Approvals and CertificationsTREX-0045-0001 Rev 3
May 2017
AMS Trex™ Device Communicator
Approvals and Certifications
Copyright and trademark information
©2017 Emerson. All rights reserved.
FOUNDATION™, HART® and WirelessHART® are marks of the FieldComm Group of Austin, Texas,USA.
The Emerson logo is a trademark and service mark of Emerson Electric Co.
All other marks are the property of their respective owners.
Notice
ImportantRead this manual before working with the Trex unit. For personal and system safety, and for optimumproduct performance, thoroughly understand the contents before using or servicing this product.For equipment service needs, contact the nearest product representative.
ImportantThis device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:(1) this device may not cause harmful interference, and (2) this device must accept any interferencereceived, including interference that may cause undesired operation.
WARNING!
If the Trex unit is used in a manner not specified by Emerson, the protection provided by the equipmentmay be impaired.
WARNING!
Do not directly connect the ports or terminals on the Trex unit to any main line voltage.
WARNING!
WARNING - POTENTIAL ELECTROSTATIC CHARGING HAZARD - SEE INSTRUCTIONS.
AVERTISSEMENT - DANGER POTENTIEL DE CHARGES ÉLECTROSTATIQUES - VOIR INSTRUCTIONS
Approvals and Certifications May 2017
2 Approvals and Certifications
AMS Trex Device Communicator overview
The Trex unit supports HART® and FOUNDATION™ fieldbus devices, so youcan configure or troubleshoot in the field or on the work bench. ElectronicDevice Description Language (EDDL) technology enables the Trex unit tocommunicate with a variety of devices independent of device manufacturer.
Depending on the attached communication module, the Trex unit lets you:
• Configure HART and FOUNDATION fieldbus devices.
• Power one HART or FOUNDATION fieldbus device.
• Measure current and voltage.
• Perform diagnostics on a 4-20 mA current loop or FOUNDATION fieldbussegment.
The Trex unit includes a color LCD touchscreen, a Lithium-Ion power module(battery pack), a processor, memory components, and optionalcommunication modules.
CAUTION!
When the Trex unit communicates with devices, follow all standards andprocedures applicable to the location. Failure to comply may result inequipment damage and/or personal injury. Understand and comply with thesections in this manual.
Precautions for the Trex unit
Before operating the Trex unit, ensure:
• The Trex unit is not damaged.
• The power module is securely attached.
• All screws are sufficiently tightened.
• The communication terminal recess is free of dirt and debris.
• The communication module is securely attached.
CAUTION!
Do not use a screen protector on an IS-approved Trex unit. Static discharge ispossible.
May 2017 Approvals and Certifications
Approvals and Certifications 3
Front view of the Trex unit
Front viewFigure 1:
A. Micro USB port (top)B. Power button (side)C. Strap connectors (side)D. TouchscreenE. KeypadF. Charger port for the AC adapter (side)
Precautions for the power module and AC adapter
Understand and follow the precautions below before using the powermodule or AC adapter.
• When transporting a Lithium-Ion power module, follow all applicableregulations.
• Ensure sufficient grounding. Ensure the personnel, working surfaces, andpackaging are sufficiently grounded when handling electrostaticallysensitive parts.
• Avoid touching the pins on the connectors or components. Dischargedenergy can affect the power modules.
Approvals and Certifications May 2017
4 Approvals and Certifications
• Protect the power module and AC adapter from moisture, and respectoperating and storage temperature limits listed in the AMS Trex DeviceCommunicator User Guide. The AC adapter is for indoor use only.
• Do not cover the power module or AC adapter while charging. Do notsubject them to prolonged periods of direct sunlight, or place them on ornext to heat-sensitive materials.
• Charge the power module with only the provided AC adapter. The ACadapter should not be used with other products. Failure to comply maypermanently damage the Trex unit and void the IS approval and warranty.
• Do not open or modify the power module or AC adapter. There are nouser-serviceable components or safety elements inside. Opening ormodifying them will void the warranty and could cause personal harm.
• Clean the AC adapter by clearing the terminal of dirt and debris, ifrequired.
• If the AC adapter is used in a manner not specified by Emerson, theprotection provided by the equipment may be impaired.
• The AC adapter comes complete with interchangeable plug heads forcountries UK, USA, EU and AU.
• The maximum operating altitude for the AC adapter is 2000 meters.
Charge the power module
Fully charge the power module before using it in the field. The Trex unit isfully operable when the power module is charging. An overcharge conditionwill not occur if the AC adapter is connected after charging completes. Youcan charge the power module when it is attached to or detached from theTrex unit.
To maintain performance, charge the power module frequently, preferablyafter each use. Limit full discharges, if possible.
If you experience communication issues when working with a device, removethe AC adapter from the Trex unit.
WARNING!
Do not install, remove, or charge the Lithium Ion (Li-Ion) power module in ahazardous area environment.
1. Plug the AC adapter into a power outlet.
2. Attach the AC adapter cable to the charger port on the lower left side ofthe Trex unit.
A full charge takes approximately three to four hours.
May 2017 Approvals and Certifications
Approvals and Certifications 5
Communication modules
The Trex unit has two communication modules.
Device Communicator communication module
The Device Communicator communication module can connect to andcommunicate with HART and FOUNDATION fieldbus devices on anexternally-powered HART loop or fieldbus segment. The DeviceCommunicator communication module has unique terminals for both HARTand FOUNDATION fieldbus devices.
Device Communicator communication moduleFigure 2:
A. Connect to externally-powered FOUNDATION fieldbus devices.B. Connect to externally-powered HART devices.
Device Communicator Plus communication module
The Device Communicator Plus communication module can connect toHART and FOUNDATION fieldbus devices, measure current and voltage, andpower a device.
Approvals and Certifications May 2017
6 Approvals and Certifications
Device Communicator Plus communication moduleFigure 3:
A. Power a FOUNDATION fieldbus device. You need to connect the FOUNDATIONfieldbus Power Plug to the FF pwr and the positive FF terminals.
B. Connect to a FOUNDATION fieldbus device that is externally-powered orpowered by the Trex unit.
C. Measure current on a 4-20 mA current loop.D. Power and connect to a HART device. The HART+pwr terminals can measure
the current output of a connected transmitter or control the current input toa connected positioner. The terminals also have a loop resistor for devicecommunication.
