Hygienic Process Connections
Michael Mies
Slide 3
Basis of hygienic instrumentation
The instrument should not effect the process.
The instrument should not effect the cleaning.
The process and the cleaning should not effect the instrument.
This basics where used by the EHEDG to develop their test.
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 4
Hygiene
What has an effect:
Material
Surface
Design, inside and housings
Installation
Organisations
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 5
DIN 11851
DIN 11864
SMS
DRD
Varivent
APV
E+H 65/85 mm
Clamp
E+H -“F“
Adapter DN 25
Hygienic process connections
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 7
For Sensors:
Deltapilot S DB 50L
Multicap DC 11/DC21
Micropilot II
Liquiphant FTL 330 L/H; FTL
360/361
Cerabar PMC 135/531
Cerabar S PMC631/731
Deltabar S FMD 633
Promag H
Promass F/M/I
various. temp. sensors
pH-assembly CPA 441, 465
CLS 21, 30, 51, 52
SMARTEC CLD 130, Smartec S
The parts of the dairy process
connection DIN 11851
from left to right:
nut, thread part, conical part , seal
PRO
Widely used
Low cost connection
Quick mounting /dismounting
CONTRA
No hygienic design
No flushmount sensors possible
Dairy process connection DIN 11851
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 10
Magmeter Promag H with DIN 11851
process connection in the dairy
Massflowmeter Promass M with
DIN 11851 process connections at the
milk separator
DIN 11851 - Installation examples
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 12
process connections - Stericom / GEA
PRO
EHEDG conform design
Fix or quick mount version
For inline and insertion sensors
CONTRA
Not widly used yet
More expensive as DIN 11851
Aseptic process connection DIN 11864
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 15
Series 2 - sealing version
A, flange
We also have to differentiate between flange and screwed connection.
For both versions, two series each exist.
Series 1 is the screwed,
series 2 is defined as the flange version.
The sealing is equal for both.
Series 1 - sealing version
A, screwed
Aseptic connection acc. to DIN 11864
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 17
For Sensors:
Deltapilot S DB 50L
Multicap DC 11/DC21
Micropilot II
Liquiphant FTL 330 L/H; FTL 360/361
Cerabar PMC 135/531; Cerabar S
PMC631/731
Deltabar S FMD 633
Promag H
Promass F/M/I
various. temp. sensors
pH-assembly CPA 441, 465
CLS 21, 30, 51, 52, SMARTEC CLD 130,
Smartec S
SMS connection
PRO
Widely used in Scandinavia and
France
Low cost connection
Quick mounting /dismounting
CONTRA
No hygienic design
No flushmount sensors possible
Process connection SMS
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 18
Threat, nut, sealing
Process connection SMS
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 20
For Sensors:
Deltapilot S DB 50L
Liquiphant FTL 330 L
Micropilot II
Cerabar PMC
135/135Z/531
Cerabar S PMC 631/731
Deltabar S FMD 633
Flush mounting DRD-flange Deltapilot S DB50L
with DRD-flange
PRO
Flush mount sensors possible
Common in food industry for
process sensors
Hygienic design
CONTRA
For tanks only
No inline sensor solution
Due to big diameter not suitable for
cones
DRD welding adapter for tanks
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 21
CO2 - head pressure measurement at the fermentation tank in
the brewery with Deltapilot S DRD-flange
Installation exsamples
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 22
For Sensors:
Deltapilot S DB 50L
Liquiphant FTL 330 L
Micropilot II
Cerabar PMC 135/135Z/531//
Cerabar S PMC 631/731
various. temp. sensors (ST
14/44/74/76)
pH-assembly Probfit CPA 441, CPA
465
CLS 51, 52, SMARTEC CLD 130/132
PRO
Can hold two sensors in one
housing
Options for small pipes
Quick connct with clamps
CONTRA
For insertion sensors only
Varivent Inline housing
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 23
The speciality of the inline housing
is that there is one process
connection size for several pipe
sizes. This means:
• Varivent Typ T/L is valid for pipe
diameters of DN 10 and DN 15
• Varivent Typ F is valid for pipe
diameters of DN 25 and DN32
• Varivent Typ N is valid for pipe
diameters from DN 40 up to
DN 162
This means e.g. that a Cerabar
pressure transmitter with Varivent
connection can be mounted in pipe
sizes of DN 40 - DN 162.
E+H sensors are not available for all sizes.
Varivent housing
We have to supply only 3 internal diameters:
31 mm, 50 mm and 68 mm.
