an in vitro acoustic analysis and comparison of popular

12
© 2019 Weiss et al. This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms. php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php). Medical Devices: Evidence and Research 2019:12 41–52 Medical Devices: Evidence and Research Dovepress submit your manuscript | www.dovepress.com Dovepress 41 ORIGINAL RESEARCH open access to scientific and medical research Open Access Full Text Article http://dx.doi.org/10.2147/MDER.S186076 An in vitro acoustic analysis and comparison of popular stethoscopes Daniel Weiss 1,2 Christine Erie 1,2 Joseph Butera III 2 Ryan Copt 2 Glenn Yeaw 2 Mark Harpster 2 James Hughes 2,3 Deeb N Salem 4 1 Department of Medicine, Charles E. Schmidt College of Medicine, Florida Atlantic University, Boca Raton, FL, USA; 2 Bongiovi Medical & Health Technologies, Inc., Port St Lucie, FL, USA; 3 Department of Surgery, University of Mississippi Medical Center, Jackson, MS, USA; 4 Department of Medicine, Tufts Medical Center, Boston, MA, USA Purpose: To compare the performance of various commercially available stethoscopes using standard acoustic engineering criteria, under recording studio conditions. Materials and methods: Eighteen stethoscopes (11 acoustic, 7 electronic) were analyzed using standard acoustic analysis techniques under professional recording studio conditions. An organic phantom that accurately simulated chest cavity acoustics was developed. Test sounds were played via a microphone embedded within it and auscultated at its surface by the stethoscopes. Recordings were made through each stethoscope’s binaurals and/or downloaded (electronic models). Recordings were analyzed using standard studio techniques and software, including assessing ambient noise (AMB) rejection. Frequency ranges were divided into those corresponding to various standard biological sounds (cardiac, respiratory, and gastrointestinal). Results: Loudness and AMB rejection: Overall, electronic stethoscopes, when set to a maximum volume, exhibited greater values of perceived loudness compared to acoustic stethoscopes. Sig- nificant variation was seen in AMB rejection capability. Frequency detection: Marked variation was also seen, with some stethoscopes performing better for different ranges (eg, cardiac) vs others (eg, gastrointestinal). Conclusion: The acoustic properties of stethoscopes varied considerably in loudness, AMB rejection, and frequency response. Stethoscope choice should take into account clinical condi- tions to be auscultated and the noise level of the environment. Keywords: stethoscope, analysis, acoustic, noise, signal Introduction The essential concept of a medical stethoscope has remained unchanged since its invention by René Laennec in 1816. The form of the stethoscope using flexible bin- aural tubes greatly improved the practicality of the instrument in the 1850s. While advances in materials, electronics, and construction methods have refined the design, no groundbreaking changes have been made since then. In parallel, as new imaging technologies appear, reliance on the human ear as a diagnostic device is decreasing. However, when imaging is not possible or is unavailable, auscultation can provide a wealth of information in a noninvasive and immediate manner. Choosing the optimal stethoscope is vital for accurate auscultation and patient care. For a stethoscope to function well, it must auscultate clinically relevant sounds and reject unwanted sounds (eg, environmental noise, nonrelevant biological sounds [BIO]). Addi- tionally, the frequency range of clinically relevant BIO varies, making one stethoscope ideal for a specific specialty and less ideal for another. 1 The stethoscope guides available Correspondence: Daniel Weiss Bongiovi Medical & Health Technologies, Inc., 649 SW Whitmore Ave, Port St Lucie, FL 34984, USA Tel +1 772 879 0578 Email [email protected]

Upload: others

Post on 16-Oct-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: an in vitro acoustic analysis and comparison of popular

© 2019 Weiss et al. This work is published and licensed by Dove Medical Press Limited. The full terms of this license are available at https://www.dovepress.com/terms. php and incorporate the Creative Commons Attribution – Non Commercial (unported, v3.0) License (http://creativecommons.org/licenses/by-nc/3.0/). By accessing the work

you hereby accept the Terms. Non-commercial uses of the work are permitted without any further permission from Dove Medical Press Limited, provided the work is properly attributed. For permission for commercial use of this work, please see paragraphs 4.2 and 5 of our Terms (https://www.dovepress.com/terms.php).

Medical Devices: Evidence and Research 2019:12 41–52

Medical Devices: Evidence and Research Dovepress

submit your manuscript | www.dovepress.com

Dovepress 41

O R i g i n a l R E s E a R c h

open access to scientific and medical research

Open Access Full Text Article

http://dx.doi.org/10.2147/MDER.S186076

an in vitro acoustic analysis and comparison of popular stethoscopes

Daniel Weiss1,2

christine Erie1,2

Joseph Butera iii2

Ryan copt2

glenn Yeaw2

Mark harpster2

James hughes2,3

Deeb n salem4

1Department of Medicine, charles E. schmidt college of Medicine, Florida atlantic University, Boca Raton, Fl, Usa; 2Bongiovi Medical & health Technologies, inc., Port st lucie, Fl, Usa; 3Department of surgery, University of Mississippi Medical center, Jackson, Ms, Usa; 4Department of Medicine, Tufts Medical center, Boston, Ma, Usa

Purpose: To compare the performance of various commercially available stethoscopes using

standard acoustic engineering criteria, under recording studio conditions.

Materials and methods: Eighteen stethoscopes (11 acoustic, 7 electronic) were analyzed

using standard acoustic analysis techniques under professional recording studio conditions.

An organic phantom that accurately simulated chest cavity acoustics was developed. Test

sounds were played via a microphone embedded within it and auscultated at its surface by the

stethoscopes. Recordings were made through each stethoscope’s binaurals and/or downloaded

(electronic models). Recordings were analyzed using standard studio techniques and software,

including assessing ambient noise (AMB) rejection. Frequency ranges were divided into those

corresponding to various standard biological sounds (cardiac, respiratory, and gastrointestinal).

Results: Loudness and AMB rejection: Overall, electronic stethoscopes, when set to a maximum

volume, exhibited greater values of perceived loudness compared to acoustic stethoscopes. Sig-

nificant variation was seen in AMB rejection capability. Frequency detection: Marked variation

was also seen, with some stethoscopes performing better for different ranges (eg, cardiac) vs

others (eg, gastrointestinal).

