JULY 2019 Met One Instruments, Inc., Phone: 541.471.7111 Web: www.metone.com, Sales: [email protected] Service: [email protected] AIR QUALITY PARTICULATE MONITORS Met One Offers Four BAM Models PAGE 1 OF 7 Beta Attenuation Mass Monitors “BAMs” are regarded as the standard for continuous ambient particulate mass measurement in air quality monitoring because of their simplicity, reliability, and unsurpassed accuracy. Almost all large air quality monitoring networks worldwide use BAMs for this reason. In addition, because of their uncompromising accuracy and ease of use BAMs are used in the following applications: Beta Attenuation Beta attenuation is the most widely used method worldwide for regulatory monitoring of particulate matter. Over the past decade BAMS have largely supplanted other methods such as the “TEOM” because they are generally more reliable, less complex, and less expensive to operate. BAM-1020 BETA ATTENUATION MASS MONITOR E-BAM PLUS PORTABLE BETA GAUGE E-BAM PORTABLE BETA GAUGE BAM-1022 PORTABLE CONTINUOUS BETA ATTENUATION MASS MONITOR more reliable, less complex, and E-BAM OR particu unsurp reason follow Bet Beta a matter becau E-BAM PLUS

Upload: others

Post on 05-Aug-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: AIR QUALITY Met One Offers Four BAM Models PARTICULATE ... · The BAM 1020 beta attenuation mass monitor is the defacto US standard for continuous PM monitoring. It wasfirst introduced

JULY 2019Met One Instruments, Inc., Phone: 541.471.7111 Web: www.metone.com, Sales: [email protected] Service: [email protected]

AIR QUALITYPARTICULATE MONITORS Met One Offers Four BAM Models PAGE 1 OF 7

Beta Attenuation Mass Monitors “BAMs” are regarded as the standard for continuous ambient

particulate mass measurement in air quality monitoring because of their simplicity, reliability, and

unsurpassed accuracy. Almost all large air quality monitoring networks worldwide use BAMs for this

reason. In addition, because of their uncompromising accuracy and ease of use BAMs are used in the

following applications:

Beta AttenuationBeta attenuation is the most widely used method worldwide for regulatory monitoring of particulate

matter. Over the past decade BAMS have largely supplanted other methods such as the “TEOM”

because they are generally more reliable, less complex, and less expensive to operate.

BAM-1020 BETA ATTENUATION

MASS MONITOR

E-BAM PLUS

PORTABLE BETA GAUGE

E-BAM

PORTABLE BETA GAUGE

BAM-1022 PORTABLE

CONTINUOUS BETA

ATTENUATION MASS MONITOR

more reliable, less complex, and

E-BAM

OR

particu

unsurp

reason

follow

BetBeta a

matter

becau

E-BAM PLUS

Page 2: AIR QUALITY Met One Offers Four BAM Models PARTICULATE ... · The BAM 1020 beta attenuation mass monitor is the defacto US standard for continuous PM monitoring. It wasfirst introduced

JULY 2019Met One Instruments, Inc., Phone: 541.471.7111 Web: www.metone.com, Sales: [email protected] Service: [email protected]

PAGE 2 OF 7Beta Attenuation Mass Monitors

BAM 1020• US-EPA Designated PM10, PM2.5, PM10-2.5

• Certified in China, Korea, EU, Russia

• Rack mounted or bench top

• 3rd generation

• Time-proven technology

• Gravimetrically-traceable calibration

• Cloud modem compatible

• Air quality monitoring networks

• Roadside monitoring

The BAM 1020 beta attenuation mass monitor is the defacto US standard for continuous PM

monitoring. It wasfirst introduced in 1995, received EPA designation for PM10 in 1998, EPA designation

for PM2.5 in 2008 (almost 2 years ahead of the competition) and EPA designation for PM10-2.5 in 2009.

It possesses more than a dozen international certifications including CNEMC (China), TUV (EU), Korea,

Taiwan, and Russia.

