basics of steam sterilization
TRANSCRIPT
8/3/2019 Basics of Steam Sterilization
http://slidepdf.com/reader/full/basics-of-steam-sterilization 1/6
8/3/2019 Basics of Steam Sterilization
http://slidepdf.com/reader/full/basics-of-steam-sterilization 2/6
55January 2009 MANAGING INFECTION CONTROL
Sterilization Technology
temperature of steam to kill microorganisms. Microbes die
when steam under pressure creates changes in cell structure,
and when cell protein coagulates (thickens into a mass).
The first steps in sterilization involve proper cleaning
(the removal of visible and invisible soil) and decontami-
nation (the removal or reduction of infectious organisms
or other harmful substances). Failure to adequately
prepare an item for sterilization hinders the direct steam
contact needed to destroy microorganisms. Operating
rooms run on a tight schedule, but proper decontamination
of instruments is mandatory. The concern with flash
sterilization is that in order to accommodate the surgeon
and schedule, the operating room staff will be pressured
to eliminate steps, therefore proper decontamination may
not occur between surgical procedures. The Association
for the Advancement of Medical Instrumentation (AAMI)
states that flash sterilization should only be considered if
“work practices ensure proper cleaning, decontamination,
inspection and arrangement of instruments into the
sterilization for patient use, it does render the instruments
safe for handling after the operation.
Flash sterilizers are typically located near operating room
suites. They are used to quickly sterilize dropped instruments
using cycles with minimum exposure times and temperatures
for unwrapped items. Sterilized items must be transferred
immediately from the flash sterilizer to the operating room
using aseptic technique to minimize recontamination from
pathogenic organisms. There is considerable risk that flash-
sterilized items can be contaminated during transport to the
point of use. The decision to use a flash sterilizer requires
that the benefit of having these instruments is greater than the
risks associated with the process of flash sterilizing and
transporting these instruments.
The Basics of the Steam Sterilization ProcessSteam, by itself, is inadequate to achieve sterilization. Pressure
that is greater than the atmosphere is required to increase the
8/3/2019 Basics of Steam Sterilization
http://slidepdf.com/reader/full/basics-of-steam-sterilization 3/6
56 January 2009MANAGING INFECTION CONTROL
Sterilization Technology
recommended sterilizing trays or other containment
devices before sterilization.”1
There are four critical parameters to steam sterilization:
steam, temperature, pressure and time. Steam must be of
high quality and contain no more than 3 percent moisture
and a relative humidity (the amount of water vapor) of
97 percent. The temperature depends upon the type of sterilizer used. Gravity air displacement sterilizers require
a temperature of 250°F (121°C). Dynamic air removal,
washer sterilizers, and flash sterilizers require a temperature
of 270°-275°F degrees (132°C-134°C). To achieve these
temperatures, the pressure must reach 15 pounds per square
inch (psi) for the 250°F (121°C) setting, and 27 psi to
sterilize at 270°F.
Exposure times depend upon the sterilizer manufac-
turer’s recommendations based upon load contents. Flash
sterilization times usually vary from three minutes to 10
minutes. If the load consists of a single instrument
comprised of metal (nonporous) and without lumens,
a three minute cycle should be used. If the load includesmultiple instruments or instruments with lumens, a 10
minute cycle should be used.2
The intense heat of flash sterilization can shorten the
life of power equipment (drills and saws) and batteries.
Frequently, when the flash cycle is complete, batteries
are not immediately removed from the sterilizer; hence
the batteries sit longer in the extreme heat than they
should. The condensation that is formed inside the power
equipment can increase the chance of corrosion and
damage to the internal components, therefore an additional
eight minute “drying time” may be added to help dry the
water droplets that have been created in the device.
Precautions must be taken when items are loaded intothe sterilizer to ensure that steam will contact all items to
be sterilized. All instruments should be opened and disassembled to
allow the steam to completely contact items during sterilization.
The sterilizer should not be overloaded and items should not be
allowed to contact the sterilizer’s chamber walls.
Monitoring Sterilization Cycles
Sterility should never be taken for granted. Sterilization cyclesshould be routinely monitored using mechanical, chemical and
biological indicators.
Mechanical indicators are indicators that come from the
sterilizer itself to indicate how it has functioned. This may include
thermometers, timers, gauges, recorders and/or other devices that
monitor its function. At the end of every cycle, the sterilizer
printout tape should be checked for:
Start time and date;
Operator’s initials;
Sterilizer parameters selected (Was the correct cycle and
time selected?);
Sterilizer parameters met (time, pressure, minimum temperature).