Conclusion: The acoustic properties of stethoscopes varied considerably in loudness, AMB

rejection, and frequency response. Stethoscope choice should take into account clinical condi-

tions to be auscultated and the noise level of the environment.

Keywords: stethoscope, analysis, acoustic, noise, signal

IntroductionThe essential concept of a medical stethoscope has remained unchanged since its

invention by René Laennec in 1816. The form of the stethoscope using flexible bin-

aural tubes greatly improved the practicality of the instrument in the 1850s. While

advances in materials, electronics, and construction methods have refined the design,

no groundbreaking changes have been made since then. In parallel, as new imaging

technologies appear, reliance on the human ear as a diagnostic device is decreasing.

However, when imaging is not possible or is unavailable, auscultation can provide a

wealth of information in a noninvasive and immediate manner.

Choosing the optimal stethoscope is vital for accurate auscultation and patient care.

For a stethoscope to function well, it must auscultate clinically relevant sounds and reject

unwanted sounds (eg, environmental noise, nonrelevant biological sounds [BIO]). Addi-

tionally, the frequency range of clinically relevant BIO varies, making one stethoscope

ideal for a specific specialty and less ideal for another.1 The stethoscope guides available

correspondence: Daniel WeissBongiovi Medical & health Technologies, inc., 649 sW Whitmore ave, Port st lucie, Fl 34984, UsaTel +1 772 879 0578Email [email protected]

Journal name: Medical Devices: Evidence and ResearchArticle Designation: Original ResearchYear: 2019Volume: 12Running head verso: Weiss et alRunning head recto: Acoustic analysis of popular stethoscopesDOI: http://dx.doi.org/10.2147/MDER.S186076

Page 2: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12submit your manuscript | www.dovepress.com

Dovepress

Dovepress

42

Weiss et al

focus on choosing a stethoscope on the basis of chest piece

material, tubing durability, and ear piece comfort.2 Often, the

decision to buy a particular stethoscope is influenced by its

appearance rather than its acoustic properties and capability.3

Price may be a factor, with stethoscopes ranging from under

US $100–1,000. Digital stethoscopes may be better suited to

situations where amplification of the sound is important, or

where storage and transmission of auscultated sounds (such

as in telemedicine) are required. This study compared the

performance of the various leading commercially available

stethoscopes using standard acoustic engineering criteria,

under recording studio conditions. To maintain uniformity

and control of testing conditions, performance was assessed

using a biological phantom with acoustic properties similar

to those of the human thorax.

Materials and methodsacoustic phantom creationTo accurately analyze and compare various stethoscopes in

vitro, care was taken to simulate and normalize the entire

auscultation system. As no “gold standard” exists to com-

pare stethoscopes against each other, an organic phantom

that accurately simulated chest cavity acoustics was created

( Figure 1). The organic phantom consisted of a lightweight

plastic container (Ziploc Tupperware; SC Johnson, Racine,

WI, USA) comprising an open cell foam (Carry Cases Plus,

Paterson, NJ, USA), an acoustically transparent plastic grid

(3D printed at the lab), castor oil (Mineral oil; CVS, Woon-

socket, RI, USA), and a 0.5 cm layer of ballistics gel (Perma-

Gel Personal Ballistic Gel Test Kit; Perma-Gel, Inc.; Albany,

OR, USA) as the “skin.” Food-grade castor oil was chosen

because its sound wave propagation speed (1,490 m/s) is

similar to that of the body.4 A 5 W transducer (DAEX25QLP;

Dayton Audio, Springboro, OH, USA) driven by a 100 W

Class A amplifier (RA300; Alesis, Cumberland, RI, USA)

was embedded within the organic phantom to reproduce

test sounds of varying frequency ranges corresponding to

specific bodily sounds.

The stethoscopes evaluated in this study were auscultated

at the phantom’s gel surface (Figure 2). The stethoscope head

was placed on the surface of the phantom with a standard-

ized weight of 227 gm corresponding to the typical pressure

applied by a hand when auscultating a chest.

The pressure of the chest piece on the phantom must be

normalized within a study cohort to ensure electrical imped-

ance of the transducer system is equal for all test subjects.

A weight of 227 g was calibrated for each stethoscope by

placing a digital scale beneath the chest piece, and then filling

the container with an appropriate number of copper balls to

equal the total target weight for the study.

A total of 18 stethoscopes were evaluated (Table 1). Of

those, 11 were acoustic and 7 were electronic stethoscopes.

Acoustic stethoscope modes include diaphragm (D) mode

or bell (B) mode, while electronic stethoscope modes can

Figure 1 Phantom container system.Notes: a simulation was constructed using materials to mimic the biological structure and sounds reproduced by an organic chest cavity. layers of foam were used to support and stabilize the transducer. The castor oil used shared similar sound wave propagation properties just like the human body.

Plastic container Ballistics gel “skin” 0.5 cm

Foam layer 1 cm

Foam layer 4 cm

Foam layer 2 cm

Support grid

Transducer

Page 3: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12 submit your manuscript | www.dovepress.com

Dovepress

Dovepress

43

acoustic analysis of popular stethoscopes

vary among D, B, wide (W), high (H), and low (L) modes.

Each stethoscope was recorded in all of its available modes.

With a system created to mimic a true chordate body, the

final consideration was to ensure that wave propagation and

transmission is possible across all relevant frequency ranges.

Frequency ranges are provided in Table 2. Figure 3 shows the

recorded frequency responses for each stethoscope. It is clear

that there is a significant transfer of all relevant frequencies

through the system. Since this paper seeks only to assess the

relative differences in response between various stethoscopes,

it is not necessary that there be a flat frequency response

through the phantom. Rather, it is only necessary that the

shape of the phantom’s frequency response curve remains

consistent between individual stethoscope tests.

Figure 2 Experimental setup.Note: To minimize frequency interference, a shock-absorbing structure suspended the phantom container using wide elastic bands.

Weight container

Elastic suspension

Support structure

Stethoscope Phantomcontainer

Stabilizing armature

Table 1 stethoscope models evaluated

Acoustic stethoscopes D B Electronic stethoscopes D B W H L

adscope Platinum X aDc adscope acc X X X Disposable Blue X cardionic E scope Disposable Yellow X Jabes analyzer X X X heine gamma 3.2 X X littman 3200 X X X littman cardiology iii X X Thinklabs da32a+ X X littman cardiology iV X X Thinklabs One X Mabis legacy sprague lc X X Welch allyn Meditron acc X XMavis spectrum X Omron sprague Rappaport X X Ultrascope X Welch allyn harvey Elite X X

Notes: The acoustic and electronic stethoscopes are listed, with each mode tested indicated by an “X.” D stands for diaphragm mode; B stands for bell mode, while electronic stethoscope modes can vary among diaphragm, bell, wide (W), high (h) and low (l) modes.

