RevA MN302000 1of38
WebCalConfigurationSoftwarefor:
EchoPodDL10,DL14,DL24,DL34,DS14&DX10SeriesEchoSonicLU23,LU27,LU28&LU29Series
FlowlineInc.
10500HumboltStreetLosAlamitos,CA90720Tel:(562)5983015Fax:(562)4318507www.flowline.com
-
2of38 MN302000 RevA
INTRODUCTION/TABLEOFCONTENTS StepOneWebCalPCconfigurationsoftwareisautilityprogramthatallowsuserstoeasilyconfiguretheirEchoPodorEchoSonicultrasoniclevelsensor.Developyourconfigurationusingpreprogrammedfunctionmenusasthetankgraphicandsetpointfieldsautomaticallychangetomatchyourconfiguration.TheninputyourlevelsetpointvaluesandclicktheWritetoUnitbutton.Yourconfigurationwillbeuploadedintothesensorinjustafewseconds. Last,clicktheWiringDiagrambuttontoopenawiringschematicofyourconfiguration inaPDFformat.Printthedocument,disconnectthesensorandwireitpertheschematic.Itsthatsimple!NewFeatures
Volumetrictankshapelinearization SampleProgramforallWebCalconfigurablesensors ViewCurvetoolenablingtheviewingofnonlinearoutputs
TableofContentsSystemRequirements:...........................................................................................................................................3SafetyPrecautions:................................................................................................................................................4GettingStarted:......................................................................................................................................................5 USBFobInterface:...................................................................................................................................6WebCal: .........................................................................................................................................................7 RelayConfigurationDL14,DL24,DL34,DS14Series:..............................................................................8 RelayConfigurationDL10,LU23,LU27,LU28&LU29Series:...............................................................15 RelayConfigurationDX10Series:..........................................................................................................17 TankShapeSelection:..............................................................................................................................20 Example:DimensionalEntryVerticalCylindrical:.....................................................................21 Example:DimensionalEntryHorizontalCylindrical:..................................................................23 TankLevels:..............................................................................................................................................25 WritetoUnit:...........................................................................................................................................26 Wiring:.......................................................................................................................................................27 WiringDiagram:...........................................................................................................................27 WiringtoEchoPodandEchoSonic:..........................................................................................27 AdvancedFeatures:.................................................................................................................................28Appendix: .......................................................................................................................................................30 SampleProgram:......................................................................................................................................30 UpdatingWebCalSoftware:..................................................................................................................31 UpdatingSensorFirmware:.....................................................................................................................32 DemoPage:..............................................................................................................................................33 StrappingTable:.......................................................................................................................................34 Linearvs.NonLinear:..............................................................................................................................35 Example#1VolumeOutput:.....................................................................................................35 Example#2CurrentOutput:.....................................................................................................36Troubleshooting:..................................................................................................................................................37Warranty,ReturnsandLimitations:.....................................................................................................................38
-
RevA MN302000 3of38
SYSTEMREQUIREMENTS StepTwo ComputerSystemRequirementsforWebCal
o Windows2000,XP,Vista,7,8o 32or64bitsystemo 1USB2.0porto 10mBharddrivespaceo 256mBRAMo Internetconnection
SensorRequirementsEchoPodSeriesSensorDL10Series DL1000,DL1001,DL1010&DL1011DL14Series DL1400,DL1401,DL1410&DL1411DL24Series DL2400,DL2401,DL2410&DL2411DL34Series DL3400,DL3401,DL3410&DL3411DS14Series DS1400,DS1401,DS1410&DS1411DX10Series DX1000,DX1001,DX1010&DX1011
EchoSonicSeriesSensorLU23Series LU2300,LU2301,LU2310,LU2311,LU2340,LU2341,LU2350&LU2351LU27Series LU2700,LU2701,LU2710,LU2711,LU2740,LU2741,LU2750&LU2751LU28Series LU2800,LU2801,LU2810,LU2811,LU2840,LU2841,LU2850,LU2851,LU2860,LU2861,LU2870&LU2871LU29Series LU2900,LU2901,LU2910,LU2911,LU2940,LU2941,LU2950,LU2951,LU2960,LU2961,LU2890&LU2971
InterfacingRequirementso KeyFob,PartNumberLI991001
Dependingonthesensorpartnumberordered,aKeyFobmayneedtobeorderedinadditiontothesensor.
AKeyFob(LI991001)isrequiredtointerfacethesensorwiththecomputer.
-
4of38 MN302000 RevA
SAFETYPRECAUTIONS StepThree AboutthisManual:PLEASEREADTHEENTIREMANUALPRIORTOINSTALLINGASENSORORUSINGTHIS
PRODUCT.ThismanualincludesinformationontheWebCalconfigurationsoftwarefromFLOWLINE.Pleaserefer to the sensor part number located on the sensor label to verify the exactmodel configuration andcompatibilitytotheWebCalsoftware.
UsersResponsibilityforSafety:FLOWLINEmanufacturesabroadrangeoflevelsensingtechnologies.Whileeachofthesesensors isdesignedtooperate inawidevarietyofapplications, it istheusersresponsibilitytoselectasensormodel that isappropriate for theapplication, install itproperly,perform testsof the installedsystem,andmaintainallcomponents.Thefailuretodosocouldresultinpropertydamageorseriousinjury.
Safety Installationshouldbedonebyproperlytrainedstaff Supplyvoltageshouldneverexceedamaximumof28VDC Makesurethesensorischemicallycompatiblewithyourapplication Designafailsafesystemthataccommodatesthepossibilityofsensorand/orpowerfailure Thissensorshouldnotbeusedinclassifiedhazardousenvironments
MakeaFailSafeSystem: Designa failsafesystem thataccommodates thepossibilityofsensorand/orpower failure. FLOWLINErecommends theuseofredundantbackupsystemsandalarms inaddition to theprimarysystem.
Flammable,ExplosiveorHazardousApplications:WebCalshouldnotbeusedwithinclassifiedhazardousenvironments.
-
RevA MN302000 5of38
GETTINGSTARTED StepFourEchoPod and EchoSonic are configured throughWebCal, a PC software program. WebCal is a freedownloadfromFlowlineswebsite. YoumustdownloadandinstallWebCalpriortopluggingintheUSBFob.Pleasegotowww.flowline.com/webcal.php,andselectyourlanguageversion.
