    Description of Analyzerperformance

    (Accuracy & Precision)

  • 8/8/2019 Analyzers Accuracy

    2/15

    Accuracy & Precision

    What do our customers expect from our analyzers?

    How will they convince themselves?

    0

    5

    10

    15

    20

    25

    30

    35

    40

    45

    0 5 10 15 20 25 30 35 40 45

    Accepted Standard Laboratory test

    FT-NIRA

    nalyzerresult

    Lab error

    Analyzer error

    Measure a single sample many times

  • 8/8/2019 Analyzers Accuracy

    3/15

    Testing the performance with a single sample

    0

    5

    10

    15

    20

    25

    30

    35

    40

    45

    0 5 10 15 20 25 30 35 40 45

    Accepted Standard Laboratory test

    F

    T-NIRA

    nalyzerresult

    Lab error

    Analyzer error

    Error in repeated measurements= Repeatability

    Below we see different results for the FT-NIR analyzer and the AcceptedStandard laboratory test

    We can see that the analyzer agrees with the Accepted Standard

    Laboratory test because both are centered at 25.

  • 8/8/2019 Analyzers Accuracy

    4/15

    Accuracy

    Common (old) Definition of accuracy Accuracy is measure of deviation from true value.

    Who knows the true value!!

    Replace Accuracy with Agreement (NIST directive)

    Error in agreement between analyzer

    result and result obtained with acceptedstandard reference method.

  • 8/8/2019 Analyzers Accuracy

    5/15

    Agreement

    How do we determine it?

    Measure a sample or process product with FT-NIR analyzer and in

    laboratory using accepted standard method

    Compare the results

    For one measurement result can be offset or error

    Repeat procedure many times

    Many = at least 20 times

    20 comparisons using the same sample 20 comparisons tested by same operator on same day

    20 comparisons with different samples on different days tested in (different)

    lab by different operators

  • 8/8/2019 Analyzers Accuracy

    6/15

    Agreement

    Test Agreement between analyzer and Accepted Standard Laboratory test

    with many samples

    y = 0.999x + 0.014

    20

    22

    24

    26

    28

    30

    32

    34

    36

    38

    20 22 24 26 28 30 32 34 36 38

    Accepted Standard Laboratory test

    FT-NIR

    analyzer

  • 8/8/2019 Analyzers Accuracy

    7/15

    Estimation of agreement error

    0

    5

    10

    15

    20

    25

    30

    35

    40

    45

    0 5 10 15 20 25 30 35 40 45

    Accepted Standard Laboratory test

    FT-N

    IRA

    nalyzerresult

    Combined error

    The agreement error between analyzer and Accepted Standard Method is the vector

    sum of the repeatability of each.

    The agreement error is always greater than the repeatability of the the analyzer or theAccepted Standard Method

    When repeatability of both are same, agreement= 1.41x error of Accepted Standard

    Method

  • 8/8/2019 Analyzers Accuracy

    8/15

    Summary on Agreement

    (guaranteed) specification for agreement between our analyzer and the AcceptedStandard Method

    We must know the repeatability of the Accepted Standard Method.

    As estimated in the protocol

    As used by the customer

    For several operators, different samples, different labsetups etc.

    It is best to obtain real up to date statistical data from the customer Use work sheet for determination of standard error of duplicate determinations

    (Pooled standard deviation)

    We must also know the repeatability of the analyzer.

    Often much better than Accepted Standard Method. Rule of Thumb1: Check performance of customer lab carefully

    Rule of Thumb2: Leave some margin in guaranteed spec.

    Quote between 2 and 1.5 times the Standard error of duplicate determinations of

    Accepted Standard Method.

  • 8/8/2019 Analyzers Accuracy

    9/15

    Summary on Agreement (cont.)

    Agreement specification is needed when product is tested with Accepted

    Standard Method.

    By Government inspector to comply with regulatory quality

    By customer to provide custody transfer certification By customer to insure product in process quality

    You can see with term agreement, we are not bound by Accepted

    Standard Method.

    We can establish agreement specification between any 2 methods of

    analysis as long as we know their repeatability spec.

  • 8/8/2019 Analyzers Accuracy

    10/15

    Repeatability and Reproducibility

    Repeatability is specification of error for repeat determinations with allconditions the same

    Formerly called precision

    Reproducibility is specification of error for repeat determinations wheresome conditions may be different

    Different operators

    Different labs

    Different Spectrometers

    Our specification of absorbance reproducibility fromspectrometer to spectrometer gives us the possibility ofcalibration transfer

  • 8/8/2019 Analyzers Accuracy

    11/15

    Summary on Repeatability & Reproducibility

    Repeatability indicates how precise an analyzer can analyze theproduct

    The repeatability error is commonly much smaller than the agreement

    error In process monitoring it may be more important than agreement

    It insures that the product is always the same to within a small error.

    Reproducibility of our analyzers is important for customers who want toproduce the same product at different sites.

  • 8/8/2019 Analyzers Accuracy

    12/15

    Agreement

    OHvalue

    40

    45

    50

    55

    60

    65

    70

    75

    80

    85

    40 45 50 55 60 65 70 75 80 85

    Accepted Standard lab method

    FT-NIR

    analyzer

    RMSD=0.7 (2 Factors)

  • 8/8/2019 Analyzers Accuracy

    13/15

    Repeatability

    OH value

    70

    71

    72

    73

    74

    75

    76

    77

    78

    79

    80

    70 71 72 73 74 75 76 77 78 79 80

    Actual

    Predicted

    cal samples

    Validation

    Linear (cal samples)

    Repeatability RMSD 0.05

  • 8/8/2019 Analyzers Accuracy

    14/15

    Repeatability

    control chart 10 weeks

    y = 0.0002x + 57.202

    56

    56.2

    56.4

    56.6

    56.8

    57

    57.2

    57.4

    57.6

    57.8

    58

    0 20 40 60 80 100 120 140

    days

    O

    H

    value

    RMSD= 0.05

  • 8/8/2019 Analyzers Accuracy

    15/15

    Reproducibility

    System to System Reproducibility

    55

    56

    57

    58

    59

    60

    1 2 3 4 5 6

    FTLA2000-160 unit

    OH

    value

    STDEV (from unit to unit) = 0.11