Test soundsAssessment of the harmonics within the phantom system

included a 20 Hz–20 kHz sine wave sweep reproduced

by the embedded transducer and recorded at the surface

of the phantom. All stethoscopes evaluated in this study

were tested with calibrated audio levels to ensure that

each stethoscope received the same acoustic energy at the

auscultation point.

Pink noise (acoustical energy distributed uniformly by

an octave throughout the audio spectrum) is useful for deter-

mining the broadband performance of an acoustic system for

human use. To assess a stethoscope’s frequency response of

BIO, pink noise was reproduced through the transducer and

auscultated at the phantom’s surface under quiet conditions. To

Page 4: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12submit your manuscript | www.dovepress.com

Dovepress

Dovepress

44

Weiss et al

assess a stethoscope’s ability to reject ambient noise (AMB), a

separate recording was then created by reproducing pink noise

through loudspeakers in the room at 80 dB sound pressure

level. While auscultating at the phantom’s surface, no sound

was reproduced through the phantom’s embedded transducer.

Microphones (TC-30; Earthworks, Glimanton, NH,

USA) were attached to both binaurals and simultaneously

recorded. All recordings were made using studio quality

transformerless mic preamps (C84; Seventh Circle Audio,

Oakland, CA, USA) and analog-to-digital conversion method

Table 2 clinically relevant frequency ranges

Frequency ranges (Hz) Relevant bodily sounds and pathologies

Full range general loudness capabilitysynthesized heart sound a relatable metric for general clinical use20–200 standard cardiac sounds (s1 + s2), gallops (s3 + s4), innocent heart murmurs, mitral stenosis200–700 Ventricular septal defect, atrial septal defect, aortic stenosis, pulmonary stenosis, tetralogy of Fallot, mitral

regurgitation, patent ductus arteriosus, pulmonary crackles, rhoncus, interstitial pulmonary fibrosis, arterial bruits, egophony

700–2,500 general respiratory sounds, general wheezes100–1,000 general gastrointestinal2,500–8,000 (Partial) blood clots, (partial) swallow sounds, (partial) tracheal sounds

Note: Frequency ranges listed with corresponding relevant bodily sounds and clinical pathologies.

Figure 3 Frequency responses for each stethoscope.Note: The phantom is capable of transmitting the full range of audible frequencies.

14172023262932353841444750535659

6568717477808386899295

101

8

SpectraFoo

16 31 62 125 250 500 1k

62

(Aurora 8; Lynx Studio Technology, Inc., Costa Mesa, CA,

USA). Audio recording and playback were accomplished

using Sonar X2 professional recording software (Cakewalk,

Inc., Boston, MA, USA).

All recordings obtained were uncompressed with a 48

kHz sample rate and 24-bit word length. This format was

supposed to be of sufficient resolution for the frequency

ranges reproduced by the stethoscopes in the study. Electronic

stethoscopes were recorded using binaural microphones to

simulate human use, as well as their own electrical output

Page 5: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12 submit your manuscript | www.dovepress.com

Dovepress

Dovepress

45

acoustic analysis of popular stethoscopes

to represent a machine interface, if available. All electronic

stethoscopes were recorded at the maximum volume.

Data analysisloudness and pink noise measurementsPink noise results were analyzed for ITU BS 1170-3 per-

ceived loudness using Adobe Audition CS6 (Adobe Sys-

tems, Inc., San Jose, CA, USA). These measurements are

represented in Loudness Units relative to digital Full Scale

(LUFS). This unit takes into account the nonlinear frequency

response of human hearing. It should be noted that a 10 dB or

LUFS decrease or increase in volume correlates with halving

or doubling of perceived loudness, respectively.

aMB rejectionThe degree of AMB rejection was derived by subtracting the

LUFS value of AMB from the LUFS value of BIO (DLUFS)

(Figure 4). The loudness of any sample auscultated from the

phantom is arbitrary and does not have a numerical reference.

The AMB rejection amount of a stethoscope should only be

compared to the rejection amount of other stethoscopes in

this study. No AMB rejection (0 DLUFS) would indicate that

Figure 4 aMB rejection.Note: The spectrogram depicts pink noise signals auscultated (left column) using a Thinklabs stethoscope and the aMB detected (right column) by microphones attached to both binaurals.Abbreviations: aMB, ambient noise; l, left; R, right.

dB

–9

–9–3

–8

dB

–9

–9–3

Hz10 k

4 k

2 k

1 k

600

400

200100

Hz10 k

4 k

2 k

1 k

600

400

200100

–8

L

R

Page 6: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12submit your manuscript | www.dovepress.com

Dovepress

Dovepress

46

Weiss et al

the stethoscope system detects 80 dB sound pressure level of

broadband noise in the atmosphere at the same level as the

test signal presented in the phantom.

clinically relevant frequency rangesThe full spectrum pink noise samples were divided into

frequency ranges derived from the literature5–24 to assess the

performance of each stethoscope for specific clinical condi-

tions (Table 2). To accomplish this, BIO and AMB recordings

were duplicated and then band-pass filtered into each range.

LUFS measurements were taken of each sample.

While the spectrographic analysis is useful for an exten-

sive analysis into the specifics of a stethoscope’s frequency

response, researchers focused on frequency band loudness

measurements as a more useful metric for correlation with

clinical utility.

ResultsOverall perceived loudness and aMB rejectionOverall, electronic stethoscopes, when set to a maximum

volume, exhibited greater values of perceived loudness com-

pared to acoustic stethoscopes. The three loudest electronic

stethoscopes were the Thinklabs ds32a+ (–20.66 LUFS in B

mode), the Jabes Analyzer (–23.91 LUFS in B mode), and the

Littmann 3200 (−22.30 LUFS in B mode). Of the electronic

stethoscopes evaluated, the Littmann 3200 exhibited the most

AMB rejection (21.40 DLUFS).

