    TechnicalInformationTI 193C/24/ae

    Conductivity Measurement

    liquisys M CLM 223 / 253

    Conductivity and Resistance Controller

    Liquisys M CLM 253Liquisys M CLM 223

    Due to the modularity of its design, theLiquisys M CLM 223 / 253 controller canbe adapted to a wide range of customerneeds. The basic version, which pro-

    vides simple measuring and alarmsignaling functions, can be equippedwith additional software and hardwaremodules to match specific applications.Retrofitting of expansion modules is alsopossible.

    Areas of application Ultrapure water Water treatment Ion exchangers Reverse osmosis Cooling water desalinization USP and WFI water

    Benefits at a glance Measuring controller in field or

    panel-mounted housing

    Universal application Conductivity and specific resistance

    via software switching

    Transmitters for contacting orelectrodeless sensors

    Ultrapure water compensation acc.to IEC 746 (contacting only)

    USP 24 Stage 1 compliant Simple handling

    Logically arranged menu structurewith plain text in 6 languagesfacilitates instrument configuration

    Large, two-line display indicatesmeasured value and temperature atthe same time

    Calibration via CAL key Safe Operation

    Overvoltage (lightning) protectionaccording to EN 61000-4-5

    Direct access for manual contactcontrol

    User-defined alarm configuration foralarm contact and error current

    CSA GP and CE approved

    The basic unit can be extended with:

    2 or 4 contacts for use as:

    Limit contacts (also for temperature)

    PID controller

    Timer for simple rinse processesComplete cleaning with Chemoclean

    Plus package:

    Any current output configuration viatable

    Automatic initiation of cleaning incase of alarm or limit violation

    Ultrapure water monitoring withpre-alarm acc. to USP 24 (contactingonly)

    Polarization detection (contacting only

    Concentration measurement(% / TDS)

    Temperature compensation via table Live check of measuring cell

    HARTor Profibus PA and DP 2nd current output for temperature

    The Power of Know How

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    2/12

    Details Liquisys M CLM 223 / 253 provides asolution for all drinking water

    processing, process water and wastewater treatment applications.

    Contacting or electrodeless

    Two instrument versions formeasurement with contacting (two-

    electrode) measuring cells or

    electrodeless sensors are available. Theuse of electrodeless sensors that are

    less sensitive to soiling than contactingmeasuring cells is recommended for

    concentration measurement or adheringmedia.

    Features of the basic version (CD / ID):

    Measurement of conductivity or

    specific resistanceThis is selected via the menu. During

    measurement, the value measured can

    be displayed in the other measuringmode. The temperature is displayed at

    the same time if desired.

    Temperature compensation selectionsavailable are linear ( = 0 means no

    temperature compensation) or specialultrapure water compensation (NaCl

    curve) according to IEC 746. Thereference temperature is 77 F (25 C).

    Different alarms are required depending

    on application and operator. Therefore

    the Liquisys M CLM 223 / 253 permitsindependent configuration of the alarm

    contact and error current for eachindividual error. Unnecessary or

    undesirable alarms can be suppressedin this manner.

    Up to four contacts can be used as limit

    contacts (also for temperature), to

    implement a PID controller or forcleaning functions.

    Direct manual operation of the

    contacts (bypassing the menu) providesquick access to limit, control or cleaning

    contacts, permitting speedy correction ofdeviations.

    The serial numbers of the instrumentand modules and the order code can be

    called up on the display.

    The cell constant can be edited andeven calibrated for demanding special

    applications.

    The CS / IS version provides additionalfunctions:

    Electrodeless sensors must normally be

    installed in pipes at a required minimumdistance from the pipe wall. If this

    minimum distance is not observed, themeasured value changes. The built-in

    adaptive calibration feature can be

    used to compensate for this once thesensor is installed.

    In order to display wide measuring

    ranges while still achieving a highresolution in certain ranges, the current

    output can be configured as requiredvia a table. This permits bilinear or

    quasi-logarithmic curves.