BAM 1022• US-EPA Designated PM2.5

• Lightweight, portable

• Self-contained: no shelter/enclosure necessary

• Simultaneous 1-hour and real-time output

• Time-proven technology

• Gravimetrically-traceable calibration

• Cloud modem compatible

• Air quality monitoring networks

• Roadside monitoring

• Emergency responder applications

• Community monitoring

• Ideal field audit device

The BAM-1022 represents the future of regulatory air quality monitoring. It is self-contained; no

shelter or enclosure is required. This relieves the user of having to appropriate the space, the power,

the expense, and the hassle of having to set up an environmentally controlled shelter on-site in order

to perform a simple PM measurement. Unlike most air quality monitors, which are generally set up and

perform their measurements at or near room temperature inside of a shelter or enclosure, the

BAM-1022 performs its measurement under near ambient conditions. This approach can improve the

accuracy of the measurement

Page 3: AIR QUALITY Met One Offers Four BAM Models PARTICULATE ... · The BAM 1020 beta attenuation mass monitor is the defacto US standard for continuous PM monitoring. It wasfirst introduced

JULY 2019Met One Instruments, Inc., Phone: 541.471.7111 Web: www.metone.com, Sales: [email protected] Service: [email protected]

E BAM PLUS• US-EPA Designated PM10

• Lightweight, portable

• Self-contained: no shelter/

enclosure necessary

• Simultaneous 1-hour and real-time output

• Time-proven technology

E BAM• Lightweight, portable

• Easily operated from battery or solar power

• Self-contained: no shelter/enclosure necessary

• Simultaneous 1-hour and real-time output

• Time-proven technology

• Gravimetrically-traceable calibration

• Real-time output: 1-minute time resolution

• Roadside monitoring

• Emergency responder applications

• Community monitoring

BAM-1020 BAM-1022Parameter E-BAM PLUS E-BAM

Regulatory Designation US-EPA PM2.5 Equivalant Method US-EPA PM10 Equivalant MethodUS-EPA PM10-2.5 Equivalant MethodTUV, CNEMCKorea, Other InternationalCerifications

US-EPA PM2.5 Equivalant Method

US-EPA PM10 Equivalant Method

China Patter Approvel

Installation Rack or Bench Mount Portable, Self-Contained Portable, Self-Contained Portable, Self-Contained

Measurement Cycle 1-Hour Dual Oututs:1-Hour, Continuous (1-Minute Minimum Resolution)

Dual Oututs:1-Hour, Continuous (1-Minute Minimum Resolution)

Dual Oututs:1-Hour, Continuous (1-Minute Minimum Resolution)

Lower Limit of Detection 4.8 μg/m3 (1-Hour)1 μg/m3 (24-Hour)

4.8 μg/m3 (1-Hour)1 μg/m3 (24-Hour)

4.8 μg/m3 (1-Hour)1 μg/m3 (24-Hour)

6 μg/m3 (1-Hour)1.2 μg/m3 (24-Hour)

Pump Type AC-Powered Rotary Vane or AC-Powered Reciprocating

AC-Powered Rotary Vane or AC-Powered Reciprocating

AC-Powered Rotary Vane or AC-Powered Reciprocating

DC-Powered Internal Diaphragm

Communications RS-232, Cellular Modem, Ethernet,USB

RS-232, USB, Ethernet, Cellular Modem, Modbus

RS-232, Cellular Modem, USB, Modbus

RS-232, Cellular Modem, USB, Modbus

Power Requirements 115-240 VAC 5.4 Peak Amps(Shelter Not Included)

100-230 VAC 50/60 Hz, 300 W

100-230 VAC 50/60 Hz 150 W

12VDC AC Power Supply:100-230 VAC, 50/60 Hz, 102W

Data Storage 182 Days Expandable 22,528 Records(2.6 Years @ 1 Rec./Hour)

1.3 Years @ 60 Minute Average

2.6 Years @ 60 Minute Average

• Gravimetrically-traceable calibration

• Cloud modem compatible

• Air quality monitoring networks

• Roadside monitoring

• Emergency responder applications

• Community monitoring

• Ideal field-audit device

Lightweight, Portable PAGE 3 OF 7

Page 4: AIR QUALITY Met One Offers Four BAM Models PARTICULATE ... · The BAM 1020 beta attenuation mass monitor is the defacto US standard for continuous PM monitoring. It wasfirst introduced