The loudest acoustic stethoscopes were the Welch Allyn

Harvey Elite (–39.02 LUFS in B mode), the Littmann Cardi-

ology III (–36.52 LUFS in D mode), and the Heine Gamma

3.2 (−38.55 LUFS in B mode). Of the acoustic stethoscopes

analyzed in this study, the Heine Gamma had the best AMB

rejection (8.00 DLUFS).

Frequency range-specific resultsgastrointestinal sounds (97–1,034 hz)Among the acoustic stethoscopes, Littman Cardiology IV

(45.982 LUFS in B mode), Mabis Spectrum (44.51 LUFS

in D mode), and Littman Cardiology III (41.95 LUFS in D

mode) demonstrated the highest perceived loudness to sounds

within the frequency range correlating to gastrointestinal

sounds (Figure 5). Overall, the electronic stethoscopes

examined were more sensitive to sounds in the 97–1,034

Hz frequency range than most of the acoustic stethoscopes.

Many of the electronic stethoscope models similarly per-

ceived loudness values in this frequency range (39.486–32.6).

The electronic stethoscopes which demonstrated the most Figu

re 5

sen

sitiv

ity t

o g

i sou

nds.

Not

e: T

he p

erce

ived

loud

ness

of a

cous

tic s

teth

osco

pes

(left)

and

ele

ctro

nic

stet

hosc

opes

(ri

ght)

to

freq

uenc

y ra

nge

97–1

,034

hz

was

mea

sure

d to

com

pare

the

ir s

ensi

tivity

to

gene

ral g

i sou

nds.

Abb

revi

atio

ns: g

i, ga

stro

inte

stin

al; l

UFs

, lou

dnes

s U

nits

rel

ativ

e to

dig

ital F

ull s

cale

.

05101520LUFS + 70 DB

253035404550

Acou

stic

ste

thos

cope

sen

sitiv

ity to

GI s

ound

s

Stet

hosc

ope

Littman Cardiology IV-B

Mabis Spectrum-D

Littman Cardiology III-D

Ultrascope-D

Littman Cardiology IV-D

Littman Cardiology III-B

Heine Gamma 3.2-D

Heine Gamma 3.2-B

Welch Allyn Harvey Elite-B

Disposable Blue-D

Adscope Platinum-D

Omron Sprague Rappaport-D

Welch Allyn Harvey Elite-D

Mabis Legacy Sprague LC-Bx

Omron Sprague Rappaport-B

Mabis Legacy Sprague LC-Dx

Disposable Yellow-D

Elec

troni

c st

etho

scop

e se

nsiti

vity

to G

I sou

nds

0

Jabes Analyzer-D

ADC Adscope Acc-D

ADC Adscope Acc-W

Jabes Analyzer-B

Jabes Analyzer-W

Littman 3200-W

Littman 3200-D

Thinklabs da32a+-D-NA

Welch Allyn Meditron Acc-L

Welch Allyn Meditron Acc-H

ADC Adscope Acc-B

Thinklabs da32a+-B-NA

Cardionics EScope

Thinklabs One-W

Littman 3200-B

5101520

Loudness

253035404550

Stet

hosc

ope

Page 7: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12 submit your manuscript | www.dovepress.com

Dovepress

Dovepress

47

acoustic analysis of popular stethoscopes

perceived loudness included the Jabes Analyzer (39.486 in

D mode), ADC Adscope Acc (39.466 in D mode), and the

ADC Adscope Acc (39.408 in W mode).

The Littman 3200, in all three of its modes, exhibited the

most noise rejection (Figure 6) of all stethoscopes. Overall,

the noise rejection capability of the acoustic stethoscopes

was superior to the electronic stethoscope noise rejection.

Among the acoustic stethoscopes, the Welch Allyn Harvey

Elite (16.816 LUFS in B mode) has the best noise rejection

followed closely by the Littmann Cardiology III (16.44 LUFS

in B mode) and the Heine Gamma 3.2 (16.202 LUFS in D

mode). Stethoscopes with relatively poor AMB rejection at

frequency 97–1,034 Hz include the Omron Sprague Rappa-

port (1.776 LUFS in B mode), ADC Adscope Acc (2.824 in

B mode), and Welch Allyn Meditron Acc (–3.884 in H mode

and –11.968 in L mode).

cardiac sounds (22–281 hz: s1–s4 murmurs, mitral stenosis [Ms])Within the 22–281 Hz frequency range, the acoustic stetho-

scopes Mabis Spectrum (in 44.948 LUFS in D mode), Litt-

man Cardiology IV (42.168 LUFS in B mode, 42.048 LUFS

in D mode), and Littman Cardiology III (41.55 LUFS in D

mode) had the highest loudness values. Of the electronic

stethoscopes, the Littman 3200 (39.416 in the B mode),

Thinklabs One (38.29 in W mode), and Welch Allyn Meditron

Acc (37.456 in H mode) exhibited the highest perceived

loudness (Figure 7).

Of all stethoscopes evaluated, the Littman 3200 rejected

the most noise within the frequency range associated with

S1–S4, innocent murmurs, and MS cardiac sounds. However,

many of the electronic stethoscopes exhibited poor noise

rejection. Of the electronic stethoscopes, the Welch Allyn

Meditron Acc (–19.004 in L mode) rejected the least noise.

The Omron Sprague Rappaport (–18.198 in B mode) and the

Mabis Legacy Sprague LC (–12.704 in B mode) are acoustic

stethoscopes which exhibited poor noise rejection (Figure 8).

cardiac sounds (205–775 hz: ventricular septal defect [VsD], atrial septal defect [asD], aortic stenosis [as], and pulmonic stenosis [Ps])Among acoustic and electronic stethoscopes, the Littman

Cardiology IV (44.324 LUFS in B mode) exhibited the

most perceived loudness of sounds within frequencies

205–775 Hz. In this frequency range, many of the electronic

stethoscopes show a similar performance, with the Jabes

Analyzer (39.404 in D mode) showing the most sensitivity

to cardiac pathology sounds such as VSD, ASD, AS, and PS. Figu

re 6

noi

se r

ejec

tion

of g

i sou

nds.