E. Connect to an externally-powered HART device. The HART terminals also havean optional loop resistor for enabling HART communications on 4-20 currentloop and optional current control for moving a positioner.
CAUTION!
• Before you insert or remove a communication module, ensure the Trexunit is powered off.
• Ensure sufficient grounding. Ensure the personnel, working surfaces, andpackaging are sufficiently grounded when handling electrostaticallysensitive parts.
• Avoid touching the pins on the connectors or components. Dischargedenergy can affect the modules.
• When you insert/attach the communication module to the Trex unit, donot over tighten the screws. Use 0.5Nm maximum torque load.
• Remove the USB cable from the Trex unit before connecting to a device.
May 2017 Approvals and Certifications
Approvals and Certifications 7
WARNING!
• The Trex unit cannot power a 4-wire device. Do not connect Trex unit tothe power terminals of a 4-wire device. This can blow a fuse inside theTrex unit. The repair/replacement will need to be completed at anauthorized service center.
• Do not connect lead sets to the HART and HART + pwr terminals at thesame time. If the lead sets are connected to devices, this increases thechance of wiring mistakes and could create a short in the HART loop.
• Do not add any external power to the device when the Trex unit ispowering the device. This can blow a fuse inside the Trex unit. The repair/replacement will need to be completed at an authorized service center.Ensure the device is disconnected from the loop/segment and no otherwires are connected to the device before providing power from the Trexunit.
• Do not use the Trex unit to power a WirelessHART device. Providingpower to a WirelessHART device may damage the device.
• Do not connect the mA terminals (ammeter) in parallel with a powered4-20 mA current loop. Ammeters have low resistance. This can disruptthe loop and cause devices to report incorrect values or positioners tomove unexpectedly.
• Do not connect the mA terminals on the Trex unit to a power supply thatis not current limited to 250 mA. This can blow a fuse inside the Trex unit.The repair/replacement will need to be completed at an authorizedservice center.
Power on or off
1. To power on, press and hold the power button on the upper left side ofthe Trex unit for one second.
2. To power off, do one of the following:
• Quickly press the power button, and then tap Turn Off.• Tap Settings or the status bar at the top of the screen, and tap More >
Power Management > Turn off.
Device connections
Use the provided lead set and the Field Communicator application tocommunicate with a device. The appropriate device description is alsorequired. If the Trex unit does not have the HART device description revision,then the device is displayed in generic mode. This mode does not display alldevice functionality. If the Trex unit does not have the fieldbus devicedescription, the device cannot be configured. See the wiring diagrams in theAMS Trex Device Communicator User Guide for more information.
Approvals and Certifications May 2017
8 Approvals and Certifications
CAUTION!
The Trex unit draws approximately 12 mA from the fieldbus segment when itis online. (The Trex unit draws 0 mA when it is offline.) Ensure the powersupply or barrier on the fieldbus segment has the capacity to provide thisadditional current when the Trex unit is online. If a heavily loaded fieldbussegment is drawing near the capacity of the segment’s power supply,connecting the Trex unit may result in loss of communication.
Maintenance and repair
Any maintenance, repair, or replacement of components not listed belowmust be performed by specially trained personnel at an authorized servicecenter. You can perform common maintenance procedures listed below:
• Clean the exterior. Use only a dry, lint-free towel or dampen the towelwith an alcohol or mild soap and water solution.
• Clean the touchscreen.
• Install, remove, or charge the power module.
• Remove and replace the stand.
• Ensure that all exterior screws are sufficiently tightened.
• Ensure the communication terminal recess is free of dirt and debris.
• Install and remove the communication module.
Technical support
Contact your local representative or go to the AMS Trex DeviceCommunicator website for technical support contact information.
Product certifications
See the AMS Trex Device Communicator website for the latest certificates,declaration of conformity, and approval information.
Approved manufacturing location
R. STAHL HMI Systems GmbH - Cologne, Germany
Labels
Each Trex unit has a main unit label. An Intrinsically Safe (KL option) Trex unithas another label on the side. If the Trex unit does not have this label, it isconsidered non-IS approved.
May 2017 Approvals and Certifications
Approvals and Certifications 9
Certifications and approvals
European directive information - CE compliance
ATEX(2014/34/EU)
This equipment complies with the ATEX Directive. Applicable standards areEN 60079-0:2012 / A11:2013 and EN 60079-11:2012
Certification No.: SIRA 16ATEX2171
II 2 G (1GD) Ex ia [ia Ga] [ia Da IIIC] IIC T4 Gb (Ta = -20°C < Ta < +50°C)
0158
Electro MagneticCompatibility(EMC)2014/30/EU
Tested to the EN 61326-1:2013-07 and ETSI EN 301489-17:2012-09specification.
Low Voltage2014/35/EU
Tested to the IEC 61010-1:2010 specification.
RED (2014/53/EU) This equipment is in conformity with the Radio Equipment Directive (RED)Directive, ETSI EN 300328: 2015-02, and IEC 62209-2: 2010-01 standards.
RoHS(2011/65/EU)
Product is compliant with the RoHS Directive.
International certifications
IECEx Certification No.: SIR 16.0057
Ex ia [ia Ga] [ia Da IIIC] IIC T4 Gb (Ta = -20°C < Ta < +50°C)
North American certifications
CanadianStandardsAssociation -cCSAus
Class I, Division 1, groups A, B, C, D, T4. Class 1, Zone 1 AEx ia [ia Ga] [ia DaIIIC] IIC T4 Gb.
CSA Ex ia [ia Ga] [ia Da IIIC] IIC T4 Gb
Hazardous areas
A Trex unit that meets the Intrinsic Safety requirements (IS-approved) can beused in Zone 1, or Zone 2, for Group IIC, and Class I, Division 1 and Division 2,Groups A, B, C, and D locations.
An IS-approved Trex unit may be connected to loops or segments that areattached to equipment located in Zone 0, Zone 1, Zone 2, for Group IIC; Zone20, Zone 21, Zone 22, and Class I, Division 1 and Division 2, Groups A, B, C,and D locations.
An IS-approved Trex unit can be ordered by selecting the KL option. The Trexunit has a label that lists the approvals.
Approvals and Certifications May 2017
10 Approvals and Certifications
CAUTION!
Do not use a screen protector on an IS-approved Trex unit. Static discharge ispossible.
WARNING!
Do not install, remove, or charge the Lithium Ion (Li-Ion) power module in ahazardous area environment.
