basic of orbscan -dr.movahedan
TRANSCRIPT
5/14/2008
1
In the name of Godthe most kind and merciful
Basic of orbscan
H.Movahedan MD
5/14/2008
2
What is Topography?
Topographic Technologies• Placido disk –based topography
• AstraMax : ( three-dimensional topography )
• Elevation-based topography:– Slit-scanning topography (orbscan)–– Scheimpflug imaging (PentacamScheimpflug imaging (Pentacam--GalileiGalilei--Precisio)Precisio)
• Artemis : (ultrasound digital topography)
5/14/2008
3
Elevation Based Topography
• Orbscan:– Placido disc & slit scanning
• Pentacam:– Scheimpflug imaging
• Galilei :– Dual scheimpflug imaging & placido disc
• Precisio: – Scheimpflug imaging
5/14/2008
4
Schematic picture of the Scheimpflug PrinciplePicture / Focus Plane
Objective Plane
Film PlaneObjective Plane
Film Plane
Picture / Focus Plane
Point of Intersection
Schematic picture of a Camera
Problem of a normal Camera:
limited depth of focus
Advantage of the Scheimpflug Camera:
•higher depth of focus,
•sharp picture, but distorted
The Scheimpflug law says: To get a higher depth of focus, move the three planes, provided that the picture plane, the objective plane and the film plane has to cut each other in one line or one point of intersection.
Scheimpflug Principle
5/14/2008
5
October 1997 Orbscan II Concept 9
The ORBSCAN II Concept
• Slit-scan triangulation and perspective reflection are complementary technologies: where one is weak the other is strong.
• ORBSCAN II exploits a synergistic combination of the two technologies.
• The result is unequaled accuracy in the elevation and curvature of clinically-relevant, complex surfaces.
October 1997 Orbscan II Concept 10
Hybrid Technology of ORBSCAN II
1. Measure surface elevation directly by slit-scan triangulation.
2. Measure surface normal directions directly using perspective reflection, supplemented with triangulated elevation.
3. Unify triangulated and reflective data to obtain accurate surfaces in elevation, slope, and curvature.
5/14/2008
6
October 1997 Orbscan II Concept 11
Slit-Scan Triangulation locates surface points individually.
Triangulate an edge point in camera object space (x, y, z)by mathematically intersecting the diffuse reflectedcamera edge raywith the calibrated slit-beam surface.
complex object
VideoCamera
October 1997 Orbscan II Concept 12
Deriving Elevation
Reflection• Requires integration.• Accuracy degrades
peripherally.• Can not traverse
discontinuities.• Placido skew ray
error is significant in clinically relevant abnormalities (e.g., keratoconus).
Triangulation• Primary
measurement.• Uniformly accurate
over the measured surface.
• Measures complex surfaces.
5/14/2008
7
Reflective and Slit-scan Technologies
• One image, one surface.
• Angle-dependent specular reflection.
• Measures slope (as a function of distance).
• Multiple images, multiple surfaces.
• Omni-direction diffuse backscatter.
• Triangulates elevation.
Two prevailing technologies have been used for corneal topography. Both have advantages and disadvantages. ORBSHOT employs Placido reflective technology. ORBSCANemploys slit-scan technology. ORBSCAN II employs a hybrid combination encompassing the best of both worlds.The overwhelming advantage of slit-scan systems is that they measure multiple ocular surfaces.
Integrates multiple disparate technologies: slit-scan, reflective, ultrasound.
How ORBSCAN is Multidimensional
Measures multiple ocular surfaces: anterior cornea, posterior cornea, anterior iris, anterior lens.
Displays multiple and complete mathematical surfaces: curvature, power, elevation, thickness.
Multidimensionality makes ORBSCAN powerful. The wide variety of maps can make ORBSCAN bewildering.The first purpose of this basic map reading course is to introduce, both technically and clinically, the most powerful ORBSCAN maps and measures.
5/14/2008
8
Color Problem• Contours are evenly spaced (the normal band being
exceptional).• Unfortunately, adjacent colors are NOT perceived as
uniformly equidistant.• An interpretation should never be based on color.