WebCalSystemRequirementsWindows2000,XP,Vista,7,8
32or64bitsystem1USB2.0port
10mBharddrivespace256mBRAM
Internetconnection
-
6of38 MN302000 RevA
GETTINGSTARTED(continued) StepFourUSBFobInterface:EchoPodandEchoSoniccommunicatewithWebCalthroughaUSBinterfacecalledaFob.Note:BeforepluggingyourFobintoyourcomputersUSBport,besurethatyouhaveinstalledWebCalonyourcomputer. Connectthered,green,whiteandblackwiresfromsensor intothecorrectterminalsontheFob.TightenthescrewsontheterminalsandplugyourFobintotheUSBportofyourcomputer.
WiringisidenticalforallEchoPodandEchoSonicseriesUseonlytheRed,Black,GreenandWhitewires. Themaximumcabledistancebetweenthecomputerandsensoris15.Thisonlyapplieswhenconfiguring
theEchoPodorEchoSonic. Once the sensor is configured and prior to installation, isolate thewhite and greenwires from active
powertopreventashortoftheconfigurationcircuit. Note:WhenusingtheFob,donotaddVDCorVACpower.TheFob,whenconnectedtothecomputer,will
providetherequiredpowertotheEchoPodorEchoSonic.
-
RevA MN302000 7of38
WEBCAL StepFiveWithEchoPodorEchoSonicconnected toyourcomputer;open theWebCalsoftwarebyclickingon theWebCal icon. Followthestepsbelowtoconfigurethesensor. ClickHelp inthe lowerrighthandcornerandopen thehelpmenuofWebCal for instructionsonWebCal. IfyouneedadditionalassistanceusingWebCal,pleasecontactaFlowlineapplicationsengineerat(562)5983015.
TabsforConfiguration,
UpdatesandDemomode
ModelSelector(SampleProgramOnly)
ConfigurationSelector
SensorDetails
LevelConfiguration
ActionButtons
TankSelector
AbovescreencaptureisfortheEchoPodDL24series.
ConfiguringEchoPodorEchoSonicwithWebCal
1. RelayConfigurationa. Configurestherelaysintermsofpump/valveoperationsaswellashighorlowalarms.b. Setsfailsafefortherelaysandthesensorsoutput(current,voltageorfrequency).
2. TankShapeSelectiona. Definestheshapeofthetankaswellasthedimensionalinformationforthetankwithrespect
tothesensorslocationonthetank.3. TankLevelConfiguration
a. Entersthesettingsfortherelayactivationpointsaswellasconfirmstheoperationalrange.4. WritetoUnit
a. Uploadstheconfigurationintothesensor.b. Providesacustomwiringdiagramspecifictothesignaloutputand/orrelayconfiguration.
-
8of38 MN302000 RevA
WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFiveThissectionofWebCaliswhereyouselectthelevelconfigurationsettings.Startfromthetopandworktothe bottom, choosing the selections that are applicable to your configuration. Not Applicable willautomaticallyshowwhenaselectiondoesntapplytoyourconfigurationsettings,andyoumaymoveon.AllconfigurationsettingsmustbeselectedorhaveNotApplicablebeforeyoucancontinuetothenextstep.Note:PressingtheClearScreenbuttonwillresettheconfigurationtableandallowaccesstoallfeatures.
NumberofPumpsThisfeatureallowsyoutoselectthenumberofpumpsorvalvesusedwithEchoPod.Thissettingactivatesthecontrolcapabilitiesofoneortworelays.Controlrelaysareoftenreferredtoaslatchingrelays.
o Switch/AlarmsOnlyThe relayswillbe standard singlepointnonlatchingrelays.Usethissettingforhighand/orlowalarms.
o 1Pump/Valve One relaywill be configured as a control orlatchingrelay (relaywillhaveastart levelandaseparatestoplevel). Use this setting to control one pump or valve forautomaticfillingoremptyingofatank.
o 2Pumps/Valves Two relays are configured as control orlatchingrelays. Eachrelaywillhaveauniquestart levelandacommon stop level. Use this setting tocontrol twopumpsorvalvesforautomaticfillingoremptyingofatank.
o 420mATransmitterOnlyThissettingwilldisengagealloftherelays. Use this function if you are not using any relays andusingonlythe420mAcurrentoutput.
Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.
-
RevA MN302000 9of38
WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive
Pump/ValveActionThisfeatureallowsyoutoselectwhetherthepumpsorvalveswillbeusedtoautomaticallyfilloremptythetank.For2Pump/Valvemode,bothdevicesmustbeused inthesame(automaticfillorempty)way.Youcannotsetonerelayforfillandtheotherforempty.
o EmptiesTankSetstherelay(s)toautomaticallyemptyatank.ThestartlevelwillbeabovetheStoplevelforeachrelay.
o Fills Tank Sets the relay(s) to automatically fill a tank. ThestartlevelwillbebelowtheStoplevelforeachrelay.
o NotApplicable Appearswhen this function is not availablebasedonpreviousselections.
EmptiesTank FillsTank (AutoEmpty) (AutoFill)
Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.
-
10of38 MN302000 RevA
WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive
Pump/ValveModeThisfeatureallowsyoutoselectthecontrolmodeforalatchingrelay.Pump/Valve mode is not active for Switch/Alarms Only or 420 mATransmitterOnly.
o Simplex Allows the relay to be used for automatic fill orempty.Thisisthedefaultandonlymodewhen1Pump/Valveisselected.
o SimplexusedtoEmptyTank
o Lead/LagAllowstworelaystohaveuniquestart levelsanda
commonstoplevel.Thefirstrelaywillbeidentifiedastheleadrelayandthesecondrelayasthelag.Eachtimetheleadlevelisreached,thefirstrelaywillalwaysstart.Thelagrelaywillonlystartwhen the lag level is reached. All relayswill stopat thecommonofflevel.
o Lead/LagusedtoEmptyTank
o Duplex Allows two relays tohave twodifferent start levels,a
common stop levelandwillalternate the relayswhen the firststartlevelisreached.Thetworelayswillalternateeachtimetheleadlevelisreachedandtheremainingrelaywillstartwhenthelaglevelisreached.Allrelayswillstopatthecommonofflevel.
o DuplexusedtoEmptyTank
o NotApplicable Appearswhen this function is not available
basedonpreviousselections.Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.
