application handbook fast gc/gcms · 2018. 11. 1. · application handbook fast gc/gcms...

1

Upload: others

Post on 26-Feb-2021

16 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Application Handbook

Fast GC/GCMSChromatography Volume 2

Page 2: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Fast GC and GC/MS using narrow-bore columns: Principles and Applications

Prof. Luigi Mondello, University of Messina, ItalyDr. Hans-Ulrich Baier, Shimadzu Europa GmbH, Germany

Application BookFast GC/GCMS

Page 3: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

I.

II.

III.

IV.

V.

VI.

VII.

VIII.

IX.

Content

Overview 7

Basics of separation in fast GC 111. Column and stationary phase2. Length and peak resolution3. Inner diameter and peak resolution4. Film thickness and peak resolution5. Type of carrier gas and average linear velocity6. GC hardware requirements

Injection requirements in fast GC 15

Detector requirements in fast GC 191. Conventional detectors2. Mass spectrometric detector (quadrupole MS)

Applications in food analysis 231. Organophosphorous pesticides in food matrices

(GC-FPD, GC-FTD, GC/MS)2. The analysis of organochlorine pesticides in food matrices

using GC-ECD and NCI GC/MS. 3. Analysis of butter fatty acid methyl esters (GC-FID)

Application in the petrochemical field 31Kerosene analysis (GC-FID)

Applications in the fields of flavors & fragrances 321. Potential allergens in perfumes 2. Quality control of flavors (GC-FID)

Application in the environmental field 37PCB analysis with GC-ECD detection

Summary 39

Page 4: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential
Page 5: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2I. OVERVIEW

8

as chromatography is a well-established technique, employedin many applicational fields suchas environmental, food, petro-chemical and pharmaceuticalanalysis [1].

The capillary columns used mayhave solid fillings (PLOTcolumns) or a liquid film as a stationary phase. The latter areemployed in gas liquid chroma -tography (GLC), which is thesubject of the present booklet. In GLC, analyte separation isachieved by means of a dynamicthermal equilibrium between thestationary phase (liquid) andmobile phase (gas). The dimen-sions of the columns employed,which vary depending on theapplication, generally range frominner diameters between 0.25 to0.53 mm and lengths from 25 to60 m. The film thickness select-ed, dependent on the boilingpoints of the analytes, usuallyranges from 0.25 to 1 μm (insome cases up to 3 μm or evenmore).

Nowadays in daily routine work,apart from increased analyticalsensitivity, demands are alsomade on the efficiency in termsof speed of the laboratory equip-ment. Regarding the rapidity ofanalysis, two aspects need to beconsidered:

1. The costs in terms of timerequired, for example in qualitycontrol analysis

2. The efficiency of the analyticalequipment employed

It is well known that analyticaltime costs may be very high instandard GC applications. In thelast few years, fast GC methodshave been shown to be suitablefor routine laboratory work. Inthis respect, the employment ofnarrow-bore columns is of par-ticular interest [2 -5].

These analytical tools enable aconsiderable reduction of analy-ses times while peak resolution isessentially maintained. Anotherapproach worthy of mention isthe use of multicapillary columns[6-8] with more than 900 capil-lary tubes with inner diametersof about 40 μm.

The present overview is focussedon the use of narrow-borecolumns in fast gas chromatogra-phy. As mentioned, this type ofapproach enables substantialreductions in GC run-timeswithout compromising the chro-matographic resolution.

An example of the validity ofthis approach is illustrated inFigure I.1 (lower chromatogram)

which reports a fast GC analysis(RTX-5 10 m, 0.1 mm ID, 0.4 μmfilm thickness) on a capillary col-umn test mixture (Grob test mix-ture, Macherey und Nagel). Theconditions were as follows: injec-tion mode: split (200 :1), averagelinear carrier gas velocity: 120cm/s (502 kPa, H2), temperatureprogram: from 140 °C at 60°C/min to 280 °C. A convention-al application (DB 5 30 m, 0.25mm ID, 0.25 μm film thickness)on the same matrix is also shownin the diagram (upper chro-matogram). As apparent, thedegree of separation is even bet-ter in the fast GC applicationwhile a speed gain of approxi-mately 8 times is attained.

However, the GC hardware hasto meet some requirements inorder to enable the efficient useof narrow-bore columns. In thefollowing chapters, general andfundamental issues relative tofast gas chromatography are dis-cussed (injection systems, capil-lary columns, carrier gas anddetector characteristics).

All the chromatograms shownwere recorded with a ShimadzuGC-2010 or GCMS-QP2010with an AOC-20i autoinjectorand GCsolution or GCMSsolu-tion software for data elabora-tion.

G

Page 6: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

9

Chromatography Volume 2 I. OVERVIEW

6.0 7.0 8.0 9.0

Minutes

3

5

1

10.0 11.0 12.0 13.0 14.0 15.0

x 106

12

34

5

6 7

8

9

0.25 0.50 0.75 1.00

Minutes

2

4

0

1.25 1.50 1.75

x 104

FWHM = 0.48 s

12

3

4 5

6

78

9

FWHM = 3 s

6

Figure I.1:

Top: Chromatogram recorded with aGrob test mixture using a DB5 30 m,0.25 mm, 0.25 μm column.Bottom: Chromatogram recorded with the Grob test mixture using aRTX-5 10 m, 0.1 mm, 0.4 μm column.

Peak identification:1) n-Decane2) n-Octanole3) n-Undecane4) 2.6-Dimethylphenole5) 2.6-Dimethylaniline6) Decane acid methyl ester 7) Undecane acid methyl ester8) Dicyclohexylamine9) Dodecane acid methyl ester

Page 7: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential
Page 8: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

N = 16 • ––––wb

tR2

Both the polarity and the boilingpoints of the sample analytesmust be considered during theselection of a specific stationaryphase material. Nowadays, sever-al phase materials are availablefor typical fast GC applications.

An important issue in fast GCregards the narrow-bore column

Chromatography Volume 2

1. Column and stationary phase

II. BASICS OF SEPARATION IN FAST GC

12

sample capacity, which is consid-erably lower than with conven-tional capillaries (this aspect willbe discussed later).

Apart from the choice of the sta-tionary phase, other importantcolumn parameters to be optimi -zed are length, internal diameterand film thickness.

The number of theoretical pla -tes N of a column is a measure ofits separation efficiency and isdefined as [9]:

2. Length and peak resolution

R = ––––4

� N �––––� – 1

k’2––––k’2 + 1

k’ = –––– ,tM

t’R � = –––– = ––––k’1k’2

t’R1

t’R2

L = column lengthN = –––––––HETP

L

R � � LR = ––––––––––––

_12 [wa + wb]

� t

where wb is the base peak widthand tR is the retention time. Peakresolution is of particular interestas the main aim is to reduce theanalysis time while maintainingthe column resolving power (Fig-ure II.2). The resolution betweentwo peaks is defined as:

Resolution (R) on the other handdepends on the capacity factor kand the plate number N in thefollowing equation [9]:

Wa Wb

�t

Figure II.1: Peak resolution parameters for two subsequent peaks in a chromatogram

with k 2 = capacity factor of themore retained solute and � = separation factor:

where tM is the time required foran unretained analyte to passthrough the column (generallyreferred to as dead time). Therelationship between the lengthof the column, N and height of atheoretical plate (HETP) is asfollows:

thus the resolution is propor-tional to the square root of thecolumn length:

Consequently, doubling the col-umn length achieves a gain ofresolution of a factor of only1.41. Therefore typical columnlengths in fast GC are within the10 - 15 m range.

• Column length for fast GC: 10 - 15 m

1

2

3

4

5

Page 9: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

� � ––––dI4df

KD = k’ • �

13

Chromatography Volume 2 II. BASICS OF SEPARATION IN FAST GC

In general, the chromatographicpeak width increases with theretention time due to resistanceto mass transfer in the mobileand stationary phase contribu-tions. The influence of the innerdiameter with respect to the theoretical plate number is nowconsidered. The basic equationfor the retention of a substance is [9]:

The linear relationship betweenpeak resolution and stationaryphase thickness requires the lat-ter to be as thin as possible infast GC.

On the other hand, it should alsobe suited to the boiling points ofthe analytes (see also �). Sincethe column sample capacity alsodecreases with the film thickness(less amount of stationary phase),the split ratio needs to be adjust-ed depending on the concentra-tion of the sample. High splitratios ensure that the sample is

3. Inner diameter and peak resolution

4. Film thickness and peak resolution

HETPmin prop dI

where:dI,df = inner diameter/film

thickness of the column KD = distribution constant� = phase ratio

The distribution coefficient KDis dependent on the temperatureand the physical and chemicalproperties of the analytes andstationary phase. The retentiontime is therefore determined bythe inner diameter, film thicknessand length of the column forfixed material properties. Fortypical inner diameters and filmthicknesses such as 0.25 mm and

0.25 μm, the phase ratio is equalto approximately 250. This is atypical phase ratio value for amedium boiling point analyterange. In general, this parameteris used by the analyst to make afirst selection of the columndimension for a given distribu-tion of analyte boiling points.Column manufacturers usuallyoffer schematic tables in theircatalogues giving an overview oftypical � values for differentapplications. The value for � typ-ically varies between 50 (columnswith thick films for volatiles) and530 (wide-bore columns, highboilers).

For further elucidation, the socalled van Deemter curves [9]will be discussed in the followingchapter where HETP values fordifferent capillary columns areplotted versus the average linearcarrier gas velocity [u].

