atomic force microscopy’s path to atomic …atomic force microscopy, invented1 and also...
TRANSCRIPT
arX
iv:c
ond-
mat
/050
3671
v1 [
cond
-mat
.mtr
l-sci
] 29
Mar
200
5
Atomic force microscopy’s path to atomicresolution
Franz J. Giessibl
We review the progress in the spatial resolution of atomic force microscopy(AFM) in vacuum. After an introduction of the basic principle and a conceptualcomparison to scanning tunneling microscopy, the main challenges of AFMand the solutions that have evolved in the first twenty years of its existence areoutlined. Some crucial steps along the AFM’s path towards higher resolutionare discussed, followed by an outlook on current and future applications.
Review Feature, submitted to Materials Today Feb 10 2005, 2nd revised ver-
sion
Experimentalphysik VI, EKM,Institute of Physics, Augsburg University,86135 Augsburg, Germany,Email: [email protected]
Atomic force microscopy, invented1 and also introduced2 in 1985/86, can be viewed
as a mechanical profiling technique that generates three-dimensional maps of surfaces by
scanning a sharp probe attached to a cantilever over a surface. The forces that act between
the tip of the cantilever and the sample are used to control the vertical distance. AFM’s
potential to reach atomic resolution was foreseen in the original scientific publication,2 but
for a long time, the spatial resolution of AFM was inferior to the resolution capability of
its parent, scanning tunneling microscopy (STM). The resolution limits of STM and AFM
are given by the structural properties of the atomic wavefunctions of the probe tip and
the sample. STM is sensitive to the most loosely bonded electrons with an energy at the
Fermi level while AFM responds to all electrons, including core electrons. Because the
1
electrons at the Fermi level are spatially less confined thancore electrons that contribute
to AFM images, in theory AFM should be able to achieve even greater spatial resolution
than STM. Today, experimental evidence emerges where in simultaneous AFM/STM stud-
ies, AFM images reveal even finer structural details than simultaneously recorded STM
images. The experimental advances that made high-resolution AFM possible started with
the introduction of frequency modulation AFM (FM-AFM), whe re the cantilever oscillates
at a fixed amplitude and the frequency is used as a feedback signal. Early implementa-
tions of FM-AFM utilized silicon cantilevers with a typical spring constant of 10 N/m that
oscillate with an amplitude on the order of 10 nanometers. The spatial resolution could be
increased by the introduction of quartz cantilevers with a stiffness on the order of 1 kN/m,
allowing the use of sub-nm amplitudes. The direct evaluation of higher harmonics in
the cantilever motion has enabled a further increase in spatial resolution. Because AFM
can image insulators as well as conductors, it is now a powerful complement to STM for
atomically resolved surface studies. Immediate applications of high resolution AFM have
been demonstrated in vacuum studies relating to materials science, surface physics and
-chemistry. Some of the techniques developed for ultrahigh-vacuum AFM may be appli-
cable to increase AFM resolution in ambient or liquid environments, such as required for
studying biological or technological specimens.
1 Introduction
In terms of the operating principle, atomic force microscopy (AFM)1, 2 can be viewed as an
extension of the toddlers way of ‘grasping’ the world by touching and feeling as indicated in
Figure 1 of Binnig and Rohrer’s article ‘In touch with atoms’,3 where a finger profiles an atomic
surface. Likewise, one could argue that stylus profilometryis a predecessor of AFM. However,
AFM and stylus profilometry have as much in common as a candle and a laser. Both of the
2
latter generate light, and even though candles are masterpieces of engineering,4 the laser is a
much more advanced technological device requiring a detailed knowledge of modern quantum
mechanics.5 While stylus profilometry is an extension of human capabilities that have been
known for ages and works by classical mechanics, AFM requires a detailed understanding of
the physics of chemical bonding forces and the technological prowess to measure forces that
are several orders of magnitude smaller than the forces acting in profilometry. Only the spectac-
ular spatial resolution of scanning tunneling microscopy (STM) could trigger the hope that the
force acting between any STM tip and sample might lead to atomic force microscopy capable
of true atomic resolution. The STM, established in 1981, is the first instrument that has allowed
to image surfaces with atomic resolution in real space.6, 7 The atomic imaging of the 7×7 re-
construction of Si (111) by STM in 19838 has later helped to solve one of the most intriguing
problems of surface science at that time and establish the dimer-adatom-stacking fault model by
Takayanagi et al.9 The capability of atomic resolution by STM provided immediate evidence
for the enormous value of this instrument as a tool for surface scientists. STM can only be used
on conductive surfaces. Given that many surfaces of technological interest are conducting or
at least semiconducting, this may not seem to be a severe shortcoming. One might think that
an STM should be capable of mapping the surface of a metallic surface at ambient conditions.
However, this is not feasible, because the pervasive layer of oxides and other contaminants
occurring at ambient conditions prevents stable tunnelingconditions. Electrical conductivity
is a necessary, but not a sufficient condition for a surface tobe imaged by STM with atomic
resolution, because surfaces need to be extremely clean on an atomic level. Except for a few
extremely inert surfaces such as graphite, atomic resolution is only possible in an ultra-high
vacuum with a pressure on the order of 10−8 Pa and special surface preparation. The invention
of the AFM by Binnig1 and its introduction by Binnig, Quate and Gerber2 opened the possibility
of obtaining true atomic resolution on conductorsand insulators. Indeed, it took only a short
3
time after the AFM’s invention before apparently atomic resolution on conductors10 and insu-
lators11–13 was obtained. While these early results reproduced the periodic lattice spacings of
the samples that were studied, single defects or step edges were not observed. Also, the forces
that acted between tip and sample were often orders of magnitudes larger than the forces that
a tip with a single front atom was expected to be able to sustain. Therefore, it was commonly
assumed thatmanytip atoms interacted with the surface at the same time in these early experi-
ments. The difference betweenapparentandtrue atomic resolution of a tip with many atomic
contacts can be illustrated by a macroscopic example: When profiling an egg crate with a single
egg, its trajectory would represent the overall periodicity of the crate as well as a dented hump
or a hole. However, when profiling one egg crate with another egg crate, again its periodicity
would be retained, but holes or dented humps would pass undetected. A similar effect can occur
when an AFM tip probes a surface. As long as single defects, steps or other singularities are
not observed, a clear proof fortrue atomic resolution is not established. Even though atomic
resolution was hardly ever achieved in initial AFM experiments, the technique was readily ac-
cepted and found many technological and scientific applications. The installed base of atomic
force microscopes rapidly outnumbered their STM counterparts. A recent survey14 about the
ten most highly cited publications ofPhysical Review Lettersranks the original AFM publica-
tion2 as number four (4251 citations as of Mar 11 2005 [ISI]) – in good company with other
breakthroughs in theoretical and experimental physics that have shaped our scientific life. Most
of these citations refer to AFM where the spatial resolutionis ‘only’ in the nanometer-range,
but the large number proves the vast applications of AFM. In spite of the rapid growth of AFM
applications, matching and even exceeding the spatial resolution of STM, its parent, had to wait
for new experimental developments.
