ANALYSIS OF PROSTAGLANDINS WITH THE QUATTRO PREMIER XE Kate Yu and Donald Kwet Waters Corporation, Milford, MA, USA INTRODUCTION The accurate and precise measurement of candidate pharmaceuticals in biological fluids, environmental pollutants in soils and water or the measurement of toxins in various food matrices requires highly sensitive and specific assays. Over the past 10 years, LC/MS and LC/MS/MS using multiple reaction monitoring (MRM) has become the technique of choice for the trace analysis of analytes in complex mixtures. This has been made possible by the advent of atmospheric pressure ionization sources, such as ESI and APCI. The highly sensitive and specific nature of the tandem quadrupole mass spectrometer facilitates simple method development and highly selective assays. The Waters ® Micromass ® Quattro Premier XE mass spectrometer with its T-Wave ™1 technology has been specifically designed with high sensitivity analysis in mind. This new mass spectrometer incorporates enhanced electronics and a new high sensitivity detector for improved performance. Figure 1. Comparative response for prostaglandin F 2a on the Quattro Premier and Quattro Premier XE. The superior negative ion sensitivity of the Quattro Premier XE is illustrated using the analysis of three prostaglandins (D 2 , E 2 and F 2a ), which are acidic steroidal compounds. INCREASING NEGATIVE ION SENSITIVITY The negative ion sensitivity of the Quattro Premier XE was compared to that of the Quattro Premier. The prostaglandin compounds were analyzed by LC/MS/MS using the Waters ® ACQUITY UPLC System. A 5 µL aliquot of the sample was loaded onto an ACQUITY UPLC BEH 2.1 x 50 mm C 18 , 1.7 µm column. The column was eluted with a gradient of 10–95% B over 2 minutes at 0.8 mL/min, where A=10% acetonitrile in 0.1% formic acid and B=80/20% acetonitrile/methanol. The column eluent was monitored by negative ion ESI, using the MRM transitions 351>271 for prostaglandins D 2 and E 2 , and the MRM transition 353>193 for prostaglandin F 2a . The resulting XIC chromatograms for prostaglandin F 2a using the Quattro Premier XE and Quattro Premier are displayed in Figure 1. Here we can see that with the Quattro Premier XE peak area of the analyte is increased by a factor of approximately 10 compared to the Quattro Premier. A ~9-fold increase in peak area was obtained for prostaglandins E 2 and ~10-fold for D 2 (data not shown). Peak Area = 6 PtP S/N = 11.23 PGF2a, 353.5>193.1 Time 0.50 1.00 1.50 2.00 2.50 3.00 % 4 0.50 1.00 1.50 2.00 2.50 3.00 % 4 MRM of 1 Channel ES- 2.74e3 MRM of 1 Channel ES- 2.74e3 Peak Area = 59 PtP S/N = 61.22 Peak Area = 6 PtP S/N = 11.23 PGF2a, 353.5>193.1 Time 0.50 1.00 1.50 2.00 2.50 3.00 % 4 0.50 1.00 1.50 2.00 2.50 3.00 % 4 MRM of 1 Channel ES- 2.74e3 MRM of 1 Channel ES- 2.74e3 Peak Area = 59 PtP S/N = 61.22 Peak Area = 6 PtP S/N = 11.23 PGF2a, 353.5>193.1 Time 0.50 1.00 1.50 2.00 2.50 3.00 % 4 0.50 1.00 1.50 2.00 2.50 3.00 % 4 MRM of 1 Channel ES- 2.74e3 MRM of 1 Channel ES- 2.74e3 Peak Area = 59 PtP S/N = 61.22 Quattro Premier Quattro Premier XE Peak Area Increased 9.8 Times Peak-to-Peak (PtP) S/N Increased 5.4 Times The Waters Micromass Quattro Premier XE Mass Spectrometer with the Waters ACQUITY UPLC System 1 The traveling wave device described here is similar to that described by Kirchner in US Patent 5,206,506 (1993).