WARNING!
Explosions can result in serious injury or death.
Use in an explosive environment must be in accordance with the appropriatelocal, national, and international standards, codes, and practices. Pleasereview the Technical specifications and Product certifications sections of theAMS Trex Device Communicator User Guide for any restrictions associated withsafe use.
Electrical shock can result in serious injury or death.
Intrinsically Safe electrical parameters
Table 1: Device Communicator communication module
FOUNDATION™ fieldbus FOUNDATION™ fieldbus
HART®(non-FISCO) (FISCO)
FF + and - FF + and - HART + and -
Ui 30 Vdc 30 Vdc 30 Vdc
Ii 380 mA 215 mA (IIC) 380 mA (IIB) 200 mA
Pi 1.3 W 1.9 W (IIC) 5.3 W (IIB) 1.0 W
Ci 0 0 0
Li 0 0 0
Uo 1.89 V 1.89 V 1.89 V
Io 32 µA 32 µA 32 µA
Po 61 µW 61 µW 61 µW
Co 14.3 µF 14.3 µF 14.3 µF
Lo 100 mH 100 mH 100 mH
May 2017 Approvals and Certifications
Approvals and Certifications 11
Table 2: Device Communicator Plus communication module
mAinterfac
e
FOUNDATION™
fieldbus
HART®
FOUNDATION™ fieldbus
(non-FISCO) (FISCO)
mAFF pwrand F- FF + and -
HART +pwr
HART +and -
FF pwr andF - FF + and -
Ui 30 Vdc 17.5 Vdc 30 Vdc 30 Vdc 30 Vdc 17.5 Vdc 30 Vdc
Ii 200 mA 380 mA 380 mA 200 mA 200 mA 380 mA 215 mA (IIC)380 mA (IIB)
Pi 1.0 W 1.3 W 1.3 W 1.0 W 1.0 W 1.3 W 1.9 W (IIC)5.3 W (IIB)
Ci 0 231 nF 0 0 0 231 nF 0
Li 0 0 0 0 0 0 0
Uo 0 17.31 V 1.89 V 25.69 V 1.89 V 17.31 V 1.89 V
Io 0 199 mA 32 µA 105 mA 1.9 mA 199 mA 32 µA
Po 0 0.94 W 61 µW 668 mW 3.6 mW 0.94 W 61 µW
Co - See table3
14.3 µF See table4
14.3 µF See table 3 14.3 µF
Lo - See table3
100 mH See table4
100 mH See table 3 100 mH
Table 3: Co and Lo values for FF pwr and F-
Co [nf] 19 69 115
Lo [µH] 100 50 30
Table 4: Co and Lo values for HART + pwr
Co [nf] 57 64 75 102
Lo [µH] 1000 750 500 100
AC adapter specifications
Electrical Specification
AC input voltage range 90-264 VAC
Frequency 47 - 63Hz
Input current 1.6A max. at low line input and full load output
Inrush current 60 A max. at 230VAC input and 25°C ambient temperature
Leakage current <0.25 mA
Input protection 1.6A 250VAC Fuse
Approvals and Certifications May 2017
12 Approvals and Certifications
Electrical Specification
No load power consumption <0.5W max. at nominal input and no load output
Output voltage 15 V
Output current 4.33 A
Ripple and noise 1% Vout
Hold up time 10ms at nominal input and full load output
No load operation Yes, to protect power supply and system from damage
Short circuit protection Auto recovery, cannot exceed 8A max. after 1 minute atnominal input line.
Over current protection 150-200%, foldback type, auto recovery
Over voltage protection 110-140%, voltage limiting, recycle input to recover
Operating temperature range 0°C to + 50°C
Operating humidity 8-90% RH non-condensing
Operating altitude Maximum of 2000 meters
Storage temperature range -20°C to +70°C
Storage humidity 5-95% RH non-condensing
Cooling Free air convection
Dimensions 119 x 54x 36mm
Weight 0.33 kg
Withstand voltage I/P-O/P (FG): 3kVAC / 10 mA / 1 minute
EMI EN55022: 2006 Class B, EN61000-3-2:2006, EN61000-3-3:1995+A1: 2001+A2:2005 EN55024:1998+A1: 2001+A2:2003, IEC61000-4-2, -3, -4, -5, -6+A1:2004, -8, -11
Safety CB (IEC60950-1: 2001), TUV GS (EN60950-1: 2005 +A1 +A2), cUL, CE, VCCI
Approvals
cUL North America, TUV GS Europe, Japan PSE, IRAM Argentina, Russian EAC,Kazakhstan EAC, South Africa SANS IEC 60 950, Korean EK, China CCC.
WEEE 2012/19/EU, RoHS (2011/65/EU)
AMS Trex Device Communicator Control Drawings
See the AMS Trex Device Communicator website at http://www.emerson.com/Trex to view the latest full-size control drawings.
May 2017 Approvals and Certifications
Approvals and Certifications 13
1K
eine Änderung ohne P
rüfung durch Prüfstelle
(z.B.U
L,F
M,C
SA
)!N
o ch
ang
es p
rior to
Ce
rti�catio
n B
od
y app
roval
(e.g
.U
L,
FM
,C
SA
)!
