all qiba phantoms

3
All QIBA PHANTOMS •C52 Transducer held in ring stand – verasonics system •Phantom on turntable – 8 rotation angles •7 transducer locations •3 focal depths (40, 60, 80 mm) •SWS estimates: –Bandpass filter 100-1000 Hz –Ransac algorithm – fov ~ 2 cmx2cm –Error bars mean +/- 1 std. dev from 56 measures

Upload: hayes-fuentes

Post on 30-Dec-2015

32 views

Category:

Documents


4 download

DESCRIPTION

All QIBA PHANTOMS. C52 Transducer held in ring stand – verasonics system Phantom on turntable – 8 rotation angles 7 transducer locations 3 focal depths (40, 60, 80 mm) SWS estimates: Bandpass filter 100-1000 Hz Ransac algorithm – fov ~ 2 cmx2cm - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: All QIBA PHANTOMS

All QIBA PHANTOMS

• C52 Transducer held in ring stand – verasonics system

• Phantom on turntable – 8 rotation angles• 7 transducer locations • 3 focal depths (40, 60, 80 mm)• SWS estimates:

–Bandpass filter 100-1000 Hz–Ransac algorithm – fov ~ 2 cmx2cm–Error bars mean +/- 1 std. dev from 56 measures

Page 2: All QIBA PHANTOMS

Results

1786-X

1787-X

40

6080

Page 3: All QIBA PHANTOMS

Sim Results – Lateral Regression Window Location

Window positionsBlue 0-12 mmRed 1-12Green 2-12Black 4-12

•Conclusion – you must be outside the ROE (near field of shear wave) to get consistent speeds •Also- depth dependence here was within numerical error???