Applications of Piezo Film Lab Amp
2008-10-13R H Brown
![Page 2: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/2.jpg)
Just imagine…
...the possibilities of piezo film
![Page 3: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/3.jpg)
The need for a lab ampThe need for a lab ampThe need for a lab amp
Piezo Film: No DC Response…but how low can we go?
Many technologies from MEAS have response down to and including DC: piezoresistive MEMS, Microfused, capacitive humidity, magnetoresistive…
But the piezoelectric effect generates charge within a capacitive source. The charge leaks away, either through the dielectric itself, or into the external circuit. This meanswe apply a high-pass filter to every signal…
![Page 4: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/4.jpg)
Example sensor: SDT1-028K
The SDT1-028K has a capacitance of 2.6 nF. With no load, the sensor "tries" to respond down to < 0.1 mHz. When connected to different input resistances, the cut-off frequency of the high-pass filter varies as shown here…
C: 2n6 Load R -3 dB freq Hz1T 0.00011G 0.061
100M 0.61210M 6.121M 61.2
The need for a lab ampThe need for a lab ampThe need for a lab amp
![Page 5: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/5.jpg)
The need for a lab ampThe need for a lab ampThe need for a lab amp
Influence of Input Impedance selection on low frequency response (voltage mode) The table below shows the theoretical -3 dB frequency (in Hz) of the high-pass filter that is formed when a sensor of capacitance listed in first column is connected to the specified input impedance. Note: the shaded cells have a -3 dB frequency that is below the range of the Low Frequency Selector,and in this case, the Low frequency Selector will determine the performance. In the case of the unshaded cells, the low frequency limit will be determined either by the Low Frequency selector, or by the data above, whichever is the higher.
1 M 10 M 100 M 1 G 100 p 1592 159 16 1.6 220 p 723 72 7.2 0.72 470 p 339 34 3.4 0.34
1 n 159 16 1.6 0.16 2.2 n 72 7.2 0.72 0.072 4.7 n 34 3.4 0.34 0.034 10 n 16 1.6 0.16 0.016 22 n 7.2 0.72 0.072 0.0072 47 n 3.4 0.34 0.034 0.0034
100 n 1.6 0.16 0.016 0.0016
![Page 6: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/6.jpg)
Instrumentation: digital oscilloscope
Most conventional instruments offer a 1 M input resistance as standard. If we connect the SDT1-028K piezo film sensor directly to this input, we create a high-pass filter at 61 Hz.
C: 2n6 Load R -3 dB freq Hz1T 0.00011G 0.061
100M 0.61210M 6.121M 61.2
The need for a lab ampThe need for a lab ampThe need for a lab amp
![Page 7: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/7.jpg)
-4
-3
-2
-1
0
1
2
3
4
0 2 4 6 8 10 12 14 16 18 20
-4
-3
-2
-1
0
1
2
3
4
0 2 4 6 8 10 12 14 16 18 20
Effect of the high-pass filter
The presence of the high-pass filter can affect the observed waveform in different ways:
1) Change in waveshape2) Change in amplitude3) Change in phase
The need for a lab ampThe need for a lab ampThe need for a lab amp
![Page 8: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/8.jpg)
Simulated waveform examples:
1) rectangular pulse (1 V pulse high for 1 s, 10 ms r/f time)
2) Sawtooth (ramp from 0 to 1 V in 1 s, 10 ms fall)
3) Continuous sine (1 Hz, 2 V pk-pk)
Example of real-life waveforms:
4) Chest strap (SDT1-028K in tension, elasticated strap)
The need for a lab ampThe need for a lab ampThe need for a lab amp
![Page 9: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/9.jpg)
Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse
Load R pos pk V neg pk V
1T 1 -0.0004
1G 0.998 -0.321
100M 0.981 -0.96
10M 0.828 -0.828
1M 0.257 -0.257
![Page 10: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/10.jpg)
Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse
Load R pos pk V neg pk V
1T 1 -0.0004
1G 0.998 -0.321
100M 0.981 -0.96
10M 0.828 -0.828
1M 0.257 -0.257
![Page 11: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/11.jpg)
Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse
Load R pos pk V neg pk V
1T 1 -0.0004
1G 0.998 -0.321
100M 0.981 -0.96
10M 0.828 -0.828
1M 0.257 -0.257
![Page 12: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/12.jpg)
Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse
Load R pos pk V neg pk V
1T 1 -0.0004
1G 0.998 -0.321
100M 0.981 -0.96
10M 0.828 -0.828
1M 0.257 -0.257
![Page 13: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/13.jpg)
Case 1: rectangular pulseCase 1: rectangular pulseCase 1: rectangular pulse
Load R pos pk V neg pk V
1T 1 -0.0004
1G 0.998 -0.321
100M 0.981 -0.96
10M 0.828 -0.828
1M 0.257 -0.257
![Page 14: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/14.jpg)
Case 2: sawtoothCase 2: Case 2: sawtoothsawtooth
Load R pos pk V neg pk V
1T 0.999 -0.001
1G 0.83 -0.17
100M 0.255 -0.73
10M 0.026 -0.81
1M 0.0026 -0.25
![Page 15: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/15.jpg)