WA
RN
ING
-D
ON
OT
US
ET
HE
US
BC
ON
NE
CT
OR
INH
AZ
AR
DO
US
LO
CA
TIO
NS
-N
EJA
MA
ISU
TIL
ISIE
RL
EC
ON
NE
CT
EU
RU
SB
DA
NS
DE
SZ
ON
ES
ÀR
ISQ
UE
D‘E
XP
LO
SIO
NA
VE
RT
ISS
EM
EN
T
WA
RN
ING
-IN
STA
LL,C
HA
RG
E,O
RR
EM
OV
ET
HE
PO
WE
RM
OD
UL
EO
NLY
INA
NO
NH
AZ
AR
DO
US
LO
CA
TIO
NS
- LIN
STA
LL
AT
ION
,LE
CH
AR
GE
ME
NTA
INS
IQ
UE
LE
NL
ÈV
EM
EN
TD
UM
OD
UL
ED
AL
IME
NTA
TIO
NE
ST
PO
SS
IBL
EA
VE
RT
ISS
EM
EN
T'
''
UN
IQU
EM
EN
TE
ND
EH
OR
SD
ES
ZO
NE
SÀ
RIS
QU
ED
EX
PL
OS
ION
'
WA
RN
ING
-AD
DIT
ION
OR
SU
BS
TIT
UT
ION
OF
CO
MP
ON
EN
TS
MA
YIM
PA
IRIN
TR
INS
ICS
AF
ET
Y- L
AJO
UT
OU
LA
SU
BS
TIT
UT
ION
DE
CO
MP
OS
AN
TS
PE
UT
CO
MP
RO
ME
TT
RE
LA
SÉ
CU
RIT
ÉIN
TR
INS
ÈQ
UE
AV
ER
TIS
SE
ME
NT
'
WA
RN
ING
-B
EF
OR
EC
ON
NE
CT
ING
FF
pwr;D
ISC
ON
NE
CT
AL
LF
UR
TH
ER
PO
WE
RS
UP
PL
IES
AN
DT
ER
MIN
AT
OR
S-
AV
AN
TD
EC
ON
NE
CT
ER
Ffpw
r,DÉ
CO
NN
EC
TE
RD
AB
OR
DT
OU
TE
SL
ES
AL
IME
NTA
TIO
NS
ET
LE
SD
ISP
OS
ITIF
SD
ET
ER
MIN
AIS
ON
AV
ER
TIS
SE
ME
NT
'
WA
RN
ING
-C
ON
NE
CT
TH
ET
RE
XD
EV
ICE
CO
MM
UN
ICA
TO
RA
CC
OR
DIN
GT
OT
HE
SC
HE
ME
SIN
TH
ISC
ON
TR
OL
DR
AW
ING
-R
AC
CO
RD
ER
CE
TA
PP
AR
EIL
TR
EX
DE
VIC
EC
ON
NU
NIC
AT
OR
SE
LO
NL
EP
LA
NF
OU
RN
IA
VE
RT
ISS
EM
EN
T
WA
RN
ING
S
Trex Device C
omm
unicatorC
ontrol Draw
ing
Trex Device C
omm
unicator comm
unication module:
-HA
RT
-FO
UN
DA
TIO
Nfieldbus (F
ISC
Oand non-F
ISC
O)
Trex Device C
omm
unicator Plus com
munication m
odule:-H
AR
T-m
AM
eter-F
OU
ND
AT
ION
fieldbus (FIS
CO
and non-FIS
CO
)
WA
RN
ING
S:
AT
EX
/IEC
Ex:II2G
(1GD
) Ex ia [ia G
a] [ia Da
IIIC]IIC
T4 G
bcC
SA
us: Class I, D
ivision 1, Groups
A, B
, C and D
,T4
Class I, Z
one 1,AE
x ia [ia Ga] [ia D
aIIIC
]IICT
4 Gb
CS
A: E
x ia [ia Ga] [ia D
aIIIC
]IICT
4 GbM
odification/Nam
e
1234
Date
Date
Nam
eD
rawing-N
umber
Page
Pages
checkedissued
checked
12
34
56
7
Alle Rechte vorbehalten! Diese Zeichnung darf ohne unsere ausdrückliche Zustimmungweder vervielfältigt, noch Dritte zugänglich gemacht werden. Außerdem darf sie durch denEmpfänger oder durch Dritte nicht in anderer Art und Weise missbräuchlich verwertet werden.
2016 19 50 0
Trex Device C
omm
unicatorIS
Installation Draw
ingR
.S
TA
HL
HM
ISyste
ms G
mb
HA
dolf-Grim
me-A
llee 8D
- 50829 Köln (C
ologne), Germ
anyT
EL: +49 (0)221-76806-1000
FA
X: +49 (0)221-76806-4100
1510.06.2016 W
n10.06.2016 D
ue
Approvals and Certifications May 2017
14 Approvals and Certifications
Inp
ut Param
eter
Ui (V
max) = 30 V
dcIi (Im
ax) = 200 mA
Ci = 0 uF
Vt or Voc of loop m
ust be <= 30 Vdc
It or Isc of loop must be <= 200 m
A
Device does not add inductance to the loop
TAB
LE1 -
CS
AE
NT
ITY
PA
RA
ME
TE
RS
forH
AR
Tw
ithTrex D
evice Com
municator com
munication m
odule
Unclassified
Area
HA
RT
Connection
Terminals
Pow
erS
upply
HA
RT
Field Device
Trex Device C
omm
unicatorcom
munication m
odule
Unclassified
Area
Approved
Barrier or
Converter,
see Note 4
Pow
erS
upply
HA
RT
Field Device
+-
V
Device does not add capacitance to the loop
Li = 0 mH
Pi (P
max) = 1 W
Hazardous C
lassifiedA
rea
Hazardous C
lassifiedA
rea
Ou
tpu
t Parameter
Uo (Voc) = 1.89 V
dcIo (Isc) = 32 µAP
o = 61 µWC
a = 14.3 uFLa = 100 m
H
Approved
Barrier or
Converter,
see Note 4
Classification:
T4
Max.A
mbient tem
p:
HA
RT
Connection
Terminals
Trex Device C
omm
unicatorcom
munication m
odule
Current I/O
Voltage I/O
ZO
NE
1,IIC,T4
Class I, D
ivision 1,G
roupsA
, B, C
and D
ZO
NE
1,IIC,T4
Class I, D
ivision 1,G
roupsA
, B, C
and D
See Security
Ad
vice on
Page 14
Lire le avertissemen
t de sécu
rité14
sur la p
age
for device classification,see separate approvalof device
for device classification,see seperate approvalof device
12
34
56
78
910
Alle Rechte vorbehalten! Diese Zeichnung darf ohne unsere ausdrückliche Zustimmungweder vervielfältigt, noch Dritte zugänglich gemacht werden. Außerdem darf sie durch denEmpfänger oder durch Dritte nicht in anderer Art und Weise missbräuchlich verwertet werden.
Date
Date
Nam
eD
rawing-N
umber
Page
Pages
checkedissued
checked
2K
eine Änderung ohne P
rüfung durch Prüfstelle
(z.B.U
L,FM
,CS
A)!
No
chan
ges p
rior to
Certi�catio
n B
od
y app
roval
(e.g.
UL
,FM
,C
SA
)!
1234
Modification/N
ame
Connection draw
ing forH
AR
Tw
ithTrex D
evice Com
municator
comm
unication module
-20°CTa +50°C
<<
2016 19 50 0
Trex Device C
omm
unicatorIS
Installation Draw
ingR.
