“SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT
JASON WENDEROTH
PRINCE OF WALES AND LIVERPOOL HOSPITALS
SYDNEY
![Page 2: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/2.jpg)
TWO MAJOR CHALLENGES IN STENT DEPLOYMENT
• Stent torsion
• Stent prolapse into aneurysm
![Page 3: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/3.jpg)
THE TORSION PROBLEM
• Tortuous anatomy
• ”Round-the-world” access
• Long stents
![Page 4: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/4.jpg)
THE TORSION PROBLEM
• Torque on catheters transmitted to FDS
• Mid/proximal stent twists
• Non-opening
• Angioplasty-resistant
![Page 5: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/5.jpg)
THE PROLAPSE PROBLEM
• Broad-necked/giant aneurysms
• Aneurysms occupying >50% circumference of vessel
• Insufficient proximal, distal or lateral support during “push” phase of stent deployment
![Page 6: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/6.jpg)
FDS TORSION PROBLEM
• Alternate unsheathing of stent delivery catheter and stent
• Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent
• Intermediate catheter can be used to “nudge” proximal end of stent for optimal opening
![Page 7: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/7.jpg)
65M ex-smoker
![Page 8: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/8.jpg)
FIRST PROCEDURE
• Aneurysm crossed “round the world”, microcatheter in M2
• Intermediate catheter proximal to aneurysm
• Failed FDS deployment:
• PED 4.75 x 30mm
• PED 4.5 x 30mm
• FRED 4.5 x 26mm
• Acute stroke presented – procedure abandoned; rebooked
![Page 9: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/9.jpg)
![Page 10: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/10.jpg)
![Page 11: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/11.jpg)
THE ”SHEPHERD” TECHNIQUE
• Initial access beyond aneurysm via microcatheter
• Direct or “round the world”
• Standard or balloon microcatheter
• Torque accrues on primary access catheter
• Intermediate catheter advanced distally over primary catheter
• Stent delivery catheter (no torque) via intermediate catheter +/- buddy wire alongside if concerned re stent prolapse
![Page 12: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/12.jpg)
Intermediate catheter
• Distal balloon access
• Inflate balloon for stability
• Torque stored in loop
![Page 13: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/13.jpg)
• Tension on balloon catheter to remove loop
• Torque transmitted to balloon shaft
![Page 14: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/14.jpg)
• Balloon shaft straight, tension stored
• Use balloon shaft as rail
![Page 15: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/15.jpg)
• Advance intermediate catheter over balloon shaft
![Page 16: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/16.jpg)
• Place intermediate catheter well distal to aneurysm
• Can tension balloon shaft to cross aneurysm “lip”
![Page 17: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/17.jpg)
• Final position of intermediate catheter
![Page 18: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/18.jpg)
• Deflate balloon and remove
• May leave “buddy” wire in position if very difficult/tortuous access
![Page 19: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/19.jpg)
Buddy wire
![Page 20: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/20.jpg)
• Remove balloon
• Prepare stent-delivery catheter
![Page 21: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/21.jpg)
• Advance stent delivery catheter as distally as is safe
• Maintain intermediate catheter position +/-buddy wire
![Page 22: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/22.jpg)
• Commence unsheathing of stent lead wire
• Maintain intermediate catheter position
![Page 23: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/23.jpg)
• Withdraw intermediate catheter
• Expose stent delivery catheter in preparation for stent delivery
![Page 24: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/24.jpg)
• Gradual retraction of delivery catheter until stent opens distally and apposes wall
![Page 25: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/25.jpg)
• Slow unsheathing of delivery catheter to permit unsheathing of next stent segment
• Buddy wire can be very helpful here…
![Page 26: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/26.jpg)
• Next segment of stent delivered
• Manipulation by freely-moving delivery catheter with intermediate catheter support
![Page 27: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/27.jpg)
• Gradual retraction of delivery catheter and intermediate catheter until stent opens distally and apposes wall
![Page 28: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/28.jpg)
• Next segment of stent delivered
• Manipulation by freely-moving delivery catheter with intermediate catheter support
![Page 29: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/29.jpg)
65M ex-smoker
![Page 30: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/30.jpg)
65M ex-smoker
![Page 31: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/31.jpg)
65M ex-smoker
![Page 32: “SHEPHERD” TECHNIQUE FOR STENT DEPLOYMENT · 2017. 6. 16. · •Intermediate catheter (+/- buddy wire) provides support for delivery catheter and stent •Intermediate catheter](https://reader035.vdocuments.net/reader035/viewer/2022063009/5fc0f8609efa44054a2d1ba7/html5/thumbnails/32.jpg)
SUMMARY
• You’re probably doing this already
• If not, try it – it works!
• We find it useful with difficult anatomy to get our stents open