With regard to the experimentalcurves, the HETP values arederived from the peak widths(eq. 1 and 4) and are plotted as afunction of the average linearvelocity of the carrier gas. A typical curve is shown in Fig-

ure II.2 for 2 columns of differ-ing inner diameter [10]. The plateheight is characterized by a mini-mum value (HETPmin) at theoptimum linear velocity [9]. Both columns have a � value ofapproximately 250. For columnswith such a high phase ratio val-ue, the resistance to mass transferin the stationary phase can beneglected and HETPmin cantherefore be approximated to thecolumn ID [10]:

Average Linear Velocity (cm/sec)

Conventional and Ultra Fast

Van Deemter Curves C16 : 0 FAME

isothermal (150 °C)

0.00 50 100 150 200

0.10.20.30.40.50.60.70.80.91.0

HET

P (m

m)

30 m x 0.25 mm i.d. 0.25 μm film

10 m x 0.10 mm i.d. 0.10 μm film

Figure II.2: Van Deemter curves [10] for 0.25 mm and 0.1 mm ID columns. The phase ratio is 250 in both cases.

transferred rapidly from the glass insert to the column (seeIII. Injection).

As a conclusion, a typical filmthickness in fast GC of 0.1 μm(dI = 0.1 mm, � = 250) is used. In specific cases of highly vola -tile compounds film thicknessesof up to 0.4 μm (� = 62.5 for 0.1 mm dI) may be used.

• Columns in fast GC: film thicknesses are typically 0.1 μm (≤ 0.4 mm)

As seen from equation 7 and Fig-ure II.2 the minimum theoreticalplate height decreases linearlywith the capillary column innerdiameter. Furthermore, theascending part of the curves risesmore gradually, enabling theapplication of higher than opti-mum velocities with little loss interms of resolving power. Toconclude, the inner diameter ofsuitable columns for fast GC are:

• Columns for fast GC: inner diameter ≤ 0.15 mm

6

7

Page 10: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2II. BASICS OF SEPARATION IN FAST GC

14

In fast GC applications, particu-lar attention must be paid to thetype of carrier gas used. The VanDeemter curves have differentminima for different gas types[9]. This factor is shown in detailin Figure II.3.

As can be seen, the minimumHETP for the hydrogen curve isattained at a higher average linearvelocity when compared to othercarrier gas types. In addition, theascending part of the curve risesmore gradually for the hydrogencurve. The data illustrated in Figure II.3 were derived from a0.25 mm ID column. In the caseof smaller inner diameters, asalready referred to, the ascendingpart rises even more gradually asmentioned in section 3.

5. Type of carrier gas and average linear velocity

The aforementioned issues have direct consequences on GC instru-mentation requirements such as injection system, carrier gas pneu-matics, column oven heating/cooling and detection. These aspects aresummarized as follows:

GC instrument requirements

• fast injection �� autoinjector, rapid injection, highsplit ratios (see chapter III.)

• reduced internal diameter �� large head pressure, high maxi-mum operational head pressure(up to 970 kPa)

• temperature program �� average linear carrier gas velocityshould be constant over theentire analysis (the constant lin-ear velocity mode requires apressure program)

• high separation efficiency �� linear temperature ramps arehigher compared to standard GC[70 °C/min linear and higher(ballistic)]

6. GC hardware requirements

In most applications, a lineartemperature program is employ -ed rather than isothermal analy-sis. Under isobaric conditions,

4020 60

N2

HETP (mm)

[u] cm/s

HETP – [u] curve for C17 at 175 °C with k = 4.95

Capillary column: OV-101 25 m, 0.25 mm ID. 0.4 μm film

0.4

0.8

80

He

H2

Figure II.3: Van Deemter curves measured using different carrier gases (Shimadzu Corporation, Kyoto, Japan)

average linear velocity decreaseswith increasing temperatures [9].This is due to the fact that thegas viscosity increases linearly

• high final temperature �� fast cooling of the GC oven(2 min from 280 °C to 100 °C)

• high concentration range �� high split ratios should be available (total flow of up to1200 mL/min)

• sharp peaks in fast GC �� high sampling rates and low filter time constants (250 Hz, 4 ms-conventional detectors)

The data in brackets refer to specifications of the GC-2010.

In the example illustrated in Figure I.1 (see page 9), a 120 cm/s aver-age linear velocity was applied. In order to obtain this velocity an initial pressure of 502 kPa was required at the initial temperature of140 °C. The gas chromatograph maintains the average linear velocityconstant over the temperature program (60 °C/min up to 280 °C).

This is achieved by the application of a pressure program automa -tically calculated by the GC software, and leads to a final pressure of 660 kPa at 280 °C. The cooling time for this analysis is below 2 minutes and the cycle time is under 4 minutes.

with temperature. Hence, theseparation efficiency is reducedas the temperature increases (theoptimum velocity is generallyselected for the initial tempera-ture) and the analysis time is alsoprolonged when compared toconstant velocity conditions.

Consequently, the gas chromato-graph should have an automaticpressure regulator in order tokeep the average linear velocityconstant over the entire tempera-ture program. This issue leadsdirectly to fast GC hardwarerequirements.

Page 11: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential
Page 12: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2III. INJECTION REQUIREMENTS IN FAST GC

16

As can be observed in Figure I.1(see page 9), the peak widths athalf height (FWHM) are below0.5 s, much less if compared toconventional GC (about 3 s).The peak width is affected by thetime needed to transfer the sam-ple, vaporized in the glass insert,onto the column. The sampletransfer must be fast enough(movement of syringe plunger,purge speed of the glass insert).

In the case of a relatively slowsample transfer rate (such as insplitless injection, see FigureIII.1), this results in an initialsample band broadening on thefirst part of the column [11], which has a considerable effecton the whole separation efficien-cy. The well-known solventeffect indicated by Grob [11] isnot applicable as the column in -itial temperature has to be setbelow the solvent boiling point,making the analysis time longer,and in addition the film thicknessis very small. This requires thatthe glass insert is not overloadedwith vapor after liquid injection[11].

The sharpness of the peaks isalso affected. For example, a 1 μL n-hexane injection into a

Injection requirements in fast GCsplit/splitless injector results in avapor volume of about 175 μL at280 °C and a head pressure of100 kPa. In the case of acetone as solvent the vapor volumeunder the same conditions isabout 310 μL. Typical glass insertvolumes on the other hand arewithin 1-2 mL.

Whenever splitless injection isapplied in fast GC, a reducedvolume glass insert and highpressure injection are required.In general, sample transfer has aprofound influence on peakshape and needs to be optimized:

Samples with a high concentra-tion: high split ratio

At low concentration ranges (lowsplit ratio or splitless injection):

1. Use of glass insert with smallinner diameter to enhancepurging with carrier gas.

2. In the case of splitless injec-tion: high pressure injection.

1. Glass insert

Figure III.2 shows the effect of glass insert dimensions onpeak width. Here the Grob testmix was analyzed with a stan-

0.5 1.0 1.5

n-D

ecan

e

Glass insert with smallinner diameter

Standard glass insert

2.0 2.5 3.0 3.5 4.0

Dod

ecan

e ac

id m

ethy

l est

er

Dic

yclo

hexy

lam

ine

Und

ecan

e ac

id m

ethy

l est

er

Dec

ane

acid

met

hyl e

ster

Dim

ethy

lani

line

Dim

ethy

lphe

nole

n-U

ndec

ane

n-O

ctan

ole

Minutes

Figure III.2: Chromatograms relative to a Grob test mixture.Top: reduced ID (inner diameter) glass insert = 1 mm.Bottom: standard glass insert, 4 mm ID.

Figure III.1: Schematic of asplit/splitless injector (SPL-17): Fc: carrier gas, Fs: split flow, Fl: Flow through glass insert

Fc

Fs

Splitting point

Fl

Glass insert

Carrier Purge

Split

Column flow

Page 13: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

17

Chromatography Volume 2 III. INJECTION REQUIREMENTS IN FAST GC

dard glass insert (4 mm innerdiameter – bottom) and a glassinsert with a 1 mm ID (top). Allother parameters remainedunchanged. In particular, thepeak widths of the less retainedvolatile components were nar-rower when the glass insertdimension was reduced. Theeffect of purging, which wasmore effective for the reduced IDglass insert, is clearly demon-strated. The split ratio was 40 :1in both cases. For increased splitratios the peak widths were alsonarrower when using the stan-dard glass insert.

In the chromatogram illustratedin Figure I.1 (see page 9), a 200 :1split ratio was selected. For peakno 1, the peak width measuredwith the standard glass insert wasFWHM = 0.42 s (split ratio200 :1) in contrast to 1.56 s inFigure III.2 – bottom (split ratio40 :1).

2. High pressure injection

As discussed in chapter II (sec-tion 5), the carrier gas averagelinear velocity has a great influ-ence on the column separationefficiency. The applied head pres-sure is therefore of great impor-tance after sample transfer to thecolumn. In this respect, it is pos-sible to apply an increased pres-sure just for sample transfer (typ-ically for 30 s to 1 min). Afterthis period the pressure isdropped automatically to theanalytical pressure to ensure peakresolution. This operation mode

is defined as high pressure injec-tion. Obviously, the purging ofthe glass insert is increased whencompared to standard conditions.In chapter V (section 1), whereanalysis of organophosphorouspesticides is discussed, high pres-sure injection (600 kPa) wasapplied in combination with asplitless injection technique inorder to optimize peak shapes.

3. Sample capacity of capillarycolumns

The maximum amount of samplewhich may be introduced onto acapillary column without over-loading is lower for narrow-borecolumns. When excessive sampleamounts are injected onto a cap-illary, peak fronting effects areobserved [9]. The generation ofasymmetric peaks may obviouslylead to a substantial loss in reso-lution. Consequently, the inject-ed sample amount must be belowa threshold value, as may beobserved in Figure III.3 [10].Here the plate number (calculat-ed from the peak width accord-ing to eq. 1) and hence separationefficiency is plotted as a functionof sample amount for twocolumns of different internaldiameter. It can be seen that forstandard columns (0.25 mm ID)up to 50 ng solute quantities maybe injected with little or no lossof separation efficiency. For anarrow-bore column with a 0.1mm ID, analyte amounts of up to1 ng may be injected without los-ing separation efficiency.

