analysis ofa gmr-basedplethysmograph transducer and its ... · analysis ofa gmr-basedplethysmograph...

4
Analysis of a GMR-Based Plethysmograph transducer and its Utility for Real-time Blood Pressure Measurement Vinit Kumar Chugh', Kubera Kalyan', Anoop C. S2, Amit Patra l , Shubham Negi' 'Department of Electrical Engineering, TIT Kharagpur, India 2Department of Avionics, lIST, Trivandmm Abst,{/ct- The lIaller l>resents and analysis of a Giant Magneto Resistance (GMR)-based magneto I,lethysmograllh and illustrates its effieaev as a tool for real-time cuff-less measurement of Blood (BP). The l>r0llosed scheme emlllo)'s two GMR sensors and associated biasing and signal conditioning in its architecture. The delay between outllUt of the GMR sensors is used to estimate the BP. The methodology, circuits and signal processing stages used arc described in the lIaller. A Ilrotol)'lle of the GMR-sensing solution is develolled and tested. Initially, tests arc carried out to determine the quality and characteristics of the IIlethysmograllhs IIrOdueed b)' develolled sensor unit, in different conditions such as various body Ilositions, bias current etc. Good quality bio-signals were obtained during the above tests. Then, the cXllcrimcnts were conducted ou 29 volunteers to find the of develol,ed scheme as a BP monitor. The results obtained show that the lIerformanee of develolled BP monitor is within aeeelltable limits. Keywords- Blood Pressure, Pulse Transition Time, Pulse Wave Velocit)" Real time Blood Pressure measurement, Non- Invasive Blood Pressu re measu rement. I. INTRODUCTION The most important parameters for primary health diagnosis include Respiration Rate (RR), Heart Rate (HR) and Blood Pressure [1,2,3]. Non-invasive devices capable of measuring above health parameters have gained considerable importance in the recent years. Desirable features for such bio- instnunents include compactness and low-cost, good accuracy and sensitivity, low power operation, capability for real-time operation, low amount of calibration, etc. In this paper, we analyze the perfonnance of a magneto-plethysmograph and investigate its application for real-time blood pressure measurement. The magneto-plethysmograph is based on giant magneto resistance principle and the overall sensing solution is shown 10 possess the aforementioned desired features of a bio-inslmment. Different methods [4] are reported for monitoring of HR, RR and BP. Many of these lechniques are particular for single body parameter, while some oUlers could be used as a universal solution. Out of Ulese parameters, BP is Ule most important tool for diagnosis of cardiovascular diseases [5]. COImnonly used devices for BP measurement are Mercury Sphygmomanometer, Aneroid and Oscillometric devices. Mercury Sphygmomanometer is considered Ule benciunark for office assessment of BP [6]. Mercury Sphygmomanomeler teclmique possesses a constant threat of mercury spills. The accuracy of Aneroid sphygmomanometer is well established, however this device requires greater maintenance and more frequent calibrations [7]. Oscillometric Sphygmomanometer is an automated device and does not require observer participation for BP measurement [8]. The above methods are cuff based methods. They are accurate, however may cause discomfort 10 Ule examinee. These methods are also not suitable for continuous BP monitoring. Pulse Wave Velocity (PWV) is defined as velocity of arterial pulse Ihrough cardiovascular syslem and can be used as a good estimator of BP [9,10]. Based on PWV and BP correlation, cuff-less solutions for BP estimation using Phonocardiogram and Photoplethysmograph teclu1iques have been developed [II]. In Utis work we develop a GMR sensor system for PWV and BP measurement. The proposed system does not require Ule use of cardiogram signals and associated signal processing stages to estimate PWv. GMR sensor can provide Magneto Plethysmographic (MPG) signal equivalent to disturbance caused by arterial blood flow in magnetic field. The similar GMR sensing modality can be used for HR [12], RR [13] measurement. The system shown in Fig. I gives good MPG signals at different positions along the ann length. The dual GMR sensor configuration is used. GMR sensor outputs from different radial artery positions are similar, but differ in phase. This phase difference infom13tion is then utilised for estin13tion of Pulse Transition Time (PTT) [14]. PWV can be calculated as a division of distance between signal acquisition Fig. I. Schematic Diagram of the developed Prototype showing typical PIT ror a volunteer and detailed analog conditioning circuit. 978-1-5090-2809-2117/$31.00 ©2017 IEEE 1704

