1
SHIMA PARTO Co.Always the best solution
Principle of X-Ray SystemsPrinciple of X-Ray Systems
Hossein Ebrahimi NasabBiomedical Engineer (M.Sc)
SHIMA PARTO Co.Always the best solution
X Ray Imaging
SHIMA PARTO Co.Always the best solution
History
Wilhelm Conrad ROENTGEN, a German physicist, discovers accidentally unknown rays, able to pass through his hand and image the bones and his ring
November 8, 1895
He called them…. “X-rays”
He was awarded the first Nobel Prize for physics in 1901
• One year later, Elihu Thomson demonstrates use of X-rays for diagnosis of fractures and location of foreign objects and Eddy C. Jerman starts the first school for Radiographers
SHIMA PARTO Co.Always the best solution
PHYSICS OF X-RAYS
Nature of X-rays
Energy unitEnergy unit
Interaction with matter
SHIMA PARTO Co.Always the best solution
SHIMA PARTO Co.Always the best solution
PARTICLES ENERGY (Definition)
When an electron is submitted to an electrical field, it is accelerated and gets a speed (an energy) proportional to the applied voltage.
This energy is expressed in “eV” : Electron Volt
When this electron reaches the positive electrode, it has a kinetic energy of 1000eV or 1 KeV
KeV is a unit of energy used for all kinds of particles
SHIMA PARTO Co.Always the best solution
INTERACTION WITH THE MATTER
In vacuum:photon move along a straight line
In materials photon can beIn materials, photon can be- transmitted (no interaction)
- absorbed (they disappear, transferring their
energy to the material),
- scattered: they are deviated, with or without
loss of energy
SHIMA PARTO Co.Always the best solution
INTERACTION WITH THE MATTER
Absorption increases with the density of the materialit is what we want to assess, and also what causes biological damage
Transmission is what we can see photographicallyp g p yIt is what we use to assess the amount of radiation absorbed
Scatter is undesirableIt interferes with the measurement of the transmitted radiation
High density tissues (bones) result in low transmissionScattered radiation reduces the quality of the image
SHIMA PARTO Co.Always the best solution
PRODUCING AN IMAGE
The X-ray image displays the part of the X-ray beam not absorbed by the tissues. The less the absorption, the darker the film.
The radiograph displays these differences in tissue density
2
SHIMA PARTO Co.Always the best solution
Technology
The heated filament produces a cloud of free electrons
These electrons are attracted d l t d t d thand accelerated towards the
anode by the electrical field (kV)
The flow of electrons is an electrical current (mA)
The electrical power (100 kV x 100 mA = 10 kW) is converted into heat (99%) and X-ray power (1%)
SHIMA PARTO Co.Always the best solution
X-ray Spectrum: Variation with Anode Material
• it changes the emission efficiencyy(continuous spectrum)
• it determines the presence of characteristic radiations and their energy
SHIMA PARTO Co.Always the best solution
X-ray Spectrum:Variation with kV
kV changes the emission efficiency
it changes the shape of theit changes the shape of the spectrum : maximum energy of photons in the spectrum (in keV) has the same value as tube HV (in kV)
it has a significant effect on the relative amplitude of characteristic radiations
SHIMA PARTO Co.Always the best solution
X-ray Spectrum:Variation with mA
• At a given energy, the rate of photon production is proportional to the mA1 mA10 mA50 mA100 mA500 mA1000 mA
• the shape of the spectrum remains the same when mA is varied
SHIMA PARTO Co.Always the best solution
FILTRATIONBeam hardening (low energy filtration)
Any material in the beam path absorbs more of the lower energy radiation (harmful for the patient) than the higher energy
Regulation require a minimum of 2.5 mm Aluminum equivalent filtration
SHIMA PARTO Co.Always the best solution
Take Away• 99% of the electrical power is converted into heat
• The X-ray beam is multi-energetic (spectrum)– Continuous spectrum – Specific peaks
• kV control mainly the maximum energy
• mA control the intensity of the beam
• Anode material defines the specific peaks
• Anode/filter combination enhances specific energies
SHIMA PARTO Co.Always the best solution
The Main ComponentX-ray tubeProduction of X-rays
CollimatorBeam size control
HV GeneratorPower supply Exposure Exposureparameters control
Image receptor• Static (Radiography)
Film• Dynamic (Fluoroscopy)
Image Intensifier + video camera
• Static & dynamicDigital detector+ Workstation
SHIMA PARTO Co.Always the best solution
THE X-RAY TUBE
insert Tube casing (housing)
SHIMA PARTO Co.Always the best solution
Tube casing (housing)The x-ray tube insert is mounted inside a protective housing. The
housing is made of aluminum and lined with lead. The housing:
C l d l k di i• Controls scatter and leakage radiation• Isolates high voltage• Cools the tube
3
SHIMA PARTO Co.Always the best solution
The tube is inserted inside a casing and immersed in oil for electrical insulation and cooling
The casing also shields x-rays emitted in all directions
SHIMA PARTO Co.Always the best solution
Insert Compeonent
Tube insert consist of the:
FrameFrame
Cathod
Anode
SHIMA PARTO Co.Always the best solution
AnodeFilament and Cathode
Electrons e- beam
X-Rays beam
SHIMA PARTO Co.Always the best solution
X-RAY TUBE COMPONENT: Frame
The outer part of the insert (sometimes called the enclosure or frame) can be made of glass, steel, or copper.