4
2 Challenges of atomic resolution AFM
The technological foundations for the feasibility of STM with atomic resolution (theory of elec-
tron tunneling, mechanical actuation with pico-meter precision, vacuum technology, surface-
and tip preparation, vibration isolation, ...) were probably available a few decades before 1981,
but it took the bold approach by Gerd Binnig, Heinrich Rohrer, Christoph Gerber and Edi
Weibel to pursue atomic resolution in real space. Binnig andRohrer were rewarded with the
Nobel Prize in Physics in 1986 (together with Ernst Ruska, the inventor of Electron Microscopy)
and the AFM principle was published in the same year. The challenges of AFM with true atomic
resolution are even more daunting than the hurdles that troubled STM. To start our discussion
of the special AFM challenges, we first look at the physics behind STM. Figure 1 (a) shows a
schematic view of a sharp tip for STM or AFM close to a crystalline sample and Fig. 1 (b) is
a plot of the tunneling current and forces between tip and sample. When tip and sample are
conductive and a bias voltage is applied between them, a tunneling current can flow. The red
curve in Fig. 1 (b) shows the distance dependence of the tunneling currentIt. The exponential
decay ofIt with distance at a rate of approximately one order of magnitude per 100 pm distance
increase is the key physical characteristic that makes atomic resolution STM possible. Because
of its strong decay rate, the tunneling current is spatiallyconfined to the front atom of the tip
and flows mainly to the sample atom next to it (indicated by redcircles in Fig. 1 (a)). A second
helpful property of the tunneling current is its monotonic distance dependence. It is easy to
build a feedback mechanism that keeps the tip at a constant distance: if the actual tunneling
current is larger than the setpoint, the feedback needs to withdraw the tip and vice versa. The
tip sample forceFts, in contrast, does not share the helpful key characteristics of the tunneling
current. First,Fts is composed of long-range background forces depicted in light-blue in Fig. 1
(b) and originating from the atoms colored light-blue in Fig. 1 (a) and a short-range component
5
depicted in blue in Fig. 1 (b) and confined to the atoms printedin blue in Fig. 1 (a). Because the
short-range force is not monotonic, it is difficult to designa feedback loop that controls distance
by utilizing the force. A central task to perfect AFM is therefore the isolation the front atom’s
force contribution and the creation of a linear feedback signal from it.
Even if it was possible to isolate the short-range force, a more basic problem needs to be
solved first: how to measure small forces. For example, commonly known force meters such
as precise scales are delicate and expensive instruments and even top models rarely exceed
a mass resolution of 100µg, corresponding to a force resolution of 1µN. In addition, high-
precision scales take about one second to acquire a weight measurement so the bandwidth is
only 1 Hz. The force meters in AFM, in contrast, require a force resolution of at least a nano-
Newton at a typical bandwidth of 1 kHz. Most force meters determine the deflectionq′ of a
spring with given spring constantk that is subject to a forceF with F = q′/k. Measuring
small spring deflections is subject to thermal drift and other noise factors, resulting in a finite
deflection measurement accuracyδq′. The force resolution is thus given byδF = δq′/k, and
soft cantilevers provide less noise in the force measurement. In contact-mode AFM, where the
tip feels small repulsive forces from the sample surface, the cantilever should be softer than
the bonds between surface atoms (estimated at≈ 10 N/m), otherwise the sample deforms more
than the cantilever.15 Because of noise and stability considerations, spring constants below
1 N/m or so have been chosen for AFM in contact mode. However, atomic forces are usually
attractive in the distance regime that is best suited for atomic resolution imaging (approximately
a few hundred pm before making contact), and soft cantilevers suffer from a ”jump-to-contact”
phenomenon, i.e. when approaching the surface, the cantilever snaps towards the surface ended
by an uncontrolled landing. While true atomic resolution bycontact-mode AFM has been
demonstrated on samples that are chemically inert,16, 17 this method is not feasible for imaging
reactive surfaces where strong attractive short-range forces act. The long-range attractive forces
6
have been compensated in these experiments by pulling at thecantilever (negative loading force)
after jump-to-contact16 or by immersing cantilever and sample in water to reduce the van-der-
Waals attraction.17 Howald et al.18 could partially solve the reactivity problem by passivating
the reactive Si tip with a thin layer of poly-tetra-fluor-ethylen (teflon). The unit cell of Si(111)-
(7×7) was resolved, but atomic resolution was not reported withthat method of tip passivation.
In summary, AFM shares the challenges that are already knownfrom STM and uses many of
its design features (actuators, vibration isolation etc.), but nature has posed four extra problems
for atomic resolution AFM: 1. Jump-to-contact, 2. Non-monotonic short range forces, 3. Strong
long-range background forces and 4. Instrumental noise in force measurements.