Upload: others

Post on 01-Aug-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Analysis ofa GMR-BasedPlethysmograph transducer and its ... · Analysis ofa GMR-BasedPlethysmograph transducer and its Utility for Real-timeBlood Pressure Measurement ... {/ct-The

Analysis of a GMR-Based Plethysmograph transducer and itsUtility for Real-time Blood Pressure Measurement

Vinit Kumar Chugh', Kubera Kalyan', Anoop C. S2, Amit Patra l, Shubham Negi'

'Department of Electrical Engineering, TIT Kharagpur, India

2Department of Avionics, lIST, Trivandmm

Abst,{/ct- The lIaller l>resents stud~' and analysis of a GiantMagneto Resistance (GMR)-based magneto I,lethysmograllhand illustrates its effieaev as a tool for real-time cuff-lessmeasurement of Blood P~essure (BP). The l>r0llosed schemeemlllo)'s two GMR sensors and associated biasing and signalconditioning in its architecture. The delay between outllUt of theGMR sensors is used to estimate the BP. The methodology,circuits and signal processing stages used arc described in thelIaller. A Ilrotol)'lle of the GMR-sensing solution is develolledand tested. Initially, tests arc carried out to determine the qualityand characteristics of the IIlethysmograllhs IIrOdueed b)'develolled sensor unit, in different conditions such as variousbody Ilositions, bias current etc. Good quality bio-signals wereobtained during the above tests. Then, the cXllcrimcnts wereconducted ou 29 volunteers to find the feasibilit~, of develol,edscheme as a BP monitor. The results obtained show that thelIerformanee of develolled BP monitor is within aeeelltablelimits.

Keywords- Blood Pressure, Pulse Transition Time, PulseWave Velocit)" Real time Blood Pressure measurement, Non­Invasive Blood Pressu re measu rement.

I. INTRODUCTION

The most important parameters for primary health diagnosisinclude Respiration Rate (RR), Heart Rate (HR) and BloodPressure [1,2,3]. Non-invasive devices capable of measuringabove health parameters have gained considerable importancein the recent years. Desirable features for such bio­instnunents include compactness and low-cost, good accuracyand sensitivity, low power operation, capability for real-timeoperation, low amount of calibration, etc. In this paper, weanalyze the perfonnance of a magneto-plethysmograph andinvestigate its application for real-time blood pressuremeasurement. The magneto-plethysmograph is based on giantmagneto resistance principle and the overall sensing solutionis shown 10 possess the aforementioned desired features of abio-inslmment.

Different methods [4] are reported for monitoring of HR,RR and BP. Many of these lechniques are particular for singlebody parameter, while some oUlers could be used as auniversal solution. Out of Ulese parameters, BP is Ule mostimportant tool for diagnosis of cardiovascular diseases [5].COImnonly used devices for BP measurement are MercurySphygmomanometer, Aneroid and Oscillometric devices.Mercury Sphygmomanometer is considered Ule benciunark foroffice assessment of BP [6]. Mercury Sphygmomanomelerteclmique possesses a constant threat of mercury spills. Theaccuracy of Aneroid sphygmomanometer is well established,however this device requires greater maintenance and morefrequent calibrations [7]. Oscillometric Sphygmomanometeris an automated device and does not require observerparticipation for BP measurement [8]. The above methods arecuff based methods. They are accurate, however may causediscomfort 10 Ule examinee. These methods are also notsuitable for continuous BP monitoring.

