bioe 485 (imaging) notes

Upload: firstdawn02

Post on 14-Apr-2018

215 views

Category:

Documents


0 download

TRANSCRIPT

  • 7/29/2019 BIOE 485 (Imaging) Notes

    1/5

    Imaging Notes

    Lecture 1Mathematics and Physics of RadiologyIntro lectureexposure to different kinds of imaging techniques. Transmission imaging, ionizing

    radiation inherent in X-ray and analogs.

    Lecture 2Interaction of radiation with matter

    Radiationinteraction with matterElectromagnetic wave

    Particle

    ExcitationIonization

    Bremstrahlung

    Types of x-ray interaction (bold most likely)

    Rayleigh scatterVery low energies (can occur in mammography)

    Compton scatterinelastic scattering (low contrast, most likely due to low z) (low energy ->

    higher probability)

    Photoelectric effectprobability protional to Z3/E

    3.Also, greater probability if incident energy is

    just higher than binding energy.

    Pair productionincident energy of up to 1.02 MeV

    Human body made up of low Z elements

    Lecture 3Interaction: attenuation and dose; Image quality: Resolution, Noise, Contrast,

    SamplingEnergy ranges1.02 MeV (pair production), 100-110 KeV normal x-ray, fluoro, CT, 20 KeV

    Mammography

    I = I*exp(-x)

    Linear attenuation coefficient sum of all individual linear coefficients for each type of interaction(Rayleigh, photoelectric, Compton, etc.)

    Also, linear coefficient of water > linear coefficient of ice > linear coefficient of water vapor

    Mass attenuation coefficient

    X-ray

    Energy ~ 100 kEV

    Current: ma

    Low contrast, no depth

    Quick, low dose - stationary

    Fluoroscopy (movie x-ray)

    Energy ~ 100-200 kEV

    Current: ma

    Low contrast, no depth

    Quick, low dose - stationary

    Mammography

    Energy ~ 20 kEV

    Current: ma

    Low contrast, no depth

    Breast must be compressed to reduce

    scatter

    CT

    Energy ~ 100 kEV

    Current: variable

    high contrast, depth

    360 view, image reconstruction

  • 7/29/2019 BIOE 485 (Imaging) Notes

    2/5

    For a given thickness, the probability of interaction is dependent on number of atoms per

    volume. It is the linear attenuation coefficient divided by density, thus the mass coefficient for all

    phases is the same.Half Value Layer defined as the thickness of material required to reduce the intensity of an x or

    gamma ray beam to one half of its initial value.

    HVL = 0.693/linear attenuation coefficient (from equation above)Beam hardeninglower energy photons of a polyenergetic beam cannot pass through matter,thus these values are filtered to reduce patient exposure.

    Kerma = kinetic energy released in matter, defined as KE transferred to charged particles by

    indirectly ionizing radiation. Units are J/Kg or gray. 100 rads in a gray.Doseenergy deposited by ionizing radiation per mass

    Exposureamount of electrical charge produced by ionizing electromagnetic radiation per mass.

    C/kg

    Equivalent dosein Sieverts, radiation weighting factor applied to absorbed dose to account forrelative effectiveness of various rays

    Effective dosein Sieverts, tissue weighting factor to determine relative dose of various tissue.

    Lecture 4Image Quality; X-Ray Production/Properties

    Contrastdifference in the image gray scale between closely adjacent regions on an image

    Digital image contrast (Contrast to Noise ratio (CNR)assessment of digital image quality.

    Looks at Density at Region A minus Density at Region B divided by noise term.Spatial domaintwo spatial dimensions of an image (x and y axis)

    Point spread functionimage produced from a single point stimulus to a detector. Can be

    isotropic (symmetrically spread out) or non-isotropic. If the PSF is measured at many differentlocations and is the same regardless of location, the system is said to be stationary.

    Line spread function(LSF)linear set of PSF

    Modulation transfer functionplot of the imaging systems modulation versus spatial frequency.

    Good representation of the resolution properties of an imaging system.Illustrates the fraction of an objects contrast that is recorded by the imaging system as a function

    of size (spatial frequency). Note that spatial frequency increases for smaller objects and

    decreases for larger objects.MTF can be found through the Fourier transform of the LSF.

    Noisemany sources. Use statistics as an example, essentially fluctuations about the mean

    (standard deviation = square root of mean). Example is the Gaussian distribution described bymean and standard deviation while Poisson distribution is only described by mean. We usually

    talk about Poisson and use it as a standard for determining covariance = std. dev/mean = 1/square

    root(mean) where variance is equal to mean.

    Contrast Detail curvedescribes the Contrast and Detail for images and allows for comparison

    Lecture 5X-Ray Production; Radiography

    Xray tubeelectron source and target. Generatorvoltage source to excite electrons. Collimator

    defines x-ray field. Tube housingshielding and coolant for xray tube.Brehmstralung spectrumx-ray photon with equal energy to the kinetic energy lost by the

    electron

    Highly inefficient, but increasing energy of incident electrons increases efficiency.