    The live check issues an alarm whenthe sensor signal does not change over

    a defined period of time. This may be

    caused by blocking, passivation, orseparation from the process.

    Ultrapure water monitoring according to

    USP 24 means that the uncompensatedconductivity and the temperature are

    measured and compared to a table. TheLiquisys M CLM 223 / 253 (contacting)comes with this function, and the

    user-adjustable pre-alarm indicatesundesirable operating values in due

    time. Full compliance with USP 24requires the use of a precisely calibrated

    measuring cell, for example, the CLS 15.

    The conversion from conductivity toconcentration is effected using four

    user-definable concentration curves.

    This permits concentrations to bedisplayed directly in % or ppm, i.e. as

    TDS (total dissolved salts solids).

    E 017

    E 123

    - - - -

    on

    off

    on

    off

    on

    off

    2.4 / 22 mA

    off

    on

    Timer

    t

    off

    off

    on

    on

    Chemoclean

    t

    Instrument DATA

    eS r Nr I1

    1 5 6 3 48 3

    t0

    20

    mA

    2

    0

    20

    mA

    The sensor and process checksystem also indicates any polarization

    of the contacting measuring cell. Since

    there is no subdivision into severalmeasuring ranges (the full range from

    0 to 2000 mS/cm is available at alltimes), the measuring range is limited by

    the cell constant only. Polarization

    detection then determines whether ornot the working range determined by the

    measuring cell is exceeded.

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    3/12

    ENDRESS+HAUSERLIQUISYS

    CAL REL

    REL1

    REL1

    ALARM

    REL2

    REL2E

    ENDRESS+HAUSERLIQUISYS

    CAL REL

    REL1

    REL1

    ALARM

    REL2

    REL2E

    Basic Version and

    Plus Package

    Measuring and Control

    System

    A complete measuring system

    comprises:

    the conductivity controller Liquisys MCLM 223 or CLM 253

    a conductivity measuring cell with orwithout integrated temperature sensor

    an immersion, flow or retractableholder

    Options:

    an extension cable CYK 71 (contactingsensors) or CLK 5 (electrodeless

    sensors) and

    junction box VBM

    Up to 4 relays

    Alarm contact

    Signal output 1:

    conductivitySignal output 2:temperature (optional)

    Aux. 15 V supply voltage

    2 digital inputs

    Power supply (mains)

    Complete measuring systemswith Liquisys S CLM 223 / 253

    3

    Immersionholder CYA 611

    Liquisys M CLM 223

    Liquisys M CLM 253

    CLS 15

    CLS 50

    MeasuringCable

    Measuring Ranges of

    Conductivity Sensors

    Basic Version With plus package (CS / IS version)

    Alarm signaling Alarm signaling

    MEASUREMENT

    CALIBRATION of cell constant Adaptive CALIBRATION (electrodeless sensors)CALIBRATION of residual coupling (toroidal sensors)

    POLARIZATION detectionRead instrument DATA (contacting cells)

    LIVE CHECK of sensor

    Linear CURRENT OUTPUTCURRENT OUTPUT simulation CURRENT OUTPUT programmable (table)

    1 programmable ALARM CONTACT(contact and error current)

    Additional features Additional features

    Controlling Controlling

    2 CHANGEOVER CONTACTS for

    - Measuring parameter limit Measurement according to USP 24 with- Temperature limit pre-alarm (contacting cells)- PID controller Automatic cleaning triggered by alarm or limit- Timer for cleaning violation

    Additional features Additional features

    Cleaning Cleaning

    2 more CHANGEOVER CONTACTS (total of 4) for

    - Measuring parameter limit- Temperature limit Cleaning triggered externally or automatically- PID controller by alarm or limit violation- Chemoclean cleaning

    (water and cleaning agent)

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    4/12

    1 7 3 4m S/cm

    ENDRESS+HAUSER

    LIQUISYS M

    CAL REL

    REL1

    REL1

    ALARM

    REL2

    REL2E

    Setpoint

    Operation

    Liquid crystal displayfor display ofmeasured values,configuration dataand current menu field

    4 main control keys forcalibration and instrumentconfiguration

    LED indicatorsfor switched limitoutputs

    Field for userlabeling

    Key for manualrelay switching

    LED indicator foralarm function

    User interface:Display and keys

    Everything at a glance

    The display simultaneously shows the

    current measured value and thetemperaturethe essential processdata. Brief informational texts in theconfiguration menu provide assistance

    with parameter configuration.