Upload: others

Post on 11-Aug-2020

0 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ANALYSIS OF PROSTAGLANDINS WITH THE QUATTRO PREMIER …€¦ · prostaglandins (D 2, E 2 and F 2a), which are acidic steroidal compounds. INCREASI NG NEGATIVE IO SENSITIVITY The negative

ANALYSIS OF PROSTAGLANDINS WITH THE QUATTRO PREMIER XE

Kate Yu and Donald KwetWaters Corporation, Milford, MA, USA

INTRODUCTION

The accurate and precise measurement of candidate pharmaceuticals in biological fluids, environmental pollutants in soils and water or the measurement of toxins in various food matrices requires highly sensitive and specific assays. Over the past 10 years, LC/MS and LC/MS/MS using multiple reaction monitoring (MRM) has become the technique of choice for the trace analysis of analytes in complex mixtures. This has been made possible by the advent of atmospheric pressure ionization sources, such as ESI and APCI. The highly sensitive and specific nature of the tandem quadrupole mass spectrometer facilitates simple method development and highly selective assays.

The Waters® Micromass® Quattro Premier™ XE mass spectrometer with its T-Wave™1 technology has been specifically designed with high sensitivity analysis in mind. This new mass spectrometer incorporates enhanced electronics and a new high sensitivity detector for improved performance.

Figure 1. Comparative response for prostaglandin F2a

on the Quattro Premier and Quattro Premier XE.

The superior negative ion sensitivity of the Quattro Premier XE is illustrated using the analysis of three prostaglandins (D2, E2 and F2a), which are acidic steroidal compounds.

INCREASING NEGATIVE ION SENSITIVITY

The negative ion sensitivity of the Quattro Premier XE was compared to that of the Quattro Premier. The prostaglandin compounds were analyzed by LC/MS/MS using the Waters® ACQUITY UPLC™

System. A 5 µL aliquot of the sample was loaded onto an ACQUITY UPLC BEH 2.1 x 50 mm C18, 1.7 µm column. The column was eluted with a gradient of 10–95% B over 2 minutes at 0.8 mL/min, where A=10% acetonitrile in 0.1% formic acid and B=80/20% acetonitrile/methanol. The column eluent was monitored by negative ion ESI, using the MRM transitions 351>271 for prostaglandins D2 and E2, and the MRM transition 353>193 for prostaglandin F2a. The resulting XIC chromatograms for prostaglandin F2a

using the Quattro Premier XE and Quattro Premier are displayed in Figure 1. Here we can see that with the Quattro Premier XE peak area of the analyte is increased by a factor of approximately 10 compared to the Quattro Premier. A ~9-fold increase in peak area was obtained for prostaglandins E2 and ~10-fold for D2

(data not shown). Peak Area = 6

PtP S/N = 11.23

PGF2a, 353.5>193.1

Time0.50 1.00 1.50 2.00 2.50 3.00

%

4

0.50 1.00 1.50 2.00 2.50 3.00

%

4

MRM of 1 Channel ES-

2.74e3

MRM of 1 Channel ES-

2.74e3

Peak Area = 59

PtP S/N = 61.22

QP

QP-XE

Peak Area Increased 9.8 TimesPtP S/N Increased 5.4 Times

Peak Area = 6

PtP S/N = 11.23

PGF2a, 353.5>193.1

Time0.50 1.00 1.50 2.00 2.50 3.00

%

4

0.50 1.00 1.50 2.00 2.50 3.00

%

4

MRM of 1 Channel ES-

2.74e3

MRM of 1 Channel ES-

2.74e3

Peak Area = 59

PtP S/N = 61.22

QP

QP-XE

Peak Area = 6

PtP S/N = 11.23

PGF2a, 353.5>193.1

Time0.50 1.00 1.50 2.00 2.50 3.00

%

4

0.50 1.00 1.50 2.00 2.50 3.00

%

4

MRM of 1 Channel ES-

2.74e3

MRM of 1 Channel ES-

2.74e3

Peak Area = 59

PtP S/N = 61.22

QP

QP-XE

Peak Area Increased 9.8 TimesPtP S/N Increased 5.4 Times

Quattro Premier

Quattro Premier XE

Peak Area Increased 9.8 TimesPeak-to-Peak (PtP) S/N Increased 5.4 Times

The Waters Micromass Quattro Premier XE Mass Spectrometer with the Waters ACQUITY UPLC System

1The traveling wave device described here is similar to that described by Kirchner in US Patent 5,206,506 (1993).