Pulse Wave Velocity (PWV) is defined as velocity ofarterial pulse Ihrough cardiovascular syslem and can be usedas a good estimator of BP [9,10]. Based on PWV and BPcorrelation, cuff-less solutions for BP estimation usingPhonocardiogram and Photoplethysmograph teclu1iques havebeen developed [II]. In Utis work we develop a GMR sensorsystem for PWV and BP measurement. The proposed systemdoes not require Ule use of cardiogram signals and associatedsignal processing stages to estimate PWv. GMR sensor canprovide Magneto Plethysmographic (MPG) signal equivalentto disturbance caused by arterial blood flow in magnetic field.The similar GMR sensing modality can be used for HR [12],RR [13] measurement. The system shown in Fig. I gives goodMPG signals at different positions along the ann length. Thedual GMR sensor configuration is used. GMR sensor outputsfrom different radial artery positions are similar, but differ inphase. This phase difference infom13tion is then utilised forestin13tion of Pulse Transition Time (PTT) [14]. PWV can becalculated as a division of distance between signal acquisition

Fig. I. Schematic Diagram of the developed Prototype showing typical PIT ror a volunteer and detailed analog conditioning circuit.

978-1-5090-2809-2117/$31.00 ©2017 IEEE 1704

Page 2: Analysis ofa GMR-BasedPlethysmograph transducer and its ... · Analysis ofa GMR-BasedPlethysmograph transducer and its Utility for Real-timeBlood Pressure Measurement ... {/ct-The

ce.

Fig. 3. Experimental selup for real time blood pressure cstin13tiOll

FFT is implemented on one of the MPG signals to find tileheart rate and so ll,e pulse period of volmueer. Peaks aredetected by maxilll3 detection in every consecutive pulseperiod and the time stamps of occurrence o[ correspondingpeaks are recorded. PTT is calculated as llle difference of peaktime stamps of two MPG signals. PIT varies from beat to beatso an average is taken over 5 pulse periods to obtain a stablePIT. PIT values have a quantization error of 0.5 msec. Forour e,,-perimentation the distance between two measurementsites is fixed so PIT is an indirect measure of PWv. Forestimation of Blood Pressure from PIT we use llle equation

BP = P + piT

where P and Qare constants estimated using linear regressionover previously obtained data. Simplified Block diagram ofPeak detection algorilllln is shown in Fig. 2. The results havebeen discussed in following sectioll

W. EXPERIMENTAL SET-UP AND RESULTS

The developed e,,-perimental prototype is shown in Fig. 3.GMR sensors and biasing magnets combined are used forsignal acquisitioll Conditioning circuits I and 2 are used foranalog domain signal processing of two MPG signals. TypicalMPG waveforms are displayed on oscilloscope (GDS-2102Afrom GW l115tek). Arduino Mega 2560 is used [or digitaldomain processing and algorithmic implementation. LCD isnsed to display estimated blood pressure in real time. Tests areconducted on llle developed prototype. These tests include (I)per[onuance study of GMR plethysmograph when placed atdifferent positions, (2) cffect of bias current variation onplelllysmograph perfonnance and (3) efficacy study of GMRbased BP monitor.

A. Pel!ormance ofdeveloped plethysmograph at differentbody locations

The GMR-based plethysmographs are known to give goodquality signals when llle sensor unit is placed at radial arteryposition. Here, we study the quality of GMR signals atdifferent body locations. In this test the relative position ofGMR lC and magnet were fixed over a non-magnetic plate.The plate was gradually displaced in perpendicular directionto the radial artery as shown in Fig. 4(a). The observations arerccorded in Table I with related wavefonns in Fig. 4(b) andFig. 4(c). 11 can be observed ll13t ll10 placement of sensor on

sampling 256 Point Pulse MaximaRate

FFT .. Period .. (peak)1 kHz Cakulation Calculation Detection

•(" Blood

PTT~

Peak time~. ....""'''' .- stamp. estimation calculation detection

LCD

Fig. 2. Block diagram of Blood Pressure estimation algorilluu.

