analyze challenging polar compounds with confidence · 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9...

2
Hydrophilic interaction chromatography (HILIC) allows you to retain and separate polar analytes using a standard LC system and common reversed-phase solvents. New Agilent InfinityLab Poroshell 120 HILIC chemistries combine innovative particle technology and bonded phases to improve retention and peak shape for polar analytes. So you can quickly and efficiently separate challenging, highly charged polar compounds—and achieve superior analytical sensitivity in MS mode using MS-friendly eluents. InfinityLab Poroshell 120 HILIC-Z columns Novel zwitterionic stationary phase bonded to 2.7 µm Poroshell 120 particles PEEK-lined column option for excellent peak shape and recovery of particularly challenging charged compounds High pH and temperature stability: Up to pH 12 and 80 ºC InfinityLab Poroshell 120 HILIC-OH5 columns Novel poly-hydroxy fructan phase bonded to 2.7 µm Poroshell 120 particles Offer alternate selectivity to HILIC and HILIC-Z phases Analyze Challenging Polar Compounds with Confidence Agilent InfinityLab Poroshell 120 HILIC columns Complete your HILIC workflow The InfinityLab Deactivator Additive and InifinityLab Poroshell 120 HILIC Columns provide a complete HILIC solution. This simple workflow reduces analyte interaction with metal in the system-increasing instrument sensitivity and saving time. www.agilent.com/chem/ infinitylab-additive Excellent retention, peak shape, and analytical sensitivity: Separation of underivatized amino acids by LC/MS Counts (%) vs. Acquisition Time (min) 1 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5 Phenylalanine Leucine Isoleucine Histidine Arginine Lysine Methionine Tyrosine Valine Proline Alanine Threonine Glycine Serine Glutamic Acid Aspartic Acid Cysteine Cystine InfinityLab Poroshell HILIC-Z 2.1 x 100 mm, 2.7 µm Conditions: Mobile phase A: 20 mM ammonium formate in H20, pH3 Mobile phase B: 90:10 ACN/H20 with 20 mM ammonium formate, pH3 Gradient: 100% B to 70% B over 10 min, return to 100% B Flow rate: 0.8 ml/min Temp: 30 ºC MS detection: Agilent MS-QQQ, MS2 SIM mode

Upload: vuongdung

Post on 25-Aug-2018

216 views

Category:

Documents


0 download

TRANSCRIPT

Hydrophilic interaction chromatography (HILIC) allows you to retain and separate polar analytes using a standard LC system and common reversed-phase solvents.

New Agilent InfinityLab Poroshell 120 HILIC chemistries combine innovative particle technology and bonded phases to improve retention and peak shape for polar analytes. So you can quickly and efficiently separate challenging, highly charged polar compounds—and achieve superior analytical sensitivity in MS mode using MS-friendly eluents.

InfinityLab Poroshell 120 HILIC-Z columns

– Novel zwitterionic stationary phase bonded to 2.7 µm Poroshell 120 particles

– PEEK-lined column option for excellent peak shape and recovery of particularlychallenging charged compounds

– High pH and temperature stability: Up to pH 12 and 80 ºC

InfinityLab Poroshell 120 HILIC-OH5 columns

– Novel poly-hydroxy fructan phase bonded to 2.7 µm Poroshell 120 particles

– Offer alternate selectivity to HILIC and HILIC-Z phases

Analyze Challenging Polar Compounds with Confidence

Agilent InfinityLab Poroshell 120 HILIC columns

Complete your HILIC workflow The InfinityLab Deactivator Additive and InifinityLab Poroshell 120 HILIC Columns provide a complete HILIC solution. This simple workflow reduces analyte interaction with metal in the system-increasing instrument sensitivity and saving time.

www.agilent.com/chem/infinitylab-additive

Excellent retention, peak shape, and analytical sensitivity: Separation of underivatized amino acids by LC/MS

Counts (%) vs. Acquisition Time (min)

1

2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9 9.5 10 10.5 11 11.5 12 12.5 13 13.5

Phen

ylal

anin

e

Leuc

ine

Isol

euci

ne

His

tidin

e

Argi

nine

Lysi

ne

Met

hion

ine

Tyro

sine

Valin

e

Prol

ine

Alan

ine

Thre

onin

e

Gly

cine

Serin

e

Glu

tam

ic A

cid

Aspa

rtic

Aci

d

Cyst

eine Cy

stin

e

InfinityLab Poroshell HILIC-Z 2.1 x 100 mm, 2.7 µm

Conditions: Mobile phase A: 20 mM ammonium formate in H20,

pH3 Mobile phase B: 90:10 ACN/H20 with 20 mM

ammonium formate, pH3 Gradient: 100% B to 70% B over 10 min, return

to 100% B Flow rate: 0.8 ml/min Temp: 30 ºC MS detection: Agilent MS-QQQ, MS2 SIM mode

Ordering Information

Learn more about separating polar compounds using reversed-phase mobile phases www.agilent.com/chem/poroshell-120

For Research Use Only. Not for use in diagnostic procedures.