The enclosure:
• Provides a structural base for the cathode and anode• Provides high-voltage insulation• Allows a vacuum to be maintained
SHIMA PARTO Co.Always the best solution
X-RAY TUBE COMPONENT: Cathode(1)The cathode provides the electron stream for x-ray production.
The cathode consists of:Filaments: which are coils of tungsten wire that, when
heated, provide the electron stream for x-ray productionheated, provide the electron stream for x ray production
Focusing cup: which surrounds the filament and concentrates, or focuses, the electron beam
SHIMA PARTO Co.Always the best solution
X-RAY TUBE COMPONENT: Cathode(2)Dual Focus CupSome focusing cups have two filaments.
Th l fil• The large filament:used for higher power levels needed for dense anatomy. (High mA, high kV, or both)
• The small filament:used for lower dosage levels, which provides better resolution.
SHIMA PARTO Co.Always the best solution
X-RAY TUBE COMPONENT: Anode
X-rays are produced by energy conversion when electrons strike the anode.
The anode consists of 3 major components:•A target, which receives a stream of electrons from the cathode
•A rotor of an induction motor, which holds the target and drives its rotation at a rate of approximately 10,000 rpm. This rotation dissipates the heat generated during the production of x-ray
•A bearing assembly, which supports the rotor and target assembly, and provides smooth rotation of the target
SHIMA PARTO Co.Always the best solution
Rotore
SHIMA PARTO Co.Always the best solution
X-ray Generation
The process by which the x-ray tube produces x-rays is very inefficient.
Only one percent of the energy produced is converted to x-rays; the rest is dissipated as heat.
4
SHIMA PARTO Co.Always the best solution
Focal Spot Characteristics
Actual Focal Spot:The electron beam from the tube filament lands on the angled target surface and creates the focal spot This areacreates the focal spot. This area(rectangular area) called the actual focal spot.
The actual focal spot is the area on the anode target where electrons strike, and where x-rays and heat are produced
SHIMA PARTO Co.Always the best solution
Focal Spot Characteristics
Effective Focal Spot:
The area from which the x-ray beamtappears to come.
Viewed from the perspective of the centralray.
The x-ray tube's effective focal spot is commonly called the focal spot.
SHIMA PARTO Co.Always the best solution
Focal Spot CharacteristicsFocal Spot Size and Shape:
Length of the tube filament Diameter of the filament wire coil Width and length of the focusing cup slotg g pDepth of the focusing cup
Focal Spot Size and Target Angle:as the target angle increases, the effective focal spot size increases.
The slope of the anode target allows a larger area to be heated while keeping the apparent area from which x-rays are produced as small as
possible.