JULY 2019Met One Instruments, Inc., Phone: 541.471.7111 Web: www.metone.com, Sales: [email protected] Service: [email protected]

Trustworthy ResultsBeta gauges provide unparalleled accuracy in measuring PM worldwide. Beta gauges, unlike other

methods such as optical techniques, are sensitive only to PM mass and not to unmeasurable and

unknowable factors such as the chemical composition and physical nature of the aerosol being

measured. Optical methods for PM measurement are impacted by these factors and since they cannot

be known ahead of time, they must make assumptions, which are not always correct. This is

especially true during extraordinary events such as wildfires, structure fires, industrial accidents or

other instances where there is a heightened need to accurately measure PM concentrations.

7.15.12 0:00

7.20.12 0:00

6.25.12 0:00

7.10.12 0:00

7.5.12 0:00

6.30.12 0:00

80

70

60

50

40

30

20

10

0

Mira Loma CA BAM-1022, BAM-1020 PM2.5

PM

2.5 H

OURL

Y C

ONCE

NTR

ATIO

N n

g/m

3

BAM-1020 A BAM-1020 B BAM-1020 C BAM-1022 A BAM-1022 B BAM-1022 C

PAGE 4 OF 7

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

50

45

40

35

30

25

20

15

10

5

0

Baton Rouge Louisiana BAM-1022 Hourly PM2.5

PM

2.5 H

OURL

Y C

ONCE

NTR

ATIO

N n

g/m

3

BAM-1022 A BAM-1022 B BAM-1022 C

5.8.13 5.8.13 5.8.13 5.8.13 5.8.13 5.8.13

50

45

40

35

30

25

20

15

10

5

0

Baton Rouge Louisiana BAM-1022 Hourly PM2.5

PM

2.5 H

OURL

Y C

ONCE

NTR

ATIO

N n

g/m

3

BAM-1022 A BAM-1022 B BAM-1022 C

Unparalleled Accuracy

7.15.12 0:00 7.20.12 0:00 6.25.12 0:00 7.10.12 0:007.5.12 0:00 6.30.12 0:00

80

70

60

50

40

30

20

10

0

Mira Loma CA BAM-1022, BAM-1020 PM2.5

PM

2.5 H

OURL

Y C

ONCE

NTR

ATIO

N n

g/m

3

BAM-1020 A BAM-1020 B BAM-1020 C BAM-1022 A BAM-1022 B BAM-1022 C

Page 5: AIR QUALITY Met One Offers Four BAM Models PARTICULATE ... · The BAM 1020 beta attenuation mass monitor is the defacto US standard for continuous PM monitoring. It wasfirst introduced

JULY 2019Met One Instruments, Inc., Phone: 541.471.7111 Web: www.metone.com, Sales: [email protected] Service: [email protected]

Reliability and Ease of OperationBAMs are easy to operate and easy to maintain. The equipment is simple, not overly complex, unlike

PM monitors based on the TEOM technology which employ multiple flow splitters, flow controllers,

valves and NAFION dryers. The more complex the measurement, the more there is to go wrong.

Gravimetric CalibrationOur BAMs are calibrated to a gravimetric standard. In other words, when you run one of our BAMs

against a reference sampler, such as our E-SEQ-FRM or our E-FRM you can be confident that the

results will match. Light-based “spectrometers” can make no such claim as there is direct traceability

to mass. Instead, they estimate the chemical aerosol composition of the air and estimate a

conversion factor between light signal and particle mass. When the aerosol composition becomes

something unexpected, such as might occur during forest fires, industrial accidents or perhaps even

when the prevailing wind direction changes, they can be spectacularly wrong – not just slightly off.

