Biosafety and Monitoring the CSSD is Priority and responsibility of everyone
in Health care • The central sterile supply
department (CSSD) plays a key role in providing the items required to deliver quality patient care. To support infection control within the healthcare facility, the CSSD staff members must be well-trained and skilled, and committed to “doing what’s right” The shortcuts are never made and that processes and practices are consistently followed.
• Dr.T.V.Rao MD
Dr.T.V.Rao MD 2
![Page 3: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/3.jpg)
What is Central sterile supply department
• Central sterile supply department( CSSD) is a service unit in a hospital that processes, issues, and controls the sterile stores supply to all departments of the hospital. It can be defined as that service, with in the hospital, catering for the sterile supplies to all departments , both to specialized units as well as general wards and OPDs
Dr.T.V.Rao MD 3
![Page 4: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/4.jpg)
We Need constant Monitoring of the Biosafety
of the Instruments and materials used in Hospitals
• The most common method for sterilization of medical instruments worldwide is steam sterilization. However, steam sterilization or autoclaving may not be 100% effective at killing all microorganisms under all circumstances. Therefore, the quality of sterilized products should be assessed by physical, chemical, and biological indicators.
Dr.T.V.Rao MD 4
![Page 5: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/5.jpg)
What CDC Recommends
• According to CDC and other medical associations, to ensure heat penetration to all instruments during each cycle, a chemical indicator should be placed inside and in the center of a load of unwrapped instruments, routinely in every sterilization process. Finally, biological indicators are used periodically for monitoring of the quality of sterilization processes
Dr.T.V.Rao MD 5
![Page 6: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/6.jpg)
Monitoring CSSD• The sterilization
procedure should be monitored routinely by using a combination of mechanical, chemical, and biological indicators to evaluate the sterilizing conditions and indirectly the microbiologic status of the processed items.
Dr.T.V.Rao MD 6
![Page 7: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/7.jpg)
Care of Mechanical Monitors
• The mechanical monitors for steam sterilization include the daily assessment of cycle time and temperature by examining the temperature record chart (or computer printout) and an assessment of pressure via the pressure gauge.
Dr.T.V.Rao MD 7
![Page 8: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/8.jpg)
Monitoring Ethylene Oxide
Monitors • The mechanical monitors for
EO include time, temperature, and pressure recorders that provide data via computer printouts, gauges, and/or displays. Generally, two essential elements for EO sterilization (i.e., the gas concentration and humidity) cannot be monitored in healthcare EO sterilizers.
Dr.T.V.Rao MD 8
![Page 9: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/9.jpg)
Why Chemical Indicators are Preferred
• Chemical indicators are convenient, inexpensive which indicate that the item has been exposed to the sterilization process. But, Chemical indicators were more likely than biological indicators to inaccurately indicate sterilization even during very short sterilization times (e.g., 2 minutes).
Dr.T.V.Rao MD 9
![Page 10: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/10.jpg)
Criteria of Classification • These classifications
are based on Spaulding’s classification of criticality of materials with regard to infection control and cost considerations
Dr.T.V.Rao MD 10
![Page 11: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/11.jpg)
Earle Spaulding Defines the Criteria of classification
• A young doctor from Philadelphia, Earle Spaulding, took on the task of testing germicides, antiseptics, soaps, and cleansers against specific germs. In a paper he published in 1939, he proposed a classification system that is still used today: The Spaulding Classification Scheme. He proposed that classifying medical devices based on the degree of risk for infection would lead to a better understanding of the degree of cleanliness required, and as a result, improve compliance and reduce infection. His simple and logical
categories are "Critical," "Semi critical," and "Noncritical."
Dr.T.V.Rao MD 11
![Page 12: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/12.jpg)
Dr.T.V.Rao MD 12
![Page 13: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/13.jpg)
Still We Need Biological Indicators
• Chemical indicators should therefore only be used in conjunction with biological indicators. It should not replace them, because they give incorrect indication of sterilization even at marginal sterilization time.
Dr.T.V.Rao MD 13
![Page 14: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/14.jpg)
Biological monitoring system
• The biological monitoring system depends on live nonpathogenic bacterial spores and stringent sterilization requirements.• The biological indicators are
prepared using a live bacterial spore strip containing a minimum of 1 million live spores.
Dr.T.V.Rao MD 14
![Page 15: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/15.jpg)
Criteria for Selection of Biological Indicators
• All of the biological indicators are approved by theAmericanType Culture Collection and are also tested using the biological indicator evaluator resistometer. The D-value of biological indicator is defined as “the time required in minutes at a certain temperature (121°C) to reduce the number of viable microorganisms by a factor of 10,”5 and the Z-value of a biological indicator is defined as “the number of degrees in Centigrade to reduce the D-value with a factor of 10,
Dr.T.V.Rao MD 15
![Page 16: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/16.jpg)
What is Sterilization
Assurance Level The sterilization assurance level is the “probability of a single unit being non-sterile after it has been subjected to sterilization.” Finally, the entire set of biological indicators after sterilization is sent to the
microbiology department for sterility assurance testing.
