attachment a ch2m hill soil boring log guidelines · (uscs), astm d 2487-85 (figures 3 and 4). the...

84
Attachment A CH2M HILL SOIL BORING LOG GUIDELINES

Upload: others

Post on 08-May-2020

9 views

Category:

Documents


0 download

TRANSCRIPT

Attachment ACH2M HILL SOIL BORING LOG GUIDELINES

SOIL BORING LOG GUIDELINE

SOIL BORING LOG POLICY

This soil boring guideline will be used for all CH2M HILLprojects in which soil boring techniques are used duringgeotechnical field exploration. The purpose of the guide-line is to assist CH2M HILL staff in accurately recordingand presenting all field data that are necessary to suffi-ciently describe, label, and package recovered soil samplesin a consistent manner. The guideline establishes theminimum kinds of information that must be recorded in thefield to adequately characterize recovered soil samples.

Because each of our projects is unique and because job re-quirements can vary widely, the minimum standards presentedin this guideline may need to be supplemented with addi-tional technical descriptions or field test results. How-ever, all soil boring field logs, regardless of specialproject circumstances, must include information addressed inthis guideline to achieve minimum acceptable standards re-quired by CH2M HILL,

All CH2M HILL staff members are encouraged to present theirsuggestions for clarification or improvements to this guide-line. Please submit all suggestions or comments in writingto the Geotechnical Discipline Group Director.

RECORDING SOIL BORING FIELD DATA

CH2M HILL Standard Form D1586, the Soil Boring Log form,will be used on all CH2M HILL projects for field logging(see Figure 1). Adherence to a standard format for record-ing data will help streamline our project efforts and ensurea consistent presentation of factual subsurface data. Allheading information must be completely filled out on eachlog sheet used, and all technical items in each column mustbe addressed in the field.

The boring log should be completed in the field according tothe attached instructions. Forms should be filled out neatlyand completely. Laboratory testing, if required, should beinitiated immediately after completion of the field work.Field classifications of samples should be checked againstthe laboratory test results, and corrections should be noted,initialed, and dated on the field log.

INSTRUCTIONS FOR COMPLETING SOIL BORING LOG, FORM D1586

Form D1586 is a standard CH2M HILL form that is available inweatherproof paper from all regional form distributors.

PD253.016.1

PROJECT NUMBER BORING NUMBER

SOIL BORING LOG

PROJECT. .OCATION

ELEVATION DRILLING CONTRACTOR.

DRILLING METHOD AND EQUIPMENT.

WATER LEVEL AND DATE—______ .START. FINISH. .LOGGER.

DE

PTH

MLO

WSU

RFA

CE

(FT)

-

-

-

-

-

-

-

.

-

-

-

SAMPLE

INTE

RVA

L

TYPE

AN

DN

UM

BER

REC

OVE

RY

|(F

T»STANDARD

PENETRATIONTEST

RESULTS

r-r-rINI

SOH. DESCRIPTION

SOIL MAMC. COLOR. MOISTURE CONTENT.RELATIVE OCNSITV OR CONSISTENCY. SOILSTRUCTURt. MINERALOGY. USCS GROUPSYMBOL

-

-

-

-

-

-

*

^

-

SYM

BO

LIC

LOO

COMMENTS

DEPTH OF CASINGDRILLING RATEDRILLING FLUID LOSSTESTS ANDINSTRUMENTATION

'

-

1

J

\

-I

-1

-

•1

JJ

11

]Figure 1SOIL BORING LOG.FORM 01586

Following are instructions for completing the log forms inthe field. See Figure 2 for an example of a field log com-pleted according to the instructions.

Field personnel should review logs on completion for accu-racy, clarity, and thoroughness of detail. Samples shouldalso be checked to see that information is correctly re-corded on both jar lids and labels and on the log sheets.

If any changes to the soil classification are made on thelog forms after completion of the field work, they should bedone in red, then initialed, and dated.

Heading Information

Project Number. Use standard region code, contract I.D.(5-digit), and point number designated for field explorationor geotechnical services.

Boring Number. Enter the boring number. A numbering systemshould be chosen that does not conflict with informationrecorded for previous exploratory work done at the site.Number the sheets consecutively for each boring. If rockcore log sheets are also used, continue the consecutive num-bering.

Project. Fill in the name of the project or client.

Location. If stationing, coordinates, mileposts, or similarproject layout information is available, indicate the posi-tion of the boring with respect to that system, using modi-fiers such as "approximate1* or "estimated" as appropriate.If this information is not available, identify the clientfacility (e.g., Richland STP, center of Clarifier No. 2site) or the town and state.

Elevation. Enter the elevation. If it is estimated from atopographic map, or if it is roughly determined using a handlevel, use the modifier "approximate." If the elevationsare to be surveyed later, or if the elevation is unknown,enter this information.

Drilling Contractor, Enter the name of the drilling companyand the city and state where the company is based.

Drilling Method and Equipment. Identify the bit size andtype, drilling fluid(e.g./ mud, Revert), and method ofdrilling (e.g., rotary, hollow-stem auger, air track).Information on the drilling equipment (e.g., CME 55, MobileB61) should also be entered here.

Water Level and Date. Enter the depth at which groundwateris first encountered. If frequent water measurements aretaken, the information should be recorded in the Comments

PD253.016.3

column. If water is not encountered during drilling, orcould not be detected because 01 the drilling method, thisinformation should be noted. Generally, water levels shouldbe measured each morning before resuming drilling and at thecompletion of each boring. Record date and time of day (fortides, river stage, etc.) of each water level measurement.

pate of Start and Finish. Enter the dates the boring wasbegun and completed.Time of day may be added if severalborings are performed on the same day.

Logger. Enter the first initial and full last name.

Technical Data

Depth Below Surface. Use a depth scale that is appropriatefor the sample spacing and for the complexity of subsurfaceconditions.

Sample Interval. Draw horizontal lines at the top and bot-tom depth of each sample interval. These lines shouldextend to the soil description column. For a very shortsample interval, the bottom line can be lowered after theinterval column to provide room for writing the information(see Figure 2). Enter the depth at the top and bottom ofthe sample interval.

Sample Type and Number. Enter the sample type and number.For instance/ SI = split spoon, first sample. Number sam-ples consecutively regardless of type. Enter a samplenumber even if no material was recovered in the sampler.

Sample Recovery. Enter the length to the nearest tenth of afoot of soil sample recovered from the sampler.

Standard Penetration Test Results. In this column enter thenumber of blows required for each 6 inches of sampler pene-tration and the "N" value, which is the sum of the blows inthe last two 6-inch penetration intervals. A typical stan-dard penetration test involving successive blow counts of 2,3, and 4 would be recorded as 2-3-4 and (7). The standardpenetration test will be terminated if the sampler encoun-ters refusal. Refusal is a penetration of more than6 inches but less than 12 inches with a blow count of 100,or a penetration of less than 6 inches with a blow count of50. A partial penetration of 50 blows for 4 inches is re-corded as 50/4".

Soil Description. The soil classification should follow theformat described in the section below entitled "Field Class-ification of Soil."

Symbolic Log. This column is usually omitted during field-work.

PD253.016.5

Comments. Include all pertinent observations {e.g./ changesin drilling fluid color, rod drops, drilling chatter, rodbounce as in driving on a cobble, damaged Shelby tubes, andequipment malfunctions). Also note if casing was used, thesizes and depths installed, and if drilling fluid was addedor changed. You should instruct the driller to alert you toany significant changes in drilling (e.g., changes in mate-rial, occurrence of boulders, and loss of drilling fluid).Such information should be attributed to the driller andrecorded in this column.

Specific information might include:

• The date and the time drilling began and endedeach day

• The depth and size of casing and the method ofinstallation

• Depth of rod chatter

• Depth and percentage of drilling fluid loss

• Depth of hole caving or heaving

• Depth of change in material

• Drilling interval through a boulder

• The results of pocket penetrometer or torvane testreported as:respectively

'PP TSF" or "TV TSF,"

The depth of piezometers and the results of in situ testsshould be noted in the Comments column.

FIELD CLASSIFICATION OF SOIL

This section presents the format for the field classifica-tion of soil. All descriptions of soil will be done inaccordance with the Unified Soil Classification System(USCS), ASTM D 2487-85 (Figures 3 and 4). The approach andformat for classifying soils should conform, in general, toASTM D 2488-84, Visual-Manual Procedure for Description andIdentification of Soils.

The Unified Soil Classification System is based on numericalvalues of certain soil properties, which are measured bylaboratory tests. It is possible, however, to estimatethese values in the field with reasonable accuracy usingvisual-manual procedures. Also, some elements or a completesoil description, such as the presence of cobbles or boul-ders, changes in strata, and the relative proportions of

PD253.016.6

GROUPSYMBOL GROUP NAME

* ••«• >* •>••» -" I.M* «H. «m

,r*M M». 200<~

^% «M<«pw«

% MM <% *»M«

LL250

/LL———«*—— \———————— *- 0-7»

\LL-MI*M| /

GROUP SYMBOL GROUP NAME

Figure 3FLOWCHART FORCLASSIFICATION OFFINE-GRAINED SOIL

GROUP SYMBOL GROUP NAME

MAVIL

IAMO

Figure 4FLOWCHART FOR CLASSIFICATIONOF COARSE-GRAINED SOIL feTCIHL'ff

soil types in a bedded deposit, can be obtained only in thefield. Corrections and additions to the field classifica-tion can be provided, when necessary, by laboratory testingof the soil samples.

Soil descriptions should be precise and comprehensive with-out being verbose. The correct overall impression of thesoil should not be distorted by excessive emphasis on in-signiticant details. In general, similarities between con-secutive samples should be stressed rather than differences.

Soil descriptions shall be recorded in the Soil Descriptioncolumn for every soil sample collected. The format and orderfor soil descriptions should be:

1. Soil name (synonymous with ASTM D 2487-85 GroupName) with appropriate modifiers

2. Color

3. Moisture content

4. Relative density or consistency

5. Soil structure or mineralogy

6. Group symbol

This order follows, in general, the format described in ASTMD 2488-84. Examples of soil descriptions are provided inTable 1.

Soil Name

The basic name of a soil shall be identical to the ASTMD 2487-85 Group Name based on visual estimates of gradationand plasticity. The soil name should be capitalized. Theonly acceptable soil names are those listed in Figures 3 and4, which are trom ASTM D 2487-85.

Examples of acceptable soil names are illustrated by thefollowing:

A soil sample is visually estimated to contain 15percent gravel, 5b percent sand, 30 percent fines(passing No. 200 sieve). The tines are estimated aseither low or highly plastic silt. This sample isSILTY SAND WITH GRAVEL, with a Group Symbol of (SM).

Another soil sample has the following visual estimate:10 percent gravel, 30 percent sand, and 60 percentfines (passing the No. 200 sieve). The fines areestimated as low plastic silt. This sample is SANDYSILT. The gravel portion is not included in the soil

PD253.016.9

Table 1EXAMPLE SOIL DESCRIPTIONS

POORLY GRADED SAND, fine, well-rounded, light brown, moist, loose (SP)

FAT CLAY, dark gray, moist, stiff (CH)

SILT, light greenish gray, wet, very loose, some mica, lacustrine (ML)

WELL-GRADED SAND WITH GRAVEL, subangular gravel to 0.6 inches max,reddish brown, moist, dense, (SW)

POORLY GRADED SAND WITH SILT, white, wet, medium dense (SP-SM)

ORGANIC SILT WITH SAND, dark brown to black, wet, firm to stiff butspongy undisturbed, becomes soft and sticky when remolded, many fineroots, trace of mica (OH)

SILTY GRAVEL WITH SAND, subrounded gravel to 1.2 inches max, brownishred, moist, very dense, (GM)

INTERLAYERED SILT (60 percent) AND CLAY (40 percent): SILT WITH SAND,nonplastic, sudden reaction to shaking, medium greenish gray, layersmostly 1.5 to 8.3 inches thick; CLAY, dark gray, firm and brittleundisturbed, becomes very soft and sticky when remolded, layers 0.2 to1.2 inches thick (ML and CH}

SILTY SAND WITH GRAVEL, weak gravel to 1.0 inches max, light yellowishbrown, compact, moist, very few small particles of coal, fill (SM)

SANDY ELASTIC SILT, very light gray to white, wet, stiff, weakcalcareous cementation, (MH)

SILTY CLAY WITH SAND, dark brownish gray, moist, stiff (ML-CL)

WELL-GRADED GRAVEL WITH SILT, rounded gravel to 1.0 inches max, brown,moist, very dense (GW-GM)

PDP253.021.1 10

name because the gravel portion was estimated as lessthan 15 percent. The Group Symbol is (ML).

The gradation of coarse-grained soil (more than 50 percentretained on No. 200 sieve) is included in the specific soilname in accordance with ASTM D 2487-85. There is no need tofurther document the gradation. However, the maximum sizeand angularity or roundedness of gravel and sand-sizedparticles should be recorded. For fine-grained soil (50percent or more passing the No. 200 sieve), the name ismodified by the appropriate plasticity/elasticity term inaccordance with ASTM D 2487-85.

Interlayered soil should each be described starting with thepredominant type. An introductory name should be used suchas "Interlayered Sand and Silt." Also, the relativeproportion of each soil type should be indicated (seeTable 1 for example).