Amount injected (ng)

0.1 1.0 10.0

60

70

80

90

100

Plate Number X 100025 m x 0.25 mm i.d. 0.25 μm film thickness

10 m x 0.10 mm i.d. 0.10 μm film thickness

100.0 1,000.0

Figure III.3: Number of theoretical plates as a function of sample amount for two different columns

Page 14: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential
Page 15: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2IV. DETECTOR REQUIREMENTS IN FAST GC

20

Filter time constant and sampling frequency

1. Conventional detectors

Signal height = 63.2 %

Time

Signal

Figure IV.1: Definition of filter time constant �

0.030 0.031 0.032

uV (x 10,000)

Minutes

0.0

0.5

1.0

1.5

2.0

0.033 0.034 0.035 0.036

4 ms

20 ms

50 ms

100 ms 200 ms

Filter time constant

Figure IV.2: FID signal of chlorodecane. The filter time constant � was changedfrom 4 to 200 ms. The sampling frequency was 250 Hz in all cases.

In this section, conventionaldetectors such as the flame ion-ization detector (FID), electroncapture detector (ECD), flamephotometric detector (FPD) andflame thermionic detector (FTDor NPD) are discussed.

Mass spectrometric detection isdescribed in the next section. As mentioned, typical fast GCpeak widths are 0.5 s or less. The detection system must meeta series of requirements in orderto maintain the peak width atcolumn outlet. In this section therequirements for FID are dis-cussed, but the principles arevalid for all of the conventionalGC detectors. In general, detec-tor peak broadening may becaused by two factors. Firstly,the possible presence of deadvolumes in the sample path.Modern GC detectors are gener-ally characterized by low deadvolumes. Furthermore, make-upgas is employed to optimize peakshapes through an additionalflow (typically 30 mL/min). Formass flow dependent detectors

(FID, FPD, FTD) this featureenhances sensitivity [9].

In modern GC, the detector sig-nal is digitalized in the detectorelectronics and the A/D convert-er passes digital data to the PC.Due to the presence of RCchain-containing circuits, anysignal amplifier has a certainresponse time with the presenceof some smoothing effect. Thisshould not influence the peakwidth which is determined bythe chromatographic process prior to the detector (injectionsystem, column). The parameterof detector electronics regardingthe response time is usuallyreferred to as filter time constant(FTC) [12]. The electronic detec-tor signal amplifier cannot followa signal in an infinitely fast way.It needs a certain time to followrapid signal changes. In FigureIV.1 this parameter is visualizedgraphically [12]. A theoreticalstep increase of signal is shownhere and the response of detectoramplifiers indicates that it fol-lows the signal with some time

delay (1-1/et). The time intervalwhich the detector electronicsneeds to reach the (1-1/e) frac-tion of such a step signal isdefined as filter time constant.

This parameter should beselectable [12] together with thenumber of data points recordedacross a peak. In an experiment,the influence of this parameteron real peak shapes was meas-ured. The effects of the injectionsystem and column on bandbroadening were reduced as fol-lows:

1. A 0.1 mm ID glass insert and a0.2 μL injection volume of justone compound (chlorodecane)with a high split ratio (600 :1).

2. Use of a 1 m column with a 0.1 mm ID and 0.1 μm film thickness; isothermal analysis at 200 °C.

3. Detector make-up gas flow of 60 mL/min.

Chlorodecane was diluted inmethanol. Figure IV.2 shows thechlorodecane peak which elutesat 0.0332 min under isothermalconditions. The filter time con-stant � was varied from 4 ms to200 ms. It can be clearly obser -

Filter time �

Page 16: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

21

Chromatography Volume 2 IV. DETECTOR REQUIREMENTS IN FAST GC

2. Mass spectrometric detector (quadrupole MS)

0.2 0.4 0.6 0.8 1.0 1.2 1.4

Filter time constant �

� = 100 ms

1.6 1.8 2.0 2.2 2.4 2.6 0.2 0.4 0.8 1.0 1.2 1.4

� = 10 ms

1.6 1.8 2.0 2.2 2.4 2.6

Figure IV.3: Chromatograms relative to a kerosene sample recorded at a filter time constant � of 100 ms and 10 ms respectively

5 ms 10 ms

Nor

mal

ized

SN

rat

io

Time constant

0

0.2

0.4

0.6

0.8

1

1.2

20 ms 50 ms 100 ms 200 ms 500 ms 1 s 2 s4 ms

Half width 10 s

Half width 1 s

Half width 0.1 s

Figure IV.4: Signal-to-noise ratio as a function of the filter time constant and for different peak widths (source: GC-2010 operation manual)

ved that increases in the filtertime constant affects both thepeak retention time and shape aspredicted from theory [12]. Thefull width at half maximum for a� value of 200 ms (a standardparameter in many GC systems)is about 3 s, whereas for a � valueof 4 ms the FWHM value isabout 40 ms. The effect of the filter time constant on peakshape obviously has a drasticeffect on resolution as shown inFigure IV.3. The figure illustratestwo kerosene very fast GC chro-matograms with applied filtertime constants of 100 and 10 msrespectively. The improvement interms of peak resolution is evi-dent in the 10 ms FTC analysis.This experiment shows theimportance of detector parameteroptimization in terms of peakresolution. In addition, the FTChas a considerable influence onthe signal-to-noise ratio (S/N).Figure IV.4 shows the normal-ized signal-to-noise ratio as afunction of the filter time con-stant. These curves are plottedfor different peak widths. For apeak width of 0.1 s the optimumfilter time constant is 20 ms withregard to the S/N ratio. Theoptimum FTC value under suchconditions is also dependent onthe sample concentration andranges between 4 and 20 ms.

A further electronic parameter ofimportance is the sampling rateor, in other words, the number ofdata points acquired per secondwhich are transferred from theA/D converter to the PC. Thenumber of points must be suffi-cient for the proper reconstruc-tion of a Gaussian peak (mini-mum 10 to 15 data points). If thenumber of data points is too low,data reproducibility is reduced.The filter time constant and thesampling frequency parametersnecessary for the range of peakwidths observed in gas chro-matography were summarized byJ.V. Hinshaw [12]. The latterreported that for peak widthsbetween 0.1 and 0.5 s the filtertime constant and sampling rateshould be about 10 ms and 20 to100 Hz respectively.

Minutes Minutes

The most common mass spectro-metric detector used in gas chro-matography is the quadrupoleMS. With regard to fast GC/MSanalysis, the MS detector obvi-ously needs to be fast enough forthe rapid eluting analyte bands.For identification purposes, GC/MS applications are carried

out in the full scan mode inorder to perform a MS librarysearch. In this operation mode,every data point corresponds to acomplete mass spectrum over themass range selected (scan).

After peak assignment, the SIMmode (selected ion monitoring)

may be used in order to increasethe analytical sensitivity.

In order to attain this data, twoparameters of the GC/MS hard-ware must be considered: firstly,the mass range selected by theanalyst must cover all the analyteion fragments in order to �

Page 17: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2IV. DETECTOR REQUIREMENTS IN FAST GC

22

3.87 3.88 3.89 3.90 3.91 3.92 3.93 3.94 3.95 3.96 3.97 3.98

6,086,2713.922

TIC*1.00

Minutes

m/zR.Time: 3.921 (Scan#: 3026)

m/zR.Time: 3.922 (Scan#: 3027)

Figure IV.5: Definition of scan speed, inter-scan delay and sampling frequency for a quadrupole MS

100

40 80 120 160 200 240 280

100

40 80 120 160 200 240 280

Scan N

m1

Intervall I:

1/I = sampling frequency

Inter-scan time

RF setup

m2 Scan N +1

2.4min. 2.5 2.4min. 2.5

Linalool

61,307,110 61,307,110TIC TIC

2.4min. 2.5

61,307,110 TIC

20 60 100 140 180

m/z

WileySI = 97

20 60 100 140 180

m/z

WileySI = 98

20 60 100 140 180

m/z

WileySI = 97

Figure IV.6: Linalool peak generated bya narrow-bore column (FWHM 0.6 s)and the spectra recorded at the begin-ning, apex and end of the peak afterbackground subtraction. Spectra simi-larities are 94, 96, 94 using the WileyMS library.

perform a library search for com-ponent identification (scan). Inthis case, the scan speed shouldbe high enough so that the rela-tive intensities of the fragmentsdo not change significantly frompoint to point across a narrowpeak. In this respect, a low scanspeed would result in low qualitylibrary matches. Secondly, thesampling frequency, which cor -responds to the number of scans(or the number of SIM datapoints) per second, is, as afore-mentioned, a fundamental para -meter. Fast GC/MS applicationsrequire a high quadrupole scanrate (up to 10,000 amu/s) and, inaddition, a low inter-scan deadtime (reset time of the electro -nics) so that the sampling rate may be sufficiently high (up to50 Hz/100Hz in scan/SIM modesfor the GCMS-QP2010). Thissituation is shown in Figure IV.5.

The true scanning speed of thequadrupole is calculated by thescan range divided by the intervaltime minus the RF setup timewhich ranges from 5 to 10 ms for the GCMS-QP2010. In Fig-ure IV.6 this is demonstrated byexperimental data. A linaloolpeak was scanned with a 20 Hzscanning frequency and a select-ed mass range of 30 to 350 amu.The analysis was carried out witha SPB5 10 m, 0.1 mm, 0.1 μm col-umn. The peak width at halfheight is about 0.6 s and about1.2 s at the base resulting in 24 data points across the peak.As can be seen, the quality of thespectra in three points across thechromatographic peak is good (a Wiley library search was per-formed). The spectra similarities,94, 96, 94, indicate that there isno skewing.