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 24
Example: the complete family of Varivent®-process connections
Varivent®-Family
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 25
Pressure measurement in a
pharmaceutical unit with Cerabar S
PMC731, Varivent-housing
Connection of a inductive conductivity
sensor CLS 51 with Varivent-housing
Installation examples
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 27
For Sensors:
Deltapilot S DB 50L
Liquiphant FTL 330 L
Micropilot II
Cerabar PMC 135/135Z/531// Cerabar
S PMC 631/731
various. temp. sensors (TST
14/44/74/76)
pH-assembly Probfit CPA 441, CPA 465
CLS 51, 52, SMARTEC CLD 130/132
Connection of a Cerabar in the APV
Inline housing
PRO
Can hold two sensors in one
housing
Flush mount sensors
CONTRA
For insertion sensors only
No quick connect design
APV Inline housing
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 29
Deltapilot S DB 50L Flush process adapter 65mm for Deltapilot S
PRO
Small dimensions
Flush mount sensors
Quick connect
CONTRA
E+H standard only
For Deltapilot and Cerabar only
sensitive to deformation when welding
Endress+Hauser process adapter 65mm
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 30
Deltapilot S DB 50L Flush process adapter 85mm for Deltapilot S
PRO
Small dimensions
Flush mount sensors
Simple to weld in
CONTRA
E+H standard only
For Deltapilot and Cerabar only
Bigger then 65 mm version
Endress+Hauser process adapter 85mm
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 31
Level measurement at the
fermentation tank in the brewery
with Deltapilot S DB50L
CO2-head pressure measurement at
the fermentation tank with
Deltapilot S DB50L
Endress+Hauser adapter - installation examples
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 32
For Sensors:
Deltapilot S DB 50L
Multicap DC21/11/12
Liquiphant FTL 330 L/H, FTL 360/361
Micropilot II
Cerabar PMC 135/ Cerabar S PMC 631/731
Deltabar S FMD 633
Promag A/ Dosimag A/Promag H
Promass F/M/I
pH-assembly Probfit CPA 465
CLS 15,51,52
SMARTEC CLD 130, Smartec S CLD 132
PRO
Quick connection
Widely used
For inline or insertion sensors
CONTRA
Many standards
No flush mount design
Clamp connection DIN32676 and ISO2852
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 33
1"
4"
3"
2 1/2"
2"
1 1/2"
-0000
-0001
-0002
-0003
-0004
-0005
22,6
35,6
48,6
60,3
72,9
97,6
318786-0000
318786-0003
318786-0005
318786-0007
318786-0009
318786-0011
28,50
28,50
28,50
28,50
28,50
28,50
50,5
50,5
64,0
77,5
91,0
119,0
ISO 2852
ISO 2852
ISO 2852
ISO 2852
ISO 2852
ISO 2852
121,0
147,0
168,0
75,0
92,0
105,0
68,50
68,50
68,50
68,50
68,50
68,50
22,1
34,8
47,5
-
-
-
Var. DN d
Schweissnaht sauber
verschliffen
! y D
X
Zch.-Nr.
Einlegeteil
ISO 2852
C D
C (40)
N6
! y d
s
L
1
Anschluss-Flansch
A Kennz.
A
! y A
nur fuer Var. 0000
! y d
L d1
0001
0002
3
! y d
1
(2/1) Detail X 0, 5 A
S - TIG
geschweisst unter Formiergas
2
B ± 1
Kennzeichnung
(Hoehe: 6 mm)
Zulaessige Variante
0, 2 B
DIN/ISO Clamp connection - Promag H
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 34
Tri-Clamp connection - Promag H
1"
4"
3"
2 1/2"
2"
1 1/2"
-0000
-0001
-0002
-0003
-0004
-0005
22,1
34,8
47,5
60,2
72,9
97,4
28,57
28,57
28,57
28,57
318786-0000
318786-0003
318786-0005
318786-0007
318786-0009
318786-0011
28,57
28,57
50,4
50,4
63,9
77,4
90,9
118,9
Tri-Clamp
Tri-Clamp
Tri-Clamp
Tri-Clamp
Tri-Clamp
Tri-Clamp
121,0
147,0
168,0
68,57
68,57
68,57
68,57
68,57
68,57
75,0
92,0
105,0
Schweissnaht sauber
verschliffen
Var. DN
! y D
d
N6
Zch.-Nr.
Tri-Clamp
Einlegeteil
! y d
D C
s A
! y A
Anschluss-Flansch
A Kennz. L
S - TIG geschweisst unter Formiergas
0, 5 A
2
B
C
L
(40) Kennzeichnung
±
(Hoehe: 6 mm)
1
1
0, 2 B
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 35
Phase separation with inductive
conductivity sensor CLS 51, Tri- ,
DIN/ISO Clamp connection.
Tri-, DIN/ISO Clamp connection - installation examples
Single place installation doesn´t
need reflection to the differences
between Tri-Clamp and DIN/ISO
Clamp.
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 36
For Sensors:
Liquiphant FTL 333L/H,Liquiphant M
Liquiphant FTL 330H
with flush process
connection "F"
Liquiphant FTL 330H with flush
process connection "F"
PRO
Quick connection
Very small dimensions
Sensor can be rotated
Flush mount design
CONTRA
For Liquiphant only
Endress+Hauser process adapter “F”
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 37
Level switch at a fermentation tank with Liquiphant FTL 330L
process connection "F"
Endress+Hauser process adapter “F”
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 38
For Sensors:
pH-assembly CPA 460, 465
pH-armature CPA 465
with Adapter DN 25
PRO
Small dimensions
Hygienic design
CONTRA
For CPA‘s only
Not established in
food yet
Process adapter DN25
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 39
Installation recommendations for
hygienic process connections
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 40
Absolutely must to keep a plant hygienic
Self draining installation!
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 41
Hygienic problematic design Hygienic Design
Installation
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 42
Mind the direction of production and cleaning!
The depth should not be more than 1,5 times the diameter
Hygienic installation: Dead Ends
15/11/2005
Hygienic Process Connections
Michael Mies
Slide 44
Varivent® Inline-housing
GEA Tuchenhagen
Delta Inline-housing
APV
Pressure transmitter
Cerabar M Conductivity sensor
Smartec S
Pressure transmitter
Cerabar M for diameter
10-15 mm
Temperature transmitter
TST 14
Good examples for the installation of instruments in pipes
Endress+Hauser – recommendation
for flush mounted pipe adaptation