  • 7/29/2019 BIOE 485 (Imaging) Notes

    3/5

    Intensity varies with frequency, essentially higher current increases area under the curve while

    higher voltage increases the cut-off frequency

    Lecture 6Radiography; Mammography

    Lecture 7Mammography, Digital Detectors

    Lecture 8Mammography, Fluoroscopy, CT

    Lecture 9CT and Exam Review

    1Figure 13.15

    2One dimensional CT detector array, Figure 13.14

    3Figure 13.11

    4 - Figure 13.105Single array CT, multiple detector CT, cone beam CT. Differences in slice width

    6Factors affecting resolution p 368-369

    CT dependent on Compton scatter, can use this principle to work in reverse i.e. determinedensity of sample.

    Development is inverse of radiography (in radiography x-rays that pass through will darken film

    while those that dont will have lighter areas (bone is white on an x-ray).

    CT # Discussion see notes

    ProblemBeam hardening can occur, average energy through tissue increases due to

    susceptibility of lower energy beams to attenuation. This worsens with prosthetics which can

    completely block x-ray

    Lecture 10 (Chapter 14)MRIsoft tissue contrast incredibly high, superconducting air-core system.

    Magnetic susceptibilityParamagnetic agent: augments local magnetic field

    Diamagnetic agent: depletes local magnetic field

    Change in magnetic field

    Nucleus has magnetic properties due to protons and neutrons

    Hydrogen atoms most important/abundant and is the key to MRI

    Magnetic Field and sample magnetization

    Thermal energy agitates and randomizes spins in the sample

    Under external field B0, protons organize in low (parallel) and high(anti-parallel) quantization

    energy levels.Element in question dictates spin and gyromagnetic ratio

    Precession frequency

    1T = 42.58 MHz1.5T = 63.86

  • 7/29/2019 BIOE 485 (Imaging) Notes

    4/5

    2T = 127.74

    All protons spinning at Larmor frequency

    Frame of ReferenceApplied magnetic field is parallel to z axis

    Think of how a top spins and falls, but has enough energy to restore its original position.

    Free Induction Decay

    Rotating frame

    T2 Versus T1 (T2 always smaller than T1)

    Pulse sequences

    Excitation, relaxation, echo formation, data acquisition

    Image contrast generated from characteristics of the tissue based upon differences of

    T1time needed for Mz decayT2

    Spin Echo90 degree excitation, 180 degree refocusing pulses

    Inversion recovery180 degree inversion pulse

    Pulse sequences are a combination of excitation, relaxation, echo formation, and data acquisition

    TRtime period between initial RF excitationsTEdelay between excitation and echo formation

    TI: inversion time (inversion recovery) to manipulate spin lattice recovery curves and null signal

    from specific tissues

    Flip angle: amount of excitation by RF pulse

    Spin Echo Pulse Sequence

    ExcitationTEEchoFID signal gradually decays with rate constant T2 -> Spin Echo peak amplitude depends on T2

    T1 > T2 > T2*

    TR Time of Repetition Time between similar angle scans

    Multiple Spin Echo (T2 versus T2*)

    Lecture 11Important aspect of MR is the soft tissue contrast, demonstrated how those parameters could be

    changed to increase contrast

    TRRadio frequency time period

    TEecho timeFIDfree-induction decay See p 391-393.

    T1 = time needed for longitudinal (Mz) regeneration of net magnetization (spin-lattice

    interaction)T2 = Time needed for transverse (Mxy) (spin-spin interaction).

  • 7/29/2019 BIOE 485 (Imaging) Notes

    5/5

    T1 and T2 are tissue dependent because it depends on how these interactions manifest.

    T1 Weightedmust have short TR and TE (intrinsically linked to TR)

    T2 Weightedmust have long TR and long (but not too long) TE (intrinsically linked to TE)Proton DensityLong TR, short TE. Enhances intensity of tissue image, can be acquired while

    gaining a T2 image.

    Spin Echo Pulse90, 180, and then 180 degree pulses, can see on p 399. 1

    st

    echo is the firstwaveform after initiation. TR between the two 90 degree pulses (RF in the second wave)Inversion Recovery (IR)Emphasizes T 1 relaxation times of the tissues by extending the

    amplitude of the longitudinal recovery by a factor of 2. After a delay (time of inversion) TI a 90

    degree RF pulse rotates the recovered fraction of Mz spins into the transverse plane to gain theFID. p. 400. Inversion Time related to T1 and Transverse Decay is T2.

    18090180 degree pulses (TR between 180 degree pulses, second 180 pulse in RF)

    Short tau inversion recovery (STIR)very short TI and magnitude signal processing to eliminate

    tissue such as fat.Fluid attenuated inversion recovery (FLAIR)use of longer TI to reduce signal level of tissue

    like CSF

    Gradient Recalled ECHO (GRE)magnetic field gradient inducing formation of an echo ratherthan the 180 degree pulse. It relies on purposeful phasing and dephasing of the FID. Spin-spin

    interaction and interaction with external inhomogeneities of the magnetic field cause

    degradation. This causes faster degradation (T2*)

    Image acquisition Look over this

    Apply RF frequency to excite certain frequencies in FOV. Apply a gradient of RF values to

    select each slice of the patient.Slice select gradient

    Frequency encode gradient

    Go up to page 437 in Chapter 15 and then look at motion artifacts.

    Process is time intensive, serial MRI is a good way to help reduce this time.