Page 18: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential
Page 19: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2V. APPLICATIONS IN FOOD ANALYSIS

24

1. Organophosphorous pesticides in food matrices

Organophosphorous pesti-cides (OPPs) are widelyused in agriculture

in order to protect a broad rangeof products. Unfortunately, thepresence of these toxic compoundsis often detected in foods.

The maximum allowable concen-trations of OPPs in foods is,

3.00

5

Minutes

Trichlorphon

1.75

1.50

1.25

1.00

0.75

0.50

0.25

0.00

2.00

10 15 20

RTX-5: 10 m, 0.1 mm, 0.4 μm

RTX-35: 30 m, 0.25 mm, 0.5 μm

Phosphamidon (E)

Azinphos - MeEthionFenthionPhosphamidon (Z)

Carbophenothion

ProfenophosMethidathion

Tolclophos - Me

DisulfotonDioxathion

3.00

1.50

1.25

1.00

0.75

0.50

0.25

0.00

1

4

Minutes

6 8

1.75

2.00

2.25

2.50

2.75

11

10

9

8

7

6

5

4

3

2

13

uV (x 1,000.000)

Figure V.1: Top: conventional GC chromatogram of organophosphorous pesticide standards in a tomato extract. Concentrations vary beween 125 and 1,000 pg. Bottom: fast GC organophosphorous pesticide analysis carried out with aRTX-5 : 10 m, 0.1 mm, 0.4 μm column.

Peak identification: 1. Dichlorphos, 2. Trichlorphon, 3. Dioxathion, 4. Phosphamidon (E), 5. Disulfoton, 6. Phosphamidon (Z), 7. Tolclophos methyl, 8. Fenthion, 9. Methidathion, 10. Profenophos, 11. Ethion, 12. Carbophenothion, 13. Azinophos methyl

however, strictly controlled [13].In Germany, the well-knownDFG S19 (Deutsche Forschungs-gemeinschaft) is employed forthe determination of traces ofthese substances in foods. ThisStandard is also present in Euro-pean regulations as DIN EN1528-3, DIN EN 12393-2. Inaddition, these methods define

the procedure for the determina-tion of organochlorine, nitrogen-containing and other types of pes-ticides. The procedure steps aremainly: 1. extraction and partition,2. gel permeation chromatography,3. mini silicagel chromatographyand 4. analysis with GC-FTD,GC-FPD or GC/MS.

A chromatogram relative to anOPP-spiked (amount rangebetween 125 and 1,000 pg) toma-to matrix is shown in Figure V.1.The column used was a RTX-3530 m, 0.25 mm, 0.5 μm. The fig-ure also shows the resultobtained with a RTX-5 (5 %phenyl), 10 m, 0.1 mm, 0.4 μmcolumn under fast GC condi-tions [14] (temperature programfrom 100 °C (1 min) at 80 °C/min to 220 °C at 10 °C/min to240 °C at 50 °C/min to 320 °C;carrier gas: H2 (120 cm/s). Theinjection volume was 2 μL in thesplitless mode. In order toincrease the sample transferspeed (see above) a 600 kPa highpressure injection was appliedfor 1 minute. The peak widths(FWHM) are approximately 3 sin the standard analysis and,below, 0.5 s in the chromatogramobtained with the narrow-borecolumn. As the area in general isdetermined by the sampleamount which was not changed(both runs were carried out inthe splitless mode) the signal-to-noise ratio was higher in the fastGC experiment. The retentiontime of azinophos methyl was 22minutes in the standard analysisand 9 minutes in the fast result.

A chromatogram derived from ablack tea analysis is shown inFigure V.2. The analytical resultindicated a contamination ofdioxathion (225.2 pg), phos-phamidon (67.1 pg), disulfoton(217 pg), dichlorphos (243 pg),ethion (145.1 pg) and azinophosmethyl (244.8 pg).

The limit of detection in a fastGC application may be derived,for example, from the results

12

Page 20: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

MethamidophosMethomylDiazinonChlorpyriphos-OEtCaptanEndosulfan sulfatePiperonyl butoxidePotasan

--

169313, 315, 212, 214

150, 14997, 386

-328, 192

NCI dominant Peaks(Da)Substance

25

Chromatography Volume 2 V. APPLICATIONS IN FOOD ANALYSIS

7.0 8.0 9.0

Dic

hlor

phos

uV (x 1,000,000)

Minutes

0.00

0.25

0.75

0.50

1.25

1.00

1.50

6.05.04.03.02.0

Azi

noph

os m

ethy

l

Ethi

on

Phos

pham

idon

(Z)

Dis

ulfo

ton

Phos

pham

idon

(E)

Dio

xath

ion

Figure V.2: Fast GC chromatogram of a black tea sample. For concentrations see text.

0 1 2 3 4 5

Ethy

lpar

athi

on

Minutes

95,00090,00085,00080,00075,00070,00065,00060,00055,00050,00045,00040,00035,00030,00025,00020,00015,00010,0005,000

0- 5,000

Bro

mop

hos

met

hyl (

i.s.)

Figure V.3: Fast GC chromatogram of an olive oil sample. For the quantitativedetermination of ethylparathion, bromophos methyl was added as internal stan-dard; concentration ethylparathion: 3 ppb.

4.0 5.0 6.0

Dia

zino

n

(x 10,000,000)

0.1

0.3

0.4

0.5

0.6

0.7

0.8

1.0

0.9

1.2

1.1

1.3

7.0

Pota

san

Pipe

rony

l but

oxid

e

Endo

sulfa

n su

lfate

Capt

an

Chlo

rpyr

opho

s

100 °C init. temp.

He 50 cm/sec

NCI 50 pg Splitless

EI with combi source 5 ng Split 20 : 1

100.0 125.0 150.0

0.0

2.5

5.0

7.5

175.0 200.0 225.0 250.0 275.0 300.0 325.0 350.0 375.0 400.0

EI

(x 10,000)

m/z 407.00 Abs. Inten

Base Peak: 272 / 79,712

102 121

0.2

133 143 157169

170 193 206217 226

231

229

237

242 263

270

272

276279 289

301 311 341349357 379

386

387

390395

0 Rel. Inten 0.00

100.0 125.0 150.00.0

2.5

5.0

175.0 200.0 225.0 250.0 275.0 300.0 325.0 350.0 375.0 400.0

(x 1,000,000)

m/z 351.70 Abs. Inten

Base Peak: 97 / 2,332,743

145

97

337392388371

11599 185158

386

209200 219 234 290270254 306 314350

86.363 Rel. Inten 3.70

Scan 50 - 400 amu

2000 amu / sec

Figure V.4: Total ion chromatograms (TIC) of an OPP standard mixture analysis carried in the EI and NCI modes

shown in Figure V.3 [15]. Thedata shown were obtained froman olive oil sample. A 3 ppb con-tamination of ethylparathion wasdetermined by using internalstandard (bromophos methyl)calibration. During samplepreparation, 2 g of olive oil wereadded to 2 mL of hexane. Theinternal standard (0.01 ppm bromophos) was then added. Aliquid extraction with acetonitrilewas then carried out and, follow-ing this, 7.5 mL was taken fromthe acetonitrile phase and dried.

Finally, 2 mL of hexane wasadded before GC injection.

GC/MS data:

In GC/MS, the electron impact(EI) mode for analyte ionizationis most widely employed. How-ever, in recent years the chemicalionization mode has becomemore popular. Sensitivity may beconsiderably enhanced for elec-trophilic molecules by negativechemical ionization (NCI) �

94, 141, 126, 111, 110, 12858, 105, 88, 73, 115, 162

304, 179, 137, 152, 276, 199314, 187, 258, 244, 97, 125, 35179, 117, 149, 264, 301, 182, 236

229, 272, 239, 207, 170, 387176, 177, 149, 178

192, 328, 176, 300, 272, 148

EI Peaks (Da)

Table V.1: EI and NCI dominant peaks observed with the organophosphorouspesticides

NCI

Page 21: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Equality

AldinThiobencarbPermethrinPyributicarbp,p'-DDDChlorpyrofosBitertanolIsofenphosQuinalphosp,p'-DDEPirimiphos methylp,p'-DDTHalfenproxEthoprophosDichloros

p,p'-DDTDimethoate�-LindaneCafenstroleFluvalinate-1Pyrifenox-EDiflufenicanHexaconazoleThiometonDiazinonProthiofosDifenoconazolePyridabenTeraconazoleChinomethionateBifenthrinInabenfideMyclobutanil

BifenoxFenvalerateChlorfenapyrCyfluthrinPyrethrin-2MEPEPNFenpropathrinAcrinthrinTrifluralinPhosaloneCyhalothrin�-CVPFlucythrinateButamifos�-BHCCaptanImibenconazole

> 10 times

MepronilPropiconazolLenacilTebucanzoleEtoxazoleTebufenpyradPyriproxyfenPyrimidifenMefenacetTriadimenolEtrimfosChlorobenzilateEsprocarbThenylchlorTricyclazoleIsoprocarb

Chromatography Volume 2V. APPLICATIONS IN FOOD ANALYSIS

26

3.0 4.0 5.0

Chlo

rpyr

ofos

(x 1,000,000)

0.00

0.50

0.75

1.00

1.25

1.50

1.75

2.25

2.00

2.75

2.50

0.25

3.00

3.25

6.0 7.0 8.0 9.0

Endo

sulfa

n su

lfate

Black Tea

Mix 0.5 pg

Figure V.5: Black tea and standard compound mixture (0.5 ppb of each compound) TIC chromatograms (NCI mode)