JI. MEASUREMENT METHODOLOGY

The prepared GMR based system comprise three mainparts (a) Signal acquisition stage, (b) Analog front-end, and (c)Digital back-end. These parts are explained in following sub­sections.

A. Signal Acquisition stageA GMR sensor when biascd properly can gcnerate a

differential signal analogous to the minute changes insurrounding magnetic field. This technique can be ntilised toget MPG signal corresponding to volumetric changes made inblood fiow due to cardiac activity. In our experimentation. wehave used two GMR sensors AA002-02 (15] from NVECorporation. Two pennanent magnets (Amazing Magnets ­D063D-N35) are used to bias ll10 GMR sensors in the linearregion o[ operation. The sensors are placed 9 cm apart on anon-magnetic base at radial artery position as can be seen inFig. 3. TllO two MPG signals thus observed have an inherentphase difference due to time taken by tllO pulse to reach from0110 measurement site to anolller. TllO separation betweensensors also ensures that tllO biasing magllOtic field of onesensor doesn·t affect tI1C otller. These differential signals arethen conditioned using analog front-end as described below.

B. Analogfront-end

The differential outputs of the MPG sensors are amplifiedusing Instntmentation Amplifier (rNA 129) with gain o[ 800.Thus obtained single ended amplified signals are passedthrough second order sallen key low-pass filter willl cutofffrequency 10Hz to suppress higher frequency noises andpower line interference. These filtered signals contain both HRand RR components. To remove RR component, we usesecond order sallen key high-pass filter with cutoff frequency0.7 Hz. Anolller second order low-pass filter was used tofurtJler reduce high frequency noises. The filters weredesigned using op-amp OP07 [rom Texas Instruments. Theconditioned signals thus obtained are shifted to 0-5V and areforwarded to digital back-end [or PWV computation.

C. Digital back-endThe analog signals (having primary fTequency component

1-2 Hz) are sampled at sampling frequency I kHz. 256 point

MPG signals

sites by PIT. Such system would be efficient in real timemeasurement of PWV in non- invasive IIl3JUler and would befree from setbacks of cuff based methods. Viability of theproposed method is established by building a prototype andtesting on 29 volunteers. Detailed working of the proposedsystem is described in the next section. Section (III) deals withExperimental set-up and Results obtained and section OV)provides the concluding remarks to the experimentation andreferences.

1705

Page 3: Analysis ofa GMR-BasedPlethysmograph transducer and its ... · Analysis ofa GMR-BasedPlethysmograph transducer and its Utility for Real-timeBlood Pressure Measurement ... {/ct-The

00 00 ~Fig. 5. MPG wavcfonns observed at Wrist, Lower Ann, and Upper Ann arc shown in (a), (b) and (c) respectively.

(a)"'~0.6IllA (b)lb~0.3IllA (c)"'~O.IIllA

Fig. 6. MFG wavcfonns observed for a volunteer at three different biasing current Clb).

different wrist positions was giving similar results. Theobservation can be supported by presence of Brachial andRadial artery across wrist. These observations prompted asimilar sensor displacement study along am), thecorresponding wavefonns are shown in Fig. 5. The waveformsobserved at wrist, lower-ann and upper-aml have similarsignal strength. It can be inferred from above study that theGMR sensor can provide good quality MPG signal even atpositions where arteries are a bit distant from surface of body.