This information is subject to change without notice.

© Agilent Technologies, Inc. 2017, 2018 Published in the USA, July 1, 2018 5991-8547EN

Column description Part number

InfinityLab Poroshell 120 HILIC-OH5, 2.1 x 50 mm, 2.7 µm 689775-601

InfinityLab Poroshell 120 HILIC-OH5, 2.1 x 100 mm, 2.7 µm 685775-601

InfinityLab Poroshell 120 HILIC-OH5, 2.1 x 150 mm, 2.7 µm 683775-601

InfinityLab Poroshell 120 HILIC-OH5, 4.6 x 50 mm, 2.7 µm 689975-601

InfinityLab Poroshell 120 HILIC-OH5, 4.6 x 100 mm, 2.7 µm 685975-601

InfinityLab Poroshell 120 HILIC-OH5, 4.6 x 150 mm, 2.7 µm 683975-601

UHPLC Guard, HILIC-Z, 2.1 mm x 5 mm, 2.7 µm, 3 pk 821725-947

UHPLC Guard, HILIC-Z, 3.0 mm x 5 mm, 2.7 µm, 3 pk 823750-948

UHPLC Guard, HILIC-Z, 4.6 mm x 5 mm, 2.7µm, 3 pk 820750-933

InfinityLab Poroshell 120 HILIC-Z, 2.1 x 50 mm, 2.7 µm 689775-924

InfinityLab Poroshell 120 HILIC-Z, 2.1 x 100 mm, 2.7 µm 685775-924

InfinityLab Poroshell 120 HILIC-Z, 2.1 x 150 mm, 2.7 µm 683775-924

InfinityLab Poroshell 120 HILIC-Z, 2.1 x 50 mm, 2.7 µm (PEEK lined) 679775-924

Column description Part number

InfinityLab Poroshell 120 HILIC-Z, 2.1 x 100 mm, 2.7 µm (PEEK lined) 675775-924

InfinityLab Poroshell 120 HILIC-Z, 2.1 x 150 mm, 2.7 µm (PEEK lined) 673775-924

InfinityLab Poroshell 120 HILIC-Z, 3.0 x 50 mm, 2.7 µm 689975-324

InfinityLab Poroshell 120 HILIC-Z, 3.0 x 100 mm, 2.7 µm 685975-324

InfinityLab Poroshell 120 HILIC-Z, 3.0 x 150 mm, 2.7 µm 683975-324

InfinityLab Poroshell 120 HILIC-Z, 4.6 x 50 mm, 2.7 µm 689975-924

InfinityLab Poroshell 120 HILIC-Z, 4.6 x 100 mm, 2.7 µm 685975-924

InfinityLab Poroshell 120 HILIC-Z, 4.6 x 150 mm, 2.7 µm 683975-924

Additive description Part numberInfinityLab Deactivator Additive, 50 mL 5191-4506

InfinityLab Deactivator Additive, 25 mL 5191-3940

Retention and separation of water soluble vitamins

Agilent InfinityLab is an optimized portfolio of LC instruments, columns, and supplies designed to work together in perfect harmony. Combined with Agilent OpenLAB software and Agilent CrossLab Services, Agilent InfinityLab provides you with end-to-end support to make every day more productive. Learn more at www.agilent.com/chem/infinitylab

This example demonstrates alternate selectivity across Poroshell HILIC phases using the same method conditions. Separations can be further optimized on each column by altering method conditions.

Conditions: A: 100 mM Ammonium Acetate + 0.5% Acetic Acid (pH ~4.6) in H2O,

B: CH3CN0.5 mL/min, 87% B for 1 min, 87-50% B in 4 min, 3 min re-equilibration

1 μL injection, 40 ºC, UV detection at 260 nm, 80 Hz

Water soluble vitamins (B&C)thiamine (B1)riboflavin (B2)niacin/nicotinic acid (B3)pyridoxide/pyridoxol (B6)folate/folic acid (B9)4-aminobenzoic acid/PABA (B10)cyanocobalamin (B12)asorbic acid (C)

250

200

150

100

50

0

0.5 1 1.5 2 2.5 3 3.5 min

mAU

250

200

150

100

50

0

0.5 1 1.5 2 2.5 3 3.5 min

mAU

250

200

150

100

50

0

0.5 1 1.5 2 2.5 3 3.5 min

mAU

B10

B6

B3

B2 B1 B9B12C

B10

B6

B1,B3

B2 B9B12C

B10

B6

B2 B1B9

B12C

B3

InfinityLab Poroshell 120 HILIC-OH5 2.1 x 100 mm 2.7 μm

InfinityLab Poroshell 120 HILIC-Z 2.1 x 100 mm 2.7 μm