SHIMA PARTO Co.Always the best solution
SHIMA PARTO Co.Always the best solution
Anode Heat Capacity
Energy= Kv*mA*time
SHIMA PARTO Co.Always the best solution
KV mAS HU
55 5 275
60 5 30060 5 300
70 16 1100
75 50 3750
)( 80 50 4000
90 100 9000
SHIMA PARTO Co.Always the best solution
• 300KHU• 40KHU/min• 1250KHU
• 3000HU 100
• 40KHU/min 4
SHIMA PARTO Co.Always the best solution
Take AwayAn X-ray tube is characterized by:
• Focal spot size (single or dual)– The smaller the focus, the better the spatial resolution
• Anode rotation speed
• Maximum heat storage capacity (kHU or kJ)
SHIMA PARTO Co.Always the best solution
The Generator
Control kV applied to the x-ray tube, thereby controlling x-ray beam penetration
The main functions of the x-ray generator are:
Control the amount of mA in the tube, thereby controlling the quantity of x-ray beams
Control exposure time and switching
5
SHIMA PARTO Co.Always the best solution
Types Of Generators(1)• Single phase generator• Three phase generator• High frequency generator• Battery powered generator• Capacitor discharge
SHIMA PARTO Co.Always the best solution
Types Of Generators(2)
Single phase generator:
SHIMA PARTO Co.Always the best solution
Types Of Generators(3)Three phase generator:
SHIMA PARTO Co.Always the best solution
Types Of Generators(4)High frequency generator:
SHIMA PARTO Co.Always the best solution
Types Of Generators(5)Battery powered generator:
SHIMA PARTO Co.Always the best solution
Types Of Generators(6)
Capacitor discharge:
SHIMA PARTO Co.Always the best solution
Generator Control ConsoleThe control console variables include:
– kV– mA– Exposure timeExposure time
SHIMA PARTO Co.Always the best solution
Generator Parameters
Importance of High KV & mA
Power of Generator = Kilo Watt (KW) = KV x mA / 1000
mAs = mA x Time (mA x sec)
KV Penetration Good For Thick Patient
mA Resolution High Contrast Image
SHIMA PARTO Co.Always the best solution
Power Voltage Tube Current
Exposure Time
Radiation Amount
Unit kW kV mA sec mAs
Clinical M i No Penetration Contrast
Shorter Time avoids Accumulated
D A tMeaning No Penetration Contrast avoidsmotion blur Dose Amount
Correlation W = VxA mAs=mAxsec
6
SHIMA PARTO Co.Always the best solution
Generator Classes
Middle Power Generators• Around 50kW
High Power Generators• Around 80~100kW• For long time exposure• For high mA Fluoroscopy
Low Power Generators• Example : Less than 20kW• Limited with Radiography
• For General Purpose• Fluoroscopy is possible
SHIMA PARTO Co.Always the best solution
General Radiology
SHIMA PARTO Co.Always the best solution
System Overview
SHIMA PARTO Co.Always the best solution
SHIMA PARTO Co.Always the best solution
Collimator(1)The collimator:
• Controls the dimensions of the x-ray beam
• Aligns the x-ray beam, the patient, and the bucky prior g y , p , y pto exposure
• Limits patient and technician exposure to radiation
• Reduces scatter radiation
SHIMA PARTO Co.Always the best solution
Collimator(2)Two pairs of shutters control the beam dimensions.
Patient Centering
A light and mirror are built into theA light and mirror are built into the collimator to pre-position the x-ray beam.
The simulation light allows the technologist to pre-center the beam on anatomical landmarks.
Notice the cross hair on the patient's shoulder.
SHIMA PARTO Co.Always the best solution
SHIMA PARTO Co.Always the best solution
Cassette Tray• The cassette tray is a mechanical
component of the x-ray system.
Purpose :Purpose :– Holds the cassette – Centers the cassette
SHIMA PARTO Co.Always the best solution
Bucky A bucky provides the housing for three assemblies:
– Cassette tray– Grid– Automatic exposure control
A b k b i h i t l– A bucky can be in a horizontal(table) or vertical (chest stand) orientation.
7
SHIMA PARTO Co.Always the best solution
Automatic Exposure Controller (A.E.C)
Automatic Exposure Control measures the dose ofradiation that strikes the X-ray film behind the patient, andturns the X-ray system off when the predetermined dosefor that screen-film combination has been reached. Thisassures that only the smallest required dose isadministered. The resulting images all show a uniformblackening, and the danger is reduced that the X-rayexamination might have to be repeated owing to an error inthe image. In this way, automatic exposure timing alsoindirectly reduces the dose.
SHIMA PARTO Co.Always the best solution
GridIn 1913, Gustav Bucky introducedthe radiographic grid, a deviceplaced between the patient andthe cassette.
The purpose of the grid is to:
Reduce the amount of scatterradiation reaching the filmIncrease contrast in the x-ray imageScatter radiation is a noise factor. Itcauses a fogging which impairsradiographic quality, lesseningimage contrast. Scatter providesno useful information.