Not

e: n

oise

rej

ectio

n of

gi s

ound

s. a

cous

tic (

left)

and

ele

ctro

nic

(rig

ht)

stet

hosc

ope

nois

e re

ject

ion

was

mea

sure

d at

freq

uenc

y ra

nges

(97

–1,0

34 h

z) o

f gi s

ound

s.A

bbre

viat

ions

: gi,

gast

roin

test

inal

; lU

Fs, l

oudn

ess

Uni

ts r

elat

ive

to d

igita

l Ful

l sca

le.

Welch Allyn Harvey Elite-B

Littman Cardiology III-B

Heine Gamma 3.2-D

Heine Gamma 3.2-B

Mabis Spectrum-D

Littman Cardiology III-D

Ultrascope-D

Littman Cardiology IV-B

Disposable Blue-D

Adscope Platinum-D

Omron Sprague Rappaport-D

Littman Cardiology IV-D

Mabis Legacy Sprague LC-Bx

Mabis Legacy Sprague LC-Dx

Welch Allyn Harvey Elite-D

Disposable Yellow-D

Omron Sprague Rappaport-B

05 –5 –10 –15 –20101520LUFS + 70 DB

25

Acou

stic

ste

thos

cope

noi

se re

ject

ion

to G

I sou

nds

Stet

hosc

ope

Littman 3200-B

Littman 3200-W

Littman 3200-D

ADC Adscope Acc-W

Cardionics EScope

Thinklabs da32a+-B-NA

Jabes Analyzer-B

Jabes Analyzer-D

Jabes Analyzer-W

ADC Adscope Acc-D

Thinklabs da32a+-D-NA

Thinklabs One-W

ADC Adscope Acc-B

Welch Allyn Meditron Acc-H

Welch Allyn Meditron Acc-L

05 –5 –10 –15 –2010152025Loudness

Elec

troni

c st

etho

scop

e no

ise

reje

ctio

n to

GI s

ound

s

Stet

hosc

ope

Page 8: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12submit your manuscript | www.dovepress.com

Dovepress

Dovepress

48

Weiss et al

Figu

re 7

sen

sitiv

ity t

o ca

rdia

c so

unds

22–

281

hz.

Not

e: T

he p

erce

ived

loud

ness

of a

cous

tic s

teth

osco

pes

(left)

and

ele

ctro

nic

stet

hosc

opes

(ri

ght)

to

freq

uenc

y ra

nge

22–2

81 h

z w

as m

easu

red

to c

ompa

re t

heir

sen

sitiv

ity t

o ca

rdia

c so

unds

.A

bbre

viat

ion:

lU

Fs, l

oudn

ess

Uni

ts r

elat

ive

to d

igita

l Ful

l sca

le.

Mabis Spectrum-D

Littman Cardiology IV-B

Littman Cardiology IV-D

Littman Cardiology III-D

Disposable Blue-D

Ultrascope-D

Welch Allyn Harvey Elite-B

Heine Gamma 3.2-B

Littman Cardiology III-B

Heine Gamma 3.2-D

Adscope Platinum-D

Omron Sprague Rappaport-D

Welch Allyn Harvey Elite-D

Mabis Legacy Sprague LC-Dx

Omron Sprague Rappaport-B

Mabis Legacy Sprague LC-Bx

Disposable Yellow-D

05101520LUFS + 70 DB

253035404550

Acou

stic

ste

thos

cope

sen

sitiv

ity to

car

diac

sou

nds

Stet

hosc

ope

Welch Allyn Meditron Acc-H

ADC Adscope Acc-B

ADC Adscope Acc-D

ADC Adscope Acc-W

Littman 3200-W

Littman 3200-D

Cardionics EScope

Littman 3200-B

Thinklabs one-W

Welch Allyn Meditron Acc-L

Jabes Analyzer-B

Jabes Analyzer-W

Jabes Analyzer-D

Thinklabs Da32a+-B-NA

Thinklabs da32a+-D-NA

05101520Loudness

253035404550

Elec

troni

c st

etho

scop

e se

nsiti

vity

to c

ardi

ac s

ound

s

Stet

hosc

ope

Figu

re 8

noi

se r

ejec

tion

of c

ardi

ac s

ound

s 22

–281

hz.

Not

e: a

cous

tic (

left)

and

ele

ctro

nic

(rig

ht)

stet

hosc

ope

nois

e re

ject

ion

was

mea

sure

d at

freq

uenc

y ra

nges

(22

–281

hz)

of c

ardi

ac s

ound

s su

ch a

s s1

–s4,

inno

cent

mur

mur

s, a

nd M

s.A

bbre

viat

ions

: lU

Fs, l

oudn

ess

Uni

ts r

elat

ive

to d

igita

l Ful

l sca

le; M

s, m

itral

ste

nosi

s.

05 –5 –10 –15 –20101520LUFS + 70 DB

25

Acou

stic

ste

thos

cope

noi

se re

ject

ion

to c

ardi

ac s

ound

s(S

1–S4

& M

S)

Stet

hosc

ope

Welch Allyn Harvey Elite-B

Littman Cardiology IV-B

Heine Gamma 3.2-D

Heine Gamma 3.2-B

Mabis Spectrum-D

Littman Cardiology III-D

Ultrascope-D

Littman Cardiology III-B

Disposable Blue-D

Adscope Platinum-D

Omron Sprague Rappaport-D

Littman Cardiology IV-D

Mabis Legacy Sprague LC-Bx

Mabis Legacy Sprague LC-Dx

Welch Allyn Harvey Elite-D

Disposable Yellow-D

Omron Sprague Rappaport-B

Littman 3200-B

Littman 3200-W

Littman 3200-D

ADC Adscope Acc-W

Cardionics EScope

Thinklabs da32a+-B-NA

Jabes Analyzer-B

Jabes Analyzer-D

Jabes Analyzer-W

ADC Adscope Acc-D

Thinklabs da32a+-D-NA

Thinklabs One-W

ADC Adscope Acc-B

Welch Allyn Meditron Acc-H

Welch Allyn Meditron Acc-L

05 –5 –10 –15 –2010152025

Loudness

Elec

troni

c st

etho

scop

e no

ise

reje

ctio

n to

car

diac

sou

nds

(S1–

S4 &

MS)

Stet

hosc

ope

Page 9: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12 submit your manuscript | www.dovepress.com

Dovepress

Dovepress

49

acoustic analysis of popular stethoscopes

Although the Littman 3200 (38.87) in D mode is sensitive,

the same stethoscope (22.376) in B mode is not as sensitive

in this frequency range (Figure 9).

