TechnicalInformationTI 193C/24/ae
Conductivity Measurement
liquisys M CLM 223 / 253
Conductivity and Resistance Controller
Liquisys M CLM 253Liquisys M CLM 223
Due to the modularity of its design, theLiquisys M CLM 223 / 253 controller canbe adapted to a wide range of customerneeds. The basic version, which pro-
vides simple measuring and alarmsignaling functions, can be equippedwith additional software and hardwaremodules to match specific applications.Retrofitting of expansion modules is alsopossible.
Areas of application Ultrapure water Water treatment Ion exchangers Reverse osmosis Cooling water desalinization USP and WFI water
Benefits at a glance Measuring controller in field or
panel-mounted housing
Universal application Conductivity and specific resistance
via software switching
Transmitters for contacting orelectrodeless sensors
Ultrapure water compensation acc.to IEC 746 (contacting only)
USP 24 Stage 1 compliant Simple handling
Logically arranged menu structurewith plain text in 6 languagesfacilitates instrument configuration
Large, two-line display indicatesmeasured value and temperature atthe same time
Calibration via CAL key Safe Operation
Overvoltage (lightning) protectionaccording to EN 61000-4-5
Direct access for manual contactcontrol
User-defined alarm configuration foralarm contact and error current
CSA GP and CE approved
The basic unit can be extended with:
2 or 4 contacts for use as:
Limit contacts (also for temperature)
PID controller
Timer for simple rinse processesComplete cleaning with Chemoclean
Plus package:
Any current output configuration viatable
Automatic initiation of cleaning incase of alarm or limit violation
Ultrapure water monitoring withpre-alarm acc. to USP 24 (contactingonly)
Polarization detection (contacting only
Concentration measurement(% / TDS)
Temperature compensation via table Live check of measuring cell
HARTor Profibus PA and DP 2nd current output for temperature
The Power of Know How
-
7/31/2019 Analizadores de Conductividad Endress Hauser
2/12
Details Liquisys M CLM 223 / 253 provides asolution for all drinking water
processing, process water and wastewater treatment applications.
Contacting or electrodeless
Two instrument versions formeasurement with contacting (two-
electrode) measuring cells or
electrodeless sensors are available. Theuse of electrodeless sensors that are
less sensitive to soiling than contactingmeasuring cells is recommended for
concentration measurement or adheringmedia.
Features of the basic version (CD / ID):
Measurement of conductivity or
specific resistanceThis is selected via the menu. During
measurement, the value measured can
be displayed in the other measuringmode. The temperature is displayed at
the same time if desired.
Temperature compensation selectionsavailable are linear ( = 0 means no
temperature compensation) or specialultrapure water compensation (NaCl
curve) according to IEC 746. Thereference temperature is 77 F (25 C).
Different alarms are required depending
on application and operator. Therefore
the Liquisys M CLM 223 / 253 permitsindependent configuration of the alarm
contact and error current for eachindividual error. Unnecessary or
undesirable alarms can be suppressedin this manner.
Up to four contacts can be used as limit
contacts (also for temperature), to
implement a PID controller or forcleaning functions.
Direct manual operation of the
contacts (bypassing the menu) providesquick access to limit, control or cleaning
contacts, permitting speedy correction ofdeviations.
The serial numbers of the instrumentand modules and the order code can be
called up on the display.
The cell constant can be edited andeven calibrated for demanding special
applications.
The CS / IS version provides additionalfunctions:
Electrodeless sensors must normally be
installed in pipes at a required minimumdistance from the pipe wall. If this
minimum distance is not observed, themeasured value changes. The built-in
adaptive calibration feature can be
used to compensate for this once thesensor is installed.
In order to display wide measuring
ranges while still achieving a highresolution in certain ranges, the current
output can be configured as requiredvia a table. This permits bilinear or
quasi-logarithmic curves.
The live check issues an alarm whenthe sensor signal does not change over
a defined period of time. This may be
caused by blocking, passivation, orseparation from the process.
Ultrapure water monitoring according to
USP 24 means that the uncompensatedconductivity and the temperature are
measured and compared to a table. TheLiquisys M CLM 223 / 253 (contacting)comes with this function, and the
user-adjustable pre-alarm indicatesundesirable operating values in due
time. Full compliance with USP 24requires the use of a precisely calibrated
measuring cell, for example, the CLS 15.
The conversion from conductivity toconcentration is effected using four
user-definable concentration curves.
This permits concentrations to bedisplayed directly in % or ppm, i.e. as
TDS (total dissolved salts solids).