3 Frequency modulation atomic force microscopy
Dynamic AFM modes19–21 help to alleviate two of the four major AFM challenges. Jump-to-
contact can be prevented by oscillating the cantilever at a large enough amplitudeA such that
the withdrawing force on the cantilever given byk × A is larger than the maximal attractive
force.22 Because the noise in cantilever deflection measurements hasa component that varies in
intensity inversely with frequency (1/f -noise), dynamic AFM modes are less subject to noise
than quasistatic operating modes. Non-monotonic interactions and strong long-range contribu-
tions are still present.
In amplitude modulation AFM,19 the cantilever is driven at a constant frequency and the
vibration amplitude is a measure of the tip-sample interaction. In 1991, Albrecht et al. have
shown that frequency modulation (FM) AFM20 offers even less noise at larger bandwidth than
amplitude modulation AFM. In FM-AFM, a cantilever with a high quality factorQ is driven
to oscillate at its eigenfrequency by positive feedback with an electronic circuit that keeps the
amplitudeA constant. A cantilever with a stiffness ofk and effective massm has an eigen-
frequency given byf0 = 1/(2π)√
k/m. When the cantilever is exposed to a tip-sample force
7
gradientkts, its frequency changes instantly tof = f0 +∆f = 1/(2π)√
k′/m with k′ = k+kts
(see Fig. 2). Whenkts is small compared tok, the square root can be expanded and the fre-
quency shift is simply given by20
∆f(z) =f0
2kkts(z). (1)
This formula is only correct ifkts is constant over the distance range fromz − A to z + A that
is covered by the oscillating cantilever. The force gradient kts was probably almost constant
within the oscillation interval in the first application of FM-AFM in magnetic force microscopy
by Albrecht et al.,20 where magnetic recording media with magnetic transitions spaced by about
2µm were imaged with a cantilever with a stiffness of about 10 N/m oscillating at an amplitude
of about 5 nm. In contrast, in the newer application of FM-AFMin atomic-resolution AFM, the
force gradient varies by orders of magnitude throughout theoscillation of the cantilever. Using
frequency modulation AFM, true atomic resolution on Si(111)-(7×7), a fairly reactive sample,
could be achieved in 1994.23 Figure 3 shows the topographic image of this data. The fast scan-
ning direction was horizontal, and the atomic contrast is rather poor in the lower section, quite
good in a narrow strip in the center part and vanishing in the top section. These changes in
contrast were due to tip changes, indicating fairly strong interaction during the imaging pro-
cess. A piezoresistive cantilever made of Si24 as shown in Fig. 4(a) with a stiffness of17 N/m
was used to obtain this image. The amplitude of the cantilever can be freely adjusted by the
operator, and while it was even planned to use the thermally excited amplitude25 (≈ 10 pm),
the empirically determined optimal amplitude values were always around 10 nm – on a similar
order of magnitude as the value ofA = 34 nm used in the data of Fig. 3. The chemical bonding
forces that are responsible for the atomic contrast in imaging Si by AFM have a range on the
order of 100 pm,26 and the amplitude was 340 times as large. The requirement of such a large
amplitude is in stark contrast to intuition. Imagine an atommagnified to a size of an orange
8
with a diameter of 8 cm. The range of the bonding force is then only 4 cm or so, and the front
atom of the cantilever would approach from a distance of 20 m and only in the last few cen-
timeters of its oscillation cycle would feel the attractivebonding forces from the sample atom
next to it. On the other hand, force gradients can be quite large in chemical bonds. According
to the well-known Stillinger-Weber potential,27 a classic model potential the interaction of Si
atoms in the solid and liquid phases, a single bond between two Si atoms has a force gradient of
kts ≈ +170 N/m at the equilibrium distance ofz = 235 pm andkts ≈ −120 N/m when the two
Si atoms are at a distance ofz = 335 pm. Because of the relatively large values of interatomic
force gradients, even cantilevers with a stiffness on the order of 1 kN/m should be subject to
significant frequency shifts when oscillating at small amplitudes (page 5 in Ref28).
Nevertheless, the large-amplitude FM technique has celebrated great successes by imaging
metals, semiconductors and insulators with true atomic resolution.29–33
4 The search for optimal imaging parameters
In order to understand why these large oscillation amplitudes were necessary, a quantitative
analysis of the physics of large amplitude FM-AFM was necessary, starting with a calculation
of frequency shift for large amplitudes. Ifkts is not constant over one oscillation cycle, Eq. 1
no longer holds and perturbation theory such as the Hamilton-Jacobi theory34 can be used to
find the relationship between frequency and tip sample forces.22 Other perturbative approaches
have confirmed the result,35–38and an instructive representation of the formula is
∆f(z) =f0
πk
∫ 1
−1kts(z − uA)
√1 − u2du. (2)
This equation is key to a physical understanding of FM-AFM allowing to evaluate the impact
of various force components on∆f , the experimental observable. On a first glance, the large-
amplitude result resembles Eq. 1 wherekts(z) is replaced by an averaged value. The average
9
force gradient is computed by convolutingkts(z) in the intervalz − A to z + A with a semi-
spherical weight function. The weight function has its maximum atu = 0 – a distanceA away
from the minimal tip-sample distance. The minimal tip sample distancezmin is an important
parameter in any STM or AFM experiment, because while a smallvalue ofzmin is desirable
for optimal spatial resolution, both tip and sample can be damaged ifzmin is too small. We
can now ask, if we keepzmin constant and varyA, what happens to our signal, the frequency
shift ∆f? The answer is given in Eq. 2: as long as the gradient of the tip-sample interaction
kts remains constant as the tip of the cantilever moves over az-range fromzmin to zmin + 2A,
∆f stays constant. However, asA reaches the decay lengthλ of the interaction, the frequency
shift drops sharply at a rate∝ (λ/A)3/2. It turns out that for amplitudes larger thanλ, ∆f
is no longer proportional to the force gradient, but to the product of force and the square root
of λ39 (or, equivalently to the geometric average between potential and force40). In FM-AFM
with amplitudes large compared to the interaction range, itis useful to define a quantityγ =
∆fkA3/2/f0.22 The ’normalized frequency shift’γ connects the physical observable∆f and
the underlying forcesFts with rangeλ, whereγ ≈ 0.4Ftsλ1/2 (see Eqs. 35-41 in33). For
covalent bonds, the typical bonding strength is on the orderof -1 nN with λ ≈ 1 A, resulting
in γ ≈ −4 fN√
m, where a negative sign indicates attractive interaction.The crossover from
the small-amplitude approximation in Eq. 1 to the large-amplitude case in Eq. 2 occurs for
amplitudes on the order of the interaction rangeλ.