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

5.8.13

0.3500

0.3000

0.2500

0.2000

0.1500

0.1000

0.0500

0.0000

Phoenix Arizona BAM-1020 PM10Hourly

PM

10 H

OURL

Y C

ONCE

NTR

ATIO

N n

g/m

3

BAM-1020 A BAM-1020 B BAM-1020 C

7.9.11

7.1.11

7.7.11

7.5.11

7.3.11

7.11.11

700.0

600.0

500.0

400.0

300.0

200.0

100.0

0.0

Hamilton Ontario Collocated E-BAM PLUS Monitors

PM

2.5 H

OURL

Y C

ONCE

NTR

ATIO

N n

g/m

3

E-BAM PLUS 1 E-BAM PLUS 2 E-BAM PLUS 3

7.9.11 7.1.11 7.7.11 7.5.11 7.3.11 7.11.11

700.0

600.0

500.0

400.0

300.0

200.0

100.0

0.0

Hamilton Ontario Collocated E-BAM PLUS Monitors

PM

2.5 H

OURL

Y C

ONCE

NTR

ATIO

N n

g/m

3

E-BAM PLUS 1 E-BAM PLUS 2 E-BAM PLUS 3

Not Overly Complex…Simple

5.8.13 5.8.13 5.8.13 5.8.13 5.8.13 5.8.13 5.8.13 5.8.13 5.8.13 5.8.13 5.8.13

0.3500

0.3000

0.2500

0.2000

0.1500

0.1000

0.0500

0.0000

Phoenix Arizona BAM-1020 PM10Hourly

PM

10 H

OURL

Y C

ONCE

NTR

ATIO

N n

g/m

3

BAM-1020 A BAM-1020 B BAM-1020 C

H ilt O t i C ll t d E BAM PLUS M it

PAGE 5 OF 7

Page 6: AIR QUALITY Met One Offers Four BAM Models PARTICULATE ... · The BAM 1020 beta attenuation mass monitor is the defacto US standard for continuous PM monitoring. It wasfirst introduced

JULY 2019Met One Instruments, Inc., Phone: 541.471.7111 Web: www.metone.com, Sales: [email protected] Service: [email protected]

EMERGENCY RESPONDER APPLICATIONS

PAGE 6 OF 7

Air Quality Monitoring NetworksThe vast majority of PM monitors used in governmental air

quality monitoring networks are BAMs. PM monitors used

in these networks generally require government

certification for PM10 or PM2.5. The BAM-1020, BAM-1022

and E-BAM PLUS monitors are government certified for

operation in such networks.

Emergency Responder ApplicationsForest fires, structure fires, or industrial accidents can lead to the release of massive amounts of

particulate matter into the air thereby imperiling the health of local residents and emergency responders.

Because of the unplanned nature of such events, it is often impossible to set up air quality monitoring

equipment in order to accurately assess health

threats to the surrounding community. The

E-BAM is the only portable beta gauge available

that may be easily and economically operated on

battery or solar power. If EPA-designated

measurements are required, then either the

BAM-1022 or the E-BAM PLUS may be quickly

and easily deployed.

GOVERNMENT AIR QUALITY

MONITORING NETWORKS

Government Certified

Page 7: AIR QUALITY Met One Offers Four BAM Models PARTICULATE ... · The BAM 1020 beta attenuation mass monitor is the defacto US standard for continuous PM monitoring. It wasfirst introduced

JULY 2019Met One Instruments, Inc., Phone: 541.471.7111 Web: www.metone.com, Sales: [email protected] Service: [email protected]

ROADSIDE MONITORING

COMMUNITY AND FENCE-LINE MONITORING

PAGE 7 OF 7

Community and Fence-line MonitoringCommunities in close proximity to stationary

emission sources such as petroleum refineries,

petrochemical plants, waste incineration

facilities, or ports may be subjected to pollutant

levels higher than is seen at air quality

monitoring stations located in the surrounding

areas. A portable carbon monitor could be

useful in identifying fugitive emissions and

easily identifying local sources of pollution.

Roadside MonitoringOften there is an interest in monitoring

mobile-source emissions originating from motor

vehicle exhaust in close proximity to heavily

traveled roadways. For these applications it is

generally important to measure only BC which

suggests multi-wavelength carbon monitors might

not be cost effective. In addition, at such sites often

external power is limited thereby making the

deployment of air quality monitors requiring an

environmental shelter impractical. For these

applications, the BC-1060 could track in real-time

BC levels in a convenient, cost effective manner.

Identify Fugitive Emissions and Local Sources