> 10 times

DichlofluanidFenalimolPendimethalinDieldrinEdifenphosTefluthrinCypermethrinThifluzamideAcetamipridPAP, PhenthoatePyrethrin-1EndrinDeltamethrin�-BHC�-BHC

Dicofol BunkaibutuMalathionParathionDimethylvinphosPyrifenox-ZParathion methylKresoxim methylFolpetCyanazineCadusafosTolclofos methylPretilachlorMalathionCaptafolFlutolanil

FosthiazateAcephateTerbacilPyraclofosPenconazolUniconazolepFensulfothionCyhalofop buthylButachlor

ChlorfentezinedegMetamidophosEPTCPropamocarbChlorprophamTerbufosBenfuresateDimethenamidAlachlorMetolachlorDiethofencarbFenthionFludioxonilMethoprenePaclobutrazolFlusilazoleCyproconazole

2 - 10 times 2 - 10 times

Comparison of the sensitivities in EI and NCIHigher sensitivity in NCI mode Higher sensitivity in EI mode

Table V.2: Classification of sensitivities with EI and NCI modes for pesticides

using iso-butane, methane orammonia as reagent gas. Theselectivity is also helpful whenpeak coelutions are observedwith other matrix components[16]. In order to identify pesti-cides the EI mode of theNCI/PCI/EI combi source wasused in these experiments. TheNCI masses of the compoundswere then determined by com-parison of the EI and NCI scandata. In Figure V.4, the GC/MSresults attained for a standardusing a RTX-5 10 m, 0.1 mm, 0.1 μm column and using bothmodes are shown. The resultingspectra similarity is comparablewith the result obtained usingthe standard method. In thisexperiment, the scan range was50 to 400 amu with a scan fre-quency of 20 Hz. The amountsof the two standards were 5 ng(EI) and 50 pg (NCI). While theMS response is enhanced drasti-cally for diazinon, chlorpyrofos,endosulfan sulfate and potasan,the signal increase is less for cap-tan, while for piperonyl butoxidethere is almost no signal at all. To illustrate the different spectragenerated by the two modes, thecorresponding spectra of endo-sulfan sulfate are also shown.

With regard to EI, a typical frag-mentation is observed; in the caseof NCI only the masses 384, 386,388 together with a dominant 97fragment are present. Table V.1reports data concerning differentdominant fragments observed inEI compared with NCI for thecompounds measured here. Fig-ure V.5 shows a TIC chromato -gram of a real sample (black tea)indicating a contamination ofchlorpyrofos and endosulfan sulfate. Also shown is a chro-matogram relative to a standardmixture containing 0.5 pg of eachcomponent.

Table V.2 shows a list of pesticidesand a classification in terms of sen-sitivity in each mode [16]. As NCIis characterized by a very highselectivity, it can also prove use-ful when matrix componentsinterfere with the peaks of inter-est. For a more complete over -view of pesticides screened witha combination of EI and NCI,please consult reference [16].

In conclusion, a combination ofEI and NCI as well as mass spec-tra library search for both modesis a powerful tool for pesticideanalysis.

Page 22: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

27

Chromatography Volume 2 V. APPLICATIONS IN FOOD ANALYSIS

2. The analysis of organochlorine pesticides in food matrices using GC-ECD and NCI GC/MS

Organochlorine (OCP)pesticides are commonlymeasured using ECD or

MS (NCI mode) detectors.

In conventional GC analysisusing standard columns of about30 m length with a 0.25 mm innerdiameter and 0.25 μm film, thetypical run time for an OCP stan-dard solution containing 23 com-pounds is about 30 minutes (Fig-ure V.6; for quantitative resultsrefer to Table V.3). The retentiontime of p,p’-DDD is about 21.5minutes. The column used in thisapplication was a 5 % phenyl; thetemperature program was as fol-lows: from 100 °C (1 min) to 170°C, at 50 °C/min (1 min), then to220 °C at 5 °C/min, then to 260°C at 10 °C/min, then to 280 °C(10 min) at 20 °C/min using N2 ascarrier gas with an initial pressureof 77 kPa corresponding to anaverage linear velocity of 23 cm/s.The injection was carried out inthe splitless mode (1 μL).

The application was then carriedout using a fast GC method witha CPsil 8 9 m, 0.1 mm, 0.1 μmcolumn and H2 as carrier gas. The result is shown in Figure V.7.As can be seen, all 23 compoundsare better separated and the reten-tion time of p,p’-DDD was lessthan 3.6 minutes. The tempera-ture program applied was as fol-lows: from 80 °C (1 min) to 280°C (3 min) at 60 °C/min with aninitial head pressure of 324 kPaand an average linear velocity of100 cm/s (H2 ) which was con-stant during the entire analysis.The selected filter time constantand sampling frequency were 20 ms and 63 Hz respectively. Theinjection volume was 1 μL with asplit ratio of 40 :1. The ECDmake-up gas flow was 80 mL/min.The signal-to-noise ratio of �-HCH in the rapid application isabout 440 :1 (split) while the samemeasured 220:1 in the splitlessconventional application, indica -ting an increased sensitivity.

10.0 12.5 15.0 17.5 20.0 22.5

Tecn

azen

e

Minutes25.0 27.5

- 0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

6.5

7.0

7.5

8.0uV (x 1,000)

Mir

ex

p,p’

-DD

D

cis-

Chlo

rdan

etr

ans-

Chlo

rdan

eB

rom

opho

s et

hyl

tran

s-H

epta

chlo

roep

oxid

eci

s-H

epta

chlo

roep

oxid

eIs

odri

neB

rom

opho

s m

ethy

lIs

oben

zaneH

epta

chlo

r

Pent

achl

orbe

nzen

e

Ald

rine

Pent

achl

oran

iline

delta

-HCH

Lind

ane

beta

-HCH

HCB

Pent

achl

oran

isol

e

alph

a-H

CH

Ben

flura

line

Figure V.6: Conventional GC analysis of an OCP standard mixture (23 compounds) using a RTX-5 30 m, 0.25 mm ID, 0.25 μm column. For operational conditions refer to text.

2.25 2.50 2.75 3.00 3.25 3.50

Tecn

azen

e

Minutes3.75 4.00

- 0.250.000.250.500.75

1.251.501.75

2.252.50

3.003.25

3.754.00

4.504.755.00

5.50uV (x 10,000)

Mir

ex

p,p’

-DD

Dci

s-Ch

lord

ane

tran

s-Ch

lord

aneB

rom

opho

s et

hyl

tran

s-H

epta

chlo

roep

oxid

eci

s-H

epta

chlo

roep

oxid

eIs

odri

neB

rom

opho

s m

ethy

l

Isob

enza

ne

Hep

tach

lor

Pent

achl

orbe

nzen

e Ald

rine

Pent

achl

oran

iline

epsi

lon-

HCH

Lind

ane

beta

-HCH

HCB

Pent

achl

oran

isol

e

alph

a-H

CH Ben

flura

line

2.00

5.25

3.50

1.00

2.75

4.25

Figure V.7: Fast analysis of the OCP standard mixture containing 23 compounds with a CP SIL 8 9 m, 0.1 mm, 0.1 μm column.For retention times and concentrations see Table V.3.

The full width at half maximumof �-HCH, for example, is about 0.5 s which is a typical value forthese types of columns, demon-strating the suitability of theECD-2010 for fast analysis in the

field of organochlorine pesticides.The limit of detection for �-HCH(considering a signal-to-noise ratioof at least 3) was about 0.1 ppb.The method was then applied to a grape sample. The resulting

chromatogram is shown in FigureV.8. Contaminations of chlor -pyrofos at a concentration of 0.53 ng/mL (corresponding to 0.48 ng/kg grapes) and �

Page 23: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

TargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTarget

RT2.232.372.58

2.5482.6132.6312.5982.6622.8182.6832.842

2.963.1093.0653.1393.1633.1963.2063.2913.3253.4693.5083.961

01 Pentachlorobenzene02 Tecnazene03 Benfluraline04 �-HCH05 HCB06 Pentachloroanisole07 �-HCH08 Lindane09 �-HCH10 �-HCH11 Pentachloroaniline12 Heptachlor13 Aldrine14 Isobenzane15 Bromophosmethyl16 Isodrine17 cis-Heptachloroepoxide18 trans-Heptachloroepoxide19 Bromophosethyl20 trans-Chlordane21 cis-Chlordane22 p,p’-DDD23 Mirex

Chromatography Volume 2V. APPLICATIONS IN FOOD ANALYSIS

28

1.5 2.0 2.5 3.0 3.5 4.0

Minutes

4.5- 0.5

0.0

1.0

1.5

2.0

3.0

3.5

4.5

5.0

6.0

6.5

7.5

8.0

8.5

9.5

uV (x 10,000)

Cypermethrin

Chlorpyrofos

2.5

9.0

5.5

0.5

4.0

7.0

Figure V.8: Fast GC-ECD result for a grape sample: chlorpyrofos 0.53 ng/mL (corresponds to 0.48 mg/kg grapes) and cypermethrin 0.55 ng/mL (corresponds to 0.5 mg/kg)

1.25 1.75 2.25 2.75 3.25 3.75

Tecn

azen

e

Minutes

4.25 4.50

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

5.0

5.5

6.0

6.5

7.0

7.5

8.0

uV (x 100,000)M

irex

cis-

Chlo

rdan

e

tran

s-Ch

lord

ane

Bro

mop

hose

thyl

tran

s-H

epta

chlo

roep

oxid

eci

s-H

epta

chlo

roep

oxid

eIs

odri

neB

rom

opho

smet

hyl

Isob

enza

ne

Pent

achl

orob

enze

ne

Ald

rine

Pent

achl

oroa

nilin

ede

lta-H

CHLi

ndan

ebe

ta-H

CH

HCB

Pent

achl

oran

isol

e

alph

a-H

CHB

enflu

ralin

e

4.003.503.002.502.001.50

p,p’

-DD

D

Hep

tach

lor

epsi

lon-

HCH

Lind

ane

Figure V.9: Fast GC chromatogram relative to an OCP standard mixture. Injection volume: 1 μL, in the splitless mode, high pressure pulse 400 kPa. Column: RTX-5 10 m, 0.18 mm, 0.4 μm.

cypermethrin at a concentrationof 0.55 ng/mL (corresponding to0.5 ng/kg) were found.