Page 2: ANALYSIS OF PROSTAGLANDINS WITH THE QUATTRO PREMIER …€¦ · prostaglandins (D 2, E 2 and F 2a), which are acidic steroidal compounds. INCREASI NG NEGATIVE IO SENSITIVITY The negative

IMPROVING THE LIMIT OF DETECTION

To further illustrate the superior performance of the Quattro Premier XE, the Limit of Detection (LOD) was compared to that of the Quattro Premier for both Prostaglandin D2 and Prostaglandin E2. The resulting data is displayed in Figure 2. It is shown that with the same on-column loading, the peak-to-peak signal-to-noise value obtained from the Quattro Premier increased from 9.4 to 47.1 for Prostaglandin D2 and from 11.2 to 35.7 for Prostaglandin E2. Therefore, from this data we can ascertain that the new Quattro Premier XE produces a 3- to 5-fold increase in signal-to-noise and a 10-fold increase in peak area sensitivity.

Figure 2. Comparing the LOD of the Quattro Premier and Quattro Premier XE.

CONCLUSION

The new Quattro Premier XE mass spectrometer provides the ultimate in sensitivity in both positive and negative ion modes. This performance is further enhanced by the T-Wave-enabled fast scanning and fast MRM switching capabilities. The Quattro Premier XE, when combined with the superior chromatographic performance of the ACQUITY UPLC makes this the ideal LC/MS/MS platform for high sensitivity, high throughput analysis.

PGD2, 351.57>271.38

Time0.50 1.00 1.50 2.00 2.50 3.00

%

6

0.50 1.00 1.50 2.00 2.50 3.00

%

6

MRM of 1 Channel ES-

2.63e3

PtP S/N = 9.36

MRM of 1 Channel ES-

2.63e3

QP

QP-XE

PGE2, 351.44>271.26

Time0.50 1.00 1.50 2.00 2.50 3.00

%

4

0.50 1.00 1.50 2.00 2.50 3.00

%

4

MRM of 1 Channel ES-

3.20e3

MRM of 1 Channel ES-

3.20e3

PtP S/N = 35.66

PtP S/N = 11.15

PtP S/N = 47.12

Peak Area Increased 9.7 Times Peak Area Increased 8.8 Times

PGD2, 351.57>271.38

Time0.50 1.00 1.50 2.00 2.50 3.00

%

6

0.50 1.00 1.50 2.00 2.50 3.00

%

6

MRM of 1 Channel ES-

2.63e3

PtP S/N = 9.36

MRM of 1 Channel ES-

2.63e3

QP

QP-XE

PGE2, 351.44>271.26

Time0.50 1.00 1.50 2.00 2.50 3.00

%

4

0.50 1.00 1.50 2.00 2.50 3.00

%

4

MRM of 1 Channel ES-

3.20e3

MRM of 1 Channel ES-

3.20e3

PtP S/N = 35.66

PtP S/N = 11.15

PtP S/N = 47.12

Peak Area Increased 9.7 Times Peak Area Increased 8.8 Times

PGD2, 351.57>271.38

Time0.50 1.00 1.50 2.00 2.50 3.00

%

6

0.50 1.00 1.50 2.00 2.50 3.00

%

6

MRM of 1 Channel ES-

2.63e3

PtP S/N = 9.36

MRM of 1 Channel ES-

2.63e3

QP

QP-XE

PGE2, 351.44>271.26

Time0.50 1.00 1.50 2.00 2.50 3.00

%

4

0.50 1.00 1.50 2.00 2.50 3.00

%

4

MRM of 1 Channel ES-

3.20e3

MRM of 1 Channel ES-

3.20e3

PtP S/N = 35.66

PtP S/N = 11.15

PtP S/N = 47.12

Peak Area Increased 9.7 Times Peak Area Increased 8.8 Times

PGD2, 351.57>271.38

Time0.50 1.00 1.50 2.00 2.50 3.00

%

6

0.50 1.00 1.50 2.00 2.50 3.00

%

6

MRM of 1 Channel ES-

2.63e3

PtP S/N = 9.36

MRM of 1 Channel ES-

2.63e3

QP

QP-XE

PGE2, 351.44>271.26

Time0.50 1.00 1.50 2.00 2.50 3.00

%

4

0.50 1.00 1.50 2.00 2.50 3.00

%

4

MRM of 1 Channel ES-

3.20e3

MRM of 1 Channel ES-

3.20e3

PtP S/N = 35.66

PtP S/N = 11.15

PtP S/N = 47.12

Peak Area Increased 9.7 Times Peak Area Increased 8.8 Times

Quattro Premier

Quattro Premier XE

Prostaglandin D2 Prostaglandin E2

Quattro Premier

Quattro Premier XE