Of acoustic stethoscopes evaluated, the Heine Gamma 3.2

(16.504 in D mode) rejected the most noise, while the Welch

Allyn Harvey Elite (4.338 in D mode) rejected the least noise

in the frequency range associated with VSD, ASD, AS, and

PS. The Littman 3200 is the electronic stethoscope which

exhibited the most noise rejection (20.328 in W mode) and

Welch Allyn Meditron Acc (–3.978 in H mode, –8.952 in L

mode) exhibited the least noise rejection at this frequency

range (Figure 10).

Respiratory sounds (689–2,584 hz)Among the acoustic stethoscopes, respiratory sounds with

a frequency range 689–2,584 Hz could be heard best by

the Mabis Spectrum (22.938 in D mode) and the Litt-

man Cardiology IV (22.504 in B mode). The Welch Allyn

Meditron Acc (29.245 in L mode) and the ADC Adscope

Acc (29.006 in W mode, 29.096 in D mode) are the elec-

tronic stethoscopes which exhibited the most perceived

loudness (Figure 11).

Most electronic and acoustic stethoscopes exhibit poor

AMB rejection when auscultating respiratory sounds. Of all

stethoscopes examined, the Littman 3200 exhibited the most

noise rejection (14.162 in W mode and 9.452 in D mode).

The acoustic stethoscope which rejected the most noise was

the Mabis Spectrum (6.382 in D mode; Figure 12).

Tracheal sounds and swallow sounds (2500–8,000 hz)All stethoscopes demonstrate a large amount of attenuation

in this range. The software used to measure LUFS outputs

a null result for all models due to this attenuation. It should

also be noted that only very loud sounds (gastrointestinal,

swallow, and tracheal sounds) may be audible in this range

on some stethoscopes.

DiscussionThis study has demonstrated a considerable variation

among commercially available stethoscopes, both in terms

of capturing clinically relevant sounds and in the ability to

reject AMB. The Jabes Analyzer electronic stethoscope, for

example, had excellent frequency response but poorer AMB

rejection. While the Littman 3200 electronic stethoscope

exhibited exceptional AMB rejection, it was not among the

most sensitive stethoscopes.

Figu

re 9

sen

sitiv

ity t

o ca

rdia

c m

urm

ur s

ound

s 20

5–77

5 h

z.N

ote:

The

per

ceiv

ed lo

udne

ss o

f aco

ustic

(le

ft) a

nd e

lect

roni

c st

etho

scop

es (

righ

t) t

o fr

eque

ncy

rang

e 20

5–77

5 h

z w

as m

easu

red

to c

ompa

re t

heir

sen

sitiv

ity t

o ca

rdia

c m

urm

ur s

ound

s.A

bbre

viat

ion:

lU

Fs, l

oudn

ess

Uni

ts r

elat

ive

to d

igita

l Ful

l sca

le.

05101520

LUFS + 70 DB253035404550

Acou

stic

ste

thos

cope

sen

sitiv

ity to

car

diac

mur

mur

sou

nds

Stet

hosc

ope

Littman Cardiology IV-B

Mabis Spectrum-D

Littman Cardiology III-D

Ultrascope-D

Littman Cardiology IV-D

Littman Cardiology III-B

Heine Gamma 3.2-D

Heine Gamma 3.2-B

Welch Allyn Harvey Elite-B

Disposable Blue-D

Adscope Platinum-D

Omron Sprague Rappaport-D

Welch Allyn Harvey Elite-D

Mabis Legacy Sprague LC-Bx

Omron Sprague Rappaport-B

Mabis Legacy Sprague LC-Dx

Disposable Yellow-D

Jabes Analyzer-D

ADC Adscope Acc-D

ADC Adscope Acc-W

Jabes Analyzer-B

Jabes Analyzer-W

Littman 3200-W

Littman 3200-D

Thinklabs da32a+-D-NA

Welch Allyn Meditron Acc-L

Welch Allyn Meditron Acc-H

ADC Adscope Acc-B

Thinklabs da32a+-B-NA

Cardionics EScope

Thinklabs One-W

Littman 3200-B

05101520Loudness

253035404550

Elec

troni

c st

etho

scop

e se

nsiti

vity

to c

ardi

ac m

urm

ur s

ound

s

Stet

hosc

ope

Page 10: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12submit your manuscript | www.dovepress.com

Dovepress

Dovepress

50

Weiss et al

Figu

re 1

0 n

oise

rej

ectio

n of

car

diac

sou

nds

205–

775

hz.

Not

e: a

cous

tic (

left)

and

ele

ctro

nic

(rig

ht)

stet

hosc

ope

nois

e re

ject

ion

was

mea

sure

d at

freq

uenc

y ra

nges

(20

5–77

5 h

z) o

f car

diac

sou

nds

such

as

VsD

, asD

, as,

and

Ps.

Abb

revi

atio

ns: a

s, a

ortic

ste

nosi

s; a

sD, a

tria

l sep

tal d

efec

t; lU

Fs, l

oudn

ess

Uni

ts r

elat

ive

to d

igita

l Ful

l sca

le; P

s, p

ulm

onic

ste

nosi

s; V

sD, v

entr

icul

ar s

epta

l def

ect.

0

Welch Allyn Harvey Elite-B

Littman Cardiology IV-B

Heine Gamma 3.2-D

Heine Gamma 3.2-B

Mabis Spectrum-D

Littman Cardiology III-D

Ultrascope-D

Littman Cardiology III-B

Disposable Blue-D

Adscope Platinum-D

Omron Sprague Rappaport-D

Littman Cardiology IV-D

Mabis Legacy Sprague LC-Bx

Mabis Legacy Sprague LC-Dx

Welch Allyn Harvey Elite-D

Disposable Yellow-D

Omron Sprague Rappaport-B

5 –5 –10 –15 –20101520LUFS + 70 DB

25

05 –5 –10 –15 –2010152025

Elec

troni

c st

etho

scop

e no

ise

reje

ctio

n to

car

diac

sou

nds

(VSD

, ASD

, AS,

PS)

Acou

stic

ste

thos

cope

noi

se re

ject

ion

to c

ardi

ac s

ound

s(V

SD, A

SD, A

S, P

S)

Stet

hosc

ope

Littman 3200-B

Littman 3200-W

Littman 3200-D

Cardionics EScope

Thinklabs da32a+-B-NA

Jabes Analyzer-B

Jabes Analyzer-D

Jabes Analyzer-W

Thinklabs da32a+-D-NA

Thinklabs One-W

ADC Adscope Acc-B

ADC Adscope Acc-D

ADC Adscope Acc-W

Welch Allyn Meditron Acc-H

Welch Allyn Meditron Acc-L

Loudness

Stet

hosc

ope

Figu

re 1

1 se

nsiti

vity

to

resp

irat

ory

soun

ds.

