ADVANCES IN MICRO GAS CHROMATOGRAPHY FOR FIXED GASES AND LIGHT HYDROCARBON ANALYSIS ON A MICRO GC FUSION

Introduction

• Presented by Debbie Alcorn – Tuesday January 27, 2015

• History of the Micro GC

• 3000 Micro GC

• The Next Generation – Micro GC Fusion

• Micro GC Fusion Features

• Advances in Applications

• Future Developments

2 INFICON Advances in Micro Gas Chromatography

INFICON Global Presence

900 employees; offices in 17 countries

3 INFICON Advances in Micro Gas Chromatography

Micro GC History

• The 3000 Micro Gas Chromatograph (GC) has over ten years of proven

excellent analytical performance

• Micro GC Fusion® expands the capability of the 3000 Micro GC

4 INFICON Advances in Micro Gas Chromatography 03/17/14

2001

3000 Micro GC

Introduced

2013

Micro GC Fusion

Introduced

2010

INFICON Acquired

3000 Micro GC

3000 Micro GC

• Microelectromechanical systems (MEMS) based GC

• Isothermally heated capillary columns

• Micro Thermal Conductivity Detector (µTCD)

• Modular

5 INFICON Advances in Micro Gas Chromatography

3000 Micro GC Applications

The Next Generation – Micro GC Fusion

• Micro GC Fusion builds on proven 3000

Micro GC technology • Utilizes the same MEMS injectors and TCD

• Traditional Micro GC technology is limited

due to: • Isothermal column temperature, leading to potential

carryover issues

• Ambient temperature effects

• Software and operating system compatibility

• Micro GC Fusion addresses these issues

using new technology

7 INFICON Advances in Micro Gas Chromatography

Micro GC Fusion Features

• Temperature programmable columns

• Plug and play modules

• Integrated touch panel display

• Web-based user interface

• Wi-Fi capability

• Integrated sample conditioner

8 INFICON Advances in Micro Gas Chromatography

Temperature Programming

• Fast temperature ramping based on resistive column heating technology

• Ability to ramp up to 300 ºC/min (5 ºC/sec)

• Column length dependent

• Ramping profile is independently optimized for each module

• Compared to one oven for multiple columns in a bench top GC

• Cool down time is optimized to achieve short cycle times

• 1-2 minutes, ramp and column length dependent

9 INFICON Advances in Micro Gas Chromatography

Temperature Programming

• Benefits • Faster analysis of later eluting

compounds

• Improved peak shape and

height

• Improves sensitivity

• Rapid column cleaning

Specifications

Ramp rate Column length dependent, up to 300 ºC/min (5 ºC/sec)

Maximum Temperature Column phase dependent, 190°C to 300°C

Cooldown Column and method dependent, 1 to 2 minutes

10 INFICON Advances in Micro Gas Chromatography

Isothermal vs. Temperature Programming

Isothermal

• Results in broad, late eluting peaks

Temperature programming

• Sharpens peak shape

• Shortens run time

• Improves sensitivity

Example – Propane

• Isothermal ~160 seconds

• Ramping ~50 seconds

• Improved peak shape and height

11 INFICON Advances in Micro Gas Chromatography

Column: RT-Q-Bond

Isothermal column temperature: 55°C

Column: RT-Q-Bond

Temperature Ramp:

50°C(10s)90°C240°C, 2°C/s, 2.2°C/s

Isothermal vs. Temperature Programming

Isothermal – late eluting compounds may interfere with the next run

12

Temperature Programming – elutes undesired peaks much faster to avoid carryover

INFICON Advances in Micro Gas Chromatography

Column: RT-Molsieve 5A

Isothermal column temperature: 100°C

Column: RT-Molsieve 5A

Temperature Ramp: 100°C(10s)200°C(2°C/s)

Plug and Play Modules

• Modules can be swapped out for easy maintenance or application

changes

• No “commissioning” or “decommissioning” modules

GC connector

13 INFICON Advances in Micro Gas Chromatography

User Interface

• Web-based software eliminates operating system

compatibility issues

• License-free

• Works with any platform (Android, Apple, Windows)

• On-board hard drive stores hundreds of thousands of

runs

Internet Explorer

Version 10 and above

Mozilla

Firefox

Google

Chrome

Apple

Safari

14 INFICON Advances in Micro Gas Chromatography

Micro GC Fusion Connectivity

• Micro GC Fusion can connect via Wi-Fi to: • Smart phones

• Tablets

• Computers with wireless cards

• Micro GC Fusion can be operated using the touch panel display to: • Run a method or sequence