-
RevA MN302000 11of38
WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive
RelayFailSafeThisfeatureallowsyoutoselectthefailsafemodefortherelaysintheevent that the sensor looses echo confidence. When the sensorregains echo confidence, the output current will revert back to thecurrentlevelcondition.
o RelaysOffTherelayswillreverttotheOFFstate.ThisappearswhenSwitch/AlarmsOnlyisselected.
o RelaysOnTherelayswillreverttotheONstate.ThisappearswhenSwitch/AlarmsOnlyisselected.
o HoldState The relay(s)will remain in the same stateas thelast confident echo detected. When the sensor regains echoconfidence,therelayswillreverttothecurrentlevel.
o Pump/ValvesOffTherelayswillreverttotheOFFstate.Thisappearswhen1Pump/Valveor2Pumps/Valvesareselected.
o Pump/ValvesOn The relayswill revert to theON state.Thisappearswhen1Pump/Valveor2Pumps/Valvesareselected.
o NotApplicable Appearswhen this function is not availablebasedonpreviousselections.
Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.
-
12of38 MN302000 RevA
WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive
Switch/AlarmConfigurationThis featureallowsyou to select the relayoperation for the switch /alarm(usedasahighorlowalarm).Thenumberofavailablerelaysisbasedupontheprevioussettings.
o NoAlarmTurnsOFFallremainingrelays.o HighAlarmsSets1to4HighAlarms(1High,2High,3High,
4High).o LowAlarms Set1to4LowAlarms(1Low,2Low,3Low,4
Low).o Combination Alarms Sets a combination of High and Low
Alarms (1Low 1High, 1Low 2High, 2Low 1High, 2Low 2High,1Low3High,3Low1High).
o NotApplicable Appearswhen this function is not availablebasedonpreviousselections.
HighAlarm LowAlarms CombinationAlarms (4High) (4Low) (2Highand2Low)Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton.
-
RevA MN302000 13of38
WEBCAL(RELAYCONFIGURATION)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive
SwitchHysteresis/DeadbandThis feature allows you to select a hysteresis or dead band for theremaininghighand/orlowalarms.
o Options forHysteresis/DeadbandNoHysteresis,,,1,2,cm,1cm,2cm,5cmorNotApplicable.
o HighAlarmsRelayactivatesabove the setpoint. Relaywilldeactivatewhen the level goes below the set point plus thevalueofthehysteresis.
o LowAlarmsRelay activatesbelow the setpoint. Relaywilldeactivatewhen the level goes above the set point plus thevalueofthehysteresis.
HighAlarmw/Hysteresis LowAlarmw/HysteresisNote:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton
-
14of38 MN302000 RevA
WEBCAL(LOOPFAILSAFE)DL14,DL24,DL24,DL34&DS14SeriesOnly StepFive
LoopFailSafe(NotAvailableonDS14series)Thisfeatureallowsyoutoselectthefailsafecurrentoutputifthesensorloosesechoconfidence.Whenthesensorregainsechoconfidence,theoutputcurrentwillrevertbacktothecurrentlevelcondition.
o HoldLastValue Theoutputwill remain in thesamestateasthelastconfidentechodetected.Example:Iftheoutputwas6.7mA just prior to the lost signal, the sensor will continue tooutput6.7mAuntilechoconfidenceisregained.
o Empty The output will revert to the current value for anemptycondition.When4mAatBottomisselected,thesensorwill output 4 mA during a failsafe condition. If 20 mA atBottom isselected,thesensorwilloutput20mAduringafailsafecondition.
o Full The output will revert to the current value for a fullcondition. When4mAatBottom is selected, the sensorwilloutput20mAduringafailsafecondition. If20mAatBottomis selected, the sensor will output 4 mA during a failsafecondition.
o Overfill(21mA)Thesensorwilloutput21mAduringafailsafecondition.
o Overfill(22mA)Thesensorwilloutput22mAduringafailsafecondition.
OutputatEmpty(NotAvailableonDS14series)Thisfeatureallowsyoutoselecttheorientationofthe4to20mAoutput(4to20mAor20to4mA). Choosewhichoutputsettingbestfitstheapplication. Typicalapplicationsaresetwith4mAatBottom. Factorydefaultis4mAatbottomand20mAattop.Whenconnectingyoursensortoadisplay,youmustaccountforyouroutputorientationsetting.
o 4mA at Bottom The output currentwill be 4mAwhen thesensormeasures an empty tank and 20mAwhen the sensormeasuresafulltank.
o 20mAatBottom Theoutputcurrentwillbe20mAwhen thesensor measures an empty tank and 4mA when the sensormeasuresafulltank.
Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton
-
RevA MN302000 15of38
WEBCAL(LOOPFAILSAFE)DL10,LU23,LU27,LU28&LU29SeriesOnly StepFive
LoopFailSafeThisfeatureallowsyoutoselectthefailsafecurrentoutputifthesensorloosesechoconfidence.Whenthesensorregainsechoconfidence,theoutputcurrentwillrevertbacktothecurrentlevelcondition.
o HoldLastValue Theoutputwill remain in thesamestateasthelastconfidentechodetected.Example:Iftheoutputwas6.7mA just prior to the lost signal, the sensor will continue tooutput6.7mAuntilechoconfidenceisregained.
o Empty The output will revert to the current value for anemptycondition.When4mAatBottomisselected,thesensorwill output 4 mA during a failsafe condition. If 20 mA atBottom isselected,thesensorwilloutput20mAduringafailsafecondition.
o Full The output will revert to the current value for a fullcondition. When4mAatBottom is selected, the sensorwilloutput20mAduringafailsafecondition. If20mAatBottomis selected, the sensor will output 4 mA during a failsafecondition.
o Overfill(21mA)Thesensorwilloutput21mAduringafailsafecondition.
o Overfill(22mA)Thesensorwilloutput22mAduringafailsafecondition.
OutputatEmptyThisfeatureallowsyoutoselecttheorientationofthe4to20mAoutput(4to20mAor20to4mA). Choosewhichoutputsettingbestfitstheapplication. Typicalapplicationsaresetwith4mAatBottom. Factorydefault is 4mA at bottom and 20mA at top. When connecting yoursensortoadisplay,youmustaccountforyouroutputorientationsetting.
o 4mA at Bottom The output currentwill be 4mAwhen thesensormeasures an empty tank and 20mAwhen the sensormeasuresafulltank.
o 20mAatBottom Theoutputcurrentwillbe20mAwhen thesensor measures an empty tank and 4mA when the sensormeasuresafulltank.
Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton
-
16of38 MN302000 RevA
WEBCAL(STARTUPCONDITON)DL10,LU23,LU27,LU28&LU29SeriesOnly StepFive
StartupConditionThisfeatureallowsyoutoselectthestartupcurrentwhenpowerisfirstapplied to the sensor. The sensorwill consume the selected powerwhile it isacquiringthe liquid level. Whenthecorrect levelhasbeenidentified,theoutputwilladjusttothe leveloutput. Usethisfeaturetoavoidfalsealarmswiththecontrollerwhenpowerisfirstappliedtothesensor.
o Empty Thecurrentoutputwillreverttothecurrentvalueforan empty condition. When 4mA at Bottom is selected, thesensorwilloutput4mAwhilethesensorpowersup. If20mAatBottom isselected, thesensorwilloutput20mAwhile thesensorpowersup.
o Mid Tank (12mA) The sensorwilloutput 12mAwhile thesensorpowersup.
o Full The output will revert to the current value for a fullcondition. When 4 mA at Bottom is selected, the sensor willoutput20mAwhilethesensorpowersup.If20mAatBottomisselected,thesensorwilloutput4mAwhilethesensorpowersup.
o Overfill(22mA)Thesensorwilloutput22mAwhilethesensorpowersup.
Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton
-
RevA MN302000 17of38
WEBCAL(SELECTOUTPUT)DX10SeriesOnly StepFive
SelectOutputThis featureallowsyoutoselecttheoutput fortheDX10series. Theoutputcaneitherbevoltageorfrequency.
o VoltageSelect thisoption foravoltageoutput. Youwillbeabletoselectvoltagescaleof02.5V,03.3V,05Vor010V.
o Frequency Select this option for a frequency/pulse output.Theoutputwillbefixedfrom976to2000Hz.
MaximumOutputThis sets the maximum output for the DX10 series. If voltage isselected, then the maximum value will be in volts. If frequency isselected,thenthemaximumvaluewillbeinhertz.
o VoltageChoosefrom:2.5volts,3.3volts,5voltsor10volts.o FrequencyThisvalueisfixedat2000Hz.
MinimumOutputThis sets the minimum output for the DX10 series. If voltage isselected, then the minimum value will be in volts. If frequency isselected,thentheminimumvaluewillbeinhertz.
o VoltageChoosefrom:0.25volts,0.5volts,0.833volts,1voltor2volts.
o FrequencyThisvalueisfixedat976Hz.Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton
-
18of38 MN302000 RevA
WEBCAL(OUTPUTATEMPTY&FAILSAFEOUTPUT)DX10SeriesOnly StepFive
OutputatEmptyThis feature allows you to select the orientation of the voltage orfrequencyoutput(ex.Empty=1voltandfull=10voltsvs.empty=10volts and full = 1 volt). Choose which output setting best fits theapplication.TypicalinstallationsaresetwithMinimumOutput.Whenconnectingyoursensortoadisplay,youmustaccountforyouroutputorientationsetting.
o MinimumOutput The outputwill be the smallest value. Ifreadinginvoltage,thiswillbetheminimumoutputvoltage.Ifreadinginfrequency,thiswillbe976Hz.
o MaximumOutput The outputwill be the largest value. Ifreadinginvoltage,thiswillbethemaximumoutputvoltage.Ifreadinginfrequency,thiswillbe2000Hz.
FailSafeOutputThis feature allows you to select the failsafe output if the sensorlooses signal confidence. When the sensor regains echo confidence,theoutputwillrevertbacktothecurrentlevelcondition.
o HoldLastValue Theoutputwill remain in thesamestateasthelastconfidentechodetected.Example:Iftheoutputwas6.7volts just prior to the lost signal, the devicewill continue tooutput6.7voltsuntilechoconfidenceisregained.
o Empty The outputwill revert to the set value for an emptycondition. WhenMinimumOutput isselected,thesensorwilloutputthe lowestvoltageorfrequencyvalueduringafailsafecondition. If Maximum Output is selected, the sensor willoutputthehighestvoltageorfrequencyvalueduringafailsafecondition.
o FullTheoutputwillreverttothesetvalueforafullcondition.WhenMinimumOutput isselected,thesensorwilloutputthehighestvoltageor frequencyvalueduringa failsafecondition.If Maximum Output is selected, the sensor will output thelowestvoltageorfrequencyvalueduringafailsafecondition.
Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton
-
RevA MN302000 19of38
WEBCAL(STARTUPCONDITION)DX10SeriesOnly StepFive
StartupConditionThisfeatureallowsyoutoselectthestartupoutputwhenpowerisfirstappliedtothesensor.Thesensorwilloutputtheselectedsignalwhileit is acquiring the liquid level. When the correct level has beenidentified,theoutputwilladjusttothe leveloutput. Usethisfeaturetoavoidfalsealarmswiththecontrollerwhenpowerisfirstappliedtothesensor.
o Empty The outputwill revert to the set value for an emptycondition. WhenMinimumOutput isselected,thesensorwilloutput the lowestvoltageor frequencyvalueuntil the level isacquired. If Maximum Output is selected, the sensor willoutputthehighestvoltageorfrequencyvalueuntilthe level isacquired.
o Mid Tank The output will hold at the mid level of theMinimumandMaximumOutputuntilthelevelisacquired.
o FullTheoutputwillreverttothesetvalueforafullcondition.WhenMinimumOutput isselected,thesensorwilloutputthehighestvoltageorfrequencyvalueuntilthelevelisacquired.IfMaximumOutputisselected,thesensorwilloutputthelowestvoltageorfrequencyvalueuntilthelevelisacquired.
Note:Rightclickonanyitemtoopenthehelpmenu.Note:Toresettheconfigurationtable,presstheClearScreenbutton
-
20of38 MN302000 RevA
WEBCAL(TANKSHAPESELECTION) StepFiveStep2 Tank Selection: The sensormaybe configured in volumetricunits (Gallonsor Liters)orDistance(HeightofLiquid)units(inches,cm,feetormeters).WebCalwilldefaultinDistance(HeightofLiquid)withunitsof Inches. To change theunitsor to change fromDistance toVolume,press the Select Tank Shapebuttonlocatednearthecenterofthewindow.