Where helpful, the evaluation of plasticity/elasticity canbe justified by describing results from any of the visual-manual procedures for identitying fine-grained soils, suchas reaction to shaking, toughness of a soil thread, or drystrength as described in ASTM D 2488*84.

Color

The basic color of a soil, such as brown, gray, or red,shall be given. The color term can be modified, if neces-sary, by adjectives such as light, dark, or mottled. Thecolor description should be kept simple and should notemphasize unimportant color aspects or shades.

Moisture Content

The degree of moisture present in a soil sample should bedefined as dry, moist, or wet. Moisture content can beestimated as follows:

Dry Requires addition of considerable moisture toobtain optimum moisture content

Moist Near optimum moisture content

Wet Requires drying to obtain optimum moisturecontent

Relative Density or Consistency

Relative density of a coarse-grained (cohesionless) soil isbased on N-values (ASTM D 1586-84). If the presence oflarge gravel or disturbance of the sample makes determina-tion of the in situ relative density or consistency

PD253.016.10 11

difficult, then this item should be left out of the descrip-tion and explained in the Comments column of the soil boringlog.

Consistency of fine-grained (cohesive) soil is properlybased on results of pocket penetrometer or torvane results.In the absence of this information, consistency can be esti-mated from N-values. Relationships for determining relativedensity or consistency of soil samples are given in Tables 2and 3.

Soil Structure or Mineralogy

Discontinuities and inclusions are important and should bedescribed. Such features include joints or fissures, slick-ensides, bedding or laminations, veins, root holes, and wooddebris.

Significant mineralogical information should be noted. Ce-mentation, abundant mica, or unusual mineralogy should bedescribed, as well as other information such as organicdebris or odor.

Residual soils have characteristics of both rock and soiland can be difficult to classify. Relict rock structureshould be described and the parent rock identified if pos-sible.

Group Symbol

Each soil description is concluded with the appropriategroup symbol trom ASTM D 2487-85 (see Figures 3 and 4). Thegroup symbol should be placed in parentheses at the end ofthe description to indicate that the classification has beenestimated.

In accordance with ASTM D 2488-84, dual symbols (e.g., GP-GMor SW-SC) can be used to indicate that a soil is estimatedto have between 5 and 12 percent fines. Borderline symbols(e.g., GM/SM or SW/SP) can be used to indicate that a soilsample has been identified as having properties that do notdistinctly place the soil into a specific group. Generally,the group name assigned to a soil with a borderline symbolshould be the group name for the first symbol. The use of aborderline symbol should not be used indiscriminately. Everyeffort should be made to first place the soil into a singlegroup.

STANDARD PENETRATION TEST PROCEDURES

Standard Penetration Tests (SPT) are conducted to obtain ameasure of the resistance of the soil to penetration of thesampler and to recover a disturbed soil sample. StandardPenetration Tests should be conducted in accordance withASTM D 1586-84, Penetration Test and Split Barrel Samplingof Soils.

PD253.016.12 12

Table 2RELATIVE DENSITY OF COARSE-GRAINED SOIL t-

(Devcloped from Sowers, 1979)

RelativeBlows/Ft Density ____________Field Test_____________

0-4 Very loose Easily penetrated with S-in. steel rod pushedby hand

5-10 Loose Easily penetrated with h-in. steel rod pushedby hand

11-30 Medium Easily penetrated with h-in. steel rod drivenwith 5-lb. hammer

31-50 Dense Penetrated a foot with >i-in. steel rod drivenwith 5-lb. hammer

>50 Very dense Penetrated only a few inches with h-in. steelrod driven with 5-lb. hammer

Table 3CONSISTENCY OF FINE-GRAIHED SOIL

(Developed from Sowers, 1979)

PocketPenetrometer Torvane

Blows/Ft Consistency (TSF) (TSF) _____Field Test_______

<2 Very soft <0.25 <0.12 Easily penetrated severalinches by fist

2-4 Soft 0.25-0.50 0.12-0.25 Easily penetrated severalinches by thumb

5-8 Firm 0.50-1.0 0.25-0.5 Can be penetrated severalinches by thumb with mod-erate effort

9-15 Stiff 1.0-2.0 0.5-1.0 Readily indented by thumbbut penetrated only withgreat effort

16-30 Very stift 2.0-4.0 1.0-2.0 Readily indented by thumb-nail

>30 Hard >4.0 >2.0 Indented with difficulty bythumbnail

PD253.022 13

Equipment and Calibration

Before starting the testing, the necessary equipment shouldbe inspected for compliance with the requirements of ASTMD 1586-84. The split-barrel sampler should measure 2 inchesO.D., with 1-3/8 inches I.D., and should have a split tubeat least 18 inches long. The dimensions should conform withthose indicated on Figure 1 of ASTM D 1586-84. The minimumsize sampler rod allowed is "A" rod (1-5/8 inches O.D.). Astiffer rod, such as "N" rod (2-5/8 inches O.D.), isrequired for depths greater than 50 feet. The drive weightassembly should consist of a 140-pound hammer weight, adrive head, and a hammer guide that permits a tree fall of30 inches.

Procedures

Standard Penetration Tests should be conducted at everychange of strata or, within a continuous stratum, at inter-vals not exceeding 5 feet. Betore driving the split-barrelsampler, all loose and foreign material should be removedfrom the bottom of the borehole. It may be helpful to mea-sure the rod "stickup" to ensure that the sampler is beingdriven from the bottom of the borehole. The Standard Pene-tration Test should be performed by driving a standardsplit-barrel sampler 18 inches into undisturbed soil at thebottom of the borehole by a 140-pound guided hammer or ram,falling freely from a height of 30 inches.

The number of blows required to drive the sampler for three6-inch intervals, for a total of 18 inches, should be ob-served and recorded on the soil boring log. The sum of thenumber of blows required to drive the sampler the second andthird 6-inch intervals is considered the Standard Penetra-tion Resistance (N) or the "blow-count." It the sampler isdriven less than 18 inches, but more than 1 foot, thepenetration resistance (N) is that for the last 1 foot ofpenetration. If less than 1 foot is penetrated, the logshall state the number of blows and the traction of 1 footpenetrated. If possible, the field logger should observethe sampler being driven and count the blows for each sampleattempt.

General Considerations

The following comments and suggestions should be consideredwhen performing Standard Penetration Testing:

1. The borehole should be cleaned out before every sampleattempt. Because a minor amount ot caving can beexpected, the borehole can be considered to be ade-quately cleaned if no more than 4 inches of loose orforeign material has collected at the bottom of theborehole. A greater amount of caving is sufficientcause to require the hole to be cleaned again.

PD253.016.14 14

2. The ball check valve in the split-barrel sampler shouldbe cleaned and working properly for each sample. Bent,chipped, or damaged shoes should be replaced. The split'barrel halves should not be warped. In case of zerosample recovery (i.e., if the sample is lost duringfirst attempt), a spring catcher should be used duringsubsequent attempts to facilitate recovery.

3. During SPT sampling, it is important that all rodconnections be tight and that the hammer guide be con-nected securely to the drill rods. If the hammer guideconnection becomes loose, much of the hammer energy maybe lost because of deflection of the hammer coupling.The lifting rope should not rub against the mast. Eachhammer fall should be 30 inches.

4. During SPT sampling, it is important that the drillrods be positioned at the center of the drill hole.This is necessary to preclude the development of fric-tion between drill rods and the walls of the boreholeor casing.

5. If the hammer weight is raised by means ot a cathead,generally two wraps on the cathead should be used. Theoptimum number of wraps will vary with the condition ofthe rope and cathead and the weather. Most impor-tantly, the driller should exercise care to preventfriction of the rope on the cathead during the fall ofthe hammer.

6. Occasionally, nonstandard procedures or equipment areused for obtaining samples (such as 3-inch O.D. split-barrel samplers, or 300-pound hammers). Any nonstan-dard practice should be described on the boring logform. Also, the blow counts should be clearly markedas not conforming to SPT values.

SAMPLE LABELING AND PACKAGING

The samples recovered from the borehole are an importantpart of the boring record and must be properly packagedand labeled. Samples that are improperly or inadequatelylabeled are useless. The following description outlines theminimum requirements for packaging and labeling of samples.

Disturbed samples should be placed in jars that are markedboth on the jar lid and on a label on the side of the jar.Standard CH2M HILL jar labels are available (Form No. 131,Soil Sample Labels) for this purpose. The following infor-mation should be clearly marked on the jars: job number,boring number, sample number/ sample depth, blow counts,sample recovery, and date. Use an indelible marker or ametal scriber on the jar lid. If moisture content tests areanticipated, jar samples should be tightly sealed, then sent

PD253.016.15 15

to the laboratory and the testing initiated as soon aspossible (within one week). See Figure 5 for labelingdetails.

Boxes containing the jars should be labeled on top and onone end with the following information: job name, jobnumber, boring number, sample numbers and sample depths,date, and name. It is helpful to start a new box for eachnew boring if the boxes are at least one-half full.

Shelby tubes should be cleaned of mud and moisture. Whendry, use an indelible marker to label them with the follow-ing information: an arrow indicating which way is up, }obnumber, boring number, sample number and depth, amount ofrecovery, and date. The top and bottom of the sample can becircumscribed on the outside of the tube with a marker.

Waxing of Shelby tubes is essential if sample testing willnot occur within a few days. In all cases, lids should beplaced on the ends and taped with airtight tape. Makecertain that the holes in the top of the tube are sealed.The open portion of the tube, above the sample, should bepacked to prevent shifting of the soil. Dampened newspaperis generally adequate for this purpose, but it should beseparated from the soil sample by a wax seal or an invertedcap. See Figure 5 for labeling and packing details.

FIELD EQUIPMENT CHECKLIST

Table 4 lists equipment and supplies that are necessary oruseful for soil boring exploration.

PD253.016.16 16

END CAPS TAPEDAND WAXED

1- TO 2- WAX SEALOR INVERTED CAP

PACKING DETAILDAMPENED ANDCRUMPLED NEWSPAPER

OMICE

PROJECT NO

TM R'/OI

JAR SAMPLES

SAMPLE

DEPTH;o.o'-/ / .s '

BLOWS

INCHES

76

/o6

19

(p

FORM *131

SIDE LABEL

LID

Figure 5SAMPLE LABELING DETAILS

Table 4FIELD EQUIPMENT CHECKLIST FOR SOIL BORING LOGGING

Siting

__Lath, flagging, and orange spray paint__Lumber crayon__100-foot tape__Brunton or Silva compass

Logging Equipment

__Soil Boring Guideline__Clipboard__Form D1586 on all-weather paper__Pens/pencils__Engineer's pocket tape measure with tape lock__Field notebook on all-weather paper__Squirt bottle with water__Spatula__HCL, 10 percent solution

Sampling and Packaging

__Jars with lids and labels (Form #131)__Shelby tubes and plastic end caps__Airtight tape (e.g., electrical)__Newspaper__Wax, stove, melting pot, and matches__Indelible fine felt-tipped markers (e.g., "Sharpie

brand)

Test Equipment

__Pocket penetrometer__Torvane__Well sounder

Other

__Camera, film__Hand lens__Rags__Ear protectors__Screwdrivers__Hard hat__Sunscreen__Insect repellent

PD253.016.18 18

SELECTED REFERENCES

American Society for Testing and Material. 1986. ASTM D1586. Standard Method for Penetration Test and Split-Barrel Sampling of Soils. Annual Book of ASTM Stan-dards, Section 4, Vol. 4.08, pp. 298-303.

American Society for Testing and Material. 1986. ASTM D1587. Standard Method for Thin-Walled Tube Sampling ofSoils. Annual Book of ASTM Standards, Section 4,Vol. 4.08, pp. 304-307.

American Society for Testing and Material. 1986. ASTM D2487. Standard Test Method for Classification of Soilsfor Engineering Purposes. Annual Book of ASTM Stan-dards, Section 4, Vol. 4.08, pp. 397-410.

American Society for Testing and Material. 1986. ASTM D2488. Standard Practice for Description of Soils(Visual-Manual Procedure). Annual Book of ASTM Stan-dards, Section 4, Vol. 4.08, pp. 411-425.

American Society of Civil Engineers. 1976. SubsurfaceInvestigation for Design and Construction of Founda-tions of Buildings. American Society of Civil Engi-neers Manual on Engineering Practice, No. 56, 53 pp.

Bell, F. G.Rocks.

1981. Engineering Properties of Soils andLondon: Butterworth Publishers, Inc.

Burmister, D. M. 1949. Principles and Techniques of SoilIdentification. Proceedings of the Highway ResearchBoard, pp. 402-433.

Casagrande, A. 1947. Classification and Identification ofSoils. American Society of Civil Engineers Transac-tions, pp. 901-991.

Kovacs, W. D., L. A. Salomone, and F. Y. Yokel. 1981.Energy Measurement in the Standard Penetration Test.U.S. Dept. of Commerce, National Bureau of Standards,Building Science Series 135.

Matula, M. 1981. Rock and Soil Description and Classifica-tion tor Engineering Geological Mapping, Report by theIAEG Commission on Engineering Geological Mapping.Bulletin of the International Association of Engineer-ing Geology, No. 24, pp. 235-274.