Dr.T.V.Rao MD 16
![Page 17: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/17.jpg)
Why Still we Need Biological Indicators
•Only a biological indicator consisting of resistant spores can measure the microbial killing power of the sterilization process. Chemical indicators are affixed on the outside of each pack to show that the package has been processed through a sterilization cycle, but these indicators do not prove sterilization has been achieved. Preferably, a chemical indicator also should be placed on the inside of each pack to verify sterility penetration.
Dr.T.V.Rao MD 17
![Page 18: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/18.jpg)
What are Chemical Indicators
• Chemical indicators usually are either heat-or chemical-sensitive inks that change color when one or more sterilization parameters (e.g., steam-time, temperature, and/or saturated steam; EO - time, temperature, relative humidity and/or EO concentration) are present.
Dr.T.V.Rao MD 18
![Page 19: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/19.jpg)
Chemical Indicator Classes Defined:
Dr.T.V.Rao MD 19
![Page 20: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/20.jpg)
Classification of Chemical Indicators
• Chemical indicators have been grouped into five classes based on their ability to monitor one or multiple sterilization parameters
Dr.T.V.Rao MD 20
![Page 21: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/21.jpg)
Class 1 : Process Indicators
• “Process indicators are intended for use with individual units, (e.g., packs, containers) to indicate that the unit has been directly exposed to the sterilization process and to distinguish between processed and unprocessed units. They shall be designed to react to one or more of the critical process variables.”
• .
Dr.T.V.Rao MD 21
![Page 22: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/22.jpg)
Process Indicators
Applications• Applications : Indicator
tapes, indicator labels, and load cards are examples of externally visible Chemical Indicators that are Process Indicators used for exposure control.
Dr.T.V.Rao MD 22
![Page 23: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/23.jpg)
Class 2 : Indicators for use in Specific Tests
• “Class 2 indicators are intended for use in specific test procedures as defined in relevant sterilizer/sterilization standards.”
• Applications : Bowie-Dick type tests are specific tests used for equipment controlto evaluate the sterilizer performance.
Dr.T.V.Rao MD 23
![Page 24: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/24.jpg)
Class 3 : Single Variable Indicators
• “A single variable indicator shall be designed to react to one of the critical variables and is intended to indicate exposure to a sterilization process at a stated value (SV) of the chosen variable.”
Dr.T.V.Rao MD24
![Page 25: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/25.jpg)
Applications• An example of a Single Variable
Indicator is a temperature tube that contains a chemical pellet that melts at a specific temperature. Single variable indicators may be used for pack control monitoring but would not provide as much information as a Class 4 or Class 5 Chemical Indicator.
Dr.T.V.Rao MD 25
![Page 26: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/26.jpg)
Class 4 : Multi-variable Indicators
• “A multi-variable indicator shall be designed to react to two or more of the critical variables and is intended to indicate exposure to a sterilization cycle at SVs of the chosen variable.”
• Applications : Multi-variable Chemical Indicators are used for pack control. These internal Chemical Indicators are usually paper strips printed with a Chemical Indicator.
Dr.T.V.Rao MD 26
![Page 27: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/27.jpg)
Class 5 : Integrating Indicators :
• “Integrating indicators shall be designed to react to all critical variables. The SVs are generated to be equivalent to, or exceed the performance requirements given in the ISO 11138 series for BIs.”
• Applications : Integrating Indicators are the most accurate of the internal Chemical Indicators. Integrating Indicators are used for pack control monitoring. They can also be used as an additional monitoring tool to release loads that do not contain implants.
Dr.T.V.Rao MD 27
![Page 28: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/28.jpg)
Class 5 : Integrating Indicators
• For this additional monitoring the Class 5 Integrating Indicator must be used in the appropriate challenge test pack or Process Challenge Device (PCD). These indicators must now have SVs at 121°C/250°F, 135°C/276°F, and at least one more temperature in between. Also, the SV at 121°C MUST be greater than 16.5 minutes to ensure performance is comparable to BIs in saturated steam.
Dr.T.V.Rao MD 28
![Page 29: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/29.jpg)
Class 6 : Emulating Indicators • Emulating indicators
are cycle verification indicators which shall be designed to react to all critical variables for specified sterilization cycles. The Svs are generated from the critical variables of the specified sterilization process.”