E 017
E 123
- - - -
on
off
on
off
on
off
2.4 / 22 mA
off
on
Timer
t
off
off
on
on
Chemoclean
t
Instrument DATA
eS r Nr I1
1 5 6 3 48 3
t0
20
mA
2
0
20
mA
The sensor and process checksystem also indicates any polarization
of the contacting measuring cell. Since
there is no subdivision into severalmeasuring ranges (the full range from
0 to 2000 mS/cm is available at alltimes), the measuring range is limited by
the cell constant only. Polarization
detection then determines whether ornot the working range determined by the
measuring cell is exceeded.
-
7/31/2019 Analizadores de Conductividad Endress Hauser
3/12
ENDRESS+HAUSERLIQUISYS
CAL REL
REL1
REL1
ALARM
REL2
REL2E
ENDRESS+HAUSERLIQUISYS
CAL REL
REL1
REL1
ALARM
REL2
REL2E
Basic Version and
Plus Package
Measuring and Control
System
A complete measuring system
comprises:
the conductivity controller Liquisys MCLM 223 or CLM 253
a conductivity measuring cell with orwithout integrated temperature sensor
an immersion, flow or retractableholder
Options:
an extension cable CYK 71 (contactingsensors) or CLK 5 (electrodeless
sensors) and
junction box VBM
Up to 4 relays
Alarm contact
Signal output 1:
conductivitySignal output 2:temperature (optional)
Aux. 15 V supply voltage
2 digital inputs
Power supply (mains)
Complete measuring systemswith Liquisys S CLM 223 / 253
3
Immersionholder CYA 611
Liquisys M CLM 223
Liquisys M CLM 253
CLS 15
CLS 50
MeasuringCable
Measuring Ranges of
Conductivity Sensors
Basic Version With plus package (CS / IS version)
Alarm signaling Alarm signaling
MEASUREMENT
CALIBRATION of cell constant Adaptive CALIBRATION (electrodeless sensors)CALIBRATION of residual coupling (toroidal sensors)
POLARIZATION detectionRead instrument DATA (contacting cells)
LIVE CHECK of sensor
Linear CURRENT OUTPUTCURRENT OUTPUT simulation CURRENT OUTPUT programmable (table)
1 programmable ALARM CONTACT(contact and error current)
Additional features Additional features
Controlling Controlling
2 CHANGEOVER CONTACTS for
- Measuring parameter limit Measurement according to USP 24 with- Temperature limit pre-alarm (contacting cells)- PID controller Automatic cleaning triggered by alarm or limit- Timer for cleaning violation
Additional features Additional features
Cleaning Cleaning
2 more CHANGEOVER CONTACTS (total of 4) for
- Measuring parameter limit- Temperature limit Cleaning triggered externally or automatically- PID controller by alarm or limit violation- Chemoclean cleaning
(water and cleaning agent)
-
7/31/2019 Analizadores de Conductividad Endress Hauser
4/12
1 7 3 4m S/cm
ENDRESS+HAUSER
LIQUISYS M
CAL REL
REL1
REL1
ALARM
REL2
REL2E
Setpoint
Operation
Liquid crystal displayfor display ofmeasured values,configuration dataand current menu field
4 main control keys forcalibration and instrumentconfiguration
LED indicatorsfor switched limitoutputs
Field for userlabeling
Key for manualrelay switching
LED indicator foralarm function
User interface:Display and keys
Everything at a glance
The display simultaneously shows the
current measured value and thetemperaturethe essential processdata. Brief informational texts in theconfiguration menu provide assistance
with parameter configuration.
Intelligent and simple
All instrument control functions are
arranged in a logical menu structure.Following access code entry, the
individual parameters can be easilyselected and modified as needed.
Overview of Liquisys MCLM 223 / 253 menu.This covers all theoptions that can beinstalled (see page 2Details).
4
Key for switchingbetween automatic andmanual relay operation
-
7/31/2019 Analizadores de Conductividad Endress Hauser
5/12
15 V
85
86
S
10 to 50 V
93
94
L+
L
L
N
~ =
41
42
43
47
57
51
54
48
58
52
55
49
59
53
56
mA
10 to 50 V
81
34
33
82
J
JJ
mA
32
31
11
84
84
15
16
83
83
12
13
k
Electrical Connection
Left:Liquisys M CLM 223,connections on the rearof the instrument
Right:Liquisys M CLM 253,rear of instrument withreplaceable electronics
box (E)
Electrical connection ofLiquisys M CLM 223 / 253
All connections to the panel-mountedinstrument CLM 223 are established via
the terminal strips on the rear.
In the case of the field instrument CLM
253, all wires (including the measuringcell cable) are connected to terminals in
a separate wiring compartment.
All the wiring can remain in place if theinstrument needs repair because all
repairs are limited to assemblyreplacement. So, dismantling the
instrument and rewiring are no longernecessary.