SHIMA PARTO Co.Always the best solution
• GRID TYPES
• GRID SPECIFACTIONS
SHIMA PARTO Co.Always the best solution
Grid Design
• A radiographic grid is composed ofstrips of lead, separated by aluminum, amaterial that is relatively transparent to x-rays. The graphic shows how gridremoves scatter radiation.
• Because the scatter radiation,represented by dotted lines, moves invarious directions, it is largely absorbedby the lead strips.
• The transmitted x-rays pass untouchedthrough the grid and reach the film.
SHIMA PARTO Co.Always the best solution
Grid Pattern(1)X-ray grids are available with either focused or parallel lead strips.
Focused Grid:The lead strips of this grid are angled.A line drawn through each lead strip intersects at apoint a specified distance from the grid, as shownhere.here.
Two important characteristics ofthe focused grid are:
• The distance from the grid to the point of intersection is called the focal distance of the Grid
• The density of the x-ray image is uniformfrom edge to edge, provided the grid is properlycentered and leveled relative to the
SHIMA PARTO Co.Always the best solution
Grid Pattern(2)
Parallel GridWhen the lead strips are not angled, they are all perpendicular to the face of the grid. This type of grid is termed parallelis termed parallel.
Parallel grids have limited use because:• The uniform density of an x-ray exposure is limited to a width of about 10 cm
The advantage of a parallel grid is:• significant only for an extremely long focal distance
SHIMA PARTO Co.Always the best solution
A E C
SHIMA PARTO Co.Always the best solution
The X-Ray FilmTHE RADIOGRAPHIC FILM CASSETTE:
SHIMA PARTO Co.Always the best solution
Latent Image Formation
When the radiation interacts with the silver halide crystals in the film emulsion, the image on the film is produced.
The image, which is not visible before processing, is called g , p g,the latent image.
An example of another type of latent image is fingerprints. If you touch an item, you leave your fingerprints even though you cannot see them on that item. When that item is treated, your fingerprints become visible.
8
SHIMA PARTO Co.Always the best solution
Film Speed
• “Film speed” refers to the amount of radiation required to produce a radiograph of standard density (darkness).y ( )
• Film speed is determined by the following factors: – The size of the silver halide crystals – The thickness of the emulsion– The presence of special radiosensitive dyes
(Cont’d)
SHIMA PARTO Co.Always the best solution
(Cont’d)
• The film speed determines how much exposure time is required to produce the image on the film.
• A fast film requires less radiation; the film responds more
Film Speed
q ; pquickly because the silver halide crystals in the emulsion are larger.
• The larger the crystals, the faster the film speed. This is the same principle as film speed on photographic film.
• F-speed film, the newest and fastest film on the market today, reduces radiation exposure to the patient by 20% to 60% compared with E-speed and D-speed film.
SHIMA PARTO Co.Always the best solution
• An intensifying screen intensifies or increases the effect of the radiation and thus decreases the amount of exposure time needed.
Intensifying Screen
• The intensifying screen is coated with a material called phosphor that gives off light when struck by x-radiation.
• The film inside the cassette is sandwiched between the intensifying screens and is affected by both the light produced by the phosphor and the x-radiation.
SHIMA PARTO Co.Always the best solution
X-Ray Film Processing• Processing is a series of steps that changes the latent image
on the exposed film into a radiograph by producing a visible image on the film.
• Proper processing is just as important as exposure technique in producing diagnostic-quality radiographs.
• Radiographs that are nondiagnostic because of poor processing techniques must be retaken, exposing the patient to unnecessary radiation.
SHIMA PARTO Co.Always the best solution
The Five Steps in Processing
• Development • Rinsing • Fixation• Fixation• Washing • Drying
SHIMA PARTO Co.Always the best solution
Developing• Developing is the first step in processing films.
• A chemical solution called the developer is used.
• The purpose of the developer is to chemically reduce the exposed silver halide crystals to black metallic silver.
• The developer solution also softens the film emulsion during this process.
SHIMA PARTO Co.Always the best solution
• Rinsing of the films is necessary to remove the developer from the film so that the development process stops.
• Usually agitating the film rack for 20 seconds is sufficient
Rinsing
• Usually, agitating the film rack for 20 seconds is sufficient.
• This must be done under safelight conditions.
SHIMA PARTO Co.Always the best solution
• The acidic fixing solution removes the unexposed silver halide crystals from the film emulsion.