TABLE I. DISPLACEMENT STUDY ACROSS WRIST

Displacement (mm)

Signal Strength (V)

B. Effect ofbias current on sensor unit

With keeping in mind the low power implemenlation forbuilding a compact prototype, the effect of different biascurrents on GMR sensor performance was studied. The supplycurrent to GMR Ie was provided through a simple ClUTentmirror circuit. Potentiometer was used in CIUTent mirror togenerate a variable current source. Bias current was measuredusing multimeter (Fluke 87V) in series with current mirror.The plethysmographic signals were recorded for current as lowas 271lA. The observed readings are reported in Table 2 withcorresponding waveforms in Fig. 6. It can be inferred fromFig. 6 that the GMR sensor used at low biasing currents can

also give quality output and hence is a fit transdncer fordevelopment of low power health monitoring system.

TABLE [I. BIAS CURRENT STUDY

Bias C,lrrellt0.9 0.6 0.4 0.3 0.2

(mAl0.1 0.05 0.027

Sigllal Strength 6.6 4.4 3.2 2.0 1.3 0.64 0.34 0.14(V)

SNR (in dB) 83 66 58 55 53 52 51 49

C. l:-]ficacy study of(jAlR based BP monitor

The feasibility of developed prototype (having 2 GMRsensors) for BP estilnation was tested for 29 volunteers.Typical MPG wavefonns obtained at two measurement sitesare shown in Fig. 7 for 3 out of the 29 vohmteers whoparticipated in the experiments. Systolic (SBP) and Diastolic(DBP) BP were recorded using Omron HEM-7120 as areference monitor [16]. PWV is measured by application ofPeak detection algorithm using developed prototype. Meanarterial pressure (MAP) for a volunteer was calculated as anaverage of SBP and DBP. PWV was plotted against SBP, DBPand MAP and a best linear fit for the data was obtained as canbe seen in Fig. 8, 9 and 10. The correlation coefficient (r) forPWV and BP was observed to be 0.61 for SBP, 0.70 for DBPand 0.74 for MAP. Thus the developed prototype can be usedfor cuff-less estimation of Blood Pressure.

1706

Page 4: Analysis ofa GMR-BasedPlethysmograph transducer and its ... · Analysis ofa GMR-BasedPlethysmograph transducer and its Utility for Real-timeBlood Pressure Measurement ... {/ct-The

(a) (b)Fig. 7. MPG signals showing Pulse Transition Time for three different volunteers.

(c)

10,---_--_--_--_--___,

°oo!:---::,00:;---':-:'''"0---,,~20;;----:,-:::30::-----=-',4·0

Systolic Blood Pressure (mm Hg)Fig. 8. Plot for Pulse Wave Velocity vs Systolic blood pressufC.

10,--_--_--~-~--_-___,

8

~ 6

~2

PWV= 0.lxSBP-7.91r =0.61

optimization of system, bias current study was done on sensor.Tltis Stlldy reflected that the design can provide significantoutput for extremely low bias currents. Thus the proposedsystem could be used to develop a low power, real-time, cuff­less solution for mgged measurement of BP. P and Qparameters are estimated using linear regression on preobtained data points, hence per patient calibration is requiredfor BP estin1ation using proposed modality. Detailed study onmedium and long tem1 stability of these parameters is thefuture exent of this work.

References

10,--_-_-_-_-_-_-_~

[IJ M. A Cretikos, R. Bellol11o, et at, "Respimlory mte: the neglected vitalsign," in Proc. MlA, 2008.

[2] Seccarcccia, Fulvia, ct aI., "Heart rale as a predictor of mortality: theMATISS project," in Proc. AJPH, 2001, pp. 1258-1263.

[3] Verdecch..ia, Paolo. cl al., "Ambulato')' blood pressure. An independelltpredictor of prognosis in essential hypertension," Hyper/ensioll, 1994,pp. 793-80 I.

[4] A. Arshad, S. Khan, CL al. , "A study 011 Health Monitoring System:Recent Advancements," in Proc. DUM Engineering Journal, 2014.

[5] World Health Fedemtion report [onlinel Available:hllp:/lwww.world-heart-fcderation.org/c,udiovascular-hea 11 h/e::l rd iov:lsclll:l r-d isease-risk-f:lelors/hypc r1c nsionl

[6J Ogedegbe, Gbenga, and T. Pickering, "Prineiples and Teclmiques ofBlood Pressure Measurement:" in Proc. Cardiology clinics, 2010, pp.571-586.