Sowers, G. F. 1979. Introductory Soil Mechanics and Foun-dations: Geotechnical Engineering. MacMillan Publish-ing Co., New York, 4th edition. 621 pp.

U.S. Bureau of Reclamation. 1974. Earth Manual. 2nd ed.Washington, D*C.: U.S. Government Printing Office.

PD253.016.19 19

Appendix cFIELD MEASUREMENT OF pH

IJ

IJ INSTRUCTION MANUAL

model 2111 digital pH meter

| ORION RESEARCH11

1

contentsIntroduction 3Instrument description 3instrument set-up 4

support rodpower sourcemeter check-outconnecting electrodes

measurement procedures 6general measurement techniquepH measurement

single buffer standardizationtwo-buffer standardization

battery replacement 8recorder output 8repair and service 8accessories 9specifications 10notice of compliance 11

repair/serviceFor Inlormallon on repair or replacement ol Ihis instrument, contact OrionResearch toll-free. Ask lor Customer Service.

ORION RESEARCH INCORPORATEDCustomer Service840 Memorial DriveCambridge, Massachusetts 02139 U.S.A.800-225-1480 (Continental U.S.)617 864-5400 (Massachusetts. Alaska, Hawaii. Canada)Telex: 921466

CtMl Odon B**wch Inctxpofaiw)OAION • U • i •el*l*(«d lrad*m*ili ol Otion Retaarch lnc<Mpo4*l«d

fo(inlM2M'3«60PitnledtnUSA

instrument set-upsupport rod

1. Insert steel support rod Into the hole in Ihe support rod clip on side ol themeter.

2. Mount electrode holder on the rod by pinching to compress the spring.Release to hold In place.

power sourceThe Model 211 operates on six nonrechargeable 1.5 voll batteries or on 110 or220 i 20% Vwith an approved AC adapter (specify voltage when ordering). Lowballery is Indicated by Ihe BAT LOW Indicator on Ihe display.NOTE: Batleries are not rechargeable - use of line adapter whenever possiblewill prevent Ihe anil's batteries Irom being discharged. II ballery operation isdesired, follow installation instructions under ballery replacement.

meter check-outi.

2.

Install six AA batteries in Ihe meter. Orient the (+) and (-) battery termi-nals to match the orientation shown In the battery compartment.Depress ON button on the front panel. If Ihe BAT. LOW Indicator on Ihe(root display lights up, the batteries must be replaced.II battery mode Is not to be used, disregard steps 1 and 2. Insert pin end ofappropriate AC line adapter Into the meter, and Ihe olher end into the ap-propriate grounded AC line receptacle.Attach BNC shorting plug to BNC input on Ihe bottom side of Ihe meter.Depress ON button on Ihe front panel. Turn CALIB knob so display readsa steady 7.00. If this cannot be done consult ORION Technical Service.Remove Ihe snorting plug. Successful completion ol steps 1-4 show the

' " I t leady tor use.

connecting electrode1. Insert the BNC connector into the electrode jack on the bottom panel ol Ihe

meter. Turn connector clockwise until it seats firmly.2. Mount electrode in the electrode holder by spreading Ihe electrode clip

open and sliding Ihe electrode into the holder so lhat Ihe clip closes onelectrode cap. See figure 2.

3. Follow measurement procedures Io use Ihe meter to measure pH.4. Disconnnect electrode by turning connector counterclockwise until releas

ed from pin.

figure 2

figure 1

it SATlow

ORION RESEARCH modal 211/dlgllal pHmeltr

ON

OFF

torn

CALIB

20

O

l«g«nd1. strip chart recorder binding posts1. DAT LOWi. LC display4. support lod clipft. towptnlun Indicator control

6. AC line adapter input7. I unction control8 calibration control9. electrode connector

10. slope control

introductionThe Model 211 is a battery- or line-operated (110/220 V AC adapter) digital phmeter lor Held or laboratory use. The meter is complete with strip chart recoi derbinding posts and is supplied with an unbreakable, gel filled combination pHelectrode, one packet ol pH 7 bullet powder, one bollle lor pH 7 buller. one bolllefor distilled water, support rod, electrode holder. AC adapter, six 1.5 V ballenes.shorting plug, and carrying case.

instrument descriptionSee figure 1.

1. strip chart recorder binding posts: black post is low (ground) and red post ishigh Input side of recorder. See page 0.

2. BAT LOW: an arrow pointing lowards BAT LOW appears on the displaywhen battery requires replacement.

3. LC display: pH display over the range of 0 - 14 with ± .01 pH uniis resolution4. support rod clip: holds steel rod used lo mount electrode holder.5. temperature Indicator control (TEMP *C): compensates lor variation m

electrode slope or temperature changes. Used in two buller calibration6. AC Una adapter Input: jack used lo Insert AC line adapter- With AC line

adapter operational, the internal battery is bypassed.(unction control: rocker switch with three positions - ON. OFF and (ON)Depress (ON) for a momentary reading. The switch will return lo OFF whenreleased.calibration control (CALIB): used lo calibrate the meter with buffers ofknown pH.electrode connaclor: accepts 8NC connector Irom pH electrode.

7.

a.

9.10. slope control: screwdriver adjustment used to set second butler In two

buffer calibration.

measurement procedures

general measurement techniquetemperature: All samples and buffers should be at the same temperature, assmall variations In temperature can cause errors In measurement. The slope ofthe pH electrode, the potential of the reference electrode, and the pH of the buf-fer are temperature-dependent.cleaning electrodes: Electrode should be rinsed and shaken between measure-ments to remove drops and to prevent solution carryover.stirring: Stir measured solutions moderately to obtain good contact betweenthe glass bulb and the solution. Insert electrode to a depth ol about 3 cm.

pH measurementssingle-butter standardization(where maximum precision is not required)NOTE: For maximum accuracy It is recommended that a Iwo-bulfer calibrationbe performed once at the beginning of each day (see page 7), This procedure en-sures the correct setting of the slope control. Subsequent measurements duringihe day may be made using a single point calibration.

1. Place the electrode In a butler solution whose pH Is near Ihe expected pH ofthe sample. Insert electrode to a depth ol about 3 cm and stir moderately.

2. Set Ihe temperature Indicator control to the temperature of the butter.3. Set the function control to ON and allow the bulfer reading to stabilize. Ad-

just the CALIB so that Ihe display indicates Ihe pH of the buffer at the solution temperature. See Table 1.

4. Remove the electrode from the buller solution and rinse by stirringmoderately in distilled water. Shake off excess drops ol waler.

5. Place electrode in the sample lo a depth ol about 3 cm and stir moderately.Set the function control lo ON and allow Ihe reading lo stabilize Record Ihesteady pH leading.

two-buffer standardization(where maximum precision is required)

1. Select two buffers to bracket Ihe expected pHol the sample, with one bulfer having a pH ol 7.

2. Place Ihe electrode in the pH 7 bulfer lo a depth of about 3cm and stirmoderately. Set Ihe temperature Indicator control to the temperature ol thebutler. Set Ihe function control lo ON and allow the reading lo stabilizeTurn CALIB until ihe display indicates Ihe pH ol Ihe buller at Ihe solutiontemperature. See table 1.

3. Remove electrode from Ihe lirsl buller and rinse by stirring moderately indistilled water. Shake off excess drops of waler.

4. Place the electrode In the second buller lo a depth ol about 3 cm and slirmoderately. Set the function control lo ON and adjust the slope control un-til Ihe pH at Ihe solution temperature is displayed. See Table 1.

5. Remove Ihe electrode and rinse by stirring moderately in distilled waterShake off excess drops ol waler.

6. Place Ihe electrode in Ihe sample lo a depth of about 3 cm and slirmoderately. Set Ihe function control to ON and allow the reading tostabilize. Record Ihe steady pH reading.

TEMP (*C)S

101520253035405060

TABLE 1

pH 7.00 Buffer pH 47087.067.037017.006.98696697657698

.01 Buffer4.004.004004004014024024034.08409

pH 10.01 Buffer1025101810.1210081001 •9.97 •9.93969983--

•i

*•;-•'

battery replacementTo replace the baMerles, remove Ihe panel on the back of Ihe meter. Be sure toobserve Ine polarity marking when Inserting new batteries.

recorder outputThe red and black binding posts at the side ol Ihe meter provide an output lorstrip chart recording of absolute mV Independent ol function mode. Forrecorders with Input Impedance ol 100 Kllohms or greater, Ihe output Is fixed toabout 100 mV/pH. pH 14.00 output Is 1.40 V. Lower Impedance recorders may beused but lull-scale output Is reduced.

1. Connect Ihe lead from Ihe high (Input aide of the recorder) to the red bindingpost and the lead Irom Ihe low (ground) side to Ihe black binding post.

2. Proceed according lo directions In Ihe strip chart recorder Instructionmanual.

repair and serviceORION warranty covers failures due lo manufacturer's workmanship or material•lelect Irom Ihe dale of purchase by Ihe user. User should return Ihe warrantycard to ORION and retain proof of purchase. Warranty is void II product has beenabused, misused, or repairs attempted by unauthorized persons.Warranties herein are lor products sold/installed lor use only In the United Stalesand Canada. For ORION producis purchased tor use in all other countries consultlocal in-country, authorized ORION sales agent/distributor for product warrantyinformation.A Return Authorization Number must be obtained Irom ORION Laboratory Pro-ducis Customer Service before returning any product lor in-warranty repair,replacement or credit.

"No Lemon" Instrument WarrantyThe instrument is covered by Ihe ORION "No Lemon" warranty. II Ihe instrument(alls within twelve months Irom dale ol purchase lor any reason other thanabuse. Ihe purchaser may elect to have it repaired or replaced al no charge. Thiswsiianly covers the original or replacement/repaired meler from date of originalmeter purchase; the warranty is not extended beyond Ihe buyer's original war-ranty dale.

accessories

815600 Rosslu epoxy body, bulb guard combination pH electrode91CMBN Laboratory grade combination pH electrode (BNC connector)910600 GX-series epoxy body, gel-filled combination electrode (BNC

connector)912600 GX-series epoxy body, gel filled flask combination electrode

(BNC connector)913600 GX-series epoxy body, gel-filled Hal surface combination pH

electrode (BNC connector)915600 RX series refutable, epoxy body combination pH electrode (BNC

connector)9162BN Combination pH electrode wilh rugged bulb (BNC connector)9163BN Combination pH electrode with needle shape (BNC connector)910004 pH 4 buffer packets, box of 25 packets, each packet making

200 ml of buffer910007 pH 7 buffer packets, box ol 25 packets, each packet making

200 ml of butler910009 pH 9 buffer packets, box ol 25 packets, each packel making

200 ml of buffer910104 pH 4.01 buffer. 475 ml bottle910107 pH 7.00 buller. 475 ml bo I He910110 pH 10.01 buller. 475 ml bottle970699 Dissolved oxygen electrode910002 Electrode holder020030 Shorling plug020120 110V AC line adapter020121 220V AC line adapter

specifications

packagecontents

rang*resolution

temperaturecompensation

Isopotentlalpoint

powerrequirement

dimensionsweight

model 211 digital pH meter, with model 910600 gel (tiledunbreakable combination pH electrode, support rod, elec-trode holder, bottles lor pH 7 bufler and distilled water, onepacket pH 7 buffer powder. AC adapter, six 1.5 V batteries,and carrying caseOtoHpH± .01 pH

manual (0 lo 100'C)

pH 7 (fixed)

sin 1.5V batteries;battery Hie: 3000 len second inter mil lent measurementswhen line adapter Is not used.Hne adapter 110 or 220V ±20%. 50160 Hz14 cm high x 9 cm wide x 4.5 cm deep0.4kg

•pocHlcalloin lutojicl lo riming wUfcoul nolle*

notice of compliance

The Model 211 may generate radio frequency energy and il not installed andused properly, that is, in strict accordance wilh the manufacturer's Inslruclions, may cause interference to radio and television reception. It has beentype tested and found lo comply wilh the limils toi a Class B computing devicein accordance with specifications in Subpart J of. Part 15 ol FCC Rules, whichare designed to provide reasonable protection against such interference in aresidential installation. However, there Is no guarantee that interference willnot occur in a particular installation. If the Model 211 does cause interferenceto radio or television reception, which can be determined by turning the unitoff and on. the user is encouraged to try to correct the interference by one ormore ol the following measures:- reorient the receiving antenna- relocate the Model 211 with respect to the receiver- move the Model 211 away from the receiver- plug the Model 211 into a different outlet so that the meter and receiver a<e

on different branch circuitsII necessary, the user should consult the dealer or an experienced radio/lelevision technician tor additional suggestions. The user may find I he followingbooklet prepared by the Federal Communications Commission helpful:"How to Identify and Resolve Radio-TV Interference Problems"This booklet Is available from the U.S. Government Printing Oil ice, Washington,DC 20402, Slock No. 004000X10345-4.

11

Appendix DFIELD MEASUREMENT OF SPECIFICCONDUCTANCE AND TEMPERATURE

Target Compound List (TCL) andContract Required Quantitation Limits (CRQL) *

Quantltation Limits**Water Low Soil/Sediment*

1.2.3.4.5.

6.7.8.9.

lu .

11.12.13.14.15.

16.17.18.19.20.

21.22.23.24.

25.

26.27.28.29.30.