Equation 2 determines the influence of the oscillation amplitude on AFM challenge number
3 (disturbing contribution of long-range forces) outlinedin the second section: Imagine an
AFM tip at a minimal distancezmin = 0.3 nm to a surface, where the total tip-sample force is
composed of a chemical bonding force with an exponential distance dependence and a given
range with a long-range force with the same strength and a tentimes longer range (see caption
of Table 1 for details). In large-amplitude AFM (here,A > 1 nm), the signal is proportional to
10
the normalized frequency shiftγ, and the long-range contribution to∆f is√
1 nm/100 pm≈ 3-
times larger than the short-range contribution. For small amplitudes (here,A < 100 pm),∆f is
proportional to the force gradient and the long-range component is only100 pm/1 nm = 1/10
of the short-range contribution. Therefore, small amplitude AFM helps to reduce the unwanted
contribution of long-range forces.
Even stronger attenuation of the unwanted long-range contribution would be possible if
higher order force derivatives could be mapped directly. For example, if we could directly mea-
sure∂2Fts/∂z2, the long range component was only 1/100 of the short-range contribution, and
for a direct mapping of the third order gradient∂3Fts/∂z3, the relative long range component
would reduce to a mere 1/1000. Higher force gradients can be mapped directly by higher-
harmonic AFM, as shown further below.
AFM physical short-range long-range relative short-method observable contribution contribution range contribution
quasistatic force 1 nN 1 nN 50 %large amplitude FM γ ≈ 0.4×force×
√range 4 fN
√m 12 fN
√m 25 %
small amplitude FM force gradient 10 N/m 1 N/m 91 %higher-harmonic n-th force gradient 10n+9(n−1) N/mn 109(n−1) N/mn ≈ 100 %(1 − 10−n)
Table 1: Short- and long-range contributions to AFM signalsin different operating modes.This model calculation assumes a chemical bonding forceF (z) = F0e
−z/λ with a strength ofFshort range(zmin) = 1 nN and a range ofλshort range = 100 pm and an equally strong long-rangebackground force withFlong range(zmin) = 1 nN and a range ofλlong range = 1 nm. Dependingon the mode of AFM operation, the short-range part has a different weight in the total interactionsignal. Higher-harmonic AFM offers the greatest attenuation of long-range forces.
Because the forces that act in AFM are small, optimizing the signal-to-noise ratio is crucial
for obtaining good images. Frequency noise in FM-AFM is inversely proportional to ampli-
tude.19, 20, 33, 41 As discussed above, the signal stays constant untilA reachesλ and drops pro-
portional to(λ/A)3/2 for larger amplitudes. Therefore, the signal-to-noise ratio is maximal for
amplitudes on the order of the decay length of the interaction that is used for imaging.42 For
11
atomic imaging, amplitudes on the order of 100 pm are expected to be optimal.
As a conclusion of these calculations, we find that the use of small amplitudesA ≈ λ would
have two advantages:
1. Increased signal-to-noise ratio42
2. Greater sensitivity to short-range forces.33
So why was it not feasible to use small amplitudes in the initial experiments? Two reasons,
related to the mechanical stability of the oscillating cantilever, can be identified. First, jump-to-
contact is prevented if the withdrawing force of the cantilever when it is closest to the sample
given byk × A is larger than the maximal attraction.22 Second, because tip-sample forces are
not conservative,43 random dissipative phenomena with a magnitude ofδEts cause amplitude
fluctuationsδA = δEts/(kA).42, 44 Both problems can be resolved by utilizing cantilevers with
sufficient stiffness. Stability considerations propose a lower threshold fork that depends on the
tip-sample dissipation as well as theQ-factor of the cantilever. Because the frequency shift is
inversely proportional to the stiffness (Eqs. 1 and 2),k should still be chosen as low as permitted
by the stability requirements. Stiff cantilevers were not commercially available when we real-
ized their potential advantages, therefore we built cantilevers with a stiffness ofk = 1800 N/m
from quartz tuning forks45, 46(see Fig. 4(b)). A secondary advantage of quartz cantilevers is their
greater frequency stability with temperature, which leadsto lower frequency drift in particular
if a quartz stabilized frequency detector is used (we used the EasyPLL by Nanosurf AG, Liestal,
Switzerland). Other small-amplitude approaches with stiff home-built tungsten cantilevers have
been demonstrated in Ragnar Erlandsson’s47 and John Pethica’s groups.48–50 As predicted by
theoretical considerations, the stiff cantilever allowedto use sub-nm amplitudes, resulting in an
improved signal-to-noise ratio, a strong attenuation of the disturbing long-range forces and the
possibility of stable scanning at very small tip-sample distances. For these reasons, the spatial
12
resolution was increased as shown in Fig. 5. The image shows avery clear picture of Si with
a defect and very large corrugation. The adatoms of Si which should be spherically symmetric
showed subatomic details that are interpreted as orbitals in the tip atom.51, 52 This AFM image
seemed to show greater resolution than what was known from STM. According to the ‘Stoll-
formula’,53 a theoretical estimate of the vertical corrugation and thusthe lateral resolution of
STM images, two physical parameters are crucial for the highspatial resolution of STM: a)
the very short decay length of the tunneling current and b) a small tip-sample distance. Three
likely reasons have been identified that may explain why dynamic AFM could provide better
resolution than STM:54
1. In dynamic AFM, the minimal tip sample distance can be muchsmaller than in STM
without destroying the tip, because the shear forces that act on the front atom during
scanning are much smaller in the oscillation phase where thetip is far from the sample.
2. When using large gap voltages, a variety of states can contribute to the tunneling current,
smearing out the image.