ShapeSelectionWindow:ThiswindowwillshowsthedifferenttankshapeoptionsavailableinWebCal. VerticalCylindrical VerticalCylindricalwithConeBottom HorizontalCylindricalwithEndcaps HorizontalCylindricalwithSphericalEnds Spherical Rectangular StrappingTableUse this feature formanualentry
ofmeasuredtankdistancesandvolumes.Select any of the above tank shapes and press OK toconfirm.
-
RevA MN302000 21of38
WEBCAL(TANKSHAPESELECTION) StepFiveDimensionalEntryVerticalCylindricalExample: Choose theSensorOutputUnitsasDistanceorVolume.AfterchoosingtheSensorOutputUnits,selecttheunitsofmeasurementinthepulldowntotheleft.
UnitsofMeasurementDistance VolumeInchesCmFeet
Meters
GallonsLiters
DistanceSensorOutputUnits:Enterthedimensionsofthetank. Youmustenterdatainallfieldsshown.SensorHeight:Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight:Distancefromthebottomofthetanktothemaximumliquidheight.RiserHeight:Distancethesensorisrecessedwithinariser,measuredfromthebottomofthesensortotheinsideofthetank.
VolumeSensorOutputUnits:Enterthedimensionsofthetank. Youmustenterdatainallfieldsshown.SensorHeight:Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight:Distancefromthebottomofthetanktothemaximumliquidheight.RiserHeight:Distancethesensorisrecessedwithinariser,measuredfromthebottomofthesensortotheinsideofthetank.TankHeight: Distance from the bottom ofthetanktothetopofthestraightsidewall.Diameter: Distance of the inside tankdiameter.
-
22of38 MN302000 RevA
WEBCAL(TANKSHAPESELECTION) StepFiveVolumeTankCapacity: Afterentering thedimensions,press theCapacitybutton toshow theCalculatedCapacityofthe tank. If theCalculatedCapacity isslightlydifferent than theexpectedcapacity,clickon theAdjustCapacityboxandentertheexpectedcapacityofthetank.IftheAdjustedCapacityismorethan10%oftheCalculatedCapacity,recheckthedimensionsinformationenteredabove.
Whenalldimensionsareentered,presstheApplybuttontoreturntothepreviousConfigurationwindow.
ApplyTransfersthedimensionstotheoriginalConfigurationwindow. TanksReturnstothepreviousShapeSelectionwindow. CancelReturnstotheConfigurationwindowwithoutsavinganyinformation. HelpJumpstotheHelpmenu.
-
RevA MN302000 23of38
WEBCAL(TANKSHAPESELECTION) StepFiveDimensional Entry Horizontal Cylindricalwith End Caps Example: Choose the SensorOutputUnits asDistanceorVolume. After choosing the SensorOutputUnits, select theunitsofmeasurement in thepulldowntotheleft.
UnitsofMeasurementDistance VolumeInchesCmFeet
Meters
GallonsLiters
DistanceSensorOutputUnits:Enter thedimensionsof the tank. Youmustenterdatainallfieldsshown.SensorHeight:Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight: Distancefromthebottomofthetanktothemaximumliquidheight.RiserHeight:Distancethesensorisrecessedwithinariser,measured from thebottomofthesensortotheinsideofthetank.
VolumeSensorOutputUnits:Enter thedimensionsof the tank. Youmustenterdatainallfieldsshown.SensorHeight:Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight: Distancefromthebottomofthetanktothemaximumliquidheight.RiserHeight:Distancethesensorisrecessedwithinariser,measured from thebottomofthesensortotheinsideofthetank.TankHeight:Distancefromthebottomofthetanktothetopofthestraightsidewall.CylinderLength:Distanceofthestraightlengthofthetank.EndCapLength:Distanceofoneendcap.Bothendcapswillbeusedinthevolumecalculation.
-
24of38 MN302000 RevA
WEBCAL(TANKSHAPESELECTION) StepFiveVolumeTankCapacity: Uponenteringthedimensions,presstheCapacitybuttontoshow theCalculatedCapacityofthe tank. If theCalculatedCapacity isslightlydifferent than theexpectedcapacity,clickon theAdjustCapacityboxandentertheexpectedcapacityofthetank.IftheAdjustedCapacityismorethan10%oftheCalculatedCapacity,recheckthedimensionsenteredabove.
Whenalldimensionsareenterediscompleted,presstheApplybuttontoreturntothepreviousConfigurationwindow.
ApplyTransfersthedimensionsbacktotheoriginalConfigurationwindow. TanksReturnstothepreviousShapeSelectionwindow. CancelReturnstotheConfigurationwindowwithoutsavinganyinformation. HelpJumpstotheHelpmenu.
-
RevA MN302000 25of38
WEBCAL(TANKLEVELSELECTION) StepFiveStep3TankLevels:ThissectionofWebCaliswhereyouconfirmtheSensorHeightandFillHeightsettingswhichapplytoallsensorseriesandentertherelaysettingsifapplicable.TheDimensionalentrywindowwilltransferthesesettingstotheConfigurationwindow.IfVolumewasselected,thentheSensorHeightandFillHeightsettingswillbegrayedout.
For theDL10,DL14,DL24 andDL34 series, the SensorHeight and FillHeight values determine theoperationalrangeforthe420mAoutput.
FortheDX10series,theSensorHeightandFillHeightvaluesdeterminetheoperationalrangeforthevoltageorfrequencyoutput.
SensorHeight: Distancefromthebottomofthetanktothebottomofthetransducer.FillHeight:Distancefromthebottomofthetanktothemaximumliquidheight.
WebCalRelayValues(DL14,DL24,DL34andDS14seriesonly):ThissectionoftheConfigurationwindowiswhereyouentertheoperationalvaluesfortherelays. Youmustentervalueswithinallfieldsshown. Allvaluesmustbe in thepreviously selectedunitsofoperation and arebaseduponunitsof liquid. All relayvaluesmustbegreaterthananemptytank(0units)andlessthanafulltank(FillHeightorCapacityVolume).
Example: Thisisa4HighLevelAlarmoperation. Asthe levelincreases from the bottomup,eachalarmwillactivateat4increments.Example: This is a 2PumpLead/Lag operation with aHigh and Low alarm. Thepumps will automaticallyempty the tank and thealarmswillactivateifthelevelgetstoohighorlow.