Contactingmeasuringcell
Signal output 1:conductivity
Signal output 2:temperature(optional)
Digital input 1
Digital input 2
Aux. voltageoutput
Alarm*
Relay 1*
Relay 2*
Relay 3*
Relay 4*
optional
*Contactstatus shown:no current orfault present
Power supply
E
5
(alternative)
Electrodelesssensor
Temperaturesensor Pt 100
-
7/31/2019 Analizadores de Conductividad Endress Hauser
6/12
ENDRESS+HAUSER
LIQUISYS
CAL REL
REL1
REL1
ALARM
REL2
REL2E
ENDRESS+HAUSERLIQUISYS
CAL REL
REL1
REL1
ALARMREL2
REL2E
Dimensions
3.54(90)
5.47(139)
0.24
(6)
3.78(96)
Dimensions ofLiquisys M CLM 253
Dimensions ofLiquisys M CLM 223
6
9.72(247)
0.24(6)
4.53
(115)
6.18(157)
0.43
(11)
2.75(70)
2.75
(70)6.06(154)6.69(170)
-
7/31/2019 Analizadores de Conductividad Endress Hauser
7/12
Mounting of CLM 223 The panel-mounted version is mountedusing the supplied tensioning screws.
The required overall mounting depth isapproximately 6.89 in. (175 mm).
Installation of panel-mounted housing
7
3.62+0.02
(92+0.5)
3.62+0.02
(92+0.5)
3.78(96)
max. 1.77(45)
0.24(6)
5.47(139)
approx.0.98(25)
1.30(33)
2.24(57)
3.54(90)
Flat gasket(included in delivery)
Mounting cutoutaccording toDIN 43700
-
7/31/2019 Analizadores de Conductividad Endress Hauser
8/12
6.18(157)
0.31(8)
0.63(16)
4.53(115)
Mounting of
Liquisys M CLM 253
8
Wall mountingScrews: 0.24 (6 mm)Anchors: 0.31 (8 mm)
Mounting on cylindricalpipes (shown withweather protection coverinstalled on right)
Post mounting kit forinstallation of field housingon cylindrical pipes(horizontal or vertical,max. 2.36 (60 mm);see Accessories)
3.54(90)
1.18(30)
0.31(8)
2.36(60)
0.24(M 6)
3.54(90)
2.76(70)
0.31(8)
0.24(6)2.76
(70)3.54(90)
2.76(70)
-
7/31/2019 Analizadores de Conductividad Endress Hauser
9/12
Manufacturer Endress+Hauser
Product designation Liquisys M CLM 223Liquisys M CLM 253
Measured quantities conductivity, resistance, temperature
Contacting conductivity/resistance measurement
Measuring range conductivity: 0 to 2000 mS/cm (uncompensated)resistance: 0 to 86 mcm
Usable cell constant k= 0.0025 to 99.99 cm-1
Maximum length of cable to measuring cell conduct ivity: 330 ft . (100 m)resistance: 50 ft. (15 m)
Measuring frequency conductivity: 300 to 1000 Hzresistance: 30 to 400 Hz
Toroidal conductivity measurement
Measuring range 0 to 2000 mS/cm (uncompressed)
Usable cell constant k= 0.0025 to 99.99 cm-1
Maximum length of cable to measuring cell 200 ft. (60 m) (CLK 5)
Measuring frequency 2 kHz
Minimum span for 0 / 4 to 20 mA signal meas. value betw. 0 and 199.9 S/cm: 20S/cmmeas. value betw. 200 and 1999S/cm: 200S/cmmeas. value betw. 2 and 19.99 mS/cm: 2mS/cmmeas. value betw. 20 and 2000 mS/cm: 200 mS/cm
Temperature measurement
Temperature sensor Pt 100
Measuring range -31 to +482 F (-35 to +250 C)
Temperature offset range 68 F ( 20 C)
Temperature compensation
Compensation types linear, NaCl, table,(with contacting cells only: ultrapure water)
Range -31 to +482 F (-35 to +250 C)
Reference temperature 77 F (25 C)
Digital inputs 1 and 2
Voltage 10 to 50 VDC
Current consumption max. 10 mA
Conductivity/resistance signal output
Current range 0 / 4 to 20 mA, galvanically separated;error current 2.4 / 22 mA
Load max. 500
Output range adjustable
Isolation voltage max 350 Vrms
/ 500 V DC
Overvoltage (lightning) protection acc. to EN 61000-4-5: 1995
Temperature signal output (optional)
Current range 0 / 4 to 20 mA, galvanically separated
Load max. 500
Output range adjustable, 10 to 100% of upper range value
Isolation voltage max 350 Vrms / 500 V DCOvervoltage (lightning) protection acc. to EN 61000-4-5: 1995
Auxiliary voltage output
Output voltage 15 V 0.6 V
Output current max. 30 mA
Contact outputs (floating changeover contacts)
Switching current with ohmic load (cos = 1) max. 2 A
Switching current with inductive load (cos = 0.4) max. 2 A
Switching voltage max. 250 V AC, 30 V DC
Switching power with ohmic load (cos = 1) max. 1250 VA AC, 150 W DC
Switching power with inductive load (cos = 0.4) max. 500 VA AC, 90 W DC
9