– Instead perceive the topography behind the colors, contours, and reference objects.
• The third purpose of this course is to warn you about numerous interpretation pitfalls:– contour and color problems– axial artifacts– elevation distortion and relativity– power and curvature confusion
• high power• short focal length
Red
BlueMin
Max
• low power• long focal length
(+)
(+ +)
A high power lens has a short focal length. Red is high power, and blue is low power.
Optical Power
5/14/2008
9
Elevation can be positive or negative. Red is high and positive (anterior to the reference surface).
• high• anterior to the
reference surface
• low• posterior to the
reference surface
Red
BlueMin
Max
reference level
(+)
(-)
anterior
posterior
Elevation (from a reference surface)
Corneal thickness equals posterior corneal depth. Red is thinner (the posterior cornea is higher).
• thin
Red
BlueMax
Min
• thick
(+)
(+ +)
Thickness
5/14/2008
10
Global Perspective
12,000 mi12,000 microns
The elevation topography of both globes is small in comparison to the entire surface
12,000 miles
Topographical Elevation is Relative
5.6 mi 5.6 miles
12,000 miles
On a local scale these features are significant
5/14/2008
11
Although they appear large up close, terrestrial mountains are very small (1000 times smaller) when compared to the size of the earth.Similarly, local clinical features (microns high) are a thousand times smaller than the cornea (millimeters in radius).This disparity in scale makes it IMPOSSIBLE to simultaneously map both the global curvature and clinically significant local elevation features of the cornea.To see clinically relevant features, corneal global curvature must first be removed. This is done by measuring elevation relative to some close-fitting reference surface, a process that inevitably introduces distortion.
Removing Global Curvature
Local clinical features (microns high) are a thousand times smaller than the cornea (millimeters in radius).This disparity in scale makes it IMPOSSIBLE to simultaneously map both the global curvature and clinically significant local elevation features of the cornea.To see clinically relevant features, corneal global curvature must first be removed. This is done by measuring elevation relative to some close-fitting reference surface, a process that inevitably introduces distortion.
Removing Global Curvature
5/14/2008
12
Close-Fitting Reference Surfaces
Topographic maps of terrestrial landscapes are displayed in the form of constant-elevation contours, measured from the “mean sea-level” of the earth.Corneal topography differs from terrestrial topography in that the reference surface is not some fixed “mean sea-level”, but is movable.
For the cornea, a reference surface (typically, a sphere) is constructed by fitting the reference surface as close as possible to the data surface.A best-fit minimizes the square difference (always a positive number) between the two surfaces, but only within a specified region known as the fit-zone.
Fit-zoneReference surface (sphere)
Data surface(cornea)
Default settings:• shape = sphere
Elevation Relativity
Click for more on reference surface shapesClick for more on reference sphere alignments
Changing the shape, size, or alignment of the reference surface changes the relative elevation, just like changing the mean sea level would change the heights of mountains.The default reference surface is the best-fit sphere. Other reference surface shapes are available for special applications.The sphere is especially useful, because it is only reference surface with no unique symmetry axis (i.e., every diameter is a symmetry axis).Two parameters define the best-fit sphere. Radius specifies size. Center location specifies alignment.The default sphere alignment is floating, which means its center location is unconstrained. Other alignments may force the reference sphere to lie on the map axis.
• alignment = floating
5/14/2008
13
Myopic Pre-Post Elevation Maps
Elevation (sphere) Elevation (sphere)
These are pre-op (left) and post-op (right) elevation maps of a myopic with-the-rule astigmatic eye corrected with Lasik.The post-operative central “sea” is not a concavity but a central flattening.The ring of relatively highest terrain is not absolutely higher (more anterior) than the “sea” bottom near the map center.
Different Fit-Zone Elevation Maps
Elevation (sphere)10 mm fit-zone
Elevation (sphere)5 mm fit-zone
These elevation maps are the same exam of a normal eye having with-the-rule astigmatism. They differ only by the size of the fit-zone used to construct the reference sphere.Both maps contain exactly the same information. They look different, because altering the fit-zone changes the size and alignment of the reference sphere.For a normal cornea, a smaller fit-zone results in a smaller reference sphere that better fits the central cornea.