Volatl les

ChloromethaneBr OB ome thaneVinyl ChlorideChl ore* thaneMethylene Chloride

AcetoneCarbon Disu l f idei , 1-Dichloroethene1 , 1 -Dichl or o« thane1 ,2-Dichloroethene ( to ta l )

Chloroform1 ,2-Dichloro*thane2-Butanone1,1, 1-TrichloroethaneCarbon Tetrachloride

Vinyl AcetateBromodi chl or one thane1 ,2-Dichloropropanecis-1 , 3-DichloropropeneTrichloroethene

Dibronochl or one thane1 , 1 , 2-Tr i chl or oe thaneBenzenetrans-1 ,3-Dichloropropene

Bromof oro

4-Methyl-2-pentanone2-BexanoneTetrachloroetheneToluene1, 1,2,2-Tetrachloroethane

CAS Number

74-87-374-83-975-01-475-00-375-09-2

67-64-175-15-U75-35-475-34-3

540-59-0

67-66-3107-06-278-93-371-55-656-23-5

108-05-475-27-478-87-5

10061-01-579-01-6

124-48-179-00-571-43-2

10061-02-675-25-2

108-10-1591-78-6127-18-4108-88-379-34-5

ua/L

101010105

105555

55

1055

105555

5555

5

1010555

U K / K g

101010105

105555

55

1055

105555

5555

5

1010

555

(continued)

C-l 7/87 Rev,

Quantitation Limits**__ Lot* Soil/Sediment*

Volatiles______________CAS Number "ue/L ___ u£/Ks

31. Chlorobenzene 108-90-7 5 532. Ethyl Benzene 100-41-4 5 533. Styrene 100-42-5 5 534. Xylenes (Total) 1330-20-7 5 5

•Medium Soil/Sedlaent Contract Required Quantitation Limits (CRQL) for VolatileTCL Compounds are 125 times the individual Low Soil/Sediment CRQL.

•Specific quantitation Halts are highly matrix dependent. The quantisationlimits listed herein are provided for guidance and may not always be achievable

"Quantitation limits listed for toll/sediment are based on wet weight. Thequantitation limits calculated by the laboratory for soil/sediment, calculatedon dry weight basis as required by the contract, will be higher.

C-2 7/87 Rev,

Contract Requlrtd Quantitation Limits (CRQL)*

Quanrltatlon Limits**

35.36.37.38.39.

40.41.42.43.

44.

45.

46.47.48.49.

50.51.52.

5354.

55.56.57.58.

59.

60.61.62.63.64.

Semivolatiles

Phenolbis(2-Chloroethyl) ether2-Chlorophenol1, 3-Dichlorobenzene1 , 4-D1 chl or obenzene

Benzyl alcohol1,2-Dl chl or obenzene2-Methylphenolbis(2-Chlorolsopropyl)ether

4-Methylphenol

N-Nitroso-dl-n-dlpropylamine

Hexa chl or oe thaneNitrobenzeneIsophorone2-Nitrophenol

2 , 4-DlmethylphenolBenzole acidbis C 2-Chl oroe thoxy )methane

2,4-Dichlorophenol1 ,2,4-Trichlorobenzene

Naphthalene4-ChloroanillneHexachlorobutadiene4-Chloro-3-methylphenol

(para-chloro-meta-cresol)2-Methylnaphthalene

Hexachlorocyclopentadlene2,4 , 6-Trl chl orophenol2,4, 5-Tri chl orophenol2-Chloronaphthalene2-Nitroanlllne

CAS Number

108-95-2111-44-495-57-8

5*1-73-1106-46-7

100-51-695-50-195-48-7

108-60-1106-44-5

621-64-767-72-198-95-378-59-188-75-5

105-67-965-85-0

111-91-1120-83-2120-82-1

91-20-3106-47-887-68-3

59-50-791-57-6

77-47-488-06-295-95-491-58-788-74-4

Waterug/L

1010101010

101010

1010

1010101010

1050

101010

101010

1010

1010501050

Low Soil/Sedimentb

ui/Kg

330330330330330

330330330

330330

330330330330330

3301600

330330330

330330330

330330

330330

1600330

1600

(continued)

C-3 1/87 Rev,

Quantltatlon Lialts**

65.66.67.68.69.

70.71.72.73.74.

75.

76.77.78.79.

80.61.82.83.84.

85.66.87.86.89.

90.91.92.93.94.

Scmlvolatiles

Dime thy IphchalateAcenaphthylene2,6-Dlnltrotoluene3-NltroanilineAcenaphchcne

2,4-Dinitrophenol4-Nitrophenol.Dibenzofuran2,4-DinitrotolueneDiethylphthala te

4-Chlorophenyl-ph«nylether

Fluorene4-Nlt roani l ine4, 6-Dlni tro-2-ne thy 1 phenolN-nitrosodiphenylamine

4-Bromophenyl-phenyl etherUexachlorobenzenePentachlorophenolPhenanthreneAnthracene

Di-n-butylph thai ateFluoranthenePyreneButyl be nzylphtha l ate3,3'-Dichlorobenzldine

Benzo(a)anthraceneChrysenebls(2-Ethylhexyl)phthalateDl-n-octy IphchalateBenzo(b)f luoranthene

CAS Number

131-11-3208-96-8606-20-299-09-283-32-9

51-28-5100-02-7132-64-9121-14-284-66-2

7005-72-386-73-7

100-01-6534-52-186-30-6

101-55-3118-74-187-86-585-01-8

120-12-7

84-74-2206-44-0129-00-085-68-791-94-1

56-55-3218-01-9117-81-7117-84-0205-99-2

Waterug/L

1010105010

5050101010

1010505010

1010501010

1010101020

1010101010

Low Son/Sediment^ui/K*

330330330

1600330

16001600

330330330

330330

16001600330

330330

1600330330

330330330330660

330330330330330

(continued)

C-4 1/87 Rev,

Quantltatlon Llmlta**____ Low Soil/Sediment^

Semlvolatiles GAS Number

95. BenzoCtOfluoranthene 207-08-9 10 33096. B«nzo(a)pyr*ne 50-32-8 10 33097. Indeno(l,2,3-cd)pyrene 193-39-5 10 33098. Dlbenz(a,h)anthracene 53-70-3 10 33099. Benzo(g,h,l)perylene 191-24-2 10 330

^Medium Soll/Sedlaent Contract Required Quantltatlon Llmlti (CRQL) for Semi-Volatile TCL Compounds are 60 tlmei the individual Low Soil/Sediment CRQL.

•Specific quantltatlon limit* are highly matrix dependent. The quant1tationlimits listed herein are provided for guidance and may not always be achievable

"Quantitation limits lifted for toll/sediment are based on wet weight. Thequantitation limits calculated by the laboratory for aoil/sediment, calculatedon dry weight basis as required by the contract, will be higher.

C-5 10/86

Target Compound List (TCP andContract Required Quantitatlon Limits (CRQL)*

Quantltation Limits**

100.101.102.103.104.

105.106.107.108.109.

110.111.112.113.114.

115.116.117.118.119.

120.121.122.123.124.

125.126.

Pesticides/PCBs

alpha-BHCbeta-BHCdelta-BHCgamma-BHC (Lindane)Heptachlor

AldrlnHeptachlor epoxideEndosulfan IDieldrin4, 4 '-DDE

EndrinEndosulfan 114, 4 '-ODDEndosulfan sulface4, 4 '-DOT

MethoxychlorEndrin keconea Ipha-Chl ordanegamma-Chl ordaneToxaphene

Aroclor-1016Aroclor-1221Aroclor-1232Aroclor-1242Aroclor-1248

Aroclor-1254Aroclor-1260

CAS Number

319-84-6319-85-7319-86-858-89-976-44-8

309-00-21024-57-3959-98-860-57-172-55-9

72-20-833213-65-9

72-54-81031-07-850-29-3

72-43-553494-70-55103-71-95103-74-28001-35-2

12674-11-211104-28-211141-16-553469-21-912672-29-6

11097-69-111096-82-5

Waterug/L

0.050.050.050.050.05

0.050.050.050.100.10

0.100.100.100.100.10

0.50.100.50.51.0

0.50.50.50.50.5

1.01.0

Low Soil/SedimentcU£/K£

8.08.08.08.08.0

8.08.08.016.016.0

16.016.016.016.016.0

80.016.080.080.0160.0

80.080.080.080.080.0

160.0160.0

CMedium Soil/Sediment Contract Required Quantltation Limits (CRQL) for Pesticide/PCBTCL compounds are 15 times the Individual Low Soil/Sediment CRQL.

•Specific quantitatlon limits are highly matrix dependent. The quantitationlimits listed herein are provided for guidance and may not always beachievable.

"Quantltation limits listed for Boll/sediment are based on vet weight. The quan-titatlon Limits calculated by th* laboratory for soil/sediment, calculated on dryweight basis as required by th* contract, will be higher.

C-6 1/87 Rev.

INORGANIC TARGET ANALYTE LIST (TAL)

Analyte

Contract RequiredDetection Limit

(ug/L)

AluminumAntimonyArsenicBariumBerylliumCadmiumCalciumChromiumCobaltCopperIronLeadMagnesiumManganeseMercuryNickelPotassiumSeleniumSilverSodiumThalliumVanadiumZincCyanide

2006010200

55

50001050251005

5000150.240

5000510

500010502010

(1) Subject to the restrictions specified in the first page of Part G,Section IV of Exhibit D (Alternate Methods - Catastrophic Failure) anyanalytical method specified in SOW Exhibit D may be utilized as Long asthe documented instrument or method detection limits meet the ContractRequired Detection Limit (CRDL) requirements. Higher detection limitsmay only be used in che following circumstance:

If che sample concentration exceeds five tines the detection limit ofthe instrument or method in use, che value may be reported even thoughChe instrument or method detection limit may not equal the ContractRequired Detection Limit. This is illustrated in the example below:

For lead:Method in use - TCPInstrument Detection Limit (IDL) - 40Sample concentration - 220Contract Required Detection Limit (CRDL) - 5

C-l 7/87

Tht value of 220 may b« reported even though instrument detection limitis greater than CRDL. The instrument or method detection limit must bedocumented as described in Exhibit E.

(2) The CRDL are the instrument detection limits obtained in pure waterthat must be met using the procedure in Exhibit E. The detectionlimits for sanples may be considerably higher depending on the samplematrix.

C-2 7/87

FIELD MEASUREMENT OF SPECIFIC CONDUCTANCEAND TEMPERATURE

Method: Specif ic Conductance, umhos 9 25*C

Reference: EPA 1979, Page 120.1, Standard Methods, 15th edition, pp 70-73

Detection Limit: 1 umno/cm 0 25*C

Optimum Range: 0.1 - 100,000 umhos/cm

Sample Handling: Determine on-site or within 24 hours

Reagents and Apparatus:

1. Conductivity meter ( Y S I ) and electrodes.

2. Deionized water in squirt bottle.

3. Standard potassium chloride solution, 0.0100 N.

Procedure:

YSI Conductivity Meter

1. With mode switch at off position, check meter zero. If not zeroed,use meter screw and adjust to zero.

2. Plug probe into jack on side of meter.

3. Turn mode switch to red line, and turn red line knob until needlealigns with red line on dial. Change batteries if cannot be aligned.

4. Totally immerse probe 1n sample. Do not allow the probe to touchthe sample container.

5. Turn mode switch to appropriate conductivity scale, X100, X10, or XI.Use a scale that will give a mid-range output on the meter.

6. Wait for needle to stabilize (about 15 sec.) and record conductivitymultiplying by scale setting.

7. While gently agitating the probe, take sample temperature ( *C) andrecord.

3. Rinse probe with deionized water.

9. Record specific conductivity (1st column) and temperature on F.O.S.sheet.

Notes:

1. Calculate conductivity using following formula:

GT______Ll + 0.02 IT^25JI

G25 3 Conductivity at 25*C, umhos/cm

T = Temperature of sample, *C

GT * Conductivity of sample at temperature T, umhos/cm

2. Report results for the standard solution with each data set.

3. Record on field sheet which meter and probe were used. Meter shouldbe wiped clean as necessary.

4. After returning to lab, compare results with previous data. Reportproblems to lab personnel.

Reagent Preparation:

1. Stock Potassium Chloride Solution. 1.00 N: Dissolve 74.555 g. K Clin Milli-Q water and dilute to 1,000 ml. in a volumetric flask.

2. Standard Potassium Chloride Solution, Q.Q1QQN: Dilute to 10.0 mis.of stock solution to 1,000 mis. with Milli-Q water using a volumetricpipet and flask.

