratorylimbwaterreservoír
Ports and connectors
Exhalation: 22mmODISO15mm IDtaper
Inhalation: 22mmODISO15mm IDtaper
Bagport: 22mmOD
Pressure gauge
Scalerange: 0to10kPa/ –20to100cmH20
Bag-to-Ventilator switch
Type: Bi-stable
Control: Controlsventilatorand directionofbreathinggas withinthecircuit
Integrated Adjustable Pressure Limiting (APL) valve
Range: 0.8to70cmH2O
Tactileknob indicationat: 30cmH2Oandabove
Adjustmentrangeofrotation: 0.8to30cmH2O(0to230°) 30to70cmH2O(230to330°)
Materials
Allmaterialsincontactwithexhaledpatientgasesareautoclavable,exceptdisposableflowsensorsandO2cell.(Autoclavableflowsensorsoptional).
Allmaterialsincontactwithpatientgasarefreeofnaturalrubberlatex.
Breathing circuit parameters
Compliance: Bagmode: 1.82mL/cmH2O
Mechanical mode: Automaticallycompensatesforcompression losseswithintheabsorberandbellows assembly
Circuitvolume: 2.7LVentMode 1.2LBagMode
Expiratoryresistance: PexpBagMode PexpVentMode Flowrate Pressuredrop Pressuredrop
10L/min 0.78cmH2O 0.77cmH2O
30L/min 1.59cmH2O 1.71cmH2O
60L/min 3.48cmH2O 3.88cmH2O
Note:Withpatientcircuitandwyepieceadd+0.89cmH2O
Anesthetic gas scavenging
Type Hospitalsystemrequired Machineconnection
Activelowflow: Highvacuum36L/min DISSevac (300mmHg)@12inHg
Passive: Passiveorexternally 30mm/0.5in attachedactivesystem MISOtaper
Activeadjustableflow: >30L/mi
©2011GeneralElectricCompany–AllRightsReserved.
GeneralElectricCompanyreservestherighttomakechangesinspecificationandfeaturesshownherein,ordiscontinuetheproductdescribedatanytimewithoutnoticeorobligation.ContactyourGErepresentativeforthemostcurrentinformation.
GEandGEMonogramaretrademarksofGeneralElectricCompany.
*Aespire,Cardiocap,Dash,SelectatecandTecaretrademarksofGeneralElectricCompany.
Datex-Ohmeda,Inc.,aGeneralElectriccompany,doingbusinessasGEHealthcare.
AN-0237-08.11-EN-US DOC0000000
About GE Healthcare GEHealthcareprovidestransformationalmedicaltechnologiesandservicesthatareshapinganewageofpatientcare.Ourbroadexpertiseinmedicalimagingandinformationtechnologies,medicaldiagnostics,patientmonitoringsystems,drugdiscovery,biopharmaceuticalmanufacturingtechnologies,performanceimprovementandperformancesolutionsserviceshelpourcustomerstodeliverbettercaretomorepeoplearoundtheworldatalowercost.Inaddition,wepartnerwithhealthcareleaders,strivingtoleveragetheglobalpolicychangenecessarytoimplementasuccessfulshifttosustainablehealthcaresystems.
Our“healthymagination”visionforthefutureinvitestheworldtojoinusonourjourneyaswecontinuouslydevelopinnovationsfocusedonreducingcosts,increasingaccessandimprovingqualityaroundtheworld.HeadquarteredintheUnitedKingdom,GEHealthcareisaunitofGeneralElectricCompany(NYSE:GE).Worldwide,GEHealthcareemployeesarecommittedtoservinghealthcareprofessionalsandtheirpatientsinmorethan100countries.FormoreinformationaboutGEHealthcare,visitourwebsiteatwww.gehealthcare.com.
GEHealthcareP.O.Box7550Madison,WI53707-7550U.S.A.
www.gehealthcare.com