7.16 mm radius 6.88 mm radius
A small fit-zone bring outs small central anomalies, like central islands.A large fit-zone emphasizes global shape. By default, the largest 10 mm diameter fit-zone is used.
5/14/2008
14
Spherical reference surface
Elevation DistortionData surface profile
As an example of distortion, consider the corneal surface (i.e., data surface) following refractive surgery for myopic correction. It is centrally flattened by the surgery.To see surface features (which are highly exaggerated in this drawing), elevation must be measured with respect to some reference surface.When relative elevation is mapped, the reference surface is effectively flattened. This distorts the relative elevation profile.This relative elevation peak is NOT the highest point on the cornea.This apparent central "concavity" does NOT exist.
Relative elevation profile
Elevation Topology: Central Hill
The normal cornea is prolate, meaning that meridional curvature decreases from center to periphery.Prolateness of the normal cornea causes it to rise centrally above the reference sphere. The result is a central hill.
Sharp center
Flat periphery
Immediately surrounding the central hill is an annular seawhere the cornea dips below the reference surface.In the far periphery, the prolate cornea again rises above the reference surface, producing peripheral highlands.
5/14/2008
15
Elevation Topology: Central Saddle
The regular astigmatic cornea is toric, meaning that meridional curvature has maxima and minima in different directions, 90 degrees apart.The sharp (or steep) profile falls below the reference surface.The flat profile rises above the reference surface.The resulting elevation topology is a central saddle.A saddle is shaped like a riding saddle, rising peripherally in one direction, and falling peripherally in a mutually perpendicular direction.
Prolateness and Toricity Balance
Fit-zone sizeCorneal radius
T, ToricAmplitude
E, ProlateShape Factor
Central Hill Central Saddle
Central elevation topology is determined by the balance of corneal prolateness and toricity.
5/14/2008
16
Prolate Patterns
Central Hill
When prolateness dominates, a central hill will appear.
Toric Patterns
Central Saddle
When toricity dominates, a central saddle will appear.
5/14/2008
17
Central SaddleCentral Hill
Prolateness-Toricity Fulcrum
The position of the prolateness-toricity fulcrum is determined by the size of the fit-zone with respect to the mean corneal radius.
Large fit-zones favor prolate shapes, preferentially producing central hills.
Small fit-zones favor toric shapes, preferentially producing central saddles.
Elevation topology is only invariant when fit-zone size and reference sphere alignment are both fixed.
T, ToricAmplitude
Fit-zone sizeCorneal radius
E, ProlateShape Factor
Basic Map Reading 10:Optical Diagnosis
Point spread functionImage convolutionRaytrace analyses
Case report: Lasik ODCase report: Bilateral Lasik
29 points
5/14/2008
18
Point Spread Function
Raytrace analysis through one or both corneal surfaces can be used to calculate the Point Spread Function (PSF), which is the retinal image of a single point of light. This PSF is aberrated.
The main focus (pink and red) is spread horizontally and irregularly. A lower intensity blur pattern (green and blue) spreads far beyond the main focus. Extra foci appear at the vertices of the blur pattern.
Image Convolution
Original PictureSimulatedRetinal Image
The easiest way to interpret a PSF is to directly see how is affects the simulated retinal image.The original picture is disassembled into points, each point is spread out into the PSF pattern, and the patterns are reassembled to form the simulated image.
5/14/2008
19
Lasik OD: Raytrace OD
This is the aberrated total corneal PSF of a patient that underwent Lasik OD. The procedure did not correct the original myopia and resulted in severe subjective symptoms (blurred vision, glare, etc.).
The simulated retinal image lets us see what the patient sees.
Original PictureSimulatedRetinal Image
Summary
• Corneal Elevation Topography is viewed relative to a reference surface
• Standardization of the reference surface is necessary to allow meaningful comparisons
5/14/2008
20
Thank you for your attention