-
26of38 MN302000 RevA
WEBCAL(WRITETOUNIT) StepFiveWritetoUnitAfteryouhaveenteredconfigurations,selectedandconfiguredtheTankShapeandenteredthe TankValues, click Write toUnit and load theconfiguration into thememoryofthesensor. Whencompleted, this configurationwill remain inside thesensormemoryandwillnotchangeunlessthesensorisconnectedtoWebCalandanewconfiguration iswrittentothesensor. Lossofpowerwillnotchangeorlosetheconfigurationwithinsensormemory.
WritetoUnitWiringdiagramSaveConfigFile
Next,usethefilemanagementfeaturestosaveyourconfigurationbyclickingSaveConfigFileandprintyourwiringdiagrambyclickingWiringDiagram.SaveConfigFilewillsavethisconfigurationasatextfilewhichcanbeloadedbackintoWebCalbypressingthe Open Config File button. It is good practice to save the configuration file for each differentconfigurationwithauniquename foreasy identification. Ifusingmultiplesensors in identicalapplications,thenuseofasingleconfigurationfileisrecommended.WiringDiagramwilldisplayaPDF fileshowing theuniquewiring for thespecificconfigurationcreated inWebCal.ThePDFcanbeprintedorEmailed.Itisgoodpracticetosavethewiringdiagramasabackup.TechTips:
420mAOutputOnly:Ifonlythe420mAoutputisrequiredwithaDL14,DL24orDL34(andnorelaysareused),underNumberofPumps,select420mATransmitterOnlytosimplifytheconfiguration.
RelaySetPoints: Neversetrelayssetpointsattheextremeendoftheoperationalrange(emptyorfull).Arelayrequirestheleveltopassthesetpointbeforeitswitches.ExampleIfalowalarmissetto0gallons, itwillnevertriggerbecausethe levelmustpassbelow0gallons for ittoactivate. Bestpracticeistosetthelowalarmwithaslightbuffer.
AlarmPrevention:Alwayssetrelayalarmsetpointswherethereistimetoreacttopreventanissue.ExampleSettinganalarmwhereatankoverflowsisnotadvisable.Bestpracticeistosetahighlevelalarmwhereoperatorshaveplentyoftimetopreventanoverflow.
-
RevA MN302000 27of38
WEBCAL(WIRINGDIAGRAM) StepFiveWiringDiagramSample
Diagramschangebaseduponeachuniquesensorconfiguration.
Thediagramshownaboveisonlyasampleandshouldnotbeusedasawiringdiagram.SensorWiring: Once the sensorhasbeen installed, follow theWiringDiagramprovidedby theWebCalsoftware. A typical wiring diagram is shown above. Flowline recommends using a qualified licensedelectriciantowireEchoPodandEchoSonicwithyourapplicationscomponents.
Configure your sensor with WebCal and use the wiring diagram button to view the appropriatediagram. Eachconfigurationwillhave itsownuniquediagram. Thediagramshownabove isonlyasampleandshouldnotbeusedasawiringdiagram.
Alwaysusestepperrelaysbetweenthesensorandexternalloads.ForDCcircuits,alwaysuseacatchdiodesuchas1N4148,shownontheWiringdiagramabovesuppliedbyWebCal.
Once the sensor is configured, isolate thewhiteandgreenwires fromactivepower topreventashortoftheconfigurationcircuit.
Generalnotesforelectricalconnections,usageandsafety: Wherepersonalsafetyorsignificantpropertydamagecanoccurduetoaspill,theinstallationmust
havearedundantbackupsafetysystem. Wiringshouldalwaysbecompletedbyalicensedelectrician. Supplyvoltageshouldneverexceed28VDC. Protectthesensorfromexcessiveelectricalspikesbyisolatingthepower,wheneverpossible. Thesensormaterialsmustbechemicallycompatiblewiththeliquidstobemeasured. Designafailsafesystemforpossiblesensorand/orpowerfailure. Neverusethesensorinenvironmentsclassifiedashazardous.
-
28of38 MN302000 RevA
WEBCAL(ADVANCEDFEATURES) StepFiveAdvancedFeatures: Theadvancedfeaturessettingsaredesignedtohelpsolveperformanceoroperationalissuesforspecificapplications.Changingthesesettingwillalterthefactorydefaultperformanceoroperation,ofyoursensor.PleasereadthroughthisHELPfiletoassistyouinmakingadjustmentsorifyouarestillunclearaboutaspecificissue,pleasecontactFLOWLINEapplicationsengineering.
Note:WhentheAdvancedButtonishighlightedwithaREDborder,thisindicatesyouhaveselectedanadvancedfeature.
-
RevA MN302000 29of38
WEBCAL(ADVANCEDFEATURES) StepFive
IncreaseOutputFiltering:Placingacheckmarkintheboxwillincreasethefiltering(averaging)ofthe
analogoutput.Usethisfilterifthe4to20mAoutputrequiresasmootheroutputfortheapplicationsuchasopenchannelflowmeasurement.
DecreaseOutputFiltering:Placingacheckmarkintheboxwilleliminateallfiltering(averaging)oftheanalogoutputwhichenablesapulsebypulselevelreading.Usethisfiltertoseechangesinlevelaftereveryechopulse.Note:Nevercheckincreaseoutputfilteringanddecreaseoutputfilteringatthesametime.
StabilizeOutputinDeadBand:PlacingacheckmarkintheboxwillactivateafiltertoholdtheoutputatFullifthelevelentersthedeadbandoftheEchoPodandEchoSonic.Thisfilterrequirestheleveltoleavethedeadbandatasmoothandsteadyrate.
InvertRelayStates:Placingacheckmarkinanyofthefourboxeswillreversethestateofthatrelay.Forexample,ifrelay4isahighalarmthatenergizesabove50.0ofliquid,checkingtheinvertboxwillreverseitsstatesotherelaywillenergizewhenthelevelisbelow50ofliquid.Invertingtherelaywillalsoinvertthefailsafeoftherelay.IftherelayisfailsafeOn,invertingtherelaywillmakeitfailsafeOff.
FailSafewillalso invertunlessyoucheckbelow: Placingacheckmark intheboxwillnot invertthefailsafewhenarelayisinverted.