Chemical indicators are indicators that verify exposure to
a sterilization process by changing color. Chemical indicators or
integrators can be used in flash sterilizers. They will change color
when the exposure or parameters have been met. The chemical
indicators/integrators should be placed in each tray among the
instruments with every cycle.
Biological indicators (BI) monitor the adequacy of the
sterilization process and the destruction of microorganisms. They
consist of an inoculated carrier (Geobacillus stearothermophilus
for steam sterilization) that provides a defined resistance to the
sterilization process in which they will be used. Sterilizer operators
should verify that the date on the BI has not expired, and should
be placed in the middle of the sterilizer above the drain. AAMIrecommends that a biological test be performed at least weekly,
preferably every day that the sterilizer is used. A positive result
for growth would indicate that the sterilizer did not function
appropriately. The sterilizer must be taken out of service immedi-
ately and the instruments must be re-sterilized. In fact, all items that
were processed in that sterilizer since the last negative BI are
considered non-sterile. This is why documentation of all flash
sterilization cycles must be traceable to a patient.
ConcernsAAMI and the Centers for Disease Control and Prevention
(CDC) warn against flash sterilizing of implantables. AAMI states
that “implantables should not be flash-sterilized. The possibleconsequences to the patient from placing even a minimally
contaminated device in an essentially avascular environment and
leaving it there at the conclusion of the procedure are potentially
severe.”3 In addition, implants should be quarantined until the
results of the BI are known.Printout tape from the sterilizer verifies that parameters
have been met.
8/3/2019 Basics of Steam Sterilization
http://slidepdf.com/reader/full/basics-of-steam-sterilization 4/6
57January 2009 MANAGING INFECTION CONTROL
Sterilization Technology
When the sterilization cycle is completed, the contents
should be removed immediately for use. Items are considered
sterile when necessary conditions have been met during a
sterilization process. The circulating nurse should assure that
all parameters have been met before releasing the load. The scrub
nurse should visually inspect the chemical indicator for color change.
Biological indicators need to be processed and read in a timely manner
according to manufacturer’s instructions.
Attention should be made to avoid heat injuries when unloading
the sterilizer. The instrument(s) must be removed and transported
using aseptic technique. Transferring the sterilized item to the sterile
field is difficult without contaminating it. The scrub nurse needs to pay
attention to how they reach into the sterilizer to remove the contents
and not touch the edges of the sterilizer. The circulator should escort
the scrub nurse to the flash sterilizer eliminating traffic or opening
of doors from the inner core during transportation of the sterile
instruments. The risk of contamination increase if “transported
through areas where personnel are scrubbing or washing their hands,
creating splashing or aerosolization.”4 The work area should be
arranged so that it ensures direct delivery of sterilized items to point of
use without physical hazards and possible contamination.
Another concern of flash sterilization is being able to assure
that all items are allowed to cool before using. To meet theScrub nurses need to verify color change on chemical
indicators.
8/3/2019 Basics of Steam Sterilization
http://slidepdf.com/reader/full/basics-of-steam-sterilization 5/6
58 January 2009MANAGING INFECTION CONTROL
Sterilization Technology
not used on the procedure, the documentation must be
completed for that particular cycle. Accountability for each
load must be maintained. Sterilization records are considered
legal documents and, since the length of time they must be
kept varies, each healthcare facility must develop record-
retention policies based upon state and local regulations and
legal considerations.
Sterilizer MaintenanceSterilizer maintenance is important. Sediment screens
should be removed and cleaned daily because clogged strainers
prevent air from being removed from the unit. Chamber
interiors should be cleaned weekly with a mild detergent.
Note: Sterilizers must be allowed to cool for eight hours
before cleaning. Maintenance requirements differ depending
on the model used, and the sterilizer manufacturer’s instruc-
tions should always be followed.
Sediment screens need to be checked daily for
proper maintenance.
Ongoing ConsiderationsTo flash or not to flash, that is the question. According to the
Association for periOperative Registered Nurses (AORN), it
states that flash sterilization should be minimized. It should only
“be used in selected clinical situations and in a controlled
manner.”5 It should not be used as a substitute for not having
enough instrumentation. Scheduling of cases in a manner that
does not require back-to-back usage of the same instrument trays
needs to be addressed. Managers who have budget constraintsoften end up with insufficient instrument inventory. However,
purchasing of additional instruments to maintain the daily
caseload should be a priority. Enforcing policies to have on
time arrivals of vendor pans, brought in the day before surgery,
is critical for providing the best care for the patient.
demands of the surgical procedure, scrub nurses frequently
try to cool the instruments by immersing them in cool
sterile water or wrap in sterile wet towels. Depending on
the item, this cooling process can cause harm to the instru-
ment. For example, if attempting to cool autoclavable
scopes too quickly, it can crack the lenses inside the scope.