In order to apply a splitless injec-tion technique, high pressureinjection in combination with aslightly thicker film needs to beused (Figure V.9). For this appli-cation, a 10 m, 0.18 mm, 0.4 μm (5 % phenyl) column wasused with a 100 °C initial temper-ature for 1 min then at 60 °C/minup to 280 °C (3 min). The carriergas was H2 and the linear velocitywas 120 cm/s over the entire run.All other parameters wereunchanged. Again considering �-HCH, the calculated detectionlimit is about 0.01 ppb in thiscase.

The determination of organochlo-rinated pesticides in food matricescan be performed very well usingfast GC-ECD with the ECD-2010. The measured detectionlimits were below 0.1 ppb for sev-eral compounds using a split of40 :1 and about 0.01 ppb using thesplitless technique.

In the case of GC/MS, negativechemical ionization may be con-sidered as ideal in order to pro-mote a thermal electron captureprocess inside the ion sourceusing a reagent gas (methane).Low energy secondary electrons

originate from the impact of theelectrons emitted from the MS fil-ament onto the reagent gasmolecules. The ions producedthen pass through the mass filterjust as in EI. Only the polaritiesof the instrument are reversed. Itis very interesting to compare theresults obtained by employingNCI GC/MS with those obtainedwith the ECD. Figure V.10 illus-trates data relative to six organo -

Table V.3: Concentration of the OCP standard. Chromatogram shown in Figure V.6.

chlorine compounds recorded inthe SIM mode. The concentra-tions employed were between 0.4 and 0.6 ppb. The signal-to-noise ratios observed for t-hepta -chloroepoxide and p,p’-DDE are28:1 (mass 71) and 210:1 (mass70) respectively. If a minimumsignal-to-noise ratio of 3 :1 is con-sidered, the limits of detection aresimilar to the ECD results. Thecalibration curve for t-hepta -

chloroepoxide is also illustrated,showing a good correlation.

In conclusion, EI and NCI incombination is a very powerfultool for screening of organophos-phorous or organochlorine pesti-cides.

Conc (ppb)21.322.552.822.124.120.620.428.823.2

126

30.421.7

522.6

22.042525

50.3655

22.421.84

Page 24: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

29

Chromatography Volume 2 V. APPLICATIONS IN FOOD ANALYSIS

4.9 5.15.0 5.2

25,424

316

318

237

ID#:1

Name: Oxychlordane

R.T.: 5.008

Conc.: 0.44200 ppb

5.8 6.05.9 6.16.1

45,640

404 406

242

ID#:5

Name: Endosulfan I

R.T.: 5.888

Conc.: 0.60456 ppb

4.9 5.0 5.1 5.2

30,396

71

280

282

ID#:2

Name: t-Heptachloroepoxide

R.T.: 5.062

Conc.: 0.48018 ppb

5.95.8 5.0 6.0 6.16.1

54,371

308

272

274

5.3 5.4 5.4

66,700

242

406

244

ID#:3

Name: Endosulfan II

R.T.: 5.350

Conc.: 0.51491 ppb

6.65.6 6.7 6.8 6.9 7.0

11,109

70

308

310 ID#:7

Name: p,p’-DDE

R.T.: 6.597

Conc.: 0.63673 ppb

5.5 5.6 5.7

7,135

346

237

344 ID#:4

Name: Dieldrin

R.T.: 5.585

Conc.: 0.45632 ppb

2.0 5.0 ppb0.0

[*10 ^4]

Calibration

ID#:2

Name: t-Heptachloroepoxide

Mass: 71.00

0.0

2.0

5.0

Figure V.10: NCI (methane) GCMS data recorded in SIM mode (figures on the graphs refer to the selected mass traces) with an organochlorine standard containingoxychlordane, t-heptachloroepoxide, endosulfan I, dieldrin, endosulfan II, endrin and p,p’-DDE. Concentration of each compound is 0.4 to 0.6 ppb.

ID#:6

Name: Endrin

R.T.: 6.011

Conc.: 0.50638 ppb

Page 25: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2V. APPLICATIONS IN FOOD ANALYSIS

30

1.61.31.12.80.23.5

11.80.91.1

32.01.80.60.3

10.124.2

2.70.6

C4 : 0C6 : 0C8 : 0

C10 : 0C10 : 1C12 : 0C14 : 0C14 : 1C15 : 0C16 : 0C16 : 1C17 : 0C17 : 1C18 : 0C18 : 1C18 : 2C18 : 3

1.41.41.12.70.23.5

11.61.01.3

32.01.80.70.3

10.724.2

3.00.5

Conventional(run time: 33.0 minutes)

Ultra-fast(run time: 2.0 minutes)

3. Analysis of butter fatty acid methyl esters (GC-FID)

Fatty acids (FAs) have essen-tial physiological functionsin the human body and

therefore have an important influ-ence on human health.

In particular, �3 unsaturated fattyacids are of great importance asthey can not be produced by thehuman metabolism. FAs such aseicosanic acid (EPA) and docosa-hexanioc acid (DHA) have a ben-eficial influence on immunologi-cal, heart and allergic diseases,diabetes and cancer [17]. As aconsequence, the determination oflipidic profiles in foods is veryimportant. The composition offats and oils in natural matricesmay be very complex and conven-tional GC separations can there-fore be time-consuming. It mustbe added that prior to GC injec-tion, the FAs must be convertedinto their respective methyl estersin order to increase volatility anddecrease polarity.

A conventional GC applicationcarried out with butter FAMEs isillustrated in Figure V.11 [18]. The separation may be consideredto be quite complex and with thecolumn type used (Carbowax 30 m, 0.25 mm ID, 0.25 μm film),the total analysis time exceeds 30 minutes.

In contrast, the result obtainedwith a fast GC method is shownin Figure V.12 (Supelcowax 10 m,0.1 mm ID, 0.1 μm film) [18]. Theresolution may be considered tobe approximately the same butthe total analysis time is reducedby a factor of 17. The parametersapplied in the conventional analy-sis were as follows: injection vol-ume: 1 μL (1 :10 in hexane); splitratio: 1:50 (250 °C); temperatureprogram: 50 °C to 250 °C at 3.0°C/min; head pressure: 53 kPa atconstant average linear velocity;carrier gas: H2; [u]: 36.2 cm/s;detector: FID (250 °C), H2: 50mL/min, air: 400 mL/min, make-up: 50 mL/ min kPa (N2); sam-

5.0 15.0 25.0Minutes

0 e3

50 e3

100 e3

150 e3

200 e3

250 e3

300 e3

350 e3

uV

30.020.010.0

1

2

3

4

5

6

7

8 9

10

111213 14

15

1618

19

21

22

23

252628

29

30

33

3536

Figure V.11: Conventional GC chromatogram of butter FAMEs using a wax column: 30 m, 0.25 mm, 0.25 μm. For experimental conditions refer to text.

0.2 1.0 1.4Minutes

0 e 3

10 e 3

30 e 3

40 e 3

60 e 3

70 e 3

90 e 3

110 e 3

uV

1.80.6

1

2

3

4

5

6

7

8 9

10

111213 14

15

161819 21

22

23

2526

28

29

30

33

35 36

20 e 3

50 e 3

80 e 3

100 e 3

Figure V.12: Fast GC chromatogram of butter FAMEs using a wax column: 10 m, 0.1 mm, 0.1 μm. For experimental conditions refer to text.

Table V.4: Comparison of relative peak areas

for conventional and fast GC. Area reproducibility over 5 successive

runs was below 1 % RSD.

pling rate: 40 ms; filter time con-stant: 200 ms.

The parameters applied in the fastanalysis were as follows: injectionvolume: 0.2 μL (1:20 in hexane);split ratio: 1:200 (250 °C); tem-perature program: 50 °C to 250°C at 90.0 °C/min; head pressure:400 kPa at constant average linearvelocity; carrier gas: H2; [u]: 116.0cm/s; detector: FID (250 °C) H2:50 mL/min, air: 400 mL/min,make-up: 50 mL/min kPa (N2);sampling rate: 4 ms (250 Hz); filter time constant: 20 ms.

The correlation between fast andconventional GC data is verygood, as can be derived fromTable V.4, which reports relativepeak area percentages for bothapplications. For more details seereference [18].

1) C4 : 0 19) C15 : 02) C5 : 0 20) C15 : 13) C6 : 0 21) C16 : 0 iso4) C7 : 0 22) C16 : 05) C8 : 0 23) C16 : 16) C9 : 0 24) C17 : 0 anteiso7) C10 25) C17 : 0 iso8) C10 : 1 26) C17 : 09) C11 : 0 27) C17 : 110) C12 : 0 28) C18 : 0 iso11) C12 : 1 29) C18 : 012) C12 : 1 isomer 30) C18 : 1 �913) C13 : 0 31) C18 : 1 �714) C14 : 0 iso 32) C18 : 2 �?15) C14 : 0 33) C18 : 2 �616) C14 : 1 34) C18 : 2 �?17) C15 : 0 anteiso 35) C18 : 3 �318) C15 : 0 iso 36) C18 : 2 con.