-
30of38 MN302000 RevA
APPENDIX StepSixSampleProgram: AsampleversionofWebCal isavailableanytimeasensor isnotattached toWebCal.TheSampleProgramshowsallthefeatures intheConfigurationTabofWebCal. Anyconfigurationcanbeopened (OpenConfigFile),Saved (SaveConfigFile)orPrinted (PrintConfigFile)with theSampleProgram.SampleProgramcannotbeviewedifasensorisattachedtothecomputerviatheLI991001Fob.ToviewtheSampleProgram,startWebCalwhenasensorisnotattachedtothecomputer.Attheopeningscreen,selectSampleProgram.
Selectyourmodeltypeintheupperrighthandcorner.Note:Whensavingoropeningaconfiguration,makesuretheModelNumbermatchesthesensorsyouintendtouse.
-
RevA MN302000 31of38
APPENDIX StepSixUpdatingWebCal Software: WebCal can be updated directly fromwithin the software. Click on theUpdatesTabatthetopofthewindowandpresstheDownloadbutton. MakesurethatyourcomputerhasaccesstotheInternet.Ifnot,anerrorwindowwillappear.PresstheDownloadbuttonandthesoftwarewillchecktheversionofsoftwareyouareusingwiththemostrecentsoftwareversionatFlowline.Iftheversionsarethesame,awindowwillopenindicatingthatthemostrecentversionisinstalled.Ifnot,thenawindowwillopenaskingyoutodownloadthelatestversion.Followtheinstructionsforinstallingthelatestversion.
-
32of38 MN302000 RevA
APPENDIX StepSixUpdatingSensorFirmware: WebCalcanalsobeusedtoupdatethefirmware insideyoursensorwiththelatest featuresandenhancements. FirstopenWebCalwitha sensor connectedand the latestversionofWebCalinstalledonyourcomputer.
Whenupdatingthesensorfirmware,disconnectthesensorfromallotherdevicesincludingdisplays,controllers,power supplies,PLCs,pumps,valvesandalarms. Connect thedevicesbackafter thefirmwarehasbeenupdated.
ClickontheUpdatesTabandthenclickonSelectProgramtoselectthefirmwareupdate.
SelectthelatestversionofthefirmwarefileandclickOK.
ConfirmthattheaddressiscorrectandthenclickonUpdateSensortobeginthefirmwareupdate.Thisstepshouldtake lessthan1minute. Youcan followtheprogresswiththestatusbartotherightoftheUpdateSensorbutton.Whencompleted,clickontheConfigurationtabtoconfigurethesensor.Remember,whenthefirmwarehasbeenupdated,theunitwillreturntoitsoriginalfactorysettings.
Ifthereisacommunicationinterruptionduringtheupdate,theprocesswillstop.ItisthenOKtoclickonUpdateSensortostarttheprocessoveragain.
-
RevA MN302000 33of38
APPENDIX StepSixDemoPage:TheDemoPagegivesyoutheabilitytotestthesensorsoperationagainstafixedtarget,includingtheconfirmationoftheswitchingpointsfortherelays.
Thelargetopnumbershowstheliquidlevelvalueintheunitslistjustbelowtotheright. Thebottomsmallernumbershowsthedistancefromthesensorfacetothetargetinthesameunits.
o AddingbothnumberswillequaltheSensorHeightvalue. TheActivitybuttonwilllighteverytimethesensorpulses. TheRelaybuttons(Relay1,2,3&4)willlightwhentheircorrespondingrelayisenergized.
o Note:Thisfeatureisonlyasimulation.Therelaysarenotphysicallyopeningandclosing.
Note:Inthismode,thevoltagesuppliedbythecomputerisbelowthesensorsspecifiedlimits.Therefore,thesensormaybesluggish,relaysmaynotoperateandtheunitmaynotoperatetofullrange.Thissimulationisintendedtogiveyouaquickfunctionalcheckafterupdatingthefirmware.
-
34of38 MN302000 RevA
APPENDIX StepSixStrappingTable: WebCal featuresastrapping table thatenablesyou toenterup to16customreferencepoints insteadofusingthestandardtankshapes. Thisfeature is idealforoddshapedtanksortankswherespecificlevelsareknownvolumesofliquid.ToaccesstheStrappingTable,clickonStrappingTableintheShapeSelectionWindowandpressOK. Enter the SensorHeight, FillHeight,
RiserHeight and TankHeight. Thisinformation isused to configure thesensortothetank.
The Strapping Table also has twocolumnsof16pointsforenteringtheknowntankdata.
Selectthedimensionsand/orvolumeat the top of the two columns andenterthetankdata.
When done, press Apply totransfer the data and return to theConfigurationwindow.
-
RevA MN302000 35of38
APPENDIX StepSixLinearvs.NonLinear:Twooftheshapes(VerticalCylinderTankandRectangularTank)willalwaysprovidealinearoutput,regardlessofselectingDistanceorVolume. Theremainingfourshapes(VerticalCylinderTankwithConeBottom,HorizontalCylinderTankwithEndCaps,HorizontalCylinderTankwithSphericalEndCapsandSphericalTank)willhavealinearoutputwhenDistanceisselected,butwillhaveanonlinearoutputwhenvolumeisselected.
VerticalCylindricalTank
In theabove illustration,10of liquidwillalwaysbeequalto100gallonsofliquid(1=10gallons).
HorizontalCylinderTankwithEndCaps
In the above illustration, 1 of liquid does notequal 10 gallons. The 10 at the bottomrepresentsariseof62.8gallonswherethechangebetween 10 and 20 represents an increase of109.6gallons.
Whenvolumeisselected,the420mAoutputfromthesensorwillbeproportionaltothevolumeofthetank,nottheheightofthetank.Thismeansthatthecurrentoutputwilltrackthevolumeofthetank(ingallonsorliters)withinanonlineartank (VerticalCylinderTankwithConeBottom,HorizontalCylinderTankwithEndCaps,HorizontalCylinderTankwithSphericalEndCapsorSphericalTank).Whenconnectingthesignaloutputtoadisplay,thesignalwillfollowthevolumeofthetank.Thedisplaywillalsoreflectthevolumeofthetankandnottheheightoftheliquid.Example#1(VolumeOutput):intheaboveillustrations,@20ofliquid,thedisplaywillshow200.0gallonsintheVerticalCylindricalTank. However, intheHorizontalCylinderTankwithEndCaps,thesame levelof20wouldshow172.4gallons.