    Intelligent and simple

    All instrument control functions are

    arranged in a logical menu structure.Following access code entry, the

    individual parameters can be easilyselected and modified as needed.

    Overview of Liquisys MCLM 223 / 253 menu.This covers all theoptions that can beinstalled (see page 2Details).

    4

    Key for switchingbetween automatic andmanual relay operation

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    5/12

    15 V

    85

    86

    S

    10 to 50 V

    93

    94

    L+

    L

    L

    N

    ~ =

    41

    42

    43

    47

    57

    51

    54

    48

    58

    52

    55

    49

    59

    53

    56

    mA

    10 to 50 V

    81

    34

    33

    82

    J

    JJ

    mA

    32

    31

    11

    84

    84

    15

    16

    83

    83

    12

    13

    k

    Electrical Connection

    Left:Liquisys M CLM 223,connections on the rearof the instrument

    Right:Liquisys M CLM 253,rear of instrument withreplaceable electronics

    box (E)

    Electrical connection ofLiquisys M CLM 223 / 253

    All connections to the panel-mountedinstrument CLM 223 are established via

    the terminal strips on the rear.

    In the case of the field instrument CLM

    253, all wires (including the measuringcell cable) are connected to terminals in

    a separate wiring compartment.

    All the wiring can remain in place if theinstrument needs repair because all

    repairs are limited to assemblyreplacement. So, dismantling the

    instrument and rewiring are no longernecessary.

    Contactingmeasuringcell

    Signal output 1:conductivity

    Signal output 2:temperature(optional)

    Digital input 1

    Digital input 2

    Aux. voltageoutput

    Alarm*

    Relay 1*

    Relay 2*

    Relay 3*

    Relay 4*

    optional

    *Contactstatus shown:no current orfault present

    Power supply

    E

    5

    (alternative)

    Electrodelesssensor

    Temperaturesensor Pt 100

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    6/12

    ENDRESS+HAUSER

    LIQUISYS

    CAL REL

    REL1

    REL1

    ALARM

    REL2

    REL2E

    ENDRESS+HAUSERLIQUISYS

    CAL REL

    REL1

    REL1

    ALARMREL2

    REL2E

    Dimensions

    3.54(90)

    5.47(139)

    0.24

    (6)

    3.78(96)

    Dimensions ofLiquisys M CLM 253

    Dimensions ofLiquisys M CLM 223

    6

    9.72(247)

    0.24(6)

    4.53

    (115)

    6.18(157)

    0.43

    (11)

    2.75(70)

    2.75

    (70)6.06(154)6.69(170)

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    7/12

    Mounting of CLM 223 The panel-mounted version is mountedusing the supplied tensioning screws.

    The required overall mounting depth isapproximately 6.89 in. (175 mm).

    Installation of panel-mounted housing

    7

    3.62+0.02

    (92+0.5)

    3.62+0.02

    (92+0.5)

    3.78(96)

    max. 1.77(45)

    0.24(6)

    5.47(139)

    approx.0.98(25)

    1.30(33)

    2.24(57)

    3.54(90)

    Flat gasket(included in delivery)

    Mounting cutoutaccording toDIN 43700

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    8/12

    6.18(157)

    0.31(8)

    0.63(16)

    4.53(115)

    Mounting of

    Liquisys M CLM 253

    8

    Wall mountingScrews: 0.24 (6 mm)Anchors: 0.31 (8 mm)

    Mounting on cylindricalpipes (shown withweather protection coverinstalled on right)