• The fixer also hardens the film emulsion during this process. • For permanent fixation the film is kept in the fixer for a
Fixing
• For permanent fixation, the film is kept in the fixer for aminimum of 10 minutes.
• However, films may be removed from the fixing solution after 3 minutes for viewing.
• Films that are not properly fixed will fade and turn brown in a short time.
• Leaving films in the fixer for a long time (e.g., over a weekend) can remove the image from the film.
SHIMA PARTO Co.Always the best solution
Washing• After fixation, a water bath is used to wash the film.
• The washing step requires about 20 minutes to thoroughly ll h i l f th l iremove all excess chemicals from the emulsion.
9
SHIMA PARTO Co.Always the best solution
• The final step in film processing is the drying of the films.
• Films may be air-dried at room temperature in a dust-free area or placed in a heated drying cabinet
Drying
area or placed in a heated drying cabinet.
• Films must be completely dried before they can be handled for mounting and viewing.
SHIMA PARTO Co.Always the best solution
The Automatic Processor Automatic film processing is a fast and simple method used to process x-ray films.
Other than opening the film packet, all steps of film processingp g p , p p gare handled by the automatic processor.
Automatic film processing requires only 4 to 6 minutes for the development, fixing, washing, and drying of a film, whereas manual processing techniques require approximately 1 hour.(Cont’d)
SHIMA PARTO Co.Always the best solution
The Automatic Processor(Cont’d)
The automatic processor maintains the correct temperature of the solutions and adjusts the processing time.
Proper maintenance of the automatic processor reduces the chance of errors during film processing.
SHIMA PARTO Co.Always the best solution
Components of the Automatic Processor
• The processor housing covers all of the component parts. • The film feed slot is for the unwrapped films to be inserted
into the automatic processor. • The roller film transporter is a system of rollers that• The roller film transporter is a system of rollers that
rapidly moves the film through the compartments. • The developer and fixer compartments holds the
solutions.The film is transported directly from the developer into the fixer without a rinsing step.
• The water compartment holds circulating water. • The drying chamber holds heated air and dries the wet
film.
SHIMA PARTO Co.Always the best solution
SHIMA PARTO Co.Always the best solution
CHARACTERISTICS OF X-RAY FILM
• In general radiography, the X-ray image is first converted to a light image using intensifying screens, which in turn produce a visible pattern of metallic black silver on the X-ray film. Ultimately, the degree of blackening is related to the intensity of the radiation reaching the intensifying screen. The amount of blackness on the film is called the optical density, D, which is defined in Figure 5. For example, if 100 light photons are incident on a film and only one is transmitted the film density would be log10(100) or 2. Useful densities in diagnostic radiology range from about 0.2 to about 2.5. High density means black films.
SHIMA PARTO Co.Always the best solution
Characteristic Curve• If the relationship between the logarithm of the radiation exposure and the
optical density is plotted we obtain a curve known as the Characteristic Curve. For film exposed with an intensifying screen, this curve is essentially sigmoidal in shape It is characterised by:
• The important part of the curve diagnostically is the approximately linear p p g y pp yregion between the toe and the shoulder where the density is proportional to the logarithm of the exposure.
SHIMA PARTO Co.Always the best solution
• The information content resulting from the radiograph arises from differences in the film density, which we can define as radiographic contrast. Radiographic contrast depends on subject contrast and film contrast. For the moment you should recall that subject contrast depends on the differential attenuation of the X-ray flux as it passes through the patient and is affected by thickness, density and atomic number of the irradiated parts of the subject, the kVp, the presence of contrast medium and scattered radiation. For example, relatively few X-ray photons pass through bone compared with soft tissue but care must be taken in selecting the correct kVp in order to produce an X-ray image of high information content for the screen-film to record. That is, the kVp influences the magnitude of the subject contrast.
• Film contrast depends on four factors:– the characteristic curve of the film,– the film density,– use of intensifying screens or direct exposure and– the film processing.
• The slope of the straight line portion of the characteristic curve tells us how much change in film density will occur as exposure changes. The slope or gradient of the curve may be measured and the maximum gradient is called the film gamma,which tells us how well the film will amplify the subject contrast.
SHIMA PARTO Co.Always the best solution
Image Quality(1)INFLUENCE OF ENERGY (kV) The higher the energy,
the higher the transmission ratio,the better the penetration
• High energy may be necessary in dense or thick areas (abdomen, pelvis)• Very low energies are rarely of use and are harmful to the patient