STA
HL
HM
ISystem
s Gm
bH
Adolf-G
rimm
e-Allee 8
D- 50829 K
öln (Cologne), G
ermany
TE
L: +49 (0)221-76806-1000FA
X: +49 (0)221-76806-4100
1510.06.2016 W
n
10.06.2016 Due
May 2017 Approvals and Certifications
Approvals and Certifications 15
Unclassified AreaHazardous Classified Area
FOUNDATION fieldbusConnection Terminals
Any ApprovedAssociatedApparatus,see Note 4
MULTIPLEFOUNDATIONfieldbus DEVICES( Number islimited by therequirement tomeet all otherIS requirementsfor the network. )
Ii (Imax) = 380 mA
Ci = 0 µF
Uo (Voc) = 1.89 Vdc
Io (Isc) = 32 µA
Li = 0 mH
Pi (Pmax) = 1.3 W
Leakage current:less than or equal to 50 µA
Classification: T4
Max. Ambient temp:
Po = 61 µW
Ui (Vmax) = 30 Vdc
For Non-FISCO installation
Ca = 14.3 µF
La = 100 mH
3Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
Trex Device Communicatorcommunication module
Connection drawing for FOUNDATIONfieldbus (Non-FISCO) on theTrex Device Communicatorcommunication module
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
TABLE 2 -CSA ENTITY PARAMETERS for FOUNDATION fieldbus (Non-FISCO)with Trex Device Communicator communication module
-20°C Ta +50°C<<
Any ApprovedFOUNDATION fieldbus
FISCO Device. See Note 5
Any ApprovedFOUNDATION fieldbus
FISCO Device. See Note 5
Any ApprovedFOUNDATION fieldbus
FISCO Device. See Note 5
Any ApprovedFOUNDATION fieldbus
FISCO Device. See Note 5
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
1
2
3
4
1 2 3 4 5 6 7
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Modification/Name Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
Trex Device Communicatorcommunication module
FOUNDATION fieldbusConnection Terminals
MULTIPLEFOUNDATIONfieldbus devices(Number islimited by therequirement tomeet all otherIS and FISCOrequirementsfor the network. )
Leakage current:less than or equal to 50 µA
Classification: T4
Max. Ambient temp:
For FISCO installation
Ii (Imax) = 215 mA(IIC)380 mA(IIB)
Uo (Voc) = 1.89 Vdc
Io (Isc) = 32 µA
Pi (Pmax) = 1.9 W(IIC)/5.3W(IIB)
Po = 61 µW
Ui (Vmax) = 30 Vdc
Unclassified AreaHazardous Classified Area
4
Ci = 0 µF
Li = 0 mH
Ca = 14.3 µF
La = 100 mH
Any ApprovedFISCO Associated
Apparatus,see Note 4
Any ApprovedFOUNDATION fieldbus
FISCO Device. See Note 5
Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
TABLE 3 -CSA ENTITY PARAMETERS for FOUNDATION fieldbus (FISCO)with Trex Device Communicator communication module
-20°C Ta +50°C<<
See Security Advice for the FISCOconcept on Page 15
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
Lire le avertissement de sécurité15pour le concept FISCO sur la page
Any ApprovedFOUNDATION fieldbus
FISCO Device. See Note 5
Any ApprovedFOUNDATION fieldbus
FISCO Device. See Note 5
Any ApprovedFOUNDATION fieldbus
FISCO Device. See Note 5
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
1
2
3
4
1 2 3 4 5 6 7
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Modification/Name Date
Date Name Drawing-Number Page
Pagescheckedissued
checked Connection drawing for FOUNDATIONfieldbus (FISCO) withTrex Device Communicatorcommunication module
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
Approvals and Certifications May 2017
16 Approvals and Certifications
FF
FFpwr
mA HART+ pwr HART
Input Parameter
Ui (Vmax) = 30 VdcIi (Imax) = 200 mA
Ci = 0 uF
Vt or Voc of loop must be <= 30 VdcIt or Isc of loop must be <= 200 mA
Device does not add inductance to the loop
TABLE 4 - CSA ENTITY PARAMETERS for HART with Trex Device Communicator Plus communication module
Connection drawing for HART withdevice Trex Device CommunicatorPlus communication module
Unclassified Area
HARTConnectionTerminals
PowerSupply
HARTField Device
Trex Device Communicator Pluscommunication module
Unclassified Area
ApprovedBarrier orConverter,see Note 4
PowerSupply
HARTField Device
+ - V
Device does not add capacitance to the loopLi = 0 mH
Pi (Pmax) = 1 W
Hazardous Classified Area
Hazardous Classified Area
Output Parameter
Uo (Voc) = 1.89 VdcIo (Isc) = 1.9 mAPo = 3.6 mWCa = 14.3 uFLa = 100 mH
ApprovedBarrier orConverter,see Note 4
FF
FFpwr
mA HART+ pwr HART
HARTConnectionTerminals
Trex Device Communicator Pluscommunication module
Voltage I/O
Classification: T4
Max. Ambient temp:-20°C Ta +50°C<<
Current I/O
ZONE 1, IIC, T4
Class I, Division 1,Groups A, B, C and D
ZONE 1, IIC, T4
Class I, Division 1,Groups A, B, C and D
for device classification,see separate approvalof device
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
1 2 3 4 5 6 7
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
5Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
1
2
3
4
Modification/Name
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
FF
FFpwr
mA HART+ pwr HART
Input Parameter
Ui (Vmax) = 30 VdcIi (Imax) = 200 mA
Ci = 0 uF
Vt or Voc of loop must be <= 30 VdcIt or Isc of loop must be <= 200 mA
Device does not add inductance to the loop
TABLE 5 - CSA ENTITY PARAMETERS for HART pwr with Trex Device Communicator Plus communication module
HART + pwrConnectionTerminals
HARTField Device
Trex Device Communicator Pluscommunication module
Device does not add capacitance to the loopLi = 0 mH
Pi (Pmax) = 1 W
Hazardous Classified Area
Output Parameter
Uo (Voc) = 25.69 VdcIo (Isc) = 105 mAPo = 668 mWCa = * nF
La 1000 µH 750 µH 500 µH 100 µH
Classification: T4
Max. Ambient temp:
Connection drawing for HART withTrex Device Communicator Pluscommunication module and internalpower supply
-20°C Ta +50°C<<
Current I/O
Ca 57 nF 64 nF 75 nF 102 nF
* Output parameter L/C
La = * µH
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
1 2 3 4 5 6 7 8 9 10
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
6Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
1
2
3
4
Modification/Name
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
May 2017 Approvals and Certifications
Approvals and Certifications 17
7Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
FF
FFpwr
mA HART+ pwr HART
Unclassified Area
mAConnectionTerminals
PowerSupply
Trex Device Communicator Pluscommunication module
Hazardous Classified Area
ApprovedBarrier orConverter,see Note 4
TABLE 6 - CSA ENTITY PARAMETERS for the mA Meter with Trex Device Communicator Plus communication module
Input Parameter
Ui (Vmax) = 30 VdcIi (Imax) = 200 mA
Ci = 0 uF
Vt or Voc of loop must be <= 30 VdcIt or Isc of loop must be <= 200 mA
Device does not add inductance to the loopDevice does not add capacitance to the loop
Li = 0 mH
Pi (Pmax) = 1 W
Output Parameter
Uo (Voc) = 0 VdcIo (Isc) = 0 mAPo = 0 mWCa = - uFLa = - mH
Classification: T4
Max. Ambient temp:-20°C Ta +50°C<<
ZONE 1, IIC, T4
Class I, Division 1,Groups A, B, C and D
HARTField Device
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
Modification/Name
1
2
3
4
Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
1 2 3 4 5 6 7 8 9 10
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Connection drawing for mA terminalson Trex Device Communicator Pluscommunication module
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
8Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
FF
FFpwr
mA HART+ pwr HART
Unclassified Area
mAConnectionTerminals
PowerSupply
Trex Device Communicator Pluscommunication module
Hazardous Classified Area
ApprovedBarrier orConverter,see Note 4
TABLE 7 - CSA ENTITY PARAMETERS for the mA Meter with Trex Device Communicator Plus communication module
Input Parameter
Ui (Vmax) = 30 VdcIi (Imax) = 200 mA
Ci = 0 uF
Vt or Voc of loop must be <= 30 VdcIt or Isc of loop must be <= 200 mA
Device does not add inductance to the loopDevice does not add capacitance to the loop
Li = 0 mH
Pi (Pmax) = 1 W
Output Parameter
Uo (Voc) = 0 VdcIo (Isc) = 0 mAPo = 0 mWCa = - uFLa = - mH
+ - test
FF
FFpwr
mA HART+ pwr HART
Unclassified Area
mAConnectionTerminals
PowerSupply
Trex Device Communicator Pluscommunication module
Hazardous Classified Area
ApprovedBarrier orConverter,see Note 4
+ - test
Classification: T4
Max. Ambient temp:-20°C Ta +50°C<<
HART Field Device(Test terminal referenced to +)
HART Field Device(Test terminal referenced to -)
ZONE 1, IIC, T4
Class I, Division 1,Groups A, B, C and D
ZONE 1, IIC, T4
Class I, Division 1,Groups A, B, C and D
for device classification,see separate approvalof device
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
Modification/Name
1
2
3
4
Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
1 2 3 4 5 6 7 8 9 10
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Connection drawing for mA terminalson the Trex Device Communicator Pluscommunication module
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
Approvals and Certifications May 2017
18 Approvals and Certifications
Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
FF
FFpwr
mA HART+ pwr HART
Unclassified Area
mAConnectionTerminals
PowerSupply
Trex Device Communicator Pluscommunication module
Hazardous Classified Area
ApprovedBarrier orConverter,see Note 4
TABLE 8 - CSA ENTITY PARAMETERS for the mA Meter with Trex Device Communicator Plus communication module
Input Parameter
Ui (Vmax) = 30 VdcIi (Imax) = 200 mA
Ci = 0 uF
Vt or Voc of loop must be <= 30 VdcIt or Isc of loop must be <= 200 mA
Device does not add inductance to the loopDevice does not add capacitance to the loop
Li = 0 mH
Pi (Pmax) = 1 W
Output Parameter
Uo (Voc) = 1.89 VdcIo (Isc) = 1.9 mAPo = 3.6 mWCa = 14.3 µFLa = 100 mH
- test
FF
FFpwr
mA HART+ pwr HART
Unclassified Area
mAConnectionTerminals
PowerSupply
Trex Device Communicator Pluscommunication module
Hazardous Classified Area
ApprovedBarrier orConverter,see Note 4
+ - test
Classification: T4
Max. Ambient temp:-20°C Ta +50°C<<
HART Field Device(referenced to -)
9
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
Connection drawing for mA terminalson the Trex Device Communicator Pluscommunication module
HARTTerminals
+
HARTTerminals
HART Field Device(referenced to +)
for device classification,see separate approvalof device
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
Modification/Name
1
2
3
4
Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
1 2 3 4 5 6 7 8 9 10
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
10Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
FF
FFpwr
mA HART+ pwr HART
Unclassified Area
mAConnectionTerminals
PowerSupply
Trex Device CommunicatorPlus communication module
Hazardous Classified Area
ApprovedBarrier orConverter,see Note 4
TABLE 9 - CSA ENTITY PARAMETERS for the mA Meter with Trex Device Communicator Plus communication module
Input Parameter
Ui (Vmax) = 30 VdcIi (Imax) = 200 mA
Ci = 0 uF
Vt or Voc of loop must be <= 30 VdcIt or Isc of loop must be <= 200 mA
Device does not add inductance to the loopDevice does not add capacitance to the loop
Li = 0 mH
Pi (Pmax) = 1 W
Output Parameter
Uo (Voc) = 1.89 VdcIo (Isc) = 1.9 mAPo = 3.6 mWCa = 14.3 µFLa = 100 mH
+ - test
Classification: T4
Max. Ambient temp:-20°C Ta +50°C<<
HARTField Device
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
Connection drawing for mA terminalson the Trex Device Communicator Pluscommunication module
HARTTerminals
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
Modification/Name
1
2
3
4
Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
1 2 3 4 5 6 7 8 9 10
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
May 2017 Approvals and Certifications
Approvals and Certifications 19
FF
FFpwr
mA HART+ pwr HART
Unclassified AreaHazardous Classified Area
FOUNDATION fieldbusConnection Terminals
Any ApprovedAssociatedApparatus,see Note 4
MULTIPLEFOUNDATIONfieldbus devices(Number islimited by therequirement tomeet all otherIS requirementsfor the network. )
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
Ii (Imax) = 380 mA
Ci = 0 µF