[7] Y. Ma, M. Tcmprosa, el aI., "Evaluating the Accuracy of an AneroidSphygmomanometer in a Clinical Trial Selling," in Proc. AJH, 2009,pp.263-266.

fS] 1. N. Amoorc, "Oscillometric sphygmomanometers: a critical appraisalof currenI tcchnology;" blood pressure moniroring, 2012, pp. 80-88.

[9J E. J Kim, C. G. Park, el at, "Relationship between blood pressureparameters and pulse wave velocity in nonllotensive aIxI hypertensivesubjects: invasivc study," in Proc. J. HUIIl. Hypertens., 2007.

[I OJ S. S. Naliar, A Seuteri, et aI., "Pulse Wave Velocity Is an IndependentPredictor of the Longitudinal Increase in Systolic Blood Pressure andof Incidcnt Hypertension in thc Baltimore Longitudinal Study ofAging," in Proe. lACC, 2008, pp. 1377-t383.

[11] S. N. Shukla, K. Kakwani, A. Patm, ci aI., "Noninvasive cuillcss bloodpressure measurement by vascular transit time," in Proc. IntemationalConference on VLSI Dcsign, Bangalore, 2015, pp. 535-540.

[I2J K. Klilyan, V. K. Chugh and C. S. Anoop, "Non-invasive heart mtemonitoring systcm using giant magncto resistance sensor:' in Proc.IEEE EMBC, 2016, Orlando, FL, 2016, pp. 4873-4876.

[13J V. K. Chugh, K. Kalyan and C. S. Anoop, "Feasibility sludy of a giantMagnelo-ResiSlllllce based respimtion mle monitor:" in Pmc. IEEEEMBC, 2016, Orlando, FL, 2016, pp. 2327-2330.

[I4J P. M. Nabeel, J. Joseph, and M. Sivapmkasam, "Magneticplcthysmograph transducers for local blood pulse wave vclocitymeasuremenl," in Proc. tEEE EMBC, 2014, pp. 1953-1956.

[151 Appliealion Notes for GMR Sensors, NVE Corp. [online]Available: http://www.nve.com/Downloadslapps.txlf.

[16] Omron Blood Pressure levcl indicator lonline] Availablc:hUDs:"ww,,, .omronhcalthcarc.il1/producl-dctail?md=HEM-7120.

95

••

PVW =0.17xMAP - 12.6r = 0.74

pvw= 0.17xDBP- 9.61r = 0.70

•065 70 75 80 85 00

Diastolic Blood Pressure (mm Hg)Fig. 9. Plot for Pulse Wave Velocity vs Diastolic blood pressure.

IV CONCLUSION

Mean Arterial Pressure (mm Hg)Fig. 10. PIOI for Pulse Wave Velocity vs Mean Arterial pressure.

080 85 00 95 100 105 110 115 120

A non-invasive cuff-less measurement scheme forestimating PWV and BP was presented in tltis paper. Thescheme uses a twin GMR-sensor configuration as the basicsensing clement. The scheme docs not require cardiogramsignals opposed to most ofexisting cuff-less BP measurementmodalities. The outputs were processed using analog front-endto obtain stable pletl1ysmograph signals. The ability of theplethysmograph to produce good quality bio signals wereillustrated using experimentation on developed prototype.Later, these signals were processed using simple, but efficientpeak detection algorithm and moving window approach toobtain delay between the sensor outputs as well as PWV andBP. The performance of the developed BP monitor was testedon vohmteer trials (29 volunteers). The correlation coefficientwas observed to be 0.61 for SBP, 0.70 for DBP and 0.74 forMAP. Worst case error for MAP estimation was observed tobe ±9 mm Hg (Fig. 10). As a potential direction of low power

1707