3. Tip-sample forces also have repulsive components with a very short decay length.
The first two characteristics can be fulfilled in STM as well byusing a very small tunneling bias
voltage and oscillating the STM tip similar to an AFM tip. Figure 6 shows an image obtained
in dynamic STM where a Co6Fe3Sm magnetic tip was mounted onto a qPlus sensor, imaging
Si.55, 56 Each Si adatom looks like a fried egg with a sharp center peak surrounded by a halo.
The radius of the center peak is only on the order of 100 pm, showing that higher-momentum
states57 must have been involved in this image. The experiment was repeated with pure Co, Fe
and Sm tips, and only pure Sm tips yielded similar images as Figure 6, we therefore concluded
that a Sm atom acted as the tip atom in this experiment.55 In atomic samarium, the electrons at
highest occupied state are in a 4f state. If one assumes, thatthe electronic states at a Sm surface
13
atom of bulk Co6Fe3Sm are similar to atomic states in Sm, it appears likely that the crystal field
around the front atom creates a state close to 4fz3 symmetry that is responsible for the tunneling
contrast. Interestingly, very small tip-sample distancescould only be realized with oscillating
tips. When the oscillation was turned off, the current setpoint had to be reduced otherwise the
tip would not survive the small tunneling distances.
Operation at small oscillation amplitudes not only resultsin greater resolution, it also facil-
itates simultaneous STM and AFM imaging. A straightforwardimplementation of combined
current- and force measurements uses the constant-height mode, where thez-position of the tip
is held constant relative to the plane connecting the surface atoms. A simultaneous measurement
of tunneling current and frequency shift allows to compare forces and tunneling currents. Figure
7 shows a comparison of current and repulsive force on graphite58 observed by simultaneous
AFM and STM in vacuum at liquid helium temperatures (4.9 K). STM only sees the electrons
at the Fermi level, while repulsive forces act wherever the local charge density is high (i.e. over
everyatom) for small enough distances. In graphite, only every second surface atom conducts
electricity, but every surface atom exerts repulsive forces. Therefore, AFM “sees more” than
STM and allows to correlate topography to local conductance. This method is promising for
other materials with more than one basis atom in the elementary cell. The images have been
taken with a low-temperature AFM/STM operating at 4.9 K in ultra-high vacuum.59, 60
While a strong bias dependence holds both for atomic-resolution STM61 as well as AFM
images,60, 62 one pronounced difference is that the direction of the tunneling current is not ac-
cessible in STM, while the direction of the measured force isdetermined by the orientation of
the cantilever. Usually, AFM senses forces that are normal to the surface, but it is also possible
to perform lateral force microscopy63 by measuring the forces acting parallel to the surface. In a
quasistatic mode, lateral forces can be recorded simultaneously with normal forces. In dynamic
modes, it is easier to rotate the attachment of the cantilever by 90 degrees and detect lateral
14
forces. Figure 8 shows a measurement of the lateral force gradients between a tip and a Si
surface. Parallel motion between tip and cantilever also allows to use extremely soft cantilevers
without suffering jump-to-contact to probe the limits of force resolution, as demonstrated by
Rugar et al. in single spin detection by magnetic resonance force microscopy.64
5 Higher-harmonic atomic force microscopy
Can we increase the spatial resolution of AFM any further? When decreasing the amplitude
from A >> λ to A << λ, the frequency shift changes from a proportionality ofFts
√λ to
Fts/λ. As outlined above, an experimental observable that is proportional to a higher force
gradient should allow even higher spatial resolution than small-amplitude FM-AFM. Luckily,
there is a physical observable that couples directly to higher force gradients. When the can-
tilever oscillates in the force field of the sample, a shift infrequency is not the only change
in the cantilever’s motions. The oscillation of the cantilever changes from a purely sinu-
soidal motion given byq′ = A cos(2πft) to an oscillation that contains higher harmonics with
q′ =∑
∞
n=0 an cos(2πnft + φn). For amplitudes that are large with respect to the range of
Fts, the higher harmonics are essentially proportional to∆f .37 However, for small amplitudes,
Durig has found thatFts could be recovered immediately within the distance range from zmin
to zmin + 2A if the amplitudes and phases of all higher harmonics of the cantilever’s motion
were known.65 Moreover, higher harmonics bear even more useful information: direct cou-
pling to higher force gradients.66 Similar to Eq. 2, we can express the magnitude of the higher
harmonics by a weighted average of a force gradient – a gradient of ordern > 1 this time:
an =2
πk
1
1 − n2
An
1 · 3 · ... · (2n − 1)
∫ 1
−1
dnFts(z + Au)
dzn(1 − u2)n−1/2du. (3)
The weight function changes from the semi-spherical shapew∆f(u) = (1 − u2)1/2 in Eq. 2 to
functionswn(u) = (1 − u2)n−1/2 that are more and more peaked with increasingn. For this
15
reason, the use of small amplitudes is of even greater importance in higher-harmonic AFM than
in FM-AFM. The magnitude of the higher harmonic amplitudesan is rather small compared
to the fundamental amplitudea1 = A, therefore higher harmonic AFM works best at low
temperatures, where the detection bandwidth can be set to very small values.
The spatial resolution of AFM and STM is fundamentally neither limited by the mechanical
vibration level nor by thermal vibrations, but by the spatial extent of the experimental objects
that are observed – electrons at the Fermi level in STM,67 and something close to the total charge
density in repulsive AFM.68 When probing the resolution limits of AFM, we first have to find
an object with the desired sharply localized electronic states. Pauling69 has noted, that transi-
tion metals show a covalent bonding character, and should therefore expose lobes of increased
charge density towards their neighbors. Indeed, while the surface atoms of W(001) expose a
large blurred charge cloud at the Fermi level fork-vectors perpendicular to the surface (Fig. 8
in Ref.70), the total charge density shows four distinct maxima (Fig.3 in Ref.70 and Fig. 3(a) in
Ref.71). Figure 9 shows a direct comparison of the simultaneously recorded tunneling current
and the amplitudes of the higher harmonics. As expected, thehigher harmonic data shows much
greater detail.