Page 26: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2 VI. APPLICATION IN THE PETROCHEMICAL FIELD

31

Kerosene analysis (GC-FID)

10 20 30

Chromatogram

Minutes

0 e 3100 e 3200 e 3300 e 3400 e 3500 e 3

700 e 3

1100 e 3

900 e 3

1300 e 31200 e 3

1400 e 3

600 e 3

1000 e 3

800 e 3

605040

uV

Figure VI.1: Conventional GC chromatogram recorded with a kerosene sample: RTX-1MS, 30 m, 0.25 mm ID, 0.25 μm

0.25 0.5 0.75

Minutes

0 e 350 e 3

200 e 3

300 e 3

400 e 3

500 e 3

700 e 3

350 e 3

150 e 3

550 e 3

450 e 3

650 e 3600 e 3

250 e 3

100 e 3

1.51.251.0

uV

750 e 3

2.252.01.75

Figure VI.2: Fast GC chromatogram relative to a kerosene sample: RTX-1MS, 10 m, 0.1 mm ID, 0.1 μm

GC analysis of kerosenemay be considered to bea very common quality

control procedure in the petro-chemical field.

An increase in sample through-put is of the highest interest inthis industrial sector. In FigureVI.1, the chromatogram recor dedwith a conventional 30 m columnis shown (temperature program:40 °C (2 min) at 2 °C/ min to 220 °C (3 min); carrier gas H2,linear velocity: 59 cm/s). Thetotal analysis time is about 72min. The predominant peaks, inan equally spaced order, arekerosene paraffines. The transi-tion to a fast method wasachieved using a RTX-1 MS, 10 m length, 0.1 mm ID, 0.1 μmstationary phase thickness col-umn. The result of the fast GCanalysis is shown in Figure VI.2.

In this case the analysis time isabout 2.3 min with a 31.3 foldspeed gain. The applied averagelinear velocity was 85.3 cm/s (H2as carrier gas) with a greatlyaccelerated temperature program.The split ratio was set at 800 :1with the gas saver function acti-vated (10 :1 after 1 min).

The temperature program was set from 40 °C up to 150 °C at 80 °C/ min, and up to 250 °C at70 °C/ min and finally up to 350 °C at 50 °C/min; the FID filter time constant and samplingfrequency were set to 10 ms and125 Hz respectively.

The chromatographic resolutionis very comparable with that ofthe conventional analysis. Thepeak widths of the narrowestpeaks are less than 0.5 s (FWHM).

Page 27: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential
Page 28: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

01 Benzyl alcohol02 Limonene03 Linalool04 Methyl heptin carbonate05 Citronellol06 Citral (Neral)07 Cinnamic aldehyde08 Geraniol09 Citral (Geranial)10 Anisyl alcohol11 Hydroxy citronellal12 Cinnamyl alcohol13 Eugenol

14 Coumarin15 Iso-eugenol16 Methyl gamma ionone17 Lilial18 Amyl cinnamic aldehyde19 Lyral20 Amyl cinnamic alcohol21 Farnesol 122 Farnesol 223 Hexyl cinnamic aldehyde24 Benzyl benzoate25 Benzyl salycilate26 Benzyl cinnamate

1. Potential allergens in perfumes

In the 7th Amendment to theEuropean Cosmetics Directive,as published in the Official

Journal N° L 66 of the EuropeanUnion on March 11, 2003, 26 com-pounds are defined as potentialallergens. This regulation requiresthat the presence of these 26 fra-grance ingredients must be indi-cated on the label of finished cos-metic products if their concentra-tions exceed a threshold of 0.01 %for rinse-off and 0.001 % for leave-on products. These compounds arereferred to as potential allergens, asthey may or may not induce anallergic reaction. Among the 26materials indicated, two are naturalextracts (oak moss and tree moss)so the method as specified below isrestricted to the determination of24 volatile chemicals [19]. Thesecompounds are listed in TableVII.1.

It must be emphasized that thedetermination of these compo-nents, which is usually carried outusing quadrupole GC/MS equip-ment, is very complex and time-consuming. An example of such anapplication carried out with purestandard compounds is illustratedin Figure VII.1. The concentrationof each compound was about 400 ppm, with a total of 26 peaks(including isomers) separated on aCP SIL 5 50 m column with 0.25mm ID and 0.25 μm film thickness[temperature program: 50 °C

(1 min) to 210 °C at 2 °C/min, to280 °C at 10 °C/min (10 min). Carrier gas: He; average linearvelocity 34.4 cm/s; split ratio300 :1]. As the analysis time ismore than 75 minutes, the aim isto decrease this extensive GC runtime in order to increase the sam-ple throughput.

Figure VII.2 shows a fast GC/MSapplication carried out on a mix-ture of standard allergens with aRTX-5 10 m, 0.1 mm ID, 0.1 μmcolumn [temperature program: 70°C (1 min) to 180 °C at 25 °C/minto 280 °C (1 min) at 80 °C/min.Carrier gas: He; constant linearvelocity: 40 cm/s; split 300:1]. Inorder to evaluate the resolutionattained in both applications, twochromatographic expansions arecompared in Figure VII.3. As canbe seen, the degree of peak separa-tion is better in the fast application(citral and anisyl alcohol areresolved) while a speed gain of afactor of about 11 is observed.Peak FWHM values were on aver-age approx. 0.5 s; a mass range of30 to 350 amu and a 20 Hz acqui-sition rate were applied. The quali-ty of the resulting spectra, provid-ed by the GCMS-QP2010, wasvery high with similarities rangingbetween 94 and 98 (Wiley library).Linearity was measured between 4 and 400 ppm and a regressioncoefficient of 0.9995 was attainedwhich indicates excellent analytical

Table VII.1: Compounds defined by the IFRA as potential allergens

Figure VII.1: Conventional GC/MS chromatogram relative to an allergen standard mixture carried out with a CP SIL-5 50 m, 0.25 mm, 0.25 μm column

(x 1,000,000)

0.25

0.50

0.75

1.00

1.25

1.50

1.75

2.20

2.25

2.50

2.75

3.00

25.0 50.0 75.0

Figure VII.2: Fast GC/MS chromatogram relative to an allergen standard mixture carried out with a RTX-5 10 m, 0.1 mm, 0.1 μm column

(x 10,000,000)

1.0

2.0

3.0

4.0

5.0

6.0

7.0

8.0

9.0

2.0 3.0 4.0 5.0 6.0

33

Chromatography Volume 2 VII. APPLICATIONS IN THE FIELDS OF FLAVORS & FRAGRANCES

Page 29: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Chromatography Volume 2VII. APPLICATIONS IN THE FIELDS OF FLAVORS & FRAGRANCES

34

27.5 30.0

Minutes

0.250.50

1.25

1.75

2.25

2.75

3.75

2.00

1.00

3.00

2.50

3.503.25

1.50

0.75

35.032.5

TIC4.00

40.037.5

(x1,000,000)

4 / M

ethy

l Hep

tin C

arbo

nate

12 /

Cinn

amyl

Alc

ohol

11 /

Hyd

roxy

Citr

onel

lal

10 /

Ani

syl A

lcoh

ol9

/ Citr

al (G

eran

ial)

8 / G

eran

iol

7 / C

inna

mic

Ald

ehyd

e6

/ Citr

al (N

eral

)5

/ Citr

onel

lol

13 /

Euge

nol

3.00

Minutes

0.51.0

2.5

3.5

4.5

5.5

7.5

4.0

2.0

6.0

5.0

7.06.5

3.0

1.5

3.503.25

TIC8.0

4.003.75

(x10,000)

4 / M

ethy

l Hep

tin C

arbo

nate

12 /

Cinn

amyl

Alc

ohol

11 /

Hyd

roxy

Citr

onel

lal

10 /

Ani

syl A

lcoh

ol

9 / C

itral

(Ger

ania

l)

8 / G

eran

iol

7 / C

inna

mic

Ald

ehyd

e

6 / C

itral

(Ner

al)

5 / C

itron

ello

l

13 /

Euge

nol

CP SIL-5 50 m, 0.25 mm ID, 0.25 μm RTX-5 10 m, 0.1 mm ID, 0.1 μmSegment S1

Figure VII.3: Chromatographic expansions relative to Figure VII.1 and Figure VII.2

0 100 200

Limonene

R = 0.9995

Area (x 1,000,000)

con. (ppm)0.0

0.5

1.0

1.5

2.0

2.5

3.5

3.0

4.0

300 0 100 2000.0

1.0

2.0

5.0

3.0

4.0

Area (x 100,000)

Lyral

R = 0.9995

con. (ppm)

Figure VII.4: Calibration curves determined with fast GC/MS for limonene and lyral

2.0 3.0 4.0 5.0 6.0 7.6 7.6

Eau de Toilette

Minutes

TIC

16

17

19

25

Figure VII.5: Fast GC/MS analysis of allergens in a sample of Eau de Toilette. The peak numbers refer to Table VII.1.

precision. To illustrate, calibrationcurves for limonene and lyral areshown (Figure VII.4).

The method was then applied toreal-world samples. The TICresult of an Eau de Toilette sam-ple, diluted in acetone (1:100), isshown in Figure VII.5. The aller-gens found (concentrations referto the dilution) are methyl gammaionone (74.5 ppm), lilial (129.7ppm), lyral (99.4 ppm) and benzylsalicylate (96.6 ppm). The spectralsimilarities were between 93 and98 using a commercial MS library(Wiley).

Page 30: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

35

Chromatography Volume 2 VII. APPLICATIONS IN THE FIELDS OF FLAVORS & FRAGRANCES

2. Quality control of flavors

In the flavor industry, routinequality control analyses areapplied to a high number of

samples. In this field, the analysisof essential oils is widespread andmay be considered as a complextask. Chromatograms relative togeranium and limette essential oilanalysis carried out on a DBWAX 10 m, 0.1 mm, 0.2 μm col-umn are reported in Figure VII.7and Figure VII.8.