Cooling power equipment (drills/saws) quickly can causemore condensation to form on the inside. Most importantly,
the hot instruments can cause injury to the patient and/or
physician. Depending on the thickness of the metal, the
scrub nurse might think it is cool enough to use, but instead
it heats up in the surgeon’s hands or while on the patient.
Some retractors are thick and the heat continues to leech
out during the case. Burns may not be noticed until the
retractor is removed from the incision site. We must
remember that while the patient is under anesthesia and
unable to alert the staff to undesirable conditions, we must
act as their advocate and prevent these accidents from
happening. Let’s not forget the danger to the employees
themselves. When carrying hot and heavy instruments fromthe flash sterilizer, poor body mechanics are frequently
used in holding the hot items out and away from their body
to prevent burning. These poor body mechanics lead to the
straining of the lower back and arms.
DocumentationDocumentation of the flash sterilization process should
be consistent with requirements for sterilizing wrapped
items. However, flash sterilized items require documentation
linking the patient. The following information should be
documented for each flash sterilization cycle:
Date;
Sterilizer number and sterilizer load number; Sterilizer operator;
Chemical indicator;
Contents of load (including BI if used);
Patient information;
Reason for flash sterilization;
Initialed printout tape of cycle parameters.
Effective documentation ensures that the sterilization
process is monitored as it occurs, that required cycle param-
eters have been met, and that accountability is established.
At the end of each day, each sterilizer should have complete
documentation for each load. A common problem with
flash sterilization is incomplete documentation. Operatingroom employees are overwhelmed with the other duties and
tasks pertaining to their job and sterilization records take a
low priority. Chemical control cards or integrators are lost;
contents of the load are not documented and/or patient
information not included. Even if the instruments were
8/3/2019 Basics of Steam Sterilization
http://slidepdf.com/reader/full/basics-of-steam-sterilization 6/6
59January 2009 MANAGING INFECTION CONTROL
Sterilization Technology
Carol Petro, RN, BSN, CRCST,is the clinical educator for
Clarian Health sterile processing
departments at Methodist Hospital,
University Hospital and Riley
Hospital for Children in
Indianapolis. She has worked in
the sterile processing department
for two years and the surgery
department for 25 years. She has a
Bachelor of Science Degree in Nursing from University of
Indianapolis. She received her technical training as a
Certified Registered Central Service Technician (CRCST)
from IAHCSMM in 2006.Previous positions held at Clarian included Oral/
Maxillofacial/Ophthalmology Specialty Coordinator from
1992 to 2001 and Special Projects Coordinator/SurgiNet
Informatics Project Leader from 2001 to 2005. She may
be reached at [email protected].
Alternatives need to be created for the doctor’s specials that arecarried from the operating room of one hospital to another. Can we eval-
uate what is contained in these special instrument sets and provide them
in one of our existing sets, so that they are sterile prior to the procedure?
Operating room personnel must maintain high standards of
sterile technique. Every individual is accountable for their role in
infection control. Work practices need to ensure that proper
cleaning and decontamination is performed and that procedures
are developed and followed in regard to flash sterilization. ✛
References1. ANSI/AAMI ST79:2006. Comprehensive Guide to Steam Sterilization and
Sterility Assurance in Health Care Facilities. Arlington, VA.2006, page 2-3.2. International Association of Healthcare Central Service and Materiel
Management. Central Service Technical Manual. Seventh Edition. 2007.Point of Use Processing, chapter 14, page 269.
3. ANSI/AAMI ST79:2006. Comprehensive Guide to Steam Sterilization andSterility Assurance in Health Care Facilities. Arlington, VA.2006, page 3.
4. ANSI/AAMI ST79:2006. Comprehensive Guide to Steam Sterilization andSterility Assurance in Health Care Facilities. Arlington, VA.2006, page 2.
5. Association of PeriOperative Registered Nurses (AORN). Standards,Recommended Practices, and Guidelines. Denver, CO. 2007, page 675.
Reader Service No. 40