6 Summary and Conclusion
We have emphasized the enormous usefulness of AFM by referring to the numerous references
to the original publication2 in the introduction. While most AFM applications are currently not
in the atomic resolution regime, the enhancement in spatialresolution is likely to add significant
value in most AFM studies in physics, chemistry, biology andmaterials science. Recently, true
atomic resolution by FM-AFM has been observed at ambient pressure in an N2 atmosphere,72
showing that some of the concepts of vacuum AFM are applicable in ambient environments.
Although STM resolution can benefit from oscillating the tip, a concept that has originated in
16
AFM, Fig. 9 shows that AFM has now clearly reached and even surpassed the resolution ca-
pability of STM. Figure 10 shows the evolution of the resolution of AFM from large amplitude
AFM in 1994 (a) to small amplitude AFM in 2000 (b) and higher harmonic AFM in 2004 (c).
While the structures within single atoms shown in Fig. 10 (b)and (c) originate in the front
atom of the probe, other examples where AFM shows more atomicdetails of specimens than
STM such as the observation of the rest atoms in Si(111)-(7×7)73, 74 or the observation of all
dangling bonds on the Si/Ge(105) surface75 establish the improved spatial resolution of AFM
over STM in special cases. Atomic- and molecular structuring has been the domain of STM for
a long time, starting with the first demonstration of manipulating single atoms76 to a variety of
nano-structuring methods by STM.77 Recently, it has been shown that atomic manipulation by
AFM is possible even at room temperature.78
We have not been able to discuss the phenomenal success of AFMin biology, a field with
a much more immediate impact on the human condition. It can beexpected that at least some
of the concepts that have been developed for AFM in vacuum will enable greater resolution in
biological AFM applications as well.79, 80
Acknowledgments
I wish to thank Jochen Mannhart for support and editorial suggestions and my current
and former students Martin Breitschaft, Philipp Feldpausch, Stefan Hembacher, Markus Herz,
Christian Schiller, Ulrich Mair, Thomas Ottenthal and Martina Schmid for their contributions
towards the progress of AFM. I also thank Gerd Binnig, CalvinF. Quate and Christoph Ger-
ber for kicking off the fun of the AFM field and for ongoing inspiring interactions. Special
thanks to Heinrich Rohrer, Calvin Quate and Christoph Gerber for critical comments and to
German Hammerl for help with LaTeX. Supported by the Bundesministerium fur Forschung
und Technologie under contract 13N6918.
17
References
1. Binnig, G.Atomic Force Microscope and Method for Imaging Surfaces with Atomic Reso-
lution (1986)US Patent 4,724,318
2. Binnig, G., Quate, C. F., and Gerber, C.Phys. Rev. Lett.(1986)56, 930
3. Binnig, G. and Rohrer, H.Rev. Mod. Phys.(1999)71, S324
4. Faraday, M.Chemical History of a Candle. reprinted in 2003, Dover Publications, Mineola,
NY (1861)
5. Siegman, A. E.Lasers. University Science Books, Herndon, VA, USA (1986)
6. Binnig, G., Rohrer, H., Gerber, C., and Weibel, E.Phys. Rev. Lett.(1982)49, 57
7. Hofer, W. A.Materials Today(2002)October, 24
8. Binnig, G., Rohrer, H., Gerber, C., and Weibel, E.Phys. Rev. Lett.(1983)50, 120
9. Takayanagi, K., Tanishiro, Y., Takahashi, M., and Takahashi, S. J. Vac. Sci. Technol. A
(1985)3, 1502
10. Binnig, G., Gerber, C., Stoll, E., Albrecht, T. R., and Quate, C. F.Europhys. Lett.(1987)3,
1281
11. Albrecht, T. R. and Quate, C. F.J. Appl. Phys.(1987)62, 2599
12. Meyer, G. and Amer, N. M.Appl. Phys. Lett.(1990)56, 2100
13. Meyer, E., Heinzelmann, H., Rudin, H., and Guntherodt,H.-J. Z. Phys. B(1990)79, 3
14. Riordon, J.APS News(2003)May issue, 3
18
15. Rugar, D. and Hansma, P.Physics Today(1990)43, 23
16. Giessibl, F. J. and Binnig, G.Ultramicroscopy(1992)42-44, 281
17. Ohnesorge, F. and Binnig, G.Science(1993)260, 1451
18. Howald, L., Luthi, R., Meyer, E., and Guntherodt, H.-J. Phys. Rev. B(1995)51, 5484
19. Martin, Y., Williams, C. C., and Wickramasinghe, H. K.J. Appl. Phys.(1987)61, 4723
20. Albrecht, T. R., Grutter, P., Horne, H. K., and Rugar, D.J. Appl. Phys.(1991)69, 668
21. Durig, U., Zuger, O., and Stalder, A.Journal of Applied Physics(1992)72, 1778
22. Giessibl, F. J.Phys. Rev. B(1997)56, 16010
23. Giessibl, F. J.Science(1995)267, 68
24. Tortonese, M., Barrett, R. C., and Quate, C.Appl. Phys. Lett.(1993)62, 834
25. Giessibl, F. J.Jpn. J. Appl. Phys.(1994)33, 3726
26. Perez, R., Stich, I., Payne, M. C., and Terakura, K.Phys. Rev. Lett.(1997)78, 678
27. Stillinger, F. H. and Weber, T. A.Phys. Rev. B(1985)31, 5262
28. Giessibl, F. J.German Patent(1996) Patentschrift DE 196 33 546
29. Morita, S., Wiesendanger, R., and Meyer, E., eds.Noncontact Atomic Force Microscopy.
Nanoscience and Technology. Springer Berlin Heidelberg New York (2002)