Technical DataGeneral specifications
Input
Output
-
7/31/2019 Analizadores de Conductividad Endress Hauser
10/12
Technical Data
(continued)Output (continued)
Accuracy
Ambient conditions
Physical data / design
Power requirements
10
Limit contactor
Pickup / dropout delay 0 to 2000 seconds
Controller
Function (adjustable) pulse length / pulse frequency controller
Controller response PID
Proportional band Kp: 0.10 to 10.00
Period for pulse length controller 0.5 to 999.9 seconds
Frequency for pulse frequency controller 60 to 180 min *
Alarm
Function (switchable) steady / fleeting contact; normally closed / open
Alarm threshold adjustment range conductivity / resistance / temperature:complete measuring range
Alarm delay 0 to 2000 seconds
Conductivity measurement
Deviation of indication1 max. 0.5% of measured value 4 digits
Reproducibility max. 0.2% of measured value 2 digits
Measurement deviation1, conductivity signal output max. 0.75% of current output range
Temperature measurement
Resolution 0.2 F or 0.1 C
Deviation of indication1 max. 1.0% of measuring range
Measurement deviation1, temperature signal output max. 1.25% of current output range
Nominal operating temperature 14 to 131 F (-10 to +55 C)
Ambient temperature ( limit operating conditions) -4 to 140 F (-20 to +60 C)
Storage and transport temperature -13 to 149 F (-25 to +65 C)
Relative humidity (nominal operating conditions) 10 to 95%, non-condensing
Protection class of panel-mounted unit NEMA 3S (IP 54 front, IP 30 housing)
Protection class of field housing NEMA 4X (IP 65)
Electromagnetic compatibility interference emission acc. to EN 50081-1: 1992interference immunity acc. to EN 50082-2: 1995
Dimensions of panel-mounted unit (H x W x D) 3.78 x 3.78 x 5.71 in. (96 x 96 x 145 mm)
Installation depth approx. 6.89 in. (175 mm)
Dimensions of field housing (H x W x D) 9.72 x 6.69 x 4.53 in. (247 x 170 x 115 mm)
Weight of panel-mounted unit max. 1.54 lb. (0.7 kg)
Weight with field housing max 5.07 lb. (2.3 kg)
Display LC display, two lines, five and nine digits, withstatus indicators
Materials
Housing of panel-mounted unit polycarbonate
Front membrane polyester, UV-resistant
Field housing ABS PC Fr
Supply voltage 100 / 115 / 230 V AC +10 / -15%, 48 to 62 Hz24 V AC/DC +20 / -15%
Power consumption max. 7.5 VA
Fuse protection fine-wire fuse, medium time-lag, 250 V / 1 A
* acc. to IEC 746-1, for nominal operating conditions
Subject to modifications.
-
7/31/2019 Analizadores de Conductividad Endress Hauser
11/12
AccessoriesMounting accessories
Type Features Order number
Pendulum For pendulum suspension of assemblies 50080196
frame e.g. CYA 611
Material: 304 SS
Post mounting Kit for mounting of field housing on 50086842kit horizontal or vertical pipes, max.
diameter, 2.5 (60 mm)
Material: 304 SS
Holders
Type Features Applications
CYA 611 Immersion assembly with G 1, G 3/4 Waste water, water
or 3/4 NPT thread
Two-electrode conductivity measuring cells (selection)
Type Features Applications
CLS 15 Cell constant: k = 0.1 cm-1 or 0.01 cm-1 Reverse osmosis
Process connections: 3/4 NPT, 1-1/2 filter, ion exchangerTri-clamp
CLS 19 Cell constant: k = 0.1cm-1 or 0.01 cm-1 Boiler feed, control
Process connections: 3/4 NPT
CLS 21 Cell constant: k = 1 cm-1 Service water, boiler
Process connections: 1 NPT and feed water2 Tri-clamp
Electrodeless conductivity measuring sensorsType Features Applications
CLS 50 Cell constant: k=2 cm-1 Concentration
Process connections: G 3/4, 1 NPT measurement,2 ANSI flange incl. 16.4 ft (5 m) attached product monitoring,cable chemical industries
CLS 52 Cell constant: k=5.9 cm-1 Concentration
Process connections: 2 Tri-clamp measurement,and Perlick product monitoring,
chemical industries,
food industries
11
Cables
Type Features Order number
CYK 71 Standard measuring cable for connection 50085333
of two electrode measuring cells with andwithout Pt 100 (also usable for extension)
CLK 5 Extension cable for CLS 50 and CLS 52 50085473
sensors
VBM Junction box NEMA 4X 51500177
-