Uo (Voc) = 1.89 Vdc
Io (Isc) = 32 µA
Li = 0 mH
Pi (Pmax) = 1.3 W
Leakage current:less than or equal to 50 µA
Classification: T4
Max. Ambient temp:
Po = 61 µW
Ui (Vmax) = 30 Vdc
For Non-FISCO installation
Ca = 14.3 µF
La = 100 mH
11Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
Trex Device Communicator Pluscommunication module
Connection drawing for FOUNDATIONFieldbus (Non-FISCO) on theTrex Device Communicator Pluscommunication module
TABLE 10 -CSA ENTITY PARAMETERS for FOUNDATION fieldbus (Non-FISCO)with Trex Device Communicator Plus communication module
-20°C Ta +50°C<<
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
1
2
3
4
1 2 3 4 5 6 7
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Modification/Name Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
FF
FFpwr
mA HART+ pwr HART
Trex Device Communicator Pluscommunication module
FOUNDATION fieldbusConnection Terminals
Leakage current:less than or equal to 50 µA
Classification: T4
Max. Ambient temp:
For FISCO installation
Ii (Imax) = 215 mA(IIC)380 mA(IIB)
Uo (Voc) = 1.89 Vdc
Io (Isc) = 32 µA
Pi (Pmax) = 1.9 W(IIC)/5.3W(IIB)
Po = 61 µW
Ui (Vmax) = 30 Vdc
Unclassified AreaHazardous Classified Area
12
Ci = 0 µF
Li = 0 mH
Ca = 14.3 µF
La = 100 mH
Any ApprovedFISCO Associated
Apparatus,see Note 4
Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
TABLE 11 -CSA ENTITY PARAMETERS for FOUNDATION fieldbus (FISCO)with Trex Device Communicator Plus communication module
-20°C Ta +50°C<<
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
MULTIPLEFOUNDATIONfieldbus devices(Number islimited by therequirement tomeet all otherIS requirementsfor the network. )
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
for device classification,see separate approvalof device
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
See Security Advice for the FISCOconcept on Page 15
Lire le avertissement de sécurité15pour le concept FISCO sur la page
1
2
3
4
1 2 3 4 5 6 7
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Modification/Name Date
Date Name Drawing-Number Page
Pagescheckedissued
checked Connection drawing for FOUNDATIONfieldbus (FISCO) with deviceTrex device communicator Pluscommunication module
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
Approvals and Certifications May 2017
20 Approvals and Certifications
FF
FFpwr
mA HART+ pwr HART
Trex Device Communicator Pluscommunication module
Hazardous Classified Area
Ci = 231 nF
Uo (Voc) = 17.31 Vdc
Io (Isc) = 199 mA
Li = 0 mH
Leakage current:less than or equal to 50 µA
Classification: T4
Max. Ambient temp:
Po = 0.94 W
Ui (Vmax) = 17.5 Vdc
For FISCO and non-FISCO installation
Ca = * nF
La = * µH
13Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
Ii (Imax) = 380 mA
Pi (Pmax) = 1.3 W
See Security Advice on Page 14
Lire le avertissement de sécurité14sur la page
1
2
3
4
1 2 3 4 5 6 7
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Modification/Name Date
Date Name Drawing-Number Page
Pagescheckedissued
checked Connection drawing for FOUNDATIONfieldbus with device Trex DeviceCommunicator Plus communicationmodule and internal power supply
TABLE 12 -CSA ENTITY PARAMETERS for FOUNDATION fieldbus with interalpower supply with Trex Device Communicator Plus communication module
FOUNDATION Fieldbus Power plug
-20°C Ta +50°C<<
La 100 µH 50 µH 30 µHCa 19 nF 69 nF 115 nF
* Output parameter L/C
ZONE 1, IIC, T4
Class I, Division 1, Groups A, B, C and D
Any ApprovedFOUNDATION fieldbus
Device. See Note 5
FOUNDATION fieldbusConnection Terminals
for device classification,see separate approvalof device
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
14Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
nonincendive Equipment Associated ApparatusEquipment non- inflammable Appareil associé
Vmax Voc or Uo>Imax Isc or Io>Pi Po>Ci + Ccable Ca (or Co)<Li + Lcable La (or Lo)<
1. No revision to drawing prior to certification body.La modification des plans n'est pas autorisée sans l'accordde l'organisme notifié.
2. The Associated Apparatus must be NRTL approved in accordanceto IEC 60079-11.Les appareils associés doivent avoir une certification NRTL selon
.IEC 60079-11
3. Manufacturer’s installation drawing must be followedwhen installing associated apparatus.Les plans d´installation des produits des appareils associés doiventêtre respectés s'ils sont installés.
4. Interconnection of nonincendive equipment apparatus withassociated apparatus is allowed when the followingis true:L'interconnexion des équipements non- inflammable est correcteuniquement si les paramètres suivants sont respectés:
WARNING:- Substitution of components may impair Safety.
AVERTISSEMENT:La substitution de composants peut compromettre la sécurité.-
1.) Determination of maximum possible capacitance of cable:
Cmax = Ca - Ci (associated Apparatus)
Determination of maximum possible inductance of cable:
Lmax = La - Li (associated Apparatus)
2.) Determination of maximum possible cable length bycapacitance and inductance of cable:
length C = CmaxCcable (*1)
length L = LmaxLcable (*1)
3.) Determination of maximum length of cable:
length C or lenght L, whatever is less.
Calculation of cable length
Security AdvicesAvertissements de sécurité
5. for ATEX / IECEx: Installation should be in accordance withANSI/ISA RP12.06.01 “Installation of Intrinsically safesystems for Hazardous (classified) Locations” andthe National Electrical Code (ANSI/NFPA 70)Pour ATEX/ IECEx : l'installation doit être conforme à la normeANSI/ISA RP12.06.01 (installation de systèmes à sécurité intrinsèquepour les emplacements dangereux (classés) et le NationalElectrical Code (ANSI/NFPA 70)
for cCSAus: Installation should be in accordance with CanadianElectrical Code, CSA 22.1, Part 1.Pour cCSAAus : l'installation doit être conforme au code canadiende l´électricité CSA 22.1 partie 1.