30. Meyer, E., Hug, H. J., and Bennewitz, R.Scanning Probe Microscopy. The Lab on a Tip.
Springer, Berlin Heidelberg New York (2003)
31. Garcia, R. and Perez, R.Surf. Sci. Rep.(2002)47, 197
19
32. Hofer, W. A., Foster, A. S., and Shluger, A. L.Rev. Mod. Phys.(2003)75, 1287
33. Giessibl, F. J.Rev. Mod. Phys.(2003)75, 949
34. Goldstein, H.Classical Mechanics. Addison Wesley, Reading MA (1980)
35. Baratoff, A.unpublished(1997)
36. Durig, U.Surf. and Interf. Anal.(1999)27, 467
37. Durig, U.Appl. Phys. Lett.(1999)75, 433
38. Livshits, A., Shluger, A., and Rohl, A.Appl. Surf. Sci.(1999)140, 327
39. Giessibl, F. J. and Bielefeldt, H.Phys. Rev. B(2000)61, 9968
40. Ke, S. H., Uda, T., and Terakura, K.Phys. Rev. B(1999)59, 13267
41. Hasegawa, Y., Eguchi, T., An, T.,et al.. Japanese Journal of Applied Physics(2004)43,
L303
42. Giessibl, F. J., Bielefeldt, H., Hembacher, S., and Mannhart, J.Appl. Surf. Sci.(1999)140,
352
43. Kantorovich, L. N.Phys. Rev. B(2001)64, 245409
44. Giessibl, F. J., Hembacher, S., Herz, M., Schiller, C., and Mannhart, J.Nanotechnology
(2004)15, S79
45. Giessibl, F. J.Appl. Phys. Lett.(1998)73, 3956
46. Giessibl, F. J.Appl. Phys. Lett.(2000)76, 1470
47. Erlandsson, R., Olsson, L., and Martensson, P.Phys. Rev. B(1997)54, R8309
20
48. Hoffmann, P. M., Oral, A., Grimble, R. A.,et al.. Proc. R. Soc. London A(2001)457, 1161
49. Hoffmann, P. M., Jeffery, S., Pethica, J. B.,OzgurOzer, H., and Oral, A.Physical Review
Letters(2001)87, 265502
50. Oral, A., Grimble, R. A.,OzgurOzer, H., Hoffmann, P. M., and Pethica, J. B.Appl. Phys.
Lett. (2001)79, 1915
51. Giessibl, F. J., Hembacher, S., Bielefeldt, H., and Mannhart, J.Science(2000)289, 422
52. Huang, M., Cuma, M., and Liu, F.Phys. Rev. Lett.(2003)90, 256101
53. Stoll, E.Surf. Sci. Lett.(1984)143, L411
54. Giessibl, F. J., Bielefeldt, H., Hembacher, S., and Mannhart, J.Ann. Phys. (Leipzig)(2001)
10, 887
55. Herz, M., Giessibl, F. J., and Mannhart, J.Phys. Rev. B(2003)68, 045301
56. Herz, M. Dynamische Tunnel-, Kraft- und Reibungsmikroskopie mit atomarer und sub-
atomarer Auflosung. Ph.D. thesis, University of Augsburg, Germany (2003)
57. Chen, C. J.Introduction to Scanning Tunneling Microscopy. Oxford University Press, New
York (1993)
58. Hembacher, S., Giessibl, F. J., Mannhart, J., and Quate,C. F. Proc. Natl. Acad. Sci. (USA)
(2003)100, 12539
59. Hembacher, S. F.Simultane Rasterkraft- und Rastertunnelmikroskopie bei 5K im Ultra-
hochvakuum. Ph.D. thesis, Universitat Augsburg, Germany (2003)
60. Hembacher, S., Giessibl, F. J., and Mannhart, J.Phys. Rev. Lett.(2005)94, 056101
21
61. Feenstra, R. M., Stroscio, J. A., Tersoff, J., and Fein, A. P. Phys. Rev. Lett.(1987)58, 1192
62. Arai, T. and Tomitori, M.Phys. Rev. Lett.(2004)93, 256101
63. Mate, M., McClelland, G. M., Erlandsson, R., and Chiang,C. Phys. Rev. Lett.(1987)59,
1942
64. Rugar, D., Budakian, R., Mamin, H. J., and Chui, B.Nature(2004)430, 329
65. Durig, U.New Journal of Physics(2000)2, 5.1
66. Hembacher, S., Giessibl, F. J., and Mannhart, J.Science(2004)305, 380
67. Tersoff, J. and Hamann, D. R.Phys. Rev. Lett.(1985)50, 1998
68. Ciraci, S., Baratoff, A., and Batra, I. P.Phys. Rev. B(1990)41, 2763
69. Pauling, L. The Nature of the Chemical Bond. Cornell Univ. Press, Ithaca, New York
(1957)
70. Posternak, M., Krakauer, H., Freeman, A. J., and Koelling, D. D. Phys. Rev. B(1980)21,
5601
71. Mattheiss, L. F. and Hamann, D. R.Phys. Rev. B(1984)29, 5372
72. Sasahara, A., Kitamura, S., Uetsuka, H., and Onishi, H.Journal of Physical Chemistry B
(2004)108, 15735
73. Lantz, M., Hug, H. J., van Schendel, P.,et al.. Phys. Rev. Lett.(2000)84, 2642
74. Eguchi, T. and Hasegawa, Y.Phys. Rev. Lett.(2002)89, 266105
75. Eguchi, T., Fujikawa, Y., Akiyama, K.,et al.. Phys. Rev. Lett.(2004)93, 266102
22
76. Eigler, D. M. and Schweizer, E. K.Nature(1990)344, 524
77. Rosei, F.Journal of Physics-Condensed Matter(2004)16, S1373
78. Sugimoto, Y., Abe, M., Hirayama, S.,et al.. Nature Materials(2005)4, 156159
79. Miles, M. Science(1997)277, 1845
80. Horber, J. K. H. and Miles, M. J.Science(2003)302, 1002
23
Figures
Figure 1: (a) Schematic view of a tip and a sample in a scanningtunneling microscope oratomic force microscope. The diameter of a metal atom is typically 0.3 nm. (b) Qualitativedistance dependence of tunneling current, long- and short-range forces. The tunneling currentincreases monotonic with decreasing distance, while the force reaches a minimum and increasesfor distances below the bond length.