    Post mounting kit forinstallation of field housingon cylindrical pipes(horizontal or vertical,max. 2.36 (60 mm);see Accessories)

    3.54(90)

    1.18(30)

    0.31(8)

    2.36(60)

    0.24(M 6)

    3.54(90)

    2.76(70)

    0.31(8)

    0.24(6)2.76

    (70)3.54(90)

    2.76(70)

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    9/12

    Manufacturer Endress+Hauser

    Product designation Liquisys M CLM 223Liquisys M CLM 253

    Measured quantities conductivity, resistance, temperature

    Contacting conductivity/resistance measurement

    Measuring range conductivity: 0 to 2000 mS/cm (uncompensated)resistance: 0 to 86 mcm

    Usable cell constant k= 0.0025 to 99.99 cm-1

    Maximum length of cable to measuring cell conduct ivity: 330 ft . (100 m)resistance: 50 ft. (15 m)

    Measuring frequency conductivity: 300 to 1000 Hzresistance: 30 to 400 Hz

    Toroidal conductivity measurement

    Measuring range 0 to 2000 mS/cm (uncompressed)

    Usable cell constant k= 0.0025 to 99.99 cm-1

    Maximum length of cable to measuring cell 200 ft. (60 m) (CLK 5)

    Measuring frequency 2 kHz

    Minimum span for 0 / 4 to 20 mA signal meas. value betw. 0 and 199.9 S/cm: 20S/cmmeas. value betw. 200 and 1999S/cm: 200S/cmmeas. value betw. 2 and 19.99 mS/cm: 2mS/cmmeas. value betw. 20 and 2000 mS/cm: 200 mS/cm

    Temperature measurement

    Temperature sensor Pt 100

    Measuring range -31 to +482 F (-35 to +250 C)

    Temperature offset range 68 F ( 20 C)

    Temperature compensation

    Compensation types linear, NaCl, table,(with contacting cells only: ultrapure water)

    Range -31 to +482 F (-35 to +250 C)

    Reference temperature 77 F (25 C)

    Digital inputs 1 and 2

    Voltage 10 to 50 VDC

    Current consumption max. 10 mA

    Conductivity/resistance signal output

    Current range 0 / 4 to 20 mA, galvanically separated;error current 2.4 / 22 mA

    Load max. 500

    Output range adjustable

    Isolation voltage max 350 Vrms

    / 500 V DC

    Overvoltage (lightning) protection acc. to EN 61000-4-5: 1995

    Temperature signal output (optional)

    Current range 0 / 4 to 20 mA, galvanically separated

    Load max. 500

    Output range adjustable, 10 to 100% of upper range value

    Isolation voltage max 350 Vrms / 500 V DCOvervoltage (lightning) protection acc. to EN 61000-4-5: 1995

    Auxiliary voltage output

    Output voltage 15 V 0.6 V

    Output current max. 30 mA

    Contact outputs (floating changeover contacts)

    Switching current with ohmic load (cos = 1) max. 2 A

    Switching current with inductive load (cos = 0.4) max. 2 A

    Switching voltage max. 250 V AC, 30 V DC

    Switching power with ohmic load (cos = 1) max. 1250 VA AC, 150 W DC

    Switching power with inductive load (cos = 0.4) max. 500 VA AC, 90 W DC

    9

    Technical DataGeneral specifications

    Input

    Output

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    10/12

    Technical Data

    (continued)Output (continued)

    Accuracy

    Ambient conditions

    Physical data / design

    Power requirements

    10

    Limit contactor

    Pickup / dropout delay 0 to 2000 seconds

    Controller

    Function (adjustable) pulse length / pulse frequency controller

    Controller response PID

    Proportional band Kp: 0.10 to 10.00

    Period for pulse length controller 0.5 to 999.9 seconds

    Frequency for pulse frequency controller 60 to 180 min *

    Alarm

    Function (switchable) steady / fleeting contact; normally closed / open

    Alarm threshold adjustment range conductivity / resistance / temperature:complete measuring range