Case 2: sawtoothCase 2: Case 2: sawtoothsawtooth
Load R pos pk V neg pk V
1T 0.999 -0.001
1G 0.83 -0.17
100M 0.255 -0.73
10M 0.026 -0.81
1M 0.0026 -0.25
![Page 16: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/16.jpg)
Case 2: sawtoothCase 2: Case 2: sawtoothsawtooth
Load R pos pk V neg pk V
1T 0.999 -0.001
1G 0.83 -0.17
100M 0.255 -0.73
10M 0.026 -0.81
1M 0.0026 -0.25
![Page 17: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/17.jpg)
Case 2: sawtoothCase 2: Case 2: sawtoothsawtooth
Load R pos pk V neg pk V
1T 0.999 -0.001
1G 0.83 -0.17
100M 0.255 -0.73
10M 0.026 -0.81
1M 0.0026 -0.25
![Page 18: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/18.jpg)
Case 2: sawtoothCase 2: Case 2: sawtoothsawtooth
Load R pos pk V neg pk V
1T 0.999 -0.001
1G 0.83 -0.17
100M 0.255 -0.73
10M 0.026 -0.81
1M 0.0026 -0.25
![Page 19: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/19.jpg)
Case 3: sine waveCase 3: sine waveCase 3: sine wave
Load R pk-pk Vphase deg
1T 2 0.004
1G 1.99 4
100M 1.71 32
10M 0.33 81
1M 0.033 89
![Page 20: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/20.jpg)
Case 3: sine waveCase 3: sine waveCase 3: sine wave
Load R pk-pk Vphase deg
1T 2 0.004
1G 1.99 4
100M 1.71 32
10M 0.33 81
1M 0.033 89
![Page 21: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/21.jpg)
Case 3: sine waveCase 3: sine waveCase 3: sine wave
Load R pk-pk Vphase deg
1T 2 0.004
1G 1.99 4
100M 1.71 32
10M 0.33 81
1M 0.033 89
![Page 22: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/22.jpg)
Case 3: sine waveCase 3: sine waveCase 3: sine wave
Load R pk-pk Vphase deg
1T 2 0.004
1G 1.99 4
100M 1.71 32
10M 0.33 81
1M 0.033 89
![Page 23: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/23.jpg)
Case 3: sine waveCase 3: sine waveCase 3: sine wave
Load R pk-pk Vphase deg
1T 2 0.004
1G 1.99 4
100M 1.71 32
10M 0.33 81
1M 0.033 89
![Page 24: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/24.jpg)
Case 4: chest strapCase 4: chest strapCase 4: chest strap
Real-life waveforms: chest strap
SDT1-028K sensor inserted in series with an elasticised chest strap, worn while breathing normally
![Page 25: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/25.jpg)
Case 4: chest strap (1 G)Case 4: chest strap (1 G)Case 4: chest strap (1 G)
-4
-3
-2
-1
0
1
2
3
4
0 2 4 6 8 10 12 14 16 18 20
![Page 26: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/26.jpg)
Case 4: chest strap (100 M)Case 4: chest strap (100 M)Case 4: chest strap (100 M)
-4
-3
-2
-1
0
1
2
3
4
0 2 4 6 8 10 12 14 16 18 20
![Page 27: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/27.jpg)
Case 4: chest strap (10 M)Case 4: chest strap (10 M)Case 4: chest strap (10 M)
-4
-3
-2
-1
0
1
2
3
4
0 2 4 6 8 10 12 14 16 18 20
![Page 28: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/28.jpg)
Case 4: chest strap (1 M)Case 4: chest strap (1 M)Case 4: chest strap (1 M)
-4
-3
-2
-1
0
1
2
3
4
0 2 4 6 8 10 12 14 16 18 20
![Page 29: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/29.jpg)
Case 4: chest strap (1 M – range changed)Case 4: chest strap (1 M Case 4: chest strap (1 M –– range range changed)changed)
-0.05
-0.04
-0.03
-0.02
-0.01
0
0.01
0.02
0.03
0.04
0.05
0 2 4 6 8 10 12 14 16 18 20
![Page 30: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/30.jpg)
Case 4: chest strap (1 M, 1 Hz HPF, +40dB )Case 4: chest strap (1 M, 1 Hz HPF, Case 4: chest strap (1 M, 1 Hz HPF, +40dB )+40dB )
-2
-1.5
-1
-0.5
0
0.5
1
1.5
2
0 2 4 6 8 10 12 14 16 18 20
![Page 31: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/31.jpg)
Conclusion – low frequenciesConclusion Conclusion –– low frequencieslow frequencies
The Piezo Film Lab Amp avoids the loss of low-frequency information content from piezo film sensor signals, by providing controlled load resistance of up to 1 G in voltage mode, and LF response down to 0.1 Hz
![Page 32: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/32.jpg)
The choice: Voltage or Charge mode
To detect the signal generated by a piezo sensor, we have basically two options:
1) Leave all of the charge on the sensor, leaving the voltage exactly as if nothing were connected, or
2) Remove all the charge from the sensor and transfer it elsewhere, leaving no voltage across the sensor
We describe option (1) as "voltage mode", and option (2) as "charge mode"
Piezo Film Lab AmpPiezo Film Lab AmpPiezo Film Lab Amp
![Page 33: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/33.jpg)
Charge Mode:
In a charge mode preamplifier, the input appears to be a short circuit. The potential between the two input terminals is maintained at exactly zero volts. All of the charge generated by the piezo sensor flows into the circuit, and is transferred to a feedback capacitor. The rate of charge decay is then governed by this selected capacitor, and not by the capacitance of the piezo source. Since the input is "shorted", stray capacitance has no effect.