24
Figure 2: (a) Schematic view of a vibrating tip close to a a sample in a dynamic atomic forcemicroscope. The forces that act between the tip and the sampleFts cause a detectable change inthe oscillation properties of the cantilever. (b) Mechanical equivalent of (a). The free cantileverwith stiffnessk and effective massm can be treated as a harmonic oscillator with an eigen-frequencyf0 = (k/m)1/2/(2π). The bond between tip and sample with its stiffnesskts altersthe resonance frequency tof = ([k + kts]/m)1/2/(2π). When the oscillation amplitude of thecantilever is large,kts can vary significantly within one oscillation cycle, requiring averaging(see text).
25
Figure 3: First AFM image of a reactive surface showing true atomic resolution: Si(111)-(7×7)reconstruction. Parameters:k = 17 N/m, A = 34 nm,f0 = 114 kHz, ∆f = −70 Hz andQ =28 000, scanning speed = 3.2 lines/s. Environment: ultra-high vacuum, room temperature.23
26
Figure 4: Micrographs of (a) a piezoresistive cantilever24 and (b) a ‘qPlus’ sensor46 - a cantilevermade from a quartz tuning fork. The piezoresistive cantilever has a length of 250µm, a widthof 50µm and a thickness of 4µm. The eigenfrequency is 114 kHz, the stiffness 17 N/m and theQ-factor in vacuum 28 000. The qPlus sensor has a typical eigenfrequency ranging from 10 to30 kHz (depending on the mass of the tip), a stiffness of 1 800 N/m and aQ-factor of 4000 invacuum atT = 300 K and 20 000 atT = 4 K. One of the prongs is fixed to a large substrateand a tip is mounted to the free prong. Because the fixed prong is attached to a heavy mass, thedevice is mechanically equivalent to a traditional cantilever. The dimensions of the free prongare: Length: 2.4 mm, width: 130µm, thickness: 214µm.
27
Figure 5: AFM image of the silicon 7×7 reconstruction with true atomic resolution with a stiffcantilever. Parameters:k = 1800 N/m, A = 0.8 nm, f0 = 16.86 kHz, ∆f = −160 Hz andQ = 4 000. Environment: ultra-high vacuum, room temperature.51
28
Figure 6: (a) Dynamic STM image of the silicon 7×7 reconstruction where a Co6Fe3Sm tip wasmounted on a qPlus sensor. Parameters:k = 1800 N/m, A = 0.5 nm,f0 = 19 621Hz, samplebias voltage -100 mV, average tunneling current 200 pA. (b) Schematic plot of tip and samplestates that can lead to the experimental image shown in (a). The sample state is a dangling bondof a Si adatom with its 3sp3 symmetry, while a Sm 4fz3 state is taken as a tip state. Environment:ultra-high vacuum, room temperature.56
29
Figure 7: (a) Constant height STM image of graphite and (b) a simultaneously recorded AFMimage (repulsive). Part (c) shows an estimate of the charge density at the Fermi level (visibleby STM) and (d) the total charge density (relevant for repulsive AFM) for graphite. Parameters:k = 1800N/m, A = 0.3 nm,f0 = 18076.5 Hz, andQ = 20 000.58
30
Figure 8: (a) Topographic STM image of Si(111)-(7×7) where the tip is mounted on a lateralforce sensor. The tip oscillates withA ≈ 80 pm in they-direction in the lower half of theimage, the oscillation is turned off in the upper half. (b) Corresponding lateral force gradient.On top of the adatoms, the bond between tip and sample causes an increase in frequency shift.Parameters:k = 1350 N/m, A = 80 pm (bottom),A = 0 (top),f0 = 10214 Hz. Environment:ultra-high vacuum, room temperature.56
31
Figure 9: Simultaneous constant height STM (left column) and higher-harmonic AFM images(center column) of graphite with a tungsten tip. The right column shows the proposed orienta-tion of the W tip atom. The W atom is represented by its Wigner-Seitz unit cell, which reflectsthe full symmetry of the bulk. We assume, that the bonding symmetry of the adatom is similarto the bonding symmetry of the bulk. This assumption is basedon charge density calculationsof surface atoms70, 71 In the first row, the higher harmonics show a two-fold symmetry, as result-ing from a [110] orientation of the front atom. In the second row, the higher harmonics showroughly a three-fold symmetry, as expected for a [111] orientation. In the third row, the sym-metry of the higher-harmonic signal is approximately four-fold, as expected for a tip in [001]orientation. Parameters:k = 1800N/m, A = 0.3 nm, f0 = 18 076.5 Hz, andQ = 20 000.Environment: ultra-high vacuum,T = 4.9 K.66
32
Figure 10: Progress in spatial resolution of AFM showing images of single atoms. The lateralscale in (a)-(c) is equal. (a) An adatom of the Si(111)-(7×7) reconstruction, showing up asa blurred spot. (b) An adatom of the Si(111)-(7×7) reconstruction, showing subatomic con-trast originating in the electronic structure of the tip. (c) Higher-harmonic image of a tungstenatom mapped by a carbon atom. Parameters: (a)k = 17 N/m, A = 34 nm, f0 = 114 kHz,∆f = −70 Hz andQ = 28 000 (ultra-high vacuum, room temperature), (b)k = 1800 N/m,A = 0.8 nm, f0 = 16 860Hz, ∆f = −160 Hz andQ = 4 000 (ultra-high vacuum, room tem-perature), (c)k = 1800N/m, A = 0.3 nm, f0 = 18 076.5 Hz, andQ = 20 000 (ultra-highvacuum,T = 4.9 K), higher harmonic detection. (d) Schematic view of a Si(001) tip close to aSi(111)-(7×7) surface. Because of the large amplitude and a fairly largeminimum tip-sampledistance, the blurry image (a) corresponding to this configuration is approximately symmetricwith respect to the vertical axis. (e) Similar to (d), but at acloser distance. The angular depen-dence of the bonding forces is noticeable. (f) W(001) surface close to a C atom in a graphitesurface. The charge distribution in W shows small pockets that are resolved by higher-harmonicAFM with a light-atom carbon-probe.
33