ATTACHMENT D-l

OPERATING INSTRUCTIONYS1 MODEL 33

CONDUCTIVITY METER

GLT626/5-10

*V\ -^ ** *• * \ * 4 "" ' r - *» j *

^f^^^s-^'{>i-^"-.--:.'^;*rv "

TABLE OF CONIENTS

Pa ye

GENERAI OESCRIPIION 2

SPECIHCATIONS 2

OPERATION PHOCEDUflE 4

1 Setup . 4

3 Salinity S

4 Conductance *>

5 EHOI 6CIRCUII DESCRIPTION MAINTENANCE

AND CALIBRATION 9

1 Dcstr.jHion 9

3 Call l*i,ihun 9

PROBE 12

1 Duscupiiun ul YSI 3300 CoiMhtCliviIv/TuntiKialuic- Pml>e 12

2 MditiiiMidiice 12

3 Piobe Use . • \: , )4

4 Cell CaMjidlion & Sldiulaid Stiluiiuns' '.t 14

YSI MODEL 33 AND 33M USED Wllll YSI 5IA b4 AND^b/

OXYGEN METERS "'*VV 16

WARRANTY '-^'21

)I

^

,:,-VV/- |? . • • ' • /- . v .-irfi? l ';^^:;*%fp l fe ^g l ^^ - v

GENERAL DESCRIPTIONThe YSI Model 33 and 33M SCI Meieis aie poitahtu batterypoweied. transistorized instruments designed 10 accurately measuresalinity conductivity and temperature They use a piobe consisting ofa rugged, plastic conductivity cell and a piecision YSI ihcinu&tortemperature sensor combined in a single unitConductivity with the Model 33 is expressed as miciomhos/cenii-meier (^mhos/cm), with the 33M. it's millisiemens/nieiei (mS/m)These ate measurements ol the electrical conductance the samplewould show il measured between opposite laces of a 1cm cube(Conversion information: 1 fimno/cm = 01 mS/m} Salinity is thenumber of grams of salt/kilogram of sample (%o - pans perthousand). This measurement assumes the sample contains a "stan-dard" sea water salt mm lure The sample lemperatuie is measuied indegrees CelsiusSalinity mcasuiements are manually temperature compensated bydirect dial Conductivity measurements aie not tempeiatuie compen-sated, however, a temperature function is provided on the m&iiumentlo aid with calculation of collections Also, when |usi tempeiatureand conductivity are known it is possible to calculate salinity andwhen only temperature and salinity aie known it is possible tocalculate conductivity

SPECIFICATIONSModel 33 Conductivity

Ranges

Accuracy

O 500. O 5 OOO. 0 SO OOO^mhos/cm with YSI 3300 SeriesPiobes (Noie The jjmho designations on the meier ate ashorthand foun (or >ntho/cm it 2 5% max eifoi at 60O 5 OOOand 50 OOO plus probet 30% man error ai 25O 2 50Oand 25OOO plus probeSee Error Section

L.

Readability

Model 33M Coi<ductivilyRanges

Accuiacy

Readability

Temperature Compensation

SalinityRange

Accuracy

ReadabilityTunipuiatuie Coinpensanun

2 5 fiii>l>os/i.iit on bOO ,Hi)lio/cinrange25 j<mho$/tm on 5000 fMitho/cinrange250 iiinhos/cin un 50 000

Ternperaiuie Compensation None

O 50 0 5OO 0 5 000 inS/in withYSI 33OO Sunes Pioln-s12 5% max eiior at 60 5OO and5 OOO plus piobe1 3 0% max eiioi ai ?5 250 <inc2 5OO |)lus piobeSee EIIOI Section025 tnS/rn on 5O inSAn Miuje2 5 inS/in on 500 mS/m range*25 O inS/in on 5 000 niSAn ianyeNone

0 4O1 ,- in ictiipui.iuin; i.iiuji- ol 2to i 45' CAhove 4LC i 0 9 ' , . . a i 4 0 and±O 7 UIHI ai 20"iM> plui t undue ttvilypiobeBelow 4"C i 1 I "•••• nt 40 ". diidt 0 9 "mi ai 2O ' •- plus i.omlutlivtlv

piubeSee EIIUI Sei iiun0 2 "'•• °" *J 4^ u" '-»"y«Manual by UueU dial hum 2 10145-C

::•.-:;-.;/>••. ••;•>,- ^^

TempAiaiur*RangeAccmacy

Readability

Power Supply

Piob«

Accuiacy

InstrumentAmbient Range

2 10 t50°C±0 1*C at 2*C ±06"C ai 46"Cplus piobeSee Eiiot SuciioniO 15°C ai 2°C lo ± 0 3 7 f c C ai45°C

Two D sue alkaline batteries Eveieady £95 or equivalent provide appioximaiely 200 his ol operationYSI 3300 Sei.es Conduchviiy/Ieinpcratuitt PiobeNominal Piobe Constant K = 5/cmi 2% ol reading lot conductivity andsalinityEMOI ol ±0 1°C at 0 C andiO3^C at 40"C

San&lacloiy operation 5 10 ( 45°CA maximum eiroi ol ±O 1% of Ihereading pei °C change in int.iiunienltemperature can occur This unor isneyligibte il the mblm.net.i is readju&led lo redlme lor each ruadmy

OPERATION PROCEDURE1 Setup

la) Adjusi meiei /ero III necessary) by tuni.ug i!n-screw on the ineiur lace so tlidi Hit: mL-iui iiciMlle coincideswith the /eio on the conductivity scale

(b| Calibiale the mcler by turnmy ihe Ml)D£ »uniiol toREDLINE and du|ustmg the HEDLINE cunnol bo ihe ini-u-i

needle lines up with ihe ledline on the nieler (due II Ihiscannot he accomplished leplaie the baiienus

It) Plug Ihe |Htil>u into llte piobe jack on the side ul Ihe mslmiimni

Id) Put Hie piobe in the solution to be inc'dSmiMl (Sue PiubeUse )

iheTempeiatiiiaSet the MODE conliol to UMl'LRAlUME Readletnpeiaitiie ort the Iwtiorn scale ol the meiei inCelsius Allow lime lor the probe tempeidiuie lo come loe<|uilibnum wilh that ul the walei beloie leading

Salinityla) Tianslui the tcnipeiature reading liom Slep 2 lo ihe 'C

scale on the instrument(b) SwitLb Ihe MODE control 10 ihe SAllNHY position and

read salinity on the red O 4O u.v meler range(c) Depress ihe CELL TEST iHillon The mt'ltM leaiJmy should

fall less lhan 2% il greater the probe <s touted and ihemeasurement is in euoi Clean the piohe and ic* mca*uie

Conductivity on Model 33 (Model 33M data am inparentheses.)(at Switch the MODE conliol tu the X10O scale II the le.idmy

is below 60 on the 0-5OO ratine (5 0 on ihe 0 50 oiigelswitch lo ihe XIO scale II ihe leading is bhH below 501501 switch to the XI scale Redd ihe metei scale andmulhfily Ihe leading appinpnalely The aaswei is e»

in >i mhos/cm (mS/ml Measuiuinenls aie 1101uie compensatedi Meler Readmy 24 7 (24 71

Scale XIO

Answer 24 70< 2 4 7 0 m S / m J

(b) When measuiing on ih» XIOO and XIO scales, ilepiess theCELL TEST button. The mtiier reading should fall less ihun2%: if greater, the piobe is fouled and I ha measuicineni isin eiiot. Clean the probe and re measure.

NOTE: The CELL TEST does not function on Ihe XI scale6. Error

The maximum error in a leading can be calculated by using thegraphs in Ihe following sections.(1) Temperature

The temperature scale is designed lo give Ihe minimumsalinity error when Ihe temperature leadings die used tocompensate salinity measurementsf KJUI* 1 shows total enur for probe and instrument versus*C maier reading

1 S°C

0 4°C

15"C L 0 4"C Im

Example. Molur Rii.u

ftil.il EIIUI

Acrmacy

|2) ComJuthvily on Model 33 (Mo tit; I 33M tl.ila >ne 111paienlliesus )f iijuie 2 ihows the wotst case conduLiiviiy onoi ds a lunc-liun of the contluchvily luadiiuj fui lh« piohe amimeal coinliined

Figure 2

itoo

_--LUtt

Example. Mului ncj<lmg 360 Hin

Scjte XIO

) TIIOI i 45%

in C)G mS/in)

Anuiacy 3l»OO t IrJ? >'IM|K,S1300 ( 1 6 2 mS/in)lut jiiiilif dixl iiishiiii

^. •&>;•: -•**.. --T*.

(3) SalinityThe salinity leadings aie a function of lentpeiaiuie andconductivity, theiefofe the accuracy is a (unction of both.The lempeiatuf* scale and tempeiaiuie contiol have beendesigned to mmimixe the tcmpuiatuieefior conmhuiion lothe salinity erior The eiioi shown in figuie 3 is ihe loial ofthe lempeialuie and conductivity piobe. the tempeiaiuiescale and ih« salinity scale euor

!»•*MCMMNO

Figurt 3

0/00 CA1IMHV H1AOINC

Example: Metei Reading: 1OO/00, (3 IO°C

%ofReadingEitor 6 5%

Accuiacy 1O%oi 0 65 %o foi alleriors. combined worstcase

i r ! s T-f - \ r >-8: * i t 8r Kh

OUluM

Wfcl/tl*

TL/

.- ..;. .,

•;•';':•-;%•> *

'-^r-f •• -v.lV " jr .

• ^•. -1*1"'

;*

••I/Sir

«n.•"rfvr^^V^iT*1

-^

•;i, - { .-*. _- * *"-•£• •.**»; f -^itf" - i,'.*.i ' •».•?•, * • • .-

Pfl^:*V>^ H.v^:- -lifeS*?";.''>• --.•">NS""c:,/ ' • . / ".^i*'.!.v*v •rvtb^&;^.^-.^;,.'";::(;--"' • '' -:

Wr-^:/v-'• ' ' -V^.e; v^:-;- l':^f>v^;-::fV--;:,;-..-..^^^^-^-•^^§^^^.^^ •W'iAr-- -. '- - ' '.': ••.-Vrl- - ' - ./ ;. ".,-.;.-^- ; / . ' / % . - • - •[>Xi?.v- '•' ' ' •-.. :-' •"' . :-- -V^-^;^^>f^*gjp.-"!-'.-: •;• '*-••• \ ;'.:;>;;- • .-.

CIRCUIT DESCRIPTION. MAINTENANCE AND CALIBRATION1. DescriptionThe circuit is composed of Iwo pans, a multivibrator and switchingtransistors The mulnvibiaior pioduces a square waveform voltage.The square wave is applied to two switching transistors Iliey alter-nalely apply Iwo batteries ol opposite polanly to the piobe lliuspioviding AC power which minimizes polan/aiion eHeels the meteris in series with one baiieiy and measuies the cunent dom a Thecuiieni liom the battery is piopoitional to the conductance ol the cellSalinity is measured in a special iange conductivity cucuit which in-cludes a user adjusted temperature compensator In Ihe temperature.rcdhne and XI positions the multivibrator operates at 1OO II/ In thesalmily. XIOO and X10 positions the multivibrator operates at 6OO Miand HI these ranges pushing the CELL TEST button drops the fiequcn-cy to 100 Ht allowing the operator to judge the deyiee ol probepolar uaiion2. Maintenanc*The only maintenance required is battery replacement Two O sizealkaline flashlight cells, such as Eveieady E95 or equivalent, willpiovide 2OO hrs of operation Accuracy will not he maintained il zinc -catbon "D" cells aic used Baltery replacement is indicated when I heredlme adjustment L anno I be accomplishedReplace haiteiics eveiy sin monihs to reduce the danger of tonosiondue to leaky batteries To icplace baitenes. remove the six scicwsIrom the reat plaie The battery holders are color coded The PusitiveI t- button) end must uo on ted3. Calibration of Modal 33 (Model 33M data aia in parentheses }II is possible fo> Ihe lentpcialure knob to heoimc loose or slip liumits normal position In an emergency Ihe dial can be ie positioned IImust be emphasi/ed that llus is an erneigency (iiocedure i>nly andthat the instrument should be lelurned to the Idilury 'ui jxuperrecadbiaiion at Ihe cailiesl opportunity

(a) Head the temperature and conductivity of Ihe solution Duieimine the salinity of the solution by running a line veitically onthe graph liom this conductance value until H miuisucis theappropriate *C line (interpolate as required for temperaturebetween Ihe given *C lines) From this inicisectiun emend a

.';>'>" "—i/ CM IMAIlOM CHAIir

* • I I I I | I M I | I M I | I I I I | 12 O'C

line hoiuomjily 10 Hie i?d<je of the yiapli This di'ieimini-* thesalinity loi tins sampleExample 25 OOO/i"il>os/cm and 2OUC yivus a s.iliaily ul 11(Example 2.50O tnS/in and 2O°C yves a sjlmtiy of 11 \

|b) Remove the T knoh switch to SALINITY and luin lhi> i onl»olshall until the incici needle mdicaies the sdlmiiy value dutei •mined m Step (a) In Ihe example given, the value is 1 /

(c) Swiich to I£MPtRATUR£ (Note This lempeiaiuie »catlingmust be ihc same as Step (a), if not. begin a<jam at Si<;p lal tPlace Ihe knob on Ihe control shdll (without lummy ihc coniiulshaft) wiilt thu knob poinier ai the same icmpeiaiuie as themetci leading and lighten both sel stie-ws secmuly

At earliest opportunity i calibrate using the lollowiny piutiedum orfttturn Hie msiruriHtnl to factory loi service

la) Sel the uistiumeni foi a salinity measurement as nonnai(b) Substitute a lOOO^I capacitor aitd 112 7uhmO l%iului.mce

icsisloi foi the piubeConnect Ihe tesislor and capacitor between iht> yieen wno diul tedwue on the jack connections inside the

GREEN WIRE

RED WIRE

112 7

.IX

-if-I000/,t

1011

|c) Torn the temperature dial unlil the meiei leads itdltneNow insiall ihe tempeiatuie knob with the aiiow ai 25"C Tins is atemporary cahbiaiion only Reiuin ihe instrument to ilie laciuiy loipioper *ec alibi anon

PHOBE1. Desciiption of YSI 3300 Seiies Conductivity/Temperature

ProbeThe YSI 3300 Seiies Coitduchviiy Piobes aie designed lot field useembodying constiuchon and design lot rugged, accurate serviceEach probe features a built in cell constant of & 0 1500 0/M( ±2%. aprecision YSI ihetmistof lampeiatuia sensoi ol ±0 1°C accuiacy aiO*C and ±0 3*C ai 40*C and a low capacitance cable assembly ter-minaiing in a thiee llierminal 0 25" dia phone type connecioiThe 33IO has a 10 fl cable and ihe 3311 is a SO ll veision Oiheilengths a*e available on special wderThe piobe has a nyid P V C body, plaiun/ud puie nxikel elecnodes.and a dmable cable pioviding iesisiance 10 a wide lanye ol waiei-boine substances2. Maintenance

U) CleaningWhen ihe cell test indicates low leadings ihe piobable tauie is dulyelcciiodes Haid waici deposits oils and oiganic inanei aie the mostlikely contaminantsFoi convenient noimal cleaning soak I lie uleuiodes for 5 minuteswilh a locally available bathioom tile cleaning pio|>aiaiu>n SUL!I asDow Chemical Baihioom Cleaner Hon/on liultismos Rjlly TilePoicelam. and Chiome Cleanei Johnson Wax tiny IniianiClcanei 01 Lysol Bund Basm Tub Tile Cleanei

For sliiMKjei cleaning a 5 minute soak in a solution ni.uk' ol 10 pansdistilled water. lOpaits isopiopyl alcohol and 1 pan IICI can be u^cdAlways rinse the p(»be utter cleaning and heloie stougeCAUTION Do not touch the eluc nodes inside Ihe piobe

Platiniitn black is soil and can be scraped oilII cleaning does nut testure the piohe peiloimaiice. ic pldliiu{iiiy is

|b) he PlaiimjiitgEt|iiipinertlII) YSI 03)40 Plaiinumg Solution 2 II u/ (3% platinumrhlonde dissolved m OO25% lead acetate solution)U) YSI Model 33 ur 33M S C T Meter(3) &O rnl glass breaker or equivalent bottle14) Distilled waier

Ill Clean the piobe as in Section la) either method(2) Place the cell m ihe beaker and add sullicient YSI #3140

solution to cover Ihe electrodes Do not cover the lop olIhe proue

(3) Plug the piobe into the Model 33 01 33M. switch to (heX1OO scale to platmue the clecliodc Move the piobeslightly to obtain the hiyhtisi meter reading and continuepldtinumg for the approximate lime shown below

Molct Readingfi mhos/cm mS/m

3OOOO2500O20OOO150OOIOOOO

30OO2 5OO20OOI 5OOI 000

Time(minutes)

566

I 116

12

" • ' ' ""

13

*»* fc Jyj*-.w- tyr^-»i *^-

j?$M&s$$'t

(4) After ihe elapsed lime remove the probe and rinse in freshwater,

(b) Return the solution lo Us container 2 02 ol solutionshould be sutticienl lor 50 treatments

3. Probe Us*(at Obstructions near the probe can distuib readings. At lea^t two

inches ol clearance must be allowed Irom non-metallic un-derwater obji-cls. Metallic objects such as piers or weightsshould be kept at leasl 6 inches Irom Ihe probe

Ibt Weights are attached lo the cable ol ihe YSI 3310 and 3311Probes The YSI 3327 Weights are supplied in pairs with atotal weight ol 4 ounces per pair. Should it become necessarylo add more weight lo overcome waier currents, we suggestlimiting Ihe total weight lo two pounds (8 pairs) For weightsin excess of two pounds use an independent suspensioncable. In either case, weights must be kepi ai leasl 6 inchesaway from the probe.

Ic) Gentle agitation by raising and lowering the probe severallimes during a measurement insures flow of specimen solu-tion through the probe and improves the lime response ol thetemperature sensor

4. Cell Calibration ft Standard SolutionsThe YSI 0330O Series Cells are calibrated to absolute accuracy ol± I 5% based on a standard solution Since the literature on conduclivity does not indicate a consistently accepted standardizationmethod, we have chosen Ihe O01 demal KCI solution method asdetermined by Jones and Bradshaw in 1937 as our standard Recenttextbooks, as well as the ASTM standards, concur with this choiceThe solution is prepared by diluting O 745 grams of pure dry KCI withdistilled water until the soluiion is 1 kilogram The table below showsIhe values ol conductivity this solution would have if the distilledwater were non-conductive However, since even hiyh purity distilled

water is slightly conductive*, the measiiicd conductivity will IIL* higherby an amount eijtial lu tlii> water's conductivity

ConductivityTentpeialure

15161718192021222324252627262930

The opuiator may use the staiidaid solution and the uble tu (Jink accuracy ol a cells constant 01 to deteimine an unknown constant [heformula is shown below

V mhos/cm1141 51167 51193612199124641273O1299 713266135361360814O8 11436514632149O9151871546 7

inS/m11421168119412201246127 313OO132 71354138 114O814371463149 1151 9154 7

flICi f C?| RiS.10*

where K =R =C. =C, =

Cull constantMeasured tesi&unce in '.(Conductivity m fjinhos/cmConductivity in /jmlios/cm ol llie iliMillnl waterused lo iD.'ike sululion

14

K$|5* > -:>•-' : r4fy«t^

. •(^ • >*-v.v*

Si = Conduclivily in mS/mSi = Conduclivily in mS/m ol Ihe di&iilled waiei used

to make Ihe solution

R. Ci and Ci. oc S> and Si. must euhei be detuiininud at ihe sameicmperaiuie of canceled lo Ihe same temperature 10 make (he equaIIQII validNote: For lurlbei information on conduciivily and the above siandaid mfoimaiion. icier to ASTM Standards Part 23 — SiandaidMethods of Tesl foi Elecincal Conduchvily. ot Waiei and IndusmalWaste Waiei — ASTM Designation D1125 64

YSI MODEL 33 AND 33M USED WITH YSI 61A. 64 and 67OXYGEN METERSIf ihe salinity measurement is lo be used foi sahnity correction on theSI A. the feadmg should be convened to Chtorosiiy The formula is

Salmily "00 OO3PPM Clilorosilv = —————— r^ ————— * 10'

I D

For these instruments the 003 can he neglected so the equationsimplifies to

PPM CI =SS

fl

IB 27 36 S%t

5.000 10,000 IV000 20.000 PPM CL

16

foi salinily coiieciion wliun using the Model 57 use tliuiL-diJiny direct doiri the Model 33 or 33M No cuiweision isnecessaryModel 33 and 33M salinity readings taken in conjunction wHh Model54 dissolved oxygen rcadinys can be used to con eel the Model 64 lorsalinity and lo make post mcd&uiemeiil salinity collections lo dissolved ojiyycn data Collection tables ate available hum ihe lacioiy

WARRANTYAll YSI pioducis cany a mtc year unconditional wairanly onvwoikmanship and parts, exclusive ol ball ones Dam.ii|c ihiouijh ,»LCIdent, misuse. 01 lampeung wiU l>e iepaned at a nominal clunjeII you are experiencing difficulty with any YSI puxtiicl. il nuiy boreturned lo an aulhoii/ed YSI dealer lor lepair. even if the waiiantyhas expiiud II you need factory assistance loi any u.-asou runiji;!

Scivice Depai IntentYellow Spnngs Instrument Co IncPO Box 279Yellow Springs. Ohio USAPhone (5131 767 7241

Appendix ETARGET COMPOUND LIST AND

CONTRACT REQUIRED DETECTION LIMITS

FIELD FILTERING

Reference: EPA 1979, Metals 5

Sample Handling; Filter as soon as possible after samplecollection

Reagents and Apparatus;

1. 10% HNO- solution in a squirt bottle and in a literplastic bottle

2. Milli-Q water

3. Plastic forceps

4. Millipore pressure filtration apparatus

5. 0.45 um membrane filters (142 mm)

6. Glass fiber prefilters (124 mm)

7. 0.80 um membrane filters (142 mm)

8. Compressed air

9. Pressure gauge

Procedure:

1. Using plastic forceps, place a 0.45 um filter on top ofblue grid.

2. Center the stainless steel cylinder on top of the fil-ters, with the white gasket down.

3. Place a prefilter on top of membrane filter.

4. Place top onto cylinder rim. Screw handwheel boltsdown until even and snug. Finish tightening with plas-tic wrench.

5. Attach end of PVC hosing to compressed air source.

6. Using squirt bottle, squirt about 10 to 25 ml of 10%HNO. into top opening.

7. Using the clamp, attach the hose adaptor to the topopening. Tighten clamp.

8. Place beaker under outlet.

9. Slowly increase pressure on compressed air tank. Whenliquid flows from outlet, lower pressure and bleed airfrom top plate vent valve until pressure gauge reads10 to 15 psi. Close valve and continue until flowstops. No leaks should be observed.

10. Shut off compressed air and open release valve.

11. Disconnect top clamp.

12. Rinse twice with Milli-Q water, following above pro-cedure.

13. Place clean beaker or sample container under outlet.

14. Add sample, following above procedure, except that whenflow of sample from outlet stops, increase pressure inincrements of 10 psi to 75 psi and continue until flowstops.

15. After shutting off air and opening release valve, dis-connect clamp and top plate and remove cylinder. Throwfilters away.

16. Pour sample into a 1 liter plastic bottle; label andpreserve.

17. Run a Milli-Q blank every 10 to 20 samples.

Notes:

1. Samples with high sediment can be filtered throughseveral membranes with increasing pore size and severalprefilters. The 0.45 urn membrane filter should alwaysbe on the grid and the coarsest filters on the top.

Reagent Preparation:

1. 10 percent HNO. Solution; Add about 900 ml of Milli-Qwater to a 1 liter Erlenmeyer flask. Using a graduatedcylinder, add 100 ml concentrated HNO3 to the Milli-Qwater while stirring.

CVR178/044

Appendix FSPECIAL ANALYTICAL SERVICES—REGIONAL REQUEST

SPECIAL ANALYTICAL SERVICESCH2M HILL METHOD VWF-1

U.S. Environmental Protection Agency SAS NumberHWI Sample Management Office [ ]P.O. Box 818, Alexandria, Virginia 22313PHONE: (703) 557-2490

SPECIAL ANALYTICAL SERVICESRegional Request

[X] Regional Transmittal [ ] Telephone Request

A. EPA Region and Site Name: Region V, Verona Well Field

B. Regional Representative: John Tanaka

C. Telephone Number: (312) 886-6337

D. Date of Request:

1. General description of analytical service requested:

Fourteen-day turnaround for a total of 99 soil analyses collected during the

soil boring phase of RI/FS sampling and analysis. Samples will arrive in

brass split-spoon liners and are to be composited by the laboratory to gen-

erate the analytical samples. Three brass liners will be used to prepare each

analytical sample. All samples are to be analyzed for TCL volatile organic:

compounds following CH2M HILL Method VWF-1 (copy attached) and CLP RAS proto-

col for GC/MS analysis of volatile organic compounds (VOCs).

2. Definition and number of work units involved (specify whether wholesamples or fractions; whether organics or inorganics; whether aqueous orSoil and sediments; and whether low, medium, or high concentrations):

Analyze 99 soil samples for the VOCs listed in the U5EPA CLP IFB SOW. Soil

samples wil] be shipped fromjthe Verona Welj Field Superfund Site jLn Battle

Creek, Michigan, in the brass liners in which they were obtained. From 20 to

30 brass liners containing soil samples will be shipped daily from the field

until all of the samples have been collected. Three liners will be

CVR168/062/1

composited in the laboratory to form one analytical sample for VOC analysis

following CH2M HILL Method VWF-1 (copy attached).

3. Purpose of analysis (specify whether Superfund (Remedial orEnforcement), RCRA, NPDES, ETC.):

Superfund—Remedial Action___________ __________ __

4. Estimated date(s) of collection:

June/July 1988

5. Estimated date(s) and method of shipment: Daily by overnight carrier.

6. Approximate number of days results required after lab receipt of sam-ples:

Laboratory should report results within 14 days arter receipt of samples

7. Analytical protocol required (attach copy if other than a protocol cur-rently used in this program):

Sample preparation and handling in accordance with CH2M HILL Method VWF-1

(attached). Sample analysis to follow CLP RAS medium level protocol for

GC/MS analysis (SOW-7/87.

8. Special technical instructions (if outside protocol requirements, spe-cify compound names, CAS numbers, detection limits, etc.):

Initial analyses shall be performed without sample dilution. If high_____

concentration of some compounds exceeds the calibration range, the sample

shall be diluted and reanalyzed to obtain values for the high concentration

components. All sets of results shall be reported.

9. Analytical results required (if known, specify format for data sheets,QA/QC reports, Chain-of Custody documentation, etc.). If not completed,format of results will be left to program discretion.

CVR168/062/2

During all phases of soil sampling and analysis, the full CLP documentation

data package is required. The laboratory shall also attach to the data____

package a photocopy of the Chain-of-Custody documentation.

The test procedure used by SAS will be clearly identified. Bench records and

all records of analyses and calculations for things such as samples, blanks,

duplicates, spikes, and standards, with resulting instrument inputs or concen-

tration readouts, will be provided by CLP SAS, along with worksheets used to

calculate results.

10. Other (use additional sheets or attach supplementary information, asneeded):

11. Name of sampling/shipping contact: Joseph DankoPhone: 503/752-4271

Please return this request to the Sample Management Office as soon as poss-ible to expedite processing of your request for special analytical services.Should you have any questions or need any assistance, please call the SampleManagement Office.

I. DATA REQUIREMENTS

Parameter Detection Limit Precision Desired(+/- % °r cone.)

VOA Same as IFB HAS SOW <±20%

CVR168/062/3

II. QUALITY CONTROL REQUIREMENTS

Audits Required Frequency of Audits Limits*(+/- * °r cone.)

Same as IFB RAS SOW Same as IFB RAS SOW Same as IFB RAS SOW

III. *Action Required if Limits are Exceeded:

Contact: Chuck Elly, CPMS, CRL (312) 353-9087

John Tanaka, EPA (312) 886-6337.

Joseph Danko, CH2M HILL, (503) 752-4271

CVR168/062

CVR168/062/4

Appendix GFIELD FILTERING METHOD FOR

GROUNDWATER SAMPLES

SPECIAL SAMPLE PREPARATION FOR SOIL SAMPLES

(CH2M HILL Method VWF-1)

1. SCOPE AND APPLICATION

1.1 This sample preparation method is presented as a

means to prepare composite soil samples in associa-

tion with the Remedial Investigation/Feasibility

Study (RI/FS) phase of the Verona Well Field

project.

1.2 The intent of the special sample preparation method

described herein is to maintain sample integrity

by minimizing loss of volatile constituents from

the soil.

2. SUMMARY OF METHOD VWF-1

2.1 The method has been designed to accommodate col-

lection of a representative aliquot of soil from

approximately 270 brass liners while minimizing

loss of volatile constituents. Soil samples will

be received by the laboratory in a brass liner and

preserved on ice. The samples will be prepared

and analyzed within 14 days of the time of collec-

tion.

2.2 Composite samples are prepared by extracting an

aliquot of soil from the entire length of each

brass liner by using a coring utensil similar to a

conventional laboratory cork boring tool. The

soil will be allowed to thaw and the boring device

cored through the center of the soil plug. Ali-

quots of soil obtained from each sample liner are

immediately placed into methanol to effect a sol-

vent extraction of the soil. Soil cores from

three designated brass liners are composited to

form each analytical sample destined for analysis.

A portion of the methanol extract is recovered for

analysis in accordance with the U.S. EPA Contract

Laboratory Program (CLP) Statement of Work for

GC/MS analysis of VOCs by the medium level method

for soil samples.

SAFETY

3.1 Samples are contaminated with constituents that

are either identified as, or suspected of being,

carcinogens. All samples are considered to be

hazardous. As such, they are to handled with the

utmost care and following established laboratory

technique in order to avoid harmful exposure.

3.2 Sample liners will have been stored on ice prior

to arrival at the lab. They should be handled

with hands covered first by surgical gloves and

then insulated gloves,

3.3 Samples shall be prepared in a fume hood. This

includes all subsampling and compositing activ-

ities.

4. APPARATUS AND MATERIALS

4.1 Sample coring utensil, similar to a laboratory

cork boring tool suitable for collecting a repre-

sentative aliquot of sample from the entire length

of a brass liner.

4.2 Bottles, 4-ounce (120 mis), wide-mouth amber glass

jars with teflon-lined caps, for use in extraction

of soil.

4.3 Analytical top loading balance, used to determine

the weight of sample aliquots to the nearest

0.01 grams.

4.4 Labware, pipettes, and glassware, as needed for

preparation of samples.

5. CHEMICALS AND REAGENTS

5.1 Methanol, spectrographic grade suitable for use in

accordance with CLP SOW for GC/MS analysis of VOCs.

5.2 Refrigeration, as required to keep samples at 4°C

prior to and after sample preparation.

6. SAMPLE PREPARATION STEPS

For Composites of Soil Samples Collected During the

Soil Boring Phase of RI/FS sampling.

6.1 Allow the sealed brass liner containing the sample

to thaw out such that the soil can be sampled in

the manner described in this section. It is rec-

ommended that a set of practice samples be pre-

pared from soil similar to that of the samples.

These can be used for perfecting the coring tech-

nique .

6.2 Remove the teflon-lined cap from one end of brass

liner to expose the end of the soil sample.

6.3 Using a preweighed coring utensil, immediately

collect an appropriate aliquot of sample by pushing

the utensil downward into the length of the liner.

Withdraw the coring utensil and reweigh to verify

that an adequate aliquot (i.e., 15 grams) of sample

has been obtained. This aliquot and aliquots from

other samples that make up a particular composite

should be of approximately equal weight (plus or

minus 10 percent). Record the weight of the sample

aliquot and all other soil aliquots used in the

makeup of the composite sample.

6.4 Immediately extrude the contents of the coring

utensil into a 4-ounce brown glass bottle that

contains an appropriate volume (i.e., 45 mis) of

methanol to accommodate the aliquots of each sample

to be composited. Soil cores from three designated

brass liners will be composited in the same bottle

to form each analytical sample. It is important

that approximately a 1:1 ratio of methanol volume

(mis) and total composite sample weight (gms) be

maintained to preserve the detection limit of the

analytical method. Record the volume of methanol

used in the extraction.

6.5 After addition of the sample aliquots to the meth-

anol, cap the bottle and extract the sample by

vigorous shaking for 2 minutes.

6.6 Recover a portion of the methanol extract for

analysis by U.S. EPA CLP SOW for GC/MS analysis of

VOCs by the medium level method for soil samples.

6.7 In accordance with the analytical method, the

amount of sample extract added to the sparging

cell of the purge and trap concentrator shall not

exceed 2.0 percent of the volume of water also

added to the sparging cell. However, it is impor-

tant for detection limit determinations that this

2.0-percent methanol/water relationship be ap-

proached in the analysis.

6.8 Should any delay occur between the extraction of

composite samples and the analysis, the sample

extracts shall be sealed in extract vials and

refrigerated.

6.9 All sample liners shall be recapped and placed in

the refrigerator after the sample aliquots have

been taken.

7.0 QUALITY ASSURANCE

7.1 Quality assurance associated with this special

sample preparation method is described in the

Quality Assurance Project Plan (QAPP) for this

project. QA/QC steps prescribed in the QAPP will

assess the precision and accuracy of the sample

preparation techniques.

CVR178/038

SPECIAL ANALYTICAL SERVICESCH2M HILL METHOD VWF-2

U.S. Environmental Protection AgencyCLP Sample Management OfficeP. 0. Box 818, Alexandria, Virginia 22313PHONE: (703}/557-2490 or FTS/557-2490

SAS Number

SPECIAL ANALYTICAL SERVICESClient Request

Regional Transmittal Telephone Request

A. EPA Region/Client: V/ CH2M HILL

B. RSCC Representative: jan

C, Telephone Number: (3121 353-?7?Q

D. Date of Request: __________

E, Site Name: Verona Well Field, Michigan

Please provide below a description of your request for Special Analytical Services underthe Contract Laboratory Program. In order to most efficiently obtain laboratory capability foryour request, please address the following considerations, if applicable. Incomplete orerroneous information may result in delay 1n the processing of your request. Please continueresponse on additional sheets, or attach supplementary information as needed.

1. General description of analytical service requested: Seven-day turnaround on analysis___of groundwater samples for volatile organic compounds (VOCs) with low detection limits.

Definition and number of work units Involved (specify whether whole samples orfractions; whether organics or Inorganics; whether aqueous or soil and sediments;and whether low, medium, or high concentration): Analyze 120 groundwater samples, 12 fieldduplicate samples, and 12 equipment blank samples per round for three run for VOCs. Sampleswill be shipped from the Verona Well field site in Battle Creek,Michigan, in 40 ml VOA vialsFrnm 1R t.n ?n <;amp1ps will hp shipped daily from thp field until all of the samples havebeen collected.

Purpose of analysis (specify whether Superfund {Remedial or Enforcement), RCRA,NPDES, etc.):

Superfund - Remedial Action

Rev. 7.08/88

- 2 -

4. Estimated date(s) of collection:

5. Estimated date(s) and method of shipment: Dailv bv overnight carrier__________

6. Number of days analysis and data reaulred after laboratory receipt of samples:Laboratory should analyze samples within 7 days after receipt of samples. Report

_ and data due'Wi'thin-lS davs of sample receipt._______;_________________

7. Analytical protocol required (attach copy 1f other than a protocol currently used inthis program):

CLP SOW for Organic Analysis (Hulti-Media. Multi-concentration) 8/87._____

8. Special technical instruction (if outside protocol requirements, specify compoundnames, CAS numbers, detection limits, etc.):________________________

1. Modifications to the CLP Organic SOW 8/87 in Attachment I.

Z. Required low level quantitation limits in Attachment II.

9. Analytical results required (if known, specify format for data sheets, QA/QC reports,Chain-of-Custody documentation, etc.). If not completed, format of results will beleft to program discretion.

As per CLP Organics SOW 8/87 with modifications as outlined in Attachment I.______

10. Other (use additional sheets or attach supplementary information, as needed)

11. Name of sampling/shipping contact:____.incoh

(503) 752'42718/88

3.

I. DATA REQUIREMENTS

Parameter: Detection Limit Precision Desired(+% or Cone.)

See attachment IT SPP at+nrhmont IT t ?ny.______

II. QC REQUIREMENTS

Audits Required Frequency of Audits Limits* (% or Cone.)

As per CLP Organic SOW 8/87 As per CLP Organic SOW 8/87 Exceptions to CLP OrganicSOW 8/87:See attachment I

III. ACTION REQUIRED IF LIMITS ARE EXCEEDED:

Contact Region V RSCC Jan Pels (312) 353-2720Chuck Elly (312) 353-9087

Please return this request to the Sample Management Office as soon as possible to expediteprocessing of your request for special analytical services. Should you have any questionsor need any assistance, please call the Sample Management Office. Rgv ' ? Q

o /oo

(page 1 of 2)

TO OHE OP ORCMJICS sow s/e?FOR THE REGION V RESIDENTIAL WELL SAS

Special Technical Instructions.Volatilea) Sanple volume is increased from 5 ml to 20 ml to achieve

lower detection limits.b) Acroiein and acrylonitrile are additional target

ccnpounds. Uiese ccnpounds will be added to both initialand continuing calibration mixes. Acroiein andacrylonitrile will use brcrrochlorcmethane as internalstandard. Ine prijnary and secondary quantitation ions tobe used for these confounds are:Ccnpounfl prjmary ion S-qconflary lonjs)Acroiein 56 55Acrylonitrile 53 52

c) Initial calibration: Five point initial calibration - 5,10, 20, 40 and 60 ug/L for all compounds except foracroiein and acrylonitrile, which should be analyzed at25,50,75,100 and 125 ug/L-

d) Continuing calibration: Performed at 20 ug/L for allcompounds except acroiein and acrylonitrile, which shouldbe analyzed at 50 ug/L.

e) Surrogate standards: SOW standards spiked at 20 ppb.f ) Matrix spike/matrix spike duplicate: Matrix spike

compounds at 20 ppb.

2. Analytical Results Required.a. Quantitation Limits: Organic Analysis Data Sheets, Form 1

will reflect detection limits experimentallydetermined and verified previously by the contractor. Thiswill include a method detection limit study andverification spikes at the determined MDL. Theverification spikes must meet all qualitative criteria(i.e. mass spectral ions/abundances, retention tijnes).

b. Additional volatile parameters: Results will includeall data for acrolein/acrylonitrile analogous to theother volatile target compounds in the SCW. Form Is maybe appended to include these parameters.

c. Dilutions: One contractor will request permission of theRegion to dilute any sample exceeding the initialcalibration range for any parameter. Diluted and undilutedsample data, will be included in the results as per theSCW.

(page 2 of 2)

a) Bctraction/blowdown: Detract entire liter bottle; rinsecap & bottle and add to sample. Final blowdown volumemay be decreased -op to 1/2 the SOW specifications toachieve required detection limits.

b) Surrogate standard: Spiked at 0.02 ppb.c) Matrix spike/toiatrix spiJce duplicate: SOW Matrix spite

confounds spiked at 1/5 the SOW concentration.

3. QC Requirements:a. Initial Calibration (Vblatiles/Semivolatiles):

SPCC criterion apply.OJie %RSD for the RFs for all confounds mist be £ 35%.Die RFs for all other (nan SPCC) ccnpounds must be 0.05

b. Continuing Calibration:SPCC criterion apply.

%D for the RFs for all compounds most be < 25%.RFs for all other (non SPCC) conpounds must be > 0.05

c. Matrix Spike/ Matrix Spike Duplicates:SOW criterion apply for %R and %RPDs.

d. Surrogates:Sow criterion apply for %R and corrective action.

e. Blanks:Sow criterion apply.

ATTACHMENT II 1 of 4

(ALL UNITS ORE MICRCGR£ri£/LlTE9)

CAS »ANALYTEBENZENE

Q U A N T I T A T I O N LIMIT

CAR r CM TE7RGC;-*u3RID£

CHLOROFORMCHL3 RCMETHANED I B ROmOCHLOROMETHANE1, 1-DICHLORCETHftNEI, 2-DICHLQROETHANE1. 1-OICHLOROETHENEtotal -U2-DICHLOROETHENE1. 2-DICHLQROPRCPANEci«-l, 3-DICHLOPROPROPENEtr«ns-l, 2-DICHuOROPROP£NeETHYL BENZENEMETHYLENE CHLORIDE <»>1, 1. £. 2-TETRftCMLOROETHflNETETRfiCHLOROETHENETOLUENE <»>1, 1, l-TRICHLOROETHONE1, l.S-TR I CHLOROETHftNETRICHLOROETHE.NEVINYL CHLORIDEACRQLEIN -ACHTONE <»)aCRYLONlTRILECARBON DISULFIDES-5UTftNON£ (4)VINYL ACETATE*-mETHYL-2-PENTANOME£-HEXGNQNESTYRENEm-XYUENE K*o-XYLENE *•••O-XYL£NE »*

w"s?-*75-£3-£74-S3-338-52-5

106-30-775-OO- 3

67—88-374-87-3

124—48-173-3^-3

1O7-O6-273-35-4

•78-87-5

10061-01-310061-02-S

100-41-473-09-279-34-3

127-18-4 -108-86-371-35-679-00-379-O1 -S73-O1 -4

107-oa-a67-64-1

107-13-173-15-076-93-3

108-OS-4ioa-io-1319-78-fi100-42-3108-38-3

1.51. 51.51.51.5A • v

1, 5

1.31.51.51.51.51.31.31.5£11.511.31.51.31.51.31.31-5•15'53.5355

1,5c1

~S _

COMMON LABORATORY SOLVENTBLANK LIMIT IS 3x METHOD DETECTION LIMIT

*» THE o-XYLENE AND p-XYLENE ARE REPORTED AS A TOTAL OF THE

Rev. 7.0

SPECIAL ANALYTICAL SERVICESCH2M HILL METHOD VWF-3

(Standardized SASs for Sulfate and Sulfide)

U.S. Environmental Protection Agency SAS NumberHWI Sample Management Office [ ]P.O. Box 818, Alexandria, Virginia 22313PHONE: (703) 557-2490

SPECIAL ANALYTICAL SERVICESRegional Request

[X] Regional Transmittal [ ] Telephone Request

A. EPA Region and Site Name: Region V, Verona Well Field

B. Regional Representative: John Tanaka

C. Telephone Number: (312) 886-6337

D. Date of Request:

1. General description of analytical service requested:

Analysis for sulfate in groundwater, Samples_ will be unfiltered. Results

are reported as mg/1 SO ._________________________________________

2. Definition and number of work units involved (specify whether whole sam-ples or fractions; whether organics or inorganics; whether aqueous orsoil and sediments; and whether low, medium, or high concentrations):

Analyze 8 groundwater samples, 1 duplicate sample, and 1 equipment blank sam-

ple per round for two rounds for Sulfate. Samples will be shipped from the

Verona Well Field Superfund Site in Battle Creek, Michigan, in 1-liter high-

density PE bottles. The samples will be shipped from the field over a 2- or

3-day period.

3. Purpose of analysis [specify whether Superfund (Remedial or Enforcement),RCRA, NPDES, ETC.]:

Superfund—Remedial Action __________________________________ ___

CVR178/045/1

4. Estimated date(s) of collection:

August 1988, October 1988, January 1989

5. Estimated date(s) and method of shipment: Daily by overnight carrier.

6. Approximate number of days results required after lab receipt of samples:

The laboratory should report results within 28 days after receipt of samples.

7. Analytical protocol required (attach copy if other than a protocol cur-rently used in this program):

A. EPA Method 375.2 (Colorimetrie Methylthmol Blue)—1983 ed. Note:

_______This method requires 0.75 mg/1 SO in Dilution Water (See Reagent

________Section 6.8) ._______________________________________________

____B. Method 426C of Standard Methods, 16th ed. (Turbidimetric). Note:

_______This last method provides for measurement of sulfate using two____

_______standard curves—one for sulfate concentrations between 0 and_____

________10 mg/1, and one between 10 and 40 mg/1 sulfate.________________

_______Samples will be kept at 4°C until validation of results._________

8. Special technical instructions (if outside protocol requirements, spe-cify compound names, CAS numbers, detection limits, etc.):

Sample holding time is not to exceed 28 days from date of sample collection.

Sulfate standards will be prepared daily from stock solution. Samples with

absorbances or turbidities greater than that in the highest standard will be

diluted and rerun. For Method 426C, 1) the reanalysis solution should con-

tain between 20 and 40 mg/1 sulfate, and 2) concentrations must be corrected

for background turbidity and color per Section 5d of Method 426C using pH____

adjusted sample a1iquots. Use on ly the methods specified. Calibration curves

must include at least 6 points (including a zero concentration standard) for

Method 375.2 and Buffer A of Method 426C.________________________________

CVR178/045/2

9. Analytical results required (if known, specify format for data sheets,QA/QC reports, Chain-of Custody documentation, etc.). If not completed,format of results will be left to program discretion.

The test procedure must be clearly identified. Results shall be reported as

mg/1 SO . Bench records tabulating the order of calibration standards, lab

control standards, lab blanks, samples, spikes, etc., with resulting absorb-

ances or concentration readouts, will be provided along with copies of work-

sheets used to calculate results. Background absorbances used for turbidity

corrections must be tabulated for each sample aliquot tested. A photocopy of

the instrument readout (i.e., strip charts, printer tapes, etc.) must be in-

cluded. All records of analysis must be legible and sufficient to calculate

all concentrations and results.

EPA QC reference samples, or any other reference sample or initial calibration

verification, wij l be identified as to source, lot number, and sample number.

Corresponding "true" or targgt values and associated 95 percejvt confidence

l-imits for analysis results will be provided for all reference samples used.

10. Other (use additional sheets or attach supplementary information, asneeded):

11. Name of sampling/shipping contact: Joseph DankoPhone: 503/752-4271

CVR178/045/3

I. DATA REQUIREMENTS

Parameter

Sulfate

Detection Limit

5 rog/1

Note: These are minimum

requirements. Report the

actual detection limits

used based on allowable

methodology options.____

II. QUALITY CONTROL REQUIREMENTS

Audits Required Frequency of Audits

Precision Desired(+/- % or cone.)

Method 375.2: Differences

in duplicate sample results

are to be <5 mg/j for con-^

centrations <50 mg/1, and

<10 percent for concentra-

tions £50 mg/1. _________

Method 426C: Differences

in duplicate sample results

are to be £2 mg/1 for

centrations S20 mg/1 and

£10 percent for concentra-

tions >2Q mg/1 in aliquot_

tested.

Limits(+/- % or cone,

Matrix Spike* One per group of 10 or 65 to 115 percent

fewer samples.

Lab Duplicate One per group of 10 or

fewer samples.

±(10 percent or 5 mg/1)

for Method 375.2

±(10 percent or 2 mg/1)

for Method 426C

CVR178/045/4

II. QUALITY CONTROL REQUIREMENTS (Continued)

Audits Required Frequency of Audits Limits;+/- % or cone.)

Lab Blank (0 mg/1 SO.) One per group of 10 or <5 mg/1--Method 375.2

fewer samples.________ -2 to +2 mg/1—Buffer B

Lab Blank (10 mg/1 SO.) One per group of 10 or

fewer samples.

of Method 426C or 8 to

10 mg/l-*Buffer A of

Method 426C

Calibration Verification One per group of 10 sam- 90 to 110 percent

Standard____________ pies and at end of sam- _____________

__________________ pie set once per sample _____________

set.

One set of EPA QC

Mineral Reference Sam-

ples

85 to Ijj i___p_ercent

each concentration.

*Matrix spike concentrations will be greater than 30 percent of sample con-centrations, but spiked samples shall not exceed working range of standardcurve.

III. ACTION REQUIRED IF LIMITS ARE EXCEEDED:

Take corrective action and reanalyze samples.________________________

Contact: John Tanaka (312) 886-6337 or Chuck Elly, CPMS, CRL (312) 353-9087

Please return this request to the Sample Management Office as soon as possibleto expedite processing of your request for special analytical services. Shouldyou have any questions or need any assistance, please call the Sample Manage-ment Office.

Joseph Danko (503) 752-4271

CVR178/045

CVR178/045/5

U.S. Environmental Protection Agency SAS NumberHWI Sample Management Office [ ]P.O. Box 818, Alexandria, Virginia 22313PHONE: (703) 557-2490

SPECIAL ANALYTICAL SERVICESRegional Request

[X] Regional Transmittal [ ] Telephone Request

A. EPA Region and Site Name: Region V, Verona Well Field

B. Regional Representative: John Tanaka

C. Telephone Number: (312) 886-6337

D. Date of Request:

1. General description of analytical service requested:

Analysis for sulfide in groundwater. The jetection limit using_ this method

is 1 mg/1._____________________________________________________

2. Definition and number of work units involved (specify whether whole sam-ples or fractions; whether organics or inorganics; whether aqueous orsoil and sediments; and whether low, medium, or high concentrations):

Analyze 8 groundwater samples, 1 duplicate sample, and 1 equipment blank sam-

ple per round for two rounds for Sulfide. Samples will be shipped from the

Verona Well Field Superfund Site in Battle Creek, Michigan, in 1-liter high-

density PE bottles. The samples will be shipped from the field over a 2- or

3-day period. ___ ___ _____________________

3. Purpose of analysis [specify whether Superfund (Remedial or Enforcement),RCRA, NPDES, ETC.]:

Superfund—Remedial Action ___ ______ ____ _____________________

CVR178/046/1

4. Estimated date(s) of collection:

August 1988, October 1988, January 1989

5. Estimated date(s) and method of shipment: Daily by overnight carrier.

6. Approximate number of days results required after lab receipt of samples:

The laboratoryshould report results within 28 days after receipt of samples.

7. Analytical protocol required (attach copy if other than a protocol cur-rently used in this program):

"Standard Methods for the Examination of_Water and Wastewater," 16th edition:

Method No. 427 B and D with No. 421 B 2e. for standardization of the iodine

solution and No. 421 B 2f for standardization of the sodium thiosulfates solu-

tion. No other methods may be used._________________________________

8. Special technical instructions (if outside protocol requirements, spe-cify compound names, CAS numbers, detection limits, etc.):

Collect all samples in duplicate to allow for any reanalysis and lab dupli-

cate. Avoid any headspace. Sample volumes must be 500 ml or larger. Pre-

serve samples by adding 2N zinc acetate at the rate of 1.5 ml/1 plus NaOH to

pH between 9 and 11. Samples must be analyzed within 7 days of collection.

Because of the preservation procedure!, the entire sample must be used for

analysis. Also samples must be prepared and analyzed using the filtration

technigue (Method 427 B, Sec. 3b and 427 j), Sec. 2b). The original volume of

the sample must be accurately known and must be included in calculating the

final results (427 D, Sec. 3)._________________________________________

Prepare arid standardize the following solutions each day:_________________

A. Iodine standard, per Method 427, Sec. D, la.______________________

B. Sodium thiosulfate solution, per Method 421 B, Sec. 2e, and 2f._______

CVR178/046/2

Prepare fresh potassium bi-iodate standard each day per Method 421 B,______

Sec. 2f. Complete all standardization titrations in duplicate. Duplicate

results for solution standardizations jnust agree to within 0.2 ml. Use the

average of the duplicate results for calculations._____________________

9. Analytical results required (if known, specify format for data sheets,QA/QC reports, Chain-of Custody documentation, etc.). If not completed,format of results will be left to program discretion.

Bench records tabulating all standardization^ titrations (in duplicate), sam-

ple and QA audit titrations, sample volumes, calculations, etc., will be____

provided. All records of analysis and calculations must^ be legible^ and suf-

ficient to recalculate all sample concentrations and QA audits.__________

10. Other (use additional sheets or attach supplementary information, asneeded):

11. Name of sampling/shipping contact: Joseph DankoPhone: 503/752-4271

Please return this request to the Sample Management Office as soon as poss-ible to expedite processing of your request for special analytical services.Should you have any questions or need any assistance, please call the SampleManagement Office.

CVR178/046/3

I. DATA REQUIREMENTS

Parameter

Sulfide

Detection Limit

1.0 mg/1

II. QUALITY CONTROL REQUIREMENTS

Audits Required Frequency of Audits

Precision Desired(+/- * or cone.)

Difference in duplicate

results should not exceed

±10 percent for concentra-

tions >3 mg/1 or .3 mg/1

concentrations <3 mg/1,

Report results to the near

est 0.1 mg/1 (above 1

or to two significant fig-

ures above 10 mg/1.

Field Duplicates One per group of 10 or

fewer samples.______

Limits*(+/- % or cone.

±_ (10 percent or

0.3 mg/1)_______

III. *Action Required if Limits are Exceeded:

Repeat all associate sample analyses with second sample bottle.

Contact; Chuck Elly, (312) 353-9087_______________________

________John Tanaka, (312) 886-6337______________________

________Joseph Danko, (503) 752-4271______________________

CVR178/046

CVR178/046/4