The need for a lab ampThe need for a lab ampThe need for a lab amp
![Page 34: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/34.jpg)
Simple charge mode circuitSimple charge mode circuitSimple charge mode circuit
![Page 35: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/35.jpg)
Simple charge mode circuitSimple charge mode circuitSimple charge mode circuit
Notes: •Circuit "gain" defined by C1/C2 (with Lab Amp, C2 is switch-selectable in decades from 100 pF to 100 nF)•Time constant defined by product R1C2 (with Lab Amp, the low freq limit is switch-selectable in decades from 0.1 Hz to 1 kHz)•Additional capacitance across input has no effect (except on very high frequency response)
![Page 36: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/36.jpg)
Voltage Mode:
In a voltage mode preamplifier, the input appears to be an infinite impedance. The potential between the two input terminals is maintained at the "open circuit" potential of the source network (including all stray capacitance). None of the charge generated by the piezo sensor flows into the circuit. A bleed resistor is required on the input (otherwise circuit would be unstable or saturate), and the low frequency response is governed by the total capacitance connected to the input, and this bleed resistance.
The need for a lab ampThe need for a lab ampThe need for a lab amp
![Page 37: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/37.jpg)
Simple voltage mode circuitSimple voltage mode circuitSimple voltage mode circuit
![Page 38: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/38.jpg)
Simple voltage mode circuitSimple voltage mode circuitSimple voltage mode circuit
Notes: •Circuit "gain" is unity; separate gain stage can follow (with Lab Amp, final gain stage offers 0 to +40 dB in 10 dB steps)•Time constant defined by product R1C1 (with Lab Amp, R1 is switch-selectable in decades from 1 M to 1 G)•Additional capacitance across source reduces the "open-circuit" potential, and thus the detected signal
![Page 39: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/39.jpg)
Selection of modeSelection of modeSelection of mode
Use charge mode when:
•Piezo sensor has long cable attached (or varying cable length)•Only a portion of the active sensor area is impacted or excited•"Fundamental" measurement of charge output preferred•Minimize triboelectric cable noise
Use voltage mode when:
•Lab Amp will later be replaced by simple buffer stage•"End of cable" output voltage is preferred
![Page 40: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/40.jpg)
Using the low-pass filterUsing the lowUsing the low--pass filterpass filter
In addition to the control over the low-frequency limit afforded by the input impedance selector and the LF filter control, the Piezo Film Lab Amp also has a low-pass filter that can block unwanted higher frequency components.
Setting this control to the "10 Hz" setting allows effective filtering of 50/60 Hz line frequency interference, which is commonly experienced when experimenting with piezo film sensors.
![Page 41: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/41.jpg)
Using the low-pass filterUsing the lowUsing the low--pass filterpass filter
With High Freq control set to 1 kHz:
![Page 42: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/42.jpg)
Using the low-pass filterUsing the lowUsing the low--pass filterpass filter
With High Freq control set to 10 Hz:
![Page 43: Applications of Piezo Film Lab Amp - Durham Instruments Lab Amplifier Use and... · Applications of Piezo Film Lab Amp 2008-10-13 R H Brown. Just imagine](https://reader033.vdocuments.net/reader033/viewer/2022052608/5ae267357f8b9a5d648ca816/html5/thumbnails/43.jpg)
Piezo Film Lab AmpPiezo Film Lab AmpPiezo Film Lab Amp
Key Features:
•Charge or voltage mode operation•0.01 to 1000 mV/pC sensitivity range in charge mode•1 M to 1 G input resistance, -40 to 40 dB dB gain, in voltage mode•Multi-pole low-pass and high-pass filtering, with -3dB frequencies ranging from 0.1Hz to 100kHz•BNC input and output•Internal battery (2 x 9 V) or external 24 Vdc power supply operation