Not

e: T

he p

erce

ived

loud

ness

of a

cous

tic (

left)

and

ele

ctro

nic

stet

hosc

opes

(ri

ght)

to

freq

uenc

y ra

nge

689–

2,58

4 h

z w

as m

easu

red

to c

ompa

re t

heir

sen

sitiv

ity t

o re

spir

ator

y so

unds

.A

bbre

viat

ion:

lU

Fs, l

oudn

ess

Uni

ts r

elat

ive

to d

igita

l Ful

l sca

le.

0

Mabis Spectrum-D

Littman Cardiology IV-B

Littman Cardiology IV-D

Littman Cardiology III-D

Disposable Blue-D

Ultrascope-D

Welch Allyn Harvey Elite-B

Heine Gamma 3.2-B

Littman Cardiology III-B

Heine Gamma 3.2-D

Adscope Platinum-D

Omron Sprague Rappaport-D

Welch Allyn Harvey Elite-D

Mabis Legacy Sprague LC-Dx

Omron Sprague Rappaport-B

Mabis Legacy Sprague LC-Bx

Disposable Yellow-D

5101520LUFS + 70 DB25303540 –5

Acou

stic

ste

thos

cope

sen

sitiv

ity to

resp

irato

ry s

ound

s

Stet

hosc

ope

0

Welch Allyn Meditron Acc-H

ADC Adscope Acc-B

ADC Adscope Acc-D

ADC Adscope Acc-W

Littman 3200-W

Littman 3200-D

Cardionics EScope

Littman 3200-B

Thinklabsone-W

Welch Allyn Meditron Acc-L

Jabes Analyzer-B

Jabes Analyzer-W

Jabes Analyzer-D

Thinklabs da32a+-B-NA

Thinklabs Da32a+-D-NA

5 –5101520Loudness

25303540

Elec

troni

c st

etho

scop

e se

nsiti

vity

to re

spira

tory

sou

nds

Stet

hosc

ope

Page 11: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12 submit your manuscript | www.dovepress.com

Dovepress

Dovepress

51

acoustic analysis of popular stethoscopes

Among the acoustic stethoscopes examined, the Mabis

Spectrum and the Littman Cardiology IV were the most sensi-

tive stethoscopes when auscultating frequency ranges asso-

ciated with gastrointestinal, cardiovascular, and respiratory

sounds. The perceived loudness of the electronic stethoscopes

examined varied. Gastrointestinal sounds and cardiovascular

murmurs were most audible when auscultating with the Jabes

Analyzer and the ADC Adscope. The Welch Allyn Meditron

also exhibited exceptional sensitivity to respiratory sounds.

Consistently, for all frequency ranges examined, the

Littman 3200 is the electronic stethoscope which exhibited

the most noise rejection. On the contrary, the Welch Allyn

Meditron poorly rejected AMB, which may be due to its

requirement of firm auscultating pressure to achieve normal

performance. Of the acoustic stethoscopes, Omron Sprague

poorly rejected AMB among all frequency ranges. Noise

rejection analysis showed that there was no single top per-

forming acoustic stethoscope. The Heine Gamma 3.2, Mabis

Spectrum, Littman Cardiology III, and the Welch Allyn

Harvey Elite were all capable of considerable AMB rejection.

ConclusionAn in vitro analysis of a variety of stethoscopes, both acous-

tic and electronic, has demonstrated considerable variation

in stethoscope performance. A clinician should interpret

these results for the frequency range relevant to their field

of practice or for their envisioned operating environment

(eg, noisy emergency room vs quiet office setting) to help

facilitate stethoscope selection. The Littmann Cardiology

IV and low-cost Mabis Spectrum were consistently the top

two performers across all assessed parameters, thus possibly

making both a reasonable choice for most acoustic stetho-

scope applications. The Littmann 3200 was an outstanding

performer in the electronic category especially for low

frequencies. However, it should be noted that all electronic

stethoscopes had very similar performance below 1,000 Hz.

The data generated in this study demonstrate the need for

new stethoscope technology that is better suited for capturing

higher frequency (2,500–8,000 Hz) information. Advances in

material science, mechanical design, digital signal process-

ing, and data analysis should facilitate the development of

groundbreaking stethoscope technology that breathes new

life into the art of auscultation.

DisclosureDaniel Weiss, Joseph Butera III, Ryan Copt, Glenn Yeaw,

Mark Harpster, and James Hughes are employees of Bon-

giovi Medical & Health Technologies, Inc. Deeb N Salem Figu

re 1

2 n

oise

rej

ectio

n of

res

pira

tory

sou

nds.

Not

e: a

cous

tic (

left)

and

ele

ctro

nic

(rig

ht)

stet

hosc

ope

nois

e re

ject

ion

was

mea

sure

d at

freq

uenc

y ra

nges

(68

9–2,

584

hz)

of r

espi

rato

ry s

ound

s.A

bbre

viat

ion:

lU

Fs, l

oudn

ess

Uni

ts r

elat

ive

to d

igita

l Ful

l sca

le.