• View analysis report

• View system status

• Micro GC Fusion can connect via Wi-Fi or LAN to computers for method development and advanced features

15 INFICON Advances in Micro Gas Chromatography

Integrated Sample Conditioner

• An optional integrated sample conditioner can be configured

• The sample conditioner allows for:

• Reduction of sample pressures of up to 1000 psi down to 5-10 psi

• Filtering of particulates

• Heating the sample to 100 ºC

• A quick connect replaces the standard 1/16” inlet

• Can be controlled via the user interface

• No manual operation necessary

16 INFICON Advances in Micro Gas Chromatography

Advances in Applications

• Temperature programming allows for expanded

application capability • Natural gas analysis on one module

• Extended natural gas analysis on two modules

• C2 analysis improvements

• Mud logging

17 INFICON Advances in Micro Gas Chromatography

Natural Gas on One Module

• A single RT-Q-Bond module can be used for C1 to C8 Plus analysis

Column: RT-Q-Bond

Ramp: 60°C (hold 50s) -> 127°C (hold 10s) -> 220°C (hold 110s)

(2°C/s, 2.5°C/s)

N2

methane

CO2

ethane

propane

isobutane

n-butane

isopentane

n-pentane

hexanes

heptanes octanes plus

18 INFICON Advances in Micro Gas Chromatography

Excellent Repeatability

19 INFICON Advances in Micro Gas Chromatography

Extended Natural Gas

• An Rxi-1ms module can be added to conduct

extended hydrocarbon (C1 – C11) analysis

Column: Rxi-1ms

Ramp: 50°C 130°C280°C (2°C/s, 5°C/s) ~3 ppm

20 INFICON Advances in Micro Gas Chromatography

Extended Natural Gas – C12 Analysis

• Rxi-1ms with temperature programming

• Straight chain hydrocarbons (C6 to C12) from a sample bag

overlaid with an extended natural gas calibration gas from a

cylinder

• Requires heated sample

21 INFICON Advances in Micro Gas Chromatography

C2 Analysis

• A RT-U-Bond module is typically used to

analyze C2 compounds

• Temperature programming allows for

faster run time and improved peak shape

22 INFICON Advances in Micro Gas Chromatography

Acetylene

Ethane

Ethylene

CO2

Column: RT-U-Bond

Isothermal Column Temperature: 50°C

Column: RT-U-Bond

Temperature Ramp:

50°C(30s)120°C, 3°C/s

Mud Logging

• Mud logging is the creation of detailed reports of a borehole on a

well site during active drilling

• Liquid, or “mud”, is used as the drilling fluid to bring gas, formation

fluids, and rock cuttings to the surface for analysis

• A mud log provides key geographical information, including lithology,

rate of penetration, and hydrocarbon content

• The concentration of compounds is critical and must be performed

on a continuous, 24-7 basis

• Two methods were developed:

• C1 to C5 in 30 seconds

• C1 to C8 in 60 seconds

• Still being optimized

23 INFICON Advances in Micro Gas Chromatography

Mud Logging in 30 Seconds

24 INFICON Advances in Micro Gas Chromatography

Column: RT-Q-Bond

Isothermal column temperature: 55°C

Mud Logging in 30 Seconds

25 INFICON Advances in Micro Gas Chromatography

Column: Rxi-1ms

Isothermal column temperature: 90°C

Extended Mud Logging

26 INFICON Advances in Micro Gas Chromatography

Column: RT-Q-Bond

Temperature Ramp:

55°C(15s hold)220°C(10s hold), 7°C/s

Extended Mud Logging

27 INFICON Advances in Micro Gas Chromatography

Column: Rxi-1ms

Temperature Ramp:

75°C(9s hold)220°C(12s hold), 7°C/s

Future Developments

• Four module Micro GC Fusion expands application capability • Fixed gases, light hydrocarbons up to C12, and olefins

• Refinery gas analysis

• Reactor gas analysis

• Air analysis for natural gas

• Available in 2015

• Additional applications:

• Syngas, biogas

• Sulfur analysis - SO2, H2S

• Natural gas odorants - TBM and THT

• Dissolved gas analysis

28 INFICON Advances in Micro Gas Chromatography

Conclusion

• Building on proven MEMS technology, Micro GC Fusion extends the

capability of the previous generation

• Using temperature programming, Micro GC Fusion extends

analytical capability

• Future development will include a four module instrument and

additional applications

29 INFICON Advances in Micro Gas Chromatography

Questions?

30 INFICON Advances in Micro Gas Chromatography

Visit our website www.INFICON.com for more information

31 INFICON Advances in Micro Gas Chromatography