Dr.T.V.Rao MD 29
![Page 30: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/30.jpg)
Biological Indicators
• Biological indicators are recognized by most authorities as being closest to the ideal monitors of the sterilization process because they measure the sterilization process directly by using the most resistant microorganisms (i.e., Bacillus spores), and not by merely testing the physical and chemical conditions necessary for sterilization
Dr.T.V.Rao MD 30
![Page 31: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/31.jpg)
Why Bacillus spores preferred
• Since the Bacillus spores used in biological indicators are more resistant and present in greater numbers than are the common microbial contaminants found on patient-care equipment, the demonstration that the biological indicator has been inactivated strongly implies that other potential pathogens in the load have been killed.
Dr.T.V.Rao MD 31
![Page 32: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/32.jpg)
Biological Indicators directly Monitor the sterilization Process
• Biological indicators are the only process indicators that directly monitor the lethality of a given sterilization process. Spores used to monitor a sterilization process have demonstrated resistance to the sterilizing agent and are more resistant than the bio-burden found on medical devices.
Dr.T.V.Rao MD 32
![Page 33: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/33.jpg)
B. atrophaeus spores used for monitoring
• B. atrophaeus spores are used to monitor EO and dry heat, and G. stearothermophilus spores ) are used to monitor steam sterilization, hydrogen peroxide gas plasma, and liquid per-acetic acid sterilizers. G. stearothermophilus is incubated at 55-60°C, and B. atrophaeus is incubated at 35-37°C. Steam and low temperature sterilizers (e.g., hydrogen peroxide gas plasma, per-acetic acid) should be monitored at least weekly with the appropriate commercial preparation of spores
Dr.T.V.Rao MD 33
![Page 34: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/34.jpg)
Daily use of Biological Indicators
Recommended • If a sterilizer is used
frequently (e.g., several loads per day), daily use of biological indicators allows earlier discovery of equipment malfunctions or procedural errors and thus minimizes the extent of patient surveillance and product recall needed in the event of a positive biological indicator
Dr.T.V.Rao MD 34
![Page 35: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/35.jpg)
Monitoring essential in each load …..
• Each load should be monitored if it contains implantable objects. If feasible, implantable items should not be used until the results of spore tests are known to be negative
Dr.T.V.Rao MD35
![Page 36: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/36.jpg)
What is New generation Biological Indicator
• The next generation of biological indicator was self-contained in plastic vials containing a spore-coated paper strip and a growth media in a crushable glass ampoule. This indicator had a maximum incubation of 48 hours but significant failures could be detected in ≤ 24 hours.
Dr.T.V.Rao MD 36
![Page 37: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/37.jpg)
What is rapid-readout biological
indicator • A rapid-readout biological
indicator that detects the presence of enzymes of G. stearothermophilus by reading a fluorescent product produced by the enzymatic breakdown of a non-fluorescent substrate has been marketed for the more than 10 years
Dr.T.V.Rao MD37
![Page 38: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/38.jpg)
Rapid Readout • Studies demonstrate
that the sensitivity of rapid-readout tests for steam sterilization parallels that of the conventional sterilization-specific biological indicators and the fluorescent rapid readout results reliable.
Dr.T.V.Rao MD 38
![Page 39: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/39.jpg)
Super Readout Biological Indicator
• A new rapid-readout EO biological indicator has been designed for rapid and reliable monitoring of EO sterilization processes. The indicator has been cleared by the FDA for use in the United States.
Dr.T.V.Rao MD 39
![Page 40: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/40.jpg)
Many Hospitals Going for Rapid Technology
Dr.T.V.Rao MD 40
![Page 41: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/41.jpg)
How the Read out technology works • The rapid-readout EO
biological indicator detects the presence of B. atrophaeus by detecting a fluorescent signal indicating the activity of an enzyme present within the B. atrophaeus organism, beta-glucosidase. The fluorescence indicates the presence of an active spore-associated enzyme and a sterilization process failure.
Dr.T.V.Rao MD 41
![Page 42: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/42.jpg)
Rapid-readout EO biological indicator •The rapid-readout EO biological indicator can be used
to monitor 100% EO, and EO-HCFC mixture sterilization cycles. It has not been tested in EO-CO2 mixture sterilization cycles. The standard biological indicator used for monitoring full-cycle steam sterilizers does not provide reliable monitoring flash sterilizers. Biological indicators specifically designed for monitoring flash sterilization are now available, and studies comparing them have been published.
Dr.T.V.Rao MD 42
![Page 43: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/43.jpg)
VISIT ME FOR MORE ARTICLES OF INTEREST
IN INFECTIOUS DISEASES
Dr.T.V.Rao MD 43
![Page 44: Bio Safety & MonitoringCentral Sterile Supply DepartmentCSSD](https://reader035.vdocuments.net/reader035/viewer/2022082208/58f9b118760da3da068bbd61/html5/thumbnails/44.jpg)
References • 1 CSSD CONNECT overview CSSD CONNECT is an Initiative by PCI with
the purpose of spreading awareness about the right practices to be followed in sterile departments in hospitals: