automation of epa method 1664 - environmental express 8.5 x 11 13...2013/11/25  · automation of...

4
The SPE-Express system is specifically designed for the extraction and evaporation required by EPA Method 1664 for Oil and Grease Analysis. ENVIRONMENTAL EXPRESS Call 800.343.5319 or 843.881.6560 • www.environmentalexpress.com Automation of EPA Method 1664

Upload: others

Post on 21-Mar-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Automation of EPA Method 1664 - Environmental Express 8.5 x 11 13...2013/11/25  · Automation of EPA Method 1664. Wet Chemistry: Solid Phase Extraction/Oil and Grease. SPE-Express

The SPE-Express system

is specifically designed

for the extraction and

evaporation required

by EPA Method 1664 for

Oil and Grease Analysis.

E N V I R O N M E N T A L E X P R E S SCall 800.343.5319 or 843.881.6560 • www.environmentalexpress.com

Automation of EPA Method 1664

Page 2: Automation of EPA Method 1664 - Environmental Express 8.5 x 11 13...2013/11/25  · Automation of EPA Method 1664. Wet Chemistry: Solid Phase Extraction/Oil and Grease. SPE-Express

Wet Chemistry: Solid Phase Extraction/Oil and Grease

SPE-Express is the only

system that both extracts

your sample and evaporates

the n-Hexane. And the only

one that utilizes a fluid

sensor to verify that the

sample vessel is empty.

It will reduce time and

labor in your lab, improve

accuracy plus minimize

analyst contact with

hazardous chemicals.

SPE-Express Specifications:

Footprint: 22.25"w x 26.105"d x 22.343"h

Clearance: 2" all sides

Weight: 118 lbs

Crated: 200 lbs

Electrical: Available in 120V or 240V

SPE-Express is the ONLY system that:• Both extracts the sample and evaporates the n-Hexane, eliminating the transfer step

• Uses a fluid sensor to verify that the sample vessel is empty improving accuracy

• Runs multiple samples simultaneously on up to 6 stations improving analysts efficiency

Simply load the 3-station SPE-Express with standard one-liter sampling jars and walk away. The system flushes itself clean, preps the extraction disk, then filters and extracts the sample through sodium sulfate with n-Hexane. In the final stage the SPE-Express automati-cally transfers the extract to a pre-weighed aluminum pan where it is evaporated to dryness. Remove the pan for final weighing and the process is complete. User-friendly PC software controls the functionality while recording data for downloading to LIMS.

Product Features and Benefits:

• Each system has three stations

• System is controlled with a PC using a single USB cable

• Two systems can be joined and connected to one PC for a total of 6 stations

• Each station can begin processing a new sample while a previous sample is being evaporated

• System comes with software containing one default method and offers the ability for the operator to create customized methods

• SPE-Express software allows users to have varying levels of access

• Fully enclosed unit does not require a fume hood

• System utilizes wide mouth (53mm) amber jars designated by the EPA method

• 33, 70, and 89mm adapters are available for other bottle types

• Available in 120V or 240V

Page 3: Automation of EPA Method 1664 - Environmental Express 8.5 x 11 13...2013/11/25  · Automation of EPA Method 1664. Wet Chemistry: Solid Phase Extraction/Oil and Grease. SPE-Express

SPE-Express performs the entire process through evaporation to dryness in approximately 45-50 minutes depending on the content of your sample.

• Prep – 5 Minutes

• Filter Sample – 15 minutes

The SPE-Express automates Oil and Grease Extraction and Evaporation as outlined in EPA Method 1664:

Preps the System

• Rinses all system components with n-Hexane.

• Preps the extraction disk with Methanol then replaces the Methanol with DI water to avoid co-solvent elution.

Filters the Sample

• Vacuums the aqueous sample through the filter until the sample bottle is empty and all free liquid is removed from the filter.

• Elutes the sample with n-Hexane and removes any sample residue from the sample bottle.

• Vacuum filters the eluted sample through the extraction disk then transfers the eluted n-Hexane through the drying cartridge into a glass chamber.

Evaporates the Eluted Sample

• Transfers the sample to a heated, pre-weighed aluminum pan for evaporation to dryness.

• Allows the analyst to begin processing a new sample on the same station while the first is being evaporated.

• Elution – 20 minutes

• Evaporation – 7 minutes

SPE-Express 120V G8000

SPE-Express 240V G8000-240

Equipment included:Description Catalog #

Base 3-Station Unit, Laptop with installed Software and Operation and Instruction Manual, Power Cord and USB Cable

Bottle Rack ABP110 20-Liter Carboy with Tubing and Fittings G3070 Tool Kit with Fastener Tool, Adapter Wrench, Drying Cartridge Extractor ToolCapacitive Sensor Calibration Tool, and Reagents Pressure Hose

Consumables included:Description Catalog #

2 packs 5.5 Gram Na2SO

4 Drying Cartridges, 50 per pack G1065

1 pack 105mm Aluminum Weigh Dishes, 100 per pack F93140DSH1 pack UltraFlow Disposable 47mm Filter Assemblies, 100 per pack G52601 pack O-Rings, 15 per pack G82061 pack Snip and Pour Standards for Oil and Grease, 20x10mL G30251 pack Snip and Pour Standards for MDL, 20x10mL G3023

Accessories sold separately:Description Catalog #

Dual Unit Assembly Kit for connecting two units to one PC G8050Pump - Vacuum Pressure 115VAC G3050Sample Jars, 950mL Wide Mouth (53mm) Amber Glass APC1213Smart Bottle Rack G8100Vent Hose, 4" DiameterNitrogen Source (70 psi minimum/110 psi maximum). Other inert gases are acceptable. (Supply hose must be an Industrial-Shape Quick Connect 0.25" socket in order to connect to unit.)

G5260 Filter Assembly

G1065 Drying Cartridge

F93140DSH Aluminum Evaporating Pan

Page 4: Automation of EPA Method 1664 - Environmental Express 8.5 x 11 13...2013/11/25  · Automation of EPA Method 1664. Wet Chemistry: Solid Phase Extraction/Oil and Grease. SPE-Express

Wet Chemistry: Solid Phase Extraction/Oil and Grease

The SPE-Express software is easy and intuitive to understand.

The software:

• Allows the analyst to track the progress

of each sample throughout the

extraction and evaporation process.

• Comes preprogrammed with EPA

Method 1664 but allows the analyst to

create customized methods as well.

• Allows the Administrator to assign

varying permission levels to other

users.

• Records all data through the extraction

and evaporation process for

downloading to LIMS.

Use the Smart Bottle Rack with the SPE-Express for maximum efficiency.

The Smart Bottle Rack uses special

sensors to monitor the levels of reagents

and waste in your containers and alerts you

when they require attention.

The Smart Bottle Rack makes sure you

never run out of reagents or overflow your

waste container.

E N V I R O N M E N T A L E X P R E S SCall 800.343.5319 or 843.881.6560 • www.environmentalexpress.com

Copyright, Environmental Express, Inc., April 2013