-
36of38 MN302000 RevA
APPENDIX StepSixExample #2 (CurrentOutput): In the illustrations on the previous page, the 4mA signal is set at 0 (0.0gallons)andthe20mAsignal issetto60(600.0gallons). IntheVerticalCylindricalTank,40of liquidwilloutputacurrentsignalof14.67mA.However,intheHorizontalCylindricalTankwithEndCaps,50ofliquidwilloutputacurrent signalof15.41mA. A simple loopdisplay setwith4mA=0gallonsand20mA=600gallonswillshow twodifferentvolumesbasedupon the tankshapeconfiguration. VerticalCylindricalTankwillshow400.0gallonswhileHorizontalCylindricalTankwithEndCapswillshow428.0gallons.
VerticalCylindricalTank
In theabove illustration,10of liquidwillalwaysbeequalto100gallonsofliquid(1=10gallons).
HorizontalCylinderTankwithEndCaps
In the above illustration, 1 of liquid does notequal 10 gallons. The 10 at the bottomrepresentsariseof62.8gallonswherethechangebetween 10 and 20 represents an increase of109.6gallons.
-
RevA MN302000 37of38
TROUBLESHOOTING StepSevenPROBLEM SOLUTION
NoUnitDetectedinWebCal.
WebCal cannotdetectanEchoPod orEchoSonicconnectedtothecomputer. CheckthattheFobisconnectedtotheUSBport. Checkthatallfourwires(Red,Black,WhiteandGreen)aresecurely
attachedtotheFob. Check Device Manager and confirm that both drivers (WebCal
Configuration&EchoFob)arepresent.Interneterror.Theservernameoraddresscouldnotberesolved.
ThisisawarningindicatingthatthecomputerconfiguringEchoPod orEchoSonic is not connected to the Internet. ClickOK to continue.Flowline recommends being connecting to the Internet duringconfiguration. Not being connected to the Internetwill not preventEchoPodorEchoSonicfrombeingconfigured.Toturnoffthiswarning,gototheUpdatesTabandclickonthecheckbox Automatically upload configurations. Click onNO in the newwindowandthepreviouscheckboxwillbecomeunchecked.WebCalwillnolongerattempttoconnecttotheinternet.Clickingonthecheckboxwillrestorethisfeature.
CannotaccesssomeoftheconfigurationfeaturesinWebCal.
AschoicesaremadeinConfiguration,WebCalwillbegintoeliminatefunctions that are not applicable to a configuration. To resetConfigurationorgetaccesstoallthefeatures,clickontheClearScreenbutton.
-
38of38 MN302000 RevA
WARRANTY,RETURNSANDLIMITATIONS StepEightWarrantyFlowlinewarrants to the original purchaser of its products that such productswill be free from defects inmaterialandworkmanshipundernormaluseandserviceinaccordancewithinstructionsfurnishedbyFlowlinefor a period of two years from the date ofmanufacture of such products. Flowline's obligation under thiswarrantyissolelyandexclusivelylimitedtotherepairorreplacement,atFlowline'soption,oftheproductsorcomponents, which Flowline's examination determines to its satisfaction to be defective in material orworkmanshipwithinthewarrantyperiod.Flowlinemustbenotifiedpursuanttotheinstructionsbelowofanyclaimunderthiswarrantywithinthirty(30)daysofanyclaimedlackofconformityoftheproduct.Anyproductrepairedunder thiswarrantywillbewarrantedonly for the remainderof theoriginalwarrantyperiod.Anyproductprovidedasareplacementunderthiswarrantywillbewarrantedforthefulltwoyearsfromthedateofmanufacture.ReturnsProducts cannotbe returned to Flowlinewithout Flowline'spriorauthorization. To returnaproduct that isthought to be defective, go towww.flowline.com, and submit a customer return (MRA) request form andfollow the instructions therein.Allwarrantyandnonwarrantyproduct returns toFlowlinemustbe shippedprepaidandinsured.Flowlinewillnotberesponsibleforanyproductslostordamagedinshipment.LimitationsThiswarrantydoesnotapplytoproductswhich:1)arebeyondthewarrantyperiodorareproductsforwhichthe originalpurchaserdoesnot follow thewarrantyprocedures outlinedabove;2)havebeen subjected toelectrical,mechanicalorchemicaldamageduetoimproper,accidentalornegligentuse;3)havebeenmodifiedoraltered; 4)anyoneotherthanservicepersonnelauthorizedbyFlowlinehaveattemptedtorepair;5)havebeeninvolvedinaccidentsornaturaldisasters;or6)aredamagedduringreturnshipmenttoFlowline.FlowlinereservestherighttounilaterallywaivethiswarrantyanddisposeofanyproductreturnedtoFlowlinewhere:1)thereisevidenceofapotentiallyhazardousmaterialpresentwiththeproduct;or2)theproducthasremainedunclaimedatFlowlineformorethan30daysafterFlowlinehasdutifullyrequesteddisposition.Thiswarrantycontains the sole express warranty made by Flowline in connection with its products. ALL IMPLIEDWARRANTIES,INCLUDINGWITHOUTLIMITATION,THEWARRANTIESOFMERCHANTABILITYANDFITNESSFORAPARTICULARPURPOSE,AREEXPRESSLYDISCLAIMED.Theremediesofrepairorreplacementasstatedabovearetheexclusiveremediesforthebreachofthiswarranty.INNOEVENTSHALLFLOWLINEBELIABLEFORANYINCIDENTALORCONSEQUENTIALDAMAGESOFANYKINDINCLUDINGPERSONALORREALPROPERTYORFORINJURYTOANYPERSON.THISWARRANTYCONSTITUTESTHEFINAL,COMPLETEANDEXCLUSIVESTATEMENTOF WARRANTY TERMS AND NO PERSON IS AUTHORIZED TO MAKE ANY OTHER WARRANTIES ORREPRESENTATIONSONBEHALFOFFLOWLINE.Thiswarrantywillbe interpretedpursuant to the lawsof theStateofCalifornia. Ifanyportionofthiswarranty isheldtobe invalidorunenforceableforanyreason,suchfindingwillnotinvalidateanyotherprovisionofthiswarranty.