1
2
3
4
1 2 3 4 5 6 7
Alle
Rech
te v
orb
ehalte
n! D
iese
Zeic
hnung d
arf
ohne u
nse
re a
usd
rück
liche Z
ust
imm
ung
weder
verv
ielfä
ltigt, n
och
Dritte z
ugänglic
h g
em
ach
t w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issb
räuch
lich v
erw
ert
et w
erd
en.
Modification/Name Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
Security Advices
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
May 2017 Approvals and Certifications
Approvals and Certifications 21
FISCO CONCEPT
15
THE FISCO CONCEPT ALLOWS INTERCONNECTION OF INTRINSICALLY SAFE APPARATUS TOASSOCIATED APPARATUS NOT SPECIALLY EXAMINED IN SUCH COMBINATION. THE CRITERIAFOR INTERCONNECTION IS THAT THE VOLTAGE (Ui OR Vmax), THE CURRENT (Ii OR Imax) ANDTHE POWER (Pi OR Pmax) WHICH AN INTRINSICALLY SAFE APPARATUS CAN RECEIVE AND REMAININTRINSICALLY SAFE CONSIDERING FAULTS, MUST BE EQUAL OR GREATER THAN VOLTAGE (Uo,Voc OR Vt), THE CURRENT (Io, Isc OR It) AND THE POWER (Po OR Pmax) LEVELS WHICH CANBE DELIVERED BY THE ASSOCIATED APPARATUS, CONSIDERING FAULTS AND APPLICABLEFACTORS. IN ADDITION, THE MAXIMUM UNPROTECTED CAPACITANCE (Ci) AND THE INDUCTANCE(Li) OF EACH APPARATUS (OTHER THAN THE TERMINATION) CONNECTED TO THE FIELDBUSMUST BE LESS THAN OR EQUAL TO 5 nF and 10 uH RESPECTIVELY.
IN EACH SEGMENT, ONLY ONE ACTIVE DEVICE, NORMALLY THE ASSOCIATED APPARATUS, ISALLOWED TO PROVIDE THE NECESSARY ENERGY FOR THE FIELDBUS SYSTEM. THE VOLTAGEUo (OR Voc OR Vt) OF THE ASSOCIATED APPARATUS IS LIMITED TO A RANGE OF 14 V TO24 Vdc ALL OTHER EQUIPMENT CONNECTED TO THE BUS CABLE HAS TO BE PASSIVE,MEANING THAT THEY ARE NOT ALLOWED TO PROVIDE ENERGY TO THE SYSTEM, EXCEPT ALEAKAGE CURRENT OF 50uA FOR EACH CONNECTED DEVICE. SEPARATELY POWEREDEQUIPMENT NEEDS GALVANIC ISOLATION TO ENSURE THAT THE INTRINSICALLY SAFEFIELDBUS CIRCUIT REMAINS PASSIVE.
THE CABLE USED TO INTERCONNECT DEVICES NEEDS TO HAVE THE PARAMETERS IN THEFOLLOWING RANGE:
Loop Resistance R’: 15.....150 Ohm/kmInductance per unit length L’: 0.4.....1 mH/kmCapacitance per unit length C’: 80.....200 nF/kmC’ = C’ line/line + 0.5C’ line/screen, if both lines are floating, orC’ = C’ line/line + C’ line/screen, if the screen is connected to one lineLength of trunk cable: less than or equal to 1000mLength of spur cable: less than or equal to 30mLength of spur splice: less than or equal to 1m
AT EACH END OF THE TRUNK CABLE AN APPROVED INFALLIBLE LINE TERMINATION WITH THEFOLLOWING PARAMETERS IS SUITABLE:
R = 90.....100 Ohm C = 0.....2.2 uF
ONE OF THE ALLOWED TERMINATIONS MIGHT ALREADY BE INTEGRATED IN THE ASSOCIATEDAPPARATUS. THE NUMBER OF PASSIVE APPARATUS CONNECTED TO THE BUS SEGMENT IS NOTLIMITED DUE TO I.S. REASONS. IF THE ABOVE RULES ARE RESPECTED, UP TO A TOTALLENGTH OF 1000 m (SUM OF TRUNK AND ALL SPUR CABLES) OF CABLE IS PERMITED.THE INDUCTANCE AND THE CAPACITANCE OF THE CABLE WILL NOT IMPAIR THEINTRINSIC SAFETY OF THE INSTALLATION.
Keine Änderung ohne Prüfung durch Prüfstelle(z.B. UL, FM, CSA)!
No changes prior to Certification Body approval(e.g. UL, FM, CSA)!
10.02.15 DüEPL marking for CSA added
Modification/Name
1
2
3
4
Date
Date Name Drawing-Number Page
Pagescheckedissued
checked
1 2 3 4 5 6 7 8 9 10
Alle
Rechte
vorb
ehalten! D
iese Z
eic
hnung d
arf
ohne u
nsere
ausdrü
cklic
he Z
ustim
mung
weder
verv
ielfältig
t, n
och D
ritte z
ugänglic
h g
em
acht w
erd
en.A
uß
erd
em
darf
sie
durc
h d
en
Em
pfä
nger
oder
durc
h D
ritte n
icht in
andere
rA
rt u
nd W
eis
e m
issbrä
uchlic
h v
erw
ert
et w
erd
en.
FISCO Concept
2016 19 50 0
Trex Device CommunicatorIS Installation Drawing
R. STAHL HMI Systems GmbHAdolf-Grimme-Allee 8D- 50829 Köln (Cologne), GermanyTEL: +49 (0)221-76806-1000FAX: +49 (0)221-76806-4100
1510.06.2016 Wn
10.06.2016 Due
Approvals and Certifications May 2017
22 Approvals and Certifications
Approvals and CertificationsTREX-0045-0001 Rev 3
May 2017
Emerson12001 Technology DriveEden Prairie, MN 55344 USAT 1(952)828-3000www.Emerson.com
©2017, Emerson.The contents of this publication are presented forinformational purposes only, and while every efforthas been made to ensure their accuracy, they arenot to be construed as warranties or guarantees,express or implied, regarding the products orservices described herein or their use orapplicability. All sales are governed by our termsand conditions, which are available on request. Wereserve the right to modify or improve the designsor specifications of our products at any timewithout notice.All rights reserved. The Emerson logo is atrademark and service mark of Emerson Electric Co.All other marks are property of their respectiveowners.