250 SERIES
IN PROCESS TO EXCELLENCE
D U T Iensity llage emperature nterface detector - - -
PORTABLE GAUGING INSTRUMENT
DUTI -254
Digital Gauging Device
DUTI-254
Ullage level
Interface level
Density &Temperature 1
Density &Temperature 3
Density &Temperature 2
Density &Temperature N
process
Specifications250 SERIES www.lemis-process.com
Advantages
Applications
Chemical industry
Pharmaceutical industry
Cosmetic industries
Petroleum industry
Food & Beverages
For more information please visit www.lemis-process.com
Ethanol production
Direct density and level measurement
Record spot density, average per tank
ATEX Hazloc certifcation
Safe operation, low maintenance
At any depths up to 30 meters
They’re economical and easy to operate
Measures highly viscous liquids up to 2000 cP
Rigid construction for heavy duty
Local results storage and Bluetooth
outdoor operation
data transfer
Able to download the measurements to PC; Multifunctional software allows to proceed the
measurements results in user-convenient form;
Compatible for a Windows 98/ME/2000/NT/XP.
Immediately printout the measurements; No need for PC.
Remote data transfer; Useful in field conditions; Software for data processing.
SOFTWARE BLUETOOTH PRINTER POCKET PC
Automatic temperature/viscosity
No sampling required
compensation
and depth of the sensor embedding
Measuring range:0… 3 g/cm³ (0… 3000 kg/m³)Density
Density Standard 0.6… 1.2 g/cm³ (600… 1200 kg/m³)
±0.0003 or g/cm³ ( 0.3 or kg/m³)±0.0005 ± ±0.5Accuracy:
Density
±0.00015 or g/cm³ ( or kg/m³)±0.00025 ±0.15 ±0.25Repeatability:
Density
Resolution:0.0001 g/cm³ (0.1 kg/m³)Density
Temperature 0.01 0.02°F°C ( )
Ambient temperature
Depth of submersion Up to 30 meters (up to 100 ft)
Intrinsically safe: Controller Level block with sensor
ATEX II (2G) EEx ib [ia] IIB T4ATEX II 1G EEx ia IIB T4
Power supply 2 x NiMH 3.6V-1200 mAh
Operating time without charging Appr. 12 hours
Dimensions, weight: ControllerLevel block with sensor
180 x 80 x 40 mm (7.1 x 3.2 x 1.6 in), 0.6 kg (1.3 lb) 420 x 245 x 140 mm, x 5.5 in(16.5 x 9.7 ), 4 kg (8.8 lb)
Temperature compensation Automatic
Data handling
Backlighted LCD displayLocal memory for 998 results with date/time stampedBuild in bluetooth for data transfer to printer or PCOptional Windows - based software
Delivery Delivered in compact carrying case
Supported measuring unitsReferred density: at 15° , 20°C, 60°F; API60; SG60CTables ASTM D1250
Real density: g/cm³, kg/m³, lb/gal, lb/ft³; API; SG
Tape graduation - Metric /English Temperature in °C or °F
-40… +85°C (-40... +185 )°F
Viscosity compensation Automatic
Temperature -40… +85°C (-40... +185 )°FLevel Up to 30 m (up to 100 ft)
±0.1 ±0. °F°C ( 2 )Temperature± m ±2 m ( 0.008”)Ullage/Interface detector
±0.1 ±0.2°F°C ( )Temperature±1 mm ± ( 1/16”)Level
Level 1 mm (1/16’)