The essential oils were diluted inethanol at a concentration of 5 %.The GC program was set from 40 °C (0.5 min) to 230 °C (at 50°C/min) with a 60 cm/s constantaverage linear velocity (hydro-gen). The injection volume was 1 μL with a split ratio of 400 :1.About 30 important flavor com-ponents were separated and iden-tified in less than 5 minutes.

0.5 1.0 1.5 2.0 2.5 3.0

cis-

beta

-Oci

men

e

Minutes3.5 4.0- 0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

3.5

4.0

uV (x 10,000)

Ger

anio

l

Citr

onel

lol

Ger

anyl

Ace

tate

Ger

anyl

For

mat

eCu

aiad

iene

+ C

itron

elly

le F

orm

ate

Ger

mac

rene

D

Ger

anyl

Isob

utyr

ate

Ner

ol

beta

-Car

yoph

ylen

eLi

nalo

ol

Lina

lool

oxid

e

Lim

onen

e

Isom

enth

one

Myr

cene

+ a

lpha

-Pha

ella

ndre

ne

Men

thon

eRose

Oxi

de

Para

cym

ene

Terp

inol

ene

alph

a-Pi

nene

tran

s-be

ta-O

cim

ene

gam

ma-

Terp

inen

e

4.5

Chromatogram

Rose

Oxi

de

Lina

lool

oxid

e

Ger

anyl

But

yrat

e

Citr

onel

lyle

Tig

late

Ger

anyl

Tig

late

Phen

ylet

hyle

Tig

late

10 g

amm

a-ep

i Eud

esm

ol

Figure VII.7: Fast GC/MS chromatogram (DB wax 10 m, 0.1 mm, 0.2 μm) of a geranium essential oil (5 % in ethanol)

0.5 1.0 1.5 2.0 3.0

cis-

beta

-Oci

men

e

Minutes3.5 4.0- 0.5

0.0

0.5

1.0

1.5

2.0

2.5

3.0

uV (x 10,000)

Terp

in-1

-en-

4-ol

alph

a-Te

rpin

eol

Non

anal

Dec

anal

beta

-Car

yoph

ylle

ne

Lina

lool

alph

a-Te

rpin

ene

Lim

onen

e

Oct

anal

Myr

cene

+ a

lpha

-Pha

ella

ndre

ne

beta

-Phe

lland

rene

beta

-Pin

ene

Para

cym

ene

Terp

inol

ene

alph

a-Pi

nene tr

ans-

beta

-Oci

men

e

gam

ma-

Terp

inen

e

4.5

Chromatogram

Sabi

nene

Cine

ol 1

,8

gam

ma-

Terp

ineo

l

Cam

phen

e

Etha

nol

Figure VII.8: Fast GC/MS chromatogram (DB wax 10 m, 0.1 mm, 0.2 μm) of a limette essential oil (5 % in ethanol)

2.0 3.0 4.0 5.0 6.0 9.07.0

1.0

1.5

2.0

2.5

3.0

3.5

4.0

4.5

(x 10,000,000)

6-M

ethy

l-G

amm

a-Io

none

Lilia

l

Ger

anio

lCitr

onel

lol

Lina

lool

Lim

onen

e

8.0

TIC

alph

a-H

exyl

cinn

amic

al

dehy

de

0.5

0.0

Figure VII.6: Fast GC/MS TIC data relative to a perfume and obtained with a SPB-5 10 m, 0.1 mm, 0.1 μm column

The method was also applied to a higher concentrated perfume.Figure VII.6 shows the chromato-graphic result obtained with a 10% solution (acetone) of a perfumeusing a SPB 5 10 m, 0.1 mm, 0.1μm column (injection volume: 1 μL; split ratio: 500:1; average linear velocity: 50 cm/s; tempera-ture program: 50 °C (0.5 min) to200 °C at 20 °C/min, to 280 °C at50 °C/min; MS: scan range 41 to320 amu, 20 scans/s at 10,000amu/s scan rate). The identifica-tion of the potential allergen com-pounds was carried out by thefinder function of the GCMSsolu-tion software. In this case thewhole TIC is searched for alibrary match with the compoundsof interest. This is also shown inthe same figure. Seven allergens

were identified with a high simi-larity using the aforementionedcommercial library.

SI93969999989691

NameLimoneneLinanoolCitronellolGeraniol6-Methyl-Gamma-IononeLilialalpha-Hexylcinnamic aldehyde

Ret. Time2,6693,2194,184,3425,8256,2017,482

Page 31: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential
Page 32: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

HCB18312853

Heptachlor52494435

Aldrine7470

p,p’-DDE101

99112

97110

77151149118153137138158187183156157180170198189194206209

Chromatography Volume 2VIII. APPLICATION IN THE ENVIRONMENTAL FIELD

38

3,00

3 4 5

Chromatogram

min

Figure VIII.1: GC-ECD chromatogram of 38 PCBs using a SPB5 10 m, 0.1 mm, 0.1 μm column. Concentrations are about 0.1 ppb.

Compound RT

TargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTargetTarget

Table VIII.1:List of PCBs contained

in the standard

PCB analysis with GC-ECD detection

The analysis of PCBs in dif-ferent matrices such as water or sludge is still

an important issue in the envi-ronmental field. These applica-tions are achieved by using eitherGC-ECD or NCI GC/MS. The results obtained by usingGC-ECD are shown in FigureVIII.1.

The chromatogram was attainedby injecting a mixture of 38 stan-dard PCBs (Table VIII.1). Thecolumn used was a SPB1, 10 m, 0.1 mm, 0.1 μm, with an

oven program of 90 °C (2 min) at40 °C/min to 290 °C and a con-stant average linear velocity of 50 cm/s (He). The injection vol-ume was 1 μL with a split ratioof 10 :1 (0.1 ppb for each com-pound). The ECD current wasset to 1 nA. The filter time con-stant and sampling frequencywere 20 ms and 100 Hz respec-tively. In order to avoid detectorband broadening, a make-up gasflow of 100 mL/min was applied.

4,5374,7464,9534,9575,0115,0655,1385,1615,2195,2385,2555,425,4375,5555,5775,6015,6485,6765,7035,7345,7455,8165,8695,9876,0726,1056,1226,1976,2236,3396,366,4266,4886,5486,6816,8397,0047,165

Page 33: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

39

The reduction of analyticalcycle times in routineapplications has become

increasingly important in orderto increase productivity or toacquire quick and correct resultsin quality control processes.

The narrow-bore columnapproach is a very effective wayof decreasing GC run times

while maintaining a high degreeof peak resolution. The applica-tion of these methods is increa -sing in routine laboratory work,also due to improved column andGC instrumental technology. Infact, modern-day GC systemsfulfill the analytical requirementsof narrow-bore capillaries.

Page 34: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Literature

[1] Matter, L. (Ed.), Food and Environmental Analysis by Capillary Gas Chromatography, Hüthig, 1997

[2] van Es, A., High Speed Narrow-Bore Capillary Gas Chromatography,Hüthig, Heidelberg, 1992

[3] van Ysacker, P. G.; Janssen, H.-G. Snijders, H. M. J.; Cramers C. A.; J. High Resol. Chromatogr. 18 (1995) 397

[4] Broske, A. D. et al, Abstract A05 and A21, 20th International Symposium on Capillary Chromatography, Riva del Garda, Italy, May 26-29, 1998

[5] David, F. et al., Abstract P53, 20th International Symposium on Capillary Chromatography, Riva del Garda, Italy, May 26 -29, 1998

[6] Cook, W. S., Today’s Chemist at Work 1 (1996) 16[7] Sandra, P. et al., Experiments with fast Multicapillary GC, presented

at the 18th International Symposium on Capillary Chromatography,Riva del Garda, Italy, May 20-24, 1996

[8] van Lieshout, M. et al., J. Micro. Sep. 11 (2) (1999) 155[9] Schomburg, G., Gaschromatographie, Grundlagen, Praxis,

Kapillartechnik, VCH[10] Mondello, L. et al., J. Chrom. A, 1035 (2004) 237 [11] Grob, K., Classical Split/Splitless Injection, Hüthig [12] Hinshaw, J. V., LCGC (2002) vol 15, p. 152[13] Leoni, V. and D’Alessandro De Luca, E., Essenz. Deriv. Agrum.,

1978, 48, 39-50.[14] Kempe, G. and Baier, H.-U., Organophosphorous pesticides

determined in natural matrices with GC-FPD, GC-FTD and GCMS:medium and narrow-bore columns, presented at 25th InternationalSymposium on Capillary Chromatography, Riva del Garda, Italy, May 14-17, 2002.

[15] Mondello, L., 22th International Symposium on Capillary Chromatography, Riva del Garda, Italy May 2002

[16] Kondo, S. et al., Euroanalysis conference, 2004 [17] Han, J. J. and Yamane, T., Lipids, 1999, 34, 989-995 [18] Mondello, L., J. Microc. Sep. 12 (1) 41-47, 2000[19] www.ifraorg.org

Page 35: Application Handbook Fast GC/GCMS · 2018. 11. 1. · Application Handbook Fast GC/GCMS Chromatography Volume 2. ... Applications in the fields of flavors & fragrances 32 1. Potential

Founded in 1875, Shimadzu Corporation, a leader in the development of ad vanced technologies, has a distinguished history of innovation built on the foundation of contributing to society through science and technology. We maintain a global network of sales, service, technical support and applications centers on six continents, and have established long-term relationships with a host of highly trained distributors located in over 100 countries.

For information about Shimadzu, and to contact your local office, please visit our website at www.shimadzu.eu

Shimadzu Europa GmbHAlbert-Hahn-Str. 6-10 · D-47269 DuisburgTel.: +49 - (0)203 - 76 87-0Fax: +49 - (0)203 - 76 66 [email protected]