0

Welch Allyn Harvey Elite-B

Littman Cardiology IV-B

Heine Gamma 3.2-D

Heine Gamma 3.2-B

Mabis Spectrum-D

Littman Cardiology III-D

Ultrascope-D

Littman Cardiology III-B

Disposable Blue-D

Adscope Platinum-D

Omron Sprague Rappaport-D

Littman Cardiology IV-D

Mabis Legacy Sprague LC-Bx

Mabis Legacy Sprague LC-Dx

Welch Allyn Harvey Elite-D

Disposable Yellow-D

Omron Sprague Rappaport-B

5 –5 –10 –15 –20101520LUFS + 70 DB

Acou

stic

ste

thos

cope

noi

se re

ject

ion

to re

spira

tory

sou

nds

Stet

hosc

ope

05 –5 –10 –15 –20101520

Elec

troni

c st

etho

scop

e no

ise

reje

ctio

n to

resp

irato

ry s

ound

s

Littman 3200-B

Littman 3200-W

Littman 3200-D

Cardionics EScope

Thinklabs da32a+-B-NA

Jabes Analyzer-B

Jabes Analyzer-D

Jabes Analyzer-W

Thinklabs da32a+-D-NA

Thinklabs One-W

ADC Adscope Acc-B

ADC Adscope Acc-D

ADC Adscope Acc-W

Welch Allyn Meditron Acc-H

Welch Allyn Meditron Acc-L

Loudness

Stet

hosc

ope

Page 12: an in vitro acoustic analysis and comparison of popular

Medical Devices: Evidence and Research 2019:12submit your manuscript | www.dovepress.com

Dovepress

Dovepress

Medical Devices: Evidence and Research

Publish your work in this journal

Submit your manuscript here: https://www.dovepress.com/medical-devices-evidence-and-research-journal

Medical Devices: Evidence and Research is an international, peer-reviewed, open access journal that focuses on the evidence, technology, research, and expert opinion supporting the use and application of medical devices in the diagnosis, monitoring, treatment and management of clinical conditions and physiological processes. The identification of novel

devices and optimal use of existing devices which will lead to improved clinical outcomes and more effective patient management and safety is a key feature. The manuscript management system is completely online and includes a quick and fair peer-review system. Visit http://www. dovepress.com/testimonials.php to read real quotes from authors.

Dovepress

52

Weiss et al

is a grant recipient from Bongiovi Medical & Health Tech-

nologies, Inc. Christine Erie reports no conflict of interest

in this work.

References 1. Choi S, Jiang Z. Cardiac sound murmurs classification with autoregres-

sive spectral analysis and multi-support vector machine technique. Comput Biol Med. 2010;40(1):8–20.

2. Beaumont CE. Choosing the best stethoscope. Nursing. 2007;37 Suppl Med:12–13.

3. Johnston FD. AN acoustical study of the stethoscope. Arch Intern Med. 1940;65(2):328–339.

4. Ortega R, Téllez A, Leija L, Vera A. Measurement of ultrasonic properties of muscle and blood biological phantoms. Phys Procedia. 2010;3(1):627–634.

5. Sarbandi RR, Doyle JD, Navidbakhsh M, Hassani K, Torabiyan H. A color spectrographic phonocardiography (CSP) applied to the detection and characterization of heart murmurs: preliminary results. Biomed Eng Online. 2011;10:42.

6. Kim D, Tavel ME. Assessment of severity of aortic stenosis through time–frequency analysis of murmur. Chest. 2003;124(5):1638–1644.

7. Akbari MA, Hassani K, Doyle JD, et al. Digital subtraction phonocar-diography (DSP) applied to the detection and characterization of heart murmurs. Biomed Eng Online. 2011;10:109.

8. Noponen AL, Lukkarinen S, Angerla A, Sepponen R. Phono-spectro-graphic analysis of heart murmur in children. BMC Pediatr. 2007;7:23.

9. Sung PH, Thompson WR, Wang JN, Wang JF, Jang LS. Computer-assisted auscultation: patent ductus arteriosus detection based on auditory time–frequency analysis. J Med Biol Eng. 2015;35(1):76–85.

10. Reichert S, Gass R, Brandt C, Andrès E. Analysis of respiratory sounds: state of the art. Clin Med Circ Respirat Pulm Med. 2008;2:45–58.

11. Pochekutova IA, Korenbaum VI. Diagnosis of hidden bronchial obstruc-tion using computer-assessed tracheal forced expiratory noise time. Respirology. 2013;18(3):501–506.

12. Bahoura M, Lu X. Separation of crackles from vesicular sounds using wavelet packet transform. In: IEEE International Conference on Acous-tics, Speech and Signal Processing, 2006.

13. Elphick HE, Ritson S, Rodgers H, Everard ML. When a “wheeze” is not a wheeze: acoustic analysis of breath sounds in infants. Eur Respir J. 2000;16(4):593–597.

14. Oliveira A, Marques A. Respiratory sounds in healthy people: a sys-tematic review. Respir Med. 2014;108(4):550–570.

15. Hayashi N. Detection of pneumothorax visualized by computer analysis of bilateral respiratory sounds. Yonago Acta Med. 2011;54(4):75–82.

16. Fiz JA, Jané R, Lozano M, Gómez R, Ruiz J. Detecting unilateral phrenic paralysis by acoustic respiratory analysis. PLoS ONE. 2014;9(4):e93595–e93599.

17. Vyshedskiy A, Bezares F, Paciej R, Ebril M, Shane J, Murphy R. Trans-mission of crackles in patients with interstitial pulmonary fibrosis, con-gestive heart failure, and pneumonia. Chest. 2005;128(3):1468–1474.

18. Ching SS, Tan YK. Spectral analysis of bowel sounds in intestinal obstruction using an electronic stethoscope. World J Gastroenterol. 2012;18(33):4585–4592.

19. Ranta R, Louis-Dorr V, Heinrich C, Wolf D, Guillemin F. Digestive activity evaluation by multichannel abdominal sounds analysis. IEEE Trans Biomed Eng. 2010;57(6):1507–1519.

20. Dumas J, Hill KM, Adrezin RS, et al. Feasibility of an electronic stethoscope system for monitoring neonatal bowel sounds. Conn Med. 2013;77(8):467–471.

21. Craine BL, Silpa ML, O’Toole CJ. Two-dimensional positional mapping of gastrointestinal sounds in control and functional bowel syndrome patients. Dig Dis Sci. 2002;47(6):1290–1296.

22. Cichero JA, Murdoch BE. Acoustic signature of the normal swallow: characterization by age, gender, and bolus volume. Ann Otol Rhinol Laryngol. 2002;111(7 Pt 1):623–632.

23. Hubbert L, Sundbom P, Loebe M, Peterzén B, Granfeldt H, Ahn H. Acoustic analysis of a mechanical circulatory support. Artif Organs. 2014;38(7):593–598.

24. Tavel ME, Bates JR. The cervical bruit: sound spectral analysis related to severity of carotid arterial disease. Clin Cardiol. 2006;29(10):462–465.