    Alarm delay 0 to 2000 seconds

    Conductivity measurement

    Deviation of indication1 max. 0.5% of measured value 4 digits

    Reproducibility max. 0.2% of measured value 2 digits

    Measurement deviation1, conductivity signal output max. 0.75% of current output range

    Temperature measurement

    Resolution 0.2 F or 0.1 C

    Deviation of indication1 max. 1.0% of measuring range

    Measurement deviation1, temperature signal output max. 1.25% of current output range

    Nominal operating temperature 14 to 131 F (-10 to +55 C)

    Ambient temperature ( limit operating conditions) -4 to 140 F (-20 to +60 C)

    Storage and transport temperature -13 to 149 F (-25 to +65 C)

    Relative humidity (nominal operating conditions) 10 to 95%, non-condensing

    Protection class of panel-mounted unit NEMA 3S (IP 54 front, IP 30 housing)

    Protection class of field housing NEMA 4X (IP 65)

    Electromagnetic compatibility interference emission acc. to EN 50081-1: 1992interference immunity acc. to EN 50082-2: 1995

    Dimensions of panel-mounted unit (H x W x D) 3.78 x 3.78 x 5.71 in. (96 x 96 x 145 mm)

    Installation depth approx. 6.89 in. (175 mm)

    Dimensions of field housing (H x W x D) 9.72 x 6.69 x 4.53 in. (247 x 170 x 115 mm)

    Weight of panel-mounted unit max. 1.54 lb. (0.7 kg)

    Weight with field housing max 5.07 lb. (2.3 kg)

    Display LC display, two lines, five and nine digits, withstatus indicators

    Materials

    Housing of panel-mounted unit polycarbonate

    Front membrane polyester, UV-resistant

    Field housing ABS PC Fr

    Supply voltage 100 / 115 / 230 V AC +10 / -15%, 48 to 62 Hz24 V AC/DC +20 / -15%

    Power consumption max. 7.5 VA

    Fuse protection fine-wire fuse, medium time-lag, 250 V / 1 A

    * acc. to IEC 746-1, for nominal operating conditions

    Subject to modifications.

  • 7/31/2019 Analizadores de Conductividad Endress Hauser

    11/12

    AccessoriesMounting accessories

    Type Features Order number

    Pendulum For pendulum suspension of assemblies 50080196

    frame e.g. CYA 611

    Material: 304 SS

    Post mounting Kit for mounting of field housing on 50086842kit horizontal or vertical pipes, max.

    diameter, 2.5 (60 mm)

    Material: 304 SS

    Holders

    Type Features Applications

    CYA 611 Immersion assembly with G 1, G 3/4 Waste water, water

    or 3/4 NPT thread

    Two-electrode conductivity measuring cells (selection)

    Type Features Applications

    CLS 15 Cell constant: k = 0.1 cm-1 or 0.01 cm-1 Reverse osmosis

    Process connections: 3/4 NPT, 1-1/2 filter, ion exchangerTri-clamp

    CLS 19 Cell constant: k = 0.1cm-1 or 0.01 cm-1 Boiler feed, control

    Process connections: 3/4 NPT

    CLS 21 Cell constant: k = 1 cm-1 Service water, boiler

    Process connections: 1 NPT and feed water2 Tri-clamp

    Electrodeless conductivity measuring sensorsType Features Applications

    CLS 50 Cell constant: k=2 cm-1 Concentration

    Process connections: G 3/4, 1 NPT measurement,2 ANSI flange incl. 16.4 ft (5 m) attached product monitoring,cable chemical industries

    CLS 52 Cell constant: k=5.9 cm-1 Concentration

    Process connections: 2 Tri-clamp measurement,and Perlick product monitoring,

    chemical industries,

    food industries

    11

    Cables

    Type Features Order number

    CYK 71 Standard measuring cable for connection 50085333

    of two electrode measuring cells with andwithout Pt 100 (also usable for extension)

    CLK 5 Extension cable for CLS 50 and CLS 52 50085473

    sensors